WIKA F6137 User manual

Operating instructions
Betriebsanleitung
EN
DE
Hydraulic ring force transducers - Geotechnical version
Models F6137, F6148, F6160, F6171
Hydraulische Ringkraftaufnehmer - Geotechnik-Ausführung
Typen F6137, F6148, F6160, F6171
Model F6137

2
ADPR1X914115.01 10/2020 EN/DE
Operating instructions, models F6137, F6148, F6160, F6171
EN
DE
© 10/2020 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®and tecsis®are registered trademarks in various countries.
WIKA®and tecsis®sind geschützte Marken in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Operating instructions models F6137, F6148, F6160, F6171
Page
3 - 32
Betriebsanleitung Typen F6137, F6148, F6160, F6171 Seite
33 - 63

3Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
Contents
Contents
1. General information 4
2. Design and function 5
2.1 Overview. . . . . . . . . . . . . . . . . . . . . . . .5
2.2 Description . . . . . . . . . . . . . . . . . . . . . . .6
2.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . .6
3. Safety 6
3.1 Explanation of symbols . . . . . . . . . . . . . . . . . . .6
3.2 Intended use . . . . . . . . . . . . . . . . . . . . . .7
3.3 Improper use . . . . . . . . . . . . . . . . . . . . . .8
3.4 Responsibility of the operator . . . . . . . . . . . . . . . . .8
3.5 Personnel qualification . . . . . . . . . . . . . . . . . . .9
3.6 Personal protective equipment . . . . . . . . . . . . . . . .9
3.7 Labelling, safety marks . . . . . . . . . . . . . . . . . . 10
4. Transport, packaging and storage 11
4.1 Transport. . . . . . . . . . . . . . . . . . . . . . . 11
4.2 Packaging and storage . . . . . . . . . . . . . . . . . . 11
5. Commissioning, operation 12
5.1 Mounting preparation . . . . . . . . . . . . . . . . . . 12
5.2 Mounting instructions . . . . . . . . . . . . . . . . . . 12
5.3 Mounting of the hydraulic ring force transducer . . . . . . . . . . 13
5.4 Commissioning of the hydraulic ring force transducer with digital pressure
gauge . . . . . . . . . . . . . . . . . . . . . . . . 15
5.5 Electrical connection - Hydraulic ring force transducer with pressure sensor 17
6. Faults 18
6.1 Faults of the hydraulic ring force transducer with pressure gauge/digital
pressure gauge . . . . . . . . . . . . . . . . . . . . 18
6.2 Faults of the hydraulic ring force transducer with pressure sensor . . . 19
7. Maintenance and cleaning 20
7.1 Maintenance . . . . . . . . . . . . . . . . . . . . . 20
7.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . 20
7.3 Recalibration . . . . . . . . . . . . . . . . . . . . . 20
8. Dismounting, return and disposal 21
8.1 Dismounting . . . . . . . . . . . . . . . . . . . . . 21
8.2 Return. . . . . . . . . . . . . . . . . . . . . . . . 21
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . 21
9. Specifications 22
9.1 Approvals . . . . . . . . . . . . . . . . . . . . . . 30
10. Accessories 30
10.1 Cables . . . . . . . . . . . . . . . . . . . . . . . 30
Annex: EU declaration of conformity 31

4Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
1. General information
1. General information
■
The hydraulic ring force transducers described in the operating instructions have
been designed and manufactured using state-of-the-art technology. All components
are subject to stringent quality and environmental criteria during production. Our
management systems are certified to ISO 9001.
■
These operating instructions contain important information on handling the instru-
ment. Working safely requires that all safety instructions and work instructions are
observed.
■
Observe the relevant local accident prevention regulations and general safety regula-
tions for the instrument's range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time. Pass
the operating instructions on to the next operator or owner of the instrument.
■
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
■
The general terms and conditions contained in the sales documentation shall apply.
■
Subject to technical modifications.
■
Factory calibrations/DKD/DAkkS calibrations are carried out in accordance with
international standards.
Further information:
- Internet address: www.wika.de
- Relevant data sheet: FO 52.20 - F6137
FO 52.21 - F6148
FO 52.22 - F6160
FO 52.23 - F6171
- Application consultant: Tel.: +49 9372 132-0
Fax: +49 9372 132-406
info@wika.de
Abbreviations, definitions
2-wire The two connection lines are used for the voltage supply.
The measuring signal also provides the supply current.
3-wire Two connection lines are used for the voltage supply.
One connection line is used for the measuring signal.
UB+ Positive power supply terminal
UB- Negative power supply terminal
S+ Positive output terminal
S- Negative output terminal

5Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
2. Design and function
2. Design and function
2.1 Overview
Hydraulic ring force transducer consisting of:
Pressure measuring instrument (pressure gauge or pressure sensor)
Adapter
Measuring piston case
Measuring piston (force introduction)
Electrical connection
Guard bracket
Hydraulic ring force transducer
with mounted pressure gauge
Hydraulic ring force transducer
with mounted pressure sensor

6Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
2. Design and function / 3. Safety
2.2 Description
Hydraulic ring force transducers are designed for measuring static and dynamic
compression forces.They comprise an hydraulic force transducer and a mounted
pressure measuring instrument (pressure gauge, pressure sensor).
The measuring body is manufactured from galvanised steel.The force acting on the
measuring piston is transmitted as pressure, proportional to the piston area, via the fill
fluid in the measuring piston to the pressure measuring instrument.
2.3 Scope of delivery
■
Hydraulic ring force transducer incl. mounted pressure measuring instrument
■
Operating instructions
3. Safety
3.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious
injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries
or damage to property or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for efficient
and trouble-free operation.

7Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
3. Safety
3.2 Intended use
The hydraulic ring force transducers are designed for static and dynamic compression
forces in the field of geotechnology for measuring anchor forces.
Model Nominal size Rated force
F6137 NS 82 0 ... 80 kN to 0 ... 700 kN
F6148 NS 146 0 ... 150 kN to 0 ... 1,500 kN
F6160 NS 383 0 ... 400 kN to 0 ... 3,200 kN
F6171 NS 827 0 ... 800 kN to 0 ... 6,000 kN
Applies to the hydraulic ring force transducer with mounted pressure sensor:
The instrument has been designed and tested in accordance with the relevant safety
regulations for electronic measuring instruments. Any usage outside of this is deemed to
be improper. The perfect functioning and operational safety of the force transducers can
only be guaranteed when complying with the instructions given in the operating instruc-
tions. During its use, the legal and safety regulations (e.g.VDE 0100) required for the
particular application must additionally be observed.This also applies accordingly when
using accessories. Faultless and safe operation of this measuring unit requires proper
transport, professional storage, installation and mounting and careful operation and
corrective maintenance.
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
The technical specifications contained in these operating instructions must be
observed. Improper handling or operation of the instrument outside of its technical
specifications requires the instrument to be taken out of service immediately and
inspected by an authorised service engineer.

8Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
3. Safety
Handle electronic precision measuring instruments with the required care (protect from
humidity, impacts, strong magnetic fields, static electricity and extreme temperatures,
do not insert any objects into the instrument or its openings). Plugs and sockets must be
protected from contamination.
The manufacturer shall not be liable for claims of any type based on operation contrary
to the intended use.
3.3 Improper use
WARNING!
Injuries through improper use
Improper use of the instrument can lead to hazardous situations and
injuries.
▶
Refrain from unauthorised modifications to the instrument.
Any use beyond or different to the intended use is considered as improper use.
3.4 Responsibility of the operator
The instrument is used in the industrial sector. The operator is therefore responsible for
legal obligations regarding occupational safety.
The safety instructions within these operating instructions, as well as the safety,
accident prevention and environmental protection regulations for the application area of
the instrument must be maintained.
The operator is obliged to maintain the product label in a legible condition.
To ensure safe working on the instrument, the operating company must ensure
■
that suitable first-aid equipment is available and aid is provided whenever required.
■
that the skilled electrical personnel are regularly instructed in all topics regarding
occupational safety, first aid and environmental protection and know the operating
instructions and in particular, the safety instructions contained therein.
■
that the instrument is suitable for the particular application in accordance with its
intended use.
■
that personal protective equipment is available.

9Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
3. Safety
3.5 Personnel qualification
WARNING!
Risk of injury should qualification be insufficient
Improper handling can result in considerable injury and damage to
property.
▶
The activities described in these operating instructions may only be
carried out by skilled personnel who have the qualifications described
below.
Skilled electrical personnel
Skilled electrical personnel are understood to be personnel who, based on their
technical training, know-how and experience as well as their knowledge of country-
specific regulations, current standards and directives, are capable of carrying out work
on electrical systems and independently recognising and avoiding potential hazards.
The skilled electrical personnel have been specifically trained for the work environ-
ment they are working in and know the relevant standards and regulations.The skilled
electrical personnel must comply with current legal accident prevention regulations.
3.6 Personal protective equipment
The requirements for the required protective equipment result from the ambient
conditions at the place of use, other products or the connection to other products.
The requisite personal protective equipment must be provided by the operating
company.The operator is in no way relieved of his obligations under labour law for the
safety and the protection of workers' health.
The design of the personal protective equipment must take into account all operating
parameters of the place of use.

10 Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
3.7 Labelling, safety marks
Product label
Model
Nominal size
Measuring range
Serial number, tag number
Date of manufacture calendar week/year
Output signal
Manufacturer/address
Product label for hydraulic ring force transducer with pressure gauge
Product label for hydraulic ring force transducer with pressure sensor
Observe the instructions stated on the product label of the pressure
measuring instrument.

11Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
4. Transport, packaging and storage
4. Transport, packaging and storage
4.1 Transport
Check the hydraulic ring force transducer for any damage that may have been caused
by transport. Obvious damage must be reported immediately.
CAUTION!
Damage through improper transport
With improper transport, a high level of damage to property can occur.
▶
When unloading packed goods upon delivery as well as during
internal transport, proceed carefully and observe the symbols on the
packaging.
▶
With internal transport, observe the instructions in chapter 4.2
“Packaging and storage”.
As precision measuring instruments, hydraulic ring force transducers require careful
handling during transport and mounting. Load impacts during transport (e.g. hitting
a hard surface) can lead to irreversible damage, resulting in measuring errors in the
subsequent measuring operation.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, sending for repair).
The hydraulic ring force transducer has been manufactured from coated steel.
■
The design with pressure gauge fulfils IP65 ingress protection.
■
The design with digital pressure gauge fulfils IP65 ingress protection.
■
The design with pressure sensor fulfils IP67 ingress protection.The design of the
accessory cables fulfils IP67 ingress protection.The ingress protection IP67 is only
guaranteed in the plugged-in state.
During storage the protective cap must always be on the electrical connection to avoid
the ingress of moisture and dirt.
Permissible conditions at the place of storage:
■
Storage temperature: -30 ... +60 °C
■
Humidity: 35 ... 85 % relative humidity (non-condensing)
Avoid exposure to the following factors:
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Dust, dirt, and other objects may not be deposited in such a way that they form a
force shunt with the measuring piston, since this will falsify the measurement.

12 Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
5. Commissioning, operation
5. Commissioning, operation
5.1 Mounting preparation
■
Before mounting the hydraulic ring force transducer, the anchor hole must be drilled,
the anchor positioned, concreted and allowed to harden. For this, a level installation
face must be ensured for the hydraulic ring force transducer.
5.2 Mounting instructions
CAUTION!
Damage to the instrument through improper installation
■
The sealed threaded connections of the hydraulic force transducer
must not be loosened.
■
When installing the hydraulic force transducer, pay attention to the
installation position and thus to the direction of loading.The hydraulic
force transducer must be mounted orthogonally to the anchor.
■
The hydraulic ring force transducer should only be loaded in the
intended position.
■
In the event that, on mounting, the pressure measuring instrument
(pressure gauge, digital pressure gauge or pressure sensor) lies
against the installation face, an anchor plate must be mounted under
the hydraulic ring force transducer.
■
Torsional and transverse forces must be avoided.Transverse loads
and lateral forces also include the corresponding components of the
measurands that are introduced slanted.
■
Torsional moments, eccentric loads and transverse loads or lateral
forces cause measuring errors and may permanently damage the
hydraulic ring force transducer.
■
During installation of the hydraulic ring force transducer, the
output
signal or the display instrument (force value) must always be monitored
to avoid mechanical overload.
■
When subjected to loads in other devices, a change in the zero
signal may result in permanent damage (leakage with jamming of the
measuring piston).
■
An overload must be excluded at all times.
■
Do not use the hydraulic ring force transducer as a climbing aid.

13Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
Wand
Elektrischer Anschluss
Kraftaufnehmer
Ankerplatte
Anker
Spannmutter
Wand
Elektrischer Anschluss
Kraftaufnehmer
Ankerplatte
Anker
Spannmutter
5. Commissioning, operation
Positioning of anchor plate 1 on the anchor . This must lie against the concrete pad .
Fit the hydraulic ring force transducer onto the anchor via anchor plate 1 .
5.3 Mounting of the hydraulic ring force transducer
Hydraulic ring force transducer with mounted pressure measuring instrument
(pressure gauge, pressure sensor)
not included in delivery:
Anchor plate 1 Clamping nut
Anchor Concrete pad
Anchor plate 2

14 Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
Wand
Elektrischer Anschluss
Kraftaufnehmer
Ankerplatte
Anker
Spannmutter
Wand
Elektrischer Anschluss
Kraftaufnehmer
Ankerplatte
Anker
Spannmutter
5. Commissioning, operation
Positioning anchor plate 2 on the hydraulic ring force transducer (piston side) on the
anchor .
Preloading the anchor using a hydraulic press. Screwing in the clamping nut .
If necessary, fix using another method.

15Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
5. Commissioning, operation
5.4 Commissioning of the hydraulic ring force transducer with digital pressure
gauge
Function / action
■
Switches the instrument on
■
Switches the instrument off
■
Display indicates the max. value while
the key is pressed
■
Display indicates the min. value while
the key is pressed
■
Resets the max. value to “ 0 ”
■
Resets the min. value to “ 0 ”
+
+
Display
Keys
Digital pressure gauge
Keys

16 Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
Display Bar graph display
Unit
Full scale
Max. value
Min. value
Main display
Special function
2nd display
Tare
Battery-charge-level
indicator LO-BAT
Min./max. memory
The min./max. memory is reset to the current measured value with each measuring cycle.
■
The display of the min. value is achieved by pressing the Min key.
■
The display of the max. value is achieved by pressing the Max key.
■
The resetting of the current measured value is achieved by holding down the respec-
tive key (Min or Max) and additionally briefly pressing the Menu key.
Bar graph with drag-pointer function
The additional integrated bar graph with drag-pointer function within the display
indicates the trend of the current operating pressure directly, independently from
the digital indicator. Using the drag-pointer function, the max. memory value is also
displayed in the bar graph display in the form of a bar segment, as well as in the digital
display. On resetting the min./max. value, this bar graph is then updated to the current
measured value.
Voltage supply
The digital pressure gauge is operated with 2 AA-Mignon batteries.The status of the
batteries is shown in the display .With a low battery charge level, a blinking battery
symbol is shown on the indicator display - the batteries should then be changed as soon
as possible.
■
Open the battery cover of the digital pressure gauge on the rear of the instrument.
■
Replace the batteries with 2 new mignon cells.
■
Close the battery cover again.
5. Commissioning, operation
For further information and advice see the appropriate operating
instructions DG-10.

17Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
Output 4 ... 20 mA, 2-wire
Circular connector M12 x 1, 4-pin
5. Commissioning, operation
Circular connector M12 x 1, 4-pin
Pin
UB+ 1
0V/UB- 3
S+ 1
S- 3
Shield Case
5.5 Electrical connection - Hydraulic ring force transducer with pressure sensor
To prevent interferences from coupling into the system, observe the following
information:
■
Use only shielded and low-capacitance measuring cables (for cables, see chapter 10
“Accessories”).
■
Ground the cable shield.
■
In the cables of the accessories, the cable shield is connected by means of the
knurled nut, thus connecting it to the case of the pressure sensor.
■
Do not install measuring cables in parallel to 3-phase-current cables and control
cables.
■
Avoid stray fields of transformers, motors and contactors.
■
Force transducers, amplifiers and processing or display units must not be grounded
several times. Connect all instruments to the same protective conductor.
The pin assignment of the connector or of the cable can be found on the product label.
For an extension, only shielded and low-capacitance cables should be used.The
permitted maximum and minimum lengths of cable are defined in ISO 11898-2. Care
should also be taken with the shielding to ensure a high-quality connection.
5.5.1 Pin assignment of the analogue output - Hydraulic ring force transducer
with mounted pressure sensor
For further information and advice see the appropriate operating
instructions of the mounted pressure sensor.

18 Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
5. Commissioning, operation / 6. Faults
6. Faults
CAUTION!
Physical injuries, damage to property
If faults cannot be eliminated by means of the listed measures, the hydrau-
lic ring force transducer must be taken out of operation immediately.
▶
Contact the manufacturer.
▶
If a return is needed, please follow the instructions given in chapter 9.2
“Return”.
For contact details see chapter 1 “General information” or the back page
of the operating instructions.
Faults Causes Measures
No display Leaking from the hydraulic ring
force transducer
Visual check whether the fill
fluid has leaked (glycerine-
water mixture)
Consult the manufacturer
Deviation of the zero point
display
Overload, load offset,
defective force introduction
Consult the manufacturer
Output signal constant
when changing force
Mechanical overload,
defective force introduction
Consult the manufacturer
Signal span varies Electromagnetic interference
sources in the environment;
e.g. frequency converter
Shield instrument; cable
shield; remove source of
interference
Signal span drops/too small Mechanical overload, leaking
of the hydraulic ring force
transducer
Consult the manufacturer
Pointer jerks with force
increase
Stick-slip effect (internal
pressure too low)
Consult the manufacturer
6.1 Faults of the hydraulic ring force transducer with pressure gauge/digital
pressure gauge

19Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
6. Faults
Faults Causes Measures
No output signal
or wrong output signal
No or wrong supply voltage,
current pulse
Rectify the supply voltage
Cable break Check the continuity
Wrong pin assignment Check pin assignment
Deviating zero-point signal Overload, load offset, wrong
connection
Consult the manufacturer
Constant output signal
when changing force
Mechanical overload, wrong
pin assignment, defective force
introduction
Consult the manufacturer
Leaking from the hydraulic ring
force transducer
Visual check whether the fill
fluid has leaked (glycerine-
water mixture)
Signal span varies Electromagnetic interference
sources in the environment;
e.g. frequency converter
Shield instrument; cable
shield; remove source of
interference
Signal span drops/too small Mechanical overload, leaking
of the hydraulic ring force
transducer
Consult the manufacturer
6.2 Faults of the hydraulic ring force transducer with pressure sensor

20 Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
7. Maintenance and cleaning
7. Maintenance and cleaning
For contact details see chapter 1 “General information” or the back page
of the operating instructions.
7.1 Maintenance
This instrument is maintenance-free.
Repairs must only be carried out by the manufacturer.
Only use original parts (see chapter 10 “Accessories”).
7.2 Cleaning
7.2.1 Hydraulic ring force transducer with pressure gauge/digital pressure
gauge
1. Clean the hydraulic ring force transducer with a cloth.
7.2.2 Hydraulic ring force transducer with pressure sensor
1. Prior to cleaning, disconnect the hydraulic ring force transducer from the voltage
supply and dismount it.
2. Clean the hydraulic ring force transducer with a cloth.
Electrical connections must not come into contact with moisture!
CAUTION!
Damage to the instrument
Improper cleaning may lead to damage to the instrument!
▶
Do not use any aggressive cleaning agents.
▶
Do not use any hard or pointed objects for cleaning.
7.3 Recalibration
DKD/DAkkS certificate - official certificates:
We recommend that the hydraulic ring force transducer be regularly recalibrated by the
manufacturer, with time intervals of approx. 24 months.The basic settings will be cor-
rected if necessary.