Woodley CLINISPIN HORIZON 842VES User manual

Operator’s Manual
CLINISPIN HORIZON 842VES Laboratory Centrifuge
Rev. A
p/n 7711038

TABLE OF CONTENTS
Model Description page 3
Supplied Equipment page 3
Warranty Information page 3
Specications page4
Control Panel / Parts of the Centrifuge page 5
Setup Location page 6
Initial Setup Procedure pages 6,7
Operation page 7
RotorCongurations
Performance Plus Horizontal Rotor page 7
6-Place Horizontal Rotor page 8
6-Place Fixed Angle Rotor page 9
Control Panel Functions
Time Adjustment and Timer Operation page 10
Speed Adjustment / RCF Display page 10
Starting and Stopping a Run page 10
Unlocking the Lid page 10
Advanced User Settings
Acceleration Rate page 10
Braking Rate page 11
Countdown Delay page 11
Imbalance Detection page 11
Audible Setting page 11
Memory Locations / Storing and Recalling page 12
Digital Cycle Counter page 12
Care and Preventative Maintenance page 13
RotorRemovalandInstallation page14
Troubleshooting page 15
Safety page 16
Emergency Rotor Chamber Entry page 16
Calibration and Earth Ground Testing page 16
Replacement Parts page 17
page 2

WARNING: For the safety of both the operator and service personnel,
care should be taken when using this centrifuge if handling substances
that are known to be toxic, radioactive or contaminated with pathogenic
microorganisms..WhenRiskGroupIImaterialsareused(asidentiedinthe
World Health Organization “Laboratory Bio-Safety Manual”), a Bio-Seal should
be employed. The rotor and rotor accessories should not be considered as
Bio-Seals. In the event that materials of a higher risk group are being used,
morethanonelevelofprotectionmustbeprovided.Theuseofammableor
explosive materials as well as those materials which have a vigorous chemical
reaction is prohibited.
Model Description
TheClinispinHorizon842VESisacontinuos-duty,digitally-controlledcentrifugedesignedforthe
separationofuidsamples.The842VESofferscontroloveralmosteveryaspectoftheunit’s
operation from variable acceleration and deceleration
to easy time and speed input. High speed and force
capability produces platelet-poor/free plasma and
greatly reduce processing times. The exclusive Woodley
Equipment horizontal rotor allows for quick and easy
sample loading and complete horizontal separation. Save
settings in one of 10 memory locations and recall them at
the touch of a button.
SUPPLIED ACCESSORIES (Standard)
One (1) 6-Perfor-
mance plus rotor
p/n
7786017
Also included (not shown):
•One(1)MainsLead •One(1)Operator’sManual
Six (6)
75mm Tube Holders
p/n 7713056
See the next
page for a
description of
optional
accessories.
page 3
WARRANTY:
Woodley Equipment Company warranties that this centrifuge is free from
defects in workmanship and parts for 2 years.
Old Station Park Buildings
St. John’s Street, Horwich
Bolton BL6 7NY, United Kingdom
Tel: +44 (0) 1204 669033 • Fax: +44 (0) 1204 669034
www.woodleyequipment.com Email: sales@woodleyequipment.com
Made in the USA
•Therotorandrotoraccessoriesareratedforarotationalspeedof6500RPM(aforceof5000xg).

Any use other than those specied by the Manufacturer is explicitly prohibited.
Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)
page4
1
This rotor or rotor accessory is rated for a rotational speed of 5000 RPM and a force of 3000 xg.
2
This rotor or rotor accessory is rated for a rotational speed of 6500 RPM and a force of 2500 xg.
3
This rotor or rotor accessory is rated for a rotational speed of 7000 RPM and a force of 5500 xg.
PART NO.# MAX TUBE SIZE SPEED RANGE FORCE RANGE
6-Place Horizontal (Standard) 77860611
75/100mm tube holders 771707916 100mm 1000 to 4,600 RPM 140 to 3,000 xg
6-Place Fixed-Angle Rotor 77860472
75mm tube holder 771304316 17mm x 75mm 1000 to 4,000 RPM 140 to 1,500 xg
100mm tube holder 771304016 17mm x 100mm 1000 to 4,000 RPM 140 to 1,730
xg
125mm tube holder 771304426 17mm x 125mm 1000 to 4,000 RPM 140 to 1,950
xg
6-Place Performance Plus Horizontal 77860173
75mm tube holder 771305636 13mm x 75mm 1000 to 6,500 RPM 140 to 5,000 xg
12-Place combo Fixed-Angle STAT 77860033
75mm tube locations n/a36 13mm x 75mm 1000 to 7,000 RPM 140 to 4,880 xg
100mm tube locations n/a36 13mm x 100mm 1000 to 7,000 RPM 140 to 5,700
xg
12-Place Fixed-Angle STAT 77860013
75mm tube locations n/a312 13mm x 75mm 1000 to 7,000 RPM 140 to 4,880 xg
Overall Dimensions (H x W x D): 8.0 in. x 11.0 in. x 13.5 in.
Centrifuge Motor: 1/2 H.P. Brushless DC
Protection Breaker: 4Amp.re–settable
Timer: electronic, with hold
or 0 to 99 minutes, +/- 1%
Weight: 23 lbs.
Permitted Environmental Conditions
Ambient Temperature During Operation: 2
oC-40oC
Maximum Relative Air Humidity: 90%
Electrical Requirements:
Power: 200 Watts
Voltage: NorthAmerica: 115Volts(+/-10%)
Japan: 100Volts(+/-10%)
Europe: 230Volts(+/-10%)
Frequency: 50/60 Hz
Any use other than those specied by the Manufacturer is explicitly prohibited.
Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)
Specications:
General Specications for the Horizon Model 842VES Centrifuge
125 mm
Tube Holder
p/n 7713044
2
13 x 75mm
Performance Tube
Holder
p/n 7713056
3
13 x 100mm
Tube Holder
p/n 7717079
3
Tube Holder
Cap
p/n 7713011
1” Cushion
Spacer
p/n 1525
Refer to serial number label
on rear of centrifuge to
conrmvoltagerequirement.
NOTE: Contact Drucker Diagnostics for assistance when changing the installed rotor. The installed rotor may only be
changed by qualied personnel.
.5 to 1 ml
tube Adapter
p/n 7713068
13 x 75mm
insert, p/n
7713064
13 x 100 mm
insert, p/n
7713066
1.5 to 2mL
tube Adapter
p/n 7713065
B
B
C
C
D
D

Lid Knob
Lid
Lid Safety
Interlock System
Control Panel
(see below)
Cabinet
CONTROL PANEL / PARTS OF THE CENTRIFUGE:
Up/Down RPM
Change Buttons
RPM Display
Running
Indicator
Light
Locked
Indicator
Light
Unlocked
Indicator
Light
Start Button
Open / Stop Button
Memory ButtonMemory
Preset
Display
Program
Button
Brake
Display
Up/Down Time
Change Buttons
Time Display
RCF Conversion Button
page 5

SETUP LOCATION:
1. Unpack the centrifuge and verify that all of the supplied equipment is present.
2. Choose a setup location which meets the following criteria:
a) Abenchtopclearanceheightof40cmsisrequiredinordertoopenthelid.
b) The clearance envelope is the space around the centrifuge which is required for
safety. Choose a setup location which will allow for a clearance envelope of at least
61 x 61 cms, (with the centrifuge at the center). No person or hazardous material shall
be permitted in the clearance envelope during operation. The operator time within
the envelope shall be limited to the time necessary for loading, unloading and
centrifuge operation only.
c) Proper ventilation is necessary to prevent the overheating of samples as well as
premature failure of the centrifuge. Choose an area which will allow unencumbered
airow.
d) The centrifuge is designed to rest on its four rubber feet. No adjustment is necessary
forlevelingthecentrifuge,however,thesurfaceshouldbeatandlevel.
e) Be sure the outlet is always within reach as the mains lead is the means of
emergency disconnection!
INITIAL SETUP PROCEDURE: If any problems are found during the initial setup
procedure, refer to the troubleshooting section. For further assistance, contact Woodley Equipment
Company LTD. at +44 (0) 1204 669033
1. Plug the female end of the mains Lead into the rear of the centrifuge. Plug the male end into
an approved electrical outlet. For electrical safety, the unit must always be properly
grounded.
2. For operator safety, the locking system is always active; requiring power and direction
from the user to disengage it (the lid also automatically unlocks at the end of a run when
it is safe to do so). To unlock the lid (in order to access the rotor chamber) press the
‘OPEN/STOP’buttononthecontrolpanel.The‘UNLOCKED’indicatorlightshould
illuminate. If it does not, refer to the section on troubleshooting. The lid will remain
unlockedfor15secondsafterpressingthe‘OPEN/STOP’button.
3. Turn the lock counterclockwise and open the lid.
4. Spintherotorbyhand;checkforfreeandlevelrotation.Iftherotordoesnotspinfreely,
refer to the section on troubleshooting.
5. Place the test tube holders inside the rotor and verify that they are seated properly.
6. Close the lid. Rotate the lid knob clockwise to its complete stop position. The
’LOCKED’indicatorlightshouldbeilluminated.Ifitisnot,makesurethatthelidis
locked properly. The centrifuge will not run unless the lid is latched properly and the
’LOCKED’indicatorlightisilluminated.
7. UsetheuparrowbuttonnexttotheRPMdisplaytosetthespeedto4000RPM.
8. Initiateatestrunbypressingthe‘START’button.
9. The‘RUNNING’indicatorlightwillilluminate.
10. The test tube carriers will slide up into the horizontal position (horizontal rotor only) and
the unit will accelerate to the current set speed.
11. Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If there
areanyloudorunusualsounds,stopthecentrifugebypressingthe’OPEN/STOP’button
immediately and refer to the section on troubleshooting.
12. Pressthe‘OPEN/STOP’buttontoterminatethetestrun.Therotorwilldeceleratetoa
complete stop and the lid will then unlock automatically for sixty (60) seconds.
13. Take time now to familiarize yourself with the various additional user settings. Turn to
page 11.
14. Thecentrifugeisnowreadyforoperation.
page 6

OPERATION:
NOTE: Follow the initial setup procedure before initial operation.
1. Pressthe‘OPEN/STOP’buttontounlockthelidandthenopenthelid.
2. Insert cushions (if needed) into the tube carriers (holders) for the tube size you are
using.Referto‘RotorCongurations’forassistance.
3. Place the test tube samples into the tube carriers (holders). Be sure to follow the
rules for balanced loads.
4. Closethelidandturnthelidknobclockwisetoitscompletestopposition.The
’LOCKED’indicatorlightshouldilluminatetoindicatethatthelatchisclosed
properly.Ifthelidknobisnotcompletelylatched,the‘LOCKED’indicatorlightwillnot
illuminate and the centrifuge cannot be operated.
5. Set the desired speed and run time using the appropriate up and down arrow buttons.
6. If desired, adjust the braking level, acceleration rate, countdown delay, imbalance
sensitivity and audible indication. Turn to page 11 for a description of these features.
7. Begintherunbypressingthe‘START’buttononthecontrolpanel.
8. Thecentrifugeshouldbegintospin.The‘RUNNING’indicatorlightshouldilluminate.
IF A PROBLEM IS FOUND DURING A SPIN THAT REQUIRES THE CENTRI-
FUGE TO SHUT DOWN, PRESS THE ‘OPEN / STOP’ BUTTON IMMEDIATELY !
9. Aftertimehaselapsed,the‘RUNNING’indicatorlightwillextinguishandtherotorwill
decelerate to a complete stop.
11. The‘UNLOCKED’indicatorlightwillilluminateandthelockingmechanismwill
disengage, allowing entry into the rotor chamber. If the automatic unlock times out
(after60seconds)simplypressthe‘OPEN/STOP’button.
12. Turn the lid knob counterclockwise and open the lid.
13. Remove the samples.
14. Thecentrifugemaybeusedimmediately.
page 7
Six-Place, Performance-Plus Horizontal Rotor (Standard):
Your centrifuge must contain a balanced load in order to
work properly. To ensure that the load is balanced, keep
these rules in mind when inserting test tube samples.
1. Opposing tube holders must be identical.
2. Opposing tube holders must be empty or loaded equally weighted samples.
3. If an odd number of samples is to be spun, use a water-lled tube to mate
with the unpaired one.
This rotor is designed to hold up to six (6) test tubes, up to 17mm x 75mm.
Rotor Congurations:
The next three pages will describe proper balancing and tube loading. Use the appropriate section
for the rotor installed in your centrifuge. Improper loading can result in damage to the centrifuge or
test tube samples.
Your centrifuge must contain a balanced load in order to
work properly. To ensure that the load is balanced, keep
these rules in mind when inserting test tube samples.
1. Opposing tube locations must be empty or loaded with equally
weighted samples.
2. If an odd number of samples is to be spun, use a water-lled tube to
mate with the unpaired one.
This rotor is designed to hold up to eight (8) test tubes, up to 17mm x 100mm.
12 Place, 12 Place combo Fixed-Angle STAT Rotor (optional):

page 8
BLUE
A. & B.
BLUE
A.
B.
C.
B
B
C
C
D
D
C.
75/100 mm
Tube Holder
* This part is available as an accessory. Contact The Drucker Diagnostics for assistance.
75/100 mm Tube
Holder with
1525 cushion*
Tube Holder Congurations:
The horizontal rotor is capable of spinning test tubes up to 17 mm x 100 mm with
its horizontal rotor. Use the following chart and drawing to determine which tube
holder and cushion combination should be used with your application.
DIRECTIONS:
1. Compare the tube to be spun with the three boxes shown below.
2. Findtheboxthatmostcloselymatchesthetube’slength.NOTE:The
tube length with its stopper or cap must be shorter then the chosen
boxorthetubewillnottproperlyinthetubeholder.
3. Matchtheletterfromthechosenboxwithoneofthecongurationsshown.
For Example: A tube is found to be as long as box C. Accordingly, we
can use a 75/100 mm tube holder (congurations C).

page 9
BLACK
B.
C.
D.
A. A. B. & C D.
125mm
Tube
Holder
75/100
mm
Tube
Holder
75/100
mm
Tube
Holder
with
1525
cushion*
Tube Holder Congurations:
Thexedanglerotoriscapableofspinningtesttubesupto17mmx125mm.
Use the following chart and drawing to determine which tube holder and cushion
combination should be used with your application.
DIRECTIONS:
1. Compare the tube to be spun with the three boxes shown below.
2. Findtheboxthatmostcloselymatchesthetube’slength.NOTE:The
tube length with its stopper or cap must be shorter then the chosen
boxorthetubewillnottproperlyinthetubeholder.
3. Matchtheletterfromthechosenboxwithoneofthecongurationsshown.
For Example: A tube is found to be as long as box B. Accordingly, we
can use a 125 mm tube holder with a 1525 cushion or a 75/100 mm
tube holder with no cushion, (congurations B).

CONTROL PANEL FUNCTIONS:
Time Adjustment and Timer Operation -The run time may be set from 30 seconds to 99
minutes and 30 seconds. Press the up and down arrow buttons next to the time display to adjust
the run time. Adjustments may be made prior to a run or after a run begins. A quick tap will adjust
the time by 30 seconds. Hold down the button for 1 minute adjustments - continue to hold for 5
minute adjustments.
Press the down button once while 00:30 is displayed to access the “hold” feature. When a “hold”
run is initiated, the timer will count up to keep track of total run time. The unit will run for a maximum
of 99 minutes and 30 seconds and will then shut off automatically. The “hold” run may also be
terminated by pressing the STOP/OPEN button. The total run time during a “hold” run is displayed
on the time display until the operator unlocks the lid by pressing the OPEN button or opens the lid
(during the automatic unlock after the run ends).
Speed Adjustment / RCF Display - Press the up and down arrow buttons next to the speed
display to change to run speed. Adjustments may be made prior to a run or after a run begins.
A quick tap will adjust the speed by 50 RPMs. Hold down the button for 100 RPM adjustments.
Continue to hold for 500 RPM adjustments.
Press the RCF button at any time to convert the current speed setting to the corresponding force
value (xg). The speed may be set by RCF by holding down the RCF button while using the up/down
speed set buttons. The jumps made in the RCF value while holding down the RCF button and
pressing the up and down arrow buttons correspond to 50 RPM adjustments in speed. Since the
RCF is a function of the square of the speed, this jump size will vary.
Starting and Stopping a Run - With the lid switch closed, press the START button to
begin a run. Press the STOP/OPEN button at any time during a run to terminate it.
Unlocking the Lid - ThelidisunlockedwhenevertheredUNLOCKEDindicatorlightis
illuminated. The lid unlocks automatically for 60 seconds at the end of a run. The lid can also be
unlocked for an additional 15 seconds by pressing the OPEN/STOP button while the unit is idle.
The lid cannot be unlocked while the rotor is spinning.
Imbalance Detection Reset - This centrifuge is equipped with imbalance detection. If an
imbalance is detected, the centrifuge will terminate the current run and will begin to brake to a stop.
Thewords“BALANC”and“ERROR”willashonthespeeddisplay.Oncetherotorhasstopped,
open the lid to cancel the error reporting, balance the load and begin a new run. Alternately, the
error reporting can be canceled by pressing the OPEN/STOP button.
ADVANCED USER SETTINGS:
This centrifuge allows for total control of the run by the user. You may adjust the acceleration rate,
adjust the braking rate, adjust the imbalance sensitivity level, select from several audible settings,
and use the optional countdown delay.
Acceleration Rate - The user may adjust the acceleration rate from minimum (1) to maximum
(10) using the following procedure:
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “ACCEL” is
displayed in the speed display.
3. The current acceleration rate is displayed in the time display. Use the ‘UP/
DOWN arrow buttons next to the time display to adjust the acceleration rate
from 1 (minimum) to 10 (maximum).
4. PressthePROGRAMbuttontoexitorusetheUP/DOWNarrowbuttonsnextto
the speed display to adjust additional settings.
page 10

ADVANCED USER SETTINGS (continued):
Braking Rate - The user may adjust the braking rate from minimum ( slowest deceleration, 0) to
maximum (quickest deceleration, 9) using the following procedure:
1. While the centrifuge is idle, press the PROGRAM button.
2. UsetheUP/DOWNarrowbuttonsnexttothespeeddisplayuntil“BRAKE”is
displayed in the speed display.
3. The current braking rate is displayed in the brake display. Use the UP/DOWN
arrow buttons next to the brake display to adjust the braking rate from 0
(minimum) to 9 (maximum).
4. PressthePROGRAMbuttontoexitorusetheUP/DOWNarrowbuttonsnextto
the speed display to adjust additional settings.
Countdown Delay - When the countdown delay is turned on, the timer will not begin
decrementing until the rotor has reached full speed as set by the user. This feature ensures that
the samples see full force for the total run time.
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “CNTDLY” is
displayed in the speed display.
3. The current status of this feature (ON or OFF) is displayed in the time display.
Press UP/DOWN arrows next to the time display to toggle the status between
ON and OFF.
4. PressthePROGRAMbuttontoexitorusetheUP/DOWNarrowbuttonsnextto
the speed display to adjust additional settings.
Imbalance Detection - This centrifuge is equipped with imbalance detection. This ensures that
the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction
occurs. This feature is intended to provide for operator safety and to extend the life of the
centrifuge. The LOW setting will allow for a greater imbalance before shutting the centrifuge down.
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “BALANC” is
displayed in the speed display.
3. The current status of this feature (LOW or HI) is displayed in the time display.
Press UP/DOWN arrows next to the time display to toggle the status between
LOW and HI.
4. PressthePROGRAMbuttontoexitorusetheUP/DOWNarrowbuttonsnextto
the speed display to adjust additional settings.
Audible Settings - The user may select from six different audible indicator levels:
Setting ButtonPressBeep? End-of-runaudiblenotication
1 NO NONE
2 YES NONE
3 NO YES, 5 beeps
4 YES YES,5beeps
5 NO YES, continuous, user must press STOP/OPEN button to cancel
6 YES YES, continuous, user must press STOP/OPEN button to cancel
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “BEEPER” is
displayed in the speed display.
3. The current audible selection is displayed in the time display. Use the UP/
DOWN arrow buttons next to the time display to select a setting (1 to 6).
4. PressthePROGRAMbuttontoexitorusetheUP/DOWNarrowbuttonsnextto
the speed display to adjust additional settings.
page 11

TheClinispinhorizon842VEScentrifugeiscapableofstoringupto10user-denedpresets.
Thesememorypresetscontainallofthesettingsneededtodeneaspecicrun,(runtime,speed,
accelerationrate,brakingrate,etc.).Theusercanusethesememorylocationstoquicklycongure
thecentrifugeforaspecictesttypeandensurethatthetestisruninthesamewayeachtimeby
recalling this same setting.
Store a conguration in a memory location
1. Adjust the various settings (speed, time, acceleration rate, etc.) for the
congurationtobesaved.
2. PressandholdthePROGRAMbuttonuntilthepresetdisplaybeginstoash.
3. Press the MEMORY button to increment the memory location by one. Continue
to press the memory button until the desired location is displayed.
4. Presstheprogrambuttontostorethecurrentcongurationintothememory
location that is displayed.
Recall a conguration from a memory location
1. Press the MEMORY button until the desired memory location is displayed in the
memory display.
2. As you review the presets, the settings on the control panel will change to
reectthesettingsstoredinthememorypresetcurrentlydisplayed.
3. When the desired preset is found, pause and the memory display will
momentarilyash.Thestoredcongurationisnowloaded.
Notes:
Onceacongurationhasbeenloaded,itwillremainactiveuntilanychangeismadeonthecontrol
panel.Ifforinstancetheuseradjuststhetime,thecurrentcongurationwillnolongermatchthe
previouslyloadedcongurationandthememorydisplaywillchangetoa“-”dash.Theprevious
congurationmayberecalledusingtheproceduredescribedabove.
Memory locations my be overwritten by selecting them during the store procedure.
DIGITAL CYCLE COUNTER:
This centrifuge is equipped with a digital cycle counter to monitor machine usage and for routine
maintenance.
To display the current cycle count, perform the following:
1. Unlock the lid by pressing the OPEN/STOP button.
2.OpenthelidlatchsothattheyellowLOCKEDindicatorlightisnotilluminated.
3.PressandholdtheSTARTbuttonforapproximatelyveseconds.Theword
“CYCLES” will be displayed in the speed display and the current cycle count will
be displayed in the time display. If the current cycle count exceeds 9,999
cycles,thecyclecountwillbedisplayedbyshowingtherstpartofthenumber
and then the end of the number. For example: the 1st shown is 0002 and the
2ndshownis4537represents24,537cycles.
MEMORY LOCATIONS / STORING AND RECALLING:
page 12

CARE AND PREVENTATIVE MAINTENANCE:
With proper care and maintenance your Horizon centrifuge will provide years of laboratory service.
For proper care, the following steps should be taken:
1. Provide Adequate Ventilation: For cooling purposes, the Horizon draws in ambient air
through the air intake cover on the top of the lid and exhausts this air in the rear of the base.
The centrifuge should be placed on a hard smooth surface for good air circulation.
2. Always Spin Balanced Loads: This centrifuge is equipped with imbalance detection.
However, it is still possible to run loads that are minimally imbalanced, particularly at an
imbalancesettingofLOWSENSITIVITY.Formaximumcentrifugelife,settheimbalance
detectiontoHISENSITIVITYandalwaysrunbalancedloads.Refertopages8and9onfor
additional information on balancing the load.
3. Keep the Tube Holders Clean: NOTE: Always follow the safety guidelines of your
laboratory to properly clean up and/or dispose of materials in the event that a substance
known to be potentially toxic, radioactive or contaminated with a pathogenic
microorganism is spilt in or on the centrifuge. Small glass fragments left in the tube carrier /
holder after a tube breakage may adhere to the next test tube inserted in that carrier /
holder. When this tube is handled, these fragments may puncture protective gloves and
laceratetheoperator’sngersorhand.Remainingfragmentsmayprovidestresspoints
on subsequent tubes and result in additional breakage. If a tube breakage occurs,
carefully remove the tube holder. Properly dispose of the sample and tube fragments and
thoroughly clean both the inside and outside of the tube carrier / holder. Insert a new tube
cushion (if necessary) and replace the tube carrier / holder in the rotor.
4. Motor and Electrical Maintenance: This centrifuge uses a brushless-DC motor. There
are no brushes to replace and it should not need routine servicing for the life of the
centrifuge. The electrical components are selected for high reliability and should not
need service.
5.Keep the Centrifuge Clean: The cabinet, rotor top and accessories shall be thoroughly
cleanedusingeitherisopropylalcohol,soapandwater.MustbeLysol(1/4cup/gallon)or“Cidex”.
NotallLysolwillbesufcient(mustbeaLysolAmphyldisinfectivecleanerorVirkon).Theuseof
Fully/PartiallyHalogenatedHydrocarbons,Ketones,Estersandallotherchemicalsnotprescribed
by the manufacturer may cause damage to the rotor and tube carriers / holders and shall not be
used. Before using any cleaning or decontamination methods except those recommended by the
manufacturer, users should check with the manufacturer that the proposed method will not
damage the equipment.
Ifitisnecessarytoremovetherotorforadditionalcleaningitisrequiredthataqualied
technician remove the rotor assembly.
Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in water or
other cleaning solutions as this will cause damage and void your warranty!
If hazardous material is spilt on or inside the centrifuge, the centrifuge must be decontaminated.
6. Tube Holder Replacement: All tube holders, samples, and caps must be removed from the rotor
chamber before transporting or storing the centrifuge to prevent damage and injury.
page 13

page14
ROTOR REMOVAL AND INSTALLATION:
To remove the rotor:
1. Unlockthecentrifugebypushingthe‘OPEN/EMERGENCYSTOP’buttonandunlockand
open the lid.
CAUTION: Unplug the centrifuge from the electrical outlet at this time to eliminate the
possibility of electrical shock or other injury.
2. Remove the test tube holders.
3. Remove the screw in the center of the rotor with a 1/8” hex key by turning the screw counter
clockwise. Be sure to save the screw and washer in a safe place.
4. Therotorissittingonacone-shapedadapter.Pulltherotorupandoffofthisadapter.
5. Remove the rotor from the rotor chamber.
NOTE: If you are removing the performance plus rotor, you must:1) Move the rotor towards
the front of the centrifuge; 2) Angle the back edge of the rotor; 3) Slide it through the
opening. See illustration below.
To install the rotor:
1. Place the rotor back onto the cone-shaped adapter. You may need to turn the rotor slightly to
line it up properly.
NOTE: If you are installing the performance plus rotor, angle the rotor as needed to get it back
into the rotor chamber.
2. The rotor should slide onto the rotor cone freely.
3. Onceaproperthasbeenachieved,threadthescrewandwasherbackintothecenterofthe
rotor. Use the Allen wrench to tighten the screw.
4. Replacethetubeholdersandverifythattheyareseatedproperly.
5. It is recommended that the initial setup procedures be performed to ensure that the rotor has
been installed correctly and that no damage has been done to the centrifuge during either the
rotor installation or possible rotor chamber cleaning.
Proper position for removing or installing the performance-plus rotor

TROUBLESHOOTING:
1. Problem: The rotor does not spin freely.
Solutions: •Makesurethatnothinghasfallenintotherotorchamber.
•Ifthereisnothingobstructingtherotor,themotormaybedamaged.
Contact your authorised dealer or Woodley Equipment Company for further
assistance.
2. Problem: There is excessive noise when the machine is running.
Solutions: •Checktoseethattheloadisbalanced.
•Makesurethatnothinghasfallenintotherotorchamber.
•Makesurethatthescrewinthecenteroftherotorissnug.
•Themotormaybedamaged.Contactyourauthoriseddealeror
Woodley Equipment Company for further assistance.
3. Problem: The centrifuge does not run or an error message is displayed.
Solutions: •Makesurethatthecentrifugeisgettingpower.Doesthecontrolpanelcome
on? Check the electrical outlet that the unit is plugged into. Check the
circuit breaker on the base near the rear of the centrifuge.
•Iftheunitisgettingpower,makesurethatthelidlatchisclosedproperly.
Thelatchisclosedproperlywhentheyellow‘LOCKED’indicatorlightis
illuminated.
•Ifthecentrifugestopssoonafterstart-upand‘ERROR’isdisplayedonthe
speed display, refer to the following table for fault information:
ERROR / SPEED
The centrifuge cannot reach full speed due to a problem with the rotor, an
inadequate power supply, or other electrical problems. Press the OPEN/
STOP button to cancel the error and then check the rotor and the line
voltage. Contact your authorised dealer or Woodley Equipment Company.
ERROR / BALANC
The centrifuge has detected an imbalance. Press the OPEN/STOP button to
cancel the error and balance the load. If the load is balanced, make sure
that the centrifuge is installed on a secure, level location.
4. Problem: The lid knob cannot be turned / the lid cannot be unlocked.
Solutions: •Makesurethecentrifugehaspowerandthattherotorisstopped.Pressthe
‘OPEN/STOP’button.Thered‘UNLOCKED’indicatorlightshouldilluminate
and the unlocked mechanism should disengage, allowing entry into the lid.
•Ifthelidisstilllocked,makesurethatthered‘UNLOCKED’indicatoris
illuminatedandturnthelidknobrstcompletelyclockwiseandthen
counterclockwise.
•Ifthered‘UNLOCKED’indicatorwillnotilluminateorthelocking
mechanism will not disengage, the electronics or locking mechanism may
be damaged. Have a technician service the centrifuge or contact Woodley
Equipment Company for further assistance.
page 15
For additional assistance, services and technical support
contact Woodley Equipment Company LTD.

SAFETY:
Horizon Lid Safety Switch: The Horizon lid is secured to the top of the cabinet by a locking knob
and pawl system. When the knob is rotated clockwise, the pawl grips the underside of the cabinet
opening and prevents the lid from opening. A mechanical stop positions the pawl and prevents it
fromrotatingcompletely.Whenrotatedtothestopposition,thepawlmakescontactwithamicro–
switch mounted underneath the cabinet top. The lid safety switch prevents the centrifuge from
operatingwhilethelidisopen.Theyellow‘LOCKED’indicatorlightonthefrontofthemachinewill
illuminate when the lid has been locked properly.
Horizon Lid Safety Interlock System: In addition to the Lid Safety Switch, the Horizon has a
true “0 RPM” lid locking system. The lid safety interlock system keeps the lid locked at all times,
(even during power failure), and requires that the rotor be at rest in order to unlock the lid. The
centrifuge will not allow entry into the rotor chamber unless the centrifuge has power and the rotor
is stopped. To open the lid, make sure that the centrifuge is plugged in and, with the rotor stopped,
pressthe‘OPEN/EMERGENCYSTOP’button.
Note: After the centrifuge has started spinning, it may be possible to rotate the lid knob enough
to cause the pawl to lose contact with the lid safety switch. If this happens, the centrifuge motor
may lose power, but the lid will still remain locked. If the knob is accidentally moved and this
situation should occur, rotate the knob fully clockwise to its stop position and the centrifuge will
resume operation.
Circuit Breaker: TheHorizonisprotectedwitha4Ampcircuitbreakerlocatedattherearofthe
machine mounted to the base. Any electrical short circuit will cause the breaker to cut power to
the machine.
Emergency Rotor Chamber Entry:
In the event of power failure, it may be impossible to unlock
the lid by conventional means. In this case, entry into the rotor
chamber may be made by removing the latch label and using
a pen to manually disengage the locking mechanism (see
photo). Pull the mechanism towards the control panel and
then unlatch and open the lid. If the unit is damaged, contact
your authorised dealer or Woodley Equipment Company.
Calibration and Earth Ground Testing:
It is recommended that the top speed, ground continuity and line leakage be tested every 12
months for continued safe operation. Contact Woodley Equipment Company for further information
or testing availability.
page 16
The Clinispin Horizon 842VES complies with all requirements of CE EN 61000-3-2, -3-3,
EN 61000-4-2, -4-3, -4-4, -4-5, -4-6, -4-11, EN 61010-1, -2-020 and UL standard 3101–2–
20, Can/CSA C22.2 No. 1010.1 , Can/CSA C22.2 No. 1010.2.20

Made in the USA
Old Station Park Buildings
St. John’s Street, Horwich
Bolton BL6 7NY, United Kingdom
Tel: +44 (0) 1204 669033 • Fax: +44 (0) 1204 669034
www.woodleyequipment.com Email: sales@woodleyequipment.com
Replacement Parts:
Part No. Description
7724177 RubberFoot
7751068 Lid Tray Micro-Switch
7723002 Lid Tray Solenoid
7735011 Motor, ½ H.P. Brushless DC
7717041 PCControlBoard
3056001 Step Down Transformer
3056003 Step Down Transformer,
220Vto12V
7751043 CircuitBreaker
7724033 FrontPanelLabel
7760003 Line cord
7714101 Pawl,latch,lid
7714103 Knob,latch,lid
7712263 Lid
7724071 Hinge,friction
7732018 Gasket, cabinet opening
7732019 Gasket, rotor chamber
Contact your authorized dealer or Drucker Diagnostics
to order replacement parts or accessories.
125 mm
Tube Holder
p/n 7713044
13 x 100 mm
Tube Holder
p/n 7717079
Tube
Holder Cap
p/n
7713011
6-Place
Fixed-Angle Rotor
p/n 7786047
75mm
Tube Holder
p/n 7713056 6-Place
Performance
Plus
Horizontal
Rotor
p/n 7786017
6-Place
Horizontal Rotor
p/n 7786061
12-Place combo
([6]75mm, [6]100mm
Fixed-Angle STAT Rotor
p/n 7786003
.5 to 1 ml
tube Adapter
p/n 7713068
13 x 75mm
insert, p/n
7713064
13 x 100 mm
insert, p/n
7713066
1.5 to 2mL
tube Adapter
p/n 7713065
12-Place (75mm
only)
Fixed-Angle STAT
Rotor
p/n 7786001
B
B
C
C
D
D
Table of contents
Other Woodley Laboratory Equipment manuals

Woodley
Woodley Clinispin CT25 User manual

Woodley
Woodley Clinispin MPC User manual

Woodley
Woodley Clinispin HCT II User manual

Woodley
Woodley Clinispin Horizon 6 Flex User manual

Woodley
Woodley Clinispin Horizon 853VES User manual

Woodley
Woodley Clinispin Micro III User manual

Woodley
Woodley Clinispin Micro II User manual

Woodley
Woodley Clinispin HORIZON 12 Flex User manual

Woodley
Woodley Clinispin Vet Combi User manual

Woodley
Woodley Clinispin HCT User manual