manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. YOKOGAWA
  6. •
  7. Test Equipment
  8. •
  9. YOKOGAWA 240413 User manual

YOKOGAWA 240413 User manual

Safety Precautions
■ For safe use, the following symbols are used on the instrument or in this manual:
Danger! Handle with Care.
This symbol indicates that the operator must refer to an explanation in the user’s manual in order to
avoid risk of injury or death of personnel or damage to the tester.
AC Voltage
This symbol indicates the presence of an AC voltage.
WARNING
Indicates a hazard that may result in the loss of life or serious injury of the user unless the described
instruction is abided by.
CAUTION
Indicates a hazard that may result in an injury to the user and/or physical damage to the product or other
equipment unless the described instruction is abided by.
High-voltage Terminal
This symbol indicates a dangerous voltage level (terminals with voltages exceeding 1000 volts must be
so marked). Never touch the terminals.
WARNING
■ To avoid injury, death of personnel, carefully observe and follow the warnings listed below:
1. During the measurement of insulation resistance
• A high voltage is generated in the probe during the measurement of insulation resistance. Do
not touch the measured object, the EARTH or LINE terminals or the GUARD terminal.
2. During the measurement of AC voltage
• There is AC voltage at the GUARD terminal during the measurement of AC voltage. Do not
touch the GUARD terminal.
3. Immediately after the measurement of insulation resistance
• The probes or measured object may remain highly charged. Do not touch them immediately
after measurement.
4. Probe
• Use the probe supplied by Yokogawa Corporation for the instrument concerned.
• Do not use a deteriorated or damaged probe.
• Do not attach/detach the probe to/from the instrument prior to releasing it from the measured object.
5. Protection
• If there are any cracks or other damage in the case because of being dropped or struck, the
instrument may not be safely insulated. Do not use the instrument before any remedial measures
are taken.
6. Object to be measured
• When insulation resistance or resistance is measured, turn off the power to the measured object.
• When theinstrumentisusedin a locationwherean electric currentisflowing,never touch thepowerline.
Use rubber gloves as a safety precaution.
7. Fuse
• To prevent a fire, use the rated fuse to meet the voltage, current and type (Spare Fuse).
• Prior to detaching the cover for replacing the fuse, release the probe from the measured object
and turn off the switch.
• Do not touch the measurement switch during replacement. Otherwise, a high voltage may be produced.
8. Operating Environment
• Do not operate the instrument in a flammable or explosive gas atmosphere.
• Do not operate the instrument, if there is condensation on it.
9. Disassembly
• Do not attempt to disassemble the instrument.
CAUTION
■ To avoid injury of personnel or damage to the instrument, carefully observe and follow
the cautions listed below.
1. Measurement of AC voltage
• Do not apply a voltage over the allowable limits between the terminals.
2. GUARD terminal
• The GUARD terminal is an auxiliary measurement terminal to eliminate a leakage current. Do
not apply a voltage to the terminal.
3. Power supply to the measured object
• Metals and wiring connected to the electric equipment may have a voltage being applied.
Confirm that no voltage is being applied prior to connecting the measurement terminal. The
same precaution applies to the grounding system.
4. Function select switch (except for 240415, 240416)
• Do not change the function select switch setting during measurement.
(Before changing the switch, wait 2 or 3 seconds after a measurement. Immediately after the
measurement, the electric shock charged by probes or measured objects may damage the tester
or the measured objects. This tester has the discharge feature.)
• Do not apply a voltage to the measurement terminals during the measurement of insulation
resistance or continuity test.
1. GENERAL
1.1 Description
Periodically testing of the insulation resistance of electrical machines, power distribution lines,
communication cables and other electric devices, helps to locate faults before they become
serious.
The 2404 Insulation Tester is a hand-driven type insulation resistance tester incorporating a
built-in generator and a constant DC voltage circuit to provide a stable output and give a direct
resistance reading. The insulation resistance is indicated on a moving coil type indicator with a
logarithmic scale. Not only insulation resistance testing and AC voltage measurement, but also
general continuity testing and low resistance measurement are possible using the 500Ωfull
scale low resistance measuring range of the instrument (not on Models 240415, 240416).
The black indicator scale is easy-to-read. This insulation tester is ruggedly built and uses a
shock-and heat-resistant polycarbonate resin case.
1.2 Specifications
Item Specifications
Model
Insulation Resistance
Rated Voltage/Rated
Resistance
Ac Voltage
Measuring
Range
Low Resistance Measuring Range
(continuity test range)
240413 500 V/ 100 MΩ0 to 300V 0 to 500 Ω(central scale mark 50Ω)
240414 500 V/ 1000 MΩ0 to 300 V 0 to 500 Ω(central scale mark 50Ω)
240415 1000 V/ 2000 MΩ0 to 300 V ____(See Note 1)
240416 2000 V/ 5000 MΩ0 to 300 V ____
±5% of indicated value in Effective Measuring Range 1 (See Note 2),
±10% of indicated value in Effective Measuring Range 2 (See Note 2)
Less than 0.7% of scale length at zero and infinity scale marks.
AC Voltage Measurement: ±10%of full scale.
Accuracy: ±10 % of rated voltage at infinity scale mark,
or more than 90 % of rated voltage at center scale mark.
More than 120 rpm.
Polycarbonate resin.
Approx. 184 112  105 mm (7-1 /4 4-3/8 4-1 /8")
Approx. 1.3 kg(2.9 lbs)
User’s Manual
Spare Fuse (2 pcs installed), Carrying Case
Accuracy
Measuring Terminal
Voltage
Rotational Speed of
Crank
Case Material
Dimensions
Weight
Accessories Supplied
at No Extra Cost
• Optional Accessories Available,
Measuring leadwires B9634FA
• Spare Part
Fuse (0.1 A)
Note 1 : The center scalemark is the scale position of
1, 2, 5 or multiples of 10 there of and
corresponds to a value of approximately one-
fiftieth the rated resistance.
Note 2: Effective Measuring Range 1 extends from
1/1000 of the rated resistance to the center
scale mark at 1, 2, 5 or multiples of 10 there
of which most nearly correspond to 1/2 of the
rated resistance.
Effective measuring range 2 extends from
the upper limit of effective measuring range
1 to the rated resistance value. Figure 1-1
IM 2404-10E
1st Edition Jun. 1987
5th Edition Oct. 2007 (KP)
5th Edition : Oct. 2007
All Rights Reserved, Copyright © 1987, Yokogawa M&C Corporation
Thank you for purchasing the YOKOGAWA 2404 Insulation Tester.
To optimize all the functions of the instrument, please read this manual thoroughly before
operating it.
Model: 240413, 240414, 240415, 240416
User’s
Manual Model 2404
Insulation Tester
(Hand-Driven)
IM 2404-10E <2>
2. NAMES AND FUNCTIONS OF COMPONENTS
(4)
(2)
(1)
(6)
(5)
(3)
Figure 2-1
(1) Function Select Switch :
Insulation and AC voltage measurements (MΩand V) and resistance Test (Ω) can be selected
with this switch. (except Models 240415, 240416)
(2) LINE Terminal:
To measure Insulation resistance, AC voltage and resistance use this LINE terminal and the
EARTH terminal (3) below.
(3) EARTH (ground) Terminal
(4) GUARD Terminal
(5) Crank
(6) Carrying Handle
3. OPERATION
3.1. AC Voltage Measurement
(l) Power line AC voltages can be measured using this tester. Always prior to an insulation
resistance measurement, use the tester to check whether the object to be measured is live or not.
(2) Connect measuring lead wires to the LINE and EARTH terminals. Use optional accessory
B9634FA measuring leadwires or equivalent high insulation resistance leadwires.
(3) Place the instrument on a fairly level surface with the crank facing you. Turn the carrying
handle down.
(4) Set the function select switch to V. Prior to connecting an object for an AC voltage
measurement, short-circuit the tester LINE and EARTH lead wires. Turn the crank to check to
see that the indicator shows zero ohms. This confirms that the tester is operating normally.
(5) To make a voltage measurement touch the LINE and EARTH lead wires to the line to be
measured and read the indication on the “AC V” scale as you would read in ordinary voltmeter.
During voltage measurement, do not turn the crank.
3.2 Insulation Resistance Measurement
(1) Read the NOTE below, then connect the object to be measured between the LINE and EARTH
terminals. Set the function select switch to MΩ.
(2) Turn the crank at a speed of more than 120 rpm to obtain a constant voltage output. The pointer
will indicate the measured insulation resistance value.
*Crank rotation may be either clockwise or counterclockwise.
(3) While turning the crank, read the pointer indication.
CAUTION
•When one side of the measured circuit is grounded, connect the earth side to the EARTH
terminal (positive side) of the tester. Normally, this connection gives smaller (conservative)
values of insulation resistance, to be on the safe side. In this case, the lead-wires connected to
the LINE terminal (negative side) should be kept always from earth and other objects.
• When the measured circuit is not grounded, it may be connected to either LINE or EARTH
terminal.
• The GUARD terminal is used only when volume resistance is to be measured without being
influenced by surface leakage resistance. For example, when the insulation test is conducted on
a cable such as shown in Figure 3-1, wind a bare wire around the insulating material and
connect this wire to the GUARD terminal. In this arrangement, the leakage current flowing on
the surface of the insulating material will not pass through the indicator. Thus, only the volume
resistance is measured.
GUARD
LINE
EARTH
Insulating Material
Core Wire Dead-sheathed Wire
Insulation Tester Model 2404
Figure 3-1 Connections
3.3 Low Resistance Measurement
Set the function select switch to Ω. Connect the object at be measured between the LINE and
EARTH leadwires. Turn the Crank at a speed of more than 120 rpm and read the indication on
the “Ω” scale. Low resistance measurement cannot be made using the Models 240415 (1000 V/
2000 MΩ) and 240416 (2000 V/5000 MΩ) insulation testers.
CAUTION
• Be sure not to apply a voltage between the terminals during low resistance measurement.
• If power line AC voltage is applied to the instrument in error while on the low resistance
range, the instrument is protected by an internal fuse. If the fuse is blown, replace it
according of the procedure as described in paragraph 4.2. Insulation resistance and AC
voltage can be measured normally even when the fuse is blown.
• If the fuse is blown, the pointer points to the “*” mark during low resistance measurement.
(See Figure 3-2)
• The 1k mark on the Ωscale is not accurate and may be considered as being “off scale”. Note
that there is no resistance scale mark for Rx = ∞(when test probes are not connected to
anything).
Figure. 3-2
4. MAINTENANCE
4.1 Storage
When storing the Insulation Tester, avoid area which is:
• Very humid.
• Subject to direct sunlight.
• Near high temperature heat sources.
• Subject to strong vibration.
• Very dusty, or contains corrosive gases.
4.2 Fuse Replacement
If the fuse is blown during the low resistance measurement, replace the fuse as follows. Remove
two case fixing screws on the bottom of the instrument. Then the instrument can be
disassembled into two parts as shown in Figure 4-1. Replace the fuse on the printed circuit
board with a spare fuse (0.1 A). Two spare fuses are installed inside the instrument case.
Figure 4-1 Fuse Replacement
■Notice Regarding This Manual
1. The contents of this manual are subject to change without prior notice.
2. If any questions arise or errors are found, or if any information is missing from this manual, please
inform Yokogawa.
3. Yokogawa is by no means liable for damage resulting from the misuse of this product by the user.
4. This manual is intended to explain the functions of this product Yokogawa does not warrant that
the functions will suit a particular purpose of the user.
IM3E-2006.2
Yokogawa Meters & Instruments Corporation
International Sales Dept.
Tachihi Bld. No.2, 6-1-3, Sakaecho, Tachikawa-shi,Tokyo 190-8586 Japan
Phone: 81-42-534-1413, Facsimile: 81-42-534-1426
YOKOGAWA CORPORATION OF AMERICA (U.S.A.)
Phone: 1-770-253-7000 Facsimile: 1-770-251-2088
YOKOGAWA EUROPE B. V. (THE NETHERLANDS)
Phone: 31-334-64-1611 Facsimile: 31-334-64-1610
YOKOGAWA ENGINEERING ASIA PTE. LTD. (SINGAPORE)
Phone: 65-6241-9933 Facsimile: 65-6241-2606
YOKOGAWA AMERICA DO SUL S. A. (BRAZIL)
Phone: 55-11-5681-2400 Facsimile: 55-11-5681-1274
YOKOGAWA MEASURING INSTRUMENTS KOREA CORPORATION (KOREA)
Phone: 82-2-551-0660 to -0664 Facsimile: 82-2-551-0665
YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA)
Phone: 61-2-9805-0699 Facsimile: 61-2-9888-1844
YOKOGAWA INDIA LTD. (INDIA)
Phone: 91-80-4158-6000 Facsimile: 91-80-2852-1441
YOKOGAWA SHANGHAI TRADING CO., LTD. (CHINA)
Phone: 86-21-6880-8107 Facsimile: 86-21-6880-4987
YOKOGAWA MIDDLE EAST E. C. (BAHRAIN)
Phone: 973-358100 Facsimile: 973-336100
LTD. YOKOGAWA ELECTRIC (RUSSIAN FEDERATION)
Phone: 7-095-737-7868 Facsimile: 7-095-737-7869

Other manuals for 240413

1

This manual suits for next models

3

Other YOKOGAWA Test Equipment manuals

YOKOGAWA AQ1200A User manual

YOKOGAWA

YOKOGAWA AQ1200A User manual

YOKOGAWA DL7480 User manual

YOKOGAWA

YOKOGAWA DL7480 User manual

YOKOGAWA 701921 User manual

YOKOGAWA

YOKOGAWA 701921 User manual

YOKOGAWA DLM2000 Series User manual

YOKOGAWA

YOKOGAWA DLM2000 Series User manual

YOKOGAWA HANDY CAL CA51 User manual

YOKOGAWA

YOKOGAWA HANDY CAL CA51 User manual

YOKOGAWA 2560A User manual

YOKOGAWA

YOKOGAWA 2560A User manual

YOKOGAWA MY10 User manual

YOKOGAWA

YOKOGAWA MY10 User manual

YOKOGAWA 279301 User manual

YOKOGAWA

YOKOGAWA 279301 User manual

YOKOGAWA DL7440 User manual

YOKOGAWA

YOKOGAWA DL7440 User manual

YOKOGAWA DL2700 Series User manual

YOKOGAWA

YOKOGAWA DL2700 Series User manual

YOKOGAWA CA330 User manual

YOKOGAWA

YOKOGAWA CA330 User manual

YOKOGAWA SL1400 ScopeCorder LITE User manual

YOKOGAWA

YOKOGAWA SL1400 ScopeCorder LITE User manual

YOKOGAWA CA320 User manual

YOKOGAWA

YOKOGAWA CA320 User manual

YOKOGAWA DL6000 Series User manual

YOKOGAWA

YOKOGAWA DL6000 Series User manual

YOKOGAWA AQ1210 Series User manual

YOKOGAWA

YOKOGAWA AQ1210 Series User manual

YOKOGAWA ScopeCorder DL950 User manual

YOKOGAWA

YOKOGAWA ScopeCorder DL950 User manual

YOKOGAWA ScopeCorder DL350 User manual

YOKOGAWA

YOKOGAWA ScopeCorder DL350 User manual

YOKOGAWA DLM5034 User manual

YOKOGAWA

YOKOGAWA DLM5034 User manual

YOKOGAWA 2553A User manual

YOKOGAWA

YOKOGAWA 2553A User manual

YOKOGAWA DL9040 Series User manual

YOKOGAWA

YOKOGAWA DL9040 Series User manual

YOKOGAWA AQ1200 User manual

YOKOGAWA

YOKOGAWA AQ1200 User manual

YOKOGAWA CA150 User manual

YOKOGAWA

YOKOGAWA CA150 User manual

YOKOGAWA DLM2000 Series User manual

YOKOGAWA

YOKOGAWA DLM2000 Series User manual

YOKOGAWA 720930 User manual

YOKOGAWA

YOKOGAWA 720930 User manual

Popular Test Equipment manuals by other brands

VOLTCRAFT PLUS DSO-1022 Brief instructions

VOLTCRAFT PLUS

VOLTCRAFT PLUS DSO-1022 Brief instructions

piciorgros TETRA TTS-2000A manual

piciorgros

piciorgros TETRA TTS-2000A manual

AVYCON AIVO-54T4K user manual

AVYCON

AVYCON AIVO-54T4K user manual

AlcoMate AL6000 user manual

AlcoMate

AlcoMate AL6000 user manual

Velleman EDU06 user manual

Velleman

Velleman EDU06 user manual

Dräger Testor 2100 Instructions for use

Dräger

Dräger Testor 2100 Instructions for use

Teledyne T3DAQ1-16 user manual

Teledyne

Teledyne T3DAQ1-16 user manual

SW Stahl PROFI Tools 21010L instruction manual

SW Stahl PROFI Tools

SW Stahl PROFI Tools 21010L instruction manual

Klein 3000 Series quick start guide

Klein

Klein 3000 Series quick start guide

Aeroflex 8800 Series Operation manual

Aeroflex

Aeroflex 8800 Series Operation manual

Preston PS-425 Operation and maintenance manual

Preston

Preston PS-425 Operation and maintenance manual

Agilent Technologies InfiniiVision 2000 X-Series Programmer's guide

Agilent Technologies

Agilent Technologies InfiniiVision 2000 X-Series Programmer's guide

Viavi HST-3000 manual

Viavi

Viavi HST-3000 manual

Tektronix TBS1000B Series user manual

Tektronix

Tektronix TBS1000B Series user manual

Transcat CheckMate 400 quick guide

Transcat

Transcat CheckMate 400 quick guide

Teledyne Lecroy 980 user guide

Teledyne Lecroy

Teledyne Lecroy 980 user guide

WilTec 62287 operating instructions

WilTec

WilTec 62287 operating instructions

Teledyne Lecroy PXP500 PCI Express user manual

Teledyne Lecroy

Teledyne Lecroy PXP500 PCI Express user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.