Yuwell YU100 Parts list manual

Enforcement Standard Number: Q/320505YY04
Wechat online service
Stock
symbol :002223http://www.yuwell.com
National Service Tel :400-828 7768
Manufacturer: Suzhou Yuyue Medical
Technology Co., Ltd. Address :9 Jinfeng Road,
High-tech Zone, Suzhou
Tel :0512-65810633
zip code :215163
Trustee Manufacturer: Jiangsu Yuyue Medical
Equipment Co., Ltd. Address: No.1 Yum Road,
Danyang Development Zone
Tel :0511-86900833
zip code :212300
Target date: November 2020
132012-0 A
Household health oxygen generator
Homecare Oxygen Concentrator
Specification Type: YU100
Product Usage and Technical Specification
Before using the product, please read carefully
and understand the date of production of this
specification see certificate or packing

Security
Summary
1
Items Video
Security Summary-------------------------------------------------- 1-3
Product characteristics-------------------------------------------------- 4-5
Installation and use------------------------------------------------- 6-9
Maintenance--------------------------------------------------------10-12
Other considerations-------------------------------------------------13-15
Special message:
In order to prevent power outages or possible failures of oxygen producers, those
in urgent need of oxygen must be equipped with other standby oxygen supply
devices (such as oxygen cylinders, oxygen bags, etc)! The equipment is suitable
for oxygen supplement, regardless of life support or life extension!
I. PREPARATION
Oxygen machine is forbidden to lie upside down!
If the use of power supply voltage instability, beyond the ~220 V±22 V
range, please install a voltage regulator after the use!
Please select safe and qualified socket and wiring board with
safety electrician certification! Non-professional personnel do
not open the chassis!
II. Placement
Oxygen machine should be away from flammable and explosive occasions.
Oxygen machine should be placed in indoor ventilation, and avoid direct
sunlight, around the wall, furniture and other analogues more than 10 cm.
The oxygen generator shall not be placed in the following
environment: near the heat source and the light and dark sources, wet
and unguardedBlock, smoke and pollution, too high or too low
temperature environment.
No sundries and water and oil containers shall be placed on the top of the
oxygen generator.
It is forbidden to place any sundries at the bottom of the oxygen generator
to prevent the inlet and outlet from clogging and causing the temperature to
be too high and causing the shutdown or the oxygen concentration to

Security
Summary
1
decrease.

Security Summary
Security
Summary
2
3
II. In use
Oxygen is a combustible gas, no smoking and away from light and dark
sources to avoid fire hazard.
No oil or grease shall be used in any part of the oxygen generator to avoid
contamination of oxygen and fire.
Oxygen machines shall not be placed or placed in places
where they are easy to drip into water or other liquids.The term
of use is five years. (other than vulnerable parts)
Name and content of harmful
substances in oxygen generator
Harmful
substances
PBDEs
(PBDE)
○
○
○
○
○
This form is prepared in accordance with the provisions of the SJ/T 11364.
○: indicates that the content of the harmful substance in all homogeneous materials of the component is
below the GB/T 26572 limit.
×: indicates that the content of the harmful substance in at least one homogeneous material of the component
exceeds the GB/T 26572 limit.
PBBs
(PBB)
○
○
○
○
○
Chromi
um
hexavale
○
○
○
○
○
Cadmiu
m (Cd)
○
○
○
○
○
Mercury
(Hg)
○
○
○
○
○
Lead
(Pb)
○
○
○
×
○
Part Name
Plastic case
Power cord
Internal piping
Circuit board
assembly
Molecular sieve

4
5
Product characteristics Product
characteristics
I. Scope of application of the product
Oxygen for families and individuals, for health care,
beauty care. Contraindications: Disable for patients with
oxygen poisoning and oxygen allergy
II. Structural characteristics
Oxygen machine is mainly composed of compressor, adsorption tower,
pressure regulating valve, reversing valve, circuit board, shell, filter, remote
control, nasal oxygen pipe and so on.
Full plastic shell, safe and reliable.
With remote control function: can operate the machine
remotely, easy to use. With tired function: display the
total working time through the display screen.
Compressor with thermal protector, better ensure the safety of compressor
and the whole machine.
III. Principle of oxygen production
This machine takes ~220 V power supply as the power source, the air as the
raw material, uses the high quality molecular sieve, through the pressure
swing adsorption separation method (PSA method) at room temperature,
produces the high purity oxygen.
IV. Key technical indicators
1. rated power: AC220V,50Hz
2. rated input power :135 W
3. oxygen flow rate 1 L/min, oxygen concentration 90%
Oxygen flow rate 2~7 L/min, oxygen concentration
50~26%4, noise: front 43 dB (A)(average); whole machine
50 dB (A)
Net 5.:5.5 kg
6. dimensions :22×18×29(cm)
7. altitude: no reduction in oxygen concentration from sea level to 1828
m
Less than 90% efficiency at 4000m

4
5
8. security systems:
Current overload or loose
connection, shutdown; compressor
high temperature, shutdown ;9,
electrical classification: II
equipment
10. Work system: continuous operation
11. Normal working environment:
Temperature
range :10℃~40℃ Relative
humidity range :30~75 per
cent
Atmospheric pressure range :860 hPa ~1060 hPa
Note: when the storage and transportation temperature is less than 5℃, the
equipment should be placed in the normal working temperature environment
for more than four hours before use.
12. Storage and transport environment:
Range of ambient temperature :-
10℃~+55℃ relative humidity:
not exceeding 93 per cent
Note: oxygen generator should be stored in a room without strong sunlight,
non-corrosive gas and well-ventilated, avoid shock and inversion during
transportation.

Installation and use
Installation
and use
6
7
I. Open-box inspection
Before installing and operating the product, the user first checks whether
the appearance of the product is good, and whether the variety and quantity
of the accessories are consistent with the packing list after the use. If there
is any defect, please contact the supplier or manufacturer in time.
II. Outline and Function
Remote control receiving
windows
Flow plus
Switch Flow reduction
Concentration indicator
Flow rate
Fig .3. Machine side
diagram
Filter
Oxygen outlet
Figure 1. Front view of the
machine
Flow plus
Standby
Flow reduction
Figure 4. Remote control schematic
Power
switch
Figure 2. Back view
of the machine
Note: the remote control transmitting window should be in front
of the machine remote control receiving window. (Our remote
control code is 10011421)
Warning: the remote control is equipped with battery, suitable for battery
model CR2025, if you do not use the remote control for a long time, please
take out the battery.
¶

Installation and use
Installation
and use
8
9
III. PREPARATIONS
Plug in the power cord
* Make sure that the power switch of the oxygen generator is off and then
connect the power plug to a safe grounding socket with power output.
No use of extended power cord
Press the power switch to "│", when the front of the machine
""""light, at the same time show tired, oxygen concentration and flow gear,
indicating that the machine function is normal, oxygen machine into normal
operation. After a few seconds, the indicator lights out when tired. Oxygen
machine work, every few seconds to make a "chuckle-" sound, is the normal
sound of air exchange.
VI. Remote control operations
Remote control can operate the machine remotely, its keystroke function is
consistent with the key function of oxygen generator.
Note: remote control operation, in front of the remote control receiving window.
VII. Symbols and meanings of safety requirements in this machine
IV. Oxygen inhalation operations
Press "or" on the front of the machine to the required flow rate,
where the machine front shows the corresponding oxygen concentration.
Connect one end of the oxygen tube to the outlet, the other end (oxygen
head or suction head end) with the user, you can start oxygen inhalation.
V. Shutdown
During use, if you need to temporarily stop oxygen inhalation, you can press
the "key" on the front of the machine to stop and start oxygen supply.
After use, unplug the oxygen pipe or connection pipe from the outlet, turn
off the power switch, then unplug the power plug and cut off the power
supply.
Symb
ol
Meanin
g
Symb
ol
Meanin
g
AC
Note: Access to
random documents
Type II
equipmen
Disconnect
(total power
supply)
Turn on (total
power supply)
Fear
of
rain
Up
upwa
rd
Prohibitio
n of
fireworks
Fragile
items
Stan
dby
Flow
plus
Flow
reductio
n
+
Flow
plus
-
Flow
reductio
n

Maintenance
Maintenance
10
11
Warning: before the maintenance of the oxygen generator, first cut off
the power supply, in order to avoid electric shock, can not disassemble
the housing.
I. Cleaning the enclosure
Clean the outside of the housing at least once a month: first cut off the
power supply, wipe with clean and soft wet cotton cloth or sponge, do not
infiltrate the liquid into the slot of the chassis.
II. Cleaning or replacement of filter felt, filter cotton
The cleaning and replacement of filter felt and cotton is very important to
protect compressor and molecular sieve and prolong the life of oxygen
generator. Please clean and replace it in time.
WARNING: Do not operate the oxygen generator when the felt, filter cotton
is not installed or wetlasting damage to the machine. Filter felt, filter cotton:
usually 100 hours per work cleaning or replacement.
Remove: the filter is located at the bottom of the machine, the filter cover is
pulled out, the filter felt and filter cotton can be removed directly. (Figures V
and VI)
●Filter felt, filter cotton should be based on the actual use time and
environmental impact, determine the actual cleaning time.
●If there is obvious ash on felt or filter cotton, clean or replace immediately,
regardless of the length of service.
III. Cleaning of nasal oxygen tubes
It should be cleaned every three days. The nasal suction head on the oxygen
suction tube should be cleaned after each use, soaked in vinegar for about 5
minutes and cleaned with clean water, or wiped with medical alcohol. It is
recommended that the oxygen tube be replaced every 2 months. Note: The
tube should be kept dry. )
IV. Replacement of fuse tubes
First cut off the power supply, fuse seat in the machine, with a word
screwdriver to pry the top cover, and then the fuse seat upper and lower
cover unscrew, can be removed and replaced. (Figure VII, figure VIII)(Note:
Fuse tube type: T1.6AL250V,Φ5×20)
Top
cover
fuse holder
Figure VII Figure VIII
Filter
Figure V Figure VI
Cleaning: first clean with light detergent, rinse with clean water, must be dry,
Filter felt
Filter cotton

Maintenance
Maintenance
10
11
then installed on the machine.

Maintenance
Other
considerations
12
13
V. GUIDELINES ON PRACTICAL MUSINESS AND
ADMINISTRATION
Fault
phenomenon
Possible causes
Exclusion
methods
Oxygen
generator
fuse of 1) power supply
1) replacement of fuse tube
Source, but
machine
The device
does not
work and the
display
screen does
not show
Insufficient power 2) power
outlet
1) shall not use extension lines.
Move the oxygen generator to
another power outlet or circuit
3) if the oxygen generator is still not working, contact the
equipment supplier or manufacturer
Indicator
Oxygen
generator,
1) oxygen pipeline kink or
blockage
1) piping
But the
oxygen supply
2) filter felt
2) cleaning or replacement of
filter felt
The air
volume is very
high
Less or no
3) contact suppliers or manufacturers if the phenomenon
persists
The oxygen
generator has
a power
supply, but
the machine
does not work
and the
1) compressor overheating
protection, shutdown
1) shut down, wait for the whole
machine to cool before turning on
2) if the oxygen generator is still not working, contact the
equipment supplier or manufacturer
display screen
shows
E1"
Note: if the failure of the oxygen generator can not be eliminated,
please contact the supplier or manufacturer.

Maintenance
Other
considerations
12
13
Air
Air filter
Compress
or
Condense
r
Oxygen
and
Adsorptio
n Tower
Air
exchange
Oxygen
Nitroge
n
Conversio
n valves
Nitroge
n
Pressure
regulatin
Oxygen
outlet
I. Schematic illustration of the gas path
II. Use of electromagnetic environmental guidance
At least 30 cm. between the machine and other household appliances in
use to avoid interference by strong electromagnetic environment
Non-professional personnel shall not disassemble the oxygen
machine case, professional maintenance personnel if need
product electrical schematic, please contact the
manufacturer or after-sales service point.
IV. After-sales service 7768)
This product from the date of sale, within 7 days of non-human factor
performance failure, users can choose to return, exchange or repair.
Non-human performance failure within 8-15 days, user

Other considerations
Other
considerations
14
15
Optional replacement or maintenance. Under normal circumstances, this
product machine warranty for one year, compressor warranty for three years,
users can purchase machine invoices, warranty card to enjoy free
maintenance. The warranty period shall be calculated from the date of the
invoice issued, and if the customer fails to provide a valid invoice, it shall
be calculated by 30 days according to the product fuselage number or the
date of departure. If the product exceeds the warranty period or does not
belong to the warranty scope, the company will carry out charge
maintenance.
The following conditions are not covered by warranty :(1) vulnerable and
consumable goods :(1) nasal oxygen pipe, fuse pipe, filter felt, filter
cotton ;(2) the user causes water, medicine and other liquid to enter the
machine can not work properly ;(3) the whole machine water or rain ;(4) the
failure caused by unauthorized disassembly, repair, modification of the
product ;(5) failure caused by careless drop in the process of use and
handling ;(6) failure caused by failure to operate in accordance with the
correct method of the instructions ;(7) damage caused by unforeseen natural
disasters such as fire, earthquake, flood, etc.
VI. Disposal of wastes and residues
The disposal of waste and residue shall conform to the
corresponding national laws and regulations. Note: circuit and
appearance changes due to machine improvement are subject to no
prior notice!
V. Packing list
1. oxygen generator :1
2. product usage and technical specification :1
3. nasal oxygen tube :1
4. filter felt :1
5. filter cotton :1 piece

Other considerations
Other
considerations
14
15
6. fuse tube (T1.6AL250V,Φ5×20):2
7. warranty card, certificate :1
8. remote :1
Table of contents
Other Yuwell Medical Equipment manuals

Yuwell
Yuwell 8F-3A User manual

Yuwell
Yuwell YX3 Series User manual

Yuwell
Yuwell 405A User manual

Yuwell
Yuwell 7F-5A User manual

Yuwell
Yuwell 7E-B5 User manual

Yuwell
Yuwell BreathCare PAP User manual

Yuwell
Yuwell 7F-3 User manual

Yuwell
Yuwell 7B-1 User manual

Yuwell
Yuwell YX102 User manual

Yuwell
Yuwell YX310 User manual
Popular Medical Equipment manuals by other brands

Huntleigh
Huntleigh Sonicaid FM800 Series Reference manual

Stereo Optical
Stereo Optical Optec PLUS C86 quick start guide

St. Jude Medical
St. Jude Medical Tendril MRI LPA1200M user manual

Welch Allyn
Welch Allyn Connex Quick reference guide

VitalConnect
VitalConnect VistaPoint 3.0 QuickStart and Instructions for Use

BEMIS
BEMIS BEMISSENTINEL 5008 quick start guide