Yuwell YX102 User manual

7306240-2A
Manufacturer:JIANGSU YUYUE MEDICAL EQUIPMENT & SUPPLY CO.,LTD.
Manufacturer Address:No.1 Baisheng Road Development Zone, Danyang, Jiangsu 212300 CHINA
Due to the limited size of the label, the font is too small, please put it at a suitable location for viewing.
All specifications and product configurations are subject to change without notification.
YUWELL®
Finger Pulse Oximeter
YX102/YX103
Please read the user manual carefully before using!
(The picture is for reference only, please refer to the actual
product.)

Warning And Caution
Safety:
1.Warning, Do not modify this equipment without authorization of manufacturer.
2.Warning, Keep away from the wet medical equipment such as drip or other similar liquid simulation as
far as possible.
3.Warning, Check the sensor site every half an hour to ensure adequate blood circulation, intact skin,
and appropriate sensor location. Otherwise, it may cause skin damage, compressive necrosis, or
inaccurate measurement readings.
4. Warning, With the increasing number of radio devices or other noise sources from electric
equipment in health care departments , our product may be interrupted when working because of their
interference. The closer the distance between each other is or stronger the signal is, the more serious
the interference will be.
The electromagnetic interference sources in health care departments may include:
(1).Electronic surgical instruments communications equipment (2).Mobile Phones
(3).Automotive two-way wireless communications equipment (4).Electronic apparatus
(5).High-definition television
In this interference, the measurement values may deviate, or the device may not work. When
interfered, the product may produce abnormal phenomenon: unstable reading values, outages or other
functions of error. If such a case, the use of the site should be checked to identify interference and the
elimination of the following measures:
(1)Shut down the equipment in the vicinity and then re-open in order to identify interference equipment;
(2)To change the direction or location of the interference equipment;
(3)To increase the distance between the product and interference sources.
5.Warning,Do not put the battery close to the fire or into the fire to avoid the battery explosion. Do not
use the battery when it leaks or molds.
6.Warning, Device conforms to the requirement of RoHS directive.
7.Warning, Device application component materials are certified for biological compatibility.
8. Warning, Do not leave the oximeter unattended around children or infants. Small items such as the
battery door, battery,and lanyard may become choking hazards if swallowed. Infants or children may
be entangled in the lanyard, thus causing strangulation.
9.Warning, Do not stare at the light (the infrared is invisible) which emitted from the oximeter,which is
harmful to the eyes.
10.Warning, Do not use the oximeter for purposes other than its intended use. Do not place the
oximeter on edema or fragile tissues.
11.Warning, Do not use the oximeter close to a permanent magnet.
12.Warning, Do not use the finger pulse oximeter in an MRI or CT environment.
13.Warning, Do not use the finger pulse oximeter during defibrillation and electrosurgey.
14.Warning, Do not use the finger pulse oximeter in the presence of flammable anesthetics or other
flammable substances, oxygen-enriched environments, or nitrous oxide to avoid the risk of explosion.
15.Caution,When using this product, you only need to fix it on your fingers. Excessive pressure on your
fingers may cause skin damage.
16.Caution, The maximum skin surface temperature is below 41℃(106℉) when measured in a 35℃
(95℉)environment, which has been verified by measuring the skin surface temperature via a Finger
Pulse Oximeter under the reasonable worst conditions.
17.Caution, Please pay attention to product storage to prevent damage caused by pets,pests or
children.
Performance:
18.Warning, Do not self-diagnose or self-medicate based on measurement results.It must be used in
conjunction with other methods of assessing clinical signs and symptoms.
19.Warning, Do not use the the finger pulse oximeter in situations where alarms are required. The
device has no Alarm System.
20.Warning, The effect of sensor and electrode degradation or electrode loosening may reduce the
performance of the measurement or cause other problems.
21. Warning,The thickness of the little finger may be smaller than the minimum size range for which this
product is designed, so we recommend using index finger, middle finger and ring finger instead of small
finger.
22. Warning, Do not use the oximeter if it shows signs of damage or is suspected of being damaged.
Inaccurate or no readings may result from internal damage.
23. Warning, Avoid the following to reduce the risk of SpO2 inaccuracy:
(1)Improper placement of the oximeter.
(2)Significant levels of dysfunctional hemoglobin (such as carbonyl hemoglobin or methemoglobin);
(3)Blood vessels contain dyes (such as indigo green or methylene blue).
(4)External paints and substances (e.g. nail polish, nail polish, glitter, etc.).
(5)The user is wearing a high-pressure cuff when measuring.
(6)Avoid placing the oximeter on any arm with an arterial cannula or blood pressure cuff.
(7)Elevated bilirubin levels. Severe anemia. (8)Venous congestion. Venous pulsation.
(9)Arterial perfusion levels are extremely low. (10)Excessive physical activity.
(11)During cardiac arrhythmia.
24.Warning, Keep the oximeter away from electrical equipment that emits radio frequencies to minimize
radio interference. RF may result in inaccurate or inaccurate readings.
25.Warning, Properly apply and avoid using the oximeter in strong ambient light sources, fluorescent
lights, infrared heat lamps, and direct sunlight to minimize interference that may result in unreadable or
inaccurate readings.
26.Caution, The display may not read clearly when exposed to direct sunlight or bright light.
27.Caution,Do not use a functional tester to evaluate the accuracy of the Finger Pulse Oximeter. The
functional tester shall only be used to check whether a unit is working properly.
28.Caution,SpO2 is empirically calibrated in healthy adult volunteers with normal levels of carboxyhe-
moglobin(COHb) and methemoglobin(MetHb).
29.Caution, Please replace the battery when a low battery remind appears.
Maintenance and others:
30. Warning, Please do not repair and maintain the equipment during use.
1 2

General Description
Oxyhemoglobin saturation is percentage of Oxyhemoglobin (O2Hb) capacity, compounded with oxygen, by
all combinativable haemoglobin (Hb) Oxyhemoglobin (O2Hb) capacity in blood. In other words, it is
consistence of Oxyhemoglobin in blood. It is a very important ecological parameter for Respiratory
Circulation System. Many respiratory diseases can result in oxyhemoglobin saturation being lowered in
human blood. Moreover, the following factors can also lead to problems in oxygen supply, so that human
oxyhemoglobin saturation might be reduced: Automatic Organic Regulation Malfunction caused by
Anesthesia, intensive Postoperative Trauma, hurts resulted in by some medical examination and etc. In the
situation, illnesses, such as dizziness, asthenia, emesis and etc, might happen to patients and even
endanger the patient's life. Therefore, it is very important to know oxyhemoglobin saturation of patient timely
in clinical medical aspects. So that doctors can find problems in time.
The YUWELL® Finger Pulse Oximeter features in small volume, low power consumption, convenient
operation and portable. It is only necessary for patient to put one of his fingers into a fingertip photoelectric
sensor for measurement, and then the screen will display the measured value of pulse oxygen saturation
and pulse rate. It has been proved in clinical experiments that it features in rather high precise and repeatability.
Measurement Principle
Principle of the Oximeter is as follows: An experience formula of data process is established taking use
of Lambert Beer law according to Spectrum Absorption Characteristics of deoxyhaemoglobin(HHb) and
Oxyhemoglobin (O2Hb ) in glow and near-infrared zones. Operation principle of the instrument is
Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse
Scanning and Recording Technology, so that two beams
of different wavelength of light (red light and infrared light)
can be focused onto human nail tip through perspective
clamp finger-type sensor. Then measured signal can be
obtained by a photosensitive element, information
acquired through which will be shown display through
process in electronic circuits and microprocessor.
Red and Infrared- ray
Emission Tube
Red and Infrared-ray
Receipt Tube
Figure 1 Work Principle
3 4
31.Warning,When discarding components (including the batteries) or this product, follow local
regulations to avoid contamination.
32.Warning, The device has been calibrated before leaving the factory. Except replacing batteries,
devices do not require routine maintenance and calibration, etc. Daily measure ten times, ten minutes
every time, devices can be used for five years.
33.Warning,Please use Two AAA 1.5V alkaline batteries, using other batteries may damage the
device. Please use Please replace two batteries at the same time to avoid mixing batteries with
different power levels.
34. Caution, This product can be operated by the patient, or by others to measure the patient's PR and
SpO2. The maintenance, operation and maintenance methods are the same.
This product contains batteries and recyclable electronic waste.To protect the environment,do not
dispose of it in the household waste, but take it to appropriate local collection points.
Medical device
Symbol Definitions Symbol Definitions
Serial Number
Caution
Refer to instructions manual(Background: Blue; Symbol: White)
Protection from ingress of particulates
than ≥12.5mm. Dripping water falling
within 15° of vertical will not have a
harmful effect on the pulse oximeter
per IEC 60529
IP22
Recyclable Date of manufacture
Batteries and electronic instruments
must be disposed of in accordance
with the locally applicable
regulations , not with domestic waste
Type BF applied part
Manufacturer
The device has no Alarm System
Use-by date
Temperature limit
Humidity limitation
Atmospheric pressure limitation
This way up
Fragile, handle with care
Rx only For prescription use
% SpO2Pulse rate (bpm:1/min)
Keep dry
MR MR Unsafe items should not enter
the MRI scanner room.
Equipment Symbols And Explain
PR bpm Pulse rate (bpm:1/min)

5 6
8.Measurement accuracy- in the absence of movement:
Note:The accuracy(ARMS) is calculated by the measurement values after a statistical distribution;
compared to the reference device in a control study, approximately 2-thirds of the values were at(over
or below) the accuracy(ARMS) value.
Pulse rate: 25bpm~250bpm, accuracy(ARMS): ±1% or ±1bpm(larger)
9.Operation mode: Continuous operation.
�
65
70
75
80
85
90
95
100
0 15 30 45 60 75 90 105120135150
Time�/ s
Response�Time
8�Beat�
Average
SpO2/ %
Figure 4
10.Device response time.(See Figure 4)
11.Peak wavelengths and light output power:
Emission wavelength range 600nm-1000nm, radiation intensity is less than
15mW/sr (20mA).
Information of wavelength range may be of especial use to clinical doctors.
12.Description of the effect on displayed and transmitted SpO2 and Pulse rate
data value:
1)data averaging and other signal processing.
2)the data update period:≤3 pulse rate cycles, less than 30s.
Note: Data processing and update will not affect pulse rate and SpO2.
13.The pulse waveform has been normalized, the measurement value is the best when the waveform is
smooth and stable.
14. Internally Powered ME Equipment
15.TYPE BF APPLIED PARTS
16.Degrees of protection provided by enclosures (IP code):IP22.
Technical Description
1. Determination of Oxygen Saturation Accuracy
The claimed oxygen saturation is supported by coverage of the entire range of clinical research
measurements.
2.Data Collection
In the clinical test process, data points are recorded with comparable density in the entire claimed
range.
3. Data Analysis
For each claimed range, the oxygen saturation accuracy of the Pulse Oximeter should be represented
in the form of mean root square of the difference between the measured values of oxygen saturation
and the reference value. The formula is as follows:
Arms: accuracy.
n: test sample quantity.
SpO2i : measured value of pulse oxygen saturation during the first measurement using the finger pulse
Oximeter.
SRi: reference value of pulse oxygen saturation during the ith measurement using the carbon-monox-
ide-blood-gas analyzer.
4. Characteristics of Population under Clinical Research
The group of healthy adult volunteers for the study consisted of 12 subjects - 4 women and 8 men, The
ages ranged from 18 to 50 years. The subject weights ranged from 48 to 90kg. The subject Height
ranged from 160 to 182cm. The Skin tones included in the study were as follows: 3 subjects with dark
Indication for Use
Intended use: The YUWELL® Finger Pulse Oximeter is a kind of non-invasive, non-sterile, reusable, spot
checking device which can measure and display SpO2 and pulse rate through finger. It is intended for adults
and children (weight >30 kg) and is expected for home and hospital inspection. The device is not for continu-
ous monitoring, using during motion or using with low perfusion.
Contraindications: None.
Technical Parameters
1.Display Type: LED
SpO2 Display range: 0%~100%.
Pulse Rate Display range: 25bpm~250bpm.
2.Power: Two AAA 1.5V alkaline batteries.
3.Working Current: Less than 40mA at rated voltage 3V.
4.Anti-interference ability of ambient light: Deviation in blood oxygen content is less than ± 1% when
measured under indoor nature light / existing lighting and measured in the dark room.
5.The product will automatically shut down when there is no signal detected for about eight seconds.
6.Dimension(YX102/YX103):60mm*38mm*35mm(LWH), Weight:38g approximately (without batteries).
7.Working Environments: Ambient temperature: 5℃~40℃; Relative humidity: ≤80%; Atmospheric
pressure: 860hPa~1060hPa.
Display
Symbol of oxygen saturation Low Battery
Symbol of pulse rate
Pulse bar Pulse rate
Oxygen
saturation
Figure 2 Wave signal undetected
Figure 3
Signal undetected
Signal inadequacy
(eg: 1.Finger is out.
2.Device fault )
Range
90% to 100%
80% to 90%
70% to 80%
<70%
ARMS
±2%
No definition
pigmentation, 5 subject with very light pigmentation.
The remaining subjects with light (medium) skin
tones of Asian origins.
5.Graphical Plot of SaO2 versus error (SpO2 - SaO2)
6.The technology used in Finger Pulse Oximeter has
been verified with accuracy when there is no motion
via human blood studies on healthy volunteers of
both male and female with light to dark pigmented
skin in induced hypoxia studies in the range of
70%-100% SpO2 against a laboratory co-oximeter.
7.The technology used in Finger Pulse Oximeter has
been verified with the pulse rate accuracy of
25-250bpm range in the bench top test against
SpO2 ACCURACY (ARMS)

Battery Installation
1.Thread thinner end of the lanyard through the hanging hole.
2.Thread thicker end of the lanyard through the threaded end
before pulling it tightly.
3.Install as the figures show. (See Figure )
Lanyard Installation
1.Pull the battery cover horizontally.
2.Put two AAA batteries into battery cassette in right polarities and push the
cover back.
Notes: Battery polarities must be correct. Otherwise, damage might occur to device.
Please put or remove batteries in right order, or it will damage the device bracket.
Please remove the battery if the oximeter is not used for long time.
3.Install as the figures show. (See Figure)
Remove the battery from the product if it is not required for extended
Product Operation Steps
1.Install two AAA batteries into battery cassette before closing the cover.
2.Nip the clamp as diagram to the largest.(See Figure )
3.Plug one finger into rubber hole of the Oximeter (it is best to plug the finger
thoroughly) before releasing the clamp.
4.Do not tremble while the oximeter is working. It's better that the whole body be
in still status.
Note: For normal use after long interruptions, refer to the product operation steps.
5.Read corresponding data from display screen.
7 8
1. Determination of Oxygen Saturation Accuracy
The claimed oxygen saturation is supported by coverage of the entire range of clinical research
measurements.
2.Data Collection
In the clinical test process, data points are recorded with comparable density in the entire claimed
range.
3. Data Analysis
For each claimed range, the oxygen saturation accuracy of the Pulse Oximeter should be represented
in the form of mean root square of the difference between the measured values of oxygen saturation
and the reference value. The formula is as follows:
Arms: accuracy.
n: test sample quantity.
SpO2i : measured value of pulse oxygen saturation during the first measurement using the finger pulse
Oximeter.
SRi: reference value of pulse oxygen saturation during the ith measurement using the carbon-monox-
ide-blood-gas analyzer.
4. Characteristics of Population under Clinical Research
The group of healthy adult volunteers for the study consisted of 12 subjects - 4 women and 8 men, The
ages ranged from 18 to 50 years. The subject weights ranged from 48 to 90kg. The subject Height
ranged from 160 to 182cm. The Skin tones included in the study were as follows: 3 subjects with dark
Product Operation Scope
The YUWELL® Finger Pulse Oximeter is designed for fingers(not thumb) between 0.3 and 0.9
inch (0.8-2.3cm) thick. And the finger shall be inserted into the sensor position which is in the
middle of the device.
The Finger Pulse Oximeter is NOT design for newborns and infants.
Product Properties
1.Operation of the product is simple and convenient.
2.The product is small in size, light in weight and portable.
3.The product can operate continuously for 17 hours with two brand new AAA batteries. (The operation
time may vary due to the different performance of batteries)
4.Low voltage prompt will appear on the display when the battery voltage is lower than the minimum
value of normal working voltage range.
5.The product will automatically shut down when there is no signal detected for about eight seconds.
periods of time in order to avoid damage to the oximeter resulting from a leaking battery.
Do not use the new batteries with the old ones. Alkaline battery of long service is recommended,
and do not use rechargeable batteries.
pigmentation, 5 subject with very light pigmentation.
The remaining subjects with light (medium) skin
tones of Asian origins.
5.Graphical Plot of SaO2 versus error (SpO2 - SaO2)
6.The technology used in Finger Pulse Oximeter has
been verified with accuracy when there is no motion
via human blood studies on healthy volunteers of
both male and female with light to dark pigmented
skin in induced hypoxia studies in the range of
70%-100% SpO2 against a laboratory co-oximeter.
7.The technology used in Finger Pulse Oximeter has
been verified with the pulse rate accuracy of
25-250bpm range in the bench top test against

Maintenance And Storage
Under normal conditions there is no need for special protection and maintenance when using, please
pay attention to the following points:
Caution: Using oximeter in required environment.
Caution: Avoid direct sunlight.
Caution: Avoid extreme infrared radiation or ultraviolet radiation.
Caution: Avoid organic solvent vapors, dust, and corrosive gas.
Transportation and storage conditions:
Temperature range: -20℃~+55℃
Relative humidity: ≤93%, no condensation
Atmosphere pressure: 500hPa~1060hPa
It is recommended that the product should be kept in a dry environment anytime. Moisture might
affect its lifetime and even damage the product.
Cleaning And Disinfecting
This product is a reusable non-sterile device. Please clean and disinfect according to the following
methods.
Warning:
1.Never immerse or soak the oximeter.
2.We recommend cleaning and disinfecting the oximeter before or after each use, or in accordance
with the policies established by the hospital, to avoid long-term damage to the oximeter and avoid
cross-infection.
3.Never use cleaning agents/disinfectants other than the recommended.
4.The sensor component is not cleaned and disinfected during testing.
5.Avoid the use of metals such as steel wire brush or polishing agent abrasive material which will
damage the oximeter panel .
Cleaning
The recommended cleaning agents include: water
1.Shut down the finger pulse oximeter and remove the battery.
2.Clean the oximeter with cotton or soft cloth moistened with water.
3.After cleaning, wipe off the water with a soft cloth.
4.Allow the oximeter to air dry.
The most commonly used hospital cleaning and non-corrosive liquid detergent can be used to clean
the oximeter. Pay attention to diluting cleaning detergent before use, following the manufacturer's
instructions.
Avoid the use of ethanol-based, amino-or acetone-based cleaning agent.
Oximeter shell should be maintained from dust pollution, use a soft cloth or lint-free cleaning agent
with the sponge to wipe. Make sure no liquid will enter into the equipment.
Disinfecting
The recommended disinfectants include: ethanol 70%, isopropanol 70%
1.Shut down the finger pulse oximeter and remove the batteries.
2.Clean the oximeter as instructed above.
3.Disinfect the oximeter with cotton or soft cloth moistened with one of the recommended disinfectants.
4.After disinfection, be sure to wipe off the disinfectant left on the oximeter with a soft cloth
moistened with water.
5.Allow the oximeter to air dry.
Possible Cases And Solutions
Warning:
Caution: Oximeter cover can only be opened by a professional maintenance staff. No internal
parts require opening by end users.
If you are not sure about the measurement precision, please use other methods to check
patient's pulse, to determine whether oximeter works.
Note: Do not splash, dump any liquid into the oximeter and attachments, switch and
connections, which may damage the oximeter.
Problems Possible reason Solution
3. Nail polish or paste manicure
1. Do not put finger correctly
2. Try some more times, if you can make sure about no problem exiting
in the product, please go to a hospital timely for exact diagnosis
3. Remove the nail polish or discharge manicure when measuring.
1. Try again
SpO2 or PR
can not be
shown
normally
2. Try not to move
2. Finger is trembling or patient is in
movement status.
1. Please replace batteries
2. Please reinstall the batteries
SpO2 or PR
is shown
unstably
1. Retry by plugging the finger
1. Finger might not be plugged deep
enough
2. Not used according to recommended
steps
3. The finger pulse oximeter might be damaged
The finger
pulse
oximeter can
not power on
1. Power of batteries might be inadequate
or not be there at all
2. Batteries might be installed incorrectly
3. Please contact with local customer service center
1. Normal
2. Replace the batteries
1. The product is automatically powered off when
no signal is detected longer than 8 seconds
2. Battery Low
Indication
lamp are
suddenly off
9 10

11 12
The EMC environment for this product is the home healthcare environment and professional
healthcare facility environment.
The essential performance of this product is the accuracy of SpO2 and pulse rate (SpO2 Accuracy:
±2% in the range of 70%-100% of SpO2, No definition for SpO2 under 70%;Pulse rate:
25bpm~250bpm, accuracy: ±1% or ±1bpm(larger)). When used directly near strong electromag-
netic interference (for example: near mobile phones, microwave ovens, etc.), it may be temporarily
inaccurate. If so, please keep the product away from interfering devices.
The finger pulse oximeter may exhibit temporary degradation of performance (e.g.deviation from the
performance indicated in the instructions for use during IMMUNITY testing) and the finger pulse
oximeter can recover from any interference within 30 seconds without affecting basic safety or
performance without operator intervention.The following degradations, if associated with BASIC
SAFETY or ESSENTIAL PERFORMANCE shall not be allowed:
- component failures;
- changes in programmable parameters or settings;
- reset to default settings;
- change of operating.
During measurement, The use of this equipment adjacent to or stacked with other equipment should
be avoided because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
During measurement, portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part
of the finger pulse oximeter, including cables specified by the manufacturer. Otherwise, degradation
of the performance of this equipment could result.
Electromagnetic Interference
Guidance and manufacture’s declaration – Radiated Emission
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Compliance
Group 1
Class B
Table 1 - For all ME EQUIPMENT and ME SYSTEMS
RF emissions CISPR 11
Emissions test Result
3 m
40dB(μV/m) 30MHz-230MHz
47dB(μV/m) 230MHz-1000MHz
IEC60601 test level
3 m
40dB(μV/m) 30MHz-230MHz
47dB(μV/m) 230MHz-1000MHz
Compliance level
Pass
Guidance and manufacturer's declaration – Radiated Emission
Table 2 - For all ME EQUIPMENT and ME SYSTEMS
Electrostatic discharge
(ESD) IEC 61000-4-2
±2 , ±4, ±8, ±15
Air Discharge
±8 Contact
Discharge
±2 , ±4, ±8, ±15
Air Discharge
±8 Contact
Discharge
IMMUNITY test ResultIEC60601 test level Compliance level
Pass
Guidance and manufacturer's declaration – Electrostatic Discharge
Table 3 - For all ME EQUIPMENT and ME SYSTEMS
IMMUNITY test ResultIEC60601 test level Compliance level
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
Pass
Guidance and manufacture's declaration – Radio-frequency Electromagnetic Field
Table 4 - For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING RS
IMMUNITY test ResultIEC60601 test level Compliance level
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
30 A/m
50Hz or 60Hz
30 A/m
50Hz and 60Hz Pass
Guidance and manufacturer's declaration – Power-frequency Magnetic Fields
Table 5 - For all ME EQUIPMENT and ME SYSTEMS
Test Reference Standard
CISPR 11
CISPR 11
IEC/EN 61000-3-2
IEC/EN 61000-3-3
IEC/EN 61000-4-2
IEC/EN 61000-4-3
IEC/EN 61000-4-4
IEC/EN 61000-4-5
IEC/EN 61000-4-6
IEC/EN 61000-4-8
IEC/EN 61000-4-11
Test Item
Emission Measurements
Conducted Emission
Radiated Emission
Harmonic Current Emissions
Voltage Fluctuations and Flicker
Immunity Measurements
Electrostatic Discharge
Radio-frequency Electromagnetic Field
Fast transients, Common Mode
Surges
Radio-frequency Common Mode
Power-frequency Magnetic Fields
Voltage Dips and Interruptions
Result(Pass or Fail or N/A)
N/A
Pass
N/A
N/A
Pass
Pass
N/A
N/A
N/A
Pass
N/A
Test Summary

13 14
Table 6 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment
Accessories
Warranty Card
Test
frequency
(MHz)
Maximum
power
(W)
Distance
(m)
IMMUNITY
Test LEVEL
(V/m)
Banda)
(MHz) Servicea) Modulationb)
704-787
385
800-960
380-390 TETRA 400
LTE Band
13,17
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
710
745
780
810
870
930
Pulse
modulationb)
18 Hz
Pulse
modulationb)
217 Hz
Pulse
modulationb)
18 Hz
27
450 430-470 GMRS 460,
FRS 460
FMc)
±5 kHz
deviation
1 kHz sine
28
9
28
5240
5500
5785
NOTE:If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is
permitted by IEC 61000-4-3.
a) For some services,only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave singal.
c) As an alternative to FM modulation,50% pulse modulation at 18 Hz may be used because while it
dose not represent actual modulation, it would be worst case.
1700-1990
2400-2570
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Bluetooth,
WLAN,
802.11 b/g/n
RFID 2450,
LTE Band 7
1720
1845
1970
2450
Pulse
modulationb)
217 Hz
Pulse
modulationb)
217 Hz
28
1.8
2
0.2
2
2
2
0.3
0.3
0.3
0.3
0.3
0.3 28
5100-5800 WLAN 802.11
a/n
Pulse
modulationb)
217 Hz
0.2 0.3 9
Lanyard: 1 pc
AAA batteries: 2 pcs
User Manual, Warranty card: 1 pc
After unpacking, check the items according to the accessories list, and check whether the oximeter is
mechanically damaged. If you find any problems, please contact the local customer service center
immediately.
During the warranty service, if you need to provide circuit diagrams, necessary materials, and if there
are any problems with the maintenance of electrical circuits, please contact the manufacturer.
The device and accessories are provided non-sterile.
Thank you very much for using our products.
Product name: Finger pulse oximeter
Model: YX102/YX103
S/N:
MFG.DATE:
JIANGSU YUYUE MEDICAL EQUIPMENT & SUPPLY CO., LTD.
Manufacturer Address: No.1 Baisheng Road Development Zone, Danyang, Jiangsu 212300 CHINA
www.yuwell.com
Please reserve the warranty card carefully.
This manual suits for next models
1
Table of contents
Other Yuwell Medical Equipment manuals

Yuwell
Yuwell 7F-5A User manual

Yuwell
Yuwell BreathCare PAP User manual

Yuwell
Yuwell 7F-3 User manual

Yuwell
Yuwell YH-550 User manual

Yuwell
Yuwell YU100 Parts list manual

Yuwell
Yuwell YX310 User manual

Yuwell
Yuwell BreathCare PAP User manual

Yuwell
Yuwell 7E-B5 User manual

Yuwell
Yuwell 7E-A User manual

Yuwell
Yuwell 7B-1 User manual
Popular Medical Equipment manuals by other brands

Drive DeVilbiss Healthcare
Drive DeVilbiss Healthcare iGo 2 manual

Tiens
Tiens Multifunctional Headcare Apparatus user manual

GE Medical Systems
GE Medical Systems LOGIQ ?200 Service manual

NRS Healthcare
NRS Healthcare N94561 User instructions

Kaz
Kaz PERSONAL STEAM INHALER use and care manual

Tivic Health
Tivic Health ClearUP user guide