16 Press Right installation
2 Special functions
This chapter describes the special functions of the inputs and outputs.
2.1 Configure automatic job selection
The Press Right supports automatic job selection by combining inputs or via a bar-code scanner.
The selection via inputs can be useful for the automatic recognition of the piece or of the mounted
tool.
To enable the automatic selection of jobs, select the menu TOOLS > SETUP > JOBS SELECTION.
2.1.1 Selection via inputs
With selection by inputs, the instrument changes its job when the state of the inputs changes. The
instrument ignores the job change command when the START signal is active.
The job number is derived from the binary combination of the inputs.
2.2 Connecting an external reset
When a reject is detected, the tool blocks preventing the execution of other pieces. To unlock the
instrument it is necessary to reset the deviation. To reset the deviation, press the RESET key or
activate a RESET input. It is to avoid locking the instrument with reject by disabling the keyboard
reset from the general options menu in the instrument configuration.
2.3 Additional controls
Up to six additional controls can be connected to the instrument.
Three of these are used as consent to start the press, the other three are used to determine the
good or bad outcome of the job or as other consents to start.
The first three are called ENABLE, the second three are called TEST SIGNAL. It is possible to assign
a name to each additional control from the instrument configuration menu.
It is possible to establish in each job whether to exclude the additional checks.
The TEST SIGNAL 1 control can also be associated with an output signal useful for activating the
control itself.
To use the additional controls you need to configure the instrument by indicating which inputs the
signals are connected to.
After configuring the inputs, it is possible to assign an identification name to each of the six signals.
The enabling signals are checked when the START signal is activated: if an enabling signal is
missing, the instrument displays the message absence of enabling signal.
The test signals condition the good or rejected outcome: if a test signal is missing, the piece is
rejected.
The moment in which the presence or absence of each test input is checked is individually
configurable.
The time when the test signals are checked can be chosen in the job options. Note that only the
input tests 1 all possibilities.
at the start the signal is checked when the starting input is activated
upon arrest the signal is monitored when the stop is commanded