Anyload EB100 User manual

Precision Scales
(EB100 / EB200)
Operation Manual
(V0116)
North America Toll Free: 1-855-ANYLOAD (269 5623)
Fax: +1-866-612-9088
http://www.anyload.com

1 | A N Y L O A D E B 1 0 0 / E B 2 0 0 M a n u a l ( V 0 1 1 6 )
Contents
1. Safety Precautions......................................................................................2
2. Features………………………………………………………………………….2
3. Parts Description………………………………………………………………..3
4. Preparation/Installation…………………………………………………………4
5. Weighing…………………………………………………………………………4
5.1 Before Weighing……………………………………………………….4
5.2 Error Messages………………………………………………………...4
5.3 Weighing Procedures………………………………………………….5
5.4 Changing Units…………………………………………………………5
5.5 Start Weighing………………………………………………………….5
6. Function Settings……………………………………………………………….5
6.1 Setting BAU Value……………………………………………………...5
6.2 Setting Parity Mode…………………………………………………….5
6.3 Setting Print Mode……………………………………………………...5
6.4 Setting Unit Selection…………………………………………………..5
6.5 Setting ZERO track……………………………………………………..5
7. Counting Procedures…………………………………………………………..6
8. Calibration……………………………………………………………………….6
6.1 When to calibrate……………………………………………………….6
6.2 How to calibrate…………………………………………………………6
9. RS232 Data Interface…………………………………………………………..7
10. Specifications……………………………………………………………………8
6.1 EB100…..………………………………………………………………..8
6.2 EB200…..………………………………………………………………..8
11 Error Code……………………………………………………………………….8

2 | A N Y L O A D E B 1 0 0 / E B 2 0 0 M a n u a l ( V 0 1 1 6 )
1. SAFETY PRECAUTIONS
All safety messages are identified by the following the words "WARNING" and "CAUTION".
These words have the following meanings,
△WARNING
Important information to alert you to a situation that might
cause serious injury and damage to your property if
instructions are not followed.
△CAUTION
Important information that tells how to prevent damage to
equipment.
When using the scale/balance, the following safety precautions should always be followed.
△WARNING
Use only the AC adaptor with the scale/balance. Other adaptor may cause damage. (AC adaptor
is optional)
△CAUTION
Avoid placing the scale/balance in direct sun light which may cause discoloration or malfunctions.
Do not mix the type of batteries. Replace all batteries at the same time.
If the scale/balance is not to be used for a long period of time, remove all batteries from the
battery compartment to avoid leakage which may cause damage to the instrument.
Avoid overloading that could cause damage to the scale/balance.
Do not drop water into the scale/balance that is not water-resistant. It causes damage, if inside of
the scale/balance is wet.
Matter charged static electricity could influence the weighing. Discharge the static electricity. As
example, the method is to use the electrification prevention spray, to spray on to both sides of
weighing platform.
2. FEATURES
●Auto calibration ● Auto shut off (optional)
●Auto zero tracking ● Auto backlight (optional)
● Low battery indication ● g/ct/oz/tola conversion(optional)
●Large LCD Display ● Counting function (optional)
●Net weight/stability indication
●1/120000&1/200000 division available

3 | A N Y L O A D E B 1 0 0 / E B 2 0 0 M a n u a l ( V 0 1 1 6 )
3. PARTS DESCRIPTION
RIGHT SIDE
TOP
BOTTOM
DISPLAY
AC adaptor
Socket
Display
ON/OFF key
turns the scale
power on/off.
Zero key
set display to zero
PRINT key (optional)
TARE PCS
key
UNIT key
Changes weighing units, g, ct, oz, tola……
Switch key
Battery compartment
Battery compartment cover
Batteries
Use 6 AA batteries
Low battery indicator
Negative value indicator
TARE indicator
weighing unit
STABLE indicator when
the reading is stable.
Print indicator
ZERO indicator

4 | A N Y L O A D E B 1 0 0 / E B 2 0 0 M a n u a l ( V 0 1 1 6 )
4. PREPARATION/INSTALLATION
4.1 Installing batteries/Connecting the AC adaptor
Batteries
Battery Remove the battery compartment
compartment cover and insert six batteries(R6P/LR6
/AA size) into the battery compartment,
taking extreme care that the polarities
plus and minus is observed.
R6P/LR6/AA size
Dry batteries.
Battery compartment cover (batteries are not included)
AC adapter (optional)
Plug the AC adapter to the AC
AC adaptor adapter socket on the side.
socket The AC input requirement could
be 110,120,220,230 or 240Volts
(50/60Hz) depending on the area
where used, so please verify that
the adaptor is correct
4.2 Setting up the scale/balance
△
!CAUTION
Avoid placing the scale/balance in direct sunlight, that may cause discoloration or malfunctions.
Place your scale/balance on a firm weighing table so that the scale/balance is level. (The
scale/balance will not perform accurately when it is not level.)
Place the scale/balance on the firm surface that is flat and level for an accurate weighing.
5. WEIGHING
5.1 Before weighing
Whenever possible, please allow the balance to warm up for 10 minutes after first turning to
power on, so that the balance will function properly and accurately.
5.2 Error messages
EEEE: Overload
: Low battery
Calibration may be required before weighing.
Read "CALIBRATION" first and if necessary, calibrate your scale/balance for accurate
weighing.

5 | A N Y L O A D E B 1 0 0 / E B 2 0 0 M a n u a l ( V 0 1 1 6 )
5.3 Weighing procedures
5.3.1 Press [ON/OFF] to turn on the balance.
When power is turned on, all display segments appear on LCD for a few seconds and finally
"0" will appear on the display.
5.3.2 Changing Units
Press [UNIT] to select a weighing unit, eg. "g"、" ct " 、" oz "、" tola " …….
Once the unit has been selected, the selected unit will be displayed to the weight value.
5.3.3 Start weighing
If you do not use a container for weighing
Verify the reading is "0". If not, press [ZERO] to display "0".
Place objects on the weighing platform to weigh.
When the reading becomes stable, the stable indicator is displayed.
If you use a container for weighing
Place an empty container on the platform.
Wait for the stability indicator to be displayed and press [ZERO].
Place the objects to be weighed in the container.
When the reading becomes stable, the stable indicator is displayed.
6. FUNCTION SETTINGS
Press [ON/OFF] to turn on the scale, the display will show flash digits and then --------,
6.1 Setting BAU Value
Press [PRINT], the display will show b xxx BAU value (press [UNIT] to select among 1200,
2400, 4800, 9600), our default setting is 9600.
6.2 Setting Parity Mode
Press [ZERO] to enter Parity mode, the display will show P xxx PAR, press [UNIT] to select
odd, even, none (odd stands for 7 data bits with odd parity, even stands for 7 data bits with
even parity, none stands for 8 data bit without parity)our default setting is none.
6.3 Setting Print Mode
Press [ZERO] to enter Print mode, the display will show n xxx, press [UNIT] to select CON,
OFF, KEY, STB (CON stands for continuous print, OFF stands for serial data output
disabled, KEY stands for demand mode by press [PRINT], STB stands for automatic print
when scale is stable, our default setting is STB. Press [ZERO] again to return to weighing
mode.
6.4 Setting Unit Selection
Press [UNIT] to enter unit selecting function, the display will show ON X (X stands for the
units: g,oz, lb,twt…tola..), press [UNIT] to select ON (activate the selected unit) or OFF
( inactivate the selected unit), press [ZERO] again to return to weighing mode.
6.5 Setting ZERO track
Press [TARE] key to enter ZERO track function setting, display shows ZRD x.x,press [UNIT]
key to select ZERO track value(0.0d-0.5d-1.0d-1.5d-2.0d), press [TARE] key to enter

6 | A N Y L O A D E B 1 0 0 / E B 2 0 0 M a n u a l ( V 0 1 1 6 )
ZERO track function setting display shows ZRT x.x,press [UNIT] key to select ZERO track
time(05. to 4.0 seconds, press [ZERO] key to enter ZERO setting,display shows RZD
XXX, press [UNIT] key to select ZERO value(0.8d-1.8d-2.8d-3.8d), press [ZERO] key to
enter FIL setting, press [UNIT] key to select filter range amend (0-1-2-3-4-5-6-7), “2” is
default setting, press [ZERO] key to select zero range when turn on the scale, press [UNIT]
key to select percent (PZR 2, PZR 5, PZR 10. PZR 20), press [ZERO] key to set zero range,
press [UNIT] key to select percent.
7. Counting procedures
7.1 Press [ON/OFF] to turn on the scale.
Wait for "0" to appear on the display.
7.2 Start the Count Procedure
If necessary, press [ZERO] key to set the display to “0”.
7.3 Press the [PCS] key to put the scale in PCS mode, the display will show P=X X, press UNIT
to select XX value (10 20 50 100). Place a given number of samples of an item on the pan
(the Sample Size should be either 10、20、50 or 100 pieces). The weight of these samples
will show on the display.
7.4 Press the [PCS] key, the display will show “XX pcs”, then the scale will recall the sample size
you selected and show the starting sample size on the display (you can now remove the
samples if you want to return the scale to 0).
7.5 Press the [PCS] key to return to the weighing mode.
8. CALIBRATION
8.1 When to calibrate
Calibration may be required when it is initially installed, if the scale/balance is moved a
substantial distance. This is necessary because the weight of a mass in one location is not
necessarily the same in another location. Also, with time and use, mechanical deviations may
occur.
8.2 How to calibrate
8.2.1 Before entering calibration mode, there is a switch key on the bottom of the scale. Turn
on the switch key first, press [ON/OFF] to turn the power on for 1minutes.
8.2.2 Enter calibration mode
Press [ON/OFF] to turn the power on, the display will show flash digits and then “-------
-”, Press [ZERO], the display will show “CAL” , Press [ZERO] again , the display will
show “X0000 CAL”, Xis flash digit. Press [UNIT] key the flash digit will move to right.
Press [PCS] key to increase the flash digit. (X is the calibration weight which can
be set according to your requirement).
8.2.3 Calibration by the weight
Select the calibration weight. Then place the calibration weight on platform. Then
press [ZERO] key,the display returns to the weight value. Now the calibration is
complete.
8.2.4Press [ON/OFF], turn off the power, press [ON/OFF] to turn on the power again. Place
the weight on the platform to make sure whether weighing is correct, if not, repeat steps
2-4.
8.2.5 Return to weighing mode
Press [ON/OFF] to turn the power off. Press [ON/OFF] again to turn on the power and
the scale/balance returns to the weighing mode.

7 | A N Y L O A D E B 1 0 0 / E B 2 0 0 M a n u a l ( V 0 1 1 6 )
9. RS232 Data Interface
9.1 Connector: DB9
BALANCE PC
2 2
3 3
5 5
9.2 Transmissions Settings
Mode: Simplex Asynchronous Serial Baud Rate: 9600
Data Bit: 8 Parity Bit: None
Stop Bit: 1 Data Format: ASCII
9.3 Transmission Information Format:20 Byte ,blank=20H
1~2
3
4~13
14~18
19
20
‘W: ’
P
DATA
UNIT
CR
LF
W: Start of Data Transmission=57H+3AH
P : Polarity ‘+’ = 2BH = Positive ‘-’ = 2DH =Negative
DATA: ‘ 12.345’= 20H+20H+20H+20H+31H+32H+2EH+33H+34H+35H
UNIT: ‘g’ =67H+20H+20H+20H+20H
‘oz ’ =6FH+7AH+20H+20H+20H
‘lb’ =6CH+62H+20H+20H+20H
‘dwt’ =64H+77H+74H+20H+20H
‘ozt ’ =6FH+7AH+74H+20H+20H
‘ct ’ =63H+74H+20H+20H+20H
‘tl.T’ =74H+6CH+2EH+54H+20H
‘tl.H’ =74H+6CH+2EH+48H+20H
‘tl.J ’ =74H+6CH+2EH+4AH+20H
‘GN’ =47H+4EH+20H+20H+20H
‘dr’ =64H+72H+20H+20H+20H
‘MM’ =4DH+4DH+20H+20H+20H
‘tola’ =74H+6FH+6CH+61H+20H
‘gsm’ =67H+73H+6DH+20H+20H
‘T/A/R’ =54H+2FH+41H+2FH+52H
‘T/M/R’ =54H+2FH+4DH+2FH+52H
‘pcs’ =50H+43H+53H+20H+20H
CR: = 0DH
LF: = 0AH

8 | A N Y L O A D E B 1 0 0 / E B 2 0 0 M a n u a l ( V 0 1 1 6 )
9. SPECIFICATIONS
9.1 EB100
Model number
Capacity
Division
EB -100
3000g
0.05g/0.1g
Platter
170 x 180mm
Net/gross weight
1350g/1500g
Package
Standard carton: 34×22×14(cm3)
6 Units in one box: 47.5×38.5×45(cm3)
Operating Temperature
0-40℃(32-104℉)
Power source
6xAA batteries or AC/DC
Adapter 10~12V/150mA (optional)
9.2 EB200
Model number
Capacity
Division
EB -200
600g
0.01g/0.005g
Platter
φ115mm
Net/gross weight
3800g/4700g
Package
Standard carton: 34×22×14(cm3)
6 Units in one box: 47.5×38.5×45(cm3)
Operating Temperature
0-40℃(32-104℉)
Power source
recharge or AC/DC
Adapter 10~12V/150mA (optional)
10. Error Code
Error
Code
Reason
Solution
ERR-O
Over load
Make sure the load is within the full
scale capacity
ERR-2
1).exceed Zero tracing
range when power on.
2). Forgot to place
platform when linearity
calibration
1). tack off the weight when power
on
2). place platform and then
recalibration
ERR-5
inner code unstable
when power on
1). load cell was being toughed
2). E+ power supply unstable
3). there are capacitance creepage
of C1AD~C6AD
ERR-C
forgot to input
calibration weight when
calibrating
CAL~000000~ press [UNIT] and
[PCS] (TARE) key to input
calibration weight
ERR-P
counting setting error
place sample when counting

9 | A N Y L O A D E B 1 0 0 / E B 2 0 0 M a n u a l ( V 0 1 1 6 )
setting
ERR-L
forgot to place test
weight when calibrating
place test weight when calibrating
ERE-E
there are something
wrong of memorizer
check U24c02

10 | A N Y L O A D E B 1 0 0 / E B 2 0 0 M a n u a l ( V 0 1 1 6 )
http://www.anyload.com
North America Toll Free: 1-855-ANYLOAD (269 5623)
Fax: +1 866 612 9088
This manual suits for next models
1
Table of contents
Other Anyload Scale manuals

Anyload
Anyload EB300 User manual

Anyload
Anyload EC100 User manual

Anyload
Anyload OCS-S Series User manual

Anyload
Anyload NPS Series Manual

Anyload
Anyload EC200 User manual

Anyload
Anyload ES-HB Series User manual

Anyload
Anyload OCS Series User manual

Anyload
Anyload EC100 User manual

Anyload
Anyload T15 Series User manual

Anyload
Anyload OCS-T User manual
Popular Scale manuals by other brands

Rice Lake
Rice Lake TuffSeal JB4ES installation manual

anko
anko 42616450 Instructions for use

KERN
KERN KXP_V20 Series installation instructions

Steinberg Systems
Steinberg Systems SBS-PF-100A1 user guide

Trisa electronics
Trisa electronics 1859 Instructions for use

HoMedics
HoMedics HealthStation IB-SC520 instruction manual