APPA 36II User manual

AC/DC CLAMP MULTIMETER
INSTRUCTION MANUAL
APPA 36Ⅱ/36RⅡ

1
INTRODUCTION
1-1 Unpacking and Inspection
Upon removing your new Digital Clamp Multimeter from its packing, you should have the following items:
1. Digital Clamp Multimeter.
2. Test lead set (one black, one red).
3. Carrying case.
4. Instruction manual.
5. Battery.
1-2 Meter Safety
Terms as Marked on Equipment.
ATTENTION — Refer to manual.
DOUBLE INSULATION — Protection Class Ⅱ.
DANGER — Risk of electric shock.

2
Symbols in this Manual.
This symbol indicates where cautionary or other information is found in the manual.
=Battery
1-3 Front Panel
Refer to Figure 1 and the following numbered steps to familiarize yourself with the meter’s front panel controls
and connectors.
1. Digital Display —The digital display has a 3 3/4 digit LCD readout (maximum reading 3999) plus decimal
point , AC, DC, AUTO , HOLD , MAX , <and unit annunciators.
2. Input Terminal — The black test leads is always connected to the “COM” input terminal and red test lead is
always connected to the “V-Ω” input terminal when measuring ACV or DCV or RESISTOR or CONTINUITY.
3. Function Switch —This slide switch is used to select V, V, A , A , Ω, <function.

3
4. HOLD/MAX( ) — This button has two modes one is data hold (A) and the other is maximum hold (B).
(A) Turning the rotary switch to power on, the button will work in data hold mode. Press to toggle data hold
ON/OFF.
(B) Pressing and hold the HOLD/MAX () button then turn the rotary switch to power on, this button will work in
maximum hold mode. Press to restart MAX recording. Press this switch more than 2 seconds to exit the
maximum hold mode. Maximum hold mode doesn't work in continuity mode.
5. ZERO A/ Ω/ <Button
Zero AButton — For DC current measurement, using the “ZERO A“ button to zero the reading .
Ω/ <Button — Pressing this button to measure resistor or continuity either.
Auto power off disable mode — Refer to 2-2 Electrical Specifications item (9) Auto Power Off Disable.
6. Trigger — Press the lever to open the transformer jaws. When the pressure on the lever is released , the jaws
will close again.
7. Hand Guard —Designed to protect user for safety.
8. Transformer Jaws —Designed to pick up the AC/DC current flowing through the conductor.

4
Figure 1

5
SPECIFICATIONS
2-1 General Specifications
Display : 3 3/4 Digital Liquid Crystal Display (LCD) with a maximum reading of 3999.
Polarity : Automatic polarity indicated.
Over range Indication : “OL” indicated.
Low Battery Indication : ““ is displayed when the battery voltage drops below the operating voltage.
Measuring Rate : 2 times/second, normal.
Position Error : ±1% of reading.
Type of Sensing : Hall effect sensing for AC and DC current.
Shook Proof : 4 feet drops.
Power Requirement : Standard 9V battery, NEDA 1604, JIS 006P, IEC6F22.
Battery Life : Alkaline 150 hours.
Maximum Jaw Opening : 45mm.

6
Max/Conductor Size : 35 mm diameter.
Temperature Coefficient : 0.2 x ( spec.Acc’y ) / ℃< 18℃, > 28℃.
Size : 82 mm (W) x 208 mm (L) x 41 mm (H) .
Weight : 360 grams (including battery)
Accessories : Test leads, battery, manual and carrying case.
2-2 Environmental Conditions
Indoor use.
Maximum Altitude : 2000 Meter.
Installation Category : IEC 1010-1 CAT. III 600V
Pollution Degree : 2
Operating Temperature : 0℃ to 30℃ (≦80% RH), 30℃to 40℃(≦ 75% RH), 40℃to 50℃(≦45% RH).
Storage Temperature : -20℃to 60℃.

7
2-2 Electrical Specifications
Accuracy is ±(% reading + number of digits) at 23℃±5℃at less than 80% R.H.
(1) AC Voltage : Auto-ranging
Range Resolution Accuracy Over voltage
protection
400.0mV 100μV Unspecified
600V rms
4.000V 1mV ±(1.5% reading + 5digits ) 40Hz ~ 300Hz
40.00V 10mV
400.0V 100mV
600V 1V
±(1.5% reading + 5digits ) 40Hz ~ 500Hz
Input Impedance : ≧ 9MΩ// less than 100pF.

8
AC Conversion Type : 36Ⅱ:
AC conversion are average sensing rms indication calibrated to the rms value of a sine
wave input.
36RⅡ:
AC conversions are ac-coupled, true rms responding, calibrated to the rms value of a
sine wave input . Accuracies are given for sine wave at full scale and non-sine wave
below half scale. For non-sine wave add the following Crest Factor corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.

9
(2) DC Voltage : Auto-ranging
Range Resolution Accuracy Over voltage
protection
400.0mV 100μV
600V rms
4.000V 1mV
400.0V 100mV
600V 1V
±(0.7% reading + 2 digits)
40.00V 10mV
Input Impedance : ≧ 9MΩ.

10
(3) Resistance Auto-ranging
Range Resolution Accuracy Overload
Protection
400.0Ω100mΩ ±(1.2% reading + 6 digits) *1
600V rms
4.000KΩ1Ω
±(0.9% reading +3 digits) *2
40.00KΩ10Ω
400.0KΩ100Ω
4.000MΩ1KΩ
40.00MΩ10KΩ ±(2.5% reading + 5 digits) *1 *3
±(1.2% reading + 3 digits) *2

11
* 1: The reading maybe rolling ≦6 digits when the reading is close to full scale.
* 2: The reading maybe rolling ≦3 digits when the reading is close to full scale.
* 3: The response time is approximate 20 seconds.
* : Put a low resistor in the input terminal before turn the rotary function switch to resistor and continuity function
maybe cause a buzzer sound.
(4) Continuity
Built-in buzzer sound when measured resistance is less than 50Ωand sound off when measured resistance is
more than > 300Ω.
Between 50Ωto 300Ωthe buzzer maybe sound or off either.

12
(5) DCA : Auto-ranging
Range Resolution Accuracy Over voltage protection
0 ~ 40.0A 0.1A
600A rms
40.0A ~ 400.0A 0.1A
400A ~ 600A 1A ±(1.9% reading + 10 digits)
±(1.5% reading + 10 digits)
±(1.5% reading + 7 digits) *
1. * For 200.0A ~ 400.0A add 1% to accuracy.
2. Temperature Coefficient : 0.2 x (Spec.Acc'y) /℃< 21 ℃or > 26℃.
3. Operating Temperature : 0℃to 30℃(≦80%RH) , 30℃to 40℃(≦75%RH)

13
(6) ACA : Auto-ranging
Range Resolution Accuracy Frequency
Response
Overload
Protection
0 ~ 40.0A 0.1A ±(1.9% reading + 7 digits)
50Hz ~ 60Hz
600A r.m.s.
40.0A ~ 400.0A 0.1A ±(1.9% reading + 5 digits) **
400A ~ 600A 1A ±(1.9% reading + 10 digits)
0 ~ 40.0A 0.1A ±(2.5% reading + 7 digits)
61Hz ~ 400Hz
40A ~ 400.0A 0.1A ±(2.5% reading + 5 digits) *
400A ~ 600A 1A ±(2.5% reading + 10 digits)
1. * For 200.0A ~ 400.0A add 1.1% to accuracy.
2. ** 200.0A ~ 400.0A add 1.6% to accuracy for 36RⅡ.
200.0A ~ 400.0A add 1.1% to accuracy for 36Ⅱ.

14
3. Temperature Coefficient : 0.2 x (Spec.Acc'y) /℃< 21 ℃or > 26℃.
4. Operating Temperature : 0℃to 30℃(≦80%RH) , 30℃to 40℃(≦75%RH)
AC Conversion Type : 36Ⅱ:
AC conversion are average sensing rms indication calibrated to the rms value of a sine
wave input.
36RⅡ:
AC conversions are ac-coupled, true rms responding, calibrated to the rms value of a
sine wave input . Accuracies are given for sine wave at full scale and non-sine wave
below half scale. For non-sine wave add the following Crest Factor corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.

15
(7) Maximum Hold
7-1 In maximum hold function the accuracy is changed as following.
Original Accuracy + 10 digitals/ change steps of range .
For example :
At First , the maximum hold reading on display is 100.0mV on 400.0mV range. If a voltage vibration changes the
maximum hold reading to 120.0V. The change steps of range are 3 steps ( 400.0mV to 4.000V to 40.00V to
400.0V) so the accuracy is needed to add 3 change steps of range x 10 digitals / change steps of range = 30 digits.
7-2 On maximum hold mode the accuracy of resistance is specified from 400.0Ω to 400.0kΩrange only.
(8) Auto Power Off
The meter will automatically shut itself off after approximately 30 minutes after power on.
(9) Auto Power Off Disable :
Pressing and hold the ZERO AΩ<button then turn the rotary switch from power off to Ω/<position quickly
(less than one second) that will disable the auto power off.

16
OPERATION
This instrument has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for
Electronic Measuring Apparatus and has been supplied in a safe condition. This instruction manual contains some
Information and warnings which have to be followed by the user to ensure safe operation and to retain the instru-
ment in safe condition.
3-1 Preparation and Caution before Measurement
1. If the meter is used near equipment that generates electro-magnetic interference, the display may be unstable
or indicate incorrect measurement values.
2. Make sure that the battery is properly connected.
3. The instrument should only be operated between 0℃~ 50℃and at less than 80% R.H. except current
function is operated between 0℃~ 40℃.
4. Do not use or store this instrument in a high temperature or high humidity environment and do not store the
unit in direct sunlight.

17
5. Do not replace battery with power on condition.
6. If the unit is not to be used for a long period of time , remove the battery.
7. Do not forget to turn off after use.
8. Maximum rated voltage to earth for voltage measurement terminals is 600V CAT.III
THIS INSTRUMENT MUST NOT BE USED ON UNINSULATED CONDUCTORS AT A
VOLTAGE GREATER THAN 600V ac/dc.

18
3-2 AC/DC Current Measurement
1. Turning the rotary switch at “A” or “A“ position.
2. Open clamp by pressing trigger on left side of meter.
3. Position clamp around wire or conductor and release clamp trigger smoothly , do not release quickly, make sure
that the clamp is entirely closed. Position the conductors at the center of the clamp jaws for accurate measure-
ment . The clamp must be positioned around only one conductors of a circuit .
If the clamp is placed around two or more current-carrying conductors , the meter reading will be FALSE.
4. For DC measurement , the reading is positive value when the current flows from the upper side to the lower
side of the instrument as Figure 2.
5. For accuracy measurement on DCA mode follow the procedure below
(i) Position clamp around wire or conductor and release clamp trigger smoothly, do not release quickly,
make sure that the wire or conductor is at the center of the clamp.
(ii) Remove the wire or conductor from the clamp.
(iii) Press the ZERO Abutton to zero the reading, effective in DCA mode only.
(iv) Repeat (i).
(v) The positive reading measurement can get better accuracy in DCA mode.