APPA 506 User manual

506/506B
User Manual / 使用說明書 / 使用说明书
ユーザーマニュアル
Руководство пользователя
Digital Multimeter (Bluetooth)
藍芽萬用電表
蓝芽万用电表
デジタルマルチメーター(Bluetooth)
Мультиметр цифровой (Bluetooth)
RU
EN
TC
SC
JP
LIMITE D
WARRAN TY
YEAR S

506/506B EN
Read First
Safety Information
WARNING
Understand and follow operating instructions carefully.
Use the meter only as specied in this manual; otherwise,the
protection provided by the meter may be impaired.
This identies hazardous conditions and actions that could
cause BODILY HARM or DEATH. To avoid possible danger,
follow below guidelines.
• Use the meter only as specied in this manual or the
protection by the meter might be impaired.
• Never operate the meter with the cover removed or the
case open.
• To avoid false readings that can lead to electric shock and
injury, replace battery as soon as low battery indicator.
• Use caution with voltages above 30VAC rms, 42VAC peak,
or ±30VDC. These voltages pose a shock hazard.
• When using test leads or probes, keep your ngers behind
the nger guards.
• Remove test lead from meter before opening the battery
door or meter case.
• Always use proper terminals, switch position, and range for
measurements.
• Do not apply more than the rated voltage, as marked on
meter, between terminals or between any terminal and
earth ground.
• Do not use the High Frequency Rejection (Low Pass
Filter) option to verify the presence of hazardous voltages.
Voltages greater than what is indicated may be present.
First, make a voltage measurement without the lter to
detect the possible presence of hazardous voltage. Then
select the lter function.
• To avoid possible electric shock or personal injury, never
attempt an in-circuit current measurement where the open
circuit potential to earth is greater than 1000V.
• Replace the fuse as soon as the indicator (FUSE) appears.
• Only replace the blown fuse with the proper rating as
specied in this manual.
• Do not use the meter around explosive gas, vapor or dust.
• To reduce the risk of re or electric shock do not expose
this product to rain or moisture.
1

506/506B EN
• Do not attempt a current measurement when the open
voltage is above the fuse protection rating. Suspected
open voltage can be checked with voltage function.
• Never attempt a voltage measurement with the test lead
inserted into the A input terminal.
• Disconnect the test leads from the test points before
changing the position of the function rotary switch.
• Disconnect circuit power and discharge all high-voltage
capacitors before testing resistance, continuity, diodes, or
capacitance.
• Always use proper terminals, switch position, and range for
measurements.
• Do not use the LoZ mode to measure voltages in circuits
that could be damaged by this mode’s low impedance.
• Replace the fuse as soon as the indicator (FUSE) appears.
• Never connect a source of voltage with the function rotary
switch in resistance, diode, continuity, and capacitance
position.
• Never set the meter in current function to measure the
voltage of a power supply circuit in equipment that could
result in damage the meter and the equipment under test.
This identies conditions and actions that could DAMAGE
the meter or equipment under test. To avoid possible
damage, follow below guidelines.
CAUTION
2

506/506B EN
Hazardous voltage
Risk of danger. Important information. See manual
AC (Alternating Current)
DC (Direct Current)
Bluetooth
Double insulated
Low battery
Fuse
Earth ground
Conforms to European Union directives
Do not discard this product or throw away
IEC Overvoltage Category
CAT III equipment is designed to protect against transients
in equipment in xed equipment installations, such as
distribution panels, feeders and short branch circuits, and
lighting systems in large buildings.
CAT IV equipment is designed to protect against transients
from the primary supply level, such as an electricity meter or
an overhead or underground utility service
Electrical Symbols
CAT III
CAT IV
Fuse broke. Replace fuse as soon.
Meter error. Have meter serviced.AC (Alternating Current)
Test probe alert. Displayed when the test probes are in the A
or mA terminal and the selected rotary switch position does
not correspond to the terminal being used.
Error Message
ProbE
FUSE
Er
3

506/506B EN
The Meter Description
Front panel illustration :
1. Auto backlight sense point
2. 40,000 count dual display
3. Push buttons
4. Rotary switch
5. Input terminal for voltage, frequency, resistance,
continuity,diode, capacitance and temperature measurements.
6. Return terminal for all measurements.
7. Input terminal for 0 to 400mA current measurements.
8. Input terminal for 0 to 10A current measurements.
TRUE RMS
600V
CAT.III
1000V
MAX
FUSED
400mA
MAX
10A
FUSED
CAT.IV
COM
mA V
A
Hz
AC+DC
A
AC+DC
V
AC+DC
mV
Hz
mV
LoZ
Auto V
V
Hz
OFF
P HOLDCANCEL
°C/°F
A HOLDENTERHFRRANGE
Hz
1
2
3
4
56
7
8
4

506/506B EN
Power On Options
Making Basic Measurements
When turn the power on, press the function button to
execute the below options.
• When connecting the test leads to the DUT (Device Under
Test), connect the common test lead before connecting the
live lead. When removing the test leads, remove the test
live lead before removing the common test lead.
The following sections describe how to take measurements
with the meter.
Push Buttons
Function
(Yellow)
RANGE
HFR
A-HOLD
ENTER
CANCEL
°C / °F
P-HOLD
Bluetooth
Select measurement function.
Select measurement range. Press > 1 sec to enter auto
range mode.
Enable/Disable the High Frequency Reject mode in the AC
measurements.
Enable the Peak-Hold mode in the AC or DC measurements.
In this mode, press button to select Peak-Hold MAX or MIN.
Press > 1 sec to disable the Peak-Hold mode.
Enable/Disable the Bluetooth mode.
Enable/Disable the Auto-Hold mode.
Enter menu function in pointer positon.
Cancel current menu function.
Select degrees Celsius or degrees Fahrenheit.
Display the rmware version.
Display all LCD segments.
Power On Options
ENTER
A HOLD
WARNING
5

506/506B EN
Measuring AC and DC Voltage
Make dB measurement
This meter have true rms readings, which are accurate
for distorted sine waves and other waveforms (with no dc
oset) such as square waves, triangle waves, and staircase
waves.
The ranges of measuring voltage are 40mV, 400mV, 4V,
40V, 400V and 1000V. To select the mV range, turn the
rotary switch to mV position.
The meter is capable of displaying voltage as a dB value, either
relative to 1 milliwatt (dBm), a reference voltage of 1 volt (dB).
A dBm measurement must use a reference impedance (600Ω)
to calculate a dB value based on 1 milliwatt. A dB measurement
uses a 1 volt reference voltage to compare the present
measurement against. Dene as below :
For best accuracy when measuring the DCmV, touch the
probe tips together and read the DC oset. If necessary, you
can use the relative (Δ) mode to automatically subtract this
value.
AC Voltage DC Voltage
dBm = 20log ( )
Vrms
dBm = 20log ( )
Vrms
1V
6

506/506B EN
Measuring Voltage in LoZ Mode
To use dB or dBm function, turn the rotary switch to ACV or
ACmV position. Then move the blink cursor of menu to
dB or dBm position, and press the ENTER button to enter
function.
Press the CANCEL button to exit function.
Caution
• Do not use the LoZ mode to measure voltages in circuits
that could be damaged by this mode’s low impedance.
To eliminate ghost voltages, the meter’s LoZ mode presents
a low impedance across the leads to obtain a more accurate
measurement. The ranges of measuring LoZ voltage
are 400V and 1000V. In this mode, meter will automatic
measure input signal which is AC or DC and determine
range.
To use the LoZ mode, turn the rotary switch to LoZ position.
Measuring AC and DC Current
• Always use proper terminals, switch position, and range for
measurements.
• To avoid possible electric shock or personal injury, never
attempt an in-circuit current measurement where the open
circuit potential to earth is greater than 1000V.
• Only replace the blown fuse with the proper rating as
specied in this manual.
WARNING
7

506/506B EN
Caution
• Replace the fuse as soon as the indicator (FUSE) appears.
To measure current, you must break the circuit under test,
thenplace the meter in series with the circuit.
The ranges of measuring current are 40mA, 400mA, 4A and
10A. AC current is displayed as an rms value. Insert the
black lead into the COM terminal. For currents less than 400
mA, insert the red lead into the mA terminal. For currents
above 400 mA, insert the red lead into the A terminal.
You can press the function (yellow) button to select current
measurement function.
Measuring Frequency
The meter measures the frequency of a voltage or current
signal by counting the number of times the signal crosses a
threshold level each second.
This function only can be operated in AC voltage and current
measurements. The ranges of measuring frequency are
400Hz,4kHz, 40kHz and 100kHz.
If a reading shows as 0 Hz or is unstable, the input signal
may be below or near the trigger level. The detail of
frequency trigger level refer to the electrical specications.
To use the frequency function, press function (yellow) button
to select measurement function.
8

506/506B EN
Make Peak-Hold Measurement
The Peak-Hold mode records wave peak maximum and
minimum input values. Response time of Peak-Hold is 10us.
To use the Peak-Hold mode, press the P-HOLD button to
enable the Peak-Hold mode in the AC or DC measurements.
In this mode, press P-HOLD button to select peak MAX or
MIN value. Press > 1 sec to disable the Peak-Hold mode.
For square wave, use the peak-hold mode in the DC
measurements.
Make High Frequency Rejection Measurement
• Do not use the High Frequency Rejection (Low Pass Filter)
option to verify the presence of hazardous voltages.
Voltages greater than what is indicated may be present.
First, make a voltage measurement without the lter to
detect the possible presence of hazardous voltage. Then
select the lter function.
The High Frequency Rejection mode equip a low pass lter
in the AC measurements. The cut-o frequency (-3dB point)
of low pass lter is 800Hz.
To use the HFR mode, press the HFR button to equip a low
pass lter in the AC measurements.
WARNING
9

506/506B EN
Make AC+DC Measurement
When input signal is ac and dc combinations: AC over DC or
DC over AC, the meter is capable of displaying one AC+DC
(rms) value combined. Dene as below :
To use the AC+DC function, turn rotary switch stop in DCV,
DCmV or A position, then press function (yellow) button to
select measurement function.
( AC + DC )V rms = ACV² + DCV²
( AC + DC )Arms = ACA² + DCA²
Measuring Resistance
CAUTION
• To avoid possible damage to the meter or to the equipment
under test, disconnect circuit power and discharge all high
voltage capacitors before measuring resistance.
The ranges of measuring resistance are 400Ω, 4kΩ, 40kΩ,
400kΩ, 4MΩ, and 40MΩ.
The test leads may be add 0.1Ω to 0.2Ω of error to
resistance measurements. To test the leads, touch the probe
tips together and read the resistance of the leads. For best
accuracy, you can use the relative (Δ) mode to automatically
10

506/506B EN
Continuity Check
CAUTION
• To avoid possible damage to the meter or to the equipment
under test, disconnect circuit power and discharge all high
voltage capacitors before testing continuity.
subtract this value.
High-resistance (>10MΩ) readings are susceptible to
electrical noise. To smooth out most noisy readings, enter
the MAX/MIN recording mode; then step to the average
(AVG) reading.
The continuity check features a buzzer that sounds as long
as a circuit is complete. The buzzer allows you to quick
continuity checks without watching the display.
When measuring resistance is less than threshold, the
buzzer sounds. You can setup the threshold in setup mode.
The continuity threshold is default 30Ω.
To use continuity check, turn the rotary switch to resistance
position, then press the function (yellow) button to select
measuring mode.
11

506/506B EN
Testing Diodes
CAUTION
• To avoid possible damage to the meter or to the equipment
under test, disconnect circuit power and discharge all
highvoltage capacitors before testing diodes.
Use the diode test to check diodes, transistors, silicon
controlled rectifiers (SCRs), and other semiconductor
devices.
This function tests a semiconductor junction by sending a
current through the junction, then measuring the junction's
voltage drop. A good silicon junction drops between 0.5V
and 0.8V.
For forward-bias readings on any semiconductor component,
place the red test lead on the component's positive terminal
and place the black lead on the component's negative
terminal. In a circuit, a good diode should still produce a
forward-bias reading of 0.5V to 0.8V.
For reverse-bias readings on any semiconductor
component, the meter still can measure. In a circuit, a good
diode should still produce a reverse-bias reading of -0.5V to
-0.8V.
The display shows “OL” if the diode is open or short.
To use diode test, turn the rotary switch to resistance
position, then press the function (yellow) button to select
measuring mode.
12

506/506B EN
Measuring Capacitance
CAUTION
• To avoid possible damage to the meter or to the equipment
under test, disconnect circuit power and discharge all
highvoltage capacitors before measuring capacitance.
Use the dc voltage function to conrm that the capacitor is
discharged.
The ranges of measuring capacitance are 40nF, 400nF, 4uF,
40uF, 400uF, 4mF and 40mF.
To improve the accuracy of measurements less than
1000nF, you can use the relative (Δ) mode to subtract the
residual capacitance of the leads.
To use capacitance measurement, turn the rotary switch to
resistance position, then press the function (yellow) button
to select measuring mode.
13

506/506B EN
Measuring Temperature
The meter measures the temperature of a K-Type
thermocouple.
You can press the function (yellow) button to choose
degrees Celsius (°C) or degrees Fahrenheit (°F). Display
ranges are -200°C to +1200°C and -328°F to +2192°F.
Readings outside of these ranges show “OL” on the display.
When there is no thermocouple connected, the display also
shows “OL”.
To use temperature measurement, turn the rotary switch to
temperature position, then press the function (yellow) button
to select measuring mode.
14

506/506B EN
Auto-Hold
Maximum / Minimum Record
When measuring, you can press the A-HOLD button to start
the Auto-Hold mode. In this mode, the meter holds reading
and shows it on the secondary display.
If the dierence between new reading and hold reading is
bigger than 5d (3¾-digit mode), and new reading is also
stable, then meter automatically holds a new reading on the
secondary display.
When reading is smaller than Auto-Hold limit, or reading is
OL, the Auto-Hold mode is not working.
When measuring, you can record the maximum, minimum
and average value of reading.
To use maximum / minimum record mode, move the blink
cursor of menu to MAX, MIN or AVG position, and press
the ENTER button to enter mode. In this mode, the meter
records each data to compare the maximum and minimum
value. Also, meter calculate the average of reading.
You can move the blink cursor of menu to MAX, MIN or AVG
position, and press the ENTER button to select result on the
secondary display.
When maximum / minimum record mode running, if you
wants to pause recording, press the A-HOLD button. Press
again to continue.
To exit maximum / minimum record mode, press the
CANCEL button.
To exit Auto-Hold mode, press the A-HOLD button again. If
you don’t want to use the Auto-Hold mode, you can disable
it in the setup mode. When Auto-Hold mode is disable, the
hold mode is not update any new reading.
Function Limit
V, A, LoZ, Hz, Cap 1% of range
Others No limit
15

506/506B EN
Relative Δ
Relative %
When measuring, you can use the relative (Δ) mode to
subtract the oset.
To use relative (Δ) mode, move the blink cursor of menu to Δ
position, and press the ENTER button to enter mode. In this
mode, meter records the presently reading as reference and
shows it on the secondary display. The relative (Δ) mode
subtract reference from each reading, and shows result on
the main display.
In relative (Δ) mode, the minimum range is relative (Δ)
range. For example: The ranges of measuring resistance
are 400Ω, 4kΩ, 40kΩ, 400kΩ, 4MΩ, and 40MΩ. If you
use the relative (Δ) mode in 4kΩ range, then the minimum
range is the 4kΩ range. You can use range is 4kΩ to 40MΩ,
cannot change to 400Ω.
To exit relative (Δ) mode, press the CANCEL button.
When measuring, you can use the relative (%) mode to
calculate the relative percent value. The relative percent
value is dene as below:
To use relative (%) mode, move the blink cursor of menu to
% position, and press the ENTER button to enter mode. In
this mode, meter records the presently reading as reference
and shows it on the secondary display. The relative (%)
mode calculate the relative percent value from each reading,
and shows result on the main display.
To exit relative (%) mode, press the CANCEL button.
16

506/506B EN
Memory Save / Load
When measuring, you can save the reading to memory and
load it from memory. The meter can store maximum 1000
data in memory. The recorded data amount shows on the
secondary display.
To use memory save / load mode, move the blink cursor
of menu to MEM position, and press the ENTER button
to enter mode. In this mode, you can operate the below
options:
To exit memory save / load mode, press the CANCEL
button.
Memory Options
A-SAVE
SAVE
CLR
MAX
MIN
LOAD
You can operate the Auto-Save mode to automati-cally save
new reading. When you use the probes to measure a new
reading, the meter will automati-cally save it. In some case,
the Auto-Save mode will not work. For example, the reading
is smaller than the limit (refer Auto-Hold), or the reading is
OL. Press ENTER button to start Auto-Save mode, press
CANCEL button to exit.
You can press the ENTER button to review data from
memory. Press UP or DOWN button to select data. Press the
CANCEL button to return.
Press ENTER button to save a new reading to memory.
Press ENTER button to clear all data from memory.
Press ENTER button to review the maximum data from
memory.
Press ENTER button to review the minimum data from
memory.
17

506/506B EN
Data Logger
You can record a lot of reading to memory in a long time,
then analyze and plot graph. The meter can store maximum
40,000 data in memory. The recorded data amount shows
on the secondary display.
The record rate can be set from 1 sec and 600 sec. The
error of timer is less than 3 seconds per hour.
To use data logger, move the blink cursor of menu to LOG
position, and press the ENTER button to enter mode. In this
mode, you can operate the below options :
To exit data logger, press the CANCEL button.
Data Logger Options
SAVE
LOAD
CLR
RATE
MAX
MIN
Press ENTER button to start data logger. The logger
automatically records at regular intervals. You can press
ENTER button to pause data
logger, press again to continue. To stop data
logger, press CANCEL button to return.
You can press the ENTER button to review data from
memory. Press UP or DOWN button to select data. Press
the CANCEL button to return.
You can setup the record rate of logger.
Press UP or DOWN button to select rate.
Press the CANCEL button to return.
Press ENTER button to review the maximum data from
memory.
Press ENTER button to review the minimum data from
memory.
Press ENTER button to clear all data from memory.
18

506/506B EN
Auto Power O
Backlight
Buzzer
Display Resolution
Setup
If you don't operate the rotary switch or buttons for a
specied time, the meter will turn o automatically to save
the power of batteries. The default APO timer is 10 minutes.
In setup mode, you can change the APO timer.
The backlight is automatically turned on at dark
environment. The backlight mode is default auto. In setup
mode, you can set the backlight mode.
The meter equip a 2kHz tone buzzer. Valid button press:
Beep once. And invalid button press: Beep twice. In setup
mode, you can turn on or o the buzzer. But the buzzer in
continuity check cannot be turn o.
This meter have two display resolution: normal resolution
(3¾-digit mode) and high resolution (4¾-digit mode).
The normal resolution is set to default. You can setup the
resolution in setup mode.
To use meter setup mode, move the blink cursor of menu to
SETUP position, and press the ENTER button to enter
function. Press the UP or DOWN button to select item; press
the LEFT or RIGHT to select options. When you complete
setup, press the CANCEL button to exit setup mode.
1min to 30min, or OFF
Backlight mode: Auto, ON or OFF
Buzzer ON or OFF
Auto-Hold mode ON or OFF
Continuity threshold: 10Ω to 50Ω
Display digit: Lo or Hi
Temperature default unit: °C or °F
Press ENTER to reset all setup options.
Setup Options
APO
bEEP
A.Hold
Cntin
diGit
TEMP
RESET
b.Lit
19
This manual suits for next models
1
Table of contents
Languages:
Other APPA Multimeter manuals
Popular Multimeter manuals by other brands

Gossen MetraWatt
Gossen MetraWatt METRAHIT Base operating instructions

HP
HP 3478A operating instructions

Sperry instruments
Sperry instruments SMART meter operating instructions

Hanna Instruments
Hanna Instruments HI 9829 instruction manual

Fluke
Fluke 8020A instruction manual

Milwaukee
Milwaukee 2217-20 Operator's manual