Baker Hughes Druck DPI 610 IS User manual


Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | i
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures
detailed in this manual. Do not use this equipment for any other purpose than that stated.
This publication contains operating and safety instructions that must be followed to ensure safe
operation and to maintain the equipment in a safe condition. The safety instructions are either
warnings or cautions issued to protect the user and the equipment from injury or damage.
Use suitably qualified* technicians and good engineering practice for all procedures in this
publication.
Maintenance
The equipment must be maintained using the procedures in this publication. Further
manufacturer’s procedures should be carried out by authorized service agents or the
manufacturer’s service departments.
Druck.com
For technical advice contact the manufacturer.
Symbols
ATEX Approved Models
Introduction
These instructions detail the requirements for using the DPI 610 IS and DPI 615 IS intrinsically
safe pressure calibrator in a hazardous area. Read the whole publication before starting.
WARNING Before operating this intrinsically safe instrument, read the safety
instructions and the special conditions of safe use.
Do not apply pressure greater the maximum safe working pressure to the sensor.
* A qualified technician must have the necessary technical knowledge, documentation, special test
equipment and tools to carry out the required work on this equipment.
Symbol Description
This equipment meets the requirements of all relevant European safety directives. The
equipment carries the CE mark.
This symbol, on the equipment, indicates a warning and that the user should refer to the
user manual.
Ce symbole, sur l'appareil, est un avertissement qui indique que l'utilisateur doit consulter
le manuel d'utilisation.
Do not dispose of this product as household waste. Use an approved organization that
collects and/or recycles waste electrical and electronic equipment. For more information,
contact one of these:
- Our customer service department: Druck.com
- Your local government office.

Copyright 2007 Baker Hughes Company.
ii | DPI 610/615 IS Instruction Manual–English
Markings
II 1 G......................................................................................... Equipment group & category
Ex ia IIC T4 Ga.................................................................................Hazardous location markings
BAS02ATEX1174X............................................................................................Certificate number
................................................................................................................... CE Mark
DPI 61X IS ................................................................................................Specific apparatus type
(Pressure Range in mbar or psi .............................................................Full-scale pressure rating
Druck LTD. Groby LE6 0FH, UK ..............................................Manufacturer's name and address
SN *******/YY-MM....................................... Serial number and date of manufacture, Year-Month.
Requirements and Conditions
Batteries
Power supply use only 6 x LR14 (C) batteries:
• Duracell MN1400-LR14
• Procell lndustrial MN1400-LR14
• Energizer E93.KR14.C.AM2
• Energizer Industrial EN93
• Varta 4014 LR14.C.AM2
Special Conditions for Safe Use
1. The DPI 61X Series Pressure Calibrator is not capable of withstanding the 500 VRMS
electric strength test between external connectors and frame of the equipment as required
by EN 60079-11. This must be taken into account when using the equipment for input
measurements in a system.
2. The outer enclosure may contain light metals in the form of aluminium, magnesium, titanium
or zirconium. The equipment must be operated in such a manner as to prevent the possibility
of it being subjected to impact or friction.
Note: Does not apply whilst the equipment is in transit within the optional carrying case.
Electrical Parameters
Maximum Output Parameters:
WARNING Only replace batteries in a safe area.
lin (SK1a)
a. The output parameters at sockets SK1, SK2 and SK3 do not exceed the values
specified in clause 5.7, Simple Apparatus, of EN 60079-11.
Vin (SK2a) SwitchIn (SK3a)
Uo = 1.1 V dc Uo = 1.1 V dc Uo = 1.1 V dc
Io = 0.16 mA dc Io = 0.11 µA dc Io = 12 mA dc
Po = 0.15 mW Po = 0.03 µW Po = 11 mW
Ci = 0.05 µF Ci = 0 Ci = 0.05 µF
Li = 0 Li = 0 Li = 0
1180

Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | iii
Maximum Safe Input Parameters:
Installation
• Provide additional protection for equipment that may be damaged in service.
• Installation should be carried out by qualified plant installation technicians in compliance with
the latest issue of EN 60079-14.
Declaration Requirements - EU Directive 2014/34/EU
The DPI 610 IS and DPI 615 IS pressure calibrators are designed and manufactured to meet the
essential health and safety requirements not covered by the EU Type Examination Certificate
BAS02ATEX1174X when installed as detailed above.
lout (SK6) RS232 External Transducer
Uo = 7.9 V dc Uo = 7.6 V dc Uo = 7.9 V dc
Ci = 0 Io = 82 mA dc Io = 155 mA dc
Li = 0.1 mH Po = 162 mW Po = 0.43 W
Ci = 0 Ci = 0.15 µF
Li = 0 Li = 0.9 mH
Um = 250 V Co = 8.6 µF
Lo = 0.3 mH
SK1, SK2, SK3 and SK6
Ui = 30 V
li = 100 mA
Pi = 1.0 W
WARNING Do not use tools on the pressure calibrator that might cause
incendive sparks - this can cause an explosion.

Copyright 2007 Baker Hughes Company.
iv | DPI 610/615 IS Instruction Manual–English

Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | v
Contents
1. Introduction 1
1.1 General 1
1.2 Description of Procedures 1
1.3 Using This Guide 2
1.4 Maximum Instrument Ratings 2
1.5 Operator Controls 3
1.6 Display 3
1.7 Hard Key Functions 4
1.8 Soft Keys 4
1.9 Cursor Keys 4
1.10 Electrical Connections 6
2. Getting Started 7
2.1 Fitting Batteries 7
2.2 Switching On 7
2.3 Change Pressure Units 8
2.4 Voltage and Current Measurements 8
2.5 Typical Calibration Set-up (Pressure to Voltage) 9
2.5.1 General Procedure 9
2.6 Zero Display Reading 9
3. Task Selection 11
3.1 Task Key 11
3.2 Using TASK Functions 11
3.3 Set Units 11
3.4 Cal Mode (DPI 615 versions only) 11
3.5 Basic Mode (Task BASIC) 12
4. Taking Measurements 13
4.1 Pressure Transmitter (P-I) Task 13
4.2 Voltage Output Pressure Transmitter (P-V) Task 13
4.3 Pressure Converter (Pressure to Pressure) Task 14
4.4 Current to Pressure Converter (I-P) Task 15
4.5 Pressure Switch Test (P-SWITCH) Task 16
4.6 Pressure to Display (P-DISPLAY) Task 17
4.7 Leak Test (LEAK TEST) Task 18
4.8 Transmitter Simulator (TX SIM) Task 19
4.9 Relief Valve Test (REL VALVE) Task 20
5. Advanced Task 21
5.1 General 21
5.2 Select Input 21
5.3 Ambient Temperature Measurement 21
5.4 Process Functions 21
5.4.1 Tare Process Function 22
5.4.2 Min/Max Process Function 23
5.4.3 Filter Process Function 23
5.4.4 Flow Process Function 24

Copyright 2007 Baker Hughes Company.
vi | DPI 610/615 IS Instruction Manual–English
5.4.5 % Span Process Function 24
5.5 Select Output 24
5.6 Electrical Outputs (Loop Power) 25
5.6.1 mA Step 25
5.6.2 mA Ramp 26
5.6.3 mA Value 26
5.7 Define New Task 27
5.8 Clear Task 27
6. Memory Operations 29
6.1 Saving Display or Data Log 29
6.2 Store Operations (Screen Snapshots) 29
6.3 Recalling Stored Data (Screen Snapshots) 29
6.4 Data Log Operations 29
6.4.1 Auto Log (Timer) 30
6.4.2 Manual Logging 30
6.4.3 Recall Data Log Files 30
6.4.4 Uploading Data Log Files 31
6.4.5 Delete Datalog and Procedure Files 32
6.4.6 Downloading Procedure Files 32
6.4.7 Running Procedure Files 33
6.4.8 Recalling Data Files 33
7. Using Set-up 35
7.1 General 35
7.2 Store Mode 35
7.3 Contrast 35
7.4 Settings - Select Setup Option 35
7.4.1 Units 36
7.4.2 Define Special Units 36
7.4.3 Language 36
7.4.4 RS232 37
7.4.5 Powerdown 37
7.4.6 Calibration 37
7.5 Date and Time (Real Time Clock) 38
7.5.1 Date Format 38
7.5.2 Set Date 38
7.5.3 Set Time 38
8. Calibration 39
8.1 General 39
8.2 Calibration Check 39
8.3 Calibration Adjustment 39
8.4 Guide to Calibration Procedures 39
8.5 Test Equipment 40
8.6 Using the Calibration Menu 40
8.7 PIN security 41
8.7.1 Change PIN 41
8.8 Calibrate Internal Ranges 41
8.9 Internal Pressure Range 41
8.10 Voltage Input Range (5 V) 42
8.11 Voltage Input Range (30 V) 44

Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | vii
8.12 Current Input Range (55 mA) 46
8.13 Current Output Range (24 mA) 48
8.14 Ambient Temperature Channel 50
8.15 Calibrate External Sensors 51
8.16 Add External Sensor 53
9. Hydraulic Calibrator Versions 55
9.1 Introduction 55
9.2 Safety Instructions 56
9.3 Preparation for Use 56
9.4 Bleeding the System 57
9.5 Operation 58
9.6 Draining the Hydraulic Fluid 58
9.7 Flushing - Refilling or Changing the Hydraulic Fluid 58
Appendix A. Datalog File Example 61
A.1 Typical Uploaded Data Log File (DPI 610) 61
A.2 Typical Uploaded Procedure Data File (DPI 615) 62
A.3 Record Type 62
Appendix B. “Quick Fit” Pressure Connector 63
B.1 Introduction 63
B.2 Connection Sequence 63

Copyright 2007 Baker Hughes Company.
viii | DPI 610/615 IS Instruction Manual–English

Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | 1
General
1. Introduction
1.1 General
The DPI 610 IS and DPI 615 IS intrinsically safe instruments measure and display pneumatic and
hydraulic pressure applied to the test port.Pressure measurement can be absolute, gauge and
sealed gauge and in ranges from 2.5 mbar to 700 bar (1.0 inH2O to 10000 psi).
Calibrator versions of this instrument contain pneumatic or hydraulic pressure generation
components to produce pneumatic pressure ranges between -1 to 20 bar (-14.5 psi to 300 psi)
and hydraulic pressure ranges up to 400 bar (6000 psi).
Using external electrical connections, the DPI 610 IS and DPI 615 IS intrinsically safe
instruments measure ±30 V dc and ±55 mA. An integral sensor provides measurement of ambient
temperature. Additional sensors (option B1) connect to an external connector and extend the
pressure measurement range and include differential pressure measurement. The DPI 615
instrument has an RS232 connector to enable downloading of test data to a compatible
documenting system. Six alkaline C size batteries, IEC Type LR14, power the instrument.
1.2 Description of Procedures
The procedures apply to both the DPI 610 IS and the DPI 615 IS instruments unless otherwise
stated. In the procedures in this manual, hard (fixed function) and soft (variable function) key
operations are shown in bold type: TASK and F1. These statements mean press the TASK key
and press the F1 key. Soft key operations can be assigned to both the F1 and F2 keys. Where a
specific soft function is referred to it is written in bold italics (e.g.) PROCESS.
This instrument has a number of operating modes that are described in simplified form in the
following sections. Diagrams accompanying the procedures give typical selection sequences and
shaded controls indicate that this control key should be pressed in the appropriate sequence.
Diagrams should be read from left to right, top to bottom where appropriate. A shaded display soft
box indicates that the function key immediately below that soft box should be pressed (either F1
for the left hand soft box or F2 for the right).
In the above diagram the following key sequence is indicated.
a. Press the F2 key (the key immediately below the UNITS soft box).
b. Use the Up and Down cursor keys (only) to select the required option. (If all keys shaded,
use all these keys to select or enter data).
c. Press the ENTER key.
INFORMATION Zinc-carbon and zinc-chloride cells must not be used in this
instrument. Use only the battery types as shown in “Batteries” on page ii.
SELECT UNITS
TASK : BASIC
VOLTAGE
PRESSURE INT
CURRENT PRESSURE
UNITS
SELECT UNITS
OF PRESSURE

Copyright 2007 Baker Hughes Company.
2 | DPI 610/615 IS Instruction Manual–English
Chapter 1. Introduction
1.3 Using This Guide
The following key symbols are used in the procedure diagrams:
1.4 Maximum Instrument Ratings
The following table shows the maximum measurement input ratings of the instrument which
should not be exceeded.
Table 1-1: Key Symbols
Key Description
Shaded cursor keys indicate that a combination of these four keys, Up,
Down, Left and Right should be used to (e.g.) enter an alpha numeric value
or to select a function.
Indicates the ENTER key. Used to confirm an operation or a selection.
Shading indicates key operation.
Exit key, used to clear current menu selection and return to next menu level
above current level. Used as an escape key from current operation. Shading
indicates key operation.
Hardkey (total 7). Legend beside key symbol indicates function. Shading
indicates key operation.
Table 1-2: Maximum Instrument Rating
Parameter Maximum Ratinga
a. The display flashes if the input pressure, voltage
or current over-range.
Pressure 120% of full-scale
Voltage 30 V dc
Current 55 mA dc
EXIT
RECALL

Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | 3
Operator Controls
1.5 Operator Controls
These divide into two groups, the operator/display controls (Figure 1-1 on page 3) and the
pressure/vacuum generation components (Figure 1-2 on page 5). The operator controls and a
typical display, common to all instrument versions, is shown below.
Figure 1-1: DPI 610/615 IS Key-pad
1.6 Display
The display and key-pad of the instrument divides into four distinct sections. The two main
sections of the display are used to show an input and an output. The remaining two sections show
the status display area and define the soft key functions. A typical display is shown below:
1 Display 2 Electrical Measurement Input Sockets
3 Cursor Keys 4 Enter Key
5 Function (Soft) Keys 6 Hard Keys
7 On/Off Key
1 Status Display 2 Input Display
3 Output Display 4 Soft Boxes
PRESSURE INT
VOLTAGE
CURRENT PRESSURE
UNITS
TASK : BASIC
PRESSURE INT
VOLTAGE
CURRENT
PRESSURE
UNITS
TASK : BASIC

Copyright 2007 Baker Hughes Company.
4 | DPI 610/615 IS Instruction Manual–English
Chapter 1. Introduction
1.7 Hard Key Functions
1.8 Soft Keys
Three soft keys, designated F1, EXIT and F2, are situated immediately below the display as
shown below. These keys have their function allocated by the instrument software which is
indicated in the bottom of the display (Voltage for F1 and Units for F2 in this example). They are
used to select menu (program) options and are fully described under the appropriate section
headings.
1.9 Cursor Keys
The cursor keys consist of a block of four keys, designated up , down , left , and right .
In programs where options need to be selected from a list, (e.g.) the TASK selection program,
the up and down cursor keys are used to highlight one of the options, from which it can be
Table 1-3: Hard Key Functions
Key Function
I/O This key selects the instrument ON and OFF.
SETUPa
a. These key functions are not available in BASIC mode.
The SETUP key provides access to the instrument’s general
configuration parameters that are set-up to certain default parameters
on delivery.
ZERO
The ZERO key zeros either the selected input or output display, only if
the display reading is within 5% of zero. Attempts to zero a larger offset
result in an error message, Zero too large.
INPUTaThe INPUT key selects the input parameter to be displayed.
TASK
The TASK key rapidly configures the instrument for a number of
different types of external device calibration. There are twenty task
configurations available, eleven pre-programmed configurations and
nine user defined configurations.
OUTPUTaThe OUTPUT key selects the output parameter to be displayed.
STOREa
Depending upon how the instrument’s STORE mode is set-up, this key
is used either to store up to 20 display screens in SNAPSHOT mode,
or to manually log a screen in DATALOG mode.
RECALLa
This key recalls a previously stored screen to the display. Depending
on the STORE mode set-up, operation of this key recalls either the
snapshot of a previously stored screen or data log file. In STORE
mode, selection displays the last screen stored. By using the cursor
keys, the operator can scroll either forward or back through memory
locations.
ENTER The ENTER key either enters data (accept entered data), or, in
conjunction with the soft keys, accepts a given selection.
EXIT
The EXIT key operates in conjunction with all the other hard and soft
keys to exit from the current screen or menu level, to the level
immediately preceding it. To quit completely from any menu level,
press EXIT until the MEASURE/SOURCE screen is displayed.

Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | 5
Cursor Keys
selected by the ENTER key. In TASK mode, where more than one page of options are provided,
the left , and right cursor keys will switch between pages.
Figure 1-2: DPI 610/615 IS Calibrator Controls
1 Test port, connect to unit under test 2 Hard keys
3 Cover (external interfaces) 4 Electrical inputs
5 Cursor keys 6 Function (soft) keys
7 Release valve (releases pressure through 8) 8 Vent port
9 Select positive or negative pressure 10 Pump
11 Fine pressure adjuster
7
8
9
5
6
4
3
10
11
1
2

Copyright 2007 Baker Hughes Company.
6 | DPI 610/615 IS Instruction Manual–English
Chapter 1. Introduction
1.10 Electrical Connections
Figure 1-3: Electrical System Connections
Measurement inputs and Source outputs are made via the control panel sockets as shown below:
Figure 1-4: Electrical Measurement Inputs/Source Outputs
1 Cover, closed when not using connectors 2 External transducer
3 RS232 connector 4 Temperature sensor
1 Status window 2 Input window
3 Electrical measurement input sockets 4 Output window
1
2
3
4
PRESSURE INT
VOLTAGE
CURRENT
INPUTS
OUTPUTS
PRESSURE
UNITS
TASK : BASIC

Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | 7
Fitting Batteries
2. Getting Started
2.1 Fitting Batteries
2.2 Switching On
Press the I/O switch on the front panel and proceed as follows:
The first time that the instrument is powered up, it will power-up in BASIC mode with the main
screen displaying voltage in the input display area and pressure in the source display area. To
switch to Current as input, press F1 as shown. Similarly, F1 to return to Voltage.
Note: No other keys are active in this mode and the instrument can only be reconfigured by
pressing the TASK key and selecting another mode.
1 Cover fixing screws.
2 Six alkaline C cells, see table. Only use the battery type in the table.
3 Low battery indication.
WARNING Batteries must only be fitted in a safe area.
Use only the batteries specified in “Batteries” on page ii.
CAUTION Old batteries can leak and cause corrosion. Never leave discharged
batteries in the instrument. Old batteries should be treated as hazardous waste
and disposed of accordingly.
PRESSURE INT
VOLTAGE
CURRENT PRESSURE
UNITS
TASK : BASIC
+
-
2
1

Copyright 2007 Baker Hughes Company.
8 | DPI 610/615 IS Instruction Manual–English
Chapter 2. Getting Started
2.3 Change Pressure Units
To change the pressure units proceed as follows. If the four units displayed are not the units
required, press TASK and select any task, other than BASIC, press SETUP and proceed as
detailed in Section 7.1. To return to BASIC mode, press TASK and select BASIC.
In BASIC mode, the unit is configured to carry out basic Pressure to Voltage (P to V) or Pressure
to Current (P to I) tests, a typical test procedure follows:
2.4 Voltage and Current Measurements
Connect the electrical input sockets as follows for voltage and current measurements. Use the
test leads provided and DO NOT push bare wires into the sockets.
Note: See Table 1-2 on page 2 for maximum instrument ratings.
SELECT UNITS
TASK : BASIC
VOLTAGE
PRESSURE INT
CURRENT PRESSURE
UNITS
SELECT UNITS
OF PRESSURE
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
V
55mA
Vin
mA in
mA Sink
max 30V
max
+
-

Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | 9
Typical Calibration Set-up (Pressure to Voltage)
2.5 Typical Calibration Set-up (Pressure to Voltage)
Connect a device under test to the instrument as shown below:
2.5.1 General Procedure
Use the hand-pump to pressurize the system to the required level as indicated on the display.
Allow the display to settle and screw the volume adjuster in or out as a fine adjustment to the
required pressure. Record the input: Voltage, reading at each applied pressure.
2.6 Zero Display Reading
Both the input and output readings can be set to zero by using the ZERO key and if the displayed
reading is within 5% of zero. To zero either the INPUT or OUTPUT displays, proceed as follows:
A External pressure source (indicator instruments only) B Pressure regulator
C Pressure to voltage device D Barrier
E Excitation 10 V F Safe area
PRESSURE INT
VOLTAGE
CURRENT PRESSURE
UNITS
TASK : BASIC

Copyright 2007 Baker Hughes Company.
10 | DPI 610/615 IS Instruction Manual–English
Chapter 2. Getting Started
This manual suits for next models
3
Table of contents
Other Baker Hughes Test Equipment manuals

Baker Hughes
Baker Hughes Druck PACE Series Use and care manual

Baker Hughes
Baker Hughes Druck Dry PTC 200 User manual

Baker Hughes
Baker Hughes Druck DPI 610A User manual

Baker Hughes
Baker Hughes Druck DPI 612 User manual

Baker Hughes
Baker Hughes Druck DPI 610E-PC User manual

Baker Hughes
Baker Hughes Druck PACE Use and care manual

Baker Hughes
Baker Hughes Druck DPI 620 Genii User manual

Baker Hughes
Baker Hughes Druck DPI 620 Genii User manual

Baker Hughes
Baker Hughes Druck DPI 880 User manual

Baker Hughes
Baker Hughes Druck DPI 880 User manual