Beurer IH20 User manual

WARNING: U.S. Federal Law restricts this device to sale by or on
the order of a licensed physician. Do not use unless prescribed by a
licensed physician and with the prescribed medication only.
Follow the physician’s instructions for use at all times.
ADVERTENCIA: La ley federal de los Estados Unidos limita la
venta a o por prescripción de un médico autorizado. No lo use a
menos que lo prescriba un médico y solamente con el medicamento
prescrito. Siga siempre las instrucciones de uso del médico.
Nebulizer......................... 2
Instructions for use
Inhalador....................... 23
Instrucciones para el uso
IH20
Distributed by:
Beurer North America LP • Hallandale Beach, FL 33009
www.beurer.com

2
English
1. Important Safety Notes
Signs and Symbols
The following signs appear in the Safety Section (pages 2 - 6) and in this manual on pages
9, 10, 11, 13, 14, 16, 17.
READ THIS ENTIRE MANUAL, INCLUDING THE SAFETY SECTION AND ALL
INSTRUCTIONS AND WARNINGS, COMPLETELY AND CAREFULLY BEFORE USING
THIS PRODUCT. FOLLOW ALL SAFETY INSTRUCTIONS AND WARNINGS TO AVOID
HAZARDOUS SITUATIONS AND TO MAKE CORRECT USE OF THIS PRODUCT.
DANGER DANGER indicates a hazardous situation which, if not avoided, will
result in death or serious injury.
WARNING WARNING indicates a hazardous situation which, if not avoided, could
result in death or serious injury.
CAUTION CAUTION indicates a hazardous situation which, if not avoided, may
result in minor or moderate injury.
NOTICE NOTICE addresses practices not related to personal injury, such as
product and/or property damage.
DANGER
•Always unplug this product immediately after use and make sure it is not accessible to
children.
•Do not place or store product where it can fall or be pulled into a bathtub, sink, water, or
other liquid. Risk of electric shock.
•Do not use this product while bathing. Showering, washing dishes, or close to water
sources of any kind. Risk of electric shock.
•Do not place device in or drop into water or other liquid. Do not allow any water to be
sprayed onto the unit. The unit must be operated only when it is completely dry. Risk of
electric shock.
•Do not reach for a product that has fallen into water or other liquid. If possible, safely un-
plug immediately. Do not touch a wet supply cable or plug. Do not touch the supply cable
or plug with wet extremities.
•Never touch the unit, including the power cord or plug, with wet hands while it is plugged
in. This can cause an electric shock.
•Before use, ensure that there is no visible damage to the unit or accessories. Never op-
erate the unit if it
- has any damaged parts, e.g. damaged cord, plug or housing
- has been exposed to extreme humidity or to any liquids inside the housing
- has been dropped,
- is not working properly.

3
If the unit does not function correctly or if you notice any abnormalities, e.g. an unusual
noise or smell, stop using the unit immediately. Switch off and unplug the unit immedi-
ately. When in doubt, do not use the unit and contact Beurer customer service. (Please
see warranty for service contact).
•Do not open or disassemble the nebulizer. Risk of electric shock. All disassembly and
maintenance of the device must be performed by a qualified service technician. If you
need to have the unit repaired, please contact our customer service. Please see warranty
for service contact.
•Before you clean the unit, always switch it off, unplug it and let it cool off.
•A product should never be left unattended when plugged in.
•When the switch is in the off position and the compressor is not running there will still be
electrical potential inside the unit. Risk of electric shock: To remove power from the com-
pressor, the unit must be unplugged from the power outlet.
•Do not pull the plug from the power outlet by the cord. Always grasp the AC plug and un-
plug it. Failure to do so could result in risk of electric shock, short circuiting, or fire.
•Do not pinch or bend the power cord or pull it over sharp objects. This can cause an elec-
tric shock.
•The unit and power cord must be protected and stored away from sources of heat. (e.g.
stoves, heating radiators)
•This appliance has a polarized plug (one blade is wider than the other). As a safety fea-
ture, this plug will fit in a polarized outlet only one way. If the plug does not fit fully in the
outlet, reverse the plug. If it still does not fit, contact a qualified electrician. Do not modify
the plug in any way. Do not attempt to manipulate with this safety feature.
•Do not operate this device outdoors. Risk of electric shock.
•Do not operate the device in a condensing or an ammoniac rich environment.
•The device contains an overload protection fuse. This must only be changed by author-
ized specialist personnel.
•Do not use extension cords, multi-sockets and/or voltage converters.
WARNING
•U.S. Federal Law restricts this device to sale by or on the order of a licensed physician.
Do not use unless prescribed by a licensed physician and with the prescribed medication
only. Follow the physician’s instructions for use at all times
•Using the unit is not a substitute for consultation with or treatment from a physician.
Whenever you have any pain, illness or discomfort, always contact your physician first. If
you have an emergency please call 911 immediately.
•In the case of oversensitivity of the bronchial system, medicines containing essential oils
can sometimes cause acute bronchiospasm (a sudden, cramp-like tightening of the bron-
chi accompanied by breathlessness). In case of an emergency call 911 immediately.
•If you start feeling any discomfort, pain, drowsiness or other change in your condition,
stop using the unit immediately. For medical advice please contact a licensed physician. If
you have an emergency please call 911 immediately.
•Always follow your physician’s instructions regarding the type of medicine to be used, the
dosage, frequency, and duration of inhalation. Any improper use may cause hazardous
situations.
•Apart from the prescribed medication, use only distilled water or a saline solution in the
unit. Other liquids may cause health problems. Always consult your physician to deter-
mine what will be suitable for you. Furthermore, other liquids may damage the nebulizer,
atomizer and accessories.

4
•Always use close adult supervision when administering a treatment to a child. NEVER
LEAVE CHILDREN UNATTENDED WITH THIS DEVICE.
•This device is not intended to be used by persons with restricted physical (e.g. activity
limitations, disabilities), sensory (e.g. insensitivity to pain) or mental abilities (e.g. mental
disorders) or persons lacking the required experience or knowledge for safe operation of
the device, unless closely supervised by a person responsible for their safety. NEVER
LEAVE THOSE WHO REQUIRE CLOSE SUPERVISION UNATTENDED WITH THIS
DEVICE.
•This product contains small parts that may present a choking hazard to children. Keep all
small parts out of reach of children. The tubing and power cord also present a strangula-
tion hazard. Always use close adult supervision when administering a treatment to a child.
NEVER LEAVE CHILDREN OR THOSE WHO REQUIRE CLOSE SUPERVISION
UNATTENDED WITH THIS DEVICE.
•Packaging materials are a deadly hazard for children and can cause suffocation. Remove
all packaging materials immediately and keep them away from children at all times.
•Use the unit only on human beings and for its intended purpose (aerosol inhalation) and in
the manner described in these instructions for use. Any improper use can be dangerous!
•Do not use the unit outside the specified ambient conditions. See section 8 Maintenance
and section 11 Technical Specifications for details.
•Do not operate the unit in the presence of flammable gases, oxygen and aerosol spray
products. This device has no AP or APG protection. Non-compliance may result in e.g.
fire, burns, explosions, etc.
•To reduce the risk of increased bacterial growth, infection, illness, injury from contamina-
tion or other health risks always follow all hygiene instructions (see section 8.1 Cleaning
and Disinfection) when using this unit. Unplug the unit before disinfection or cleaning the
unit.
•For hygienic reasons, the atomizer [10] and the accessories must be cleaned after each
treatment and disinfected after the last treatment of the day. Prior to initial operation, the
atomizer [10] and accessories must be cleaned and disinfected as well. For details see
section 8.1 Cleaning and Disinfection.
•Ensure that all parts are thoroughly dried, following every cleaning and disinfection. Re-
sidual moisture or water droplets can result in an increased risk of bacterial growth and
therefore represent a risk for your health.
•If several different medicines are to be inhaled in succession during therapeutic use, the
atomizer [10] must be rinsed under warm tap water after each use before using different
medication. For information about this, see section 8.1 Cleaning and Disinfection.
•Single patient use only: For hygienic reasons, each user must use their own accessories.
There is the risk of infections and communication of diseases.
•During use, hold the unit away from the eyes, as some nebulized medicines may cause
damage to the eyes.
•Do not use this unit while operating a vehicle.
•Never use while sleeping or drowsy.
•Do not administer treatment to a sleeping person, because not enough medicine will
reach the lungs.
•Before use, check the data plate on the bottom of the IH 20 Nebulizer to ensure that the
voltage and current indicated on the unit correspond to the voltage and current available.
•The IH 20 Nebulizer must only be used with suitable Beurer atomizers and with the ap-
propriate Beurer accessories. The use of other atomizers and accessories may damage
the device and can impair therapeutic efficiency.

5
•This device should not be used adjacent to, or stacked with other equipment. If adjacent
or stacked use is necessary, this device should be observed to verify normal operation.
•Before each use, check that the hose connections on the nebulizer and the atomizer are
tight. Use of an improperly assembled nebulizer could diminish or prevent adequate deliv-
ery of medication and the effectiveness of the treatment.
•Portable and mobile RF communication devices can disrupt electrical medical equipment.
Technical electromagnetic compatibility data is available on request from Beurer customer
service. (Please see warranty for service contact)
•Never block the air vents of this product or place it on a soft surface, such as a bed or
couch, where the air vents may be blocked. Keep the air vents free of lint, hair and similar
substances. Blockage could cause excessive heat buildup.
•Never cover the unit during operation. Covering could cause excessive heat buildup.
•This unit is not intended for commercial or clinical use, but only for individual, private
household use! Do not allow others to use it.
•Do not allow any objects to enter or obstruct the air vents.
•Lay the power cord in such a way that no one can trip over it. There is a danger of injury
due to e.g. stumbling, tripping, falling down.
•Do not tilt or shake the unit when in use.
•Essential oils, cough medicines, solutions designed for gargling, and drops for application
to the skin or for use in steam baths are wholly unsuitable for inhalation using a nebulizer.
These substances are often highly viscous and can impair the correct function of the de-
vice and hence affect the effectiveness of its application in the long term.
•The atomizer and accessories must not be mechanically cleaned with brushes or the like
as this can cause irreparable damage and may render the treatment ineffective.
•For additional requirements regarding the necessary hygienic preparations (hand wash-
ing, handling of medicines or inhalation solutions) in high-risk groups (e.g. cystic fibrosis
patients), contact your physician.
•Check the atomizer, mouthpiece, and any other optional components before each use.
Dirty, damaged and/or worn parts may diminish or eliminate the effectiveness of treat-
ment. Replace damaged and/or worn parts. Dirty parts must be cleaned following all hy-
giene instructions. (see section 8.1 Cleaning and Disinfection)
•When cleaning, ensure that all residues are removed. Never use any substances that may
be poisonous upon contact with the skin or mucous membranes, or when swallowed or
inhaled. When in doubt, contact your physician for advice.
•Never use an atomizer with a clogged nozzle. If the nozzle is clogged, the aerosol mist will
be reduced, altered or stopped, which will diminish or eliminate the effectiveness of the
treatment.
•Follow the instructions for changing the filter contained in these instructions. A contami-
nated filter can impair the correct function of the device and represents a risk to your
health.
CAUTION
•Do not strongly pull the elastic band of the adult mask and child mask. The ends of the
elastic band may slip out and hurt you (e.g. your eyes).
•Do not operate the nebulizer without a filter.
•Check the filter at regular intervals (at the latest after every 10th use). If the filter is very
contaminated or blocked, replace the used filter. If the filter has become wet, it must also
be replaced with a new one.
•Do not try to clean and re-use the used filter!

6
•Only use original filters supplied by the manufacturer. Otherwise, your nebulizer may be-
come damaged or treatment may not be effective.
NOTICE
•Protect the unit or the parts from strong impacts such as dropping the unit on the floor.
•Do not leave medication inside the atomizer when not in use.
•Never open or attempt to repair the unit yourself. Failure to observe this regulation shall
void the warranty.
•When disinfecting the atomizer, mouthpiece and nosepiece with the “Boiling water
method”, avoid that the parts come into contact with the hot base of the pot.
•The face masks and air hose must not be placed in hot water!
•Ensure that no water penetrates inside the unit! Never immerse the unit in water or any
other liquid, otherwise liquid can enter it and cause damage.
•Do not put the device or accessories in a dishwasher, washing machine or tumble dryer.
•Store the device and accessories in a location protected against environmental influ-
ences, i.e. in a safe and dry location. Do not expose the device and accessories to direct
sunlight, extreme hot or cold temperatures or humidity (e.g. in the bathroom). The device
and accessories must be stored in the specified ambient conditions. (see section 11
“Technical Specifications” for details)
•Protect the unit against dust, dirt and moisture. This may damage the device.
•Do not use any aggressive solvents, cleaning agents, detergents or any other strong
chemicals to clean the device.
•Unclogging the small holes of the nozzle should only be done carefully from the inside out
using a needle or pin. Replace if necessary.
•Dispose of the device, components and optional accessories according to applicable local
regulations. Unlawful disposal may cause environmental pollution.
•Use the nebulizer intermittently; the maximum operating period is 30 minutes. The nebu-
lizer must stand unused for 30 minutes between uses.
•If the medicine cup is empty, do not attempt to operate the unit.

7
Contents
1. Important Safety Notes.................. 2
DANGER................................ 2
WARNING.............................. 3
CAUTION............................... 5
NOTICE ......................................... 6
2. Package Contents ......................... 7
3. Getting to Know Your Nebulizer .... 7
4. Description of Device and
Accessories.................................... 8
5. Initial Use ....................................... 9
6. Operation......................................10
7. Changing the Filter .......................13
8. Maintenance .................................14
8.1 Cleaning and Disinfection.....14
8.2 Material Resistance ..............16
8.3 Storage and Caring for the
Device...................................16
8.4 Disposal ................................17
9. Troubleshooting............................18
10.Frequently Asked Questions ........18
11.Technical Specifications...............19
12.Warranty and Service...................21
SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE
If the instruction manual is damaged or if you no longer have the instruction manual in your
possession, please contact Beurer customer service. Please see warranty for service con-
tact.
The manufacturer shall not be held liable for damage or injuries caused by improper or in-
correct use.
2. Package Contents
•For details of the delivered content, see description of the device and accessories on page 8.
•Nebulizer •Atomizer
•Compressed-air hose •Mouthpiece
•Adult mask •Child mask
•Nosepiece •Spare filter
•These instructions for use •Quick reference guide
3. Getting to Know Your Nebulizer
Dear valued customer
Thank you for choosing this Beurer product. Our name stands for high-quality, thoroughly
tested products for applications in the areas of heat, weight, blood pressure, body tempera-
ture, pulse, gentle therapy, massage and air. Please read these instructions completely and
carefully before using this product and keep them for later reference. Be sure to make them
accessible to all other users and observe the information they contain.
For product information, questions or customer service call toll free 1-800-536-0366 or email
to info@beurer.com.
With best regards,
Your Beurer team

8
Area of application
This nebulizer is a device used only for the nebulization of liquids and liquid medication
(aerosols) approved for nebulization and prescribed by a physician for the treatment of the
upper and lower respiratory tract of humans.
The nebulization and inhalation of medication prescribed by a physician can fight against
diseases of the respiratory tract, treat the side effects of these disorders and accelerate
healing. Consult your physician for more information on possible applications.
The device is suitable for inhalation at home, as directed by a physician. Inhalation should
be performed in a calm and relaxed atmosphere. Inhale slowly and deeply to enable the
medicine to reach the small bronchi deep into the lungs. Exhale normally.
4. Description of Device and Accessories
Overview of the Nebulizer
1 Power cord
2 Cover
3 On/off switch
4 Compartment for
accessories
5 Air vents
6 Filter cap with filter
7 Hose connection
8 Holder for atomizer
Overview of Atomizer and Accessories
9 Compressed-air hose
10 Atomizer
11 Nozzle attachment
12 Medicine container
13 Mouthpiece
14 Adult mask
15 Child mask
16 Nosepiece

9
5. Initial Use
WARNING:
•Packaging materials are a deadly hazard for children and can cause suffocation. Remove
all packaging materials immediately and keep them away from children at all times.
•To reduce the risk of increased bacterial growth, infection, illness, injury from contamina-
tion or other health risks always follow all hygiene instructions (see section 8.1 “Cleaning
and Disinfection”) when using this unit. Unplug the unit before disinfection or cleaning the
unit.
Setup
•Unpack the unit. Place it on an even and
stable surface.
•Do not place on cushioned, upholstered,
carpeted or other soft surfaces. Place the
unit in an area that is clean and free of
dust.
•Ensure that the air vents are free.
•Push the cover [2] backwards to access
the accessory parts.
Before Using the Unit for the First
Time
•The atomizer and accessories must be
cleaned and disinfected before they are
used for the first time. For information
about this, see section 8.1 “Cleaning and
Disinfection”.
•Attach the compressed-air hose [9] at the
bottom of the medicine container [12].
•Connect the other end of the hose [9] to
the hose connector [7] on the nebulizer
by lightly turning it.

10
Connecting the Power Cord
DANGER:
•Never touch the unit, including the power cord or plug, with wet hands while it is plugged
in. This can cause an electric shock.
WARNING:
•Before use, check the data plate on the bottom of the IH 20 Nebulizer to ensure that the
voltage and current indicated on the unit correspond to the voltage and current available.
•Lay the power cord in such a way that no one can trip over it. There is a danger of injury
due to e.g. stumbling, tripping, falling down.
Connect the unit only to a 120 V household receptacle.
•Ensure that there is a power outlet near to
where the unit will stand.
•Insert the mains cable plug [1] into a suit-
able power outlet.
•Push the plug completely into the power
outlet.
•To unplug the nebulizer from the power
outlet after use, first switch off the device
and then remove the plug from the power
outlet.
6. Operation
DANGER
•Always unplug this product immediately after use and make sure it is not accessible to
children.
•Before use, ensure that there is no visible damage to the unit or accessories. Never oper-
ate the unit if it
- has any damaged parts, e.g. damaged cord, plug or housing
- has been exposed to extreme humidity or to any liquids inside the housing
- has been dropped
- is not working properly.
If the unit does not function correctly or if you notice any abnormalities, e.g. an unusual
noise or smell, stop using the unit immediately. Switch off and unplug the unit immedia-
tely. When in doubt, do not use the unit and contact Beurer customer service.
(Please see warranty for service contact)
•A product should never be left unattended when plugged in.
WARNING
•U.S. Federal Law restricts this device to sale by or on the order of a licensed physician. Do
not use unless prescribed by a licensed physician and with the prescribed medication
only. Follow the physician’s instructions for use at all times
•Using the unit is not a substitute for consultation with or treatment from a physician.
Whenever you have any pain, illness or discomfort, always contact your physician first. If
you have an emergency please call 911 immediately.

11
•In the case of oversensitivity of the bronchial system, medicines containing essential oils
can sometimes cause acute bronchiospasm (a sudden, cramp-like tightening of the bron-
chi accompanied by breathlessness). In case of an emergency call 911 immediately.
•If you start feeling any discomfort, pain, drowsiness or other change in your condition, stop
using the unit immediately. For medical advice please contact a licensed physician. If you
have an emergency please call 911 immediately.
•Always follow your physician’s instructions regarding the type of medicine to be used, the
dosage, frequency, and duration of inhalation. Any improper use may cause hazardous
situations.
•Apart from the prescribed medication, use only distilled water or a saline solution in the
unit. Other liquids may cause health problems. Always consult your physician to determine
what will be suitable for you. Furthermore, other liquids may damage the nebulizer, atom-
izer and accessories.
•Always use close adult supervision when administering a treatment to a child. NEVER
LEAVE CHILDREN UNATTENDED WITH THIS DEVICE.
•This device is not intended to be used by persons with restricted physical (e.g. activity limi-
tations, disabilities), sensory (e.g. insensitivity to pain) or mental abilities (e.g. mental dis-
orders) or persons lacking the required experience or knowledge for safe operation of the
device, unless closely supervised by a person responsible for their safety. NEVER LEAVE
THOSE WHO REQUIRE CLOSE SUPERVISION UNATTENDED WITH THIS DEVICE.
•To reduce the risk of increased bacterial growth, infection, illness, injury from contamina-
tion or other health risks always follow all hygiene instructions (see section 8.1 “Cleaning
and Disinfection”) when using this unit. Unplug the unit before disinfection or cleaning the
unit.
•Before each use, check that the hose connections on the nebulizer [7] and the atomizer
[10] are tight. Use of an improperly assembled nebulizer could diminish or prevent ade-
quate delivery of medication and the effectiveness of the treatment.
•Never block the air vents of this product or place it on a soft surface, such as a bed or
couch, where the air vents may be blocked. Keep the air vents free of lint, hair and similar
substances. Blockage could cause excessive heat buildup.
•Never cover the unit during operation. Covering could cause excessive heat buildup.
•Essential oils, cough medicines, solutions designed for gargling, and drops for application
to the skin or for use in steam baths are wholly unsuitable for inhalation using a nebulizer.
These substances are often highly viscous and can impair the correct function of the de-
vice and hence affect the effectiveness of its application in the long term.
•Check the atomizer, mouthpiece, and any other optional components before each use.
Dirty, damaged and/or worn parts may diminish or eliminate the effectiveness of treat-
ment. Replace damaged and/or worn parts. Dirty parts must be cleaned following all hy-
giene instructions. (see section 8.1 “Cleaning and Disinfection”)
•Never use an atomizer with a clogged nozzle. If the nozzle is clogged, the aerosol mist will
be reduced, altered or stopped, which will diminish or eliminate the effectiveness of the
treatment.
•Follow the instructions for changing the filter contained in these instructions (see section 7
“Changing the Filter”). A contaminated filter can impair the correct function of the device
and represents a risk to your health.
NOTICE
•Use the nebulizer intermittently; the maximum operating period is 30 minutes. The nebu-
lizer must stand unused for 30 minutes between uses.

12
Before use, check that the device is functioning correctly. To do this, switch on the nebu-
lizer (including the connected atomizer, but without any medication) for a short time, e.g.
for 5 – 10 seconds. If the atomizer [10] expels air, the device is functioning correctly.
1 Inserting the Nozzle Attachment
•Make sure the on/off switch [3] is in the off position.
•To open the atomizer [10], twist the upper section counter-clockwise against the
medicine container [12]. Place the nozzle attachment [11] in the medicine container
[12].
2 Filling the Atomizer
•Pour an isotonic saline solution or the medicine directly into the
medicine container [12]. Do not overfill! The maximum recom-
mended fill level is 6 ml !
•If the required volume of medication is less than 2 ml, top up this
quantity to at least 4 ml using an isotonic saline solution. Viscous
medicines may also require dilution. Strictly follow your physi-
cian’s instruction as to dosage of medication and inhalation time.
3 Closing the Atomizer
•To close the atomizer [10], twist the upper section clockwise against the medicine
container [12]. Ensure the parts are connected correctly.
4 Connecting the Accessories to the Atomizer
•Connect the required accessory part (mouthpiece [13], adult mask [14], child mask
[15] or nosepiece [16]) to the atomizer [10].
•Before commencing treatment, pull the atomizer upwards out of its holder [8].
•Start the nebulizer using the on/off switch [3].
•A flow of spray from the atomizer indicates that the device is functioning properly.
5 Treatment
•Start your treatment according to your physician’s instructions.
•Inhale the nebulized medicine deeply.
NOTE: During treatment, hold the atomizer straight (vertical), otherwise the atomiza-
tion does not work and might impair the treatment
What Special Steps Should be Taken for Babies and Children?
- For babies, the mask should cover the mouth and nose to guarantee effective
inhalation.
- For children, the mask should also cover the mouth and nose.
6 End of Inhalation:
•If the spray becomes intermittent or if the sound changes during inhalation, you can
end the treatment.
•After treatment, turn off the nebulizer at the on/off switch [3]. Unplug the power cord
from the outlet.
•After use, replace the atomizer [10] in its holder [8].
7 Clean the Device.
For information about this, see section 8.1 “Cleaning and Disinfection”.

13
NOTE: The device also features a thermal fuse that turns off the nebulizer in the event of
overheating. Should this occur, proceed as follows:
•Switch off the appliance
•Unplug the power cord from the outlet.
•Wait for a minimum of 30 minutes until the device has completely cooled down.
•Plug the power cord back into the outlet.
•Then switch the device on again, and observe whether it makes any abnormal noises. If
there are no unusual noises and smells, you can use the device again. Otherwise, contact
your customer service team. (Please see warranty for service contact)
7. Changing the Filter
DANGER:
•Always unplug this product immediately after use and make sure it is not acces-
sible to children.
CAUTION:
•Do not operate the nebulizer without a filter.
•Check the filter at regular intervals (at the latest after every 10th use). If the filter
is very contaminated or blocked, replace the used filter. If the filter has become
wet, it must also be replaced with a new one.
•Do not try to clean and re-use the used filter!
To replace the filter, proceed as follows:
•First switch off the unit and unplug it.
•Allow the unit to cool down.
1 Pull the filter cap [6] forward and turn it.
•NOTE: If the filter remains in the unit when the cap is removed, remove the filter
from the unit using tweezers or the like.
2 Replace the filter cap [6] containing the new filter.
3 Check that the filter cap is firmly in place.

14
8. Maintenance
8.1 Cleaning and Disinfection
DANGER
•Never touch the unit, including the power cord or plug, with wet hands while it is plugged
in. This can cause an electric shock.
•Before you clean the unit, always switch it off, unplug it and let it cool off.
WARNING
•To reduce the risk of increased bacterial growth, infection, illness, injury from contamina-
tion or other health risks always follow all hygiene when using this unit. Unplug the unit be-
fore disinfection or cleaning the unit.
•For hygienic reasons, the atomizer [10] and the accessories must be cleaned after each
treatment and disinfected after the last treatment of the day. Prior to initial operation, the
atomizer [10] and accessories must be cleaned and disinfected as well.
•Ensure that all parts are thoroughly dried, following every cleaning and disinfection. Re-
sidual moisture or water droplets can result in an increased risk of bacterial growth and
therefore represent a risk for your health.
•If several different medicines are to be inhaled in succession during therapeutic use, the
atomizer [10] must be rinsed under warm tap water after each use before using different
medication.
•The atomizer and accessories must not be mechanically cleaned with brushes or the like
as this can cause irreparable damage and may render the treatment ineffective.
•For additional requirements regarding the necessary hygienic preparations (hand washing,
handling of medicines or inhalation solutions) in high-risk groups (e.g. cystic fibrosis pa-
tients), contact your physician.
•When cleaning, ensure that all residues are removed. Never use any substances that may
be poisonous upon contact with the skin or mucous membranes, or when swallowed or
inhaled. When in doubt, contact your physician for advice.
NOTICE:
•When disinfecting the atomizer, mouthpiece and nosepiece with the “Boiling water
method”, avoid that the parts come into contact with the hot base of the pot.
•The face masks and air hose must not be placed in hot water!
•Ensure that no water penetrates inside the unit! Never immerse the unit in water or any
other liquid, otherwise liquid can enter it and cause damage.
•Do not put the device or accessories in a dishwasher, washing machine or tumble dryer.
•Do not use any aggressive solvents, cleaning agents, detergents or any other strong
chemicals to clean the device.
•Unclogging the small holes of the nozzle should only be done carefully from the inside out
using a needle or pin. Replace if necessary.

15
a) Atomizer and accessories
The atomizer [10] and accessories are in-
tended for multiple use. Please note that
cleaning requirements differ according to
the different application areas.
Preparation
•Directly following each treatment, all
parts of the atomizer [10] and any acces-
sories used must be cleaned of medicine
residues and contamination.
•Remove the accessory part (mouthpiece
[13], adult mask [14], child mask [15] or
nosepiece [16]) from the atomizer.
•Dismantle the atomizer by twisting the
upper section counter-clockwise against
the medicine container [12].
•Remove the nozzle attachment [11] from
the medicine container [12].
•The atomizer is later reassembled in the
reverse order.
Cleaning
Before you clean the unit, always switch it
off, unplug it and let it cool off.
•The atomizer and any accessories used,
(mouthpiece, nosepiece, adult mask,
child mask) must be cleaned thoroughly
after each treatment with hot tap water.
Carefully dry all components using a soft
cloth. Once all components are fully dry,
reassemble the atomizer and place the
components in a dry, sealed container or
carry out disinfection.
Care of the Hose
The air hose must not to be washed as
only filtered air passes through it. Keep the
outer surface of the air hose dust-free by
wiping it regularly.
Depending on the environmental condi-
tions, condensation may collect in the
hose. To prevent against bacterial growth
and to ensure effective therapeutic use,
this moisture must always be removed.
To remove moisture from the hose, pro-
ceed as follows:
•Remove the hose [9] from the atomizer
[10].
•Leave the hose plugged into the nebu-
lizer [7].
•Run the nebulizer until all the moisture
has been displaced by the flow of air.
•In the event of significant contamination,
replace the hose.
Disinfection
Follow the steps listed below carefully in
order to disinfect your atomizer and the ac-
cessories. It is necessary to disinfect the in-
dividual components after the last time they
are used - each day at the latest.
•First clean the atomizer and the accesso-
ries as described under “Cleaning”.
•You can disinfect the atomizer and any
accessories used, with the vinegar or
boiling water method.
Vinegar Method
For disinfection, using the vinegar method,
you will require a small amount of clear
vinegar and distilled water.
•Make up a vinegar solution consisting of
¼ clear vinegar and ¾ distilled water.
Ensure that there is sufficient solution to

16
fully immerse all parts, such as atomizer,
mouthpiece, nosepiece, adult mask and
child mask. The air hose [9] must not be
placed in the vinegar solution.
NOTE: Dismantle the atomizer by twist-
ing the upper section counter-clockwise
against the medicine container [12].
•Submerge the components in the vinegar
solution for 30 minutes.
•Remove the components and discard the
solution. Rinse the components with wa-
ter.
Boiling Water Method
•The atomizer should be taken apart and
placed together with the mouthpiece and
nosepiece in boiling hot water for at least
10 minutes. Avoid the parts coming into
contact with the hot base of the pot. The
face masks [14 and 15] and air hose [9]
must not be placed in hot water! The face
masks [14 and 15] must be disinfected
using the “Vinegar method”
Drying
•After disinfection carefully dry all compo-
nents using a soft cloth.
•Place the individual components on a
dry, clean and absorbent surface and al-
low to dry completely (for at least 4
hours).
•Once all components are fully dry, reas-
semble the atomizer [10] and place the
components in a dry, sealed container.
b) Nebulizer
•Clean the casing of the main unit daily by using a damp cloth, to which you can apply a
little detergent if necessary.
•Keep the unit dust-free.
8.2 Material Resistance
As with all plastic components, frequent use and cleaning of the atomizer and the accesso-
ries can lead to a certain amount of wear. Over time, this can alter the aerosol properties
and may eventually affect the efficiency of therapy. We therefore recommend that you re-
place the atomizer and the accessories at the latest once a year. In case of frequent use
earlier replacement might be necessary.
8.3 Storage and Caring for the Device
DANGER
•Always unplug this product immediately after use and make sure it is not accessible to
children.
•The unit and power cord must be protected and stored away from sources of heat. (e.g.
stoves, heating radiators).
WARNING
•The IH 20 Nebulizer must only be used with suitable Beurer atomizers and with the appro-
priate Beurer accessories. The use of other atomizers and accessories may damage the
device and can impair therapeutic efficiency.

17
NOTICE
•Protect the unit or the parts from strong impacts such as dropping the unit on the floor.
•Do not leave medication inside the atomizer when not in use.
•Store the device and accessories in a location protected against environmental influences,
i.e. in a safe and dry location. Do not expose the device and accessories to direct sunlight,
extreme hot or cold temperatures or humidity (e.g. in the bathroom). The device and ac-
cessories must be stored in the specified ambient conditions (see section 11 “Technical
Specifications” for details).
•Protect the unit against dust, dirt and moisture. This may damage the device.
The accessories can be safely stored in the accessories compartment [4].
Store the device in a dry place.
8.4 Disposal
NOTICE
•Dispose of the device, components and optional accessories according to applicable local
regulations. Unlawful disposal may cause environmental pollution.

18
9. Troubleshooting
Problem/Question Possible Cause/Remedy
1. Check if plug is properly fit into the power outlet.
Device doesn’t operate.
2. Check if the on/off switch [3] is in on position.
1. Too much or too little medicine or liquid in the atomizer.
Minimum: 0.5 ml, Maximum: 6 ml.
2. Check that the nozzles are not blocked
If necessary, clean the nozzle (e.g. by rinsing).
Use the atomizer again or replace if necessary.
3. The atomizer is not held in an upright position
4. The medicine solution is unsuitable for nebulizing (e.g. too
thick).
The medicine solution should be specified by a physician.
5. Check the hose for blockage or air leakage. Replace as
needed.
6. Check that the atomizer is properly assembled and not dam-
aged. If there is any damage, replace if necessary.
7. Check if filter is dirty; replace filter if necessary.
The atomizer produces
little or no aerosol
8. Make sure that the air hose and other components are properly
attached.
Output is too low Hose twisted, filter blocked, too much inhalation solution
Some inhalation solu-
tion remains in the at-
omizer.
This is normal and occurs for technical reasons. Stop inhalation if
the sound of the atomizer changes.
10. Frequently Asked Questions
Which medicines are
suitable for inhalation?
Consult your physician.
Essentially all medicines that are suitable and approved for
device-based inhalation can be inhaled.
Inhalation with the mask
takes longer
This is for technical reasons. Less medicine is inhaled per breath
through the mask holes than using the mouthpiece. The aerosol is
mixed with ambient air through the holes in the mask.
Why do I need to replace
the atomizer at regular in-
tervals?
There are two reasons for this:
1. To provide for a therapeutically effective particle spectrum, the
nozzle holes should not exceed a certain diameter. Due to the
thermal and mechanical stresses, the plastic is subjected to a
certain amount of wear. The nozzle attachment [11] is particularly
susceptible to this wear. This can also cause changes to the
droplet composition of the aerosol, which has a direct effect on
the effectiveness of the treatment.
2. It is also recommended that you replace the atomizer - at the lat-
est - once a year. In case of frequent use earlier replacement
might be necessary.
Does each person need
their own atomizer and
accessories?
This is absolutely necessary for hygienic reasons.

19
11. Technical Specifications
Dimensions (WHD) 11.6” x 7” x 3.7” (295 x 178 x 94 mm)
Weight 4 lbs (1.8 kg)
Operating pressure approx. 12psi (0.85 bar)
Atomizer filling volume max. 6 ml
Medicine flow rate approx. 0.2 ml/min.
MMAD 5 μm
Sound pressure max. 60 dBA (acc. to DIN EN 13544-1:2004-09 section 26)
Motor type Piston pump
Power 120 V AC, 60 Hz
Power wattage 132 VA
Power cord length 4.92 ft. (1.5 m)
Housing material ABS / PP
Accessories atomizer, air hose, mouthpiece, nosepiece, adult mask, child
mask, filters (5pcs)
Air flow approx. 6.5 lpm
Operating conditions Temperature: 50°F to 104°F (+10 °C to +40 °C)
Relative humidity: 10 % to 95 %
Atmospheric pressure 700 to 1,060 hPa (hectopascal)
Storage and transport
conditions
Temperature: -5.8°F to 158°F (-21 °C to +70 °C)
Relative humidity: 10 % to 95 %
Atmospheric pressure: 500 to 1,060 hPa (hectopascal)
Classification Application part Type BF
Protection class Internal power supply, IPX0, no AP or APG
Subject to technical modifications.
Symbols Key
The following symbols are used on the packaging and on the nameplate of the device and ac-
cessories.
WARNING! Read safety information to avoid serious injury.
Type BF applied part
Device in protection class 2
I On
O Off
SN Serial number
30 ON/
30 OFF
30 minute operation, followed by a 30 minute break before operation continues.
No AP/APG Do not operate the unit in the presence of flammable gases, oxygen and aerosol
spray products.

20
Spare Parts
Accessory Replacement Kit # 601.71 for Beurer Nebulizer IH20 # 601.70
Content Material
1 x mouthpiece PP
1 x nosepiece PP
1 x adult mask PVC
1 x child mask PVC
1 x atomizer PS / PP
1 x compressed-air hose PVC
10 x filters PP
For replacement accessory please contact Beurer customer service at 1-800-536-0366 or
All measurements were obtained with a sodium
chloride solution using a laser diffraction method.
This diagram may not be applicable for suspen-
sions or highly viscous medicines. More informa-
tion can be obtained from the relevant medicine
manufacturer.
NOTE:
If the unit is used outside of the specifications, proper function is no longer warranted! We
reserve the right to make technical changes to improve and further develop the product.
This device and its accessories comply with European standards EN60601-1 and
EN60601-1-2, EN13544-1, as well as UL1431 and are subject to special safety measures
in terms of electromagnetic tolerance.
Note that portable and mobile RF communication equipment can have an effect on this
unit. For detailed information, contact customer service. The unit conforms to the require-
ments of the European Directive for Medical Products 93/42/EC, the MPG (German Medi-
cal Units Act).