brewer Assist PRO 7500 User manual

Document #101522 RevBPrinted in USA © 2017
Installation & Operation Manual
Important Information............................................2
Safety Information.................................................3
Component Identication ......................................3
Specications ........................................................4
Installation .............................................................5
Operation ..............................................................8
Table Adjustments...............................................14
Maintenance........................................................19
Warranty..............................................................20
Accessories.........................................................20
The Brewer Company, LLC
N88 W13901 Main Street, Suite 100
Menomonee Falls, WI 53051
P 1.888.Brewer.1
F 262.251.2332
www.brewercompany.com
Model 7500
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 2 Document #101522 RevB
Service
If you require assistance with the installation or operation
of your Brewer
table,
call the Brewer Customer Service
Department at (1-888-Brewer1). Our trained staff will attempt
to assist you in correcting the problem directly over the phone.
If service is required, a factory authorized technician will be
sent to your location.
Please ll in the following information for use when calling the
Brewer Company or your distributor with questions regarding
your unit. See Figure 1 for model and serial number location.
Date of Purchase _________________________
Serial Number ____________________________
Model Number ___________________________
Authorized
Dealer Name ____________________________
Dealer
Phone Number ___________________________
Dealer Address ___________________________
IMPORTANT INFORMATION
General
Read and understand all operating instructions, safety
information, and maintenance requirements contained
in this manual prior to operating the table. Become
familiar with all of the table functions before using it with
a patient.
The Brewer AssistPRO Power Procedure Table is
designed to provide positioning and support of patients
during general examinations conducted by qualied
medical professionals.
The AssistPRO is primarily used in examination rooms
for general examinations and minor procedures. The
wide variety of positions provided by four powered
motions, adjustable legrest, adjustable headrest, stirrups
and treatment pan, create a safe and convenient patient
positioning table for most examinations and procedures
performed in the doctor’s ofce.
The table is easily adjusted to a wide variety of positions
using the convenient foot control or hand pendant
provided with the table. The table height can be
adjusted from 19” to 40”. The table seating surface can
be moved from the normal horizontal position to an angle
of 30°. The backrest can be adjusted from a at lying
down position, to an 85° angle for seating. The legrest
can be adjusted from a vertical position for seating, to a
horizontal position for laying.
The AssistPRO is also capable of storing four preferred
positions into the tables memory. Utilizing an intuitive
two step programming method, the user can recall any
programmed position with the touch of a button.
In addition to the electronically controlled positions
noted above, several manual adjustments are available:
1. The legrest can be manually adjusted from a sitting
to a kneeling position.
2. With the table in the horizontal position, the legrest
can be extended an additional 5” for tall patients.
3. Stirrups may be manually extended and positioned
innitely within the lateral range of travel.
4. A removable treatment pan is stowed beneath the
front of the seat and can be pulled out for use during
examinations.
5. The headrest can be adjusted to a variety of patient
positions using three clamp action locking levers.
6. The backrest incorporates a self-storing paper roll. Figure 1. Model and Serial Number Location
Model &
Serial Number
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 3 Document #101522 RevB
Safety Information
The primary concern of The Brewer Company is
that the equipment is operated and maintained
with the safety of the patient and healthcare staff in
mind. To ensure safe and reliable operation:
•Read and understand all instructions in this
manual before attempting to install or operate
the unit.
•Ensure that appropriate personnel are in-
formed on the manual contents. This is the
responsibility of the purchaser.
•Ensure that this manual is located near the
table, or if possible permanently afxed to the
table.
SAFETY INFORMATION
COMPONENT IDENTIFICATION
Leg Section
Headrest
Backrest
Stirrups
Seat
Treatment
Pan
Headrest Linkage
Assembly
Programmable
Foot Control
Programmable
Hand Pendant
Figure 2. Component Overview
DANGER Indicates a hazardous
situation that, if not avoided, will
result in death or serious injury.
WARNING indicates a hazardous
situation which, if not avoided, could
result in death or serious injury.
CAUTION indicates a hazardous
situation which, if not avoided, could
result in minor or moderate injury.
DANGER
WARNING
CAUTION
This is the safety alert symbol. It is used to
alert you to potential physical injury hazards.
Obey all safety messages that follow this
symbol to avoid possible injury or death.
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 4 Document #101522 RevB
SPECIFICATIONS
Power Cord
• Length...................Extends 70in. (Minimum) from table.
• Type..............................................................Detachable
• Conductors...................................18 AWG / 3 conductor
• Grade.............................................Hospital Grade Plug
Type of protection against electrical shock
Class......................….............................Class 1 Grounded
Patient Applied Parts................................................Type B
Type of protection against
ingress of water
• Table…….........................................................Ordinary
• Foot Control...........................................................IPX1
Paper Roll Sizes
• Regular.........................................................18 in. x 3.5 in.
• Long.............................................................21 in. x 3.5 in.
Certications*.....UL60601-1; CAN-CSA C22.2 No. 601.1
Operating Conditions
• Temperature Range.......................................65
ºF to 85ºF
• Humidity Range..............................................10% to 90%
Transportation and Storage
• Temperature Range.....................................-20°F to 85°F
• Humidity Range..............................................10% to 90%
Weight of Table...................................................550 lbs.
Load Rating (maximum)
• Seat/Back...........................................................450 lbs.
• Legrest (Flat & Kneeling) ...................................450 lbs.
• Headrest.............................................................100 lbs.
Back Section Range.............0° (horizontal) to +85° ± 5°
Table Top Height Range.........19 ± 0.5 in. to 40 ± 1.0 in.
Leg Section Range..........0° (horizontal) to 90° (vertical)
Table Top
Tilt Range............... 0° (horizontal) to +30° ± 5° (foot up)
Dimensions
• Upholstered Headrest...............13 in. wide x 11 in. high
• Upholstered Top..................28.0 in. wide x 44.5 in. long
• Upholstered
Leg Rest ..…………………16.5 in. wide x 11.25 in. high
• Overall Length................................................69 in. long
• With Headrest/
Footrest Extended..........................................90 in. long
Electrical Rating
..................115 VAC nominal, 60 HZ, 5.8 amps maximum
Duty Cycle (10%)
................. 2 minutes on / 18 minutes off (motor run time)
WARNING
Explosion/Fire Hazard
Do not use this table in an explosive or oxy-
gen rich atmosphere. Failure to do so may
result in serious personal injury or death.
WARNING
Shock Hazard
When performing cauterization or similar
treatment, the patient must be insulated from
the metal portions of the table by non-con-
ductive material. Failure to do so may result
in electrical shock.
*Classied in the United States and Canada
per the standards listed above.
Certications
This product has been evaluated with respect to electrical
shock, re, and mechanical hazards only in accordance with
UL60601-1; CAN/CSA C22.2 No. 601.1
NOTICE
Electromagnetic Interference (EMI)
This product is designed and built to minimize electromagnetic
interference with other devices; however, if interference is
noticed between another device and this product, remove the
interfering device from the room or plug this product into an
isolated circuit.
NOTICE
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 5 Document #101522 RevB
Overview
Perform the following sequence in order when setting up
the table:
• Uncrating
• Leveling the Table
• Installing the Power Cord
• Installing the Foot Control
• Installing the Hand Pendant
Uncrating
1. Using a 1/2” open-end wrench, or socket, unbolt the
table from the wooden shipping skid by removing the
four bolts located underneath the table’s base; two
on each side.
2. Remove the table from the shipping skid by sliding it
off the front of the skid. Be sure to lift at the correct
points as indicated in Figure 3. Position the table in
the desired room location.
3. Remove packing tape from stirrups (Figure 4). Pull
out treatment pan slide and remove packing tape
from the treatment pan assembly components.
Return treatment pan slide to its stowed position.
INSTALLATION
Figure 3. Table Lift Points
Figure 4. Stirrups and Treatment Pan
DO NOT lift at these points!
Lift table at these points
Do Not Lift by Stirrups or
Treatment Pan
NOTICE
Inspect Carton and Contents
Inspect all boxes and contents for
damage. Report any damage to the
carrier immediately.
No Sharp Tools
To avoid damaging the table’s
upholstery or painted surfaces, DO
NOT use a knife or other sharp object
to open the packaging.
NOTICE
Optimum Table Performance
For optimum table performance, allow the table to
reach room temperature before operating.
NOTICE
Table weight up to approximately 550 lbs.
NOTICE
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 6 Document #101522 RevB
Leveling the Table
A leveling screw pad (Figure 5) is located in four places
under the table’s base.Adjust the leveling pads to
achieve a solid, level installation.
Installing the Power Cord
1. Remove the power cord from the shipping box.
2. Plug the female end of the power cord into the
receptacle on the table (Figure 6). Make sure
the plug end shape is correctly oriented to the
receptacle shape. Plug the male end of the power
cord into a 115 Volt grounded receptacle.
Installing the Foot Control
1. Remove the foot control from the shipping box.
2. Insert the cord into the foot control receptacle on
the table (Figure 6). Make sure the tab on the plug
is aligned with the groove in the receptacle, and the
plug is fully seated.
3. Thread the locking collar onto the receptacle.
4. Make sure the Controls On/Off switch located on the
backrest shroud is turned to the “ON” position. See
Figure 7.
Figure 5. Leveling Screw Pad (located under table)
Figure 6. Power Cord & Foot Control Cord Receptacles
Figure 7
INSTALLATION (CONTINUED)
Locking Collar
Receptacle
Foot Control and Hand
Pendant On/Off switch
Prevent Leveler Damage
Levelers must remain in the table and be fully engaged a
minimum of (3) turns. If levelers are not fully engaged in the base,
damage to the threading in the base and/or leveler can occur.
NOTICE
NOTICE
Conrm Voltage and Plug Type
See Specications Page for electrical rating of this unit.
Thethree-prongedgroundingplugonthetablepowercord
mustbepluggedintoamatchingthree-pronged,grounded,
non-isolated, correctly polarized 115 VAC receptacle.
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 7 Document #101522 RevB
Installing the Hand Pendant
1. Remove the hand pendant from the shipping bag.
2. Insert the cord into the hand pendant receptacle
on the backrest shroud. (Figure 8). Make sure the
tab on the plug is aligned with the groove in the
receptacle, and plug is fully seated.
3. Thread the locking collar onto the receptacle.
Figure 8
INSTALLATION (CONTINUED)
Receptacle
Hand
Pendant
Cord
Locking
Collar
WARNING
Shock Hazard
After installing or moving table, make sure
there are no damaged or exposed wires that
could result in electrical shock.
Figure 9
Power On/Off switches
1. The table is equipped with two On/Off switches that
disrupt the main power to the table.
2. The switches are located on the upper left and right
hand sides of the backrest shroud. See Figure 9.
3. In the “On” position the switch is lighted green. Both
switches must be in the On, lighted position for the
table to operate.
The foot control and hand pendant are moisture resistant
and run on low voltage. To avoid temporary or permanent
damage to the controls, do not immerse in water or liquids.
NOTICE
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 8 Document #101522 RevB
Leg
Down Table
Down
Figure 10
Controls
1. See Figure 10 for the pedal and button identica-
tions on the foot control
2. See Figure 11 for the button identications on the
hand pendant
OPERATION
Program Recall
Mode
Stop
Program #3
Program #4
Tilt
Down
Back
Down
Program #2
Program #1
Program
Set
Leg
Up Table
Up
Tilt
Up
Back
Up
Unintended Operation
To avoid accidental operation, keep foreign objects away
from controls. Be sure the controls are not positioned below
the legrest or headrest linkage.
NOTICE
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 9 Document #101522 RevB
Leg Up
Leg Down
Tilt Down
Tilt Up
Program Recall #2
Program Recall #4
Back Up
Back Down
Table Down
Table Up
Program Recall #1
Program Recall #3
Figure 11
OPERATION (CONTINUED)
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 10 Document #101522 RevB
OPERATION (CONTINUED)
Figure 11. Treatment Pan
Figure 12. Headrest Handles
Treatment Pan
1. To expose the treatment pan for use, grasp the tray
handle and pull until tray is fully extended (Figure 11).
2. The pan can be easily removed for cleaning with
the tray fully extended. Fully seat pan into tray when
clean.
3. Push the tray handle as far in as it will go to fully stow
the treatment pan after use. A light clicking sound will
be heard just before the pan reaches its fully stowed
position. (The light clicking sound is created by a
switch that allows the legrest to move upward).
Headrest Positions
1. To move the headrest, rotate the three handles to
their open position. Move the headrest to its desired
location. Rotate the 3 handles to their locked position
as shown in Figure 12.
2. Each section of the headrest can be adjusted
independently by rotating the handle to the open
position, moving the desired section, and then
rotating the handle to the closed position.
NOTE: Make sure all three handles are locked. A handle
is locked correctly when it is pushed past center and
the force of the clamping action pulls the handle into the
locked position and holds it there.
Headrest Handle
(Locked)
Upper Handle
(Locked)
Lower Handle
(Locked)
Tray
Treatment
Pan
Make sure headrest is securely locked into position before
use. Failure to have the three handles correctly locked could
result in headrest falling during a procedure which could result
in serious personal injury.
CAUTION
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 11 Document #101522 RevB
Figure 14. Legrest Kneeling Position
Legrest Kneeling Position
The legrest can be rotated to a kneeling position (Figure
14).
1. Grasp the legrest cushion on both sides.
2. Rotate the cushion upward until rotation stops.
3. Slide the legrest downward and rotate the cushion
downward to engage the stop. Two height positions
are provided.
Cam positioned in slot
Lift up over-center
1 2 3
Legrest Extension
In the horizontal position, the legrest can be extended 5”
to increase the overall length of the table (Figure 13).
Lift up on the legrest and pull outward to extend the
legrest. To return the legrest to the shortest position, lift
up on the legrest and push fully inward.
Figure 13. Legrest Extension
To prevent damage to the unit, the legrest must be returned
to the shortest position prior to lowering the legrest to the
vertical position.
NOTICE
OPERATION (CONTINUED)
Make Sure Legrest is Engaged with Stops
Make sure the legrest cushion is fully engaged with
the stops on both sides. When properly engaged, the
legrest cushion is parallel to the ground.
NOTICE
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 12 Document #101522 RevB
Adjustable Stirrups
Operating the StirrupS
1. To operate the stirrups, grasp end of stirrup and pull straight
out of the table. Lift up slightly while sliding the stirrup out
(Figure 15).
NOTE: Applying any downward pressure when pulling stirrup
from table may make it difcult to move. Resistance is normal
and is part of the stirrup friction locking mechanism
2. Unfold stirrup upward to the fully open position.
3. Adjust the stirrup to the desired length by sliding it in or out.
Stirrup LateraL adjuStment
1. Lift end of stirrup and then rotate outward to
desired position (Figure 16).
2. When the desired position is achieved, lower the stirrup to
engage the lateral locking mechanism.
3. Check that lateral locking mechanism is engaged by
attempting to rotate stirrup assembly without lifting on the
stirrup end.
StOring the StirrupS
1. To store the stirrups in the table, grasp end of stirrup and pull
straight out to its full extension (Figure 17).
2. Fold stirrup down against bar, then rotate it to the inner most
lateral position and slide it back into the stowed position.
Figure 18. Afx No Attachments!
Use Stirrup Only as Designed
Do not use any part of the stirrup for attachment of any device (Figure 18).
Warranty will be voided. If knee crutches are required for use, order mounting
adapters, clamps, and rails shown in the Accessories Section.
NOTICE
Figure 15. Operating the Stirrups
Figure 16. Stirrup Lateral Adjustment
Pull Stirrup out, then unfold
Lateral
Locking
Mechanism
Figure 17. Storing the Stirrups
OPERATION (CONTINUED)
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 13 Document #101522 RevB
Figure 19. Paper Roll
Slot
Paper
Paper Roll
To change a paper roll, insert the paper through the slot
in the backrest shrouding and set the paper roll in the
holder (Figure 19).
OPERATION (CONTINUED)
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 14 Document #101522 RevB
TABLE ADJUSTMENTS
Figure 20. Table Height Positioning
Table Height Control
Table Height
The table top can be adjusted to any height between 19”
to 40”. (Figure 20)
1. To raise the table, depress and hold the Table/Up
function until the desired height is achieved.
2. To lower the table, depress and hold the Table/Down
function until the desired height is achieved.
WARNING
Area Not Clear Hazard
Before initiating movement keep personnel and equipment
clear of table to avoid personal injury or damage to the equip-
ment.
Actuator Travel Limit Protection
Each control function will automatically stop moving when
either the UP or DOWN travel limit is reached. If the foot pedal
is pressed with travel at its limit, the actuator will not operate.
Sensors prevent the actuator motor from operating when travel
limit has been reached, preventing wear of the actuator.
NOTICE
Avoid Continuous Operation
If the table is operated continuously causing the control
to exceed its allowable operating temperature, the control
will shut off. If normal operation ceases, do not attempt to
operate the table. Allow the table to cool for 10-15 minutes
before attempting to use again. The control will automatically
reset when it has reached a safe operating temperature.
NOTICE
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 15 Document #101522 RevB
TABLE ADJUSTMENTS (CONTINUED)
Tilt Function
The table top can be tilted to any angle between
horizontal and +30° (foot end of table raised).
(Figure 22).
1. To tilt the foot end of the table upward, depress and
hold the Tilt/Up function until the desired angle is
achieved.
2. To tilt the foot end of the table downward, depress
and hold the Tilt/Down function until the desired
angle is achieved. Figure 22. Tilt Function
Figure 21. Adjusting Backrest
Backrest Position Control
Figure 23. Do Not Allow Headrest to Contact Objects
Back Positions
The back can be adjusted to any angle between
horizontal and 85°. (Figure 21)
1. To adjust the back from a chair (upright) position to a
horizontal position, depress and hold the Back/Down
function until the desired angle is achieved.
2. To adjust the backrest from a horizontal position to a
chair (upright) position, depress and hold the Back/
Up function until the desired angle is achieved.
Tilt Function Control
Back Actuator-Push Only
The back is positioned by a “push only” actuator that raises
the back. When lowered, the actuator retracts and gravity
causes the back to lower.
NOTICE
Headrest Linkage Clearance
Use care when tilting the table to avoid contact between
the headrest linkage and the oor or other objects.
(Figure 23).
NOTICE
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 16 Document #101522 RevB
Figure 24. Legrest Positions
Legrest Positions
The legrest can be adjusted from a horizontal position to a
vertical position. Figure 24.
1. To adjust the legrest from a vertical position to a
horizontal position, depress and hold the Legrest/Up
function until the desired angle is achieved.
2. To adjust the legrest from a horizontal position to a
vertical position, depress and hold the Legrest/Down
function until the desired angle is achieved.
Legrest Position Control
TABLE ADJUSTMENTS (CONTINUED)
Treatment Pan Tray Must be Stowed
The legrest will not rise if the treatment pan tray is
extended. A light on the foot control labeled “Pan Out” is
illuminated. Depressing the legrest up function will cause
the light to blink and a double beep alarm to sound. If the
legrest is inoperable, check to ensure that the treatment
pan tray is in its fully stowed position.
NOTICE
Legrest Actuator-Push Only
The legrest is positioned by a “push only” actuator that
raises the legrest. When lowered, the actuator retracts and
gravity causes the legrest to lower.
NOTICE
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 17 Document #101522 RevB
Using the Program Feature
1. The table can be programmed using the hand pendant or the foot control. Up to four different positions can be
programmed.
2. The table is already programmed from the factory with four preset positions. See Figure 25.
Position 1 Position 2 Position 3 Position 4
3. Tapping any of the four blue buttons (see Figure 11), numbered 1 through 4, will move the table to the position
programmed for that button. When the table is in a programmed position, the blue button for that position will be lit
on the foot control.
4. The mode in which the programmed positions are recalled can be set by the end user. The black AUTO-RUN
button (Figure 10) is used to set the recall motion to AUTO-RUN “ON” or AUTO-RUN “OFF”.
• In the AUTO-RUN ”ON” mode, a single instantaneous touch of a blue Program Recall button will recall that
position.
• In the AUTO-RUN “OFF” mode, the blue Program Recall button must be held down continuously until the table
reaches the programmed position.
5. A light on the foot control adjacent to the black AUTO-RUN button will indicate if auto run mode is ON or OFF.
6. To change the program recall mode, depress the black AUTO-RUN button for ve seconds, until two beeps are
heard.
7. When the table is moving in the auto run mode, motion of the table can be stopped by:
a. Tapping the “Stop” button on the foot control.
b. Tapping any pedal or button on the foot control.
c. Depressing any button on the hand pendant.
d. Flipping either of the two lighted switches located at the top of the backrest. See Figure 9.
Figure 25
TABLE ADJUSTMENTS (CONTINUED)
Treatment Pan Tray Must be Stowed
If the treatment pan tray is out, the position recall will not
work if the pre-set position requires the legrest to move
upward. Five beeps will sound, and the “Pan Out” indica-
tor light will ash on the foot control.
NOTICE
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 18 Document #101522 RevB
Programming the Table with the
Hand Pendant
1. Move the table to the desired position using the buttons on the hand
pendant. Simultaneously depress the number “1” and number “2”
buttons for three seconds. Release the two buttons. Depress one of
the four Program buttons (1 thru 4) on the hand pendant to store the
position. Two beeps will sound.
2. Move the table to a new position and repeat Step 1 for the three
additional program positions.
3. To recall one of the four programmed positions, press the program
button (1 thru 4) corresponding to that position
4. To change the AUTO-RUN program recall mode using the hand
pendant, simultaneously depress the number “3” and number “4”
buttons for ve seconds until two beeps are heard.
Programming the Table with the
Foot Control
1. Move the table to the desired position using the pedals on the foot control. Tap the black “Program” button.
The four blue buttons will light for three seconds. Tap one of the blue buttons (1 thru 4). The blue light for that
button will stay lit and two beeps will sound to indicate the position has been stored for that button.
2. Move the table to a new position and repeat Step 1 for the remaining blue buttons to store up to 4 programmed
positions.
3. To recall one of the four programmed positions, tap the blue button (1 thru 4) corresponding to that position.
4. Any of the programmed positions can be changed by repeating Step 1 above.
TABLE ADJUSTMENTS (CONTINUED)
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 19 Document #101522 RevB
Preventative Maintenance Table Care
If the table malfunctions, immediately remove your hand
or foot from the controls, unplug the power cord from the
receptacle, and assist the patient from the table. Service
the table prior to use.
NOTICE
• Inspect table every 6 months.
• Periodically inspect the electrical, foot control, and
hand pendant cords to ensure they are free of cuts
or damage.
• Inspect the mechanical functions to ensure satis-
factory operation.
• Check fasteners to make sure they are present
and tightened securely.
• Lubricate moving parts (such as back, tilt, and
legrest hinges) with a food grade silicon lubricant.
Loose Fasteners and Damaged Components
Failure to perform periodic inspections of the table
could result in equipment damage.
NOTICE
Wipe Spills Immediately!
The upholstery material is resistant to most medicinal type
stains, but may be damaged by solvents and dyes. Remove
any spilled uids from the upholstery immediately.
NOTICE
For the latest in Cleaning Guidelines visit:
www.brewercompany.com
Safety
The table has a built in safety feature which helps avoid injury to the patient or physician when recalling a programmed
position in the Auto-Run “ON” mode:
1. The table will stop moving downward once the legrest (in the stowed position) reaches a height of four inches off
the ground.
2. The legrest will stop lowering to provide one inch clearance between the legrest actuator and the lift column
shrouding.
3. The pelvic tilt will stop raising after moving to an angle of 10 degrees.
WARNING
Shock Hazard
Disconnect power by removing plug from
wall receptacle before servicing table.
MAINTENANCE
TABLE ADJUSTMENTS (CONTINUED)
Shop tables & accessories at
ExamTablesDirect.com

Printed in USA © 2017 20 Document #101522 RevB
ACCESSORIES
Listed below are the accessories which are authorized for use with this table.
Description Order No. Use/Restrictions
Urology Tray w/Hose & Bucket 100253 Conveniently and safely captures uids during urology
procedures.
Side Rails (qty 2) 99501 Used to attach accessories. Can be mounted to seat or
back.
Universal Socket Clamp (qty 1) 99502 Used to mount deluxe articulating knee crutches to ac-
cessory rail.
WelchAllyn IV Light Bracket
(customer installed) 102025 For mounting of WelchAllyn IV light.
Grab Bars (qty 2) 99497 Requires 99501 Side Rails for mounting to seat.
Articulating Arm Board 99499 Requires 99501 Side Rails for mounting to backrest.
WARRANTY
For Brewer Warranty Information:
Visit www.brewercompany.com or contact the Brewer Customer Service Department at (1-888-Brewer1).
Deluxe Articulating Knee Crutches 99504 Allows leg positioning for ease of examination.
Contains one LH and one RH Knee Crutch.
Side Rails (qty 2)
99501 Provides attachment point for Knee Crutch Clamps.
Universal Socket Clamp (qty 1)
99502 Clamps Knee Crutch to Side Rail.
Must order (2) two 99502 Universal Socket Clamps.
The following items are required for the Articulating Knee Crutch accessory:
Shop tables & accessories at
ExamTablesDirect.com
Table of contents
Other brewer Medical Equipment manuals