Bruker D8 Discover Manual

1
Standard Operating Procedure:
Bruker D8 Discover X-Ray Diffractometer
(You MUST be trained by an authorized BSCMC trainer to use this instrument)
Hazards: This unit produces a high intensity X-ray beam. Take all precautions to avoid exposure to the x-rays.
Warning: Beryllium! Do not touch the front window of either detector as they both contain Beryllium. Fumes
of the dust from Beryllium and its compounds can be hazardous if inhaled!
Typical Experiment Sequence:
1. Sign in on the electronic calendar
2. Visually check the diffractometer for instrument status.
3. Open the cooling water valves and turn the x-ray generator ON.
4. Launch the D8Tools software to check the system status and ramp the power up.
5. Launch the operating software:
a. GADDS for use with the HI-STAR area detector
b. Commander for use with the Scintillation detector.
6. Check the x-ray beam intensity
7. Load and align sample
a. Use the Laser-Video System with GADDS and the HI-STAR area detector
b. Use a Z-scan and half intensity method or the Laser-Video System with
Commander and the Scintillation detector.
8. Set up scan and save data.
9. Remove sample, reduce power, turn the generator off and shut off the cooling water
10. Sign out on the Sign electronic calendar Be sure that the calendar accurately
displays the number of hours the XRD was OCCUPIED by your experiment.

2
Procedures for ALL X-Ray Diffraction Experiments on the Bruker D-8 Discover:
1. Sign in on the electronic calendar at https://boisestate.ilabsolutions.com
2. Visually check the diffractometer for instrument status. Check the machine status on
either left or right column before using the machine. The “Ready” and “On” lights
should be on. There might be an “alarm” due to the x-ray generator being OFF. The
“X-Ray ON” lights on top of machine should be on.
a. Left Control Column:
i. STOP button: If hit it switches off the control electronics. The X-Ray source is
turned off and all moving drives will stop instantly. The stop button should only be used
in emergency situations and not for normal shut down of the diffractometer system.
ii. Light button: This button turns ON/OFF the fluorescent tube located at the ceiling
of the enclosure.
iii. Fan button: This button turns the fan located on top of the enclosure ON or OFF.
iv. System Instrument-Key Switch: This key will lock/unlock the system. During normal
operation the switch must be in position “unlocked”.
v. Open Door button: This button must be pressed to open the front door. If the
Open Door button is activated while the tube window is open the X-ray shutter will close
automatically and it will disrupt the experiment.
vi. Ready light: Status of High Voltage Generator. Illuminated orange display indicates that
the high voltage generator is operating. Flashing display indicates the high voltage
generator is ramping up.
vii. On light: Illuminated green display indicates the control electronics are ready.
viii. Alarm light: Illuminated red display indicates a pending system alarm, flashing red light
indicates a pending warning.
ix. Busy light: Illuminated yellow display indicates that a measurement is in progress.
Flashing yellow display means that the system is not ready.
b. Right Control Column:
i. Stop button: Functions the same as the Stop button on left column.
“X-Rays ON”

3
ii. Power OFF button: It switches off the control electronics, the high voltage generator
and all components connected to the AC outlets.
iii. Power ON button: It switches on the control electronics, the high voltage generator
and all components connected to the AC outlets.
iv. High Voltage- Rotary Switch: Used for turning high voltage on or off.
v. Open Door button: Functions the same as the button on the left column
vi. Ready, On, Alarm and Busy lights have same functions as those on the left column.
3. Starting your session
a. Open the cooling water valves (on the wall behind the instrument) and check for any leaks.
b. Turn the x-ray generator ON by flipping the HV switch clockwise for few seconds and this will
automatically set the power to 20kV and 5 mA.
c. Open the D8 tools and heck the system status.
(→Online Status, →Online Refresh)
d. Check for any alarms
(An alarm or warning will prompt a red or orange box)
i. If they appear, click on the red or orange box to be directed to the cause
e. Ramp up the X-ray power(→X-Ray Generator, →Utilities, →Set kV/mA).
i. Set 30 kV, 10 mA and wait 2-3 minutes
ii. Set 40 kV, 15 mA and wait 2-3 minutes
iii. Set 40 kV, 30 mA and wait 2-3 minutes
iv. Set 40 kV, 40 mA
f. Check the status of the drives
i. Are the drives initiated? (Make sure the drives have significant figures (i.e. 23.047 vs.
23.---)) i. If not, initiate them by performing the following:
a. Click on the green box next to the drive to be initiated, then check the
settings of the drive (→Settings)
b. Check the reference position of the drive
Needs
initialized

4
c. Then physically look at the drive to determine which way it needs to go in
order to be at the reference position
i. If it needs to go in the positive direction to fulfill the reference
position, push the find reference up button. Likewise, if the drive
needs to go in the negative direction, push the find reference down
button (→)
NOTE: Be sure to have the mouse on the STOP icon to be able to stop the drive if it looks as
though it will crash (this will require you to watch each drive as it is initiated!)
d. Once the drive has stopped moving, push the adjust button (→)
e. Repeat steps a. and b. for any of the drives that need to be initiated.
ii. Do the drives appear to be in the indicated position? (Look at the values in D8 Tools for
θ1and θ2and compare them to the physical locations of θ1and θ2on the goniometer).
g. Close the D8Tools software.

5
Procedures for X-Ray Diffraction Experiments using the point detector:
4. Launch the operating software:
a. Open the Commander program for use with the point detector
5. Check the x-ray beam intensity (Make sure there is no sample or sample holder that
will interfere with the path of the x-rays)
NOTE: It is imperative that the copper absorbers are in place! If not, the
detector could be damaged!
a. Make sure the following slits are in place:
i. 1.2mm slit in the Goebel mirror exit slit (furthest left)
ii. 6.0mm slit with the TWO copper absorbers in the next slit
b. Set up a detector scan from -1 to 1 degree
i. Select Detector scan from the pull down menu
ii. Make sure that all of the drives are at 0
1. Type a 0 in each drive command window and then click on the Move Drives icon
to move the drives to 0 position.
Göbel
mirror
6.0 mm
slit and 2
Copper
absorbers
1.2 mm
slit
X-Ray
Tube

6
iii. Make sure the y-scale is in CPS
iv. Set the increment to 0.01°/step and the scan speed to 0.2 sec/step
v. Press the icon.
c. Make sure that the max count rate is 65000 CPS or more. If not, please contact an authorized
BSCMC trainer.
d. Make sure the peak is approximately centered about 0. If not, please contact an authorized
BSCMC trainer.

7
6. Load and align the sample (Depending on the sample and the stage, this can be done
using the Laser-Video System or using a Z-scan)
NOTE: Things to be careful of when loading your sample:
•Do not touch the sollar slits
•Be gentle with the laser/video system
oThese are precisely aligned to give you “good” data
•Do not bump the X-ray optics
oA very slight bump could easily result in a loss of 30% of your beam intensity
a. Open the doors to the XRD (Push Open Door button on either side of control column, then open
the window.) or
b. Put your sample on the XYZ table (stage will depend on preference, special instruction may be
required) and position it directly under the camera.
6.a Align the Z-height using the Laser Video System
i. Align the sample to the instrument using the Laser-Video System
NOTE: All of the components including the X-ray beam, camera, and laser have been aligned to aim at a point
which is called universal center of the goniometer. To align your sample surface at this universal center, you
need to use the Manual Controller to move the XYZ table.
ii. Activate the Manual controller and turn the laser on:
1. Press the Direct Mode icon to get to the command prompt window.
2. Type “RC1” and press “Send Command” to give the computer control.
3. Type “OC16,1” and press “Send Command” to turn the laser on.
4. Type “RC0” and press “Send Command” to give command to the manual
controller.
5. Position the stage and sample accordingly.
iii. Understanding the Manual Controller functions. The drives used are as follows (the
numbers on the controller correspond to the following drive numbers):
1. Θ1–X-ray source (aka tube)

8
2. Θ2–Detector (In this case the HI-STAR area detector)
3. φ(phi) –Rotation of the sample about the z-axis
4. χ(chi/psi)–Rotation of the sample about the Eularian cradle
5. X
6. Y
7. Z
8. Camera Zoom (1X ≤ Zoom ≤ 7X)
iv. The Manual controller has two functions:
1. Command (set point) mode (access by pushing the “F1/DRVC”button on the
controller)
a. Enter three commands (Drive #, Instruction, Location)
i. Drive # is where the drive you want to move is called on
ii. Instruction is what you want the drive to do (“Go To” is always
the number 2)
iii. Location is where you want the drive to move to
iv. (i.e. if you want the video zoom to go to 7X, you would type in
“8,2,7” and then press “ENTER”. The zoom will automatically go
to 7X)
2. Manual (video game) mode (access by pushing “SHIFT” then “F1/DRVC” on
the controller)
a. Identify the drive you want to move (1-8, see above)
b. To change the position of the drive, use the arrows (↓or ↑).
c. To change the speed of movement of the drive, use the “+” or “-“buttons.

9
!!!Warning: Make sure that none of the drives will run into each other when they are moved since there are NO
safety switches installed for this purpose. The “Emergency Stop” buttons can be used if
needed but it is preferred to use the ESC/STOP button on the controller is you see a potential problem.
v. Open VIDEO program, Turn on the light in the XRD chamber as needed
vi. Zoom the camera out (zoom = 1) to help easily find your sample in the video window.
vii. Move the X and Y drives to the area of interest on your sample (using the controller in
manual mode).
viii. Move Z to focus the laser on your sample (The laser spot should be in the center of video
window)
1. The laser might be difficult to see on reflective samples. If this is the case, move
Chi/psi to 22.5 degrees (splitting the angle between the laser and the video) to
cause the laser to reflect back into the camera.
Zoom = 1X Zoom = 7X
!!! Warning: Always leave one finger on “ESC/STOP” button when you are using command mode. Press
“ESC/STOP” button immediately if the moving part is going to hit other part.
ix. Once your sample is in focus and the laser is in the center of the video window (centered
on the crosshairs), zoom the camera in (zoom = 7) to fine tune your sample position.
x. Once the laser spot is centered, the sample is aligned and ready to take data (return
chi/phi to 0 (if applicable)).
!!!Warning: You can use the controller to move Θ1and Θ2to the desired 2θ angle. Make sure the detector will
not get hit by the direct beam (as a rule of thumb, always keep Θ1+ Θ2> 15°).

10
1. Back in the Direct Command window in Commander, type “RC1” and press
“Send Command” to give control back to the computer.
2. Type “OC16,0” and press “Send Command” to turn the laser off.
xi. Exit the Direct Command window by closing the prompt window.
6.b Align the Z-height the Z-scan half intensity method.
i. Align the sample to the instrument using a Z-scan and half beam intensity method.
ii. In the Scantype drop down menu, choose “Z”
iii. Set the scan at -1.25mm and end the scan at 1.9 mm
iv. Set the Increment to 0.01°/step and the Scanspeed to 0.2sec/step and click .
v. Note the average max intensity. (i.e. ~ 9750 CPS)
vi. Typically run a fast scan to start, then zoom in on the area of interest, click .
vii. Set the Increment to 0.005°/step and the Scanspeed to 1sec/step and click .
viii. Once the scan is completed, double click on the where the beam intensity is
approximately one half of the maximum intensity. (i.e. 4875 CPS). Once you double
click on the scan, the Z-position is automatically stored in the Z-drive command
window.

11
NOTE: All of the components including the X-ray beam, camera, and laser have been aligned to aim at a point
which is called universal center of the goniometer. To align your sample surface at this universal center, you
need to use the Manual Controller to move the XYZ table.
7. Set up a scan and save data:
a. Set up the scan that is of interest for your data.
b. Unless you are using a script, Commander DOES NOT automatically save your data! Make sure
that you save your data before you attempt to do ANYTHING further with the software!
8. Remove your sample
a. You are responsible for removing your sample as soon as your scan is finished. The Boise State
Center for Materials Characterization is NOT responsible for any samples. If your sample is
left on the instrument, the next user has the right to remove the sample and set it aside. Clean up
the stage and leave the instrument as clean, if not cleaner, then when you arrived to use it.

12
9. End of session
a. Set the power generator to 20 kV and 5 mA by setting each scroll bar to the far left and pressing
the SET icon. Make sure the generator reaches 20kV and 5mA prior to closing out of
Commander.
b. Close Commander.
c. Wait for 5 minutes
d. Turn the X-ray generator OFF by flipping the rotary switch counter clockwise for a few seconds
(Do not shut the main power down)
e. Close the valves for cooling water.
10. Sign out on the iLab electronic calendar
Be sure that the calendar accurately displays the number of hours the XRD was
OCCUPIED by your experiment.
Other manuals for D8 Discover
1
Table of contents
Other Bruker Measuring Instrument manuals

Bruker
Bruker VERTEX 70v User manual

Bruker
Bruker DektakXT Manual

Bruker
Bruker S1 TRACER User manual

Bruker
Bruker DektakXT User manual

Bruker
Bruker minispec LF Series User manual

Bruker
Bruker ultraflex Series User manual

Bruker
Bruker S1 TITAN User manual

Bruker
Bruker CTX User manual

Bruker
Bruker CTX User manual

Bruker
Bruker BMS User manual
Popular Measuring Instrument manuals by other brands

ACOEM
ACOEM Meax DU 01 user manual

Transducer Techniques
Transducer Techniques DPM-3 Operator's manual

YSI
YSI 650 MDS Operation manual

Emerson
Emerson Rosemount CT5100 Reference manual

Harbor Freight Tools
Harbor Freight Tools Cen-Tech 5645 Assembly & operating instructions

LaserLiner
LaserLiner ACL 4C operating instructions