Bruker VERTEX 70v User manual

User Manual
1007253
VERTEX 70v

1st edition 2007, publication date November 2007
© 2007 BRUKER OPTIK GmbH, Rudolf Plank Str. 27, D-76275 Ettlingen, www.brukeroptics.com
All rights reserved. No part of this manual may be reproduced or transmitted in any form or by any
means including printing, photocopying, microfilm, electronic systems etc. without our prior written per-
mission. Brand names, registered trademarks etc. used in this manual, even if not explicitly marked as
such, are not to be considered unprotected by trademarks law. They are the property of their respective
owner.
The following publication has been worked out with utmost care. However, Bruker Optik GmbH does
not accept any liability for the correctness of the information. Bruker Optik GmbH reserves the right to
make changes to the products described in this manual without notice.
This manual is the original documentation for the VERTEX 70v spectrometer.

VERTEX 70v User Manual iii
. . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
T
ABLE
OF
C
ONTENTS
1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Warning Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Delivery Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Site Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Connecting VERTEX 70v to the Power Supply . . . . . . . . . . . . . . . . . . . 11
Connecting VERTEX 70v to a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Connecting VERTEX 70v to the Vacuum Pump . . . . . . . . . . . . . . . . . . 13
Connecting VERTEX 70v to the Purge Gas Line. . . . . . . . . . . . . . . . . . 15
4 Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
External Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Internal Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Optical Path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Switching VERTEX 70v On and Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
QuickLock. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Automatic Accessory Recognition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Performing a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Evacuating and Venting the Spectrometer. . . . . . . . . . . . . . . . . . . . . . . 36
Optimizing the Vacuum Operation of the Spectrometer. . . . . . . . . . . . . 40
Purging the Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Exchanging the Beamsplitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Exchanging the Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Cooling an MCT Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6 Maintenance and Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Evacuating the MCT Detector Dewar. . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Replacing the Laser Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Replacing a defective IR Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Replacing the Fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

iv VERTEX 70v User Manual
Table of Contents
Replacing the Sample Compartment Windows . . . . . . . . . . . . . . . . . . . 69
Cleaning the Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Maintaining the Vacuum Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
7 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Diagnostic Means. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Problem - Possible Cause - Solution . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
A Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
B Consumable Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
C Default Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
D Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
E Connecting VERTEX 70v to PC . . . . . . . . . . . . . . . . . . . . . . 115
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Possible Connection Topologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Selecting Network Addresses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Assigning Network Addresses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Checking the Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
F Electronics and Power Supply . . . . . . . . . . . . . . . . . . . . . . . 125
Electronics Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Power Supply Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
G Firmware Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Updating the Firmware. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Restoring a previous Firmware Version. . . . . . . . . . . . . . . . . . . . . . . . 134
Backing up the current Firmware Version . . . . . . . . . . . . . . . . . . . . . . 135
H Sample Preparation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Sample Preparation Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
I Service Addresses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

VERTEX 70v User Manual 1
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
S
AFETY
1
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
GENERAL INFORMATION
Read the following safety instructions carefully before putting the spectrometer into
operation. Keep this manual in a suitable place for future reference.
Always observe the instructions described in this manual to ensure user safety and to
avoid property damage. Improper use or failure to follow these safety instructions can
result in serious injuries and/or property damage. Any non-observance of the precau-
tions will infringe the intended use (i.e. performing spectroscopic measurements) of the
spectrometer. In this case Bruker Optik GmbH will not assume any liability.
It is the operator’s duty to plan and implement all necessary safety measures and to
supervise their observance. Moreover, the operator must ensure that the spectrometer
is in proper functioning condition. A safe and faultless operation can only be guaranteed
if the spectrometer is transported, stored, installed, operated and maintained properly
according to the procedures described in this manual.
Never remove or deactivate any supporting safety systems during spectrometer opera-
tion. Ensure that objects and/or material not required for the measurement is out of the
spectrometer operating area.
The spectrometer complies with the IEC/EN 61010-1 safety regulations.
Protective Earthing
To avoid personal injuries and/or property damage caused by electrical power, the
spectrometer is equipped with a safety plug. Connect this plug only to a socket outlet
with earthing contact. Make sure that the socket complies with IEC (International Elec-
trotechnical Commission).
Qualified Personnel
Primary installation and all maintenance and repair works not described in this manual
should only be performed by Bruker service personnel. Only authorized operating per-
sonnel that have been briefed about the spectrometer operation and all relevant safety
aspects should operate and maintain (i.e. only maintenance works that are described in
this manual) the spectrometer.
All repairs, adjustments and alignments on any spectrometer component must be per-
formed in accordance with the safety regulations and standards applied in the country in
which the instrument is installed.

2 VERTEX 70v User Manual
1SAFETY
Warning Labels
Correct Usage
The spectrometer and its components should only be used according to the instructions
described in the manual or advised by a Bruker engineer. In case of accessories or
components made by other manufacturers and used in connection with the spectrome-
ter, Bruker does not assume any liability for safe operation and proper functioning.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
WARNING LABELS
When operating the spectrometer you have to observe a number of safety instructions
which are highlighted by various warning labels. This section describes the warning
labels and explains their meaning. All warning labels on the spectrometer must always
be kept legible. Immediately replace a worn or damaged label.
The following warning labels indicate different dangerous situations which may be
caused by improper use of the spectrometer.
Caution - General Hazard
This warning symbol indicates general hazard. Observe the safety
instructions and follow the precautions described to avoid personal injury
and/or property damage.
Caution - Electrical Shock
This warning symbol indicates electrical hazard. The symbol is located near
live parts or on enclosures behind which are live parts that represent an
accidental contact hazard. Never touch these parts. Before removing the
corresponding compartment covers and beginning any maintenance or
repair work, first turn off the mains switch and unplug the main power cable.
Ensure that all live parts do not come into contact with a conductive sub-
stance or liquid. Non-observance of these safety instructions can cause
severe personal injury and/or property damage.
Caution - Hot Surface
This warning symbol indicates components and surfaces which can
become very hot during spectrometer operation. Do not touch these com-
ponents and surfaces. Risk of skin burn! Be careful when operating near
hot components and/or surfaces.
Caution - Laser Radiation
This warning symbol indicates the existence of laser radiation. Never look
directly into the laser beam or use any kind of optical instruments to do so.
Otherwise permanent eye damage can be the result.

VERTEX 70v User Manual 3
. . . . .
SAFETY
Warning Labels
Besides the dangers described above, there can also be hazardous situations caused
by the sample material. Depending on the type of hazardous substances you work with,
you have to observe specific substance-relevant safety instructions. Put on the corre-
sponding warning label on the appropriate spectrometer position. The label must be
legible and permanently discernible. The following list contains some examples of haz-
ardous substances:
Caution - Frostbite
This warning symbol indicates cryogenic materials (e.g. liquid nitrogen)
required to operate the spectrometer (e.g. cooling detector). Skin contact
with these liquids or cooled components causes severe frostbite. Always
handle the liquids with utmost care. Observe the safety instructions for
handling of cryogenic liquids.
Caution - Harmful Material
This warning symbol indicates the existence of harmful or irritant material
(e.g. the window material BaF2). Observe the safety instructions on the
packaging, and the safety data sheets attached. Non-observance may
cause personal injury.
Caution - Toxic Material
This warning symbol indicates the existence of toxic material (e.g. the win-
dow material KRS-5). Observe the safety instructions on the packaging,
and the safety data sheets attached. Non-observance may cause severe
personal injury or even death.
Caution - Infectious Material
This warning symbol indicates the possible presence of bio-hazardous and
infectious material. When working with this kind of material always, observe
the prevailing laboratory safety regulations and take all necessary precau-
tions and disinfection measures (e.g. wearing protective clothing, masks,
gloves etc.). Failure to do so may cause severe personal injury or even
death. (For information on how to use, dilute and efficiently apply disinfec-
tants, refer to the Laboratory Biosafety Manual: 1993 by WHO - World
Health Organization.)
Caution - Radioactive Material
This warning symbol indicates the possible presence of radioactivity. When
working with radioactive material, always observe the safety regulations
and take all necessary protective measures (e.g. wearing protective cloth-
ing, masks gloves etc.). Failure to do so may cause severe personal injury
or even death.

4 VERTEX 70v User Manual
1SAFETY
Safety Instructions
Waste Disposal
Dispose all waste produced (chemicals, infectious and radioactively contaminated sub-
stances etc.) according to the prevailing laboratory regulations. Detergents and clean-
ing agents must be disposed according to the local waste regulations.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
SAFETY INSTRUCTIONS
The following chapters describe all relevant safety aspects of the spectrometer opera-
tion. Depending on the degree of hazard the safety instructions are classified as fol-
lows:
Danger indicates that death, severe personal injury or substantial property damage
WILL result if proper precautions are not taken.
Warning indicates that death, severe personal injury or substantial property damage
CAN result if proper precautions are not taken.
Caution indicates that minor personal injury or property damage CAN result if proper
precautions are not taken.
Note draws your attention to particularly important information on the product,
e.g. product operation or to a special part of the manual.
The safety instructions Danger, Warning and Caution stand out by the corresponding
warning labels.
Caution - Corrosive Substance
This warning symbol indicates the possible presence of corrosive sub-
stances. When working with corrosive substances, always observe the lab-
oratory safety regulations and take protective measures (e.g. wearing
protective masks and gloves). Failure to do so may cause severe personal
injury or even death.

VERTEX 70v User Manual 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
G
ENERAL
2
VERTEX 70v is an evacuable, fully digital FT-IR spectrometer for demanding R&D
applications. The spectrometer is equipped with a number of features such as AAR
(Automatic Accessory Recognition) ACR (Automatic Component Recognition) and Per-
formanceGuard that facilitate performing spectroscopic measurements and ensure reli-
able measurement results. The function AAR identifies automatically the accessory
installed in the sample compartment, performs several tests and loads automatically the
corresponding experiment file including the pre-defined measurement parameters. The
feature ACR recognizes automatically the currently installed optical components like
source, detector and beamsplitter. These components are electronically coded so that
the spectrometer firmware can recognize them. This information is passed on to the
application software OPUS. The purpose of ACR is to enable the user to select the right
optics parameters in OPUS. In addition, the spectrometer components are monitored
permanently to ensure that they operate within the specification range. This feature is
called Performance Guard. Its purpose is to facilitate fault diagnostics and maintenance.
The data acquisition is based on a free running delta-sigma, dual-channel A/D converter
with 24-bit dynamic range. The A/D converter is integrated into the detector preamplifier
electronics. The DigiTect technology ensures a signal transmission free from interfer-
ences and guarantees the highest signal-to-noise ratio.
VERTEX 70v can be controlled by any data system (PC workstation, notebook etc.) on
which the operating system Microsoft Windows and the spectroscopic software OPUS is
installed. The Ethernet connection provides the possibility to control the spectrometer
also via your intranet or the internet.
The standard spectrometer configuration is designed for data acquisition in the mid IR
region. Optionally, VERTEX 70v can be equipped with additional optical components to
cover the whole spectral range - starting in the far infrared or THz region at 10cm-1 up to
the ultraviolet region at 28,000cm-1. Due to the pre-aligned optical components and the
permanently aligned RockSolid interferometer, the spectral range can be changed eas-
ily. If you work with the advanced spectrometer configuration (i.e. two detector positions
and two source positions are available inside the spectrometer) you can select them
using the software. Removable vacuum-tight covers provide access to the detector and
beamsplitter if you want to exchange these components.
VERTEX 70v has five IR-beam outlet ports (on the right, front and left side) and two IR-
beam inlet ports (on the right and rear side) allowing the connection of a multitude of
optional accessories and/or components like:
• TGA-coupling
• PMA 50 (Polarization Modulation Accessory for VCD and PM-IRRAS)
• HYPERION 1000/2000 IR microscope and HYPERION 3000 imaging
microscope with FPA detector (Focal Plane Array detector system)
• IMAC module (Imaging Accessory with FPA detector)

6 VERTEX 70v User Manual
2GENERAL
• External sample compartment (XSA)
• HTS-XT module (High Throughput Screening Extension)
• Fiber optic coupling module with MIR or NIR fiber probes for solid and
liquid samples
• FT Raman module (e.g. RAM II)
• FIR bolometer
• External, water-cooled sources
There is also the possibility to connect several accessories simultaneously (e.g. a
water-cooled Hg-arc source at the rear side, the RAM II FT-Raman module at the right
side, a fibre optics coupling at the right front side, the HYPERION IR microscope at the
left side and a bolometer detector at the front side).
Diagnostic routines help to maintain optimum instrument status and performance. The
internal validation unit (IUV) is located inside the spectrometer. It contains standards
(test samples) used for the validation and testing of the instrument.
Note: Depending on the spectrometer configuration you have ordered,
your spectrometer may not include all options that are described in
this manual.
The evacuable VERTEX 70v spectrometer allows measurements under vacuum condi-
tions, i.e. unwanted atmospheric interferents (e.g. water vapor or carbon dioxide) are
eliminated nearly completely from the spectrometer interior. Evacuating the spectrome-
ter is more efficient than purging it or using desiccant cartridges. The result of an opti-
mal measurement under vacuum conditions is an IR spectrum in which no H2O or CO2
absorptions mask weak spectral features of the sample.
The spectrometer design enables a separate evacuation of the spectrometer compart-
ments, i.e. either the complete spectrometer interior (sample compartment plus the opti-
cal bench) or only the optical bench can be evacuated. Vacuum shutters (so called
flaps), which can be equipped with optical or IR windows, allow a ventilation of only the
sample compartment in order to preserve the vacuum in the rest of the optics compart-
ment during a sample exchange or an accessory installation. Evacuating and venting
the sample compartment and/or optics are computer-controlled. Moreover, the spec-
trometer is equipped with two pressure sensors providing for the display of the current
pressure inside the spectrometer optics and/or sample compartment.
VERTEX 70v is supplied with an efficient vacuum pump that can evacuate the spec-
trometer optics within a few minutes. The oil-free vacuum pump prevents the spectrom-
eter optics from being contaminated by hydrocarbons.

VERTEX 70v User Manual 7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
I
NSTALLATION
3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GENERAL INFORMATION
Unpacking and initial installation of VERTEX 70v is done by Bruker service engineers.
The operating company has to provide an installation site that meets the site require-
ments described in this chapter. (See also the technical document Installation Require-
ments for VERTEX 70v provided by Bruker Optik GmbH in advance.)
This chapter contains a list of the standard as well as the optional spectrometer compo-
nents and describes the procedures for connecting the spectrometer:
•to the power supply,
•to a PC,
•to the vacuum pump and
•to the purge gas supply line, if necessary.
For detailed information about how to install the computer, refer to the PC manual.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DELIVERY SCOPE
The basic instrument of VERTEX 70v allows upgrading with additional components and/
or accessories. The delivery scope depends on the spectrometer configuration you
have ordered.
Standard Components
The basic instrument includes the following items:
•VERTEX 70v spectrometer (including the user manual)
•Power cord
•PC compatible data system (if desired, the PC can also be provided by the
customer)
•Data cable (Cat5, crossover cable for 10Base-T Ethernet standard)
•Purge gas hose (OD: 6mm, length: approx. 5m)
•Tool kit (slot-head screw driver, cross-head screwdriver and hex keys of several
sizes, sample preparation tools, 3x spare fuses, IR sensor card, metallic cap
shown in fig. 28)
•Software package OPUS/IR (including the OPUS Reference Manual)

8 VERTEX 70v User Manual
3INSTALLATION
Delivery Scope
For installing the vacuum pump, the following items are included:
•Vacuum pump (including the user manual)
•Noise reduction hood
•Vibration absorber
•2x flexible metal hoses
•4x hose clamps
•4x sealing rings
Optional Components
Depending on the ordered spectrometer configuration, the delivery scope can also
include following optional components:
•Optional spectrometer components (e.g. optional detectors) and/or accessories
•Optional OPUS software packages (e.g OPUS/STEP) including the corresponding
manuals
Inspecting the Packaging
After the receipt of the spectrometer, inspect the packaging for damages. If there are
any signs of damage, contact your local shipping representative before opening the
shipping box.
Warning: Do not put a spectrometer into operation that shows signs of
damage. Failure to do so may result in severe personal injuries
and/or property damage.
Transportation
Due to its weight (about 105kg), VERTEX 70v has to be carried by at least four persons
using the supplied transport handles. For transportation purposes, attach these handles
properly to the right and left spectrometer side as shown in figure 1 using 12 screws
(M5 x 16). Tighten the screws using a hex kex (size 4mm). After having transported the
instrument to the desired place, you can remove the transport handles again. Alterna-
tively, you can transport the instrument with a fork lifter.
Warning: Due to the high instrument weight, improper transportation can
lead to personal injuries and/or spectrometer damage.

VERTEX 70v User Manual 9
. . . . .
INSTALLATION
Site Requirements
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SITE REQUIREMENTS
Space Requirements
The spectrometer dimensions are 85cm (w) x 71cm (d) x 32cm (h). (For exact spec-
trometer dimensions refer to appendix D.) At the rear side, the spectrometer requires a
clearance of at least 25cm (10“). The spectrometer should be placed on a stable and
horizontal base. Note that the basic instrument has a weight of about 105kg.
When preparing the installation location for the spectrometer, take into consideration
that the mains power supply connection is easily accessible at any time. The mains
power supply can be interrupted, for example, either by disconnecting the safety plug or
switching off the mains switch on the spectrometer rear side or disconnecting the pri-
mary power receptacle.
Environmental Requirements
To ensure optimum spectrometer performance and long-term reliability the following
environmental conditions are essential:
Temperature Range: 18 - 35 °C (64 - 95 °F)
In case the vacuum pump is operated with installed noise reduction hood ensure the
ambient temperature does not exceed 32°C (90°F).
Humidity (non-condensing): 80% (relative humidity)
Temperature variations can impair the results of long-term measurements. Therefore,
the temperature variations should be less than 1°C per hour and should not exceed 2°C
per day for this type of measurement.
Figure 1: Installing the Transport Handles
Transport
Handles
≤

10 VERTEX 70v User Manual
3INSTALLATION
Site Requirements
Vibration
Ideally, the spectrometer should not be installed near vibration sources (e.g. ventilation
hoods, air conditioners, motors, elevator etc.) or in rooms with intense floor vibration.
Power Supply
The spectrometer power supply unit automatically adapts to the most common power
sources.
Valid voltage range: 100 V AC to 240 V AC
Valid frequency range: 50 to 60 Hz
VERTEX 70v is an instrument of the protection class I.
Caution: To avoid personal injury and spectrometer damage, connect the
spectrometer only to a socket outlet with earthing contact.
To provide for good data quality and a long spectrometer service life, ensure that the fol-
lowing site requirements are met:
•Do not install the spectrometer near sources of potential inductive electrical
interference (e.g. pumps, switching motors, microwave ovens etc.), sources of
high energy pulses, and sources that might cause magnetic or radio frequency
interference.
•Do not place devices such as large electric motors, heaters, welding equipment,
radio transmitting equipment, units emitting pulsed NMRs, or high powered lasers
in close vicinity to the spectrometer. These devices can interfere with the
spectrometer and cause spectrometer malfunction. Ensure that these types of
devices are not connected to the same electrical circuit as the spectrometer.
•If a reliable mains power supply is a problem at your site (caused by brownouts,
power surges, frequent thunderstorms, for example), take precautions to ensure
an uninterruptible power supply.

VERTEX 70v User Manual 11
. . . . .
INSTALLATION
Connecting VERTEX 70v to the Power Supply
CONNECTING VERTEX 70V TO THE POWER
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SUPPLY
Power Cord
Before connecting the power cord, make sure that the spectrometer is switched off, i.e.
the mains switch (B in figure 2) is in the “O” position. Connect the supplied power cord
to the primary power receptacle (C in figure 2) as well as to the mains socket outlet.
The power cord length should not exceed 3m. Depending on the local conditions, the
original power cord may need to be exchanged for a power cord that complies with the
standards of the country in question. The power cord must have approbation of at least
your local authority, UL for US, CSA for Canada or VDE for Europe. The spectrometer
power supply unit automatically adapts to the local voltage and frequency range. (See
section Site Requirements.)
Component
A Ethernet port
BMainsswitch
CPrimary power receptacle (for connecting the
power cord)
Figure 2: Spectrometer Rear Side - Connections for Power Supply and PC
A
B
C

12 VERTEX 70v User Manual
3INSTALLATION
Connecting VERTEX 70v to a PC
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CONNECTING VERTEX 70V TO A PC
Data Cable
The data cable included in the spectrometer delivery scope is a CAT5 crossover cable
(labelled “Cross-over”) with two RJ-45 plugs. This cable is only used for the direct con-
nection of VERTEX 70v to a computer. If you intend to connect the spectrometer to a
network, a different type of cable (i.e. non-crossover, CAT 5 cable for the 10Base-T Eth-
ernet standard) is required. (See appendix E.) The data cable length should not exceed
100m (without repeater).
Connect one end of the data cable to the Ethernet port (ETH) (A figure 2) and the other
end of the data cable to the RJ-45 socket of the computer network interface card. (For
detailed information refer to the computer manual.)
After having set up the data cable connection, turn on the spectrometer using the mains
switch. After a few seconds, the spectrometer beeps once and starts a self test. After
the initialization has been completed successfully, the “STATUS” LED (figure 10) turns
from red to green. Now switch on the computer and the monitor. (For information on
how to install the computer and how to set up signal and power cable connections for
the computer, monitor etc. refer to the computer manual.)
Computer Setup
VERTEX 70v and the delivered PC are already configured for the stand-alone opera-
tion. The spectrometer IP address is factory-set to 10.10.0.1. In case you have not pur-
chased the computer together with the VERTEX 70v spectrometer, you have to assign
an appropriate IP address to the computer to which you want to connect the spectrome-
ter. For detailed information about how to assign an IP-address to the computer refer to
appendix E.

VERTEX 70v User Manual 13
. . . . .
INSTALLATION
Connecting VERTEX 70v to the Vacuum Pump
CONNECTING VERTEX 70V TO THE VACUUM
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
PUMP
The attachment flange (NW25 flange) for connecting the vacuum pump is at the spec-
trometer rear side. Figure 3 shows the valve block with removed cover.
The vent openings are covered by a plug made from sintered-powder metal which is air-
permeable (i.e. the spectrometer can be vented with the plugs installed on the vent
opening). (See figure 8.) The plug functions like a filter preventing particles from enter-
ing the spectrometer together with the influent air.
Note: For detailed information about the vacuum pump refer to the user
manual provided by the vacuum pump manufacturer.
Attachment Flange for the Vacuum Pump
Opening for venting
the optical bench
(Note: When purging the
spectrometer this port is
used as purge gas inlet for
the optical bench.
Valve for evacuating
the optical bench
Valve for evacuating
the sample compart-
ment
Figure 3: Valve Block (Spectrometer rear Side)
Opening for venting
the sample compart-
ment
(Note: When purging the
spectrometer this port is
used as purge gas inlet for
the sample compartment.
Valve for venting the
sample compartment
Valveforventingthe
optical bench

14 VERTEX 70v User Manual
3INSTALLATION
Connecting VERTEX 70v to the Vacuum Pump
Installation Procedure
•Remove the valve block cover shown in figure 4 by loosening the two Allen screws
using a hex key (size 3mm) and pulling off the cover.
•Install the supplied sealing ring at the attachment flange. See figure 5.
•Press the supplied flexible metal hose against the attachment flange (figure 6a)
and attach the hose to the flange using the supplied hose clamp (figure 6b).
Secure the hose clamp by fastening the wing screw.
Allen Screws Valve Block Cover
Figure 4: Removing the Valve Block Cover
Figure 5: Connecting VERTEX 70v to Vacuum Pump - Step 1
Attachment Flange Sealing Ring
Flexible
Metal Hose

VERTEX 70v User Manual 15
. . . . .
INSTALLATION
Connecting VERTEX 70v to the Purge Gas Line
During operation, the vacuum pump generates vibrations. In order to prevent these
vibrations from being transferred to the spectrometer via the flexible metal hose, the
supplied vibration absorber has to be installed between the vacuum pump and the
spectrometer. The procedure for connecting the flexible metal hose to the vacuum
pump and to the vibration absorber is identical to the procedure described above.
Note: Make sure that the vibrating metal hoses do not come into contact
with the table on which the spectrometer is placed.
During the operation, the vacuum pump produces an increased noise level. In order to
reduce the noise level install the supplied noise reduction hood over the vacuum pump.
For information about the noise reduction hood installation refer to instructions provided
by the pump manufacturer.
CONNECTING VERTEX 70V TO THE PURGE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GAS LINE
As an alternative to the vacuum operation, VERTEX 70v can be purged with either dry
air or dry nitrogen gas. The spectrometer has two purge gas inlets; one for purging the
sample compartment and the other for purging the optical bench. The purge gas inlets
are at the spectrometer rear side. See figure 7.
Figure 6: Connecting VERTEX 70v to Vacuum Pump -Step 2
Hose Clamp
Wing Screw

16 VERTEX 70v User Manual
3INSTALLATION
Connecting VERTEX 70v to the Purge Gas Line
For detailed information about the required purge gas supply conditions refer to chapter
Operation, section Purging the Spectrometer.
Installation Procedure
Note: In case the spectrometer is evacuated, first vent it before starting
the installation procedure. Otherwise, a warning message regarding
unstable pressure conditions inside the spectrometer will appear.
•To connect the spectrometer to the purge gas supply you need a stiff hose with an
outer diameter of 6mm. Remove the plug (made from sintered-powder metal) from
the purge gas inlet by pressing the lock ring inwards (figure 8) and pulling out the
plug. Connect one end of the hose to your supply line for dry air or dry nitrogen
gas and insert the other end of the hose into the purge gas inlet for either the
sample compartment or optical bench.
•If you want to purge both the sample compartment and the optical bench, you
need a T-shape connecting hose with two hose ends leading to the spectrometer.
After having connected the main end of the hose to the supply line, insert one of
the other two hose ends into the purge gas inlet for the sample compartment and
the other hose end into the purge gas inlet for the optical bench.
Figure 7: Purge Gas Inlets
Purge gas inlet for
optical bench.
(Note: In case of vacuum
operation-ventopeningfor
venting the optical bench.)
Purge gas inlet for
sample compartment.
(Note: In case of vacuum
operation - vent opening for
venting the sample com-
partment.)
Table of contents
Other Bruker Measuring Instrument manuals

Bruker
Bruker ultraflex Series User manual

Bruker
Bruker CTX User manual

Bruker
Bruker BMS User manual

Bruker
Bruker D8 Discover Manual

Bruker
Bruker S1 TRACER User manual

Bruker
Bruker DektakXT Manual

Bruker
Bruker DektakXT User manual

Bruker
Bruker ultraflex III User manual

Bruker
Bruker CTX User manual

Bruker
Bruker S1 TITAN User manual