
Application hints
Instruction Manual, Nov-18 14
•A reduction of AMD running times by up to 50 % can be reached by using "extra
thin" (100 μm) layers which require only 2 mm migration increments and shorter
drying times. Use of Licrosphere plates reduces developing times as well.
•If a polarity change of the gradient does not result in the desired resolution after 2-
3 attempts: change selectivity, i.e. the „base solvent“. If there is still poor resolution
the stationary phase must be changed, e.g. to diol, amino, cyano and RP18 W
HPTLC plates.
•Gradients with less change in volume % than shown in the table below are
comparable with multiple isocratic developments over the same distance. Such
gradients are not recommended.
Polarity change Min. change (% per step)
methanol to dichloromethane
acetonitrile to dichloromethane
t-butyl methyl ether to n-heptane
dichloromethane to n-heptane
•To avoid increasing diffusion of peaks, 5-10 steps are sufficient for isocratic
development
•Of course all solvents, glassware, HPTLC plates etc. used should be as pure as
required by the analytical task. If necessary pre-washing of HPTLC plates can be
done by immersion in very pure methanol or isopropanol for about 2 hours or by
pre-developing with the above mentioned solvents to a higher running distance
than the following gradient. Then the plates have to be dried again, e.g. in a very
pure drying cabinet at 120 °C for 20 min. It is recommended to keep prewashed
plates protected against laboratory atmosphere, e.g. in an (empty!) desiccator or
wrapped in aluminium foil
•If basic or acidic components are present or expected in a sample, it is
recommended to adjust the pH. This can be done by adding small amounts (0.01-2
%) of NH3, HCOOH, CH3COOH etc. to the polar solvent. Adding bases or acids to
dichloromethane or less polar solvents has no effect. In this case the layer has to be
conditioned by the gas phase by filling the conditioning bottle with about 200 mL
of a 0.1-4.0 N solution of the acids or bases. The rule that acids are to be
chromatographed in an acid environment and bases in an alkaline one need not be
followed with AMD chromatography and a universal gradient. Since a universal
gradient starts with a high polarity, even salts will initially migrate, i.e. salts of acids
in an alkaline medium or salts of bases in an acidified medium. Often sharper
separations are obtained in such systems. The less favourable behaviour of acids in
an acidified medium etc. may be explained as follows: in an acid medium the free,
dissociated acid migrates to the position corresponding to its polarity. Of course
this is a different position than the one assumed by the salt of this acid when