manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. CURIOSIS
  6. •
  7. Laboratory Equipment
  8. •
  9. CURIOSIS Cellpuri Installation guide

CURIOSIS Cellpuri Installation guide

This manual suits for next models

1

Other CURIOSIS Laboratory Equipment manuals

CURIOSIS Celloger Nano User manual

CURIOSIS

CURIOSIS Celloger Nano User manual

CURIOSIS FACSCOPE B Instruction manual

CURIOSIS

CURIOSIS FACSCOPE B Instruction manual

CURIOSIS Celloger Mini Plus User manual

CURIOSIS

CURIOSIS Celloger Mini Plus User manual

CURIOSIS Celloger Mini Plus Installation guide

CURIOSIS

CURIOSIS Celloger Mini Plus Installation guide

CURIOSIS FACSCOPE B User manual

CURIOSIS

CURIOSIS FACSCOPE B User manual

CURIOSIS Celloger Mini Installation guide

CURIOSIS

CURIOSIS Celloger Mini Installation guide

CURIOSIS Celloger Mini User manual

CURIOSIS

CURIOSIS Celloger Mini User manual

CURIOSIS Celloger Mini Installation guide

CURIOSIS

CURIOSIS Celloger Mini Installation guide

CURIOSIS FACSCOPE B User manual

CURIOSIS

CURIOSIS FACSCOPE B User manual

Popular Laboratory Equipment manuals by other brands

Radleys Carousel 6 Plus instructions

Radleys

Radleys Carousel 6 Plus instructions

Hemotherm 400MR Operation & technical manual

Hemotherm

Hemotherm 400MR Operation & technical manual

mrclab DVO Series Operation manual

mrclab

mrclab DVO Series Operation manual

CoolLED pE-800 quick start guide

CoolLED

CoolLED pE-800 quick start guide

Luminex Guava easyCyte Maintenance Guide

Luminex

Luminex Guava easyCyte Maintenance Guide

IKA A10 basic operating instructions

IKA

IKA A10 basic operating instructions

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Belden

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler K1223 Series Operation and instruction manual

Koehler

Koehler K1223 Series Operation and instruction manual

Globe Scientific GCM-12 quick start guide

Globe Scientific

Globe Scientific GCM-12 quick start guide

Getinge 86 SERIES Technical manual

Getinge

Getinge 86 SERIES Technical manual

CORNING Everon 6000 user manual

CORNING

CORNING Everon 6000 user manual

Biocomp GRADIENT MASTER 108 operating manual

Biocomp

Biocomp GRADIENT MASTER 108 operating manual

Cole Parmer Stuart SB3/1 instruction manual

Cole Parmer

Cole Parmer Stuart SB3/1 instruction manual

BEVS 2501/2A user manual

BEVS

BEVS 2501/2A user manual

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Quick Manual
www.curiosis.com
2
Ⅰ. Product layout
Ⅱ. Preparation
•Cellpuri
•Syringe pump
Ⅲ. Operation
Ⅳ. Warnings & Precautions
V. Specification
Table of contents
www.curiosis.com
3
I. Product layout
CellpuriParts Syringepump
Pusher
Operation button Power switch
Power cable
USB CDC (Engineer mode)
Chip holder
Syringe
Inlet
Waste
Outlet
Chip type selection
(Engineer mode)
www.curiosis.com
4
Materials
Pipette
Syringe
Cell sample
Cellpuri
Syringe pump
Power cable
Syringe holder
(Optional)
II. Preparation
Ⅰ. Cellpuri
Preparation
Take Cellpuri out of the box.
Tear the edge of individual package and remove Cellpuri from the cover.
* Ensure that Cellpuri is facing upward when placed on the table.
Load the luer-lock syringe with desired volume (2~10 ml) of cell suspension.
* Cell concentration less than 5×105cells/ml is recommended.
Insert the syringe tip into the inlet of Cellpuri and turn clockwise to connect.
Place pre-loaded Cellpuri on the chip holder.
Step 1.
Step 3.
Step 4.
Step 2.
Step 5.
www.curiosis.com
5
II. Preparation
ⅠI.Syringepump
STEP1. Put the syringe pump in the clean bench.
3
STEP 1 STEP 4STEP 2 / STEP 3
2
STEP2. Connect the power cable to the pump.
STEP3. Turn on the power switch.
STEP4. Check if the operation button lights up in blue.
www.curiosis.com
6
III. Operation
Place the pre-loaded Cellpuri on the chip holder.
Press the operation button once.
* Emergency Stop
Press the operation button once to stop (during operation)
* Return to start
Pressing the operation button in the Stop state returns to the initial
position
* Error
When an error occurs during operation, the pump will stop automatically
Eg: When loading a high concentration sample
Wait for the pusher to fully go down and return to initial position.
Gently mix and take the enriched cells from the outlet using a pipette.
Dispose the chip and syringe after use.
Step 1.
Step 3.
Step 2.
Step 4.
Step 5.
www.curiosis.com
7
IV.Warnings & precautions
- Do not store at extreme temperature and humidity.
- Keep the product at a room temperature between 1 ~ 30 °C and
relative humidity.
- Keep the product in the original box to protect from light
and physical damage.
- Cellpuri can be easily damaged by external shock.
Be careful not to drop the product.
Storing Cellpuri
Warnings & Precautions
- Cellpuri is a single-use, sterilized product. Do not reuse.
- Do not use if the package is open or damaged.
- Read the instructions carefully before use.
- Use immediately after opening the package.
- Make sure that the product is facing upward and placed
on a flat rigid surface.
- Do not add any solution other than cell suspension.
-Be careful of product’s sharp edges.
Using Cellpuri
- Operate the device following the operating condition.
-Ensure the voltage matches with the device’s power supply voltage.
- Check if the power cable is properly grounded to avoid
potential electric shock.
- Use only the components provided and authorized by Curiosis Inc.
- Do not disassemble the device and contact an authorized service
center if required.
Using Syringe pump
www.curiosis.com
8
V. Specification
Specifications
Cellpuri
Dimension 76 x 25 x 23 mm
Weight 15.3 g
Sample size 7~15 μm size cells
Sample concentration ~ 2×106cells/ml
Enrichment 20-fold ~
Yield 90% ~
Loading volume 2~20 ml
Orderinginformation
Catalog no. Product Name Description
CRCPR-CE10 Cellpuri Cell Enrichment chip (10 pc/box)
CRCPR-SP01 Syringe pump Motorized vertical syringe pump
Syringe pump
Dimension 17 x 26 x 28 cm
Weight 4 kg
Volume range 2~10 ml
Flow rate 1 ml/min
Power DC 24V, 2A
Operating condition Temperature 20~35 ℃
Relative humidity < 95 %