DEGA NSx-yL III LCD User manual

INSTRUCTION
MANUAL
Gas Detector
DEGA NSx-yL III LCD
Reproduction of this manual, or any part thereof, in any form, without the prior permission of DEGA.CZ s.r.o. is
prohibited
DEGA CZ s.r.o. reserves the right to alter the specifications of the hardware and software described in this
manual at any time and without prior notice
DEGA CZ s.r.o. bears no liable for any damage resulting from use of this device

Content
For your safety .................................................................................................................................................................................. 3
Technical data and information ........................................................................................................................................................ 3
Operational conditions ..................................................................................................................................................................... 4
Terminology ...................................................................................................................................................................................... 4
Product description ........................................................................................................................................................................... 5
Instalation, assembly and disassembly of the detector ................................................................................................................... 6
1. Assembly of the detector ...................................................................................................................................................... 6
. Replacement of the sensor module ...................................................................................................................................... 6
3. Replacement of the battery .................................................................................................................................................. 6
4. Connecting the detector via current loop to the controler DEGA UPA II/DEGA UPA III ....................................................... 7
5. Connecting the sensor via RS485 to the controller DEGA UKAIII/DEGA UPA III/DEGA UDA III ............................................ 7
6. Installation of wiring for RS485 ............................................................................................................................................. 7
7. Setting the RS485 adress of the detector ............................................................................................................................. 7
8. Terminating resistor .............................................................................................................................................................. 7
9. Communication protocol switch DEGA/MODBUS ................................................................................................................ 8
Detector functions ............................................................................................................................................................................ 8
1. Turning on the detector ........................................................................................................................................................ 8
. Gas detection ........................................................................................................................................................................ 8
3. Malfunction ........................................................................................................................................................................... 8
4. Monitoring the calibration periods ....................................................................................................................................... 8
5. Reading the record of measured concentrations and alarms .............................................................................................. 8
Detector controls .............................................................................................................................................................................. 9
1. History menu „HIST“ ........................................................................................................................................................... 10
. Information menu „INF“ ..................................................................................................................................................... 10
3. Settings menu „SET“ ........................................................................................................................................................... 11
4. Test menu „TEST“ ............................................................................................................................................................... 13
Operation, maintenance, inspection and service of the detector .................................................................................................. 14
1. Usage limits ......................................................................................................................................................................... 14
. Operation ............................................................................................................................................................................ 14
3. Operation/Maintenance ..................................................................................................................................................... 14
Accessories and basic types of detectors ....................................................................................................................................... 15
1. Calibration adapter/connection to the gas pump DEGA GAS INLET .................................................................................. 15
. Cover against splashing water DEGA WATER CAP .............................................................................................................. 15
3. Funnel for gas collection DEGA COLLECT CAP .................................................................................................................... 15
4. Additional Ex „d“ cable gland .............................................................................................................................................. 15
Basic type of detectors ................................................................................................................................................................... 15
1. Detectors with a catalytic sensor NSx-CL III LCD ................................................................................................................. 15
. Detectors with an electrochemical sensor NSx-EL III LCD .................................................................................................. 15
3. Detectors with an infrared sensor NSx-IL III LCD ................................................................................................................ 17
4. Detectors with a semiconductor sensor NSx-SL III LCD ...................................................................................................... 18
5. Detectors with a PID sensor NSx-PL III LCDPID ................................................................................................................... 18
Add-on modules .............................................................................................................................................................................. 19
Attachments .................................................................................................................................................................................... 19
1. Chart for setting the detector address ............................................................................................................................... 19
. Conversion between volumetric concentration and lower explosion limit of methane .................................................... 19
3. Table of error codes ............................................................................................................................................................ 19
4. Signalization transmitted by the current loop 4- 0 mA ..................................................................................................... 0
5. Package content .................................................................................................................................................................. 0
General warranty terms and conditions ......................................................................................................................................... 1

3
For your safety
Beware of stati ele tri ity
Electronic components are sensitive to static electricity. Do not touch them directly - they may get damanged.
The devi e is intended to be installed by a trained person
The product is designed for installation only by a certified technician. The manufacturer is not liable for damages resulting
from incorrect or improper handling.
In ase of malfun tion, immediately unplug from the power supply
If you notice an unusual smell or smoke emitting from the product, unplug it from the power supply, battery backup and all
other attachments. Continued operation could result in injury or property damage. After disconnecting, have the device
inspected at an authorized dealer or manufacturer.
Do not open the dete tor and do not repla e the sensors in the dete tor in a potentially explosive atmosphere
Opening the cover and replacing the sensor in the detector in a potentially explosive atmoshpere can cause an explosion. If
service is necessary, first unplug the device from the power supply. After that the device may be disassembled and the
sensors may be replaced (certified partner or manufacturer).
Do not disassemble the produ t and ensure against it's onta t with water
Contact with internal components of the product may cause an electric shock. In case of any malfuction entrust the servicing
of the product exclusively to a certified service centre. Contact with water can create a short circuit in the product and
consequent damange to property or personal injury.
Use appropriate able types
To ensure compliance with the parameters of the product, only use cable types recommended in this guide.
Dispose of used produ ts and trasmitter sensors with respe t to the environment
Detector sensors contain hazardous substances. Dispose of them in accordance with the current legislation on environmental
protection.
Use the dete tor only with the appropriate ertified DEGA produ ts
The device is certified as functionally and technically qualified only with original "DEGA" accessories. In case of using the
device with any other products the manufacturer is not liable for any damages that may occur.
Undertake regular fun tional he ks and alibrations of the dete tor
Perform regular "CALIBRATION" (setting the detection limits, checking the responsivness of the sensor, checking the
functionality of the detector) and "OPERATIONAL AND FUNCTIONAL CHECKS" of the entire detenction system (sensor
excitation with subsequent control of optical and audible alarms, triggering fans, shutdown technology, etc.). Perform
calibration and operational and functional checks only at certified service centers with a valid certificate of competence or the
manufacturer.
Warning: The detector automatically checks it's calibration period - the period of validity of it's calibration. After 1 months since the last
calibration (Max. calibration period) the detector will transmit this fact to the host system. The detector must be calibrated immediately at
a certified service center with a valid certificate of competance or the manufacturer. See section "Monitoring the calibration peridos".
Te hni al data and information
Supply voltage: 4 V nominal, operational range 8-30 V
Cable connections via 4- 0 mA: shielded cable 3 x 1 mm (max. 1 00 m)
shielded cable 3 x 1,5 mm (max. 400 m)
Cable connections via RS485: shielded cable 4 x 0,8 mm (max. 400 m) - see section “ instalation of cabeling
for RS485“
Output: 4 - 0 mA
RS485 - protocol DEGA, or MODBUS
Piezo buzzer
Standarts: EN60079-0: 01 , IEC60079-0: 011
EN60079-1: 014, IEC60079-1: 014
Marking according to ATEX: II G Ex db IIB+H T6 Gb Tamb:-40°C-+60°C
Marking according to IECEx: Ex db IIB+H T6 Gb Tamb:-40°C-+60°C
II – eqiupment group - non-mining environment 2 G - eqiupment category - explosive atmospheres - zone 1
Ex d - type of protection - flameproof enclosure „d“ IIB+H2 - gas group
T6 – temperature class Gb - eqiupment protection level EPL Tamb – ambient temperature

4
Dimensions without cable glands: 150 x 170 x 65 mm (WxHxD)
Weight: ,1 kg
Capacity of the internal memory of history: 44 days at 60s recording interval
Interval record storage memory: 60 s (adjustable range 1- 55 s)
Dead band: max 5 % of range
Consumption/input at 24V (output RS485)
Warm
-
Up time
DEGA NSx
-
EL III LCD
90 mA/ , W
DEGA NSx
-
EL III LCD
max. 180 s
DEGA NSx
-
CL III LCD
135 mA/3,3 W
DEGA NSx
-
CL III LCD
max. 30 s
DEGA NSx
-
IL III LCD
11
5
mA/ ,8 W
DEGA NSx
-
IL III LC
D
max. 15 s
DEGA NSx
-
SL III LCD
135 mA/3,3 W
DEGA NSx
-
SL III LCD
max. 180 s
DEGA NSx
-
PL III PID LCD
154 mA/3,4 W
DEGA NSx
-
PL III PID LCD
max. 30 s
Consumption/input at 24V (output 4
-
20mA)
Time to stabilize (>5day without pow
e
r)
DEGA
NSx
-
EL III LCD
1
10 mA/ ,7 W
DEGA NSx
-
EL III LCD
1
-
7 h
DEGA NSx
-
CL III LCD
155 mA/3,7 W
DEGA NSx
-
CL III LC
max. 1 h
DEGA NSx
-
IL III LCD
135 mA/3,3 W
DEGA NSx
-
IL III LCD
max. 30 min
DEGA NSx
-
SL III LCD
155 mA/3,7 W
DEGA NSx
-
SL III
LCD
max
.
5 h
DEGA NSx
-
PL III LCD
15
4 mA/3,4 W
DEGA NSx
-
PL III LCD
max. 30 min
Response time (T90)
Sensor lifetime in a lean environment
DEGA NSx
-
EL III LCD
max. 180 s
-
based on sensor type
DEGA NSx
-
EL III LCD
years
DEGA NSx
-
CL III LC
D
max. 15 s
DEGA NS
x
-
CL III LCD
years
DEGA NS
x
-
IL III LCD
max. 15 s
DEGA NSx
-
IL III LC
D
5 years
DEGA NSx
-
SL III LCD
max. 30 s
DEGA NSx
-
SL III LCD
years
DEGA NSx
-
PL III LCD
max. 15 s
DEGA NSx
-
PL III LCD
5 000 h
Operational onditions
Ambient temperature: -40 °C to +60 °C
Relative humidity: 0-95 % RH
Air pressure: 80-1 0 kPa
Flow of ambient air: max. m/s - flow directly to the sensor in not allowed
Protection level with a cover: IP 54, with a DEGA WATER CAP IP 66 cover
Location: BE3N - explosive atmoshpheres - zone 1 ( G)
Terminology
The marking system for dete tor DEGA NSx-yL III LCD:
Gas detector DEGA NSx-yL III LCD
Series of detectors in potentially explosive atmospheres
Type of detected gas
Sensor type with linear output
Third generation product with new DEGA design
Versions with an LCD display
DEGA NSx-EL III LCD with an ele tro hemi al sensor
They operate on the principle of change of electrical parameters on the electrodes stored in electrolyte, due to
oxidation/reduction reactions of the detected gas on it's surface. These sensors have good selectivity and the ability to detect
very low concentrations of toxic gases.
DEGA NSx-CL III LCD with a atalyti sensor (Pelistor)
They operate on the principle of catalytic combustion - gas concentration is measured based on the amount of heat released
in a controlled combustion reaction. The reaction is supported by a suitable temperature and the pressence of a catalyst.
These sensors can be used to detect a broad range of flammable gases. The sensors are characterized by fast response,a long
lifetime and high stability. A minimum of 10 % of Oxygen in the air is required for it's proper function.
DEGA NSx-IL III LCD with an infrared sensor (NDIR)
Top quality scanning method. They operate on the principle of infrared spectroscopy. The sensors have excellent selecivity in
organic matter, do not require any oxygen in the atmoshpere and are resistant to catalyst poisons (sulfur and silicon

5
compounds) which cause a change of sensitivity of catalytic sensors. The sensors are characterized by high stability and a long
lifetime.
DEGA NSx-SL III LCD with a semi ondu tor sensor
They operate on the principle of changes in electrical conductivity of semiconductors by changing the concentration of the
detected gas. Their advatage is a long lifetime in a clean environment and a wide range of different types of gases and vapors.
Their disadvatage is their low selectivity - the sensor largly responds to other gases for which it is not calibrated.
DEGA NSx-PL III LCD PID with a PID – photoionization sensor
Typical photoionization detectors measure volatile organic compounds and other gases in concentrations from sub parts per
billion to 10 000 parts per million (ppm). The photoionization detector is an efficient and inexpensive detector for many gas
and vapor analytes. PIDs produce instantaneous readings, operate continuously, and are commonly used as detectors for gas
chromatography or as hand-held portable instruments. Hand-held, battery-operated versions are widely used in military,
industrial, and confined working facilities for health and safety. Their primary use is for monitoring possible worker exposure
to volatile organic compounds (VOCs) such as solvents, fuels, degreasers, plastics & their precursors, heat transfer fluids,
lubricants, etc. during manufacturing processes and waste handling.
Produ t des ription
Magnetic control
„BACK“
Magnetic control
„FRONT“
Ex „d“ cable gland
Magnetic control
„ENTER/EXIT“
IrDa interface
OptionalEx „d“ cable
gland
Body of the
removable sensor
Cover against splashing
water (optional accessories)
Body of the removable
sensor
Ex „d“ cable gland Cover against
splashing water
(optional
accessories)
Mounting holes
PCB electronics Ex „d“ cable gland Optional Ex „d“ cable gland
Body of the
removable sensor
Power supply terminal
block+ 4V DC
Power supply
terminal block GND Signal terminal block 4- 0 mA
6
5
8
7
4
3
2
1
11
0
10
0
9
6
5
8
7
4
3
2
1

6
Terminal resstor
RS485 A
Terminal resistor
RS485 B
Expanding contact
relay ERROR Expanding contact relay 1
Expanding contact
relay Expanding contact relay
3
Jumper connector of
the terminal resistor
RS485
Jumper connector of the
communication protocol
(DEGA/MODBUS)
Battery CR 03
DIP switch for the RS485
adress
Programming
connector LCD display connector
Status LED Calibration jumpers
Instalation, assembly and disassembly of the dete tor
Before assembling, read the valid instalation standarts EN 60079- 9- (Selection, instalation, use and maintenance of
detectors for combustible gases and oxygen) and EN 45544-4 (Guildelines for the selection, installation, use and maintenance
of detectors of toxic substances).
In explosive environments the electrical installation must be performed according to DIN EN 60079.14 (Electrical instalation in
hazardours areas)
Secure that the sensor is reachable by air. The detector must be in a free area with no obstacles in its way (furniture etc.)
Ensure that the input of the sensor cannot be polluted by layers of dust or other contamination.
1. Assembly of the dete tor
The detector consists of four parts - the body of the detector, the removable sensor and Ex „d“ cable glands.
Detector assembly procedure is as follows:
a) Disassemble the detector using four screws with an internal hexagon of 5 mm.
b) Disconnect the LCD from the motherboard.
c) Mount the detector on a flat surface using four 6 mm dowels to a height above the floor according to the
specification of the detected substance with the gas inlet facing downwards.
d) Pass the cable through the gland.
e) Connect the cabling to the detector terminal board according to the chapters "connection of the detector to
control panels". It is only permitted to use an Ex "d" gland with an M 0x1.5 thread.
f) Interconnect using M4 grounding screws. A CUPAL washer must be used.
g) Connect the LCD.
h) Assemble the detector using four screws with an internal hexagon of 5 mm.
2. Replacement of the sensor module
This activity must not be performed when the detector is energized.
In case of need to replace the sensor module with a new piece, unscrew and remove the
cover. Disconnect the sensor module cable from the PCB (1), loosen the sink screw ( ) and
unscrew the steel cylinder with the sensor (3). Screw the new sensor module, retain it
with the sink screw and connect the cable to the PCB. Finally, mount the cover and
connect to the power supply.
3. Replacement of the battery
The battery lifetime in the
sensor is approximately 5 years. After
this time some functions of the
detector may not work properly. Remove the battery from the holder and replace it with one of the
recommended types. Replacing the battery in the transmitter, which is not connected to the power,
will erase the internal clock.
Panasoni BR2032
14
13
16
15
12
11
10
9
22
21
20
19
18
17

7
4. Connecting the detector via current loop to the controler DEG UP II/DEG UP III
Connect one detector to each channel of the controler as shown in the picture below
5. Connecting the sensor via RS485 to the controller DEG UK III/DEG UP III/DEG UD III
6. Installation of wiring for RS485
Wiring must be installed using bus topology and in compliance with the RS485 policy.
The maximum number of connected transmitters per controller channel is 16 (may be less depending on the configuration of
the controller), while the total length of the controller (electrical distance between the controller and the last transmitter)
should not exceed 400 meters.
Sele ting the appropriate type of able depends on the fire report and the proto ol for determining external influen es.
7. Setting the RS485 adress of the detector
Each detector must have a unique adr
ess within the entire bus, otherwise there will be communication
collisions and malfunctions.
The detector adress can either be set internally using the DegaConfig program or using the DIP switch
on the PCB.
If the position 6 (labeled ADD) is in the ON position, then the adress according to the setting of pins 1-5
is considered. Otherwise the adress set in DegaConfig is considered.
The adress can be set from range 1-31 using binary values. A table with DIP settings for individual
adresses is listed in the attachment "Chart for setting the detector adress"
8. Terminating resistor
According to the RS48
5 specifications, the last device on the bus must be ending with a terminating
resistor 1 0R. Plug a jumper on the JP connector of the last device on the bus to include the 1 0R
terminating resistor. In the default configuaration, the jumper connector is not plugged.

8
9. Communication protocol switch DEG /MODBUS
Plugging a jumper in the JP1 connector will swtich from the DEGA communication protocol to the
MODBUS communication protocol.
Dete tor fun tions
The detector's motherboard is equipped by status LEDs, which help in detecting problems during the installation.
LED „POW“ shines at correct power
LED „TxD“ flashes when transmitting a packet via RS485
LED „RxD“ flashes when a packet is correctly recieved via RS485
LED „ERR“ shines/flashes in case of malfunction or substandart situations
1. Turning on the detector
After turning on the power the LED "POW" starts shining and the LED "ERR" starts flashing, indicating a forming sequence of
the sensor and automatic testing procedures, which can take up to 180 s depending on the sensor used. The output of the
current loop is 1 mA. During this sequence, testing of internal electronics and stabilization of the sensor in order to eliminate
false alarms after turning on, is taking place. At this time, the display only shows the time to the beginning of the
measurement loop.
After completion of the formation, a 4 mA current begins to flow on the output of the current loop and the detector starts
working according to it's settings.
2. Gas detection
The detector continuously measures the detected gas concentration in the atmoshpere and converts it's current value into a
4- 0 mA signal or transmits it's value to the evaluation unit via DEGA/MOBUS protocol.
3. Malfunction
In case of malfunction the LCD displays the measured concentration, an error code and a key symbol . On the PCB this
condition is indicated by the yellow „ERR“ LED. The meaning of each error code is included in the attachment "Table of error
codes"
4. Monitoring the calibration periods
The detector continuously checks the calibration validity of the connected sensor.
After 1 months since the last calibration (Max. calibration interval) the LED "ERR" starts flashing and an inscription
starts flashing on the LCD display. The connected sensor must be calibrated immediately. The detector will
transmit the information about the ending calibration via current loop. The transmission will be the following:
10 s transmitting a 4- 0 mA signal informing about the actual gas concentration following a 1 second interval of mA current.
5. Reading the record of measured concentrations and alarms
The detector periodically after 60 s stores the current detected concentration into it's internal memory. The internal memory
retains data from the last 64 60 measurements (cca 44 days). In order to read this information, the program DegaConfig is
required. Refer to the DegaConfig program instruction.

9
Dete tor ontrols
Detectors equipped with an LCD display can be controlled by the magnetic contact located on the sides of the display
Warning: calibration
required
Warning: Service is
required
Data display
Time display
Channel number
(adress)
Activated magnetic
control
Calibration icon
Communication with the
control panel in
progress
Measured value Bar graph of the
measured value
Measured value units
Alarm PEL (exceeding
long-term exposure
limit)
Alarm STEL (exceeding
short-term exposure
limit)
Reached alarm level,
alarm levels
Alarm indication
-
„BACK“ move backwards in the menu, reduci
ng the value
-
„FRONT“move forwards in the menu, incr
e
a
sing the value
-
„ENTER“
confirm/enter the menu
Placing a magnet into the indent „ “ for 5 s, activates the magnetic control, which is displayed by the icon „ “ on the
LCD.
Rea hing level 4 alarm with atalyti and semi ondu tor sensors auses a rat het phenomenon, where even after re overy
from gas leak, the sensor is still in level 4 alarm and needs to be manually reset by pressing „ “.
Main menu stru ture:
Enter the menu by holding the magnet for 5 s
|
History->-Information->-Settings->- Test->- Exit
|-----------------------<------------------------------|
Menu
Display
Description
History
Browsing the alarm history
Information
Information about detector settings
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1

10
Se
ttings
Detector configuration
Test
Testing the detec
t
o
r funcions
Exit
Return to
normal operation
1. History menu „HIST“
Browsing the history
Menu
Display
Des ription
History of all
alarms
Use „ “ to go trough individual alarms. Displays the
date and time of the alarm. Return to the history menu
„ “
H
i
ghest
on entration in
the past hour
Displays the highest measured concentration and alarm
for the past hour. Return to History menu „ “
Highest
on entration in
the past 8 hours
Displays the
highest measured concentration and alarm
for the past 8 hours. Return to History menu „ “
Highest
on entration in
the past 12 hours
Displays the highest measured concentration and alarm
for the past 1 hours. Return to History menu „ “
Highest
on entration in
the past 24 hours
Displays the highest measur
e
d
concentration and alarm
for the past 4 hours. Return to History menu „ “
Highest
on entration
stored in the
memory
Displays the highest measured concentration and alarm,
that is stored in the memory. Return to History menu
„ “
Exit
Return to the
m
ain menu
2. Information menu „INF“
Information about settings. Sequentially displays the following information. Access previous information by pressing „ “
Menu
Display
Description
Date of next
alibration
Displays the date of the next calibration

11
Dat
e
of last
alibration
Displa
ys the date of the last calibration
Power voltage
Displays the power voltage
Temperature
Displays the chip temperature (about 15°C higher than
the ambient temperature)
RS485 devi e
adress
Displays the device adres
s
Range up to
Measured conc
entration value corresponding to 0
mA
current
Alarm level 1
Displays alarm level 1
Alarm level 2
Displays alarm level
Alarm level 3
Displays alarm level 3
A
larm level 4
Displays alarm level 4
3. Settings menu „SET“
Protected by password 0004. By enetering a wrong password, the detector returns to measurement mode.
The password can be changed in the menu SET -> PSWD.
Menu
Display
Description
Calibration
Sens
or calibration
Setting the alarms
Setting the alar
m
s
Setting the range of the 20
mA
urrent loop
Setting the range of the 0
mA
current loop

1
Changing the password
Changing the password
Setting the year
Setting the year
Setting the date
Setting
the date
Setting the time
Setting the time
Exi
t
Calibration
a) Connect fresh air to the sensor input. The icon
„ “
flashes
.
After the value stabilizes, move onto the next step using
„ “
b) Using „ “ select the concentration of the calibration gas, confirm „ “
c) Connect the calibration gas to the sensor input. The icon „ “ flashes. Wait until the value stabilizes and onfirm
„ “
d) Using „ “ swtch between „YES“ - save calibration, or „NO“ - return to the Settings menu. Confirm the selected
option „ “
e) In case of saving a new calibration, the detector will restart.
Setting the alarms
a) Using „ “ select the concentration for alarm 1. Confirm „ “
b) Using „ “ select the concentration for alarm . Confirm „ “
c) Using „ “ select the concentration for alarm 3. Confirm „ “
d) Using „ “ select the concentration for alarm 4 Confirm „ “
e) Using „ “ switch between „YES“ - save settings, or „NO“ - return to the Settings menu. Confirm the selected
option „ “
Setting the range of the 20 mA urrent loop
a) Using „ “ select the concentration appropriate to 0 mA range. Confirm „ “
b) Using „ “ switch between „YES“ - save settings, nebo „NO“ - return to the Settings menu. Confirm the selected
option „ “
Changing the password
a) Using „ “ select a number in thousands. Save the selected number „ “
b) Using „ “ select a number in hundreds. Save the selected number „ “
c) Using „ “ select a number in tens. Save the selected number „ “
d) Using „ “ select a number in units. Save the selected number „ “

13
e) Using „ “ switch between „YES“ - save settings, nebo „NO“ - return to the Settings menu. Confirm the selected
option „ “
Setting the year
a) Using „ “ select a number in thousands. Save the selected number „ “
b) Using „ “ select a number in hundreds. Save the selected number „ “
c) Using „ “ select a number in tens. Save the selected number „ “
d) Using „ “ select a number in units. Save the selected number „ “
e) Using „ “ switch between „YES“ - save settings, nebo „NO“ - return to the Settings menu. Confirm the selected
option „ “
Setting the date
a) Using „ “ select a number in thousands. Save the selected number „ “
b) Using „ “ select a number in hundreds. Save the selected number „ “
c) Using „ “ select a number in tens. Save the selected number „ “
d) Using „ “ select a number in units. Save the selected number „ “
e) Using „ “ switch between „YES“ - save settings, nebo „NO“ - return to the Settings menu. Confirm the selected
option „ “
Setting the time
a) Using „ “ select a number in thousands. Save the selected number „ “
b) Using „ “ select a number in hundreds. Save the selected number „ “
c) Using „ “ select a number in tens. Save the selected number „ “
d) Using „ “ select a number in units. Save the selected number „ “
e) Using „ “ switch between „YES“ - save settings, nebo „NO“ - return to the Settings menu. Confirm the selected
option „ “
4. Test menu „TEST“
Protected by password 0004. By enetering a wrong password, the detector returns to measurement mode.
The password can be changed in the menu SET -> PSWD.
Menu
4
-
0 mA current loop
test
Using „ “ set the output current in a 4- mA range. Return to Test menu
„ “
Digital communication
test
Using „ “ set the concentration broadcasted via RS485 in measuring range
of the sensor.. Return to Test menu „ “
Detector restart
Using „ “ switch between „YES“ - for restart, or „NO“ for returning to Test
menu. Confirm „ “.
Return to main menu
Return to main menu

14
Operation, maintenan e, inspe tion and servi e of the dete tor
1. Usage limits
To maintain proper operation of the detector it is neccesery to respect the fact, that step changes of humidity, condensation
or rapid changes of pressure can cause incorrect indication of the measured value. Each sensing technology is suited for
different methods of application, which is described below. All sensors are characterized by a smaller or larger cross-sensitivity
to other gases than those which are set. Therefore before processing project documentation we recommend to have the air in
the deployment area of the detection system analyzed.
a) atalyti sensors: Trace amount of vapors of silicon compounds and sulfur compounds cause a permanent loss of sensitivity,
which requires recalibration or replacement of the sensor. Longterm crossing of the measuring range causes a decrease in
sensitivity. In case of an atmoshpere having an oxygen content of less than 17 %, there will be an underestimaton of the
measured value. In case of an atmoshpere having an oxygen content of more than 5 %, there will be an overestimation of the
measured value.
b) ele tro hemi al sensors: Constant exposure to toxic gases or short-term exposure to gases, which greatly exceed the
maximum range of the sensor, can damage the electrochemical sensor, which requires recalibration or replacement. High
temperature along with low relative humidity have a negative effect on the sensor's lifetime. In case of an atmoshpere having
an oxygen content of less than 1 % for longer than 1 hour, there will be an underestimation of the measured value.
c) infrared sensors: Vapor acids and alkalis can etch the optical system and distort the measurements. A check or a calibration
may be neccesery.
d) semi ondu tor sensors: Short-term exposure to gases or vapors of organic solvents, which greatly exceed the maximum
range of the sensor, may damage the sensor and a recalibration or replacement may be required. In case of an atmoshpere
having na oxygen content of less than 18 %, there will be an underestimation of the measured value.
e) photoionization sensors: the UV lamp or sensor must be replaced regularly, its clogging may cause loss of signal.
2. Operation
To maintain proper operation of the detector it is neccesery to respect the fact, that the presence of certain concentrations of
gases or vapors, other than those for which the sensor is set, can cause an alarm, even if the concentration of the gas does not
exceed the set level. Given the range of disturbing gases or vapors (diluents, exhaust gases, vapors of organic substances,
disinfectants, etc.) a generally allowable concentration of interfering gases can not be determined. Data on cross-sensitivity to
certail gases are included at the apporpriate sensors. Therefore before processing project documentation we recommend to
have the air in the deployment area of the detection system analyzed
.
3. Operation/Maintenance
In case of contamination the surface can be cleaned with a slightly moistened cloth.
The seonsors have a different lifetime depending on the sensing technology used, and environmental conditions.
Characteristics of the sensors vary over time. Therefore it is required to perform regular checks and calibrations, which can be
done in two ways:
a)
1 x every six months carry out a „ alibration“ and fun tional ontrol - adjust the sesitivity of the sensor using
calibration gas and check the functionality of the system. The exact interval depends on the purity of the environment,
required accuracy and the occurrance of disturbing gases in the atmoshpere
.
b)
1 x every twelve months carry out a „ alibration“ - adjust the sesitivity of the sensor using calibration gas and check the
functionality of the system. The exact interval depends on the purity of the environment, required accuracy and the
occurrance of disturbing gases in the atmoshpere. Also carry out a „fun tional ontrol" 1 x every three months -
checking the function of the entire detection system using a test gas, which does not exceed the range of the sensor.
We recommend using gas intended for laboratory use
.
For the „fun tional ontrol“ do not use means for testing fire alarm dete tors!
Perform calibration only at certified service centers with a valid certificate of competence or the manufacturer.
For the Cze h Republi only DEGA CZ s.r.o.

15
A essories and basi types of dete tors
1. Calibration adapter/connection to the gas pump DEG G S INLET
2. Cover against splashing water DEG W TER C P
3. Funnel for gas collection DEG COLLECT C P
4. dditional Ex „d“ cable gland
M 0x1,5
Basi type of dete tors
1. Detectors with a catalytic sensor NSx-CL III LCD
Product code Detector type Detected gas Measurement
range
Measurement of
current loop
(4- 0 mA)
Resolution Calibration gas
40100158 DEGA NSM-CL 100 III LCD RE
Methane (CH4)
Natural gas
CNG
0–100 % LEL 0- 0 % LEL 0,1 % LEL Methane 0,88 %
40100159 DEGA NSP-CL 100 III LCD RE
Butane (C4H10)
Propan-Butane
LPG
0–100 % LEL 0- 0 % LEL 0,1 % LEL Butane 0,3 %
40100160 DEGA NSH-CL 100 III LCD RE
Hydrogen (H ) 0–100 % LEL 0- 0 % LEL 0,1 % LEL Hydrogen 0,8 %
40100157 DEGA NSB-CL 100 III LCD RE
Gasoline vapors 0–100 % LEL 0- 0 % LEL 0,1 % LEL Hexane 0,18 %
40100158 DEGA NSHC-CL 100 III LCD RE
Other
flammable and
combustible
gases and
vapors
according to
the selectivity
of the sensor
0–100 % LEL 0- 0 % LEL 0,1 % LEL According to the selectivity of the
sensor
2. Detectors with an electrochemical sensor NSx-EL III LCD
Product code Detector type Detected gas Measurement
range
Measurement of
current loop
(4- 0mA)
Resolution Calibration gas
40100111
DEGA NSCO-EL 1000 III LCD
Carbon
Monoxide (CO) 0–1000 ppm 0-130 ppm 1 ppm Carbon Monoxide
450 ppm
40100108 DEGA NSCO-EL 00 III LCD
Carbon
Monoxide (CO) 0– 00 ppm 0- 00 ppm 0,1 ppm Carbon Monoxide
130 ppm

16
40100109 DEGA NSCO-EL A500 III LCD
Carbon
Monoxide (CO) 0- A500 ppm 0-500 ppm 1 ppm Carbon Monoxide
130 ppm
40100110 DEGA NSCO-EL 000 III LCD
Carbon
Monoxide (CO) 0– 000 ppm 0- 000 ppm
1 ppm Carbon Monoxide
130 ppm
4010011 DEGA NSA-EL 100 III LCD
Ammonia
(NH3)
0-100 ppm
0-50 ppm
0,1 ppm
Ammonia
100 ppm
401001 8 DEGA NSA-EL 1000 III LCD
Ammonia
(NH3)
0-1000 ppm
0-300 ppm
1 ppm
Ammonia
100 ppm
40100130
DEGA NSA-EL 10000 III LCD
Ammonia
(NH3)
0-10000 ppm
0-10000 ppm
1 ppm
Ammonia
100 ppm
40100133 DEGA NSA-EL 500 III LCD
Ammonia
(NH3)
0-500 ppm
0-500 ppm
1 ppm
Ammonia
100 ppm
401001 9 DEGA NSA-EL 5000 III LCD
Ammonia
(NH3)
0-5000 ppm
0-5000 ppm
1 ppm
Ammonia
1000 ppm
4010013 DEGA NSA-EL 000 III LCD
Ammonia
(NH3)
0- 000 ppm
0- 000 ppm
1 ppm
Ammonia
1000 ppm
40100131 DEGA NSCL -EL 0 III LCD
Chlorine
(Cl ) 0- 0 ppm 0-5 ppm 0,01 ppm Chlorine
5 ppm
40100105 DEGA NSCL-EL 00 III LCD
Chlorine
(Cl )) 0- 00 ppm 0- 00 ppm 0,1 ppm Chlorine
50 ppm
40100106 DEGA NSO -EL 1 III LCD
Oxygen (O ) 0-1 % 0-1 % 0,01 % Air
40100140 DEGA NSO -EL 30 III LCD
Oxygen (O ) 0-100 % 0-30 % 0,1 % Air
40100141 DEGA NSO -EL T30 III LCD
Oxygen (O ) 0-100 % 0-30 % 0,1 % Air
4010014 DEGA NSO3-EL 5 III LCD
Ozone (O3) 0-5 ppm 0-0, ppm 0,01ppm Chlorine
5 ppm
40100144 DEGA NSO3-EL 100 III LCD
Ozone (O3) 0-100 ppm 0-100 ppm 0,1ppm Chlorine
5 ppm
40100143 DEGA NSHCL-EL 0 III LCD
Hydrogen
chloride (HCl) 0- 0 ppm 0-10 ppm 0,01 ppm Hydrogen chloride
10 ppm
401001 4 DEGA NSHCL-EL 00 III LCD
Hydrogen
chloride (HCl) 0- 00 ppm 0- 00 ppm 0,1 ppm Hydrogen chloride
5 ppm
401001 5 DEGA NSH S-EL 50 III LCD
Hydrogen sulfide
(H S) 0-50 ppm 0-15 ppm 0,1 ppm Hydrogen sulfide 5 ppm
40100119 DEGA NSH S-EL 500 III LCD
Hydrogen sulfide
(H S) 0-500 ppm 0-500 ppm 1 ppm Hydrogen sulfide 500 ppm
401001 1 DEGA NSH S-EL 100 III LCD
Hydrogen sulfide
(H S) 0-100 ppm 0-100 ppm 0,1 ppm Hydrogen sulfide 5 ppm
40100118 DEGA NSH S-EL 000 III LCD
Hydrogen sulfide
(H S) 0- 000 ppm 0- 000 ppm 1 ppm Hydrogen sulfide 000 ppm
401001 0 DEGA NSNO -EL 0 III LCD
Nitrogen dioxide
(NO ) 0- 0 ppm 0-15 ppm 0,01 ppm Nitrogen dioxide 0 ppm
40100137 DEGA NSNO -EL 100 III LCD
Nitrogen dioxide
(NO ) 0-100 ppm 0-100 ppm 0,1 ppm Nitrogen dioxide 0 ppm
40100138 DEGA NSNO -EL 500 III LCD
Nitrogen dioxide
(NO ) 0-500 ppm 0-500 ppm 1 ppm Nitrogen dioxide 100 ppm
40100139 DEGA NSNO-EL 50 III LCD
Nitric oxide (NO) 0- 50 ppm 0-15 ppm 0,1 ppm
Nitric oxide 150 ppm
40100135 DEGA NSNO-EL 5 III LCD
Nitric oxide (NO) 0- 5 ppm 0- 5 ppm 0,1 ppm
Nitric oxide 15 ppm
40100134 DEGA NSNO-EL 1000 III LCD
Nitric oxide (NO) 0-1000 ppm 0-100 ppm 1 ppm
Nitric oxide 150 ppm
40100136 DEGA NSSO -EL 0 III LCD
Sulfur dioxide
(SO ) 0- 0 ppm 0-3,5 ppm 0,01 ppm Sulfur dioxide 10 ppm
40100153 DEGA NSSO -EL 00 III LCD
Sulfur dioxide
(SO ) 0- 00 ppm 0- 00 ppm 0,1 ppm Sulfur dioxide 50 ppm
40100154 DEGA NSSO -EL 000 III LCD
Sulfur dioxide
(SO ) 0- 000 ppm 0- 000 ppm 1 ppm Sulfur dioxide 1000 ppm
40100155 DEGA NSSO -EL 100 III LCD
Sulfur dioxide
(SO ) 0-100 ppm 0-100 ppm 0,1 ppm Sulfur dioxide 50 ppm
40100150 DEGA NSSO -EL 1000 III LCD
Sulfur dioxide
(SO ) 0-1000 ppm 0-1000 ppm 1 ppm Sulfur dioxide 1000 ppm

17
40100151 DEGA NSSO -EL 10000 III LCD
Sulfur dioxide
(SO ) 0-10000 ppm 0-10000 ppm 1 ppm Sulfur dioxide 1000 ppm
4010015 DEGA NSCH O-EL 10 III LCD
Formaldehyde
(CH O) 0–10 ppm 0-0,7 ppm 0,01 ppm Carbon Monoxide 130 ppm
40100094 DEGA NSCH O-EL 1000 III LCD
Formaldehyde
(CH O) 0–1000 ppm 0-1000 ppm 1 ppm Carbon Monoxide 450 ppm
40100096 DEGA NSCH O-EL 50 III LCD
Formaldehyde
(CH O) 0–50 ppm 0-50 ppm 0,1 ppm Carbon Monoxide 130 ppm
40100095 DEGA NSC H4-EL 10 III LCD
Ethylene (C H4) 0–10 ppm 0-10 ppm 0,01 ppm Ethylene ppm
40100097 DEGA NSC H4-EL 1500 III LCD
Ethylene (C H4) 0–1500 ppm 0-1500 ppm 1 ppm Ethylene 00 ppm
40100098 DEGA NSC H4-EL 00 III LCD
Ethylene (C H4) 0– 00 ppm 0- 00 ppm 0,1 ppm Ethylene 00 ppm
40100099 DEGA NSC H4O-EL 10 III LCD
Ethylene oxide
(C H4O) 0–10 ppm 0-1,5 ppm 0,01 ppm Ethylene oxide ppm
40100100 DEGA NSC H4O-EL 100 III LCD
Ethylene oxide
(C H4O) 0–100 ppm 0-100 ppm 0,1 ppm Ethylene oxide 100 ppm
40100101 DEGA NSC H4O-EL 1000 III LCD
Ethylene oxide
(C H4O) 0–1000 ppm 0-1000 ppm 1 ppm Ethylene oxide 100 ppm
4010010 DEGA NSC H4O-EL 500 III LCD
Ethylene oxide
(C H4O) 0–500 ppm 0-500 ppm 1 ppm Ethylene oxide 100 ppm
40100103 DEGA NSH-EL 1000 III LCD
Hydrogen
(H ) 0–1000 ppm 0-400 ppm 1ppm Hydrogen
1000 ppm
40100113 DEGA NSH-EL 4000 III LCD
Hydrogen
(H ) 0–4000 ppm 0-4000 ppm 1ppm Hydrogen
1000 ppm
40100114 DEGA NSH-EL 40000 III LCD
Hydrogen
(H ) 0–400000 ppm 0–100 %
LEL
0,1 % Hydrogen 0.8% vol
40100115 DEGA NSHCN-EL 50 III LCD
Hydrogen
cyanide (HCN) 0–50 ppm 0-50 ppm 1 ppm Hydrogen sulfide 1 ppm
401001 6 DEGA NSPH3-EL 5 III LCD
Phosphine
(PH3)
0–5 ppm
0-0, ppm 0,01 ppm Hydrogen sulfide 5 ppm
40100148 DEGA NSPH3-EL 0 III LCD
Phosphine
(PH3)
0– 0 ppm
0- 0 ppm 0,01 ppm Hydrogen sulfide 5 ppm
40100145 DEGA NSPH3-EL 00 III LCD
Phosphine
(PH3)
0– 00 ppm
0- 00 ppm 0,1 ppm Hydrogen sulfide 5 ppm
40100146 DEGA NSPH3-EL 000 III LCD
Phosphine
(
PH3)
0– 000 ppm
0- 000 ppm 1 ppm Hydrogen sulfide 50 ppm
40100147 DEGA NSSIH4-EL 50 III LCD
Silane (SiH4) 0–50 ppm 0-5 ppm 0,1 ppm Hydrogen sulfide 5 ppm
40100149 DEGA NSCLO -EL 5 III LCD
Chlorine dioxide
(ClO ) 0-1 ppm 0-0,4 ppm 0,01 ppm Chlorine
5 ppm
40100107 DEGA NSH O -EL 100 III LCD
Hydrogen
peroxide
(H
O
)
0-100 ppm 0-100 ppm 0,1 ppm Sulfur dioxide 5 ppm
40100116 DEGA NSH O -EL 500 III LCD
Hydrogen
peroxide
(H
O
)
0-500 ppm 0-500 ppm 1 ppm Sulfur dioxide 50 ppm
40100117 DEGA NSHBr-EL 0 III LCD
Hydrogen
bromide
(HBr)
0- 0 ppm 0- 0 ppm 0,01 ppm Hydrogen sulfide 5 ppm
401001 DEGA NSHBr-EL 00 III LCD
Hydrogen
bromide
(HBr)
0- 00 ppm 0- 00 ppm 0,1 ppm Hydrogen sulfide 50 ppm
401001 3 DEGA NSHF-EL 10 III LCD
Hydrogen
fluoride (HF) 0-10 ppm 0-10 ppm 0,01 ppm Nitrogen dioxide 0ppm
401001 7 DEGA NSVOC-EL 0 III LCD
VOC 0- 0 ppm 0- 0 ppm 0,01 ppm VOC 0ppm
40100156 DEGA NSBR -EL 0 III LCD
Bromine (Br) 0- 0 ppm 0- 0 ppm 0,01 ppm Chlorine
5 ppm
4010009 DEGA NSBR -EL 00 III LCD
Bromine (Br) 0- 00 ppm 0- 00 ppm 0,1 ppm Chlorine
50 ppm
40100093 DEGA NSALC-EL 00 III LCD
Alkohol 0- 00 ppm 0- 00 ppm 0,1 ppm Ethanol 00 ppm
40100091 DEGA NSRCOOH-EL 100 III LCD
Organic acids 0-100 ppm 0-100 ppm 0,1 ppm Ammonia
300 ppm
40100104 DEGA NSCS -EL 100 III LCD
Carbon disulfide
(C S) 0-100 ppm 0-100 ppm 0,1 ppm Carbon disulfide 50ppm
3. Detectors with an infrared sensor NSx-IL III LCD
Product code Detector type Detected gas Measurement
range
Measurement of
current loop
(4- 0mA)
Resolution Calibration gas
40100164 DEGA NSCO -IL 5 III LCD RE
Carbon dioxide
(CO ) 0-5 % vol. 0- ,5 % vol. 0,1 % Oxid uhličitý ,5 % vol.

18
40100165 DEGA NSM-IL 100 III LCD RE
Methane (CH4)
Natural gas
CNG
0–100 % LEL 0- 0 % LEL 0,1 % Metan 0,88 % LEL
40100166 DEGA NSP-IL 100 III LCD RE
Butan / LPG
/ Propan-butan 0–100 % LEL 0- 0 % LEL 0,1 % Butan 0,3 % LEL
40100168 DEGA NSC H6-IL 100 III LCD RE
Carbon dioxide
(CO ) 0-5 % LEL 0- ,5 % LEL 0,1 % Carbon dioxide
,5 % LEL
40100169 DEGA NSC H5OH-IL 100 III LCD RE
Methane (CH4)
/ Natural gas
/Coal gas / CNG
0–100 %
LEL
0- 0 %
LEL
0,1 % Methane 0,88 %
LEL
40100173 DEGA NSC H4-IL 100 III LCD RE
Butane / LPG
/ Propane-Butane 0–100 %
LEL
0- 0 %
LEL
0,1 % Butane 0,3 %
LEL
40100171 DEGA NSC H4O-IL 100 III LCD RE
Ethane
(C H6) 0–100 %
LEL
0- 0 %
LEL
0,1 % Ethane 0 %
LEL
4010017 DEGA NSC6H14-IL 100 III LCD RE
Ethanol
(C H5OH) 0–100 %
LEL
0- 0 %
LEL
0,1 % Ethanol 0 %
LEL
40100170 DEGA NSN O-IL 1 III LCD RE
Ethylen (C H4) 0–100 %
LEL
0- 0 %
LEL
0,1 % Ethene 0 %
LEL
40100176 DEGA NSC5H1 -IL 100 III LCD RE
Etyhelen oxide
(C H4O) 0–100 %
LEL
0- 0 %
LEL
0,1 % Etyhelen oxide 0 %
LEL
40100177 DEGA NSC3H6-IL 100 III LCD RE
Hexane
(Petrol) 0–100 %
LEL
0- 0 %
LEL
0,1 % Hexane 0 %
LEL
40100175 DEGA NSHC-IL 100 III LCD RE
Nitrous oxide
(N O) 0-1 % LEL 0-0,5 % LEL 0,01 % Nitrous oxide 0,5 % LEL
4. Detectors with a semiconductor sensor NSx-SL III LCD
Product code Detector type Detected gas Measurement
range
Measurement of
current loop
(4- 0mA)
Resolution Calibration gas
40100163
DEGA NSY-SL 100 III LCD RE
Acetylene (C
H
) 0–100 % LEL
0- 0 %
LEL
0,1 % 0 % LEL
4010016 DEGA NSR-SL 000 III LCD RE
Refrigerant:
R401A, R404A,
R407C, R3 , R410A,
R1 , R
0– 000 ppm
0- 0 %
LEL
1 ppm According to the type of refrigerant
40100161 DEGA NSHFO-SL 000 III
Refrigerant:
R1 34yf
0– 000 ppm
0- 000 ppm
1 ppm According to the type of refrigerant
5. Detectors with a PID sensor NSx-PL III LCDPID
Product code Detector type Detected gas Measurement
range
Measurement of
current loop
(4- 0mA)
Resolution Calibration gas
40100178
DEGA NSVOC-PL III LCD RE PID
VOC 0–4000 ppm
0-500 ppm
0,1ppm Izobutylene 300ppm
LEL - Lowest explosion level
ALARM "PEL": Permitted exposure limit - maximal permitted value of the avarage concentration over time.
Alarm settings can be chosen freely according to customer requirements in measurement range.

19
Add-on modules
Product code
Name
Produ
ct description
40 00003
DEGA NS III Relay Modul
Internal 4
-
relay, 50 V/10 A
40 00010
DEGA NS III RS485
Internal output RS485
40 00011
DEGA
N
S
III Buzzer
Internal
buzzer on PCB, 4 VDC, 7 V
DC, 30 mA, 88 dB
Atta hments
1. Chart for setting the detector address
add
r
ess
1
3
4
5
1
ON
OFF
OFF
OFF
OFF
OFF
ON
OFF
OFF
OFF
3
O
N
ON
OFF
OFF
OFF
4
OFF
OFF
ON
OFF
OFF
5
ON
OFF
ON
OFF
OFF
6
OF
F
ON
ON
OFF
OFF
7
ON
O
N
ON
OFF
OFF
8
OFF
OFF
OFF
ON
OFF
9
ON
OFF
OFF
ON
OFF
10
OFF
ON
OFF
ON
OFF
11
ON
ON
OFF
ON
OFF
1
OF
F
OFF
ON
ON
OFF
13
ON
OFF
ON
ON
OFF
14
OFF
ON
ON
ON
OFF
15
O
N
ON
ON
ON
OFF
16
OFF
OFF
OFF
OFF
ON
2. Conversion between volumetric concentration and lower explosion limit of methane
%DMV
–
level of lower explosio
n limit concentration in %
%obj – volume sample of gas
4,4 – lower explosive limit of methane (4.4 % by volume)
1
3
4
5
17
ON
O
F
F
OFF
OFF
ON
18
OFF
ON
OFF
OFF
ON
19
ON
ON
OFF
OFF
ON
0
OFF
OFF
ON
OFF
ON
1
ON
OFF
ON
OFF
ON
OFF
ON
ON
OF
F
ON
3
ON
ON
ON
OFF
ON
4
OFF
OFF
OFF
ON
ON
5
ON
OFF
OFF
ON
ON
6
OF
F
ON
OFF
ON
ON
7
ON
ON
OFF
ON
ON
8
OFF
OFF
ON
ON
ON
9
ON
OF
F
ON
ON
ON
30
OFF
ON
O
N
ON
ON
31
ON
ON
ON
ON
ON
3
OFF
OFF
OFF
OFF
ON
3. Table of error codes
code
cause
solution
1
Sensor is not present
(EEPROM is not
communicating)
Disconnect and reconnect
the sensor, then restart the
sensor by disconnecting
and reconnecting the
power
Unknows sensor type
Contact the manufacturer
3
Type senesor type ID not
found in the table setting
LMP91000
Contact the manufact
urer
-
FW update required
4
CRC does not match the
sensor EEPROM
Sensor error
-
contact the
manufacturer
5
LMP91000 does
n
ot respond
Sensor err
or
-
contact the
manufacturer
6
The set sensor ID does n
ot
match with the connected
sensor
Checking the config
uration
of the sensor required
7
Reserved
Reserved
8
EL sensor test error
Sensor error
9
Detector is in preheating
mode
W
ait a few minutes, th
e
detector will automatically

0
enter measurement mode
10
Reserved
11
reserved
1
Error reading the internal
FLASH
Restart the detector. If the
error persists, contact the
manufacturer
13
Error reading the internal
FLASH
Restart
t
h
e detector. If the
error persists, contact the
manufacturer
14
Error reading
the internal
FLASH
Restart the detector. If the
error persists, contact the
manufacturer
15
Sensor error
Replace the sensor
16
The device exceeded the
maximum calibration
interval by 50 %
Calibrati
on is necceserry
1 9
The measured value
exceeds the range of -10%
If the error persists,
contact the manufacturer
130
The result of the AD
covertion is out of range
Contact the manufacturer
4. Signalization transmitted by the current loop 4-20 m
Measurement: The measure concentration is directly proportional to 4- 0 mA current output
Ex eeding the range of measured on entrations: Current output ranges from 0- mA
End of valid alibration: Current output transmits the actual measured concentration for 10 s in a 4- 0 mA range and
for 1 s mA current
Malfun tion: Current output will be set to 0,5 mA
Mainterna e: Current output will be set to 1 mA
5. Package content
1 x detector
1 x cable gland
This manual suits for next models
7
Table of contents
Other DEGA Gas Detector manuals