
3. Carefully mesh the splines on the angle with the splines of the Motor-to-Angle Contra-Angle
Attachment (see FIG 8).
4. Tighten knurled ring onto threads until snug.
5. Install bur according to the angle’s instruction manual.
7LWDQ(
USP/XEH)UHH
0DGHLQ WKH86$
-;;;;;
6WDU 'HQWDO
SPLINES
ANGLE
MOTOR-TO-ANGLE
CONTRA-ANGLE ATTACHMENT
TITAN-E MOTOR
TRANSMISSION GEAR
CAUTION
When angle does not seat on attachment, gears are not meshed correctly. Remove
angle; spin angle transmission gear and repeat Step 3.
Titan E
®
StarDental
®
INTERNAL
SURFACES
SWIVEL BEARING/
SPRING ASSEMBLY
CLEAN
StarDental®
Instrument Solutions
StarDental®
Instrument Solutions
7LWDQ (
6WDU'HQWDO
7LWDQ (
6WDU'HQWDO
Drive
Air
RED TIP
APPLICATOR
TROUBLESHOOTING
Problem: LACK OF POWER
Cause: Air pressure
Correction: Verify proper handpiece air pressure. Refer to Installation.
Cause: Crimped tubing
Correction: Straighten tubing inside and outside delivery head.
Cause: Swivel not threaded fully onto tubing
Correction: Refer to Installation.
Cause: Brittle motor or swivel silicone rubber seal
Correction: Replace silicone rubber seal for 2/3-line swivel with StarDental Part #203103
and 4-line with StarDental Part #202897.
Cause: Swivel O-rings worn
Correction: Replace swivel O-rings, StarDental Part #259884.
Cause: Debris build-up in handpiece
Correction: Clean handpiece. Refer to Cleaning.
Problem: SWIVEL WILL NOT DISENGAGE FROM HANDPIECE
Cause: Build-up of debris
Correction: Place 3 drops of DentaLube II, StarDental Part #262539, into the swivel/tubing
connector. Rotate swivel to circulate lube. Disengage handpiece. Contact dealer
if unable to disengage swivel. To prevent reoccurrence, refer to Cleaning.
Note: If a problem persists, contact dealer for further assistance or contact StarDental technical
service at 1-866-DTE-INFO.
Silicone Rubber Seal, 4-Line (Pkg. of 3).......................202897 ........................................202897
DentaLube II .................................................................262539 ........................................262539
ACCESSORIES
Swivel, Motor 2/3-Line ..................................................260152......................................... 260152
Swivel, Motor 4-Line .....................................................260144......................................... 260144
LubeFree Ball Bearing Auto Latch Angle ......................264146
LubeFree Ball Bearing Manual Latch Angle .................262738
LubeFree Ball Bearing Autochuck Friction Grip Angle.......264151
Ball Bearing Auto Latch Angle ......................................264182
Ball Bearing Manual Latch............................................259005
Ball Bearing Autochuck Friction Grip Angle ..................264149
CLEANING & STERILIZATION
CAUTION It is the responsibility of the user to use only FDA compliant sterilizing
equipment and accessories that are capable of meeting the manufacturers
recommendations below.
CAUTION Do not exceed 135°C
Do not use: • dry heat sterilizer • cold sterilizers • ultrasonic
cleaners • hand soaps • surface sprays • surface wipes •
chemical disinfectors • surgical milk • detergents •
immersion techniques
• sterilizer bags containing disinfectants
LIMITS ON REPROCESSING Repeated processing has minimal effect on these attachments. End of life
is normally determined by wear and damage due to use.
INSTRUCTIONS
The handpiece motor must be cleaned and sterilized before first use and
between patient use.
CONTAINMENT AND
TRANSPORTATION:
It is recommended that handpieces are reprocessed as soon as is reasonably
practical following use.
PREPARATION FOR
CLEANING:
Wear puncture-resistant gloves and protective eye wear. After each patient,
flush air by running handpiece motor for 20 seconds. Remove motor from
swivel if applicable or tubing.
CLEANING - AUTOMATED: Washer-disinfectors are not recommended.
CLEANING - MANUAL: The Titan E Motor is LubeFree and, therefore, requires no lubrication.
At times debris from the delivery system’s drive air tubing collects in the
motor. This build-up may cause sluggish motor operation. To ease, place 3
drops of DentaLube II, StarDental Part #262539 into the drive air tube of
motor (FIG 9), and alternately operate the motor in forward and reverse.
CLEANING THE SWIVEL, THE MATING MOTOR PART
Clean all internal surfaces at rear of handpiece motor with a cotton swab
soaked in isopropyl alcohol until clean (FIG. 10).
Clean all external surfaces using a small toothbrush with isopropyl alcohol.
To optimize operation of the swivel connector, clean all internal surfaces of
the motor with a cotton swab soaked in isopropyl alcohol until clean. Clean
area around bearing/spring assembly at rear of swivel using isopropyl
alcohol and a toothbrush. Dry both using syringe air.
MAINTENANCE
&LUBRICATION:
The Titan-E Motor is LubeFree and, therefore, requires no lubrication. Motor
should be routinely maintained, checked for corrosion, worn parts, etc. and
should be replaced on an as needed basis.
FIG. 8
FIG. 9
FIG. 10
DISINFECTION: Use steam sterilization exclusively for complete sterility as outlined below.
INSPECTION & FUNCTION
TESTING:
See Installation, Operating Recommendations, and/or Troubleshooting
sections of this manual.
PACKAGING: Single: A standard sterilization packaging material must be used. Ensure
that the pack is large enough to contain the item without stressing the seals.
WRAPPED INSTRUMENT
STERILIZATION:
Gravity Type Autoclave: Pre-Vacuum
Autoclave:
• Minimum of 15 minutes at 132° C • Minimum of 4
minutes at 132° C
• Minimum of 30 minutes at 121° C
Do not exceed 135° C
DRYING: Dry the motor for 30 minutes minimum using the sterilization dry cycle and
allow the motor to cool to room temperature in bag prior to use. If moisture
is present, increase your sterilizer dry cycle appropriately.
STORAGE: Store in bag after sterilization until ready to use.
ADDITIONAL INFORMATION: Do not exceed the sterilizer manufacturer’s guidelines for load capacity. Use
only paper or paper/plastic bags.
MANUFACTURER CONTACT: See first or last panels of this manual.
WARNING
REPLACEMENT PART NUMBERS
ALWAYS INSIST ON STARDENTAL PARTS FOR BEST PERFORMANCE.
DESCRIPTION
TITAN T
ITEM NUMBER
TITAN 3
ITEM NUMBER
LIMITED WARRANTY
O-Rings and Seals are specifically excluded from this warranty.
DENTALEZ warrants the Titan E Low-Speed Handpiece Motor for 1 year from the date of
purchase*.
Please note the following additional terms of our warranty policy:
• Warranty covers manufacturing defects only and does not cover defects resulting from abuse,
accidents, misuse, improper handling, cleaning, care or maintenance or the failure to follow
or observe our operating/use, maintenance and/or installation instructions. Damage resulting
from the use of chemicals, cleaners, disinfectants or sterilization is not covered under this
warranty nor is normal wear and tear. Failure to use authorized parts or an authorized repair
facility voids this warranty.
• Liability is limited to repair or replacement of the defective part(s) or products at our sole
discretion. All other liabilities, in particular liability for damages, including, without limitation,
special, exemplary, consequential or incidental damages are excluded
• THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. NO EMPLOYEE, REPRESENTATIVE OR DEALER IS AUTHORIZED
TO CHANGE THIS WARRANTY IN ANY WAY OR TO GRANT ANY OTHER WARRANTY.
*Provided conditions in warranty are met.
WARRANTY REGISTRATION
Please visit the website below to register your StarDental product. This will ensure proper registration and prompt and accurate
service should any of your purchases need evaluation or repair during the warranty period.
StarDental Products: WWW.DENTALEZ.COM/warranty
FOR NEW PRODUCTS, MANUALS AND TECHNICAL INFORMATION, CALL
1-866-DTE-INFO OR VISIT OUR WEB SITE @ WWW.DENTALEZ.COM
IMPORTANT NOTICE ABOUT RETURNING PRODUCT TO DENTALEZ INC.
Anyone wishing to return any DENTALEZ Inc. product to a DENTALEZ facility MUST obtain the proper RETURN AUTHORIZATION
NUMBER from the Customer Service Department.The RA Number must appear clearly marked on both the outside and inside of
the returned product carton. When calling for a return authorization number you will be asked to provide the following information:
To Obtain A
Return Authorization
Number Call
1-866-DTE-INFO
1. Product Name, Model Number or Part Number
2. Serial Number of Product
3. Dealer Return Purchase Order Number
4. Reason For Return
5. Copy of Invoice Showing Purchase
Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation with this procedure will
help us expedite processing of any returns.
TITAN E LOW-SPEED MOTOR