For maximum turbine life do not exceed recommended pressure.
Note: Lubricate handpiece daily with Handpiece Lubricant, StarDental Part #266188,
and prior to each sterilization procedure.
1. Disconnect handpiece.
2. Shake lubricant can well before using.
3. Thread 430 SWL RA Lubricated Lubricant Tip onto can nozzle.
4. Press handpiece onto 430 SWL RA Lubricated Lubricant Tip until handpiece stops.
5. Place rag around handpiece.
6. Holding rmly, press spray button on can for one second.
7. Remove handpiece from 430 SWL RA Lubricated Lubricant Tip.
Note: Never operate handpiece without bur in place.
8. With bur in place connect handpiece to tubing, operate for 20 seconds to expel excess lubricant.
9. Wipe off excess lubricant from handpiece.
430 SWL RA High-Speed Handpiece is designed to t a 6-pin W&H/ADEC RA-24 swivel. The 430 SWL RA connects to
both StarDental and standard 4/6-Line ISO ber optic (F.O.) tubing.
Specications
430 SWL RA Type: High speed ber optic with an RA Swivel and replaceable autochuck turbine.
(Swivel sold separately.)
Operating Air Pressure: 44 to 46 psi
Air Consumption: 1.6 SCFM @ 45 psi
No Load Speed: 410,000 rpm
Connection: 360° 6-pin W&H/ADEC RA-24 Swivel with ISO Type C (6 pin Tubing)
conguration and ber optics.
Power: 16 watts
Weight: 51g
Note: Filter air and water supplies, recommended ltration: 25 microns. Soften and deionize water. Be sure lters
operate efciently.
Handpiece Installation
Autochuck Bur Changing System
Maximum
24.0 mm
OVERALL LENGTH
Minimum
7.3 mm
Shank Length
6-pin W&H/ADEC RA-24 Swivel
Tubing
Coupler
Tubing
430 SWL RA
Handpiece
Snap 430 SWLRA handpiece onto 6-pin W&H/ADEC RA-24 Swivel.
Air and Water Adjustment
1.Air Pressure –
NOTICE Toprevent premature damage to theturbine,do notexceedrecommendedpressure(see Specications)
when using a W&H/ADEC RA-24 Swivel.
Using an air pressure gauge, StarDental®Part #262591 for 4/6-Line, set pressure to 44-46 psi.
If a gauge is not available, set pressure at delivery head allowing 1 psi pressure loss for every foot of tubing. (See
Specications for recommended operational pressure.)
For example, if you want to operate at 40 psi and you are using either 5 feet or 12 feet of tubing, then the pressure
should be set at the delivery head to compensate for the air pressure losses.
5 feet tubing 40 psi + 5 psi = 45 psi
12 feet tubing 40 psi + 12 psi = 52 psi
2. Water Pressure –
Adjust water volume and coolant air pressure at the delivery head to obtain a satisfactory spray mist.
Use standard length concentric diamond and carbide burs with .0628" nominal
diameter friction grip shanks designed for high speed handpieces. Every
StarDental Diamond Bur meets the specications at right for recommended
lengths. Contact your dealer for a complete list of StarDental Diamond Burs.
StarDental autochuck turbines have the following bur requirements:
Handpiece Operation
CAUTION Never depress the autochuck end cap button or rest the end cap against any part of the oral anatomy
during handpiece operation. Depressing the end cap while a bur is rotating can activate the chuck mechanism and
result in end cap heat-up and/or loss of bur retention.
1. To remove, grasp handpiece rmly so your thumb touches the end cap
and push until a slight ‘click’ is noticed.
2. Pull bur straight out.
3. Release end cap.
Note: Do not use burs that produce excessive noise or do not run concentrically. Do not use burs that are modied,
eccentrically designed, or are bent or ‘soft’ (non-heat treated) shanks.
Autochuck Bur Changing System
1. Insert bur into autochuck until resistance is felt.
2. Depress autochuck end cap button and continue to feed-in bur until it clicks and stops.
3. Release end cap button. Now, apply light inward force to the dental bur.
This will help set the chucking mechanism for initial start up. (Always tug
on bur after it has been installed to conrm that it is secured).
Use only burs with hardened, tempered steel shanks or carbide shanks. Every
StarDental Diamond Bur maintains required hardness. Before use, verify
concentricrotationbyoperatinghandpieceoutsidethe oralcavityina safeplace.
Sterilizable up to 135°C
135°C
DO NOT USE FOR ORAL SURGERY
BEFORE OPERATING THE HANDPIECE, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS.
SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL WARNINGS AND CAUTIONS.
Indications For Use
The 430 SWL®RA High Speed Dental Handpieces are pneumatically driven, hand-held devices intended for use
by trained dental professionals for drilling in the oral cavity. Procedures include, but are not limited to, caries
removal, restorative work and crown preparations.
866.DTE.INFO (383.4636)
www.DentalEZ.com
0344
StarDental®
1816 Colonial Village Lane
Lancaster, PA 17601 U.S.A.
Printed in the U.S.A.
©2019 DentalEZ, Inc.
Emergo Europe
Prinsessegracht 20
2514 AP The Hague
The Netherlands
EC REP
Caution: Federal law restricts this device to
sale by or on the order of a dentist.
065994 H Date of Issue 2019-03-31
430 SWL® RA
High-Speed Handpiece
Instruction Manual
Includes Lubricated and LubeFree
StarDental® Instruments
Non-Sterile
Medical Device
CAUTION
SAFETY
INSTRUCTIONS
WARNING
• Specications
• Handpiece Installation
• Handpiece Operation
• Safety Instructions
• Lubrication - For Lubricated
Handpieces ONLY
• Cleaning & Sterilization
• Troubleshooting
• Turbine Replacement
Contents
Cleaning & Sterilization
CAUTION It is the responsibility of the user to use only FDA compliant sterilizing equipment and
accessories that are capable of meeting the manufacturers recommendations below.
CAUTION Do not exceed 135°C
To prevent damage to the bur, use an autoclavable bur block.
Do not use:
• dry heat sterilizer • cold sterilizers • ultrasonic cleaners
• chemical disinfectors • immersion techniques • surgical milk
• surface wipes • hand soaps • detergents
• sterilizer bags •surface sprays
containing disinfectants
Limits On Reprocessing Repeated processing has minimal effect on this handpiece. End of life is normally
determined by wear and damage due to use.
INSTRUCTIONS
WARNING Clean and sterilize handpieces and burs as soon as practical before rst use and after each
patient.
Containment and
transportation: It is recommended that handpieces are reprocessed as soon as is reasonably practical
following use.
Preparation for cleaning: Remove handpiece from swivel. No further disassembly is required.
Cleaning - Automated: Washer-disinfectors are not recommended.
Cleaning - Manual: Cleaning Handpieces:
Internal Surface
Optical Surface
(1 of 2)
Remove debris from external surfaces using a
small brush and isopropyl alcohol. Clean
internal surfaces of the handpiece at the swivel
connection location using a small brush and
isopropyl alcohol. Optical surfaces can be
cleaned with a cotton swab soaked in isopropyl
alcohol. Dry thoroughly. Repeat after each
patient.
Cleaning Diamond Burs:
• After each use, clean the diamond bur by running it against a wet diamond cleaning stone.
• To mechanically clean, the diamond bur should be held in a solution of either enzymatic
presoak, Sparkl® Dental Instrument Cleaner StarDental Part# 205472, or water.
• Rinse the bur thoroughly and allow it to dry, then Sterilize – See Sterilization.
Lubrication: Turbine Lubrication - For Lubricated Handpieces ONLY- See 'Lubrication' instruction section
Bearing/Ring Assembly Weekly Lubrication:
StarDental®
Instrument Solutions
Red Tip
Applicator
Lubricate bearing/ring assembly with DentaLube II,
StarDental Part# 262539. Place three (3) drops as
shown. Use a small brush to distribute the lubricant.
Disinfection: Use steam sterilization exclusively for complete sterility as outlined below.
Maintenance: Use Junk-Out High Speed Chuck Cleaner Kit (263790) once a week prior to sterilization
to ensure optimum working condition of the chuck. Instructions for cleaning the chuck are
included in the Junk-Out kit.
See Turbine Replacement section of this manual for additional information.
Handpiece should be routinely maintained, checked for corrosion, worn parts, etc. and
should be replaced on an as needed basis.
Inspection & function testing: See Handpiece Operation or Troubleshooting sections of this manual.
Packaging: Single: A standard sterilization packaging material must be used. Ensure that the pack is
large enough to contain the handpiece without stressing the seals.
In sets: Handpieces may be loaded into dedicated instrument trays or general purpose
sterilization trays. Ensure that cutting edges are protected and do not exceed the tray
manufacturer’s limit per tray. Wrap the trays using appropriate packaging materials and methods.
430
SWL RA Lubricated
066186
Lubrication - For Lubricated Handpieces ONLY
• Replacement Part Numbers
• Warranty
NOTICE LubeFree Handpieces Contain A LubeFree Ceramic Bearing Turbine -
This turbine uses a sophisticated ceramic bearing design and, therefore, requires no lubrication. To
ensure the life of, and to prevent damage to, this turbine - do not lubricate it. If it gets lubricated even
once, it MUST then continue to be lubricated for the rest of its life.
To minimize risk of serious injury:
• Always insert bur the full length of the chuck. If you attempt to extend it from the chuck, the bur may dislodge during
use and possibly injure you or your patient.
• Use only recommended burs to prevent damage to the autochuck and in some cases to prevent burs from improper
or incorrect placement. An improperly clamped bur may dislodge during use, causing injury to you or your patient. A
sudden high-pitched sound or increased vibration while handpiece is operating alerts the operator to the possibility
that the chuck is not clamped properly to retain the bur or a bent bur is inserted.
To minimize risk of injury or damage to the handpiece or auto chuck:
• Replace turbine when handpiece does not run at full speed or runs noisily.
• Follow these safety instructions:
- Conrm bur is secure by tugging on it after it is loaded
- Verifyconcentric rotation byoperatinghandpieceoutsideoral cavity.Do notuse ifitdoesnot runconcentrically
or runs with excessive noise.
- Do not operate handpiece without an installed bur.
- Always use burs with overall length of less than 24mm. Shank length must be at least 7.3mm.
- When using autochuck, never depress end cap while bur is rotating.
NOTICE Anyserious incident occurringfromtheuseofthisdevice should be reportedtoStarDentalimmediately. If incident
occurred in a EU Member State the competent authority must be notied.