For maximum turbine life do not exceed recommended pressure.
Note: Lubricate handpiece daily with Handpiece Lubricant, StarDental Part #266188,
and prior to each sterilization procedure.
1. Disconnect handpiece.
2. Shake lubricant can well before using.
3. Thread 430 SWL RA Lubricated Lubricant Tip onto can nozzle.
4. Press handpiece onto 430 SWL RA Lubricated Lubricant Tip until handpiece stops.
5. Place rag around handpiece.
6. Holding rmly, press spray button on can for one second.
7. Remove handpiece from 430 SWL RA Lubricated Lubricant Tip.
Note: Never operate handpiece without bur in place.
8. With bur in place connect handpiece to tubing, operate for 20 seconds to expel excess lubricant.
9. Wipe off excess lubricant from handpiece.
430 SWL RA High-Speed Handpiece is designed to t a 6-pin W&H/ADEC RA-24 swivel. The 430 SWL RA connects to
both StarDental and standard 4/6-Line ISO ber optic (F.O.) tubing.
Specications
430 SWL RA Type: High speed ber optic with an RA Swivel and replaceable autochuck turbine.
(Swivel sold separately.)
Operating Air Pressure: 44 to 46 psi
Air Consumption: 1.6 SCFM @ 45 psi
No Load Speed: 410,000 rpm
Connection: 360° 6-pin W&H/ADEC RA-24 Swivel with ISO Type C (6 pin Tubing)
conguration and ber optics.
Power: 16 watts
Weight: 51g
Note: Filter air and water supplies, recommended ltration: 25 microns. Soften and deionize water. Be sure lters
operate efciently.
Handpiece Installation
Autochuck Bur Changing System
Maximum
24.0 mm
OVERALL LENGTH
Minimum
7.3 mm
Shank Length
6-pin W&H/ADEC RA-24 Swivel
Tubing
Coupler
Tubing
430 SWL RA
Handpiece
Snap 430 SWLRA handpiece onto 6-pin W&H/ADEC RA-24 Swivel.
Air and Water Adjustment
1.Air Pressure –
NOTICE Toprevent premature damage to theturbine,do notexceedrecommendedpressure(see Specications)
when using a W&H/ADEC RA-24 Swivel.
Using an air pressure gauge, StarDental®Part #262591 for 4/6-Line, set pressure to 44-46 psi.
If a gauge is not available, set pressure at delivery head allowing 1 psi pressure loss for every foot of tubing. (See
Specications for recommended operational pressure.)
For example, if you want to operate at 40 psi and you are using either 5 feet or 12 feet of tubing, then the pressure
should be set at the delivery head to compensate for the air pressure losses.
5 feet tubing 40 psi + 5 psi = 45 psi
12 feet tubing 40 psi + 12 psi = 52 psi
2. Water Pressure –
Adjust water volume and coolant air pressure at the delivery head to obtain a satisfactory spray mist.
Use standard length concentric diamond and carbide burs with .0628" nominal
diameter friction grip shanks designed for high speed handpieces. Every
StarDental Diamond Bur meets the specications at right for recommended
lengths. Contact your dealer for a complete list of StarDental Diamond Burs.
StarDental autochuck turbines have the following bur requirements:
Handpiece Operation
CAUTION Never depress the autochuck end cap button or rest the end cap against any part of the oral anatomy
during handpiece operation. Depressing the end cap while a bur is rotating can activate the chuck mechanism and
result in end cap heat-up and/or loss of bur retention.
1. To remove, grasp handpiece rmly so your thumb touches the end cap
and push until a slight ‘click’ is noticed.
2. Pull bur straight out.
3. Release end cap.
Note: Do not use burs that produce excessive noise or do not run concentrically. Do not use burs that are modied,
eccentrically designed, or are bent or ‘soft’ (non-heat treated) shanks.
Autochuck Bur Changing System
1. Insert bur into autochuck until resistance is felt.
2. Depress autochuck end cap button and continue to feed-in bur until it clicks and stops.
3. Release end cap button. Now, apply light inward force to the dental bur.
This will help set the chucking mechanism for initial start up. (Always tug
on bur after it has been installed to conrm that it is secured).
Use only burs with hardened, tempered steel shanks or carbide shanks. Every
StarDental Diamond Bur maintains required hardness. Before use, verify
concentricrotationbyoperatinghandpieceoutsidethe oralcavityina safeplace.
Sterilizable up to 135°C
135°C
DO NOT USE FOR ORAL SURGERY
BEFORE OPERATING THE HANDPIECE, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS.
SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL WARNINGS AND CAUTIONS.
Indications For Use
The 430 SWL®RA High Speed Dental Handpieces are pneumatically driven, hand-held devices intended for use
by trained dental professionals for drilling in the oral cavity. Procedures include, but are not limited to, caries
removal, restorative work and crown preparations.
866.DTE.INFO (383.4636)
www.DentalEZ.com
0344
StarDental®
1816 Colonial Village Lane
Lancaster, PA 17601 U.S.A.
Printed in the U.S.A.
©2019 DentalEZ, Inc.
Emergo Europe
Prinsessegracht 20
2514 AP The Hague
The Netherlands
EC REP
Caution: Federal law restricts this device to
sale by or on the order of a dentist.
065994 H Date of Issue 2019-03-31
430 SWL® RA
High-Speed Handpiece
Instruction Manual
Includes Lubricated and LubeFree
StarDental® Instruments
Non-Sterile
Medical Device
CAUTION
SAFETY
INSTRUCTIONS
WARNING
• Specications
• Handpiece Installation
• Handpiece Operation
• Safety Instructions
• Lubrication - For Lubricated
Handpieces ONLY
• Cleaning & Sterilization
• Troubleshooting
• Turbine Replacement
Contents
Cleaning & Sterilization
CAUTION It is the responsibility of the user to use only FDA compliant sterilizing equipment and
accessories that are capable of meeting the manufacturers recommendations below.
CAUTION Do not exceed 135°C
To prevent damage to the bur, use an autoclavable bur block.
Do not use:
• dry heat sterilizer • cold sterilizers • ultrasonic cleaners
• chemical disinfectors • immersion techniques • surgical milk
• surface wipes • hand soaps • detergents
• sterilizer bags •surface sprays
containing disinfectants
Limits On Reprocessing Repeated processing has minimal effect on this handpiece. End of life is normally
determined by wear and damage due to use.
INSTRUCTIONS
WARNING Clean and sterilize handpieces and burs as soon as practical before rst use and after each
patient.
Containment and
transportation: It is recommended that handpieces are reprocessed as soon as is reasonably practical
following use.
Preparation for cleaning: Remove handpiece from swivel. No further disassembly is required.
Cleaning - Automated: Washer-disinfectors are not recommended.
Cleaning - Manual: Cleaning Handpieces:
Internal Surface
Optical Surface
(1 of 2)
Remove debris from external surfaces using a
small brush and isopropyl alcohol. Clean
internal surfaces of the handpiece at the swivel
connection location using a small brush and
isopropyl alcohol. Optical surfaces can be
cleaned with a cotton swab soaked in isopropyl
alcohol. Dry thoroughly. Repeat after each
patient.
Cleaning Diamond Burs:
• After each use, clean the diamond bur by running it against a wet diamond cleaning stone.
• To mechanically clean, the diamond bur should be held in a solution of either enzymatic
presoak, Sparkl® Dental Instrument Cleaner StarDental Part# 205472, or water.
• Rinse the bur thoroughly and allow it to dry, then Sterilize – See Sterilization.
Lubrication: Turbine Lubrication - For Lubricated Handpieces ONLY- See 'Lubrication' instruction section
Bearing/Ring Assembly Weekly Lubrication:
StarDental®
Instrument Solutions
Red Tip
Applicator
Lubricate bearing/ring assembly with DentaLube II,
StarDental Part# 262539. Place three (3) drops as
shown. Use a small brush to distribute the lubricant.
Disinfection: Use steam sterilization exclusively for complete sterility as outlined below.
Maintenance: Use Junk-Out High Speed Chuck Cleaner Kit (263790) once a week prior to sterilization
to ensure optimum working condition of the chuck. Instructions for cleaning the chuck are
included in the Junk-Out kit.
See Turbine Replacement section of this manual for additional information.
Handpiece should be routinely maintained, checked for corrosion, worn parts, etc. and
should be replaced on an as needed basis.
Inspection & function testing: See Handpiece Operation or Troubleshooting sections of this manual.
Packaging: Single: A standard sterilization packaging material must be used. Ensure that the pack is
large enough to contain the handpiece without stressing the seals.
In sets: Handpieces may be loaded into dedicated instrument trays or general purpose
sterilization trays. Ensure that cutting edges are protected and do not exceed the tray
manufacturer’s limit per tray. Wrap the trays using appropriate packaging materials and methods.
430
SWL RA Lubricated
066186
Lubrication - For Lubricated Handpieces ONLY
• Replacement Part Numbers
• Warranty
NOTICE LubeFree Handpieces Contain A LubeFree Ceramic Bearing Turbine -
This turbine uses a sophisticated ceramic bearing design and, therefore, requires no lubrication. To
ensure the life of, and to prevent damage to, this turbine - do not lubricate it. If it gets lubricated even
once, it MUST then continue to be lubricated for the rest of its life.
To minimize risk of serious injury:
• Always insert bur the full length of the chuck. If you attempt to extend it from the chuck, the bur may dislodge during
use and possibly injure you or your patient.
• Use only recommended burs to prevent damage to the autochuck and in some cases to prevent burs from improper
or incorrect placement. An improperly clamped bur may dislodge during use, causing injury to you or your patient. A
sudden high-pitched sound or increased vibration while handpiece is operating alerts the operator to the possibility
that the chuck is not clamped properly to retain the bur or a bent bur is inserted.
To minimize risk of injury or damage to the handpiece or auto chuck:
• Replace turbine when handpiece does not run at full speed or runs noisily.
• Follow these safety instructions:
- Conrm bur is secure by tugging on it after it is loaded
- Verifyconcentric rotation byoperatinghandpieceoutsideoral cavity.Do notuse ifitdoesnot runconcentrically
or runs with excessive noise.
- Do not operate handpiece without an installed bur.
- Always use burs with overall length of less than 24mm. Shank length must be at least 7.3mm.
- When using autochuck, never depress end cap while bur is rotating.
NOTICE Anyserious incident occurringfromtheuseofthisdevice should be reportedtoStarDentalimmediately. If incident
occurred in a EU Member State the competent authority must be notied.
Problem: IMPROPER SPRAY
Cause: Clogged lters, tubes or water ports
Correction: Inspect and clean water lter in delivery head. Disconnect handpiece.
Depress foot control; verify water exits the swivel.
(If water does not come from swivel, refer to swivel's instruction manual)
Cause: Swivel O-rings worn
Correction: Replace o-rings
Cause: Brittle swivel rear seal
Correction: Replace rear seal
Cause: Coolant air line not working
Correction: Setup delivery unit to supply coolant air to handpiece
Cause: Clogged spray ports
Correction: Use clean-out wire to unclog ports
Problem: BUR LOCKS OR SLIPS IN AUTOCHUCK
Cause: Debris build-up in chucking mechanism
Correction: Clean with Junk-Out Chuck Cleaner Part Number 263790
Cause: Use of non-ISO burs
Correction: Always use ISO burs
Note: Always use ISO burs to prevent locking and slipping of burs in autochuck. If problems of locking and slipping
continue, replace worn autochuck turbine immediately with the appropriate StarDental Autochuck
Turbine Replacement Kit for the handpiece type (LubeFree or Lubricated) - See Replacement Part
Numbers.
If any problem persists, contact dealer for further assistance; or, contact StarDental Technical Service at 1-866-DTE-
INFO.
Problem: NO LIGHT OR INTERMITTENT LIGHT
Cause: Electrical connections
Correction: Verify connections: wall outlet-power pack, power pack-lamp control, and lamp control-tubing
connector
Cause: Crimped signal air line to lamp control
Correction: Straighten signal air line to allow air ow
Cause: Corrosion on swivel contact pins
Correction: Clean contact pins on swivel with isopropyl alcohol
Cause: Lamp burned out
Correction: Replace lamp in swivel - refer to the swivel's instruction manual.
Problem: YELLOWISH LIGHT
Cause: Debris on optical surfaces
Correction: Clean all exposed optical surfaces. (Refer to Cleaning.)
Problem: REDUCED LIGHT
Cause: Low voltage
Correction: Refer to vendor's instruction manual of electrical system
Cause: Lamp
Correction: Replace lamp in swivel - refer to the swivel's instruction manual.
Problem: LACK OF POWER
Cause: Air pressure too low
Correction: Verify proper handpiece air pressure. Refer to Handpiece Installation.
Cause: Crimped tubing
Correction: Straighten tubing inside and outside delivery head.
Cause: Debris buildup
Correction: Clean inside of housing and end cap with swab and alcohol.
Cause: Worn out turbine
Correction: Refer to Turbine Replacement.
Cause: Swivel not threaded fully onto tubing
Correction: Refer to Handpiece Installation.
Problem: WATER LEAKAGE
Cause: Swivel O-rings worn
Correction: Replace o-rings on RA Swivel with StarDental®part #266187
Cause: Swivel faulty rear seal
Correction: Replace rear seal on swivel with StarDental part #263770
Problem: SWIVEL WILL NOT DISENGAGE FROM HANDPIECE
Cause: Buildup of debris
Correction: Place 2 drops of DentaLube II into swivel/tubing connector area. Rotate swivel to circulate lube.
Disengage handpiece. Contact dealer if unable to disengage swivel. To prevent recurrence, refer to
Cleaning.
Turbine Replacement
Note: Replace the turbine when the handpiece fails to achieve full speed and/or increases sharply in noise level. The
replacement cartridge is available from your dealer and is easily replaced in the ofce. When ordering a replacement,
be sure to order the appropriate StarDental Autochuck Turbine Replacement Kit for the handpiece type (LubeFree or
Lubricated) - See Replacement Part Numbers.
1. Remove old end cap and turbine assembly.
2.Inspecthousingbefore cleaning.Removeanyobviousdebristopreventdamageandtoinsureperformanceofturbine.
3. Connect handpiece and run air through turbine housing without turbine installed for 5 seconds.
4. Clean housing thoroughly with cotton swab and isopropyl alcohol. Dry with syringe air.
5. Place turbine assembly into housing and gently insert as shown.
6. Position and engage threads of end cap into the housing by hand until snug.
7. Tighten end cap with the combination end cap wrench employing the wrench opening marked AC as shown above.
Position wrench with marked side facing end cap for best wrench control.
Autochuck Turbine Assembly
Housing
Wrench Opening
Replacement Part Numbers
Always insist on StarDental Parts for best performance.
430®SWL RA Fiber Optic Handpiece with Lubricated Autochuck Turbine........................................................................266183
Autochuck Turbine Replacement Kit for Lubricated High-Speed Handpieces..............................................................263968
430®SWL RA Fiber Optic Handpiece with LubeFree Autochuck Turbine...........................................................................266182
Autochuck Turbine Replacement Kit for LubeFree High-Speed Handpieces.................................................................263739
Autochuck Turbine End Cap....................................................................................................................................................................261294
End Cap Wrench ...........................................................................................................................................................................................203048
Air Pressure Gauge (ISO 4/6-Line) ........................................................................................................................................................262591
Cleaning Wire (Gauge: 0.008, Package of 2)....................................................................................................................................... 256719
Junk-Out High-Speed Chuck Cleaner...................................................................................................................................................263790
Handpiece Lubricant...................................................................................................................................................................................266188
6-Pin Swivel Rear Seal................................................................................................................................................................................ 263770
SPARKL®Dental Instrument Cleaner....................................................................................................................................................205472
DentaLube II...................................................................................................................................................................................................262539
Note: Swivel assemblies sold separately.
Description Item Number
Troubleshooting
Clean
debris from
inside of
housing
Bearings
Limited Warranty
O-Rings and Seals are specically excluded from this warranty.
DentalEZ®warrants the 430 SWL®RA Handpieces for 1 year from the date of purchase* with the
exceptionoftheberoptic glass whichcarriesawarrantyfor 5 years fromdate ofhandpiecepurchase*.
Please note the following additional terms of our warranty policy:
• Warranties cover manufacturing defects only and do not cover defects resulting from abuse,
improper handling, cleaning, care or maintenance, normal wear and tear or non-observance
of operating, maintenance or installation instructions. Failure to use authorized parts or an
authorized repair facility voids this warranty.
• Liability is limited to repair or replacement of the defective product at our sole discretion.
All other liabilities, in particular liability for damages, including, without limitation,
consequential or incidental damages are excluded.
• THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
NO EMPLOYEE, REPRESENTATIVE OR DEALER IS AUTHORIZED TO CHANGE THIS WARRANTY IN
ANY WAY OR TO GRANT ANY OTHER WARRANTY.
*Provided conditions in warranty are met.
For new products, manuals and technical information, call
1-866-DTE-INFO or visit our web site @ www.dentalez.com
Important Notice About Returning Product To DentalEZ Inc.
Anyone wishing to return any DentalEZ Inc. product to a DentalEZ facility MUST obtain the proper
RETURN AUTHORIZATION NUMBER from the Customer Service Department. The RA Number must appear
clearly marked on both the outside and inside of the returned product carton. When calling for a return
authorization number you will be asked to provide the following information:
To Obtain A
Return Authorization
Number Call
1-866-DTE-INFO
• Product Name, Model Number or Part Number
• Serial Number of Product
• Dealer Return Purchase Order Number
• Reason For Return
• Copy of Invoice Showing Purchase
Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation with
this procedure will help us expedite processing of any returns.
WARRANTY REGISTRATION
PLEASE VISIT THE WEBSITE BELOW TO REGISTER YOUR STARDENTAL PRODUCT. THIS WILL
ENSURE PROPER REGISTRATION AND PROMPT AND ACCURATE SERVICE SHOULD ANY OF YOUR
PURCHASES NEED EVALUATION OR REPAIR DURING THE WARRANTY PERIOD.
StarDental® Products: www.dentalez.com/warranty
Wrapped Instrument
Sterilization: Gravity Type Autoclave:
• Minimum of 15 minutes at 132°C.
• Minimum of 30 minutes at 121° C
Pre-Vacuum Autoclave:
• Minimum of 4 minutes at 132° C
Do not exceed 135° C
Drying: Dry the handpiece for 30 minutes minimum using the sterilization dry cycle and allow the
handpiece to cool to room temperature in bag prior to use. If moisture is present, increase
your sterilizer dry cycle appropriately.
Storage: Store in bag after sterilization until ready to use.
Additional information: Do not exceed the sterilizer manufacturer’s guidelines for load capacity.
Manufacturer contact: See front or back panels of this manual.
Cleaning & Sterilization, continued Troubleshooting, continued
Spray Port

This manual suits for next models

2

Other DentalEZ Dental Equipment manuals

DentalEZ StarDental Concentrix FX User manual

DentalEZ

DentalEZ StarDental Concentrix FX User manual

DentalEZ Star Titan Low-Speed User manual

DentalEZ

DentalEZ Star Titan Low-Speed User manual

DentalEZ Forest User manual

DentalEZ

DentalEZ Forest User manual

DentalEZ StarDental PROPHY STAR 3 User manual

DentalEZ

DentalEZ StarDental PROPHY STAR 3 User manual

DentalEZ Star TITAN User manual

DentalEZ

DentalEZ Star TITAN User manual

DentalEZ Star User manual

DentalEZ

DentalEZ Star User manual

DentalEZ StarDental Titan LubeFree Five Star User manual

DentalEZ

DentalEZ StarDental Titan LubeFree Five Star User manual

DentalEZ StarDental 430 SWL Flex User manual

DentalEZ

DentalEZ StarDental 430 SWL Flex User manual

DentalEZ Heritage CORE Delivery Unit User manual

DentalEZ

DentalEZ Heritage CORE Delivery Unit User manual

DentalEZ Heritage Simplicity LED Light User manual

DentalEZ

DentalEZ Heritage Simplicity LED Light User manual

DentalEZ StarDental Solara QT User manual

DentalEZ

DentalEZ StarDental Solara QT User manual

DentalEZ aXcs Basic-Duo User manual

DentalEZ

DentalEZ aXcs Basic-Duo User manual

DentalEZ Star LUBRICATED PROPHY ANGLE User manual

DentalEZ

DentalEZ Star LUBRICATED PROPHY ANGLE User manual

DentalEZ StarDental 430 SWL Torque User manual

DentalEZ

DentalEZ StarDental 430 SWL Torque User manual

DentalEZ Aeras Ramvac C3 User manual

DentalEZ

DentalEZ Aeras Ramvac C3 User manual

DentalEZ aXcs Operatory Light User manual

DentalEZ

DentalEZ aXcs Operatory Light User manual

DentalEZ Star TITAN 3 5K MOTOR User manual

DentalEZ

DentalEZ Star TITAN 3 5K MOTOR User manual

DentalEZ Star User manual

DentalEZ

DentalEZ Star User manual

DentalEZ Aeras Ramvac RV4-12 User manual

DentalEZ

DentalEZ Aeras Ramvac RV4-12 User manual

DentalEZ Simplicity LED Light Head Manual

DentalEZ

DentalEZ Simplicity LED Light Head Manual

DentalEZ Star User manual

DentalEZ

DentalEZ Star User manual

DentalEZ Star TITAN 3 User manual

DentalEZ

DentalEZ Star TITAN 3 User manual

DentalEZ aXcs Series User manual

DentalEZ

DentalEZ aXcs Series User manual

DentalEZ Aeras Star 500 Elite Handpice User manual

DentalEZ

DentalEZ Aeras Star 500 Elite Handpice User manual

Popular Dental Equipment manuals by other brands

Amanngirrbach Artex Instructions for processing

Amanngirrbach

Amanngirrbach Artex Instructions for processing

PhotoCentric Wash 15 User manual and installation guide

PhotoCentric

PhotoCentric Wash 15 User manual and installation guide

Saeyang MARATHON-N7R user manual

Saeyang

Saeyang MARATHON-N7R user manual

BG Light BLUEDENT 12 BL Operating instructions manual

BG Light

BG Light BLUEDENT 12 BL Operating instructions manual

Aseptico ALU-40TC Operation and maintenance instruction manual

Aseptico

Aseptico ALU-40TC Operation and maintenance instruction manual

GNATUS Optilight Max owner's manual

GNATUS

GNATUS Optilight Max owner's manual

DiO PROBO cure user guide

DiO

DiO PROBO cure user guide

DENTSPLY thermaprep 2 Directions for use

DENTSPLY

DENTSPLY thermaprep 2 Directions for use

Carlo De Giorgi Airbox Simplex instruction manual

Carlo De Giorgi

Carlo De Giorgi Airbox Simplex instruction manual

Dabi Atlante Bellagio owner's manual

Dabi Atlante

Dabi Atlante Bellagio owner's manual

Woodpecker i-Scan instruction manual

Woodpecker

Woodpecker i-Scan instruction manual

DEXIS NOMAD Pro 2 Operator's manual

DEXIS

DEXIS NOMAD Pro 2 Operator's manual

SCHEU IP-120 Operation manual

SCHEU

SCHEU IP-120 Operation manual

Dowell PIEZOEX1 Manual of use and maintenance

Dowell

Dowell PIEZOEX1 Manual of use and maintenance

Vatech PaX-Reve3D installation manual

Vatech

Vatech PaX-Reve3D installation manual

Dr. Hinz OrthoPreventAligner PR-M hard Patient Information

Dr. Hinz

Dr. Hinz OrthoPreventAligner PR-M hard Patient Information

IDC MIKRO ic user manual

IDC

IDC MIKRO ic user manual

B.A. International Ultimate BA-102 Operation manual

B.A. International

B.A. International Ultimate BA-102 Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.