DentalEZ Star User manual

NON-STERILE
DENTALEZ, Inc., StarDental ®Division
1816 Colonial Village Lane
Lancaster, PA 17601 USA
For maximum turbine life do not exceed recommended pressure.
NOTE: Lubricate handpiece daily with Handpiece Lubricant, StarDentalPart #266188, and prior
to each sterilization procedure. (Fig. 4)
1. Disconnect handpiece.
2. Shake lubricant can well before using.
3. Thread 430 SWL RA Lubricated Lubricant Tip onto can nozzle.
4. Press handpiece onto 430 SWL RA Lubricated Lubricant Tip until handpiece stops.
5. Place rag around handpiece.
6. Holding firmly, press spray button on can for one second.
7. Remove handpiece from 430 SWL RA Lubricated Lubricant Tip.
Note: Never operate handpiece without bur in place.
8. With bur in place connect handpiece to tubing, operate for 20 seconds to expel excess lubricant.
9. Wipe off excess lubricant from handpiece.
430 SWL RA High-Speed Handpiece is designed to fit a 6-pin W&H/ADEC
RA-24 swivel. The 430 SWL RA connects to both StarDental and standard
4/6-Line ISO fiber optic (F.O.) tubing.
430 SWL RA Type: High speed fiber optic with an RA Swivel and
replaceable autochuck turbine.
(Swivel sold separately.)
Operating Air Pressure: 44 to 46 psi
Air Consumption: 1.6 SCFM @ 45 psi
No Load Speed: 410,000 rpm
Connection: 360° 6-pin W&H/ADEC RA-24 Swivel with
ISO Type C (6 pin Tubing)
configuration and fiber optics.
Power: 16 watts
Weight: 51g
NOTE: Filter air and water supplies, recommended filtration: 25 microns.Soften and deionize water. Be sure filters operate efficiently.
HANDPIECE INSTALLATION
6-PIN W&H/ADEC RA-24 SWIVEL
TUBING
COUPLER
TUBING
430 SWL RA
HANDPIECE
Snap 430 SWL RA handpiece onto 6-pin W&H/ADEC RA-24 Swivel. (Fig. 1)
AIR AND WATER ADJUSTMENT
1. Air Pressure –
NOTICE To prevent premature damage to the turbine, do not exceed recommended pressure (see Specifications) when
using a W&H/ADEC RA-24 Swivel.
Using an air pressure gauge for 4/6-Line, set pressure to 44-46 psi.
If a gauge is not available, set pressure at delivery head allowing 1 psi pressure loss for every foot of tubing.
(See Specifications for recommended operational pressure.)
For example, if you want to operate at 40 psi and you are using either 5 feet or 12 feet of tubing, then the pressure should be
set at the delivery head to compensate for the air pressure losses.
5 feet tubing 40 psi + 5 psi = 45 psi
12 feet tubing 40 psi + 12 psi = 52 psi
2. Water Pressure –
Adjust water volume and coolant air pressure at the delivery head to obtain a satisfactory spray mist.
Use standard length concentric diamond and carbide burs with .0628" nominal diameter
friction grip shanks designed for high speed handpieces. Every StarDental Diamond
Bur meets the specifications at right for recommended lengths. Contact your dealer
for a complete list of StarDental Diamond Burs. StarDental autochuck turbines have
the following bur requirements: (Fig. 2)
Use only burs with hardened, tempered steel shanks or carbide shanks.Every StarDental
Diamond Bur maintains required hardness. Before use, verify concentric rotation by
operating handpiece outside the oral cavity in a safe place.
HANDPIECE OPERATION
Never depress the autochuck end cap button or rest the end cap against any part of the oral anatomy
during handpiece operation. Depressing the end cap while a bur is rotating can activate the chuck mechanism and
result in end cap heat-up and/or loss of bur retention.
NOTE: Do not use burs that produce excessive noise or do not run concentrically. Do
not use burs that are modified, eccentrically designed, or are bent or ‘soft’ (non-heat
treated) shanks.
AUTOCHUCK BUR CHANGING SYSTEM
1. Insert bur into autochuck until resistance is felt. (Fig. 3)
2. Depress autochuck end cap button and continue to feed-in bur until it clicks and stops.
3. Release end cap button. Now, apply light inward force to the dental bur. This will
help set the chucking mechanism for initial start up. (Always tug on bur after it
has been installed to confirm that it is secured).
4. To remove, grasp handpiece firmly so your thumb touches the end cap and push
until a slight ‘click’ is noticed.
5. Pull bur straight out.
6. Release end cap.
Sterilizable up to 135°C
135°C
The 430 SWL®RA High Speed Dental Handpieces are pneumatically driven, hand-held devices intended
for use by trained dental professionals for drilling in the oral cavity. Procedures include, but are not limited
to, caries removal, restorative work and crown preparations.
CLEANING & STERILIZATION
CAUTION It is the responsibility of the user to use only FDA compliant sterilizing equipment and
accessories that are capable of meeting the manufacturers recommendations below.
CAUTION Do not exceed 135°C
To prevent damage to the bur, use an autoclavable bur block.
Do not use:
• dry heat sterilizer • cold sterilizers • ultrasonic cleaners
• chemical disinfectors • immersion techniques • surgical milk
• surface wipes • hand soaps • detergents
• sterilizer bags • surface sprays
containing disinfectants
LIMITS ON REPROCESSING Repeated processing has minimal effect on this handpiece. End of life is normally
determined by wear and damage due to use.
INSTRUCTIONS
WARNING Handpieces and burs must be cleaned and sterilized before first use and after each patient.
CONTAINMENT AND
TRANSPORTATION:
It is recommended that handpieces are reprocessed as soon as is reasonably practical
following use.
PREPARATION FOR
CLEANING:
Remove handpiece from swivel. No further disassembly is required.
CLEANING - AUTOMATED: Washer-disinfectors are not recommended.
CLEANING - MANUAL: Cleaning Handpieces:
Internal Surface
Optical Surface
(1 of 2)
Remove debris from external surfaces using a
small brush and isopropyl alcohol. (Fig. 5)
Clean internal surfaces of the handpiece at the
swivel connection location using a small brush
and isopropyl alcohol. Optical surfaces can be
cleaned with a cotton swab soaked in isopropyl
alcohol. Dry thoroughly. Repeat after each
patient.
Cleaning Diamond Burs:
After each use, clean the diamond bur by running it against a wet diamond cleaning stone.
To mechanically clean, the diamond bur should be held in a solution of either enzymatic
presoak, Sparkl®Dental Instrument Cleaner StarDental Part# 205472, or water.
Rinse the bur thoroughly and allow it to dry, then Sterilize – See Sterilization.
LUBRICATION: Turbine Lubrication - For Lubricated Handpieces ONLY- See 'Lubrication' instruction
section
Bearing/Ring Assembly Weekly Lubrication:
StarDental®
Instrument Solutions
RED TIP
APPLICATOR
Lubricate bearing/ring assembly with
DentaLube II, StarDental Part# 262539. Place
three (3) drops as shown. Use a small brush to
distribute the lubricant.
DISINFECTION: Use steam sterilization exclusively for complete sterility as outlined below.
MAINTENANCE: Use Junk-Out High Speed Chuck Cleaner Kit (263790) once a week prior to sterilization
to ensure optimum working condition of the chuck. Instructions for cleaning the chuck are
included in the Junk-Out kit.
See Turbine Replacement section of this manual for additional information.
Handpiece should be routinely maintained, checked for corrosion, worn parts, etc. and
should be replaced on an as needed basis.
INSPECTION & FUNCTION
TESTING:
See Handpiece Operation or Troubleshooting sections of this manual.
PACKAGING: Single: A standard sterilization packaging material must be used. Ensure that the pack is
large enough to contain the handpiece without stressing the seals.
In sets: Handpieces may be loaded into dedicated instrument trays or general purpose
sterilization trays. Ensure that cutting edges are protected and do not exceed the tray
manufacturer’s limit per tray. Wrap the trays using appropriate packaging materials and
methods.
430
SWL RA Lubricated
066186
LUBRICATION - FOR LUBRICATED HANDPIECES ONLY
NOTICE LubeFree Handpieces Contain A LubeFree Ceramic Bearing Turbine -
This turbine uses a sophisticated ceramic bearing design and, therefore, requires no lubrication. To ensure the
life of, and to prevent damage to, this turbine - do not lubricate it. If it gets lubricated even once, it must then
continue to be lubricated for the rest of its life.
065994 Rev I Date of Issue 2023-03-01 ©2023 DENTALEZ, Inc.
430 SWL RA LUBRICATED
430 SWL RA LUBEFREE
DO NOT USE FOR ORAL SURGERY
Before operating the handpiece, carefully read and follow all of these instructions. Save these instructions for
future reference and observe all warnings and cautions.
INDICATIONS FOR USE
CONTENTS SPECIFICATIONS
SPECIFICATIONS
HANDPIECE INSTALLATION
HANDPIECE OPERATION
SAFETY INSTRUCTIONS
LUBRICATION
CLEANING & STERILIZATION
TROUBLESHOOTING
TURBINE REPLACEMENT
REPLACEMENT PART NUMBERS
WARRANTY
FIG. 1
AUTOCHUCK BUR CHANGING
SYSTEM
MAXIMUM
24.0 MM
OVERALL LENGTH
MINIMUM
7.3 MM
SHANK LENGTH
FIG. 2
FIG. 3
SAFETY INSTRUCTIONS
WARNING:
TO MINIMIZE RISK OF SERIOUS INJURY:
• Always insert bur the full length of the chuck. If you attempt to extend it from the chuck, the bur may dislodge during use
and possibly injure you or your patient.
TO MINIMIZE RISK OF INJURY OR DAMAGE TO HANDPIECE OR AUTO CHUCK:
• Replace turbine when handpiece does not run at full speed or runs noisily.
• Use only recommended burs. An improperly clamped or bent bur, usually indicated by sudden high-pitched sound or
increased vibration during use, may dislodge injuring you or your patient.
• FOLLOW THESE SAFETY INSTRUCTIONS:
- Confirm bur is secure by tugging on it after it is loaded
- Verify concentric rotation by operating handpiece outside oral cavity. Do not use if it does not run concentrically
or runs with
excessive noise.
- Do not operate handpiece without an installed bur.
- Always use burs with overall length of less than 24mm. Shank length must be at least 7.3 mm. This includes
most short shank burs.
(see Handpiece Operation)
- When using autochuck, never depress end cap while bur is rotating.
CAUTION:
CAUTION:
FIG. 4
FIG. 5
FIG. 6
430 RA HANDPIECE
Caution: U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
PROBLEM: IMPROPER SPRAY
Cause: Clogged filters, tubes or water ports
Correction: Inspect and clean water filter in delivery head. Disconnect handpiece.
Depress foot control; verify water exits the swivel.
(If water does not come from swivel, refer to swivel's instruction manual)
Cause: Swivel O-rings worn
Correction: Replace o-rings
Cause: Brittle swivel rear seal
Correction: Replace rear seal
Cause: Coolant air line not working
Correction: Setup delivery unit to supply coolant air to handpiece
Cause: Clogged spray ports
Correction: Use clean-out wire to unclog ports. (Fig. 8)
PROBLEM: BUR LOCKS OR SLIPS IN AUTOCHUCK
Cause: Debris build-up in chucking mechanism
Correction: Clean with Junk-Out Chuck Cleaner Part Number 263790
Cause: Use of non-ISO burs
Correction: Always use ISO burs
NOTE:Always use ISO burs to prevent locking and slipping of burs in autochuck. If problems of locking
and slipping continue, replace worn autochuck turbine immediately with the appropriate StarDental
Autochuck Turbine Replacement Kit for the handpiece type (LubeFree or Lubricated) - See
Replacement Part Numbers.
If any problem persists, contact dealer for further assistance; or, contact StarDental Technical Service at
1-866-DTE-INFO.
PROBLEM: NO LIGHT OR INTERMITTENT LIGHT
Cause: Electrical connections
Correction: Verify connections: wall outlet-power pack, power pack-lamp control, and lamp
control-tubing connector
Cause: Crimped signal air line to lamp control
Correction: Straighten signal air line to allow air flow
Cause: Corrosion on swivel contact pins
Correction: Clean contact pins on swivel with isopropyl alcohol
Cause: Lamp burned out
Correction: Replace lamp in swivel - refer to the swivel's instruction manual.
PROBLEM: YELLOWISH LIGHT
Cause: Debris on optical surfaces
Correction: Clean all exposed optical surfaces. (Refer to Cleaning.)
PROBLEM: REDUCED LIGHT
Cause: Low voltage
Correction: Refer to vendor's instruction manual of electrical system
Cause: Lamp
Correction: Replace lamp in swivel - refer to the swivel's instruction manual.
PROBLEM: LACK OF POWER
Cause: Air pressure too low
Correction: Verify proper handpiece air pressure. Refer to Handpiece Installation.
Cause: Crimped tubing
Correction: Straighten tubing inside and outside delivery head.
Cause: Debris buildup
Correction: Clean inside of housing and end cap with swab and alcohol. (Fig. 7)
Cause: Worn out turbine
Correction: Refer to Turbine Replacement.
Cause: Swivel not threaded fully onto tubing
Correction: Refer to Handpiece Installation.
PROBLEM: WATER LEAKAGE
Cause: Swivel O-rings worn
Correction: Replace o-rings on RA Swivel with StarDental part #266187
Cause: Swivel faulty rear seal
Correction: Replace rear seal on swivel with StarDental part #263770
PROBLEM: SWIVEL WILL NOT DISENGAGE FROM HANDPIECE
Cause: Buildup of debris
Correction: Place 2 drops of DentaLube II into swivel/tubing connector area. Rotate swivel to circulate
lube. Disengage handpiece. Contact dealer if unable to disengage swivel. To prevent
recurrence, refer to Cleaning.
TURBINE REPLACEMENT
NOTE: Replace the turbine when the handpiece fails to achieve full speed and/or increases sharply in noise
level. The replacement cartridge is available from your dealer and is easily replaced in the office. When
ordering a replacement, be sure to order the appropriate StarDental Autochuck Turbine Replacement
Kit for the handpiece type (LubeFree or Lubricated) - See Replacement Part Numbers.
1. Remove old end cap and turbine assembly.
2. Inspect housing before cleaning. Remove any obvious debris to prevent damage and to insure performance
of turbine.
3. Connect handpiece and run air through turbine housing without turbine installed for 5 seconds.
4. Clean housing thoroughly with cotton swab and isopropyl alcohol. Dry with syringe air.
5. Place turbine assembly into housing and gently insert as shown. (Fig. 9)
6. Position and engage threads of end cap into the housing by hand until snug.
7. Tighten end cap with the combination end cap wrench employing the wrench opening marked AC as shown
above. Position wrench with marked side facing end cap for best wrench control.
430 SWL RA Fiber Optic Handpiece with Lubricated Autochuck Turbine ................................................ 266183
Autochuck Turbine Replacement Kit for Lubricated High-Speed Handpieces...................................... 263968
430 SWL RA Fiber Optic Handpiece with LubeFree Autochuck Turbine.................................................. 266182
Autochuck Turbine Replacement Kit for LubeFree High-Speed Handpieces ....................................... 263739
Autochuck Turbine End Cap..................................................................................................................... 261294
End Cap Wrench...................................................................................................................................... 203048
Cleaning Wire (Gauge: 0.008, Package of 2)........................................................................................... 256719
Junk-Out High-Speed Chuck Cleaner ..................................................................................................... 263790
Handpiece Lubricant................................................................................................................................ 266188
6-Pin Swivel Rear Seal ............................................................................................................................ 263770
SPARKL Dental Instrument Cleaner........................................................................................................ 205472
DentaLube II ............................................................................................................................................ 262539
TROUBLESHOOTING
WRAPPED INSTRUMENT
STERILIZATION:
Gravity Type Autoclave:
• Minimum of 15 minutes at 132°C.
• Minimum of 30 minutes at 121° C
Pre-Vacuum Autoclave:
• Minimum of 4 minutes at 132° C
Do not exceed 135° C
DRYING: Dry the handpiece for 30 minutes minimum using the sterilization dry cycle and allow the
handpiece to cool to room temperature in bag prior to use. If moisture is present, increase
your sterilizer dry cycle appropriately.
STORAGE: Store in bag after sterilization until ready to use.
ADDITIONAL INFORMATION: Do not exceed the sterilizer manufacturer’s guidelines for load capacity.
MANUFACTURER CONTACT: See front or back panels of this manual.
CLEANING & STERILIZATION, CONTINUED TROUBLESHOOTING, CONTINUED
SPRAY PORT
CLEAN DEBRIS
FROM INSIDE
OF HOUSING
BEARINGS
FIG. 7
FIG. 8
AUTOCHUCK TURBINE ASSEMBLY
WRENCH OPENING HOUSING
FIG. 9
LIMITED WARRANTY
O-Rings and Seals are specifically excluded from this warranty.
DENTALEZ warrants the 430 SWL RA Handpieces for 1 year from the date of purchase* with the exception of the fiber optic
glass which carries a warranty for 5 years from date of handpiece purchase*
Please note the following additional terms of our warranty policy:
• Warranty covers manufacturing defects only and does not cover defects resulting from abuse, accidents, misuse, improper
handling, cleaning, care or maintenance or the failure to follow or observe our operating/use, maintenance and/or installation
instructions. Damage resulting from the use of chemicals, cleaners, disinfectants or sterilization is not covered under this
warranty nor is normal wear and tear. Failure to use authorized parts or an authorized repair facility voids this warranty.
• Liability is limited to repair or replacement of the defective part(s) or products at our sole discretion. All other liabilities,
in particular liability for damages, including, without limitation, special, exemplary, consequential or incidental damages
are excluded
• THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO EMPLOYEE,
REPRESENTATIVE OR DEALER IS AUTHORIZED TO CHANGE THIS WARRANTY IN ANY WAY OR TO GRANT ANY
OTHER WARRANTY.
*Provided conditions in warranty are met.
WARRANTY REGISTRATION
Please visit the website below to register your StarDental product. This will ensure proper registration and prompt and accurate
service should any of your purchases need evaluation or repair during the warranty period.
StarDental Products: WWW.DENTALEZ.COM/warranty
FOR NEW PRODUCTS, MANUALS AND TECHNICAL INFORMATION, CALL
1-866-DTE-INFO OR VISIT OUR WEB SITE @ WWW.DENTALEZ.COM
IMPORTANT NOTICE ABOUT RETURNING PRODUCT TO DENTALEZ INC.
Anyone wishing to return any DENTALEZ Inc. product to a DENTALEZ facility MUST obtain the proper RETURN AUTHORIZATION
NUMBER from the Customer Service Department. The RA Number must appear clearly marked on both the outside and inside of
the returned product carton. When calling for a return authorization number you will be asked to provide the following information:
To Obtain A
Return Authorization
Number Call
1-866-DTE-INFO
1. Product Name, Model Number or Part Number
2. Serial Number of Product
3. Dealer Return Purchase Order Number
4. Reason For Return
5. Copy of Invoice Showing Purchase
Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation with this procedure will
help us expedite processing of any returns.
REPLACEMENT PART NUMBERS
ALWAYS INSIST ON STARDENTAL PARTS FOR BEST PERFORMANCE.
NOTE: Swivel assemblies sold separately.
DESCRIPTION ITEM NUMBER

Other DentalEZ Dental Equipment manuals

DentalEZ Forest User manual

DentalEZ

DentalEZ Forest User manual

DentalEZ AERAS Star User manual

DentalEZ

DentalEZ AERAS Star User manual

DentalEZ aXcs Basic-Duo User manual

DentalEZ

DentalEZ aXcs Basic-Duo User manual

DentalEZ Aeras Ramvac RV4-12 User manual

DentalEZ

DentalEZ Aeras Ramvac RV4-12 User manual

DentalEZ aXcs Series User manual

DentalEZ

DentalEZ aXcs Series User manual

DentalEZ StarDental Titan 3 257506 User manual

DentalEZ

DentalEZ StarDental Titan 3 257506 User manual

DentalEZ StarDental Star Simplicity FX User manual

DentalEZ

DentalEZ StarDental Star Simplicity FX User manual

DentalEZ Aeras Ramvac C3 User manual

DentalEZ

DentalEZ Aeras Ramvac C3 User manual

DentalEZ Heritage CORE Delivery Unit User manual

DentalEZ

DentalEZ Heritage CORE Delivery Unit User manual

DentalEZ Simplicity LED Light Head Manual

DentalEZ

DentalEZ Simplicity LED Light Head Manual

DentalEZ Heritage CORE Chair User manual

DentalEZ

DentalEZ Heritage CORE Chair User manual

DentalEZ StarDental 430 SWL Flex User manual

DentalEZ

DentalEZ StarDental 430 SWL Flex User manual

DentalEZ StarDental PROPHY STAR 3 User manual

DentalEZ

DentalEZ StarDental PROPHY STAR 3 User manual

DentalEZ StarDental Blis-sonic K User manual

DentalEZ

DentalEZ StarDental Blis-sonic K User manual

DentalEZ StarDental Titan 3 Five Star User manual

DentalEZ

DentalEZ StarDental Titan 3 Five Star User manual

DentalEZ Forest User manual

DentalEZ

DentalEZ Forest User manual

DentalEZ TruSim User manual

DentalEZ

DentalEZ TruSim User manual

DentalEZ Star TITAN 3 User manual

DentalEZ

DentalEZ Star TITAN 3 User manual

DentalEZ StarDental 430 SWL RA User manual

DentalEZ

DentalEZ StarDental 430 SWL RA User manual

DentalEZ Star TITAN User manual

DentalEZ

DentalEZ Star TITAN User manual

DentalEZ everLight EVLCE User manual

DentalEZ

DentalEZ everLight EVLCE User manual

DentalEZ aXcs Magellan User manual

DentalEZ

DentalEZ aXcs Magellan User manual

DentalEZ RAMVAC Utility User manual

DentalEZ

DentalEZ RAMVAC Utility User manual

DentalEZ Heritage User manual

DentalEZ

DentalEZ Heritage User manual

Popular Dental Equipment manuals by other brands

Cattani TURBO SMART HP instruction manual

Cattani

Cattani TURBO SMART HP instruction manual

Refine PT-X instruction manual

Refine

Refine PT-X instruction manual

Dentsply Sirona CEREC SpeedFire operating instructions

Dentsply Sirona

Dentsply Sirona CEREC SpeedFire operating instructions

Orascoptic UNIVERSAL FLIP-UP LOUPE Instructions for use

Orascoptic

Orascoptic UNIVERSAL FLIP-UP LOUPE Instructions for use

DENTAURUM tomas torque ratchet Instructions for use

DENTAURUM

DENTAURUM tomas torque ratchet Instructions for use

Medit i700 manual

Medit

Medit i700 manual

NSK FX204 M4 Operation manual

NSK

NSK FX204 M4 Operation manual

KaVo ESTETICA E50 Life Instructions for use

KaVo

KaVo ESTETICA E50 Life Instructions for use

DENTAURUM discovery pearl Instructions for use

DENTAURUM

DENTAURUM discovery pearl Instructions for use

Durr Dental VistaScan Mini View Installation and operating instructions

Durr Dental

Durr Dental VistaScan Mini View Installation and operating instructions

Cattani HYDROCLICLONE ISO Operator's handbook

Cattani

Cattani HYDROCLICLONE ISO Operator's handbook

Carestream DENTAL CS 9600 user guide

Carestream DENTAL

Carestream DENTAL CS 9600 user guide

Dentmate Ledex WL-090+ Instructions for use

Dentmate

Dentmate Ledex WL-090+ Instructions for use

Sendoline S1-E operating instructions

Sendoline

Sendoline S1-E operating instructions

Ugin e.ON sinter fast + Instructions for use

Ugin

Ugin e.ON sinter fast + Instructions for use

EMS AIR-FLOW handy 3.0 PERIO Operation instructions

EMS

EMS AIR-FLOW handy 3.0 PERIO Operation instructions

3M ESPE Pentamix 3 operating instructions

3M ESPE

3M ESPE Pentamix 3 operating instructions

EMS FT-200 Operation instructions

EMS

EMS FT-200 Operation instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.