Replacement Part Numbers
Always insist on StarDental Parts for best performance.
Cartridge Replacement
1. Remove the end cap using the plastic wrench to turn it clockwise. (See illustration
under Cleaning.)
NOTICE Do notusepliers on the plasticbushing, damageto the bushing mightoccur.
2. Pull plastic bushing to remove cartridge from prophy head as shown below. If
cartridge is not removable by hand, thread a prophy cup or a screw-to-button
adapter in the cartridge and tug on it.
Gear
Assembly
Plastic
Bushing
Cartridge
3. Separate gear assembly from the plastic bushing by pulling apart as shown above.
4. If gears are worn, replace cartridge, StarDental Part #262719. If not, simply wipe
clean with a lint free cloth.
5. Wipetheinsideofthehousingwith alint freeclothorcottonswab,barelydampened
with isopropyl alcohol.
Note: Do not immerse head in solvent.
6. Guide the shaft of the gear assembly through the center of the plastic bushing.
Install the assembled cartridge into the angle housing.
7. Position end cap and turn counterclockwise. Hand tighten to secure. Use wrench
to fully seat – 30° maximum.
Description Item Number
Titan®LubeFree Five Star Prophy Angle
Titan 3 .............................................................................................................................................265801
Titan T ............................................................................................................................................265802
Titan LubeFree Five Star Prophy Angle Cartridge .........................................................262719
(includes cartridge and plastic bushing)
LubeFree Transmission Gear Assembly............................................................................. 262724
Titan Knuckle ............................................................................................................................... 263241
Prophy Angle End Caps (Package of 2) ..............................................................................265805
Screw-to-Button Adapter.........................................................................................................258950
End Cap Wrench (Plastic)........................................................................................................258953
Angle Disconnect Wrench.......................................................................................................203048
Titan 3 Motor-to-Angle Adapter ...........................................................................................264073
Titan T Motor-to-Angle Adapter ........................................................................................... 263970
Troubleshooting
Problem: LOW POWER
Cause: Air leak at motor
Correction: Replace gasket;
refer to Titan®Motor Instruction Manual.
Cause: Debris build-up
Correction: Refer to Cleaning.
Cause: Faulty angle
Correction: Replace cartridge.
Note: If a problem persists, contact dealer for further assistance;
or, contact StarDental®technical service at:
Limited Warranty
O-Rings and Seals are specifically excluded from this warranty.
DentalEZ®warrants the Titan®LubeFree Five Star Prophy Angle to be free of
defects in material and workmanship,under normal usage,under the following
terms:
StarDental® Products: Warranty Period*:
LubeFree Five Star Prophy Angle 1 Year from date of purchase
Please note the following additional terms of our warranty and return policy:
• Warranties cover manufacturing defects only and do not cover defects
resulting from abuse, improper handling, cleaning, care or maintenance,
normal wear and tear or non-observance of operating, maintenance or
installation instructions. Failure to use authorized parts or an authorized
repair facility voids this warranty.
• Liability is limited to repair or replacement of the defective product at our sole
discretion. All other liabilities, in particular liability for damages, including,
without limitation, consequential or incidental damages are excluded.
• THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO
EMPLOYEE,REPRESENTATIVE OR DEALER IS AUTHORIZED TO CHANGE
THIS WARRANTY IN ANY WAY OR TO GRANT ANY OTHER WARRANTY.
* Provided conditions in warranty are met.
NOTICE Do not overtighten as poor performance and damage to angle may result.
For new products, manuals and technical information, call 1-866-DTE-INFO or
visit our web site at www.dentalez.com.
Important Notice About Returning Product To The DentalEZ Inc.
Anyone wishing to return any DentalEZ Inc. product to a DentalEZ facility MUST obtain
returned product carton. When calling for a return authorization number you will be
asked to provide the following information:
To Obtain A
Return Authorization
1-866-DTE-INFO
Your cooperation with this procedure will help us expedite processing of any returns.
NOTE: Only the Gear assembly and end cap may be cleaned ultrasonically.
7. When prophy paste is consistently found in the cartridge, install a new
cartridge, StarDental Part #262719, which includes the plastic bushing.
Replacing thecartridgerenewsthe sealand protectsthegearandbushing
from abrasive paste.
8. Wipe the inside of the housing with a lint-free cloth or cotton swab.
Note: Do not immerse head in solvent.
9. Reassemble by guiding the shaft of the gear assembly through the center
of the plastic bushing. Install cartridge into angle housing.
10. Turn end cap counterclockwise and hand tighten to secure. Use wrench
to fully seat – 30° rotation maximum.
Do not overtighten as poor performance and damage to angle may result.
Cleaning
Manual
Continued:
Dry the prophy angle for 30 minutes minimum using the sterilization dry cycle and
allow the prophy angle to cool to room temperature in bag prior to use. If moisture is
present, increase your sterilizer dry cycle appropriately.
Drying:
Use steam sterilization exclusively for complete sterility as outlined below.
Disinfection:
See Installation, Operating Recommendations, and/or Troubleshooting sections of
this manual.
Inspection &
function testing:
Lubricate - Lubrication is not needed. Prophy angle is Lube Free.
Lubrication:
Angle should be routinely maintained, checked for corrosion, worn parts, etc. and
should be replaced on an as needed basis.
Maintenance:
Wrapped
Instrument
Sterilization:
Gravity Type Autoclave: Pre-Vacuum Autoclave:
Do not exceed 135° C
Single: A standard sterilization packaging material must be used. Ensure that the
pack is large enough to contain the item without stressing the seals.
Packaging:
Additional
Information:
Do not exceed the sterilizer manufacturer’s guidelines for load capacity. Use only
paper or paper/plastic bags.
Storage: Store in bag after sterilization until ready to use.
Manufacturer
Contact:
See rst or last panels of this manual.
NOTICE
WARRANTY REGISTRATION
StarDental®Products: www.dentalez.com/warranty