062731 J DateofIssue2018-03-01
StarDental®
1816 Colonial Village Lane
Lancaster, PA 17601 U.S.A.
Printed in the U.S.A.
©2018 DentalEZ, Inc.
866.DTE.INFO (383.4636)
www.DentalEZ.com
Emergo Europe
Prinsessegracht 20
2514 AP The Hague
The Netherlands
EC REP
StarDental® Instruments
0344
Titan® Five Star
LubeFree Prophy Angle
Instruction Manual
Caution: Federal law restricts this device to
sale by or on the order of a dentist.
Non-Sterile
DO NOT USE FOR ORAL SURGERY
Features
Rated for use with StarDental Motors and Attachments
Titan LubeFree Five Star Prophy Angle:
StarDental®Titan 3 Part #265801 or Titan T Part #265802
Controlled Speed Reduces Prophy Paste ‘Spin-off’
Field Replaceable Cartridge
Quick, Easy Cleaning
Compatible with Titan 3 and Titan T Systems
Steam Sterilization: 135° C Sterilizable up to 135°C
135°C
NOTE: When angle does not seat on
adapter, gears are not
meshed correctly. Remove angle; spin motor-to-angle
adapter drive gear and repeat step one.
2. Fit angle disconnect wrench, StarDental Part #203048, furnished with every angle,
to mating indentations on angle neck; tighten securely.
The prophy angle uses either a screw-on or a snap-on type polishing cup. When using
a snap-on type cup, use the screw-to-button adapter, StarDental Part #258950.
CAUTION Operate the Titan motor in the forward direction only.
If the Titan motor is operated in reverse direction, the
prophy cup will separate from the prophy angle.
Thread the screw-to-button adapter, or polishing cup, with a clockwise rotation until
snug into the hollow cartridge shaft of the prophy angle.
Unthread the adapter, or polishing cup, counterclockwise. If the screw-to-button
adapter is difcult to disengage, remove the prophy cup, and use a small screwdriver
to loosen the screw-to-button adapter.
Prophy Cup Removal
Before use, verify concentric rotation by operating the handpiece outside the oral
cavity in a safe place.
PLEASE OBSERVE SAFETY RULE
BEFORE OPERATING THE PROPHY ANGLE, CAREFULLY READ AND FOLLOW ALL OF THESE
INSTRUCTIONS. SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL
WARNINGS AND CAUTIONS.
Prophy Cup Installation
SAFETY
INSTRUCTIONS
Mating Indentations Titan
LubeFree
Prophy Angle
Disconnect
Wrench
Motor-to-Angle
Adapter
Screw-to-
Button Adapter
INSTRUCTIONS
WARNING
Limits on
Reprocessing
Repeated processing has minimal effect on these angles. End of life is
normally determined by wear and damage due to use.
Theanglemustbecleanedandsterilizedbeforerstuseandbetweenpatientuse.
CAUTION Do not exceed 135°C
Do not use:

 







Containment &
transportation:
It is recommended that angles are reprocessed as soon as is reasonably
practical following use.
Preparation
for cleaning: Wear puncture resistant gloves to prepare the prophy angle for
sterilization. Remove prophy cup or screw-to-button adapter from prophy
angle. Remove angle from motor-to-angle adaptor.
Cleaning
Automated:
Washer-disinfectors are not recommended.
1. Clean and inspect once before rst use and after each patient.
2. Clean all external surfaces with
isopropyl alcoholand a lint-free cloth
or brush.
3. Usethe plastic wrench,
StarDentalPart
#258953,
to unthread the end cap by
turning clockwise. See illustration:
Note: Older style angle - denoted by
having no marking on the end cap
of the angle (See illustrations
below) - will unthread by turning
counterclockwise.
4. Remove the cartridge per instructions in ‘Cartridge Replacement’ and
inspect as follows:
Do not use pliers on the plastic bushing,
damage to the bushing might occur.
5. If prophy paste is not found on
thecartridge,internalcleaningis
not necessary. Simply wipe off
with lint-free cloth.
6. If prophy paste is found on
the cartridge, clean off the
paste manually. The gear
assembly and end cap may be
ultrasonically cleaned.
Cleaning
Manual:
Plastic
Wrench
New Style - Unthread
End Cap in Clockwise
Direction
Cartridge
Plastic Bushing
End Cap
Housing Gear Assembly
Prophy Angle Installation for Titan 3 or Titan T Systems
1. For the Titan 3 or Titan T systems thread angle onto
motor-to-angle adapter, StarDental Titan 3 Part #264073 or
Titan T Part # 263970, by hand. Seat completely.
Installation and Operation
Motor-to-Angle
Adapter Drive Gear
NOTICE
Cleaning & Sterilization








Contents
Indications For Use
Used intraorally by trained dental professionals for drilling and preparation of dental cavities for
restoration,suchasllings,disking,cavityandcrownpreps,polishing,postandpindrillingandpinsetting.
Undernormaloperatingconditions,airisexhaustedfromthebackoftheheadofthis
handpiece. If air is directed into an open soft tissue wound or beneath the mucosa
or dermis, injury to the patient could result from air emphysema of air embolism.
This angle has no internal air or water cooling capabilities. External water for

WARNING
CAUTION 
and accessories that are capable of meeting the manufacturers recommendations
below.
WARNING
Motor Accessories Ratio Max Allowable RPM
5K MTA 1:1 5,000
20K  4:1 5,000
Old Style:

-Unthread End Cap in
CounterClockwise Direction
:
-c
34
o and L
Marked On End Cap
-Unthread End Cap in
Clockwise Direction
Replacement Part Numbers
Always insist on StarDental Parts for best performance.
Cartridge Replacement
1. Remove the end cap using the plastic wrench to turn it clockwise. (See illustration
under Cleaning.)
NOTICE Do notusepliers on the plasticbushing, damageto the bushing mightoccur.
2. Pull plastic bushing to remove cartridge from prophy head as shown below. If
cartridge is not removable by hand, thread a prophy cup or a screw-to-button
adapter in the cartridge and tug on it.
Gear
Assembly
Plastic
Bushing
Cartridge
3. Separate gear assembly from the plastic bushing by pulling apart as shown above.
4. If gears are worn, replace cartridge, StarDental Part #262719. If not, simply wipe
clean with a lint free cloth.
5. Wipetheinsideofthehousingwith alint freeclothorcottonswab,barelydampened
with isopropyl alcohol.
Note: Do not immerse head in solvent.
6. Guide the shaft of the gear assembly through the center of the plastic bushing.
Install the assembled cartridge into the angle housing.
7. Position end cap and turn counterclockwise. Hand tighten to secure. Use wrench
to fully seat – 30° maximum.
Description Item Number
Titan®LubeFree Five Star Prophy Angle
Titan 3 .............................................................................................................................................265801
Titan T ............................................................................................................................................265802
Titan LubeFree Five Star Prophy Angle Cartridge .........................................................262719
(includes cartridge and plastic bushing)
LubeFree Transmission Gear Assembly............................................................................. 262724
Titan Knuckle ............................................................................................................................... 263241
Prophy Angle End Caps (Package of 2) ..............................................................................265805
Screw-to-Button Adapter.........................................................................................................258950
End Cap Wrench (Plastic)........................................................................................................258953
Angle Disconnect Wrench.......................................................................................................203048
Titan 3 Motor-to-Angle Adapter ...........................................................................................264073
Titan T Motor-to-Angle Adapter ........................................................................................... 263970
Troubleshooting
Problem: LOW POWER
Cause: Air leak at motor
Correction: Replace gasket;
refer to Titan®Motor Instruction Manual.
Cause: Debris build-up
Correction: Refer to Cleaning.
Cause: Faulty angle
Correction: Replace cartridge.
Note: If a problem persists, contact dealer for further assistance;
or, contact StarDental®technical service at:

Limited Warranty
O-Rings and Seals are specifically excluded from this warranty.
DentalEZ®warrants the Titan®LubeFree Five Star Prophy Angle to be free of
defects in material and workmanship,under normal usage,under the following
terms:
StarDental® Products: Warranty Period*:
LubeFree Five Star Prophy Angle 1 Year from date of purchase
Please note the following additional terms of our warranty and return policy:
•  Warranties cover manufacturing defects only and do not cover defects
resulting from abuse, improper handling, cleaning, care or maintenance,
normal wear and tear or non-observance of operating, maintenance or
installation instructions. Failure to use authorized parts or an authorized
repair facility voids this warranty.
•  Liability is limited to repair or replacement of the defective product at our sole
discretion. All other liabilities, in particular liability for damages, including,
without limitation, consequential or incidental damages are excluded.
•  THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO
EMPLOYEE,REPRESENTATIVE OR DEALER IS AUTHORIZED TO CHANGE
THIS WARRANTY IN ANY WAY OR TO GRANT ANY OTHER WARRANTY.
* Provided conditions in warranty are met.
NOTICE Do not overtighten as poor performance and damage to angle may result.
For new products, manuals and technical information, call 1-866-DTE-INFO or
visit our web site at www.dentalez.com.
Important Notice About Returning Product To The DentalEZ Inc.
Anyone wishing to return any DentalEZ Inc. product to a DentalEZ facility MUST obtain


returned product carton. When calling for a return authorization number you will be
asked to provide the following information:
To Obtain A
Return Authorization

1-866-DTE-INFO

Your cooperation with this procedure will help us expedite processing of any returns.





NOTE: Only the Gear assembly and end cap may be cleaned ultrasonically.
7. When prophy paste is consistently found in the cartridge, install a new
cartridge, StarDental Part #262719, which includes the plastic bushing.
Replacing thecartridgerenewsthe sealand protectsthegearandbushing
from abrasive paste.
8. Wipe the inside of the housing with a lint-free cloth or cotton swab.
Note: Do not immerse head in solvent.
9. Reassemble by guiding the shaft of the gear assembly through the center
of the plastic bushing. Install cartridge into angle housing.
10. Turn end cap counterclockwise and hand tighten to secure. Use wrench
to fully seat – 30° rotation maximum.
Do not overtighten as poor performance and damage to angle may result.
Cleaning
Manual
Continued:
Dry the prophy angle for 30 minutes minimum using the sterilization dry cycle and
allow the prophy angle to cool to room temperature in bag prior to use. If moisture is
present, increase your sterilizer dry cycle appropriately.
Drying:
Use steam sterilization exclusively for complete sterility as outlined below.
Disinfection:
See Installation, Operating Recommendations, and/or Troubleshooting sections of
this manual.
Inspection &
function testing:
Lubricate - Lubrication is not needed. Prophy angle is Lube Free.
Lubrication:
Angle should be routinely maintained, checked for corrosion, worn parts, etc. and
should be replaced on an as needed basis.
Maintenance:
Wrapped
Instrument
Sterilization:
Gravity Type Autoclave: Pre-Vacuum Autoclave:


Do not exceed 135° C
Single: A standard sterilization packaging material must be used. Ensure that the
pack is large enough to contain the item without stressing the seals.
Packaging:
Additional
Information:
Do not exceed the sterilizer manufacturer’s guidelines for load capacity. Use only
paper or paper/plastic bags.
Storage: Store in bag after sterilization until ready to use.
Manufacturer
Contact:
See rst or last panels of this manual.
NOTICE
WARRANTY REGISTRATION



StarDental®Products: www.dentalez.com/warranty

Other DentalEZ Dental Equipment manuals

DentalEZ StarDental Solara QT User manual

DentalEZ

DentalEZ StarDental Solara QT User manual

DentalEZ Star LUBRICATED PROPHY ANGLE User manual

DentalEZ

DentalEZ Star LUBRICATED PROPHY ANGLE User manual

DentalEZ StarDental 430 SWL Torque User manual

DentalEZ

DentalEZ StarDental 430 SWL Torque User manual

DentalEZ aXcs Magellan User manual

DentalEZ

DentalEZ aXcs Magellan User manual

DentalEZ StarDental PROPHY STAR 3 User manual

DentalEZ

DentalEZ StarDental PROPHY STAR 3 User manual

DentalEZ StarDental Titan 3 Five Star User manual

DentalEZ

DentalEZ StarDental Titan 3 Five Star User manual

DentalEZ Heritage CORE Delivery Unit User manual

DentalEZ

DentalEZ Heritage CORE Delivery Unit User manual

DentalEZ RAMVAC Utility User manual

DentalEZ

DentalEZ RAMVAC Utility User manual

DentalEZ Aeras Star 500 Elite Handpice User manual

DentalEZ

DentalEZ Aeras Star 500 Elite Handpice User manual

DentalEZ StarDental Titan 3 257506 User manual

DentalEZ

DentalEZ StarDental Titan 3 257506 User manual

DentalEZ Star User manual

DentalEZ

DentalEZ Star User manual

DentalEZ Aeras 400 Flex User manual

DentalEZ

DentalEZ Aeras 400 Flex User manual

DentalEZ AERAS Star User manual

DentalEZ

DentalEZ AERAS Star User manual

DentalEZ Star Aeras 500 Elite User manual

DentalEZ

DentalEZ Star Aeras 500 Elite User manual

DentalEZ StarDental 430 SWL RA User manual

DentalEZ

DentalEZ StarDental 430 SWL RA User manual

DentalEZ Star TITAN 3 5K MOTOR User manual

DentalEZ

DentalEZ Star TITAN 3 5K MOTOR User manual

DentalEZ Star User manual

DentalEZ

DentalEZ Star User manual

DentalEZ aXcs Basic-Duo User manual

DentalEZ

DentalEZ aXcs Basic-Duo User manual

DentalEZ StarDental Star Simplicity FX User manual

DentalEZ

DentalEZ StarDental Star Simplicity FX User manual

DentalEZ Aeras Ramvac C3 User manual

DentalEZ

DentalEZ Aeras Ramvac C3 User manual

DentalEZ StarDental Concentrix FX User manual

DentalEZ

DentalEZ StarDental Concentrix FX User manual

DentalEZ Heritage Simplicity LED Light User manual

DentalEZ

DentalEZ Heritage Simplicity LED Light User manual

DentalEZ aXcs Operatory Light User manual

DentalEZ

DentalEZ aXcs Operatory Light User manual

DentalEZ Simplicity LED Light Head Manual

DentalEZ

DentalEZ Simplicity LED Light Head Manual

Popular Dental Equipment manuals by other brands

KaVo ESTETICA E30 Instructions for use

KaVo

KaVo ESTETICA E30 Instructions for use

KaVo Kerr DEXIS CariVu quick start guide

KaVo Kerr

KaVo Kerr DEXIS CariVu quick start guide

DENTAURUM Orthocryl LC Instructions for use

DENTAURUM

DENTAURUM Orthocryl LC Instructions for use

NSK Viva Ace Operation manual

NSK

NSK Viva Ace Operation manual

Dürr Dental XR 24 Easy Installation and operating instructions

Dürr Dental

Dürr Dental XR 24 Easy Installation and operating instructions

Aseptico Transport III Service manual and parts list

Aseptico

Aseptico Transport III Service manual and parts list

Miele G 7881 operating instructions

Miele

Miele G 7881 operating instructions

Morita TORQTECH CA-10RC Operation instructions

Morita

Morita TORQTECH CA-10RC Operation instructions

Renfert Vario basic 2960-0000 instruction manual

Renfert

Renfert Vario basic 2960-0000 instruction manual

Woodpecker Endo Smart+ instruction manual

Woodpecker

Woodpecker Endo Smart+ instruction manual

SCHEU MINISTAR S Operation manual

SCHEU

SCHEU MINISTAR S Operation manual

iM3 HCR-100 Operation manual

iM3

iM3 HCR-100 Operation manual

KaVo ARCUS digma operating instructions

KaVo

KaVo ARCUS digma operating instructions

Durr Dental CA 4 Installation and operating instructions

Durr Dental

Durr Dental CA 4 Installation and operating instructions

TKD TITANUS E operating instructions

TKD

TKD TITANUS E operating instructions

RAMVAC Bison Series Operation, maintenance and user manual

RAMVAC

RAMVAC Bison Series Operation, maintenance and user manual

DENTAURUM Assistent 3000 Instructions for use

DENTAURUM

DENTAURUM Assistent 3000 Instructions for use

Tiger Dental Mini Mold Instructions for use

Tiger Dental

Tiger Dental Mini Mold Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.