  9. dosatron D128R Operating instructions

dosatron D128R Operating instructions

Page 1
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 3: Unscrew and remove the black stem.
Step 2: Grab the metal wishbone and pull the piston
straight out. Set the piston aside.
This D128R Troubleshooting will cover: The medicator is clicking, but
not drawing the solution or liquid is going down into the stock tank.
D128R Troubleshooting - Part 1
NOTE: Before you start, make sure the unit is clicking.
It would be helpful to have a strap wrench, #1 Phillips
screw driver, and a large paper clip or awl.
Step 1: Unscrew and remove the lid. You may what to
use a strap wrench (Part #: TOOL01).
Scan the QR
code to watch
our D128R
Troubleshooting
video.
Page 2
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 4: Remove the top seal (Part #: J002) that is
located between the blue body and the black stem.
NOTE: Sometimes the top seal remains in the body.
If this happens, just remove it.
Step 5: Unscrew the gray hose nut if it isn’t
already removed.
Step 6: Unscrew the black cap and remove the bottom
seal (Part #: J003) between the black stem and the cap.
D128R Troubleshooting - Part 1
Page 3
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D128R Troubleshooting - Part 1
Step 9: To change the plunger rod use a # 1 Phillips
screw driver and remove the two wishbone screws and
washers.
Step 8: Next, pick up the piston and inspect the plunger
rod (Part #: P009) and slide valve (Part #: P010). The
plunger rides inside the top seal to create suction. If
there is a build up of scale or has deep scratches, it may
not draw the correct amount of solution. If your plunger
rod does have scale build up, just soak the piston
assembly in a solution of Medicator Rx
(Part #: CLEANSER-LS) for 24 to 48 hours. If the
plunger rod has scratches it will need to be replaced.
Step 7: Inspect both seals for nicks, grooves, and wear.
Make sure they aren’t gummy, sticky, or missing. If the
seals are dirty, you may be able to just clean them and
reinstall. But if they are damaged or missing, it is
denitely time to replace.
Page 4
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D128R Troubleshooting - Part 1
Step 12: Slide the actuator assembly out.
Step 11: Remove the four screws in the center of
the piston.
Step 10: Spread the wishbone and remove from the
inlet valve.
Page 5
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D128R Troubleshooting - Part 1
Step 15: Remove the plunger rod (Part #: P009) by
lifting it up slightly and out. Set aside the actuator
(Part #: MP001).
Step 14: Remove the push rod (Part #: P008) by
sliding it out.
Step 13: Unhook and remove the actuator springs
(Part #: M009).
Page 6
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 18: To reinstall the slide valve onto the new plunger
rod slide the spring over the slide valve. Push to the
bottom and grasp rmly.
Step 17: Now push the metal pin (Part #: M005) out
using the paper clip. This allows you to remove the
spring from the slide valve.
D128R Troubleshooting - Part 1
Step 16: To remove the slide valve from the plunger rod
you will need a large paper clip or awl. Push the spring
down and hold rmly.
Page 7
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 20: Pick up the actuator and install the new plunger
rod by sliding it into the actuator and pulling it down to
lock. Make sure the head of the plunger rod is ush.
Step 19: Insert the plunger and reinstall the pin. The
center hole is the 1:128 setting. If you need 1:200, install
the pin in the top hole, if you need 1:100, install the pin in
the bottom hole. Make sure the pin is ush so the spring
doesn’t get stuck.
D128R Troubleshooting - Part 1
Step 21: Install the push rod with the long portion up
making sure to insert it with the groove side inward.
Page 8
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 24: To hold the assembly in place use your index
and middle ngers to apply pressure on the base.
Step 23: Align the push rod and inlet valve inside
the piston shell and slide the actuator assembly
in completely.
Step 22: Install the actuator springs.
D128R Troubleshooting - Part 1
Page 9
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D128R Troubleshooting - Part 1
Step 27: Test the piston by pressing on the top and
bottom of the push rod. It should click both times.
Step 26: Reinstall the wishbone spring, then reinstall the
washers and screws, hand tight.
Step 25: Now reinstall the four screws and tighten in a
star pattern.
Page 10
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 2: Push the spring down and hold rmly.
NOTE: To replace the slide valve and spring you will
need a large paper clip or awl.
Step 1: Inspect the bottom seal (Part #: J003), slide
valve (Part #: P010), and spring (Part #: M003).
If worn, damaged, deformed, or broken, replace
the parts.
D128R Troubleshooting - Part 2
This D128R Troubleshooting will cover: Water going down into the
stock tank.
Step 3: Using the paper clip, push the metal pin
(Part #: M005) out.
Page 11
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 5: Push the slide valve into the plunger rod and
reinstall the pin. The center hole is the 1:128 setting.
If you need 1:200, install the pin into the top hole, if you
need 1:100, install the pin into the bottom hole. Make
sure the pin is ush so the spring doesn’t get stuck.
Step 4: Install the new slide valve and spring onto the
plunger rod. To do this slide the new spring over the top
of the slide valve and hold rmly.
D128R Troubleshooting - Part 2
Step 6: To reassemble, place the top seal with the raised
center portion facing upward inside the body.
Page 12
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 8: Insert the bottom seal, raised portion facing up,
into the black cap and thread onto the stem, hand tight.
D128R Troubleshooting - Part 2
Step 7: Thread the black stem into the body, hand tight.
Step 9: Reattach the gray hose nut.
Page 13
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D128R Troubleshooting - Part 2
Step 11: Replace the black lid and hand tighten.
Step 10: Pick up the blue body and insert the piston
assembly into the body and slide it down as far as it
will go.
Step 12: One last thing. Take a look at the top of the
suction tube. If it ares out, cut o󰀨 an inch or so and
reinstall. Or, if the suction tube is cracked, has holes,
is rigid, or looks really old, replace.
Page 14
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D128R Troubleshooting - Part 3
Step 3: Inspect the wishbone (Part #: MP014) on the
top. If you see anything missing, replace.
Step 2: Grab the metal wishbone and pull the piston
straight out. Set the blue body aside.
Step 1: Remove the lid. You may want to use a strap
wrench (Part #: TOOL01).
This D128R Troubleshooting will cover: Medicator not clicking.
Page 15
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 5: In order to check the piston shell, remove the
piston and check the upper and lower anges on the
piston shell. If they are worn or have scratches the piston
shell needs to be replaced.
Step 4: Inspect the actuator springs (Part #: M009) on
the bottom. If you see anything missing, replace.
D128R Troubleshooting - Part 3
Page 16
800-523-8499 lwww.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 8: Hand tighten the lid.
Step 7: Insert the piston back into the blue body.
D128R Troubleshooting - Part 3
Note: To watch the video go to Dosatron’s video page www.dosatronusa.com/videos.aspx or
YouTube Channel www.youtube.com/user/DosatronIntl

Other manuals for D128R

1

Other dosatron Laboratory Equipment manuals

dosatron DM11F Operating instructions

dosatron

dosatron DM11F Operating instructions

dosatron D45RE15 Operating instructions

dosatron

dosatron D45RE15 Operating instructions

dosatron D 25 F 1 Assembly instructions

dosatron

dosatron D 25 F 1 Assembly instructions

dosatron D14MZ3000 Operating instructions

dosatron

dosatron D14MZ3000 Operating instructions

dosatron D14MZ3000AFII User manual

dosatron

dosatron D14MZ3000AFII User manual

dosatron D14MZ2 Programming manual

dosatron

dosatron D14MZ2 Programming manual

dosatron D 25 F 1 Operating instructions

dosatron

dosatron D 25 F 1 Operating instructions

dosatron D8RE2 Operating instructions

dosatron

dosatron D8RE2 Operating instructions

dosatron D14 MZ 3000 User manual

dosatron

dosatron D14 MZ 3000 User manual

dosatron D8RE3000 Operating instructions

dosatron

dosatron D8RE3000 Operating instructions

dosatron D40WL3000NIEAF User manual

dosatron

dosatron D40WL3000NIEAF User manual

dosatron D14MZ2VFII User manual

dosatron

dosatron D14MZ2VFII User manual

dosatron D14MZ2 Assembly instructions

dosatron

dosatron D14MZ2 Assembly instructions

dosatron MKD14MZ3000AF User manual

dosatron

dosatron MKD14MZ3000AF User manual

dosatron D25RE2 Operating instructions

dosatron

dosatron D25RE2 Operating instructions

dosatron D14MZ10 Operating instructions

dosatron

dosatron D14MZ10 Operating instructions

Popular Laboratory Equipment manuals by other brands

Westlab 665-259 manual

Westlab

Westlab 665-259 manual

Accu-Scope DS-360 manual

Accu-Scope

Accu-Scope DS-360 manual

Drawell DW-280B 18L manual

Drawell

Drawell DW-280B 18L manual

Jouan B4i Service manual

Jouan

Jouan B4i Service manual

Dionex P680 Quick installation guide

Dionex

Dionex P680 Quick installation guide

MedKlinn Home Asens+ Series operating manual

MedKlinn

MedKlinn Home Asens+ Series operating manual

Okolab H201-LG-DLS user manual

Okolab

Okolab H201-LG-DLS user manual

AGROJECT AGROJECT user manual

AGROJECT

AGROJECT AGROJECT user manual

NACALAI TESQUE COSMOSIL quick start guide

NACALAI TESQUE

NACALAI TESQUE COSMOSIL quick start guide

Bürkle StopCock Directions for use

Bürkle

Bürkle StopCock Directions for use

ARBORJET AccuFlo 070-3100 user manual

ARBORJET

ARBORJET AccuFlo 070-3100 user manual

CEM Discover SPS installation instructions

CEM

CEM Discover SPS installation instructions

Systec VX Series operating manual

Systec

Systec VX Series operating manual

Zass ZSHG 01 operating manual

Zass

Zass ZSHG 01 operating manual

HAVER & BOECKER HAVER EML 200 Premium operating instructions

HAVER & BOECKER

HAVER & BOECKER HAVER EML 200 Premium operating instructions

Quincy lab Analog Series operating manual

Quincy lab

Quincy lab Analog Series operating manual

Xylem WTW Turb 750 IR/T operating manual

Xylem

Xylem WTW Turb 750 IR/T operating manual

Bibby Sterilin Techne FB-08 owner's manual

Bibby Sterilin

Bibby Sterilin Techne FB-08 owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.