dosatron D45RE15 Operating instructions

Page 1
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 3: You should see the liquid begin moving up the suction
hose. If you don’t see liquid moving up the hose or you didn’t
feel suction on your nger, turn the water off. Open a valve
downstream to release the pressure.
Step 2: Then, check to see if you feel suction on your nger.
If so, just cut 1/2” off your suction hose and reattach.
NOTE: To service your Dosatron D45RE15, you will need a strap
wrench, a # 1 and 2 phillips head screw driver, a at blade screw
driver and needle nose pliers.
Step 1: To begin, make sure the strainer is clear of debris. While
the unit is still installed, turn the water on. Unscrew and remove
the suction hose nut and hose.
D45RE15 Troubleshooting - Part 1
This D45RE15 Troubleshooting will cover: The injector is clicking, but not drawing
solution or liquid is going down into the stock tank.
Scan the QR code to
watch our D45RE15 -
Troubleshooting video.
TOOL-
KIT
Tool Kit
P170 Check valve spring
J051 Check valve seal
M072 Check valve cone
M221 Actuator Springs (2)
TOOL01 Strap Wrench
Page 2
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 6: Pull the injection stem assembly straight down and out of
the lower housing.
Step 5: Now, unscrew the four screws around the injection sleeve.
Step 4: Unscrew the suction hose nut and remove the suction
hose. Set both aside.
D45RE15 Troubleshooting - Part 1
Page 3
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 9: If you see damage, or don’t remember the last time
you changed the plunger seal, consider changing the seal
(Part #: J009) now. To change the plunger seal, pinch and roll it off
of the plunger.
Step 8: Take a close look at the plunger seal for knicks, scratches,
or ware.
Step 7: Grasp the plunger and pull it straight down.
D45RE15 Troubleshooting - Part 1
Page 4
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 12: Hand tighten the screws carefully.
Step 11: To reinstall the injection stem, slide it into the lower
housing and align the four screws and the metal retainer into
the body.
Step 10: Install the new seal into the groove onto the plunger.
D45RE15 Troubleshooting - Part 1
Page 5
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 13: Now, thread the suction hose nut onto the hose, slide the
hose onto the barb and hand tighten the nut. Slowly, turn the water
on and your injector should now be drawing the concentrate.
D45RE15 Troubleshooting - Part 1
Page 6
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 2
Step 2: Unscrew and remove the check valve barb.
Step 1: If your injector is pushing water down into the concentrate
tank, it’s time to investigate the check valve. To do this remove the
hose nut and hose and set aside.
Step 3: Remove the check valve seal (Part #: J051),
cone (Part #: M072) and spring (Part #: P170).
Page 7
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 2
Step 6: Now, position the new check valve seal with the groove
side up inside the check valve barb.
Step 5: Insert the new check valve spring into the new cone and
press it into the stem.
Step 4: Inspect all the check valve parts. If they are dirty, you may
be able to soak, brush off, and reinstall. If you see damage or you
don’t remember the last time you changed these parts, now is a
good time to replace them.
Page 8
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 2
Step 9: Slide it onto the barb and hand tighten the nut. This should
stop the water from dripping back into the concentrate tank.
Step 8: Thread the suction hose nut onto the hose.
Step 7: Hand tighten.
Page 9
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 3
Step 2: Using the strap wrench, loosen the body nut and remove.
Step 1: Unscrew and remove the suction hose nut and hose.
Step 3: Pull the lower housing straight down.
This D45RE15 Troubleshooting will cover: The injector isn’t clicking.
Page 10
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 3
Step 5: Inspect the white valve support on the bottom.
Step 4: Lift the piston up and out of the lower housing.
Set the lower housing aside.
Step 6: Press the push rod down and press the two actuator
springs (Part #: M221) on the top. If they are broken or missing,
they will have to be replaced.
Page 11
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 3
Step 7: You can check the piston shell two ways. First, by
inspecting the upper and lower anges to see if they are
worn or have deep scratches.
Step 8: The second test is to check the piston shell’s t inside the
housing. To do this slide the piston into the upper housing.
It should be a snug t.
Step 9: Now, slide the piston into the lower housing. It should also
be a snug t. If it feels loose or it is worn, the piston shell needs to
be replaced (Part #: MPDI170).
Page 12
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 3
Step 11: Then, remove the four screws around the plunger rod.
Step 10: To replace the piston shell, use the # 1 screw driver and
remove the valve support screw and plastic retainer.
Step 12: Press the push rod up.
Page 13
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 3
Step 14: Next, remove the four screws in the center of the piston
shell to release the actuator assembly.
Step 13: Pry the valve support off of the inlet valve with a at
blade screw driver and set aside.
Step 15: Pull the actuator assembly out and discard the old
piston shell.
Page 14
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 3
Step 17: Then, position the actuator assembly in the shell as
shown and hold rmly.
Step 16: Using the needle nosed pliers, position the four screws
in the center of the new piston shell.
Step 18: Now, hand tighten the four screws in a star pattern.
Page 15
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 3
Step 20: Reinstall the valve support by sliding it onto the
inlet valve.
Step 19: Press the white plastic portion of the actuator side by
side. If there is movement, retighten the screws.
Step 21: Next, position the retainer and hand tighten the screw.
Page 16
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 3
Step 23: Carefully hand tighten them in a star pattern.
Step 22: Then reinstall the four screws in the black plastic
retainer and align the pin in the plunger rod as shown.
Step 24: Test the piston by pressing the push rod from the top
and bottom. It should click each time.
Page 17
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 3
Step 26: Carefully slide the piston into the lower housing
about half way.
Note: This step is very important. Inspect the large o-ring in
the groove on the top of the lower housing.
Step 25: To nish up, make sure the push rod is in the
down position.
Important
Step 27: Carefully press the piston into the upper housing. Align
the housing and hand tighten the body nut.
Page 18
800-523-8499
www.dosatronusa.com
Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
D45RE15 Troubleshooting - Part 3
Step 29: Slide it onto the barb and hand tighten the nut.
Step 28: Thread the suction hose nut onto the hose.
Step 30: Slowly turn the water on and your injector should now
be clicking and drawing the concentrate.
Scan the
QR code to
go to
Dosatron’s
video page.
Scan the QR
code to go to
Dosatron’s
YouTube
Channel.

This manual suits for next models

1

Other dosatron Laboratory Equipment manuals

dosatron D25RE2 Operating instructions

dosatron

dosatron D25RE2 Operating instructions

dosatron D8RE3000 Operating instructions

dosatron

dosatron D8RE3000 Operating instructions

dosatron D14MZ3000 Operating instructions

dosatron

dosatron D14MZ3000 Operating instructions

dosatron D14MZ2 Programming manual

dosatron

dosatron D14MZ2 Programming manual

dosatron MKD14MZ3000AF User manual

dosatron

dosatron MKD14MZ3000AF User manual

dosatron D 25 F 1 Assembly instructions

dosatron

dosatron D 25 F 1 Assembly instructions

dosatron D8RE2 Operating instructions

dosatron

dosatron D8RE2 Operating instructions

dosatron D14MZ3000AFII User manual

dosatron

dosatron D14MZ3000AFII User manual

dosatron D14MZ2 Assembly instructions

dosatron

dosatron D14MZ2 Assembly instructions

dosatron DM11F Operating instructions

dosatron

dosatron DM11F Operating instructions

dosatron D 25 F 1 Operating instructions

dosatron

dosatron D 25 F 1 Operating instructions

dosatron D40WL3000NIEAF User manual

dosatron

dosatron D40WL3000NIEAF User manual

dosatron D14 MZ 3000 User manual

dosatron

dosatron D14 MZ 3000 User manual

dosatron D14MZ10 Operating instructions

dosatron

dosatron D14MZ10 Operating instructions

dosatron D14MZ2VFII User manual

dosatron

dosatron D14MZ2VFII User manual

dosatron D128R Operating instructions

dosatron

dosatron D128R Operating instructions

Popular Laboratory Equipment manuals by other brands

Westlab 665-259 manual

Westlab

Westlab 665-259 manual

Accu-Scope DS-360 manual

Accu-Scope

Accu-Scope DS-360 manual

Drawell DW-280B 18L manual

Drawell

Drawell DW-280B 18L manual

Jouan B4i Service manual

Jouan

Jouan B4i Service manual

Dionex P680 Quick installation guide

Dionex

Dionex P680 Quick installation guide

MedKlinn Home Asens+ Series operating manual

MedKlinn

MedKlinn Home Asens+ Series operating manual

Okolab H201-LG-DLS user manual

Okolab

Okolab H201-LG-DLS user manual

AGROJECT AGROJECT user manual

AGROJECT

AGROJECT AGROJECT user manual

NACALAI TESQUE COSMOSIL quick start guide

NACALAI TESQUE

NACALAI TESQUE COSMOSIL quick start guide

Bürkle StopCock Directions for use

Bürkle

Bürkle StopCock Directions for use

ARBORJET AccuFlo 070-3100 user manual

ARBORJET

ARBORJET AccuFlo 070-3100 user manual

CEM Discover SPS installation instructions

CEM

CEM Discover SPS installation instructions

Systec VX Series operating manual

Systec

Systec VX Series operating manual

Zass ZSHG 01 operating manual

Zass

Zass ZSHG 01 operating manual

HAVER & BOECKER HAVER EML 200 Premium operating instructions

HAVER & BOECKER

HAVER & BOECKER HAVER EML 200 Premium operating instructions

Quincy lab Analog Series operating manual

Quincy lab

Quincy lab Analog Series operating manual

Xylem WTW Turb 750 IR/T operating manual

Xylem

Xylem WTW Turb 750 IR/T operating manual

Bibby Sterilin Techne FB-08 owner's manual

Bibby Sterilin

Bibby Sterilin Techne FB-08 owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.