manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Dräger
  6. •
  7. Medical Equipment
  8. •
  9. Dräger PAS Colt Series User manual

Dräger PAS Colt Series User manual

PAS®Colt Series
Airline/Short duration respiratory protection equipment with PAS®ASV (airline whistle) Instructions for Use
i
PAS® is a registered trademark of Dräger 3360044 (A3-D-P) Page 1 of 3
3360044
© Dräger Safety UK Limited
Edition 05 – November 2018 (Edition 01 – June 2013)
Subject to alteration
Draeger Safety UK Limited
Ullswater Close Tel +44 1670 352 891
Blyth, NE24 4RG Fax +44 1670 356 266
United Kingdom www.draeger.com
1 For your safety
1.1 General safety statements
●Before using this product, carefully read the Instructions for Use.
●Strictly follow the Instructions for Use. The user must fully understand
and strictly observe the instructions. Use the product only for the
purposes specified in the Intended Use section of this document.
●Do not dispose of the Instructions for Use. Ensure that they are
retained and appropriately used by the product user.
●Only fully trained and competent users are permitted to use this
product.
●Comply with all local and national rules and regulations associated
with this product.
●Only trained and competent personnel are permitted to inspect, repair and
service the product. Dräger recommend a Dräger service contract for all
maintenance activities and that all repairs are carried out by Dräger.
●Properly trained service personnel must inspect and service this
product as detailed in the Maintenance section of this document.
●Use only genuine Dräger spare parts and accessories, or the proper
functioning of the product may be impaired.
●Do not use a faulty or incomplete product, and do not modify the
product.
●Notify Dräger in the event of any component fault or failure.
●The air supply shall meet the requirements for breathing air according
to EN 12021.
1.2 Definitions of alert icons
Alert icons are used in this document to provide and highlight text that
requires a greater awareness by the user. A definition of the meaning of
each icon is as follows:
WARNING
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided,
could result in physical injury or damage to the product or
environment. It may also be used to alert against unsafe practices.
NOTICE
Indicates additional information on how to use the product.
2 Description
2.1 Product overview
This variant of the Dräger PAS®Colt Series provides respiratory protection
for working in a contaminated environment using the air cylinder or an airline.
The equipment is available as a 10 minute or 15 minute short-duration
version. These are nominal durations available for self-contained use of
the equipment until a whistle sounds to warn the wearer that there is low
cylinder pressure. The nominal duration is determined by the capacity
(volume and pressure rating) of the selected air cylinder – the actual
duration is also dependent on the rate at which the wearer uses air from
the cylinder (the breathing rate).
The features of the equipment are:
●The carrying system is a bandolier shoulder harness and waist belt with
a hip mounted cylinder holster.
●The pressure reducer (Fig 1, Item 2) connects directly on to the air
cylinder, and reduces the cylinder pressure to the medium pressure
required at the lung demand valve (Fig 1, Item 1). The reducer has a
cylinder pressure gauge, and a whistle that sounds to warn the wearer
that there is low cylinder pressure.
●The airline connector (Fig 1, Item 4) is a male quick coupling that is
used to connect an independent air supply for airline use.
●The medium-pressure hose has a quick coupling that allows rapid
removal and fitting of the lung demand valve when required.
●The Dräger PAS®ASV (automatic switch-over valve), air cylinder and
lung demand valve are described below.
Optional feature:
●Dropdown cylinder holster: assists wearer movement by making the
cylinder and holster more manoeuvrable (in a confined space for
example).
2.1.1 PAS®ASV
The Dräger PAS®ASV (Fig 1, Item 3) allows the PAS®Colt to be used
with an independent air supply such as a factory airline or other breathing
air supply (e.g. Dräger PAS®AirPack). The switch-over valve
automatically switches between the airline supply and the PAS®Colt air
cylinder, maintaining an uninterrupted air supply to the wearer during
switching.
Refer to Fig 2
AL (airline) Input from an independent air supply
BA (breathing apparatus) Input from the PAS®Colt air cylinder
LDV (lung demand valve) Output to the lung demand valve
During airline use, the PAS®Colt is connected to a breathing-quality
independent air supply, and the PAS®Colt cylinder valve is open. The
PAS®ASV uses the independent air supply as the primary supply of
breathing air to the wearer, and the PAS®Colt air cylinder as the
secondary (or back-up) supply.
●If the independent air supply pressure remains above a preset switching
pressure the wearer breathes from the independent air supply.
●If the independent air supply pressure falls below the preset switching
pressure, the PAS®ASV automatically switches to supply the wearer
from the air cylinder, and the PAS®ASV whistle (Fig 2, Item 1) sounds.
If the independent air supply pressure recovers (increases above the
preset switching pressure), the PAS®ASV switches back to the
independent air supply and the whistle stops.
●If the independent air supply is disconnected, or falls to a very low
pressure, the PAS®ASV whistle stops.
See the technical data (Section 8) for the PAS®ASV operating pressures.
2.1.2 Air cylinder
Cylinders are available with a 200 bar or 300 bar working pressure rating,
and in steel or composite materials. Only air cylinders listed in the Dräger
!
!
i
i
certification are approved for use with the PAS®Colt. Contact Dräger for
further information.
2.1.3 Lung demand valve (LDV)
A variety of Dräger lung demand valves are compatible with this
equipment, with the coupling (Fig 3, Item 1) selected to match the face
mask coupling (see table below):
During use, the lung demand valve activates automatically as the wearer
breathes, and then regulates the breathing air supply into the face mask in
response to the breathing rate of the wearer.
●On positive-pressure systems, when the lung demand valve is
activated, the internal valve remains open until closed by the user.
Positive-pressure valves have a reset button (Fig 3, Item 2) that closes
the valve when required. Pressing the reset button closes the internal
valve to switch off the air flow through the lung demand valve.
●On negative-pressure systems the internal valve closes automatically
to switch off the air flow through the lung demand valve.
The lung demand valve can also be activated manually by pressing the front
button (Fig 3, Item 3) to open the internal valve and activate air flow when
required. During airline use only, the front button can be pressed to deliver
additional air (supplementary air) into the mask when required by the wearer.
2.2 Intended use
When this product is used with an approved face mask, air cylinder, lung
demand valve and independent air supply, it provides the wearer with
respiratory protection for working in contaminated or oxygen-deficient
conditions. It is intended for use in applications where a high level of
respiratory protection is required. The equipment is intended to be used for
LDV
coupling
Face
mask
coupling
Type Coupling type
A P Positive
pressure
Push-in – Dräger specific
AE PE Positive
pressure
Screw-in – M45 x 3 to EN 148-3
N RA Negative
pressure
Screw-in – 40 mm round thread
to EN 148-1
combination airline/short duration applications, but it can also be used for
short duration only applications.
The air cylinder, face mask (full face mask conforming to EN 136 Class 2
or Class 3) and other accessories used with this product must be certified
Dräger components, assembled in an approved configuration; otherwise
the operation of the device may be impaired. Contact Dräger for further
information.
2.3 Limitations on use
This product is not approved for use in CBRN applications.
Use in potentially explosive atmospheres
●The PAS®Colt Series are type tested as suitable for use in potentially
explosive atmospheres. Electronic sub-assemblies are ATEX certified.
The combinations are suitable for use in hazardous areas up to and
including zone 0 and zone 20. The combinations can be used in
atmospheres containing gases of the gas explosion group IIC, with the
exception of combinations using the f 2 range of face masks, which are
only suitable to be used in atmospheres containing gases of the gas
explosion group IIB.
●Do not charge the cylinder in a potentially explosive atmosphere.
2.4 Approvals
The European standards, guidelines, and directives according to which
this product is approved are specified in the declaration of conformity (see
declaration of conformity or www.draeger.com/product-certificates).
3 Use
WARNING
The cylinder and airline air quality shall meet the requirements for
breathing air according to EN 12021. Do not use oxygen or
oxygen-enriched air. The moisture content of breathing air should
be controlled within the EN 12021 limits to avoid freezing the
apparatus.
Carry out a risk assessment of the workplace to ensure that it is
not possible to connect to any airline supply other than breathable
air (e.g. Nitrox).
Fill the PAS®Colt air cylinder to the full rated working pressure
prior to use, and do not commence any operation using a cylinder
that is less than 90 per cent full (or greater when national
regulations dictate).
Before using airline equipment, ensure that the independent air
supply meets the air quality requirements, and complies with the
airline pressure, flow and hose requirements in the technical data
(see Section 8), and has been issued with a permit for use if
necessary.
Position the source of the independent air supply in a safe and
uncontaminated area. Dräger recommend that a controller should
monitor and maintain the independent air supply throughout any
operation.
3.1 Preparation for use
3.1.1 Visual inspection
Carry out a visual inspection, checking the full breathing apparatus
including all component parts and accessories. Check that the equipment
is clean and undamaged, paying particular attention to pneumatic
components, hoses and connectors. Typical signs of damage that may
affect the operation of the breathing apparatus include impact, abrasion,
cutting, corrosion and discoloration. Report damage to service personnel
and do not use the apparatus until faults are rectified.
3.1.2 Fitting the cylinder
1. Ensure that the cylinder is fully charged.
2. Check the threads of the cylinder valve port and the pressure reducer.
Ensure that the O-ring seal (Fig 4, Item 1) and the sintered filter (Fig 4,
Item 2) in the reducer are clean and undamaged.
3. Fully insert the cylinder into the carrying holster. To prevent damage,
ensure that the cylinder remains clear of the hand wheel of the
pressure reducer while inserting the cylinder.
4. Align the cylinder with the pressure reducer and tighten the hand wheel
hand tight (Fig 5). Do not use tools or over tighten.
3.1.3 Functional testing
WARNING
If the breathing apparatus fails to meet any of the standards or
parameters described in the functional tests, or if an immediate
leak is evident, there is a system fault. Report the fault to trained
service personnel or contact Dräger. Do not use the breathing
apparatus until the fault condition is rectified.
1. Positive-pressure systems: press the reset button (Fig 3, Item 2) to
switch off the valve.
2. Open the PAS®Colt cylinder valve slowly, but fully, to pressurize the
system. The whistle on the PAS®Colt pressure reducer will briefly
sound.
3. Ensure that cylinder is fully charged and then close the cylinder valve.
4. Wait one minute and then observe the pressure gauge and reopen the
cylinder valve. The gauge must not show an increase in pressure of
more than 10 bar (one radial marking on the gauge face). Investigate
and repair any leak before use (see Section 4). If necessary, use a
soapy solution to locate the leak.
5. Close the cylinder valve and vent the system as follows:
○Positive-pressure systems: cover and seal the lung demand valve
outlet with the palm of the hand. Press the front
button (Fig 3, Item 3) to activate air flow and then lift the hand to
very slowly vent.
○Negative-pressure systems: carefully press the front
button (Fig 3, Item 3) to very slowly vent.
6. The whistle on the PAS®Colt pressure reducer must commence in the
range 110 to 100 bar.
7. Positive-pressure systems: press the reset button (Fig 3, Item 2) to
switch off the valve.
8. Open the PAS®Colt cylinder valve slowly, but fully, to pressurize the
system. The whistle will stop.
9. Connect the independent air supply to the male
coupling (Fig 1, Item 4). If the independent air supply has a shut-off
valve, open the valve. The PAS®ASV whistle will briefly sound to
!
!
1
3726/3731
12
3
4
BA
AL
LDV
1
2
3687
3
3717
12
3
12
4
3723
Dräger
5
3727
6
3625
1
2
3
7
3626
Dräger
8
3728
9
3729
Dräger
Dräger
10
3730
BA
AL
LDV
11
3713
Instructions for Use
i
Dow Corning®and Molykote®are registered trademarks of Dow Corning Corporation 3360044 (A3-D-P) Page 2 of 3
PAS®Colt Series
Airline/Short duration respiratory protection equipment with PAS®ASV (airline whistle)
indicate that the system has switched to use the independent air
supply.
WARNING
Do not direct the airflow on to the face, eyes or skin.
10. Press the front button (Fig 3, Item 3) to activate air flow from the valve
for 3–5 seconds. Unobstructed air will flow from the outlet of the lung
demand valve.
11. Positive-pressure systems: press the reset button (Fig 3, Item 2) to
switch off the valve.
12. Isolate and disconnect the independent air supply. There should be no
audible leak. If there is any leak, investigate and repair the leak before
use. If necessary, use a soapy solution to locate the leak.
13. Close the PAS®Colt cylinder valve.
14. Test the PAS®ASV whistle and switch-over operation:
○Positive-pressure systems: cover and seal the lung demand valve
outlet with the palm of the hand. Press the front
button (Fig 3, Item 3) to activate air flow and then lift the hand to
very slowly vent.
○Negative-pressure systems: carefully press the front
button (Fig 3, Item 3) to very slowly vent.
15. The PAS®ASV whistle begins to sound at the preset switching
pressure, indicating that the valve has switched from the independent
air supply (AL) to the PAS®Colt air cylinder (BA). The whistle on the
pressure reducer will also sound as the pressure falls. Both whistles
will cease when the system is vented.
16. Positive-pressure systems: press the reset button (Fig 3, Item 2) to
switch off the valve.
3.1.4 Putting on the PAS®Colt (ready position)
See also Fig 1 which shows the PAS®Colt worn in the ready position.
1. Open the waist belt buckle and fully extend the waist and shoulder
strap.
2. Place the left arm through the shoulder harness, taking the harness
over the head and on to the right shoulder, positioning the strap
diagonally across the body with the cylinder positioned against the left
hip.
3. Loop the waist belt around the waist and fasten the buckle – do not
tighten.
4. Grip the cylinder valve with the left hand and lift until the waist belt is in
line with the waist. Then tighten the waist belt strap until the equipment
is secure and comfortable on the waist. Pull down to adjust the shoulder
strap.
5. Check that the face mask port, and the lung demand valve coupling
and O-ring are clean and undamaged.
6. Connect the lung demand valve to the face mask as follows.
○Push-in coupling: press into the port of the face mask until it latches
in position. Check the attachment by gently attempting to pull the
coupling apart.
○Screw-in coupling: screw into the port of the face mask and tighten
hand tight. When the lung demand valve is fitted to the face mask, the
connector can swivel to allow for head and body movement of the
wearer.
7. Put the neck strap of the face mask over the head, and then insert the
neck strap stud into the hole in the centre strap of the head harness.
3.1.5 Putting on the face mask
WARNING
Correct fit of the face mask can only be achieved if the complete
mask seal makes contact with skin. Head hair, facial hair (including
beard stubble and sideburns), earrings, other facial piercings and
normal spectacles will interfere with the mask seal and are not
permitted in the sealing area. Additionally, head hair that could
affect the face mask fit (buns, pony-tails, hairpieces, etc.) is not
permitted.
NOTICE
Refer also to the face mask Instructions for Use.
1. Positive-pressure systems: press the reset button (Fig 3, Item 2) to
switch off the valve.
2. Open the cylinder valve slowly, but fully, to pressurize the system. The
whistle on the PAS®Colt pressure reducer will briefly sound.
3. Detach the neck strap stud from the centre strap of the head harness.
4. Spread the head harness (Fig 6). Place the chin into the chin cup of
the face mask and pull the harness over the head locating the harness
centre plate on back of the head.
5. Referring to Fig 7, tighten both lower (1) and then upper straps (2)
evenly towards the back of the head. If necessary, tighten the centre
strap (3).
6. Breathe normally and check that the head has a full range of
movement without pulling against the lung demand valve hose. If any
resistance to movement is felt, readjust the hose routing and then
recheck. If resistance is still felt, do not use the breathing apparatus
and contact Dräger.
7. Carry out the mask function check.
3.1.6 Mask function check
1. Close the cylinder valve and breathe normally to empty the system of
air. When empty, the face mask should hold on to the face to indicate
a positive seal.
2. Immediately reopen the cylinder valve and breathe normally.
3. Inhale and hold your breath – there should be no audible leak. If a leak
is detected, readjust the head harness and retest.
4. Recommence breathing – exhaled air should flow easily out of the
exhalation valve.
When the function check has been satisfactorily completed, breathe
normally and proceed to the work area.
3.2 During use
The nominal duration (10 or 15 minutes) applies to a PAS®Colt with a fully
charged cylinder, and refers to the usage time until the whistle on the
pressure reducer sounds to warn the wearer that there is low cylinder
pressure. The wearer should be in a safe area before the whistle on the
pressure reducer sounds.
!
!
i
i
3.2.1 Self-contained use (not supplied by an airline)
WARNING
The available duration starts from the time that the cylinder valve
is opened, and is dependent on the capacity of the cylinder and the
breathing rate of the wearer.
Dräger strongly advise that the front button (Fig 3, Item 3) is not
pressed to deliver supplementary air into the mask during self-
contained use. Using supplementary air would use air from the
cylinder and reduce the operating duration of the air supply.
●Regularly check the reading on the cylinder pressure gauge.
●If the whistle on the pressure reducer sounds, there is low cylinder
pressure. Immediately leave the hazardous area by the shortest and
safest route.
3.2.2 Airline use
WARNING
Dräger recommend that the user carries out a risk assessment to
establish local procedures to be followed in the event of failure of
the independent air supply.
If the PAS®ASV whistle sounds at any time during an operation,
the PAS®ASV has switched to supply the wearer from the PAS®
Colt air cylinder. If you continue to work while breathing from the
air cylinder, regularly check the cylinder pressure gauge – ensure
that the remaining air in the air cylinder allows sufficient time to
evacuate to a safe area.
During airline use, at very high work rates the pressure in the face
mask may become negative at peak inhalation flow.
NOTICE
The PAS®Colt cylinder valve must remain open during airline use.
●Connect the independent air supply to the male coupling (Fig 1,
Item 4), and if the independent air supply has a shut-off valve, open the
valve The PAS®ASV whistle will briefly sound to indicate that the
wearer is now breathing from the airline.
●If additional air is required, briefly press and release the front
button (Fig 3, Item 3) to deliver a single jet of supplementary air into
the mask.
●When the task is complete or when advised by a controller leave the
hazardous area:
○Disconnect from the independent air supply (see Section 3.2.1), or
○Remain connected and carefully withdraw the airline hose.
Evacuation procedure (independent air supply failure)
If the independent air supply fails, the wearer should proceed in line with
established local procedures (see warnings above).
●To evacuate proceed as follows:
a. Disconnect the independent air supply from the PAS®Colt.
b. Immediately leave the hazardous area by the shortest and safest
route.
When in a safe area, remove the lung demand valve from the face mask if
necessary and continue to breathe normally.
3.3 Dropdown cylinder holster (optional accessory)
●To release the dropdown cylinder holster:
a. Hold the cylinder valve with the left hand and, with the right hand,
press and hold the red button to open the locking
mechanism (Fig 8).
b. Lift the cylinder and holster clear of the locking mechanism and
then release the red button (Fig 9).
c. Lower the cylinder until it is supported by the harness
straps (Fig 10).The holster is held by two harness straps, one
fixed and one adjustable.
d. Hold the cylinder valve to move the cylinder and holster as required.
●To reconnect the dropdown cylinder holster, align and press the roller
on the holster into the locking mechanism.
3.4 After use
WARNING
Do not remove the equipment until in safe area, clear of hazard.
CAUTION
Do not drop or throw down equipment as damage could occur.
1. Loosen the face mask straps.
○Positive-pressure systems: as the seal between the mask and the
face is broken, press the reset button (Fig 3, Item 2) to switch off
the valve.
2. Remove the face mask and fully extend all of the straps of the head
harness.
3. If the independent air supply is still connected, isolate and disconnect
the supply.
4. Fully close the cylinder valve.
5. Press the front button (Fig 3, Item 3) to fully vent the system.
6. Remove the lung demand valve from the face mask (Fig 11 – push-in
coupling shown).
7. Open the waist belt buckle, lift the shoulder strap buckle to loosen the
harness and then remove the equipment.
8. Carry out the after use maintenance tasks in the maintenance
table (see Section 5.1).
4 Troubleshooting
The troubleshooting guide shows fault diagnosis and repair information
applicable to breathing apparatus users. Further troubleshooting and
repair information is available in Instructions for Use supplied with
associated equipment (e.g. face mask and air cylinder).
Contact service personnel or Dräger when the remedy information
indicates a service task, or if the symptom remains after all remedy actions
have been attempted.
!
!
i
i
!
!
5 Maintenance
5.1 Maintenance table
Service and test the breathing apparatus, including out-of-use apparatus,
in accordance with the maintenance table. Record all service details and
testing. Refer also to the Instructions for Use for the lung demand valve,
face mask and other associated equipment.
Additional inspection and testing may be required in the country of use to
ensure compliance with national regulations.
Notes
O Dräger recommendations
1. Clean the equipment if it is dirty. If it the equipment has been exposed
to contaminants, disinfect any components that come into direct and
prolonged contact with the skin.
2. These maintenance tasks may only be carried out by Dräger or trained
service personnel. Details of the tests are contained in the technical
manual which is issued to service personnel that have attended a
relevant Dräger maintenance course.
3. For type A check the O-ring on the lung demand valve; and for type
ESA check the outer surface of the male part of the push-in connector
on the lung demand valve. As a guide, lubricant should be felt on the
fingers but not seen. If relubrication is required, lightly apply Dow
Corning®Molykote®111 (other lubricants are not tested and may
damage the equipment).
5.2 Maintenance tasks
5.2.1 Removing the cylinder
WARNING
High-pressure air release may cause injury to the user or other
personnel near the breathing apparatus. Close the cylinder valve and
fully vent the system before attempting to disconnect the air cylinder.
1. Close the cylinder valve and press the front button (Fig 3, Item 3) to
fully vent the system.
2. Disconnect the cylinder valve from the pressure reducer.
3. Carefully remove the cylinder from the holster. To prevent damage,
ensure that the hand wheel of the pressure reducer remains clear of
the cylinder.
5.2.2 Air cylinder charging
WARNING
Air quality for compressed-air cylinders must conform to
requirements of EN 12021.
●Refer also to the instructions supplied with the cylinder and the
charging unit for recharging the cylinder.
Symptom Fault Remedy
High-pressure air
leak or failed leak
test
Loose or dirty cylinder
connector
Disconnect, clean and
reconnect the connector
and retest
Faulty hose or
component
Substitute user
replaceable accessories
and retest
Air leak from
medium-pressure
hose connection at
the pressure reducer
(safety relief valve)
Faulty O-ring, retainer,
spring or pressure
reducer
Service task
High or low medium
pressure
Pressure reducer fault Service task
Poor sounding
whistle
Whistle dirty Clean whistle flute and
retest
Whistle not
functioning correctly
Activation mechanism
fault
Service task
Component/
System
Task After
use
Every
month
Every
year
Every
six
years
Complete
equipment
Visual inspection
(see Note 1 and
Section 3.1.1)
OO
Functional test
(see Section 3.1.3)
OO
Flow and static
tests (see Note 2)
O
Lung demand
valve O-ring
Check and
lubricate if
necessary (see
Note 3)
O
Pressure
reducer
Medium-pressure
check (see Note 2)
O
Renew the high-
pressure connector
O-ring and sintered
filter (see Note 2)
O
Pressure
reducer and
PAS®ASV
Overhaul – Contact
Dräger for the
Repair & Exchange
(REX) service
O
Cylinder Charge to correct
pressure (see
Section 5.2.2)
O
Check the initial
test date stamped
on the cylinder
O
Cylinder pressure
test and
recertification
Carry out in line with national
regulations
Cylinder valve Basic overhaul During cylinder pressure test or
on condition
!
!
Instructions for Use
i
3360044 (A3-D-P) Page 3 of 3
PAS®Colt Series
Airline/Short duration respiratory protection equipment with PAS®ASV (airline whistle)
●Only charge compressed-air cylinders which:
○Conform to national standards.
○Feature the original manufacturer’s test date and test mark.
○Have not exceeded the test date indicated on the cylinder by the
last testing station.
○Are not damaged.
●To prevent ingress of moisture into the cylinder, ensure that the
cylinder valve remains closed until connected to the charging unit.
●Recharge to the rated working pressure of the cylinder. Dräger
recommend a charge rate of 27 bar/minute (rapid charging will increase the
temperature resulting in an incomplete charge).
●To prevent overcharging of the cylinder, Dräger recommend using a
pressure-limiting device on the charging compressor.
5.3 Cleaning and disinfecting
CAUTION
Do not exceed 60 °C for drying, and remove components from the
drying facility immediately when dry. Drying time in a heated dryer
must not exceed 30 minutes.
Do not immerse pneumatic or electronic components in cleaning
solutions or water.
If water is trapped and then freezes inside the pneumatic system
of the breathing apparatus (such as the lung demand valve),
operation will be impaired. Prevent any liquid from entering, and
thoroughly dry the breathing apparatus after cleaning to prevent
this from occurring.
For information about suitable cleaning and disinfecting agents
and their specifications refer to document 9100081 on
www.draeger.com/IFU.
Refer also to the Instructions for Use for the lung demand valve, face mask
and other associated equipment.
●Use only clean lint-free cloths
1. Clean the breathing apparatus manually using a cloth moistened with
cleaning solution to remove excess dirt.
2. Apply disinfecting solution to all internal and external surfaces.
3. Rinse all components thoroughly with clean water to remove all
cleaning and disinfecting agents.
4. Dry all components using a dry cloth, in a heated dryer or in air.
5. Contact service personnel or Dräger if disassembly of pneumatic or
electronic components is required.
6 Storage
Store the equipment between -15 °C and +25 °C. Ensure that the
environment is dry, free from dust and dirt, and does not subject the
equipment to wear or damage due to abrasion. Do not store the equipment
in direct sunlight. Note also the following:
●Extend the shoulder strap, the waist belt and the head harness straps
of the face mask.
●For hygienic storage, place the face mask in a protective bag (contact
Dräger for supply of a suitable bag).
●Route rubber hoses in such a way that the bend radius is not too acute
and the hose is not stretched, compressed or twisted.
●Fix the apparatus securely to any raised mounting point to prevent it
from falling.
7 Disposal
When required, dispose of the PAS®Colt in accordance with national or
local regulations for waste disposal.
8 Technical data
●High-pressure connector: 200 bar or 300 bar, standard G5/8” as per
EN 144-2.
●Independent air supply (single user):
○Airline pressure and flow requirements: pressure 6 to 10 bar, air
flow rate at least 550 litres/minute. Important note: do not
exceed 10 bar.
○Airline hose requirements: maximum of 100 m of Dräger
approved hose. Important note: no more than four individual
hoses (a maximum of five hose connections) are permitted in the
airline.
●PAS®ASV: whistle activates and switch-over occurs in the range:
3.5 to 5.5 bar (airline pressure).
●Pressure reducer: whistle commences in the range: 110 to 100 bar
(cylinder pressure).
●Whistles cease in the range: 1.75 to 0 bar.
●Whistle volume: >90 dBA.
!

Other Dräger Medical Equipment manuals

Dräger MIcroVent User manual

Dräger

Dräger MIcroVent User manual

Dräger ESS II 3L User manual

Dräger

Dräger ESS II 3L User manual

Dräger Oxy K 30 HW User manual

Dräger

Dräger Oxy K 30 HW User manual

Dräger HP NO Series User manual

Dräger

Dräger HP NO Series User manual

Dräger Zeus Infinity Empowered User manual

Dräger

Dräger Zeus Infinity Empowered User manual

Dräger Oxylog 3000 plus User manual

Dräger

Dräger Oxylog 3000 plus User manual

Dräger PSS 5000 S Series User manual

Dräger

Dräger PSS 5000 S Series User manual

Dräger EVITA XL User manual

Dräger

Dräger EVITA XL User manual

Dräger Oxy 3000 User manual

Dräger

Dräger Oxy 3000 User manual

Dräger Prana W User manual

Dräger

Dräger Prana W User manual

Dräger Fabius GS SW 3.n User manual

Dräger

Dräger Fabius GS SW 3.n User manual

Dräger Primus SW 4.5n User manual

Dräger

Dräger Primus SW 4.5n User manual

Dräger WBR82 User manual

Dräger

Dräger WBR82 User manual

Dräger Draeger-Tube Mercaptan 20/a User manual

Dräger

Dräger Draeger-Tube Mercaptan 20/a User manual

Dräger Panorama Nova P User manual

Dräger

Dräger Panorama Nova P User manual

Dräger Oxylog 2000 plus User manual

Dräger

Dräger Oxylog 2000 plus User manual

Dräger Babylog VN800 User manual

Dräger

Dräger Babylog VN800 User manual

Dräger X-am 7000 User manual

Dräger

Dräger X-am 7000 User manual

Dräger PulmoVista 500 User manual

Dräger

Dräger PulmoVista 500 User manual

Dräger Fabius Tiro User manual

Dräger

Dräger Fabius Tiro User manual

Dräger Movita lift User manual

Dräger

Dräger Movita lift User manual

Dräger Isolette 8000 plus User manual

Dräger

Dräger Isolette 8000 plus User manual

Dräger CVP 5220 User manual

Dräger

Dräger CVP 5220 User manual

Dräger Babylog 1 HF User manual

Dräger

Dräger Babylog 1 HF User manual

Popular Medical Equipment manuals by other brands

ResMed Mirage Vista user guide

ResMed

ResMed Mirage Vista user guide

Otto Bock 1E56 Axtion Instructions for use

Otto Bock

Otto Bock 1E56 Axtion Instructions for use

Care of Sweden CuroCell A4 Cirrus Instructions for use

Care of Sweden

Care of Sweden CuroCell A4 Cirrus Instructions for use

Otto Bock DynamicArm Plus Instructions for use

Otto Bock

Otto Bock DynamicArm Plus Instructions for use

VOLTCRAFT BS-22 operating instructions

VOLTCRAFT

VOLTCRAFT BS-22 operating instructions

Aqua Creek Tidal Wave manual

Aqua Creek

Aqua Creek Tidal Wave manual

Hausted UNICARE Series Operating and service manual

Hausted

Hausted UNICARE Series Operating and service manual

medi protect.Lumbafix Instructions for use

medi

medi protect.Lumbafix Instructions for use

Sirona galileos Maintenance manual

Sirona

Sirona galileos Maintenance manual

Kyoto Kagaku MW64 instruction manual

Kyoto Kagaku

Kyoto Kagaku MW64 instruction manual

SPORLASTIC GENU-HiT SUPREME Fitting instructions

SPORLASTIC

SPORLASTIC GENU-HiT SUPREME Fitting instructions

OrthoPrax LSO 637 manual

OrthoPrax

OrthoPrax LSO 637 manual

MEDIAID 100 user manual

MEDIAID

MEDIAID 100 user manual

Monaghan Aerobika Instructions for use

Monaghan

Monaghan Aerobika Instructions for use

dynarex 10462 quick start guide

dynarex

dynarex 10462 quick start guide

orfit DYNASYST ORFITUBE Instructions for use

orfit

orfit DYNASYST ORFITUBE Instructions for use

Game ready Med4 Elite quick start guide

Game ready

Game ready Med4 Elite quick start guide

TytoCare TytoPro quick start guide

TytoCare

TytoCare TytoPro quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.