1.2 Specifications - General
Enclosure: Polycarbonate. Type 4X, IP65.
Note: To ensure a water-tight seal, tighten all four front panel
screws to 6 in-lbs of torque
Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10 x 5.15
in.). utout: 1/2 DIN 139mm x 139mm (5.45 x 5.45 in.)
Conduit Openings: Accepts 1/2” or PG13.5 conduit fittings
Display: Monochromatic graphic liquid crystal display.
128 x 96 pixel display resolution. Backlit. Active display
area: 58 x 78mm (2.3 x 3.0 in.).
Ambient Temperature and umidity: 0 to 55°
(32 to 131°F). Turbidity only: 0 to 50° (32 to 122°F),
RH 5 to 95% (non-condensing)
Storage Temperature Effect: -20 to 60 º (-4 to 140 °F)
Power: ode -01: 115/230 VA ±15%, 50/60 Hz. 10W.
ode -02: 20 to 30 VD . 15 W.
ode -03: 85 to 265 VA , 47.5 to 65.0 Hz,
switching. 15 W.
Note: ode -02 and -03 power supplies include four
programmable relays
Equipment protected by double insulation
Alarms relays*: Four alarm relays for process measure-
ment(s) or temperature. Any relay can be configured as a
fault alarm instead of a process alarm. Each relay can be
configured independently and each can be programmed with
interval timer settings.
Relays: Form , SPDT, epoxy sealed
Inductive load: 1/8 HP motor (max.), 120/240 VA
*Relays only available with 02 power supply (20 - 30 VD ) or 03
switching power supply (85 - 265 VA )
Inputs: One or two isolated sensor inputs
Outputs: Two 4-20 mA or 0-20 mA isolated current
outputs. Fully scalable. Max Load: 550 Ohm.
Output 1 has superimposed HART signal
(1056-0X-2X-3X-HT only)
Current Output Accuracy: ±0.05 mA @ 25 º
Terminal Connections Rating:
Power connector (3-leads): 24-12 AWG wire size.
Signal board terminal blocks: 26-16 AWG wire size.
urrent output connectors (2-leads): 24-16 AWG wire
size.
Alarm relay terminal blocks: 24-12 AWG wire size
(-02 24 VD power supply and -03 85-265VA power
supply)
Weight/Shipping Weight:
(rounded up to nearest lb or nearest 0.5 kg):
3 lbs/4 lbs (1.5 kg/2.0 kg)
RFI/EMI: EN 61326-1
LVD: EN 61010-1
azardous Location Approvals -
Options for SA: 01, 02, 03, 20, 21, 22, 24, 25, 26, 27, 30,
31, 32, 34, 35, 36, 37, 38, AN, DP and HT.
lass I, Division 2, Groups A, B, , & D
lass Il, Division 2, Groups E, F, & G
lass Ill T4A Tamb= 50 °
Type 4X, IP66 Enclosure
Non-Incendive Field Wiring (NIFW) may be used when
installed per drawing 1400325. The ‘ ’ and ‘US’ indicators
adjacent to the SA Mark signify that the product has been
evaluated to the applicable SA and ANSI/UL Standards, for
use in anada and the U.S. respectively.
Evaluated to SA Standard 22.2 No. 0-10, 0.4-04, 25-1996,
94-M1991, 142-M1987, 213-M1987, 60529-2005/2015.
ANSI/IE 60529-2004/2011. ANSI/ISA 12.12.01:2007.
UL No. 50, 11th Ed., 508 17th Ed.
lass I, Division 2, Groups A, B, , & D
lass Il & lll, Division 2, Groups E, F, & G
T4A -20° ≤ Tamb ≤ +50°
Enclosure Type 4X
CAUTION
RISK OF ELECTRICAL S OCK
Maximum Relay Current
Resistive
28 VD 5.0 A
115 VA 5.0 A
230 VA 5.0 A
POLLUTION DEGREE 2: Normally only non-conductive
pollution occurs. Occasionally, however, a temporary con-
ductivity caused by condensation must be expected.
Altitude: for use up to 2000 meter (6562 ft.)
WARNING
Exposure to some chemicals may degrade the sealing
properties used in the following devices: Zettler Relays
(K1-K4) PN AZ8-1 H-12DSEA
WARNING
ptions for FM: 01, 02, 03, 20, 21, 22, 24, 25, 26, 30, 31, 32, 34,
35, 36, 38, AN, and HT. 01 power supply not available with
turbidity sensor options.
Instruction Manual
LIQ-MAN-1056
Section 1.0: Description and Specification
April 2017
Non-Incendive Field Wiring (NIFW) may be used when installed
per drawing 1400324. Evaluated to FM standards 3600:1998,
3611:2004, 3810:2005,ANSI/NEMA 250:2003.