Technical Specifications
ENDA EPA741 PROGRAMMABLE AC/DC AMMETER
Supply Voltage Order Code
Output Internal Shunt Resistor External Shunt Resistor
SÝSEL MÜHENDÝSLÝK ELEKTRONÝK SAN. VE TÝC. A.Þ.
Yukarý Dudullu Barbaros Cad. Kutup Sok. No:20 34775 - ÜMRANÝYE/ÝSTANBUL/TÜRKÝYE
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr EPA741-E-03
1/3
ENDA
PROGRAMMABLE AC/DC AMMETER
EPA 741
SET
OUT AC DC ACDC
Read this document carefully before using this device. The guarantee will be expired by
damaging of the device if you don't attend to the directions in the user manual. Also we don't
accept any compensations for personal injury, material damage or capital disadvantages.
* 72 x 72mm sized.
* 4 digits display.
* Easy to use by front panel keypad.
* With current trafo or shunt can be used.
* Programmable scale between 5A and 9999A.
* Multifunctional alarm output (NO+NC) for upper and lower
limits.
* Having CE mark according to Europan Norms.
* Measuring type can be selected AC, DC or True RMS
Thank you for choosing ENDA EPA741 programmable AC/DC ammeter.
ENVIRONMENTAL CONDITIONS
Height Max. 2000m
%80 up to 31°C decreasing linearly %50 at 40°C.
Ambient/stroge temperature
Max. Relative humidity
0 ... +50°C/-25 ... 70°C
Rated pollution degree According to EN 60529 Front panel : IP60 , R
are panel : IP20
Do not use the device in locations subject to corrosive and flammable gases.
Supply 230V AC +%10 -%20, 50/60Hz or 24V AC %10 , 50/60Hz or 9-24V AC/DC %10± ±
Power consumption
2.5mm² screw-terminal connections
Wiring
Scale
Sensitivity
Accuracy
EMC
Safety requirements
EN 61326-1: 1997, A1: 1998, A2: 2001 (Performance criterion B for the EMC standard)
EN 61010-1: 2001 (Pollution degree 2, overvoltage category II)
ELECTRICAL CHARACTERISTICS
0A...9999A (Specified )c.tr.r c.tr.rparameter. For example:scale is 0A...5A for =5.00
0.002A x ( For example , 0.01A for =5.00 )c.tr.r c.tr.r
Max. 7VA
AC
RMS
DC ±%1
%1
(full scale) (For square wave form %2)±
(full scale)
(full scale) (For square wave form %2)±
Input Impedance EPA741Sxx 10MÙ
EPA141xx
Frequency Range DC , 10Hz - 200Hz (For square wave form 10Hz-70Hz)
Input Range EPA741Sxx -5A...5A (Device damages 10A peak and more current.)
EPA741xx -60mV...60mV (Device damages 50V peak and more voltage.)
AC and RMS
DC -999A...9999A (Specified parameter. For example:scale is -5A...5A for =5.00)c.tr.r c.tr.r
±
±%1
OUTPUTS
Alarm output Relay: 250V AC, 8A (for resistive load), NO+NC
Mechanical 30.000.000 ; Electrical 100.000 operation.
Life expectancy for relay
HOUSING
Housing type Suitable for flush-panel mounting. (According to DIN 43 700)
Dimensions W72xH72xD97mm
Weight
Enclosure material Self extinguishing plastics.
EPA741 Approx. 350g (after packing)
EPA741-24 Approx. 350g (after packing)
While cleaning the device, solvents (thinner, benzine, acid etc.) or corrosive materials must not be used.
40kÙ
230V AC +%10 -%20
24V AC %10±
Relay
Relay
-
-
EPA741-R-230VAC
EPA741-230VAC
EPA741-R-24VAC
EPA741-24VAC
EPA741S-R-230VAC
EPA741S-230VAC
EPA741S-R-24VAC
EPA741S-24VAC
9-24V AC/DC %10± Relay
-EPA741-R-SM
EPA741-SM
EPA741S-R-SM
EPA741S-SM
UP.l1
UP.l1 HYS.U+
UP.l1 HYS.U-
LO.l1
LO.l1 HYS.U+
LO.l1 HYS.U-
Out
Out
dly.U
A.out n.c.=
A.out n.o.=
dly.U dly.ldly.l
Input
t
t
t
2/3 EPA741-E-03
A
-A
0
A
0
A
0
A
-A
A
0
A
0
A
-A
0
T3T/2
T/2 2T
T3T/2
T/2 2T
T3T/2
T/2 2T
T3T/2
T/2
0
T
3T/2
T/2
2T
T3T/2
T/2 2T
dd
T2T
0.308 A
0.386 A
0.000
A
0.000
A
d
T
d
T
A
A
d
T
Ad
T2
2
0.000
1
2
A1
2
A
2
A1
2
Ap
1
2
A
1
Ap
2
A1
2
A1
2
A1
3
A1
3
A1
Dimensions
Note :
1)
2)
While panel mounting, additional distance
required for connection cables should be
considered.
Panel thickness should be maximum 10mm.
3) If there is no 90mm free space at the back side
of the device, it would be difficult to remove it from
the panel.
For removing mounting clamps:
- Push out the flush-mounting clamp
In direction 1 shown in figure below.
- Pull out the clamp in direction 2
Panel cut-out
Connection
cables
Depth
97mm
75mm
84mm
68 mm
+0.7
68 mm
+0.7
Rubber
packing
Flush mounting
clamp
Panel
1
2
Equipment is protected throughout by DOUBLE
INSULATION
Holding screw 0.4-0.5Nm.
Connection Diagram
ENDA EPA741 is intended for installation in control panels.Make sure that the device is
used only for intended purpose.The electrical connections must be carried on by a
qualified staff and must be according to the relevant locally applicable regulations.
During an installation, all of the cables that are connected to the device must be free of
energy. The device must be protected against inadmissible humidity, vibrations, severe
soiling and make sure that the operation temperature is not exceeded. The cables
should not be close to the power cables or components.
SUPPLY:
NOTE :
14
15
184-253V AC
50/60Hz 7VA
Line
Neutral
230V AC
Supply
Switch
Cable size: 1,5mm²
Fuse
F 100mA
250V AC
Fuse should be connected.
1) Mains supply cords shall meet the requirements of IEC 60227 or IEC 60245.
2) In accordance with the safety regulations, the power supply switch shall bring the identification of
the relevant instrument and it should be easily accessible by the operator.
ac Ac.dc
dc (rms)
72mm
78mm
ENDA
PROGRAMMABLE AC/DC AMPERMETER
EPA 741
SET
OUT AC DC ACDC
11
12
13
14
15
19
230V AC +10% -20%
50/60Hz 7VA
20
OUTPUT
250V AC 8A
RESISTIVE
LOAD
max.
5A peak
AC / DC
CAT II
+
SN: XXXXXXXXX
ENDA
INDUSTRIAL ELECTRONICS
EPA741S-R-230VAC
PROGRAMMABLE AC/DC AMMETER
14
15
19
230V AC +10% -20%
50/60Hz 7VA
20
Max.
5A peak
AC / DC
CAT II
+
SN: XXXXXXXXX
ENDA
INDUSTRIAL ELECTRONICS
EPA741S-230VAC
PROGRAMMABLE AC/DC AMMETER
11
12
13
14
15
19
24V AC +10%
50/60Hz 7VA
20
OUTPUT
250V AC 8A
RESISTIVE
LOAD
Max.
5A peak
AC / DC
CAT II
+
SN: XXXXXXXXX
ENDA
INDUSTRIAL ELECTRONICS
EPA741S-R-24VAC
PROGRAMMABLE AC/DC AMMETER
14
15
19
20
Max.
5A peak
AC / DC
CAT II
+
SN: XXXXXXXXX
ENDA
INDUSTRIAL ELECTRONICS
EPA741S-24VAC
PROGRAMMABLE AC/DC AMMETER
24V AC +10%
50/60Hz 7VA
SN: XXXXXXXXX
14
15
19
20
230V AC +10% -20%
50/60Hz 7VA
Max.
60mV peak
AC / DC
CAT II
+
ENDA
INDUSTRIAL ELECTRONICS
EPA741-230VAC
PROGRAMMABLE AC/DC AMMETER
SN: XXXXXXXXX
14
15
19
20
Max.
60mV peak
AC / DC
CAT II
+
ENDA
INDUSTRIAL ELECTRONICS
EPA741-24VAC
PROGRAMMABLE AC/DC AMMETER
24V AC +10%
50/60Hz 7VA
11
12
13
14
15
19
230V AC +10% -20%
50/60Hz 7VA
20
OUTPUT
250V AC 8A
RESISTIVE
LOAD
max.
60mV peak
AC / DC
CAT II
+
SN: XXXXXXXXX
ENDA
INDUSTRIAL ELECTRONICS
EPA741-R-230VAC
PROGRAMMABLE AC/DC AMMETER
11
12
13
14
15
19
24V AC +10%
50/60Hz 7VA
20
OUTPUT
250V AC 8A
RESISTIVE
LOAD
Max.
60mV peak
AC / DC
CAT II
+
SN: XXXXXXXXX
ENDA
INDUSTRIAL ELECTRONICS
EPA741-R-24VAC
PROGRAMMABLE AC/DC AMMETER
PROGRAMMING MODE
SET
&
RUNNING MODE
Upper limit value can be
adjusted between lower scale
and upper scale that specified
with c.t.r.r parameter.This
parameter can’t be lower than
( )
value.
lo.l1 + HYS.l + HYS.U
Lower limit value can be
adjusted between lower scale
and upper scale that specified
with c.t.r.r parameter.This
parameter can’t be upper than
( )
value.
UP.l1 - HYS.U - HYS.L
Hysteresis value for upper
limit can be adjusted
between 0and /20.
This parameter can’t be
higher than
( - - )
value.
c.tr.r
Up.l1 Lo.l1 Hys.l
Hysteresis value for lower
limit can be adjusted
between 0and /20.
This parameter can’t be
higher than
( - - )
value.
c.tr.r
Up.l1 Lo.l1 Hys.U
Delay time for upper
limit alarm can be
adjusted between 0
and 900 seconds.
Delay time for lower
limit alarm can be
adjusted between 0
and 900 seconds.
Alarm output position can
be adjusted or If
at
alarm condition.
n.o. N.c..
n.o. is adjusted ,
output relay is powered
Measuring type can be adjusted , orac dc
acdc..Three leds shows measuring type at
the top of the display.
Current transforming rate can be adjusted
between 5.00(/5) and (/5).If this
parameter changes upper limit value sets to
upper scale ,lower limit value sets to lower
scale and hysteresis values set to 0.
9999
3/3 EPA741-E-03
ENDA
PROGRAMMABLE AC/DC AMPERMETER
EPA 741
SET
OUT AC DC ACDC
If current transforming rate is lower than 10, measuring value can be displayed like ( ) or ( ).
If current transforming rate is between 10 and 100 ,measuring value can be displayed like ( ) or
().
If current transforming rate is higher than 100, measuring value can be displayed like only ( ).
0.00 0.0
0.0
0
0
you want by using key again.
SET
To change the designation of decimal number press key while in c.tr.r menu then select the one
SET
&
Adjusting Decimal Number
Means, measured current value is
lower than down scale.
ERROR MESSAGES
Means, measured current value is
higher than up scale.
To adjust the selected parameter hold down key then you can adjust by using and keys.
If you hold down the keys for 0.6 seconds while changing the parameters you skip to fast change mode. If you hold down the
keys more than 5 seconds the active parameter would change faster.
If you wait 25 seconds before holding down the keys or power off and power up again, you skip to operation mode.
Parameter Adjustment Method
SET
In Running mode, you can enter or lo.l1 menus by using key.Then you can adjust the
value you want by following the parameter adjustment method.
Up.l1
To skip the programme mode while in operation mode press and keys at the same time.For
returning back to operation mode you should follow the same way while you are in upper menu.
SET
(*) There is only EPA741 , EPA741-24 , ) that has no relay.1nPt menu at the models( EPA741S , EPA741S-24
*

Other ENDA Measuring Instrument manuals

ENDA EPV741 User manual

ENDA

ENDA EPV741 User manual

ENDA EDP 741C User manual

ENDA

ENDA EDP 741C User manual

ENDA EDP2041 User manual

ENDA

ENDA EDP2041 User manual

ENDA EI7412 User manual

ENDA

ENDA EI7412 User manual

ENDA EPV241A User manual

ENDA

ENDA EPV241A User manual

ENDA EPV742 User manual

ENDA

ENDA EPV742 User manual

ENDA EDP 141 User manual

ENDA

ENDA EDP 141 User manual

ENDA ETS1410 User manual

ENDA

ENDA ETS1410 User manual

ENDA EPA742 User manual

ENDA

ENDA EPA742 User manual

ENDA ETI741 User manual

ENDA

ENDA ETI741 User manual

ENDA EPV241A User manual

ENDA

ENDA EPV241A User manual

ENDA ETS762 User manual

ENDA

ENDA ETS762 User manual

ENDA EDP2041 User manual

ENDA

ENDA EDP2041 User manual

ENDA EDP 141 User manual

ENDA

ENDA EDP 141 User manual

ENDA EI7412 User manual

ENDA

ENDA EI7412 User manual

ENDA EPA141 User manual

ENDA

ENDA EPA141 User manual

ENDA EDP2041 User manual

ENDA

ENDA EDP2041 User manual

ENDA EPV141 User manual

ENDA

ENDA EPV141 User manual

ENDA ETS762 User manual

ENDA

ENDA ETS762 User manual

ENDA ETS1410 User manual

ENDA

ENDA ETS1410 User manual

ENDA EPA942 User manual

ENDA

ENDA EPA942 User manual

ENDA EI1410 NTC User manual

ENDA

ENDA EI1410 NTC User manual

ENDA EPA242 User manual

ENDA

ENDA EPA242 User manual

Popular Measuring Instrument manuals by other brands

McIntosh MPM4000 owner's manual

McIntosh

McIntosh MPM4000 owner's manual

Draminski HMM manual

Draminski

Draminski HMM manual

Electro Industries/GaugeTech 3DAA5 quick start guide

Electro Industries/GaugeTech

Electro Industries/GaugeTech 3DAA5 quick start guide

Dataq DI-700 user manual

Dataq

Dataq DI-700 user manual

Tektronix KEITHLEY 4200A-SCS-PKC quick start guide

Tektronix

Tektronix KEITHLEY 4200A-SCS-PKC quick start guide

Elma Instruments FieldSense 2.0 manual

Elma Instruments

Elma Instruments FieldSense 2.0 manual

getemed CardioMem CM 100 XT user manual

getemed

getemed CardioMem CM 100 XT user manual

Endress+Hauser Liquiphant FTL64 operating instructions

Endress+Hauser

Endress+Hauser Liquiphant FTL64 operating instructions

Oval ULTRA UF-II instructions

Oval

Oval ULTRA UF-II instructions

PCE Instruments PCE-XXM 30 user manual

PCE Instruments

PCE Instruments PCE-XXM 30 user manual

sauermann E INSTRUMENTS E6000 operating & maintenance manual

sauermann

sauermann E INSTRUMENTS E6000 operating & maintenance manual

GHM DeltaOHM LPUVA02 operating manual

GHM

GHM DeltaOHM LPUVA02 operating manual

biochrom WPA Lightwave S2000 instruction manual

biochrom

biochrom WPA Lightwave S2000 instruction manual

Dostmann Electronic P700-EX Operation manual

Dostmann Electronic

Dostmann Electronic P700-EX Operation manual

Topcon AP-L1A instruction manual

Topcon

Topcon AP-L1A instruction manual

TESTO 330 instruction manual

TESTO

TESTO 330 instruction manual

Larson Davis SoundTrack LxT user manual

Larson Davis

Larson Davis SoundTrack LxT user manual

Inficon BPG400 instruction sheet

Inficon

Inficon BPG400 instruction sheet

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.