ENDA EPV241A User manual

OUTPUTS
Alarm output Relay: 250V AC, 8A (for resistive load), NO+NC
Mechanical 30.000.000 ; Electrical 100.000 operation.
Life expectancy for relay
0 ... +50°C/-25 ... 70°C
HOUSING
Housing type Suitable for flush-panel mounting. (According to DIN 43 700)
Dimensions W77xH35xD71mm
Weight
Enclosure material Self extinguishing plastics.
ENDA EPV241A AC/DC VOLTMETER
While cleaning the device, solvents (thinner, benzine, acid etc.) or corrosive materials must not be used.
Approx. 350g (after packing)
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Yukarı Dudullu Barbaros Cad. Kutup Sok. No:20 34775 - ÜMRANİYE/İSTANBUL/TÜRKİYE
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
* 77 x 35mm sized.
* 3 digits display.
*
* Easy to configure with front panel keypad.
* Multifunctional alarm output (NO+NC) for upper and lower
limits.
* Communication feature over isolated RS485, using ModBus
RTU protocol. (Functional).
* Measuring type can be selected as AC, DC or true RMS.
* CE marked according to Europan Norms.
Values between -100V and 100 V can be indicated with one
decimal point.
* For maximum 50V AC/DC measurements, measurements
between -50V and +50V can be shown with two decimal digits
by having 10 times more accurate measurement input.
Supply 230V AC +10% -20%, 50/60Hz or 24V AC 10% , 50/60Hz or optional 9-30V DC / 7-24V AC 10% SMPS± ±
Power consumption
2.5mm² screw-terminal connections
Wiring
Scale
Sensitivity
ELECTRICAL CHARACTERISTICS
If C.InP; 500 is selected, it is 0V...500V or If C.ınP ;50 is selected, it is 0....50V
If C.InP ;500 is selected, it is -500V DC....500V DC or If C.InP;50 is selected, it is -50V DC....50V DC
0,01V (If C.InP;50 is selected )
Max. 5VA
AC and RMS:
DC :
EPV241A-E-02-R-201307
1/5
±
Accuracy
EMC
Safety requirements
EN 61326-1: 2006
EN 61010-1: 2010 (Pollution degree 2, overvoltage category II)
1%
AC
RMS
DC ±
1%
1%
(Full scale) (For square wave form 2%)±
(Full scale)
Input Impedance 870kΩ
Frequency Range
Input Range -500V...500V (If CINP 500 is selected, device breaks down at more than ±1250 DC voltages.)
-50V....50V (If CInp 50 is selected, device breaks down at more than ±125 DC voltages.)
± (Full scale) (For square wave form 2%)
±
DC , 10Hz - 200Hz (For square wave form 10Hz-70Hz)
Thank you for choosing ENDA EPV241A AC/DC voltmeter.
Read this document carefully before using this device. The guarantee will be expired by
device damages if you don't attend to the directions in the user manual. Also we don't
accept any compensations for personal injury, material damage or capital disadvantages.
Technical Specifications
.Do not use the device in locations subject to corrosive and flammable gases
ENVIRONMENTAL CONDITIONS
Ambient/stroge temperature
Max. Relative humidity
Rated pollution degree
Height
80% Relative humidity for temperatures up to 31 % °C,decreasing linearly to 50% at 40°C.
According to EN 60529 Front panel : IP65 , Rear panel : IP20
Max. 2000m
0,1V (If CINP;500 is selected and higher than -100V or lower than 100V for input values)
1V (If CINP;500 is selected and lower than -100V or higher than 100V for input values)
1-Output
R......Relay
None...No relay
2-Supply Voltage
230VAC...230V AC RSI..... Insulated ModBus (optional)
110VAC...110V AC
24VAC.....24V AC
SM...........9-30V DC / 7-24V AC
3-ModBus
SETSET
EPV 241A
ENDA
OUT AC DC ACDC
V
Order Code : EPV241A- - -
23
1

UP.l1
UP.l1 HYS.U +
UP.l1 HYS.U -
LO.l1
LO.l1 HYS.L +
LO.l1 HYS.L -
Out
Out
dly.U
A.out n.c. =
A.out n.o. =
dly.U dly.ldly.l
Input
t
t
t
Dimensions
71,5mm
29,5mm
2/5
A
-A
0
A
0
A
0
A
-A
A
0
A
0
A
-A
0
T3T/2
T/2 2T
T3T/2
T/2 2T
T3T/2
T/2 2T
T3T/2
T/2
0
T
3T/2
T/2
2T
T3T/2
T/2 2T
dd
T2T
0.308 A
0.386 A
0.000
A
0.000
A
d
Td
T
A
A
d
T
Ad
T2
2
0.000
1
2
A1
2
A
2
A1
2
Ap
1
2
A
1
Ap
2
A1
2
A12
A1
3
A1
3
A1
ac Ac.dc
dc (rms)
Panel cut-out
Connection Diagram
4
5
SUPPLY:
NOTE :
184-253V AC
50/60Hz 5VA
Line
Neutral
Note:
Cable Size: 1,5mm²;
230V AC
Supply
Switch
Fuse
F 100mA
250V AC
Fuse should be connected.
1) Mains supply cords shall meet the requirements of
IEC 60227 or IEC 60245.
2) In accordance with the safety regulations, the power
supply switch shall bring the identification of the relevant
instrument and it should be easily accessible by the operator.
4
5
Faz
Nötr
Equipment is protected throughout by
DOUBLE INSULATION
Holding screw 0.4-0.5Nm.
Panel
Depth
2
61mm 5mm
Rubber
packing
Flush mounting
clamp
1
ENDA EPV241A is intended for installation in control panels. Make sure that the device is used
only for intended purpose. The electrical connections must be carried out by a qualified staff
and must be according to the relevant locally applicable regulations. During an installation, all of the
cables that are connected to the device must be free of electrical power. The device must be protected
against inadmissible humidity, vibrations, severe soiling. Make sure that the operation temperature is
not exceeded. The cables should not be close to the power cables or components.
35mm
77mm
SETSET
EPV 241A
ENDA
OUT AC DC ACDC
V
CAT II
+
1 2 3 4 5 6 7 8 9 10
ENDA INDUSTRIAL ELECTRONICS
SN: XXXXXXXXX
EPV241A-R-230VAC-RSI
PROGRAMMABLE
AC/DC VOLTMETER
11 12
OUTPUT
250V AC 8A
RESISTIVE LOAD
COM.
B
A
8 680407 704934
230V AC +10% -20%
50/60Hz 5VA
Made in Turkey
+ Max. 500V
AC / DC
Max. 50V
AC / DC
CAT II
+
1 2 3 4 5 6 7 8 9 10
ENDA INDUSTRIAL ELECTRONICS
SN: XXXXXXXXX
EPV241A-230VAC
PROGRAMMABLE
AC/DC VOLTMETER
11 12
8 680407 704903
230V AC +10% -20%
50/60Hz 5VA
Made in Turkey
+ Max. 500V
AC / DC
Max. 50V
AC / DC
CAT II
+
1 2 3 4 5 6 78910
ENDA INDUSTRIAL ELECTRONICS
SN: XXXXXXXXX
EPV241A-230VAC-RSI
PROGRAMMABLE
AC/DC VOLTMETER
11 12
COM.
B
A
8 680407 704910
230V AC +10% -20%
50/60Hz 5VA
Made in Turkey
+ Max. 500V
AC / DC
Max. 50V
AC / DC
CAT II
+
1 2 3 4 5 6 7 8 9 10
ENDA INDUSTRIAL ELECTRONICS
SN: XXXXXXXXX
EPV241A-R-230VAC
PROGRAMMABLE
AC/DC VOLTMETER
11 12
OUTPUT
250V AC 8A
RESISTIVE LOAD
8 680407 704927
230V AC +10% -20%
50/60Hz 5VA
Made in Turkey
+ Max. 500V
AC / DC
Max. 50V
AC / DC
If C.Inp input type “500” is selected, the measurement terminals 9 and 12 of the terminals must be connected. Otherwise the measurement will be incorrect.
If C.ınp input type “50” is selected, the measurement terminals 10 and 11 of the terminals must be connected. Otherwise the measurement will be incorrect.
EPV241A-E-02-R-201307
Note : 1) Panel thickness should be maximum 6 mm.
2) There must be at least 60mm free space
behind the device, otherwise it would be
difficult to remove it from the panel.
For removing mounting clamps:
- Push the flush-mounting clamp in direction 1 as
shown in the figure left.
- Then, pull out the clamp in direction 2.

&
&
Or
SET
SET SET
SET SET
Used for displaying and configuring the selected parameter value.
Programming
key
When held down for a few seconds, configured numeric value increases faster.
.
Used for increasing the setpoint value and changing parameters.
When held down for a few seconds, configured numeric value decreases faster.
Used for decreasing the setpoint value and changing parameter.
Increment
key
Decrement
key
EPV241A PROGRAMMING DIAGRAM
pressed, upper limit
value can be adjusted.
If both keys are
Measurement value
If this key is pressed,
upper limit value
appears.
If this key is pressed,
upper limit value
appears.
If both keys are
pressed, lower limit
value can be adjusted.
ADJUSTING THE ALARM VALUE
By using and keys, upper limit value can be adjusted between -500V and 500V.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
This parameter can’t be lower than ( ) value.lo.l1 + HYS.l + HYS.U
By using and keys, upper limit value can be adjusted between -500V and 500V.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
This parameter can’t be lower than (UP.L1 ) value. - HYS.l - HYS.U
Loc mode; the menu can not be accessed and other functions.
It can be seen only the measured value
Loc
Measurement value
If key is pressed, “unL” mode is entered.
uNL mode; the menu and other functions can be accessed.
unL
Measurement value
&&
If these keys are pressed together, revision date appears as day and month.
12.03 2013
If key is pressed, revision date is displayed as year.
Measurement value
If key is pressed, “Loc” mode is entered.
&
If both & keys are pressed and held for 3 seconds, programming mode is entered. If keys are pressed while parameter names are displayed, then it returns to measured
value made.
Upper limit
value
SET
By using and keys, upper limit value can be adjusted between -500V and 500V.
( This parameter can’t be lower than ( ) value.lo.l1 + HYS.l + HYS.U
If this key is pressed,
upper limit value
appears. SET
If key is pressed, it returns to parameter name.
Hysteresis value
for upper limit
SET
By using and keys, hysteresis value for upper limit can be adjusted between 0.0V and
20.0V. This parameter can’t be higher than ( - - ) value. Up.l1 Lo.l1 Hys.l
If this key is pressed,
hysteresis value
for upper limit appears
SET
If key is pressed, it returns to parameter name.
Delay time for
upper limit alarm
SET
By using and keys, delay time for upper limit alarm can be adjusted between 0 and
900 seconds.
delay time for upper
limit appears
If this key is pressed,SET
If key is pressed , it returns to parameter name.
Lower limit
value
SET
By using and keys, lower limit value can be adjusted between -500V and 500V.
This parameter can’t be higher than ( ) value.UP.l1 - HYS.U - HYS.L
If this key is pressed,
lower limit value
appears.
SET
If key is pressed, it returns to parameter name.
SET
By using and keys, hysteresis value for lower limit can be adjusted between 0.0V and
20.0V. This parameter can’t be higher than ( - - ) value. Up.l1 Lo.l1 Hys.U
If this key is pressed,
hysteresis value
for lower limit
appears
SET
If key is pressed,it returns to parameter name.
Delay time for
lower limit alarm
SET
By using and key,s delay time for lower limit alarm can be adjusted between 0 and
900 seconds.
delay time for lower
limit appears
If this key is pressed,SET
If key is pressed ,it returns to parameter name.
Alarm
output position
SET
By using and keys, alarm output state can be adjusted to or If it is adjusted no
at alarm condition.
n.o. N.c. ,
output relay is powered
If this key is pressed,
alarm output state
appears.
SET
If key is pressed ,it returns to parameter name.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
10.0 If C.Inp;it is 50, this parameter adjustable between 0.00 and 20.00.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
If C.Inp;it is 50, this parameter adjustable between 0.00 and 20.00.
10.0
Hysteresis value
for lower limit
V
SET
ENDA
ACDC
DC
ACOUT
EPV 141
V
SET
ENDA
ACDC
DC
ACOUT
EPV 141
SETSET
EPV 241A
ENDA
OUT AC DC ACDC
V
PROGRAMMING MODE
3/5EPV241A-E-02-R-201307

*
*
Sampling time of the measurement value is shown on the screen. If the 1 selected;sampling time of the measurement is 250ms,
If 2 is selected,it is 500ms.If 3 is selected, it is 750ms.If 1 is selected,it is 1 second.
Optn
If this key is pressed,
measuring type
appears.
By using and keys, measuring type can be adjusted to , or ac dc acdc.Three leds
at the top of the display show measuring type.
Measuring
type
SET
If this key is pressed,
address of the device appears.
Adrs 1
By using and keys address of the device can be adjusted between 1-247.
SET
**
If this key is pressed,
sampling time measurement
value appears.
SET
4
Sampling time
measurement value
ACdC
SET
If key is pressed,it returns to parameter name.
SET
If key is pressed, it returns to parameter name.
SET
If key is pressed, it returns to parameter name.
SET
baud OFF
By using and keys, baudrate value of the device can be adjusted to OFF,1200,2400,4800,9600,19200.
SET
If key is pressed,it returns to parameter name.
(*) There are only
(**)The AdrS and baud parameters are only in the devices those have modbus.
,tYPE, optn, CInp parameters in the devices those have no relay.
If any key is pressed in 25 seconds or the device is powered down and powered upş,then it returns to operation mode.
NOTE:
**
If key is held down while the device is powered up, the d.PAr message will appear and the factory settings will be restored.
Means, measured current value is higher than maximum scale. Means, measured current value is lower than minimum scale.
ERROR MESSAGES
When this key is pressed,
baudrate is seen.
SET
C.Inp 500
*
When this key is pressed,
measurement input type
s seen.
If C.Inp input type “500” is selected, the measurement terminals 9 and 12 of the terminals must be connected.Otherwise the measurement is done incorrectly.
If Cınp input type “50” is selected, the measurement terminals 10 and 11 of the terminals must be connected.Otherwise the measurement is done incorrectly.
4/5
Factory settings restored after CInp;50,HYSL and HYSU are set to “1.00”
EPV241A-E-02-R-201307

Holding Register
Addresses
Decimal Hex
Data
Type
Data Content Parameter
Name
Read/Write
Permission Status
Value
0000d
0001d
0002d
0003d
0004d
0005d
0006d
0007d
0008d
0009d
0x0000
0x0001
0x0002
0x0003
0x0004
0x0005
0x0006
0x0007
0x0008
0x0009
word
word
word
word
word
word
word
word
word
word
The upper limit of the setpoint
The upper limit of the hysteresis value
Delay time for the upper limit alarm
The lower limit of the setpoint
The lower limit of the hysteresis value
Delay time for the lower limit alarm
Measurement method (0=AC,1=DC,2=ACDC)
Device address for RS485 network connection.
Adjustable between 1-247.
Baudrate (0=Off;1=1200;2=2400; 3=4800; 4=9600; 5=19200)
uPL1
dlyU
LoL1
HYSL
dLyl
HYSU
type
OPTN
aDRS
BAUD
Readable/Writable 500
0
30
0
0
30
aCdC
4
1
oFF
1.1 HOLDING REGISTERS
ENDA EPV241A DİJİTAL VOLTMETER MODBUS PROTOCOL ADDRESS MAP
Input Register
Addresses
Decimal Hex
Data
Type Data Content Parameter
Name Read/Write Permission
0000d 0x0000 word Measured voltage value -- Only Readable
1.2 INPUT REGISTERS
Discrete Input
Addresses
Decimal Hex
Data
Type
Data Content Parameter
Name
00d 0x00 Bit Relay output state (0=OFF; 1=ON) --
1.3 DISCRETE INPUTS
Coil Addresses
Decimal Hex
Data
Type Data Content Parameter
Name
Read/Write
Permission
00d 0x00 Bit Alarm output state (0=no; 1=nc)Aout
1.4 COILS
Status
Value
no
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Only Readable
Read/Write Permission
Readable/Writable
5/5
Sampling time of the measurement value.If 1 is selected,it is
250ms.If 2 is selected,it is 500ms.If 3 is selected,it is 750ms.
If 4 is selected,it is 1 second.
*Holding Register Parameter Table (No Relay Models)
*Coil and Discrete input parameters are not available in the devices those have no relay
0000d
0001d
0002d
0003d
0x0000
0x0001
0x0002
0x0003
word
word
word
word
Sampling time of the measurement value
Measurement method (0=AC,1=DC,2=ACDC)
Device address for RS485 network connection.
Adjustable between 1-247.
Baudrate (0=Off;1=1200;2=2400; 3=4800; 4=9600; 5=19200)
type
OPTN
aDRS
BAUD
aCdC
4
1
oFF
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
0001d 0x0001 word Measured voltage value.(While the
measured value of 3-digit decimal point is read as.
This value is “0” in other cases.) -- Only Readable
EPV241A-E-02-R-201307

OUTPUTS
Alarm output Relay: 250V AC, 8A (for resistive load), NO+NC
Mechanical 30.000.000 ; Electrical 100.000 operation.
Life expectancy for relay
0 ... +50°C/-25 ... 70°C
HOUSING
Housing type Rail mountable (EN60715,TH35)
Dimensions W54xH94xD68mm
Weight
Enclosure material Self extinguishing plastics.
ENDA EPV541 AC/DC VOLTMETER
While cleaning the device, solvents (thinner, benzine, acid etc.) or corrosive materials must not be used.
Approx. 50g (after packing)
2
* 54 x 94mm sized.
* 3 digits display.
*
* Easy to configure with front panel keypad.
* Multifunctional alarm output (NO+NC) for upper and lower limits.
* Communication feature over isolated RS485, using
ModBus RTU protocol. (Functional).
* Measuring type can be selected as AC, DC or true RMS.
* CE marked according to Europan Norms.
Values between -100V and 100 V can be indicated with one decimal point.
* For maximum 50V AC/DC measurements, measurements between -50V
and +50V can be shown with two decimal digits by having 10 times more
accurate measurement input.
Supply 230V AC +10% -20%, 50/60Hz or 24V AC 10% , 50/60Hz or optional 9-30V DC / 7-24V AC 10% SMPS± ±
Power consumption
2.5mm² screw-terminal connections
Wiring
Scale
Sensitivity
ELECTRICAL CHARACTERISTICS
If C.InP; 500 is selected, it is 0V...500V or If C.ınP ;50 is selected, it is 0....50V
If C.InP ;500 is selected, it is -500V DC....500V DC or If C.InP;50 is selected, it is -50V DC....50V DC
0,01V (If C.InP;50 is selected )
Max. 5VA
AC and RMS :
DC :
±
Accuracy
EMC
Safety requirements
EN 61326-1: 2012
EN 61010-1: 2010 (Pollution degree 2, overvoltage category II)
1%
AC
RMS
DC ±
1%
1%
(Full scale) (For square wave form 2%)±
(Full scale)
Input Impedance 870kΩ
Frequency Range
Input Range -500V...500V (If CINP 500 is selected, device breaks down at more than ±1250 DC voltages.)
-50V....50V (If CInp 50 is selected, device breaks down at more than ±125 DC voltages.)
± (Full scale) (For square wave form 2%)
±
DC , 10Hz - 200Hz (For square wave form 10Hz-70Hz)
Thank you for choosing ENDA EPV541 AC/DC voltmeter.
Technical Specifications
.Do not use the device in locations subject to corrosive and flammable gases
ENVIRONMENTAL CONDITIONS
Ambient/stroge temperature
Max. Relative humidity
Rated pollution degree
Height
80% Relative humidity for temperatures up to 31 % °C,decreasing linearly to 50% at 40°C.
According to EN 60529 Front panel : IP65 , Rear panel : IP20
Max. 2000m
0,1V (If CINP;500 is selected and higher than -100V or lower than 100V for input values)
1V (If CINP;500 is selected and lower than -100V or higher than 100V for input values)
1-Output
R......Relay
None...No relay
2-Supply Voltage
230VAC...230V AC RSI..... Insulated ModBus (optional)
110VAC...110V AC
24VAC.....24V AC
SM...........9-30V DC / 7-24V AC
3-ModBus
Order Code : EPV541 - - -
23
1
EPV541-E-02-201409
1/4
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
Barbaros Cad. No:18
ENDA SET
1 2 3 4 56789
1817
1615
14
13
12
11
10
A
OUT AC DC ACDC
ENDA
V
EPV541
OUT AC DC ACDC
SET
Read this document carefully before using this device. The guarantee will be expired by device
damages if you don't attend to the directions in the user manual. Also we don't accept any
compensations for personal injury, material damage or capital disadvantages.

10
11
UP.l1
UP.l1 HYS.U +
UP.l1 HYS.U -
LO.l1
LO.l1 HYS.L +
LO.l1 HYS.L -
Out
Out
dly.U
A.out n.c. =
A.out n.o. =
dly.U dly.ldly.l
Input
t
t
t
A
-A
0
A
0
A
0
A
-A
A
0
A
0
A
-A
0
T3T/2
T/2 2T
T3T/2
T/2 2T
T3T/2
T/2 2T
T3T/2
T/2
0
T
3T/2
T/2
2T
T3T/2
T/2 2T
dd
T2T
0.308 A
0.386 A
0.000
A
0.000
A
d
Td
T
A
A
d
T
Ad
T2
2
0.000
1
2
A1
2
A
2
A1
2
Ap
1
2
A
1
Ap
2
A1
2
A12
A1
3
A1
3
A1
ac Ac.dc
dc (rms)
CONNECTION DIAGRAM
ENDA EPV541 is intended for installation in control panels.Make sure that the device is used only for intended purpose.The electrical
connections must be carried on by a qualified staff and must be according to the relevant locally applicable regulations. During an installation,
all of the cables that are connected to the device must be free of electrical power. The device must be protected against inadmissible humidity,
vibrations and severe soiling. Make sure that the operation temperature is not exceeded. The cables should not be close to the power cables
or components.
SUPPLY:
NOTE :
10
11
184-253V AC
50/60Hz 5VA
Line
Neutral
230V AC
Supply
Switch
Cable size: 1,5mm²
Fuse
F 100mA
250V AC
Fuse should be connected.
1) Mains supply cords shall meet the requirements of IEC 60227 or IEC 60245.
2) In accordance with the safety regulations, the power supply switch shall bring the identification of the relevant instrument and it should be easily
accessible by the operator.
If C.Inp input type “500” is selected, the measurement terminals 1 and 4 of the terminals must be connected. Otherwise the measurement will be
incorrect.
If C.ınp input type “50” is selected, the measurement terminals 2 and 3 of the terminals must be connected. Otherwise the measurement will be
incorrect.
Line
Neutral
3
2
68mm
94mm
54mm
1
46mm
46,5mm
Rail-lock
For mounting the device on rail :
Push the device to rail in direction 1 and make
shure that rail-lock is interlocked to rail.
For removing the device from rail :
Push the rail-lock with a flat tip screwdriver in
direction 2 and pull the device in direction 3 .
INDUSTRIAL ELECTRONICS
EPV541-230VAC-RSI
AC/DC VOLTMETER
ENDA
8 680407 710539
7 8 9
COM
B
A
INDUSTRIAL ELECTRONICS
EPV541-230VAC
AC/DC VOLTMETER
ENDA
8 680407 710522
SN: XXXXXXXXX
11
10
567 8 9
230V AC +10% -20%
50/60Hz 5VA
CAT II
Made in Turkey
SN: XXXXXXXXX
11
10
56
230V AC +10% -20%
50/60Hz 5VA
CAT II
Made in Turkey
Max.
500V
AC/DC
1 2 3 4
+
Max.
50V
AC/DC
+
Max.
500V
AC/DC
1 2 3 4
+
Max.
50V
AC/DC
+
INDUSTRIAL ELECTRONICS
EPV541-R-230VAC
AC/DC VOLTMETER
ENDA
8 680407 710546
OUTPUT
250V AC 8A
RESISTIVE
LOAD
INDUSTRIAL ELECTRONICS
EPV541-R-230VAC-RSI
AC/DC VOLTMETER
ENDA
8 680407 710553
7 8 9
COM
B
A
SN: XXXXXXXXX
15
14
13
12
11
10
567 8 9
230V AC +10% -20%
50/60Hz 5VA
CAT II
Made in Turkey
OUTPUT
250V AC 8A
RESISTIVE
LOAD
Max.
500V
AC/DC
SN: XXXXXXXXX
15
14
13
12
11
10
1 2 3 4 5 6
230V AC +10% -20%
50/60Hz 5VA
+
CAT II
Made in Turkey
Max.
50V
AC/DC
+
Max.
500V
AC/DC
1 2 3 4
+
Max.
50V
AC/DC
+
Equipment is protected throughout by DOUBLE
INSULATION
Holding screw 0.4-0.5Nm.
2/4 EPV541-E-02-201409

&
Or
Used for displaying and configuring the selected parameter value.
Programming
key
When held down for a few seconds, configured numeric value increases faster.
.
Used for increasing the setpoint value and changing parameters.
When held down for a few seconds, configured numeric value decreases faster.
Used for decreasing the setpoint value and changing parameter.
Increment
key
Decrement
key
EPV541 PROGRAMMING DIAGRAM
pressed, upper limit
value can be adjusted.
If both keys are
Measurement value
If this key is pressed,
upper limit value
appears.
If this key is pressed,
upper limit value
appears.
If both keys are
pressed, lower limit
value can be adjusted.
ADJUSTING THE ALARM VALUE
By using and keys, upper limit value can be adjusted between -500V and 500V.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
This parameter can’t be lower than ( ) value.lo.l1 + HYS.l + HYS.U
By using and keys, upper limit value can be adjusted between -500V and 500V.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
This parameter can’t be lower than (UP.L1 ) value. - HYS.l - HYS.U
Loc mode; the menu can not be accessed and other functions.
It can be seen only the measured value
Loc
Measurement value
If key is pressed 3 seconds while in Loc mode , “unL” mode is entered.
uNL mode; the menu and other functions can be accessed.
unL
Measurement value
&
&
If these keys are pressed together, revision date appears as day and month.
12.03 2013
If key is pressed, revision date is displayed as year.
Measurement value
If key is pressed, 3 seconds “Loc” mode is entered.
&
If both keys are pressed and held for 3 seconds, programming mode is entered. If keys are pressed while parameter names are displayed, then it returns to
measured value made.
Upper limit
value
By using and keys, upper limit value can be adjusted between -500V and 500V.
( This parameter can’t be lower than ( ) value.lo.l1 + HYS.l + HYS.U
If this key is pressed,
upper limit value
appears.
If key is pressed, it returns to parameter name.
Hysteresis value
for upper limit
By using and keys, hysteresis value for upper limit can be adjusted between 0.0V and
20.0V. This parameter can’t be higher than ( - - ) value. Up.l1 Lo.l1 Hys.l
If this key is pressed,
hysteresis value
for upper limit appears If key is pressed, it returns to parameter name.
Delay time for
upper limit alarm
By using and keys, delay time for upper limit alarm can be adjusted between 0 and 900 seconds.
delay time for upper
limit appears
If this key is pressed,
If key is pressed , it returns to parameter name.
Lower limit
value
By using and keys, lower limit value can be adjusted between -500V and 500V.
This parameter can’t be higher than ( ) value.UP.l1 - HYS.U - HYS.L
If this key is pressed,
lower limit value
appears. If key is pressed, it returns to parameter name.
By using and keys, hysteresis value for lower limit can be adjusted between 0.0V and
20.0V. This parameter can’t be higher than ( - - ) value. Up.l1 Lo.l1 Hys.U
If this key is pressed,
hysteresis value
for lower limit
appears If key is pressed,it returns to parameter name.
Delay time for
lower limit alarm
By using and key,s delay time for lower limit alarm can be adjusted between 0 and 900 seconds.
delay time for lower
limit appears
If this key is pressed,
If key is pressed ,it returns to parameter name.
Alarm
output position
By using and keys, alarm output state can be adjusted to or If it is adjusted no
at alarm condition.
n.o. N.c. ,
output relay is powered
If this key is pressed,
alarm output state
appears. If key is pressed ,it returns to parameter name.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
10.0 If C.Inp;it is 50, this parameter adjustable between 0.00 and 20.00.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
If C.Inp;it is 50, this parameter adjustable between 0.00 and 20.00.
10.0
Hysteresis value
for lower limit
PROGRAMMING MODE
ENDA
A
EPA541
OUT AC DC ACDC
ENDA
V
EPV541
OUT AC DC ACDC
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
&
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
&
EPV541-E-02-201409
3/4

*
*
Sampling time of the measurement value is shown on the screen. If the 1 selected;sampling time of the measurement is 250ms,
If 2 is selected,it is 500ms.If 3 is selected, it is 750ms.If 1 is selected,it is 1 second.
Optn
If this key is pressed,
measuring type
appears.
By using and keys, measuring type can be adjusted to , or ac dc acdc.Three leds
at the top of the display show measuring type.
Measuring
type
If this key is pressed,
device address appears.
Adrs 1
By using and keys device address can be adjusted between 1-247.
**
If this key is pressed,
sampling time measurement
value appears.
Sampling time
measurement value
ACdC
If key is pressed,it returns to parameter name.
If key is pressed, it returns to parameter name.
If key is pressed, it returns to parameter name.
baud OFF
By using and keys, baudrate value of the device can be adjusted to OFF,1200,2400,4800,9600,19200.
If key is pressed,it returns to parameter name.
**
C.Inp 500
*
(*) There are only
(**) The AdrS and baud parameters are only in the devices those have modbus.
, tYPE , optn and C.inP parameters in the devices those have no relay.
NOTE : If key is held down while the device is powered up, d.PAr message will appear and factory settings will be restored.
i
i
For returning to measured value and saving to new settings, one of the following options can be applied ;
a) Wait for 25 seconds without pressing any key after the pogramming.
b) Press and keys together
c) Power down and power up the device .
after the pogramming.
after the pogramming
Means, measured current value is higher than maximum scale. Means, measured current value is lower than minimum scale.
ERROR MESSAGES
If C.Inp input type “500” is selected, the measurement terminals 1 and 4 must be connected. Otherwise the measurement will be incorrect.
If C.ınp input type “50” is selected, the measurement terminals 2 and 3 must be connected. Otherwise the measurement will be incorrect.
Holding Register
Addresses
Decimal Hex
Data
Type Data Content Parameter
Name
Read/Write
Permission Status
Value
0000d
0001d
0002d
0003d
0004d
0005d
0006d
0007d
0008d
0009d
0x0000
0x0001
0x0002
0x0003
0x0004
0x0005
0x0006
0x0007
0x0008
0x0009
word
word
word
word
word
word
word
word
word
word
The upper limit of the setpoint
The upper limit of the hysteresis value
Delay time for the upper limit alarm
The lower limit of the setpoint
The lower limit of the hysteresis value
Delay time for the lower limit alarm
Measurement method (0=AC,1=DC,2=ACDC)
Device address for RS485 network connection. Adjustable between 1-247.
Baudrate (0=Off;1=1200;2=2400; 3=4800; 4=9600; 5=19200)
uPL1
dlyU
LoL1
HYSL
dLyl
HYSU
typE
OPTN
aDRS
BAUD
Readable/Writable 500
0
30
0
0
30
aCdC
4
1
oFF
1.1 HOLDING REGISTERS
ENDA EPV541 DIGITAL VOLTMETER MODBUS PROTOCOL ADDRESS MAP
Input Register
Addresses
Decimal Hex
Data
Type Data Content Parameter
Name Read/Write Permission
0000d 0x0000 word Measured voltage value -- Only Readable
1.2 INPUT REGISTERS
Discrete Input
Addresses Data
Type Data Content Parameter
Name
1.3 DISCRETE INPUTS
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Read/Write Permission
Sampling time of the measurement value.If 1 is selected,it is 250ms.If 2 is selected,
it is 500ms.If 3 is selected,it is 750ms. If 4 is selected,it is 1 second.
*Holding Register Parameter Table (No Relay Models)
0000d
0001d
0002d
0003d
0x0000
0x0001
0x0002
0x0003
word
word
word
word
Sampling time of the measurement value
Measurement method (0=AC,1=DC,2=ACDC)
Device address for RS485 network connection. Adjustable between 1-247.
Baudrate (0=Off;1=1200;2=2400; 3=4800; 4=9600; 5=19200)
typE
OPTN
aDRS
BAUD
aCdC
4
1
oFF
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
0001d 0x0001 word Measured voltage value. (While the measured value of 3-digit decimal point is
read as. This value is “0” in other cases.) -- Only Readable
Decimal Hex
00d 0x00 Bit Relay output state (0=OFF; 1=ON) --
Coil Addresses
Decimal Hex
Data
Type Data Content Parameter
Name
Read/Write
Permission
00d 0x00 Bit Alarm output state (0=no; 1=nc)Aout
1.4 COILS
Status
Value
no
Only Readable
Readable/Writable
SET
SET
SET
SET
SET
SET
4
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
If this key is pressed,
baudrate value appears.
If this key is pressed,
measurement input
type appears.
By using and keys measurement input type can be selected as 500 or 50 .
If key is pressed, it returns to parameter name.
SET
SET
If input type 500 is selected, max. 500V DC or 500V AC can be applied.
If input type 50 is selected, max. 50V DC or 50V AC can be applied.
Before the factory defaults, if the C.inp 50 is selected, HySL and HysU will be applied as 1.00 after the factory defaults.
*Coil and Discrete input parameters are not available in the devices those have no relay
4/4 EPV541-E-02-201409

Read this document carefully before using this device. The guarantee will be expired by device damages if you
don't attend to the directions in the user manual. Also we don't accept any compensations for personal injury,
material damage or capital disadvantages.
PROGRAMMABLE AC/DC VOLTMETER
ENDA
EPV 741A
SET
OUT AC DC AC/DC
LOCK
OUTPUTS
Alarm output Relay: 250V AC, 8A (for resistive load), N.O.
Mechanical 30.000.000 ; Electrical 100.000 operation.
Life expectancy for relay
0 ... +50°C/-25 ... 70°C
HOUSING
Housing type Suitable for flush-panel mounting. (According to DIN 43 700)
Dimensions W72xH72xD97mm
Weight
Enclosure material Self extinguishing plastics.
ENDA EPV741A AC/DC VOLTMETER
While cleaning the device, solvents (thinner, benzine, acid etc.) or corrosive materials must not be used.
Approx. 50g (after packing)
2
* 72 x 72mm sized.
* 3 digits display.
*
* Easy to configure with front panel keypad.
* Multyfunctional alarm output (N.O.) for upper and lower limits.
* Communication feature over isolated RS485, using
ModBus RTU protocol (Functional).
* Measuring type can be selected as AC, DC or true RMS.
* CE marked according to Europan Norms.
Values between -100V and 100 V can be indicated with one decimal point.
* For maximum 50V AC/DC measurements, measurements between -50V
and +50V can be shown with two decimal digits by having 10 times more
accurate measurement input.
Supply 230V AC +10% -20%, 50/60Hz or 24V AC 10% , 50/60Hz or optional 9-30V DC / 7-24V AC 10% SMPS± ±
Power consumption
2.5mm² screw-terminal connections
Wiring
Scale
Sensitivity
ELECTRICAL CHARACTERISTICS
If C.InP; 500 is selected, it is 0V...500V or If C.ınP ;50 is selected, it is 0....50V
If C.InP ;500 is selected, it is -500V DC....500V DC or If C.InP;50 is selected, it is -50V DC....50V DC
0,01V (If C.InP;50 is selected )
Max. 5VA
AC and RMS :
DC :
±
Accuracy
EMC
Safety requirements
EN 61326-1: 2013
EN 61010-1: 2012 (Pollution degree 2, overvoltage category II)
1%
AC
RMS
DC ±
1%
1%
(Full scale) (For square wave form 2%)±
(Full scale)
Input Impedance 870kΩ
Frequency Range
Input Range -500V...500V (If CINP 500 is selected, device breaks down at more than ±1250 DC voltages.)
-50V....50V (If CInp 50 is selected, device breaks down at more than ±125 DC voltages.)
± (Full scale) (For square wave form 2%)
±
DC , 10Hz - 200Hz (For square wave form 10Hz-70Hz)
Thank you for choosing ENDA EPV741A AC/DC voltmeter.
Technical Specifications
.Do not use the device in locations subject to corrosive and flammable gases
ENVIRONMENTAL CONDITIONS
Ambient/stroge temperature
Max. Relative humidity
Rated pollution degree
Height
Max. humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C.
According to EN 60529 Front panel : IP65 , Rear panel : IP20
Max. 2000m
0,1V (If CINP;500 is selected and higher than -100V or lower than 100V for input values)
1V (If CINP;500 is selected and lower than -100V or higher than 100V for input values)
1-Output
R......Relay
None...No relay
2-Supply Voltage
230VAC...230V AC RSI..... Insulated ModBus (optional)
110VAC...110V AC
24VAC.....24V AC
SM...........9-30V DC / 7-24V AC
3-ModBus
Order Code : EPV741A- - -
23
1
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
Barbaros Cad. No:18
1/5 EPV741A-E-02-201406

72mm
78mm
PROGRAMMABLE AC/DC VOLTMETER
ENDA
EPV 741A
SET
OUT AC DC ACDC
LOCK
97mm
2
2
Panel
1
1
75mm
84mm
68 mm
68 mm
UP.l1
UP.l1 HYS.U +
UP.l1 HYS.U -
LO.l1
LO.l1 HYS.L +
LO.l1 HYS.L -
Out
Out
dly.U
A.out n.c. =
A.out n.o. =
dly.U dly.ldly.l
Input
t
t
t
2/5
V
-V
0
ac.dc
3T/2T/2 T2T 0.000
V
d
T
V
1
2
V
2
V1
1
2
V
2
V1
3
V1
3
V1
19
20
EPV741A-E-02-201406
V
-A
0
V
0
V
0
V
-V
V
0
V
0
ac dc
T3T/2
T/2 2T
T3T/2
T/2 2T
T3T/2
T/2 2T
T3T/2
T/2
0
T2T
T3T/2
T/2 2T
dd
0.308 V
0.386 V
0.000
V
d
T
V
d
T
Vd
T2
2
0.000
1
2
V
2
Vp
1
Vp
2
V12
V1
ac Ac.dc
dc (rms)
230V AC +10% -20%
50/60Hz 7VA
11
12
13
14
15
OUTPUT
250V AC 8A
RESISTIVE
LOAD
SN: XXXXXXXXX
Max. 50V
AC / DC
+
16
17
18
8680407705092
ENDA
INDUSTRIAL ELECTRONICS
EPV741A-R-230VAC-RSI
PROGRAMMABLE AC/DC VOLTMETER
CAT II
COM
B
A
Made in Turkey +
Max. 500V
AC / DC
230V AC +10% -20%
50/60Hz 7VA
19
20
11
12
13
14
15
SN: XXXXXXXXX
Max. 50V
AC / DC
+
16
17
18
8680407705078
ENDA
INDUSTRIAL ELECTRONICS
EPV741A-230VAC-RSI
PROGRAMMABLE AC/DC VOLTMETER
CAT II
COM
B
A
Made in Turkey +
Max. 500V
AC / DC
17
18
Line
Neutral
CONNECTION DIAGRAM
Equipment is protected throughout
by
DOUBLE INSULATION
Holding screw 0.4-0.5Nm.
ENDA EPV741A is intended for installation in control panels.Make sure that the
device is used only for intended purpose.The electrical connections must be carried
on by a qualified staff and must be according to the relevant locally applicable
regulations. During an installation, all of the cables that are connected to the device
must be free of electrical power. The device must be protected against inadmissible
humidity, vibrations and severe soiling. Make sure that the operation temperature is
not exceeded. The cables should not be close to the power cables or components.
SUPPLY:
NOTE :
17
18
184-253V AC
50/60Hz 7VA
Line
Neutral
230V AC
Supply
Switch
Cable size: 1,5mm²
Fuse
F 100mA
250V AC
Fuse should be connected.
1) Mains supply cords shall meet the requirements of IEC 60227 or IEC 60245.
2) In accordance with the safety regulations, the power supply switch shall bring the identification of
the relevant instrument and it should be easily accessible by the operator.
If C.Inp input type “500” is selected, the measurement terminals 13 and 16 of the
terminals must be connected. Otherwise the measurement will be incorrect.
If C.ınp input type “50” is selected, the measurement terminals 14 and 15 of the
terminals must be connected. Otherwise the measurement will be incorrect.
- Push the flush-mounting clamp
in direction 1 as shown in the
figure left.
- Then, pull out the clamp in
direction 2.
For removing mounting clamps:
Note
1) During panel mounting, additional distance is
required for connection cables.
2) Panel thickness should be maximum 10mm.
3) There must be at least 90mm free space
behind the device, otherwise it would be
difficult to remove it from the panel.
Flush mounting
clamp
Rubber packing
Connection
cables
DIMENSIONS Panel cut-out
Depth

PROGRAMMABLE AC/DC VOLTMETER
ENDA
EPV 741A
SET
OUT AC DC AC/DC
LOCK
EPV741A PROGRAMMING DIAGRAM
SET
Used for displaying and configuring the selected parameter value.
Programming
key
When held down for a few seconds, configured numeric value increases faster.
.
Used for increasing the setpoint value and changing parameters.
When held down for a few seconds, configured numeric value decreases faster.
Used for decreasing the setpoint value and changing parameter.
Increment
key
Decrement
key
&
&
Or
SET SET
SET SET
pressed, upper limit
value can be adjusted.
If both keys are
value
Measurement If this key is pressed,
upper limit value
appears.
If this key is pressed,
upper limit value
appears.
If both keys are
pressed, lower limit
value can be adjusted.
ADJUSTING THE ALARM VALUE
By using and keys, upper limit value can be adjusted between -500V and 500V.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
This parameter can’t be lower than ( ) value.lo.l1 + HYS.l + HYS.U
By using and keys, upper limit value can be adjusted between -500V and 500V.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
This parameter can’t be lower than (UP.L1 ) value. - HYS.l - HYS.U
Loc mode; the menu can not be accessed and other functions.
It can be seen only the measured value
Loc
Measurement value
If key is pressed, “unL” mode is entered.
uNL mode; the menu and other functions can be accessed.
unL
Measurement value
&&
If these keys are pressed together, revision date appears as day and month.
12.03 2013
If key is pressed, revision date is displayed as year.
Measurement value
If key is pressed, “Loc” mode is entered.
&
If these keys are pressed and held for 3 seconds, programming mode is entered. If keys are pressed while parameter names are displayed, then it returns
to measured value mode.
Upper limit
value
SET
By using and keys, upper limit value can be adjusted between -500V and 500V.
( This parameter can’t be lower than ( ) value.lo.l1 + HYS.l + HYS.U
If this key is pressed,
upper limit value
appears. SET
If key is pressed, it returns to parameter name.
Hysteresis value
for upper limit
SET
By using and keys, hysteresis value for upper limit can be adjusted between 0.0V and
20.0V. This parameter can’t be higher than ( - - ) value. Up.l1 Lo.l1 Hys.l
If this key is pressed,
hysteresis value
for upper limit appears
SET
If key is pressed, it returns to parameter name.
Delay time for
upper limit alarm
SET
By using and keys, delay time for upper limit alarm can be adjusted between 0 and
900 seconds.
delay time for upper
limit appears
If this key is pressed,SET
If key is pressed , it returns to parameter name.
Lower limit
value
SET
By using and keys, lower limit value can be adjusted between -500V and 500V.
This parameter can’t be higher than ( ) value.UP.l1 - HYS.U - HYS.L
If this key is pressed,
lower limit value
appears.
SET
If key is pressed, it returns to parameter name.
SET
By using and keys, hysteresis value for lower limit can be adjusted between 0.0V and
20.0V. This parameter can’t be higher than ( - - ) value. Up.l1 Lo.l1 Hys.U
If this key is pressed,
hysteresis value
for lower limit
appears
SET
If key is pressed,it returns to parameter name.
Delay time for
lower limit alarm
SET
By using and key,s delay time for lower limit alarm can be adjusted between 0 and
900 seconds.
delay time for lower
limit appears
If this key is pressed,SET
If key is pressed ,it returns to parameter name.
Alarm
output position
SET
By using and keys, alarm output state can be adjusted to or If it is adjusted no
at alarm condition.
n.o. N.c. ,
output relay is powered
If this key is pressed,
alarm output state
appears.
SET
If key is pressed ,it returns to parameter name.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
10.0 If C.Inp;it is 50, this parameter adjustable between 0.00 and 20.00.
If C.Inp;it is 50, this parameter adjustable between -50.00V-50.00V.
If C.Inp;it is 50, this parameter adjustable between 0.00 and 20.00.
10.0
Hysteresis value
for lower limit
LOCK
LOCK
LOCK
LOCK
SET
PROGRAMMING MODE
3/5EPV741A-E-02-201406

*
*
Sampling time of the measurement value is shown on the screen. If the 1 selected;sampling time of the measurement is 250ms,
If 2 is selected,it is 500ms.If 3 is selected, it is 750ms.If 1 is selected,it is 1 second.
Optn
If this key is pressed,
measuring type
appears.
By using and keys, measuring type can be adjusted to , or ac dc acdc.Three leds
at the top of the display show measuring type.
Measuring
type
SET
If this key is pressed,
address of the device appears.
Adrs 1
By using and keys address of the device can be adjusted between 1-247.
SET
**
If this key is pressed,
sampling time measurement
value appears.
SET
4
Sampling time
measurement value
ACdC
SET
If key is pressed,it returns to parameter name.
SET
If key is pressed, it returns to parameter name.
SET
If key is pressed, it returns to parameter name.
SET
baud OFF
By using and keys, baudrate value of the device can be adjusted to OFF,1200,2400,4800,9600,19200.
SET
If key is pressed,it returns to parameter name.
**
When this key is pressed,
baudrate is seen.
SET
C.Inp 500
*
When this key is pressed,
measurement input type
s seen.
(*) There are only
(**) The AdrS and baud parameters are only in the devices those have modbus.
, c.t.r.r , tYPE , d.Pnt , optn parameters in the devices those have no relay.
If C.Inp input type “500” is selected, the measurement terminals 13 and 16 of the terminals must
be connected. Otherwise the measurement will be incorrect.
If C.ınp input type “50” is selected, the measurement terminals 14 and 15 of the terminals must
be connected. Otherwise the measurement will be incorrect.
NOTE : If key is held down while the device is powered up, d.PAr message will appear and factory settings will be restored.
i
i
For returning to measured value and saving to new settings, one of the following options can be applied ;
a) Wait for 25 seconds without pressing any key after the pogramming.
b) Press and keys together
c) Power down and power up the device .
after the pogramming.
after the pogramming
Means, measured current value is higher than maximum scale.
Means, measured current value is lower than minimum scale.
ERROR MESSAGES
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
Barbaros Cad. No:18
4/5 EPV741A-E-02-201406

Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Readable/Writable
Decimal Hex
0000d
0001d
0002d
0003d
0004d
0005d
0006d
0007d
0008d
0009d
0x0000
0x0001
0x0002
0x0003
0x0004
0x0005
0x0006
0x0007
0x0008
0x0009
word
word
word
word
word
word
word
word
word
word
uPL1
dlyU
LoL1
HYSL
dLyl
HYSU
typE
OPTN
aDRS
BAUD
500
10.0
0
0
10.0
0
aCdC
1
oFF
1.1 HOLDING REGISTERS
Decimal Hex
0000d 0x0000 word - -
1.2 INPUT REGISTERS
Decimal Hex
00d 0x00 Bit - -
1.3 DISCRETE INPUTS
Decimal Hex
00d 0x00 Bit Aout
1.4 COILS
no
4
0000d
0001d
0002d
0003d
0x0000
0x0001
0x0002
0x0003
word
word
word
word
typE
OPTN
aDRS
BAUD
aCDC
1
oFF
4
00001 0x0001 word - -
Holding Register
Addresses Data
Type
Data Content Parameter
Name
Read/Write
Permission
Status
Value
The upper limit of the setpoint
The upper limit of the hysteresis value
Delay time for the upper limit alarm
The lower limit of the setpoint
The lower limit of the hysteresis value
Delay time for the lower limit alarm
Measurement method (0=AC,1=DC,2=ACDC)
Device address for RS485 network connection.
Adjustable between 1-247.
Baudrate (0=Off;1=1200;2=2400; 3=4800; 4=9600; 5=19200)
Readable/Writable
ENDA EPV741A DIGITAL VOLTMETER MODBUS PROTOCOL ADDRESS MAP
Input Register
Addresses Read/Write Permission
Measured voltage value Only Readable
Discrete Input
Addresses Read/Write Permission
Sampling time of the measurement value.If 1 is selected,
it is 250ms. If 2 is selected, it is 500ms. If 3 is selected,
it is 750ms. If 4 is selected, it is 1 second.
*Holding Register Parameter Table (No Relay Models)
Sampling time of the measurement value
Measurement method (0=AC,1=DC,2=ACDC)
Device address for RS485 network connection.
Adjustable between 1-247.
Baudrate (0=Off;1=1200;2=2400; 3=4800; 4=9600; 5=19200)
Measured voltage value. (While the measured
value of 3-digit decimal point is read as. This
value is “0” in other cases.)
Only Readable
Relay output state (0=OFF; 1=ON)
Coil Addresses Read/Write
Permission
Status
Value
Only Readable
Alarm output state (0=no; 1=nc)Readable/Writable
*Coil and Discrete input parameters are not available in the devices those have no relay
Data
Type Data Content Parameter
Name
Data
Type Data Content Parameter
Name
Data
Type Data Content Parameter
Name
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
Barbaros Cad. No:18
5/5 EPV741A-E-02-201406