ENDA EPA742 User manual

ENDA EPA742 Programmable AC/DC Ammeter
Thank you for choosing ENDA EPA742 programmable AC/DC ammeter.
Max. 2000m
80 Relative humidity for temperatures up to 31% °C, decreasing linearly to 50% at 40°C (should be no icing or condensation in the environment).
0 ... +50°C/-25 ... 70°C (should be no icing or condensation in the environment)
IP20 According to EN 60529
Read this document carefully before using this device. The guarantee will be expired by device demages if you don't attend to the
directions in the user manual. Also we don't accept any compensations for personal injury, material damage or capital
disadvantages.
72 x 72mm sized.
4 digits display.
5A/60 mV , CT20/30 current transformer or 1A input feature (Please Specify at Order).
Measuring type can be selected as AC, DC or True RMS.
Programmable scale range between 5A and 9999A.
Multifunctional alarm output (NO) for upper and lower limits (Optional).
0-20mA, 4-20mA, 0-10V or 1-5V output selection (Optional - Specified devices with output type as ‘A' only).
Three-way isolation architecture between input, output and supply.
Communication feature over isolated Modbus RTU protocol (Optional).
Keylock feature.
CE marked according to European Norms.
Suitable for EN60715 Standards, TH35 rail type.
W72xH72xD94mm
Self extinguishing plastics.
Avoid any liquid contact when the device is switched on.
DO NOT clean the device with solvent (thinner, gasoline, acid etc.) and / or abrasive cleaning agents.
Approx. 350g (after packing)
EN 61326-1: 2013
EN 61010-1: 2010 (Pollution degree 2, overvoltage category II)
0.002A x ( i.e. : 0.01A for c.tr.r = 5 )c.tr.r
(full scale) (± 2% For square wave form)
(full scale)
DC , 20Hz - 70Hz
Input type 5A ; -5A...5A ( Device may be damaged at 10A and above currents )
If input type is 1A, scale : -999A...9999A ( )Specified by c.tr.r parameter. i.e : scale is c.tr.r -1A...1A for = 1
%1
± %1
±
If input type is CT20/30 / 60mv :
If ityp = Ct20 , 0A...300A or Ct30 , 0A...120A ( Specified by turn parameter. i.e : scale is turn = 1 )
If ityp = 5Hnt , 0A...9999A ( )
0A..300A / 0A..120A for
Specified by c.tr.r parameter. i.e : scale is 0A...5A for c.tr.r = 5
If input type is CT20/30 / 60mV :
DC measurement can not be performed by using CT.
If ityp = 5Hnt , scale : 0A...9999A ( )Specified by c.tr.r parameter. i.e : c.tr.r scale is -5A...5A for =5
KEEP AWAY device from exposed to corrosive, volatile and flammable gases or liquids and DO NOT USE the device in similar hazardous locations.
Max. 7VA
2.5mm² screw-terminal connections
If input type is 5A / 60mV, scale 0A...9999A : ( Specified by )c.tr.r c.tr.r parameter. i.e : scale is 0A...5A for = 5
(Mechanical 30.000.000; Electrical 100.000 operation. 250V AC, 8A (resistive load).
Relay : 250V AC, 8A (for resistive load), NO
If input type is 1A, scale 0A...9999A : ( Specified by )c.tr.r c.tr.r parameter. i.e : scale is 0A...1A for = 1
If input type is 5A / 60mV, scale : -999A...9999A ( )Specified by c.tr.r parameter. i.e : scale is c.tr.r -5A...5A for = 5
90-250V AC 50/60Hz ;10-30V DC / 8-24V AC 50/60Hz SMPS
INPUTS
Input Type
AC
and
RMS
DC
AC/RMS
DC
Input type 1A ; -1A...1A ( Device may be damaged at 2A and above currents )
Input type 60V ; -60mV...60mV ( Device may be damaged at 50V and above voltages )
Input type CT20/30 ; 0 ... 150 mA
20kΩ for 60mV Input, 90mΩ for 1A Input, 12mΩ for 5A Input, 600mΩ for CT20/30 Input.
ELECTRICAL CHARACTERISTICS
OUTPUTS
0-20mA DC, or 4-20mA DC % ( Load resistance for current outputs Max. 500Ω ).±0,5
0-10V DC or 1-5V DC, ( Short circuit protected ).Max. 10mA. ±0,5%
HOUSING
ENVIRONMENTAL CONDITIONS
Housing Type
Dimensions
Weight
Enclosure Material
Height
Max. Relative Humidity
Rated Pollution Degree
Ambient/Storage Temperature
Output
Analog Output
Supply
Power Consumption
Wiring
EMC
Safety Requirements
Input Impedance
Frequency Range
Input Range
Sensitivity
Accuracy
Scale
EPA742-UV : 5A or 60mV
EPA742-UV-CT : CT20/30 current transformer or 60mV
EPA742-UV-X1 : 1A
PROGRAMMABLE AC/DC AMMETER
ENDA
EPA742
SET
LOCK
0.000
OUT AC DC AC/DC
EPA742-EN-05-220103
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
Barbaros Cad. No:18
1/5
ENDA
TM
CT20/30 should be ordered separately when required.
2 - Input Type
CT...........................CT20/30
Current Transformer input OR 60mV.
X1...........................1A
Default (Blank)......5A or 60mV.
3 - Output
R..........08A Relay
4 - Isolated Modbus
RSI........RS485 Modbus Available
( Specify at order)
Order Code : EPA742
134
2
1 - Supply Voltage
UV....90-250V AC
LV......10-30V DC /
8-24V AC

ConF Out1
tYpE
d.pnt
Optn
Adr5
bAUd
UPLL
HY5U
dLYU
LOLL
HY5L
OtYp
dLyL
& &
SET
SET
&
2/5
5dly
turn
atYp Analog Output Status ( In devices with output type "Analog"
Can be adjusted to 0-20 , 4-20 , 0-10 , 1-5 values.
Input Type ( In devices with input type "CT" )
Can be adjusted to Ct20 , CT30 , 5HNt values.
itYp
If 5Hnt is selected, 60mV input terminals must be used.
Terminal inputs :
EPA742 PROGRAMMING DIAGRAM
Programming
key
Increment
key
Decrement
key
Used for increasing the setpoint value and changing parameters.
When held down for a few seconds, configured numeric value increases faster.
When held down for a few seconds, configured numeric value decreases faster.
Used for decreasing the setpoint value and changing parameters.
Used for displaying and configuring the selected parameter value.
By pressing to keys together for 3 seconds, xcxcxc message appears and the "Programming Mode" is entered. If the keys are pressed
during "Programming Mode" or no operation is performed for 3 seconds, returns to the "Running Mode".
PROGRAMMING MODE
Current Conversion Ratio
Can be adjusted between 5(/5) and (/5). If this parameter
changes, upper limit value is set to maximum scale, minimum limit
value is set to minimum scale and hysteresis values are set to 0.1
9999
Measuring Method
Can be adjusted to , or ac dc acdc. LEDs on the top of the
display indicates the adjusted measurement method.
Decimal Indicator
0
dPnt type may change according to relay parameters value.
If measured value is lower than 10, it will be displayed as
(0.0 0) , (0.00) , (0.0) or (0) or (0).
If measured value between 10 and 100, it will be displayed as
(0.00) ,(0.0) or (0).
If measured value between 100 and 1000, it will be displayed as
(0.0) or (0).
Samping Time
If 1 ( 1 ) is selected ; sampling time of the measurement is 250ms,
If 2 ( 2 ) is selected, it is 500ms.
If 3 ( 3 ) is selected, it is 750ms.
If 4 ( 4 ) is selected, it is 1 second.
Device Address
Can be adjusted between 1 - 247 .
Baud Rate
It can be adjusted as OFF, 1200, 2400, 4800, 9600, 19200,
38400, 57600 and 115200 .
and
12
9
Out1 Output State
It can be adjusted as n.o. or If is selected, incase
of alarm, out relay is activated.
N.c. n.o
Upper Limit Value
It can be adjusted between minimum and maximum scale that
is specified with c.tr.r parameter.
This parameter can’t be lower than ( ) l0ll - HY5U - HY5U
Hysteresis Value for Upper Limit
It can be adjusted between 0 and ctrr /5 parameter. This
parameter can’t be higher than ( - - HY5L ).
When ctrr changed, HY5U gets the value of 0.1 .
Upll l0ll
Delay Time for Upper Limit Alarm
It can be adjusted between 0 and 900 .
Lower Limit Value
It can be adjusted between lower scale and upper scale that is
specified with c.tr.r parameter.
This parameter can’t be higher than (
) value.
UPLL - HY5U -
HY5L
Hysteresis Value for Lower Limit
It can be adjusted between 0 and Ctrr /5 .
This parameter can’t be higher than ( - -
) value. When ctrr is changed, HY5U gets the
value of 0.1 .
UPLL L0LL
HY5U
Delay Time for Lower Limit Alarm
It can be adjusted between 0 and 900 seconds.
Delay Time for Initial Upper Limit Alarm
It can be adjusted between 0 and 900 seconds.
SETTING UP THE PARAMETERS
If key is pressed, the current value of the parameter appears by flashing on the display.
After the setting up the parameters, if set key is pressed again, adjusted parameter
name appears on display.
By using “UP” or “DOWN” navigation keys, selected parameter can be adjusted to
the desired value.
Measured Value
L0C UL0C
0.00 0.00
LOCKING & UNLOCKING KEYPAD
SET
UPLL LOLL
0.00
0.00
+
By pressing to key for 3 seconds, quick menu is entered.
QUICK MENU
&&01.01 2015
Year
Day.Month
If these keys are pressed and held together, revision
date appears as day, month and year.
SET
0.00
REVISION NUMBER
SET
SET
Powered on device by pressing key. dPAr message appears
on display and device resets to default settings.
Measured current value is higher than maximum scale.
Measured current value is lower than minimum scale.
&&
SET
While revision information displayed and if one of the pressed key
is released, measured value is displayed again.
Measured Value
Measured Value
Measured Value
DEFAULT SETTINGS
ERROR MESSAGES
In “Running Mode” , by pressing to key for 3 seconds, keypad locked or unlocked.
SET
turn
Iin max(A)
1 2 3 4 5 6
CT20
CT30
300
Iin max(A) 150 100 75 60 50
120 60 40 30 24 20
7 8 9 10
42,8 37,5 33,3 30
17,1 15 13,3 12
CT20/30 Current Transformer & Windings Chart
Note :
If ityp parameter set to 5hnt , turn parameter is not appears.
If ityp parameter set to Ct20 or Ct30 , ctrr parameter is not
appears.
Note :
Before setting the relay parameters, the operating scale must be
determined from dpnt parameter.
If dpnt , typE and Ityp parameters are changed (if applicable),
Upll , loll , Hy5U and Hy5l values must be checked.
Number of Windings ( In devices with input type "CT" )
Number of windings of the current cable getting through the CT20/30
current transformer.
Please see "CT20/30 Current Transformer & Windings" chart on
the right side of this page.
i
SET
Conf and Out1 menus can be switched between by pressing the SET key.
key provides to access next parameter. If the keys are pressed
together, returns to the Conf menu.
This menu is available only for specified devices
with "R" (RELAY) in the order code only. Please
check "Output Type" in the order code.
typE 1000
ACdC dC 1000
AC
EPA742-EN-05-220103
PROGRAMMABLE AC/DC AMMETER
ENDA
EPA742
SET
LOCK
0.000
OUT AC DC AC/DC

1)
2) There must be at least 90mm free space
behind the device, otherwise it would be difficult
to remove it from the panel.
Panel thickness should be maximum 10mm.
- Push the flush-mounting
clamp in direction 1 as
shown in the figure left.
- Then, pull out the clamp in
direction 2.
For removing mounting clamps :
Connection
Cables
Note :
3/5
Panel Cut-out
75mm
84mm
68mm
68mm
Gasket
Flush mounting
clamp
Panel
PROGRAMMABLE AC/DC AMMETER
ENDA
EPA742
SET
LOCK
0.000
OUT AC DC AC/DC
94mm
2
1
1
2
72mm
78mm
1919
2020
1313
1414
1515
SN: XXXXXXXXXSN: XXXXXXXXX
1616
1717
1818
8680407711550
8680407711550
EPA742-UV
PROGRAMMABLE AC/DC AMMETER
EPA742-LV-CT-RSI
PROGRAMMABLE AC/DC AMMETER
Made in TurkeyMade in Turkey
1111
1212
+
ENDA
INDUSTRIAL ELECTRONICS
ENDA
INDUSTRIAL ELECTRONICS
90-250V AC
50/60Hz
7VA
5A
AC/DC
CT20/30
AC
60mV
AC/DC
60mV
AC/DC
+ +
CAT IICAT II
10-30V DC
8-24V AC
50/60Hz
7VA
Ölçme Girişi 0-120A
CT30 Akım Trafosu
CT30
B
A
ModBus
{
UP.l1
UP.l1 HYS.U +
UP.l1 HYS.U -
LO.l1
LO.l1 HYS.L +
LO.l1 HYS.L -
Out
Out
dly.U
0tYP n.c. =
0tYP n.o. =
dly.U
dly.l
dly.l
Input
t
t
t
A
-A
0
A
-A
A
-A
0
T3T/2
T/2 2T
T3T/2
T/2 2T
0
3T/2
T/2 T2T 0.000
A
0.000
A
0.000
1
2
A1
2
A
3
A1
3
A1
ac Ac.dc
dc (rms)
DIMENSIONS & CONNECTION DIAGRAM
ENDA EPA742 is intended for installation in control panels. Device must be used to according to instructions. Mounting and electrical
connections must be carried on by a qualified staff and must be according to the relevant locally applicable regulations. During an
installation, all of the cables that are connected to the device must be free of electrical power. The device must be protected against
inadmissible humidity, vibrations and severe soiling. Make sure that the operation temperature is not exceeded. The cables should not be
close to the power cables or components.
CAUTION :
Only one of the 60mV or 5A/CT20-30 inputs can be used.
Equipment is protected throughout by
DOUBLE INSULATION
Holding screw 0.4-0.5Nm.
SUPPLY:
NOTE :
17
18
90-250V AC
veya
10-30V DC /
8-24V AC
50/60Hz 7VA
Line
Neutral
UV / LV
Supply
Switch
Cable size: 1,5mm²
Fuse
F 100mA
250V AC
Fuse should be connected.
1) Mains supply cords shall meet the requirements of IEC 60227 or
IEC 60245.
2) In accordance with the safety regulations, the power supply
switch shall bring the identification of the relevant instrurment and
it should be easily accessible by the operator.
Please see page 5 for Modbus Connection Diagram EPA742-EN-05-220103

0000d
0001d
0002d
0003d
0004d
0x0000
0x0001
0x0002
0x0003
0x0004
word
word
word
word
word
word
0x00050005d
word
word
Input Type ( 0 = Ct20 , 1 = Ct30 , 2 = 5Hnt ) ityp Ct20
*0006d 0x0006
*0007d 0x0007 Number of windings for transformer
(Can be set between 1 and 10). turn 1
Current Conversion Ratio
Measurement method (0=AC, 1=DC, 2=ACDC)
Decimal point ( 0 = 0 , 1 = 0.0 , 2 = 0.00 , 3 = 0.000 )
Baudrate (0=Off;1=1200;2=2400; 3=4800; 4=9600; 5=19200
6= 38400; 7= 57600; 8= 115200)
ctrr
typE
dpnt
OPTN
aDr5
BAUD
5
aCdC
0.00
4
1
off
Sampling time of the measurement value
Device address for RS485 network connection.
Adjustable between 1-247.
R / W
R / W
R / W
R / W
R / W
R / W
R / W
R / W
* 6d and 7d addresses are available for only in EPA742-xx-CT-x-RSI CT20/30 input type devices.
0000d
0003d
0004d
0006d
0007d
0008d
0x0000
0x0003
0x0004
0x0006
0x0007
0x0008
0x0009
word
word
word
word
word
word
word
0009d
0010d word
0x000A
0011d 0x000B
0x000C
word
word
0012d
0002d 0x0002 word
0005d 0x0005 word
*0014d 0x000E word
*0015d
0x000F
word turn
ityp
1
Ct20
Number of windings for transformer
Input Type (0 = Ct20 , 1= Ct30 , 2 = 5Hnt )
*14. and *15. addresses are only used in devices with EPA742-xx-xx-R-RSI, input type CT20/30 current transformer.
0001d 0x0001 word
0013d 0x000D word
ENDA EPA742-xx-xx-R-RSI DIGITAL AMPERMETER WITH ALARM RELAY OUTPUT MODBUS PROTOCOL ADDRESS MAP
Holding Register
Adresleri Data
Type
Decimal Hex
Data Content Parameter
Name
Read/Write
Permission
Default
Value
The upper limit of the hysteresis value
Delay time for the upper limit alarm
The lower limit of the setpoint
The lower limit of the hysteresis value
Delay time for the lower limit alarm
Current replacement rate
The upper limit of the setpoint
Alarm output status
dlyU
L0Ll
HY5L
dLyl
HY5U
ctrr
0.10
5.00
0
0.10
0
no
UPLL
5
0tyP
0.00
R / W
R / W
R / W
R / W
R / W
R / W
R / W
R / W
Measurement method (0=AC, 1=DC, 2=ACDC)typE R / W aCdC
Decimal point. ( )0 = 0 , 1 = 0.0 , 2 = 0.00 , 3 = 0.000 dpnt 0.00
R / W
0PTN
Sampling time of the measurement value. If 1 is selected, it is
250ms. If 2 is selected, it is 500ms. If 3 is selected, it is 750ms.
If 4 is selected, it is 1 second.
4
R / W
aDr5 1
Device address for RS485 network connection.
Adjustable between 1-247. R / W
Baudrate (0=Off;1=1200;2=2400; 3=4800; 4=9600; 5=19200
6= 38400; 7= 57600; 8= 115200) BAUD oFF
R / W
Delay Time for Initial Upper Limit Alarm 0
5dly R / W
R / W
R / W
ENDA EPA742-xx-x-xx DIGITAL AMPERMETER WITH ANALOGUE OUTPUT MODBUS PROTOCOL ADDRESS MAP
Holding Register
Adresleri Data
Type
Decimal Hex
Data Content Parameter
Name
Read/Write
Permission
Default
Value
EPA742-EN-05-220103
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
Barbaros Cad. No:18
4/5
ENDA
TM

Input Register
Addresses
Decimal Hex
Data
Type Data Content Parameter
Name
Read / Write
Permission
0000d 0x0000 word Measured current value -- Read Only
Discrete Input
Addresses
Decimal Hex
Data
Type Data Content Parameter
Name
00d 0x00 Bit Relay output state (0=OFF; 1=ON)--
Coil Addresses
Decimal Hex
Data
Type Data Content Parameter
Name
Read/Write
Permission
00d 0x00 Bit Output state (0=no; 1=nc)0typ
Default
Value
Read/Write Permission
0001d 0x0001 word Decimal point of measured current value -- Read Only
Read Only
no
R / W
* MODBUS CONNECTION DIAGRAM
120 Ohm
120 Ohm
Master Slave - 1 Up to 127 slave devices
can be controlled.
Slave - 2
Slave - 127
-
+
A
B
-
+
A
B
-
+
A
B
-
+
A
B
Termination should be accomplished by
attaching 120 Ohm resistors to the start
and at the end of the communication line.
* Applies to devices with Modbus function.
ENDA EPA742-xx-xx-x-RSI INPUT REGISTERS FOR OUTPUT DEVICES
Note 1 : Coil and Discrete input parameters are not available in the devices those have no relay
Note 2 : 0tyP menu parameters can be used as “Holding Register” or “Coil.
Note 3 : Value read in 0th address of input register gives the measured value. Also, the 1st address of the input register specifies the
decimal part of the measured current value.
For example ;
Value read in 0th address of input register is 2842 , if value read in 1st address from input register as 1, it is 284.2
Value read in 0th address of input register is 2842 , if value read in 1st address from input register as 2, it is 28.42
Value read in 0th address of input register is 2842 , if value read in 1st address from input register as 3, it is 2.842
EPA742-EN-05-220103
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
Barbaros Cad. No:18
5/5
ENDA
TM
ENDA EPA742-xx-xx-R-RSI DIGITAL AMPERMETER WITH ALARM RELAY OUTPUT DISCRETE INPUTS
ENDA EPA742-xx-xx-R-RSI DIGITAL AMPERMETER WITH ALARM RELAY OUTPUT COILS INPUTS
Other ENDA Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Precision Digital Corporation
Precision Digital Corporation PD6908 instruction manual

Hitachi
Hitachi UST-676P instruction manual

Honeywell
Honeywell EK205 Quick reference guide

BJE
BJE 010807 Recommended Installation, Maintenance and Inspection Instructions

Sky Control
Sky Control SC8100 Configuration manual

MADDALENA
MADDALENA ElecTo Bulk Instructions for installation, use and maintenance