
Input (Continuation) Resistance to low pressures (vacuum resistance) to 10 mbarabsolute (0.15 psia)
Adjusting the span (turn down) to TD 10:1 (see also this page figure 2)
Zero point increase and Analogue
decrease Smart
PROFIBUS-PA
– ±10% within extended measurement limits
– any within extended measurement limits
– any within extended measurement limits
(see also this page figure 1)
Output Analogue signal 4…20 mA (analogue electronics)
Output signal 4…20 mA
Signal on alarm Signal overrun (>20.5 mA) or underrun (<3.6 mA)
Integration time (damping) Directly on instrument using DIP switches,
switch position "off": 0 s; "on": 2 s
4…20 mA with HART protocol (Smart electronics)
Output signal 4…20 mA with HART protocol
Resolution 1 µA
Signal on alarm optional 3.6 mA, 22 mA or "continue" (last current value will
be held)
Integration time (damping) Directly on instrument using DIP switches,
switch position "off": 0 s; "on": 2 s
or with Universal HART Communicator DXR 275
or using operating program e.g. Commuwin II: 0…40 s
PROFIBUS-PA (PROFIBUS-PA electronics)
Output signal Digital communication signal PROFIBUS-PA (Profile 3.0)
PA function Slave
Transmission rate 31.25 kBit/s
Response time Slave: approx. 20 ms
PLC: 300…600 ms for approx. 30 transmitters
(depending on segment coupler)
Integration time (damping) 0…40 ms via communication
Communication resistance none, PROFIBUS-PA termination-resistor
Physical layer IEC 61158-2
Accuracy Reference conditions DIN IEC 770 TU=25°C (+77°F)
Accuracy data adopted after entering "Low sensor
calibration" and "High sensor calibration" for lower
range-value and upper range-value.
Linearity including hysteresis and reproducibility
(based on the limit point method to DIN IEC 770)
±0.2% of set span***
Linearity at low absolute pressure ranges (due to
performance limits of currently available DKD
calibration rigs)
for ≥40 mbar absolute to <100 mbar absolute:
±0.3 % of set span***
Warm-up time Analogue
Smart
PROFIBUS-PA
– 200 ms
– 1 s
– 1 s
Rise time (T90 time) Analogue
Smart
PROFIBUS-PA
– 60 ms
– 220 ms
– 220 ms
Settling time Analogue
Smart
PROFIBUS-PA
– 180 ms
– 600 ms
– 600 ms
Long-term drift
(with reference to the set span***)
±0.1% per year
±0.25% per 3 years
Thermal effects
– with reference to the set span***
– Applies to transmitters without diaphragm seals or
capillary tubes, see also "Temperature Effects"
page 12.
Analogue electronics:
– for –10°C…+60°C (+14…+140°F): ±(0.3% x TD + 0.3%)
– for –40°C…–10°C (–40…+14°F),
+60°C…+85°C (+140…185°F): ±(0.5% x TD + 0.5%)
Smart and PROFIBUS-PA electronics:
– for –10°C…+60°C (+14…+140°F): ±(0.2% x TD + 0.2%)
– for –40°C…–10°C (–40…+14°F),
+60°C…+85°C (+140…185°F): ±(0.4% x TD + 0.4%)
Temperature coefficient (maximum TK)
– If the temperature coefficient level exceeds the
thermal change, then the thermal change
automatically becomes valid.
– Applies to transmitters without diaphragm seals or
capillary tubes, see also "Temperature Effects"
page 12.
Zero signal and span:
Analogue electronics:
– for –10…+60°C (+14…+140°F): ±0.15% of nominal value/10 K
– for –40…–10°C (–40…+14°F),
+60…+85°C (+140…185°F): ±0.2% of nominal value/10 K
Smart and PROFIBUS-PA electronics:
– for –10…+60°C (+14…+140°F): ±0.08% of nominal value/10 K
– for –40…–10°C (–40…+14°F),
+60…+85°C(+140…185°F): ±0.1% of nominal value/10 K
Vibration effects None (4 mm in path peak-to-peak 5…15 Hz,
2 g: 15…150 Hz, 1 g: 150 Hz…2000 Hz)
–1 0 0.4 +1
nominal value
set span***
measurement limits**
+5%–5%*
extended measurement limits
Figure 1: extended measurement limits
* not with absolute pressure sensors or
with overpressure sensors with lower
measurement limit of –1 bar
** measurement limits, see page 15
Examples:
0…4 bar overpressure sensor:
extended measurement limits:
–0.2…4.2 bar
0…10 bar absolute pressure sensor:
extended measurement limits:
0…10.5 bar abs
Figure 2: Turn down
Explanation of terms:
Turn down (TD) =
nominal value / set span***
Example:
nominal value = 1 bar
set span*** = 0.4 bar
TD = 1 : 0.4
*** Calibrated span for instruments
with PROFIBUS-PA electronics
16