Fieldpiece SMAN440 User manual

01
06
02
07
03
08
04
09
05
10
Outdoor Dry Bulb
Indoor Wet Bulb
Battery Life
Auto Power O Enabled
Wireless Signal Lost
Selected Refrigerant of System
Static Target Superheat Calculation
Connection Strength to Analyzer
IDWB Connection Strength
ODDB Connection Strength
Quick Start
1 Install six AA batteries into rear battery com-
partment. Batteries included in packaging.
2 Press the center blue button for 1 second to
turn on your new manifold.
3 Connect hoses and pipe clamps to the mani-
fold and the system.
4 See real-time pressureand temperaturemea-
surements all at once!
Certifications
FCC ID: VEARF915A
C-Tick (N22675)
CE
RoHS Compliant
Description
Your SMAN440 is a top of the line Wireless 4
Port Manifold for HVACR professionals. See all
your pressures and temperatures at the same
time on the redesigned large display with bright
blue backlight.
SMAN440 combines high precision, absolute
pressure sensors, a superheat/subcooling
calculator and dual temperature measurements.
Your SMAN440 calculates and displays target
superheat and actual superheat to verify proper
charge. SMAN440’s large 3/8”VAC port and true
3/8”bore throughout the block allow for quicker
recovery and evacuations.
Your SMAN440 is designed to meet the
demands of HVACR eld service with a rugged
rubber boot for durability, a strong metal
hanger for easy storage and a form tting, water
resistant, padded nylon pouch.
Use additional wireless products like
SDP2 Dual In-Duct Psychrometer to receive
temperature measurements wirelessly for real-
time target superheat calculations.
Pressure (bar)
Pressure (pounds/in2)
Pressure (kilopascals or Megapascals)
Negative Pressure (inches of mercury)
Negative Pressure (cm of mercury)
Superheat
Subcooling
Target Superheat
Suction Line Temperature
Liquid Line Temperature
Vapor Saturation Temperature
Liquid Saturation Temperature
Fieldpiece
4 Port Wireless
Manifold
OPERATOR'S MANUAL
Model SMAN440
Controls
1 Insert Type K thermocouple plugs here.
2 Temperature calibration pots.
3 Press to zero atmospheric pressure.
4 Press to calibrate to refrigerant tank.
(See Advanced Pressure Calibration section.)
5 Press to conrm selection.
6 Press/hold to cycle through refrigerants.
7 Hold 1 second to enable wireless functionality.
8 Press to scroll up to adjust values.
9 Press to change units.
10 Press to scroll down to adjust values.
11 Hold 1 second to power on or o. Press to toggle backlight.
12 Press to enterTarget Superheat setup mode.
13 Turn clockwise to close High side port.
14 Turn clockwise to close Low side port.
15 Turn clockwise to close 3/8" Vacuum port.
16 Turn clockwise to close Refrigerant port.
Functions
Superheat and Subcooling
Your SMAN440 can calculate and display both superheat and
subcooling simultaneously.
1 Select the appropriate refrigerant using the REFRIGERANT button.
2 Connect EPA approved refrigerant hoses to low and high side on
SMAN440. Plug Type K thermocouple pipe clamps to SLT and LLT.
3 Connect your SMAN440 to the system:
Superheat: Hand tighten low side hose to suction line service
port. Place the SLT pipe clamp thermocouple on the suction line
between the evaporator and compressor, no closer than 6 inches
to compressor.
Subcooling: Hand tighten high side hose to liquid line service
port. Attach the LLT pipe clamp thermocouple on the liquid line
between the condenser and expansion valve (TXV), as close to the
service port as possible.
4 After turning the system on or making any adjustments to
the system wait 15 minutes before charging by superheat or
subcooling to ensure the system is stabilized.
5To add or remove refrigerant connect a refrigerant or recovery
cylinder to the REF port on SMAN440. Follow recommneded
charging or recovery practices from equipment manufacturer. Use
the low side, high side, and REF valves on SMAN440 to charge or
recover refrigerant as needed. Let system stabilize for 15 minutes.
Note: When superheat and/or subcooling cannot be calculated an
"OL" or "-OL" will be displayed. Please check the following:
1 The correct refrigerant is selected on the SMAN.
2 The pipe thermocouples are plugged into SLT/LLT ports and are in
good working condition.
3 The pipe thermocouples are attached in the appropriate location on
the system. See step 3 above for details.
Target Superheat
Target Superheat is useful for charging xed orice airconditioning
systems. Your SMAN440 can receive real-time indoor wet bulb (IDWB)
and outdoor dry bulb (ODDB) temperatures wirelessly to calculate real-
time target superheat.The IDWB and ODDB values can also be entered
both manually if measurements are taken by other instruments, or a
combination of one wireless measurement with one manually inputted
measurement.
Receive IDWB and ODDB temperatures wirelessly
1 PressTarget SH button to enterTarget SH setup mode. SMAN will
search for last connected wireless instrument and automatically
connect if found. Press ENTER to end search and sync to a new
wireless instrument. IDWB will blink indicating it is ready for an
input.
2 Hold SYNC until beep is heard to search for a Fieldpiece wireless
instrument.
3 Set your Fieldpiece wireless instrument to connect with the
SMAN440. Refer to your Fieldpiece wireless instrument's manual
for more information.
4 Once SMAN440 is connected with Fieldpiece wireless instrument,
the real-time wet bulb temperature measurement will display.
5 Ready the Fieldpiece wireless instrument to measure indoor wet
bulb temperature and place the probe at the return side of the
evaporator between the lter and the coil.
6 On your SMAN440 use the ARROWS to setup ODDB.When ready,
ODDB will be blinking.
7 Repeat steps 2 -5 for outdoor dry bulb measurements (ODDB). For
outdoor dry bulb measurement, place the dry bulb temperature
probe onto the side of the condenser. For accurate results, keep it
shaded from direct sunlight.
8 Target superheat (TSH) is calculated and updated in real-time and
located in the center column of display.
Wireless Notes
1For real-time target superheat calculations, both IDWB and ODDB
measurements must be received wirelessly.
2 A blinking HOLD will appear to the left of the TSH calculation
when one measurement is wirelessly received and the other
manually input indicating a staticTSH calculation.
3 If the indoor wet bulb or outdoor dry bulb measurements result in
an uncalculable TSH, "OL" or "-OL" will display.
4 When the IDWB or ODDB is wirelessly connected with SMAN440,
and the connection is lost,“nSG” (No Signal) will display.
The SMAN440 will try to reconnect with Fieldpiece wireless
instrument for 2 minutes while the unit is powered ON. During
these 2 minutes IDWB or ODDB can be connected to a dierent
Fieldpiece wireless instrument but manual entry will be disabled.
Manually Input Temperatures
1 PressTarget SH button to enterTarget SH mode. Press ENTER to
end wireless search. IDWB will blink indicating it is ready for an
input.
2 Press UP or DOWN ARROW to toggle between IDWB or ODDB
input. Hold ENTER to select which temperature you want to input,
either IDWB or ODDB. The far left digit of IDWB or ODDB will begin
blinking indicating manual input mode is ready.
3 Press the ARROWS to change values and press ENTER to lock in
each digit.
4 Repeat steps 2 and 3.The calculated target superheat will show in
the center column of the display. A solid HOLD will display to the
left of the TSH calculation indicating a staticTSH calculation.
Note: If the inputted temperature is out of the calculable range for IDWB
or ODDB an "Err" will ash once and a double beep will sound. IDWB
range (40°F to 125°F, 4.4°C to 51.7°C) and ODDB range (50°F to
140°F, 10°C to 60°C). Re-input a temperature within these ranges to
calculate target superheat.
Changing Units
Your SMAN can display pressure and temperature measurements
in English, Metric or combination of both units.
1 Press UNITS to enter unit selection screen.
2 Use ARROW to select your desired pressure units. Press ENTER.
3 Use ARROW to select your desired temperature units. Press ENTER
to return to pressure units.
4 Press UNITS to return to normal SMAN display.
14
16
13
15
11
22
3 4
5
89
11
10
12
67
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

11
16
12
17
13
18
14
19
15
20
Wireless System Analyzers
Your SMAN440 can connect with the Fieldpiece HVAC Guide® HG3
for deeper diagnosis, data storage and reporting. Send all your line
pressures and temperatures, vapor and liquid saturation temperatures,
superheat and subcooling calculations and micron readings wirelessly
to the Fieldpiece HVAC Guide® HG3.
1 Connect all hoses and pipe clamp thermocouples to the
SMAN440 and the HVACR unit to be tested.
2 From main display, hold SYNC for 1 second until a beep is heard.
SMAN440 will begin searching for a FieldpieceWireless System
Analyzer.
3 Set your Fieldpiece HVAC Guide® HG3 to connect with the
SMAN440. Refer to your Fieldpiece HVAC Guide® HG3 manual for
more information on how to connect wirelessly.
Wireless Notes
1 The SMAN460 and Fieldpiece HVAC Guide® HG3 must be between
1 to 10 feet for initial sync to occur.
2 Be sure the Fieldpiece HVAC Guide® HG3 has been updated to the
latest rmware version located on our website.
Firmware Updates
Your SMAN440 rmware can be updated in the eld to ensure you
always have the most up-to-date features for your manifold. Just go
to www.eldpiece.com to periodically check for the latest rmware
version. If a newer version is available, follow the download link and
installation instructions on the website. Connect your SMAN440 to the
PC via a mini-USB to USB cable (not included) to install the update on
your SMAN.
To check your current rmware version, power o your SMAN440.
Pressand hold the blue power button for about6 seconds.TheSMAN440
rmware version will show in the top right corner of the display (X.XX).
Specifications
Mini-USB port: For updating to newer versions of rmware
Display size: 5 inches (diagonal)
Backlight: Blue (On for 3 minute unless turned o manually)
Battery: 6 x AA (Battery life below based on alkaline type)
Battery life: 350 hours (without backlight and wireless)
Low battery indication: is displayed when the battery
voltage drops below the operating level
Auto shut o: 30 minutes of inactivity when APO is activated
Over range: "OL" or "-OL" is displayed
Operating environment: 32°F to 122°F (0°C to 50°C) at <75% RH
Temperature coecient: 0.1 x (specied accuracy) per °C (0°C to
18°C, 28°C to 50°C), per 0.6°F (32°F to 64°F, 82°F to 122°F
Storage temperature: -4°F to 140°F (-20°C to 60°C), 0 to 80% RH
(with battery removed)
Weight: 4.03 lbs (1.83 kg)
Pressure
Sensor type: Absolute pressure sensors
Connector type: Standard 1/4" and 3/8" NPT male are tting
Range: 29”HgV to 580psig (English), 74 cmHgV to 0 to 4000KPa
(Metric), 4.000MPa (Metric), and 40.00bar (Metric)
Resolution: 0.1 psi/inHg; 1 kPa/cmHg; 0.001MPa; 0.01bar
Accuracy: 29”HgV to 0”HgV: ±0.2”HgV
74 cmHgV to 0 cmHgV: ±1 cmHgV
0 to 200 Psig: ±1 Psi; 0 to 1378 KPa: ±7 KPa; 0 to 1.378MPa:
±0.007MPa; 0 to 13.78 bar ±0.07 bar
200 to 580 Psig: ±(0.3% of reading+1 Psig); 1378 to 4000KPa: ±(0.3% of
reading+7 Kpa); 1.378 to 4.000MPa: ±(0.3% of reading+0.007MPa);
13.78 to 40.00bar: ±(0.3% of reading +0.07bar)
Maximum overload pressure: 800 psig
Units: Psig, kPa, MPa, bar, inHg, and cmHg
Temperature
Sensor type: Type K thermocouple
Range: -95°F to 999.9°F (-70°C to 537.0°C)
Resolution: 0.1°F/°C
Accuracy: ±(1.0°F) -95°F to 199.9°F; ±(0.5°C) -70°C to 93°C
±(2.0°F) 200°F to 999.9°F; ±(1.0°C) 93°C to 537.0°C
Note: All accuracies are after a eld calibration.
Wireless
Wireless range: 1 to 100 feet (30m)
Initial sync setup range: 1 to 10 feet
Wirelessly transmit to Fieldpiece HVAC Guide® HG3 System Analyzer
Wirelessly receive indoor wet bulb (IDWB) and outdoor dry bulb
(ODDB) from Fieldpiece transmitters
Refrigerants
The P-Tcharts of the following 45 refrigerants come pre-programmed
into your SMAN. In your SMAN the refrigerants are listed in order of most
commonlyused. Here,they are listedinnumerical orderfor easy reference.
R11, R113, R114, R12, R123, R1234YF, R124, R125, R13, R134A, R22,
R23, R32, R401A(MP39), R401B, R402A, R402B, R404A, R406A, R407A,
R407C, R407F, R408A, R409A, R410A, R414B (Hotshot), R416A, R417A,
R417C (HOT SHOT 2), R420A, R421A, R421B, R422A, R422B(NU22B),
R422C(Oneshot), R422D, R424A, R427A, R434A(RS-45), R438A(MO99),
R500, R502, R503, R507A, R508B (Suva95)
Using Dierent Refrigerants
You can use your manifold with dierent refrigerants. Be sure to
purge your manifold and hoses before connecting to a system with a
dierent refrigerant.
Maintenance
Clean the exterior with a dry cloth. Do not
use liquid.
Battery Replacement
The battery must be replaced when the
battery life indicator is empty. SMAN will
display "lo batt" and power o. Remove rear
battery cover and replace with 6 AA batteries.
Auto Power O (APO)
To conserve battery life, your SMAN will
power down after 30 minutes of inactivity. APO
is activated by default and APO displays above
the battery icon. To deactivate, press and hold
ENTER while powering on the SMAN. When
deactivated, APO will no longer show above
the battery icon.
Recommended
Accessory
Easy View Micron Gauge
Model SVG3
Measure the depth of the vacuum in microns
of mercury to verify all moisture and non-
condensibles have been removed from the
system. For the most accurate reading, connect
directly to a service port on the system or to a
Schrader core removal tool (illustrated below).
Calibration
Temperature
To calibrate your SMAN440 temperature
thermocouples, adjust the pot on the front
of the meter labeled SLT Cal or LLT Cal. The
best way to calibrate is to match to a known
temperature. Ice water is very close to 32°F and
is readily available. Accuracies of one degree or
better are obtained through eld calibration.
1 Stabilize a large cup of ice water by stirring. Pure, distilled water
will be the most accurate.
2 Immerse the temp probe in ice water from SLT and adjust the
SLT Cal pot with a at head screwdriver and let it stabilize, keep
stirring.
3 Repeat Step 2 for temp probe in LLT.
Pressure Zeroing
To calibrate your SMAN440 pressure sensors
to atmospheric pressure, ensure that your
SMAN440 is disconnected from any pressure
source and at equilibrium with the ambient
pressure.
1 Press the CAL Atmospheric Pressure button and your SMAN440
will set the zero point of pressure to the ambient pressure.
Advanced Pressure Calibration
Your SMAN440 has the ability to perform a
linear adjustment of the pressure sensors based
on refrigerant type, temperature, and pressure.
Calibration setup: For best results, rst perform both the
Temperature and Pressure Zeroing procedures. See Calibration
section for details. This will ensure pressure readings are zeroed
and thermocouple is properly calibrated to the SLT port of the
SMAN. Calibration to LLT port is not necessary for this calibration.
The refrigerant cylinder should be stored in a stable ambient
environment for at least 24 hours before calibration.
1 Plug in a Type K thermocouple into SLT. (A bead type
thermocouple, like the ATB1, is recommended.)
2 Connect the SMAN440 to a refrigerant cylinder of a known,
single refrigerant using an EPA approved service hose. Be sure
to open both HIGH and LOW side valves on your manifold and
cap the unused ports. (If caps are not available you can connect
both ends of a refrigerant hose to the two unused caps. Note:
Some refrigerant will remain in the hoses which will need to be
recovered.)
3 Press the REFRIGERANT button to match the refrigerant of the
cylinder you are using.
4 Attach bead-type thermocouple to the side of the cylinder using
tape. It is recommended to attach in the middle of the cylinder.
Important: Let the temperature of the thermocouple stabilize to
the refrigerant temperature for 1 to 2 minutes or until stable.
5 Open the refrigerant cylinder. The pressure inside cylinder should
now be displayed on both HIGH and LOW side pressure sensors.
6 Press the CALTest Pressure button. If successful, "Good" will
display for 3 seconds. If failed, "Err" will display for same time.
Your SMAN checks with its built-in P-T charts to compare the
temperature of the refrigerant in the tank to the vapor saturation
temperature based on the refrigerant you selected. If the measured
pressures on your SMAN are within ±3psi of the P-T chart pressure
corresponding to the vapor saturation temperature, the SMAN will
adjust the pressure sensor linearity to match the P-T chart.
Possible causes of failed "Err" pressure calibration:
1. Refrigerant tank was not stored in stable ambient conditions for at
least 24 hours.
2. Thermocouple attached to refrigerant tank was not properly
calibrated to SLT port of SMAN.
3. Thermocouple was plugged into wrong port LLT instead of SLT.
4. Incorrect refrigerant was selected on the SMAN.
FCC Compliance and Advisory
Statement
This device complies with Part 15 of the FCC rules. Operation is
subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.
This equipment has beentested and found to comply with the limits
for a Class B digital device, according to Part 15 of the FCC rules.These
limits are designed to provide reasonable protection against harmful
interference in aresidential installation.This equipment generates,uses
and can radiate radio frequency energy and if not installed and used
in accordance with the instructions, may cause harmful interference to
radio communications. However, there is noguarantee that interference
will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be
determined byturning the equipment o and on,the user is encouraged
to try correct the interference by one or more of the following measures:
1.Reorient the receiving antenna.
2.Increase the separation between the equipment and receiver.
3.Connect the equipment into an outlet on a circuit dierent from
that to which the receiver is connected.
4.Consult the dealer or an experienced radio/TV technician for help.
Shielded interface cables must be used in order to comply with
emission limits.
FCC Caution: To assure continued compliance, any changes or
modications not expressly approved by the party responsible
for compliance could void the user's authority to operate this
equipment.
Limited Warranty
This meter is warranted against defects in material or workmanship
for one year fromdate of purchase froman authorized Fieldpiecedealer.
Fieldpiece will replace or repair the defective unit, at its option, subject
to verication of the defect.
This warranty does not apply to defects resulting from abuse,
neglect, accident, unauthorized repair, alteration, or unreasonable
use of the instrument.
Any implied warranties arisingfrom the sale of a Fieldpiece product,
including but not limited to implied warranties of merchantability and
tness for a particular purpose, are limited to the above. Fieldpiece
shall not be liable for loss of use of the instrument or other incidental
or consequential damages, expenses, or economic loss, or for any claim
of such damage, expenses, or economic loss.
State laws vary.The above limitations or exclusions may not apply
to you.
Obtaining Service
Email Fieldpiece warranty department at fpwarranty@eldpiece.
com for current xed price repair service. Send check or money order
made out to Fieldpiece Instruments for the amount quoted. If your
meter isunder warranty therewill be no cost forthe repair/replacement.
Send your meter, freight prepaid, to Fieldpiece Instruments. Send proof
of date and location of purchase for in-warranty service.The meter will
be repaired or replaced, at the option of Fieldpiece, and returned via
least cost transportation.
For international customers, warranty for products purchased
outside of the U.S. should be handled through local distributors. Visit
our website to nd your local distributor.
© Fieldpiece Instruments, Inc 2014; v08
!
WARNINGS
DO NOT APPLY MORE THAN 800 PSI TO ANY PORT ONTHE MANIFOLD.
FOLLOW ALL EQUIPMENT MANUFACTURER'S TESTING PROCEDURES
ABOVETHOSE IN THIS MANUAL IN REGARDSTO PROPERLY
SERVICING THEIR EQUIPMENT.
Directly to
Service Port
Schrader Core
Removal Tool
Hose
To
Pump
MICRONS OF MERCURY
SVG3
VACUUM
GAUGE
MIN
ALARM
ENTER
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Other Fieldpiece Test Equipment manuals
Popular Test Equipment manuals by other brands

ELPRO
ELPRO EL-ERTK-A2 Operator's manual

PCE Health and Fitness
PCE Health and Fitness PCE-2550 user manual

Huazheng Electric Manufacturing
Huazheng Electric Manufacturing HZ-9003 user manual

insize
insize ISH-MR150 Operation manual

Tektronix
Tektronix DPO4104 instructions

Fluke
Fluke 525A Getting started guide