manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Fluke
  6. •
  7. Measuring Instrument
  8. •
  9. Fluke ASM-990 Series Instruction Manual

Fluke ASM-990 Series Instruction Manual

Other Fluke Measuring Instrument manuals

Fluke 773-II User manual

Fluke

Fluke 773-II User manual

Fluke 433 Configuration guide

Fluke

Fluke 433 Configuration guide

Fluke CableAnalyzer DSP-4000 Series User manual

Fluke

Fluke CableAnalyzer DSP-4000 Series User manual

Fluke NORMA 6003 User manual

Fluke

Fluke NORMA 6003 User manual

Fluke 9040 User manual

Fluke

Fluke 9040 User manual

Fluke 700G01 User manual

Fluke

Fluke 700G01 User manual

Fluke OptiView User manual

Fluke

Fluke OptiView User manual

Fluke VR1710 User manual

Fluke

Fluke VR1710 User manual

Fluke IntelliTone Pro 200 LAN User manual

Fluke

Fluke IntelliTone Pro 200 LAN User manual

Fluke PM6681 User manual

Fluke

Fluke PM6681 User manual

Fluke 80i-410 User manual

Fluke

Fluke 80i-410 User manual

Fluke LS-1310 Manual

Fluke

Fluke LS-1310 Manual

Fluke 945 User manual

Fluke

Fluke 945 User manual

Fluke 85RF User manual

Fluke

Fluke 85RF User manual

Fluke 902 Use and care manual

Fluke

Fluke 902 Use and care manual

Fluke 481 User manual

Fluke

Fluke 481 User manual

Fluke PRUFTECHNIK VIB 6.172 Programming manual

Fluke

Fluke PRUFTECHNIK VIB 6.172 Programming manual

Fluke DTX-CLT CertiFiber User manual

Fluke

Fluke DTX-CLT CertiFiber User manual

Fluke ProcessMeter 787 User manual

Fluke

Fluke ProcessMeter 787 User manual

Fluke Micro-Bath 7102 User manual

Fluke

Fluke Micro-Bath 7102 User manual

Fluke FS17XX User manual

Fluke

Fluke FS17XX User manual

Fluke E-DWT-H User manual

Fluke

Fluke E-DWT-H User manual

Fluke 312 User manual

Fluke

Fluke 312 User manual

Fluke 712 Use and care manual

Fluke

Fluke 712 Use and care manual

Popular Measuring Instrument manuals by other brands

Suaoki D40 user manual

Suaoki

Suaoki D40 user manual

Elster BK-G1.6 operating instructions

Elster

Elster BK-G1.6 operating instructions

Endress+Hauser Deltabar FMD71 technical information

Endress+Hauser

Endress+Hauser Deltabar FMD71 technical information

Duratool D03128 manual

Duratool

Duratool D03128 manual

Vega VEGAPULS 65 Quick setup guide

Vega

Vega VEGAPULS 65 Quick setup guide

Tonghui TH2618B Operation manual

Tonghui

Tonghui TH2618B Operation manual

Tenmars TM-801 user manual

Tenmars

Tenmars TM-801 user manual

MSR 145 user manual

MSR

MSR 145 user manual

TPS WP-82 manual

TPS

TPS WP-82 manual

PCB Piezotronics 003C30 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 003C30 Installation and operating manual

Tenma 72-10465 manual

Tenma

Tenma 72-10465 manual

Endress+Hauser Proline Promag 50 Brief operating instructions

Endress+Hauser

Endress+Hauser Proline Promag 50 Brief operating instructions

Siemens SITRANS F Coriolis FCT030 Function manual

Siemens

Siemens SITRANS F Coriolis FCT030 Function manual

KLINGER CMF V Series instruction manual

KLINGER

KLINGER CMF V Series instruction manual

Keysight M8290A Getting started guide

Keysight

Keysight M8290A Getting started guide

Acterna MS1400 Operation manual

Acterna

Acterna MS1400 Operation manual

R&S ZPH manual

R&S

R&S ZPH manual

Datacolor ColorReader DC10-2 user guide

Datacolor

Datacolor ColorReader DC10-2 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

ASM-990 Series: Operating Guidelines
Quick Reference Guide
NOTE: Consult with your hospital’s radiation safety officer or trauma/emergency coordinator for detailed instructions for evaluating,
triaging and treating patients with suspected radiation exposure.
Operating Instructions
PN 990QR January 2007
© 2007 Fluke Corporation. All rights reserved. Printed in USA.
All product names are trademarks of their respective companies.
AUDIO LIGHT
ESC SEL
START/STOP
RST/SAVE
ON/OFF
1. POWER UNIT ON with “ON/OFF” button
a) A screen displaying “Victoreen Advanced Survey Meter” will
appear for ~5 seconds
b) If this does not occur, refer to BACK PAGE (“Power/Batteries”)
PROBE INFO
MODEL #: PANCAKE
SERIAL #: 002132
CAL DATE:
CAL CYCLE: 12 MOS.
CAL DUE:
IS PROBE CORRECT?
YES
NO
2. VERIFY PROBE MODEL # is Pancake
Select “YES” with the arrows and press the “SEL” button
a) If “MODEL #” is not Pancake refer to BACK PAGE (“Probe
Selection”)
b) If “CAL DUE” date has passed, contact radiation safety officer
3. ENSURE RATE MODE is in USE (mR/hr)
Once the probe is selected, the normal operating screen is
displayed
a) Only use the “RATE” mode with “mR/hr” units (Note: may
also read “R/hr” for elevated radiation levels as 1R = 1,000
mR)
b) Verify that the “RATE”mode is in use (denoted by the Rate
Mode Indicator and “mR/hr”)
c) If this is not the case, refer to BACK PAGE (“Mode Selection”)
4. SCAN PATIENT WITH UNIT (as shown)
a) Hold UNIT an inch from patient, with the
BOTTOM of the UNIT facing the patient
b) Survey entire body until a MAXIMUM RATE
reading is identified by visual inspection
c) Record MAXIMUM RATE and Time/Date
If using Patient Tracking Tags, refer to
BACK PAGE for “Logging Data / Barcoding”
Note: refer to ASM-990 Series Operators Manual for detailed use and care instructions
Triage Guidelines
Maximum RATE
reading (mR/hr) Action(s)***
< 0.1 mR/hr No action required
0.1 – 10 mR/hr Contact radiation safety officer (RSO) or trauma
coordinator (TC)
10 mR/hr – 1 R/hr Contact RSO or TC
Limit exposure to patient
> 1 R/hr or
OVERRANGE
Contact RSO or TC
Limit exposure to patient
Maintain safe distance from patient, when possible
(greater than 2 feet)
*** NOTE: if Maximum RATE is > 0.1 mR/hr, ensure that the elevated RATE
level is not a result of a recent medical procedure
Tips for Reducing Radiation Exposure
•TIME – minimize time near radiation source
•DISTANCE – maintain maximum practical distance from radiation source
•Moving 2x the distance away from radiation source will decrease exposure
rate by a factor of 4
•SHIELDING – place patient behind concrete or lead barriers if possible
Note
If this device is used in a manner not specified by Fluke Biomedical,
the protection by the device may be impaired.
Caution
There are no user servicable parts for this unit. Should unit
require service, please contact Fluke Biomedical at
440.248.9300.
ASM-990 Series: Detailed Operating Guidelines
Quick Reference Guide
NOTE: Consult with your hospital’s radiation safety officer or trauma/emergency coordinator for detailed instructions for evaluating,
triaging and treating patients with suspected radiation exposure.
Category Description
Keypad •“AUDIO” key is used to enable/disable the audible alarm/speaker. An icon, resembling a small speaker, will
appear in the lower left corner of the LCD
•“LIGHT” key will illuminate the LCD’s backlight for a time period (user menu selected)
•“START/STOP/RST/SAVE” key is used to manually log the current reading and the physical location, for each
measurement, in barcode format
•“SEL” key, when pressed, will activate the menu system
•“ESC” key will return the user to the normal operating screen
Power/Batteries •Operates on two “D” batteries located in the unit’s handle; recommended batteries include nickel metal
hydride (NiMH), rechargeable, or alkaline
Probe Selection •If “Pancake” is not listed as the “Model #”, highlight “NO,” then press “SEL,” to display a list of additional
probes that can be selected
•On the subsequent “Probe Selection” screen, highlight “Pancake” then press “SEL”
Mode Selection •If “RATE” mode is not in use, press “SEL” to change to the “MAIN MENU” screen
•Highlight “MODES” on the “MAIN MENU” then press “SEL”
•Highlight “RATE” on the “MODES SUB-MENU” screen then press “SEL”
•Highlight “mR/hr” on the “RATE UNIT SUB-MENU” screen then press “SEL”
•This series of functions will then return you to the normal operating screen
Logging Data /
Barcoding
•“START/STOP/RST/SAVE” key is used for manually logging data / reading a barcode
•When this key is pressed, the instrument then displays the message “SPACE AVAILABLE XX” for one
second, while the current reading is being logged. The display will also show “SPACE AVAILABLE ###”
between each data point that is logged
•When the memory is full, the instrument will display “LOG BUFFER FULL.”
Data Transfer /
Upload
•Saved data points can be transferred to a PC via the IrDA port using PC based software
•Select “IrDA COMMUNICATION” from the “MAIN MENU;” at the “ENABLE IrDA PORT?” prompt, select
“YES.” Then physically place the unit within six inches of the IrDA data port on the host PC. Initialize the PC
based software.
Note
If this device is used in a manner not
specified by Fluke Biomedical, the
protection by the device may be impaired.
Caution
There are no user servicable parts for
this unit. Should unit require service,
please contact Fluke Biomedical at
440.248.9300.
Bargraph
Clock
Rate Mode
Indicator
Keypad Locked
Buffer Full
Audio
Indicator
Low Battery
Indicator
Reading
Scale Multiplier
Routine
Cleaning
•Do not immerse the unit – liquid could damage the circuits. The unit should be kept clean and free from dirt
and contamination; The unit may be cleaned by wiping with a damp cloth using any commercially available
cleaning or decontaminating agent
Note: refer to ASM-990 Series Operators Manual for detailed use and care instructions