manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Fluke
  6. •
  7. Measuring Instrument
  8. •
  9. Fluke E-DWT-H User manual

Fluke E-DWT-H User manual

Other Fluke Measuring Instrument manuals

Fluke i3000s Flex Series User manual

Fluke

Fluke i3000s Flex Series User manual

Fluke 945 User manual

Fluke

Fluke 945 User manual

Fluke 845A User manual

Fluke

Fluke 845A User manual

Fluke 1520 Use and care manual

Fluke

Fluke 1520 Use and care manual

Fluke 1550B User manual

Fluke

Fluke 1550B User manual

Fluke FiberInspector FT500 User manual

Fluke

Fluke FiberInspector FT500 User manual

Fluke FIBER OneShot User manual

Fluke

Fluke FIBER OneShot User manual

Fluke 381 User manual

Fluke

Fluke 381 User manual

Fluke VR1710 User manual

Fluke

Fluke VR1710 User manual

Fluke 3210-PR User manual

Fluke

Fluke 3210-PR User manual

Fluke VIBGUARD IIoT Guide

Fluke

Fluke VIBGUARD IIoT Guide

Fluke PRV240 User manual

Fluke

Fluke PRV240 User manual

Fluke i430-Flexi-TF-II User manual

Fluke

Fluke i430-Flexi-TF-II User manual

Fluke SimpliFiber User manual

Fluke

Fluke SimpliFiber User manual

Fluke iFlex i2500-18 Reference guide

Fluke

Fluke iFlex i2500-18 Reference guide

Fluke ScopeMeter 125 User manual

Fluke

Fluke ScopeMeter 125 User manual

Fluke 925 User manual

Fluke

Fluke 925 User manual

Fluke 1630-2 Use and care manual

Fluke

Fluke 1630-2 Use and care manual

Fluke 789 Use and care manual

Fluke

Fluke 789 Use and care manual

Fluke 411D User manual

Fluke

Fluke 411D User manual

Fluke 1630 Use and care manual

Fluke

Fluke 1630 Use and care manual

Fluke SureGrip 80PK-25 User manual

Fluke

Fluke SureGrip 80PK-25 User manual

Fluke i5s User manual

Fluke

Fluke i5s User manual

Fluke T5-600 Operating and installation instructions

Fluke

Fluke T5-600 Operating and installation instructions

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

IRIMO 984-6D-1 owner's manual

IRIMO

IRIMO 984-6D-1 owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

E-DWT-H User Guide
TEST port (DH500F)
Reservoir cap
Priming pump plunger
Reservoir cap vent valve
Reference pressure monitor (RPM4-E-DWT)
Test shut off valve and fine pressure adjustment
‘Lo Q-RPT shut off’ valve position prompt LED
Lo Q-RPT shut off valve
‘Lo Q-RPT active’ valve position prompt LED
‘Lo Q-RPT active’ caution indicator
Reservoir shut off valve
Variable volume (screw press)
Variable volume piston position indicator
Figure 1: E-DWT-H with Hi and Lo Q-RPT
Introduction
This demo guide is to assist customers and
representatives to perform an effective
demonstration or evaluation of E-DWT-H. It is
written for the more complex case of a dual Q-RPT
E-DWT-H model.
Preparation
Plug in the E-DWT-H to power up.
Attach the four variable volume handles.
Open the reservoir cap (vent port) by turning
counter-clockwise.
Open the reservoir shutoff valve.
Back out the variable volume by turning
counter-clockwise. Back it out all the way, then
turn back in (clockwise) ½ turn.
Set the Lo Q-RPT shutoff valve to the Hi
Q-RPT (clockwise).
Set the RPM4 to the Hi Q-RPT using the
[7/Range] key.
Open the test shutoff valve by turning counter-
clockwise until it is fully backed out. Turn the
handle back in (clockwise) 4 turns to bring the
valve close to mid-stroke.
DUT Connection and Priming
Remove the shipping plug from the test port on
top of the E-DWT-H enclosure.
Connect the gauge to be tested loosely into the
test port. One point in the DUT connection
should remain loose enough to bleed some test
fluid out through it. If there are multiple
adaptors used to connect the DUT, leave the
loose fitting at the highest point possible. If the
DUT is a gauge with a built-in bleed port, open
this port.
Pull up on the reservoir plunger.
Push down the reservoir plunger until clear
fluid, with no air bubbles, bleeds out the loose
DUT connection. For a system with a large
volume of air, you may need to cycle the
plunger more than once to start the fluid
flowing out the open port.
Tighten the loose DUT connection.
Push the reservoir plunger down until 30 to 50
psig is indicated on the RPM4-E-DWT display,
or until the bottom of the plunger stroke is
reached. While holding the 30 to 50 psig
pressure with the plunger, close the reservoir
shutoff valve to hold pressure in the system.
Warning
Do not generate more than 100 psig
pressure using the reservoir
priming pump plunger. The
reservoir hardware and seals are
not designed for greater pressures
and may be damaged.
Running an E-DWT-H Quick Test
Select [ENT] on the RPM4-E-DWT to begin
Quick Test setup.
Note:
The RPM4 must be in a display
mode , such as the Rate display,
that does not use the [ENT] key for
other display functions.
Choose Quick
Select units, ie. psi.
Select DUT FS range, ie. 10,000.
Enter DUT Tolerance.
Note:
As an example, the DUT tolerance
may be ±0.25%, but you may want to
enter 0.1% for the purposes of the
demonstration. It will be more likely
you’ll see an out of tolerance data
point and see E-DWT’s reaction.
Select test increments, ie. 20%
Select Up and Down points, if desired.
Confirm the test selections by pressure [ENT].
Confirm the unit’s setting of an AutoRange.
Note:
When an AutoRange is selected on
RPM4, the unit automatically sets it’s
pressure upper limit (for alarms),
stability test pass/fail criteria,
pressure display resolution and
Q-RPT range selection based on the
DUT range and tolerance entered by
the operator.
If a zero point is required for your test, open the
reservoir shutoff valve and confirm that the test
shutoff valve is open. This will produce zero
gauge pressure in the test system. After
recording this point, you should repeat step 6 of
DUT Connection and Priming to prime the
system again. Be aware that even when the
reservoir shutoff and test shutoff valves are
open, there may be a difference in pressure
between the DUT and the E-DWT due to fluid
head. For more precise instruments with lower
ranges, you may need to zero the DUT before
making final test connections.
AutoTest will guide you to set each pressure that
was defined in Running an E-DWT-H Quick Test.
Note.
Use the variable volume to control
pressures. For more precise control,
use the test shutoff/fine pressure
adjust valve. When testing DUTs,
especially analog gauges, set test
pressures according to the indication
on the DUT, then read (or let the
RPM4-E-DWT’s software read and
record) the reference pressure value
for you. When running the AutoTest,
you will not be prompted to enter
DUT indications because the DUT is
expected to be set to the nominal test
point.
Fluke. Keeping your world up and running.®
Fluke Corporation
DH Instruments Division
4765 East Beautiful Lane
Phoenix, AZ 85044-5318 USA
Ph: 602.431.9100
Fax: 602.431.9559
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries: +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.dhinstruments.com
©2010 Fluke Corporation. Specifications subject to change
without notice. E-DWT-H and E-DWT are registered trademarks of
Fluke Corporation, DH Instruments Division.
Printed in U.S.A. 5/2010 3469593A D-EN-N
Modification of this document is not permitted
without written permission from Fluke Corporation.
Once you have set a test pressure,
E-DWT-H will automatically test for stability,
and indicate to the operator when it is safe to
take data. Press [ENT] to collect a data point.
Hint:
Try collecting a data point while the
pressure is still unstable and you
have not yet seen the stability light
illuminate green – to see and prove
the unit’s reaction to a technician
trying to ignore or override the
stability test.
Note:
The stability ready/not ready limit is
equal to 10% of the DUT tolerance.
To collect test data, simply select [ENT].
E-DWT-H will either respond with a tone to
indicate that the DUT measurement at that point
is in tolerance, or a lower tone and OT
indication on the display to indicate that the
DUT measurement is out of tolerance.
When the test is complete, use [SETUP],
<5ATest>, to view the stored test results.
Performance and Safety Features to Note
•Note that in the E-DWT-H design, special
attention was given to making a system that
minimize vessel expansion and thermal effects
that can cause the system to require a long time
to reach stability. As such, an operator should
be able to set stable test points in about 30
seconds, or no more than 1 minute depending
on the DUT tolerance.
•The system of lights and audible warnings on
E-DWT-H is intended to keep operators from
applying a high pressure to RPM4-E-DWT while
the Lo Q-RPT is selected. The green lights near
the Lo Q-RPT shutoff valve indicate the position
that the valve should be in based on the current
RPM4-E-DWT selection.
•There is a relief valve built into E-DWT-H that
protects the Lo Q-RPT in the event that the
operator mistakenly applies a high pressure to
the Lo Q-RPT circuit.
•In addition to audible warnings upon near-
overpressure of a Q-RPT; if the Lo Q-RPT shutoff
valve is set to Lo and the Hi Q-RPT is selected
on the RPM4-E-DWT, an audible warning
sounds if over 300 psi is sensed because it is
likely that the operator has forgotten to change
one of the settings and is applying pressure to
the Lo Q-RPT without the ability to see the
measured pressure.