solderingjumperA1
Electric connection:
ElectricconnectionsfortheGTH2448/4orGTH2448/5arelocatedatthebackofthedevice.
Connectionismadeviascrew-type/plug-interminals(max.terminalrange1,5mm²).
Make it a rule to always mount screw-type/plug-in terminals while they are still loose and connect only later.
If terminals are mounted after connection there is a risk that soldering eyes may come loose. Please use suitable
screw-driver and do not tighten screws by force.
Supply voltage: 12VDCor24VDC
Please make sure to check if supply voltage and voltage range set
conform to each other.
Use the soldering jumper next to the connection terminal to select
supply voltage:
Jumper"A1"open: 24V(18-29VDC)
Jumper"A1"closed: 12V(8-20VDC)
Probe connection: Pt1000,2-wire
Both the connection and commissioning of the device must only be car-
ried out by skilled personnel. In case of a wrong connection, the device
may be destroyed - no warranty claims can be accepted !
Safety regulations
Makeitaruletoalwaysobservethefollowingpointstoexcludeanyriskwhatsoeverfortheoperator.
a) Incaseofanyobviousdamageand/orfunctionalproblemsdisconnectdeviceimmediately
b) Priortoopeningit,disconnectdeviceandsupplyvoltagesource.Makesurethatallpartsofthedeviceareprotectedagainst
directtouchingwhenmountingthedeviceandsettingitsconnections.
c) Pleasealwaysadheretothestandardsafetyregulationsforelectricdevices,powersystemsandlight-currentinstallations,and
makesurethatyournationalsafetyregulations(e.g.VDE0100)areobserved.
d) Ifdeviceistobeconnectedtootherdevices(e.g.viaserialinterface)thecircuitryhastobedesignedmostcarefully.Internal
connectioninthirdpartydevicesmayresultinnot-permissiblevoltages.
Skilled personnel
Thesearepersonswhoarefamiliarwiththeinstallation,mounting,commissioningandtheoperationoftheproductandhaveac-
quiredaqualicationfortheirjob:
Trainingorinstructionsorqualicationtoswitchon/off,isolate,groundandapplymarkingstocircuitsanddevices/systemsin
accordancewiththelateststateoftheartstandardsofsafetytechnology.
Trainingorinstructionsregardingthepropercareanduseofsuitablesafetyequipmentinaccordancewiththelateststateofthe
artstandardsofsafetytechnology.
Firstaidtraining.
Warning: Whenoperatingelectricdevicespartsofthesedeviceswill,asamatterofcourse,belive.
Unlessthewarningsareobservedseveredamagetolifeandlimbortopropertymaybethe
result.Makesurethatonlyskilledpersonnelisworkingwiththisdevice.Trouble-freeopera-
tionofthisdevicecanonlybeguaranteedifitisproperlytransportedandstored.Careful
installation,mounting,operationandmaintenancearevitalfactorsforthesafeoperationofthis
device.
Warning:
Donotusetheseproductassafetyoremergencystopdevices,orinanyotherapplicationwherefail-
ureoftheproductcouldresultinpersonalinjuryormaterialdamage.
Failuretocomplywiththeseinstructionscouldresultindeathorseriousinjuryandmaterialdamage.
Terminalassignment
+UV Supplyvoltage+
GND Supplyvoltage–
Pt Probe(Pt1000)
Pt Probe(Pt1000)
Disposal notes
Thisdevicemustnotbedisposedas‘residualwaste’.Todisposethisdevice,pleasesenditdirectlytous(adequatelystamped).
Wewilldisposeitappropriatelyandenvironmentallyfriendly.
Pt1000 (-60.0 ... +199,9°C)
GTH 2448 / 5
Pt
Pt