4. Mount the digital counter onto the Co2 canis-
ter and mount the canister onto the trap unit.
For easier mount, one can moisten the rubber
on the counter. There are 24 shots in the Co2
canister. Tighten the canister onto the trap
swiftly to avoid Co2 leakage. It is important to
place the counter as shown on the above pic-
ture. This is to ensure that it works correctly.
WARNING! THE TRAP IS NOW ACTIVATED
5. The safe box is mounted steadfast and secure
with the use of the pre-punched holes and
ground spikes. Alternatively, the safe box can
be mounted to a wall using the pre-punched
holes and the enclosed kit.
6. The A24 unit and trap is mounted in the safe
box. Close the lid and for extra safety, one can
add a pad lock.
COUNTER
1. Resetting the counter when mounting it.
Display must be off. Press down the button
under the display for app. 2 sec. until it blinks
and 00 is on the display.
2. Reading the number of shots. Press down
shortly on the button under the display. The
number of shots are shown on the display.
3. Changing the battery. The bottom lid on the
counter is opened using a slotted screwdriver
or the like. Change the battery to a new one
of the same type and put the lid back on.
SERVICE AND MAINTENANCE
Service and maintenance of the GN A24 in safe box
performed by trained personnel.
1. Open the lid using the supplied key. Use of
gloves is recommended to avoid leaving trails
of human scents. Take out the A24 unit.
WARNING! THE TRAP IS ACTIVATED
2. Check the number of shots on the counter.
There are 24 shots in the Co2 canister.
If there are still shots and monitoring paste
left in the trap close and lock the safe box until
the next service check.
3. Remove the Co2 canister. Always use gloves
when removing the canister as it can become
very cold due to residual leakage of Co2.
Only use original 16 gram Co2 canisters like
the one shown on the above picture. If other
types are used the warranty is voided.
4. Remove the shroud cap and monitoring paste
from the unit. Warning: There can be leftover
shots even after the Co2 canister is removed.
As a safety measure check the trap minimum
2 times from above with the key or another
tool until it no longer reacts. To refresh and
optimize the scent trail remove a bit of the
outer layer of the monitoring paste and press
out new monitoring paste from the monitor-
ing paste basket. If necessary, change the
monitoring paste to a new one. Mount the
monitoring paste and shroud cap on the A24
unit again.
5. Clean the trap if necessary.
6. Reset the counter as described earlier. Mount
it onto a new Co2 canister. It is important to
place the counter as shown on the above pic-
ture. This is to ensure that it works correctly
and counts all shots. Screw the Co2 canister
onto the A24 unit.
WARNING! THE TRAP IS NOW ACTIVATED
7. Mount the A24 unit in the safe box again.
Close the lid and use a pad lock for extra safe-
ty if necessary.
8. Check that the safe box is still securely
mounted in the ground or alternatively to a
wall.
TRANSPORT AND STORAGE
If the safe box and the A24 unit is moved to another
location, the Co2 canister must first be removed and
the trap checked for leftover shots using the key or
another tool.
When delivered the safe box is secured using a cable
tie. This is also recommended if it is later moved or
placed in storage.
If the safe box and A24 unit has been active and must
go to storage for a period, one should remove the