4GMC-I Messtechnik GmbH
1 Safety Precautions
SECUTEST BASE(10), SECUTEST PRO and
SECULIFE ST BASE(25) test instruments are
manufactured and tested in accordance with
the following safety regulations:
IEC/EN 61010-1 / VDE 0411-1, DIN VDE
0404, IEC/EN 61577 / VDE 0413-2,-4 / DIN
EN 61557-16 / VDE 0413-16
The safety of the user, the test instrument
and the device under test (electrical equip-
ment or electrical medical device) is only
assured when the instrument is used for its
intended purpose.
Read these condensed operating instructions and
the full operating instructions carefully and com-
pletely before placing your test instrument into
service (available at our homepage www.gossen-
metrawatt.com for download). Follow all instruc-
tions contained therein. Make sure that the oper-
ating instructions are available to all users of the
instrument.
Tests may only be performed by a qualified
electrician, or under the supervision and
direction of a qualified electrician. The user
must be instructed by a qualified electrician
concerning performance and evaluation of
the test.
Suitable personal safety equipment is
required.
If you use any active or passive implanted
medical devices, please consult your doctor
or the manufacturer of these devices.
Manufacturers and importers of electrical
medical devices must provide documen-
tation for the performance of mainte-
nance by trained personnel.
Observe the following safety precautions:
• The instrument may only be connected
to electrical systems (TN, TT or IT) with a
maximum of 240 V which complies with
applicable safety regulations (e.g. IEC
60346, VDE 0100) and is protected with
a fuse or circuit breaker with a maximum
rating of 16 A.
• Measurements within electrical systems
are prohibited.
• Be prepared for the occurrence of unex-
pected voltages at devices under test (for
example, capacitors can be dangerously
charged).
• Make certain that the measurement
cables are in flawless condition, e.g. no
damage to insulation, no cracks in
cables or plugs etc.
• When using a test probe with coil cord
(SK2W):
Grip the tip of the test probe firmly, for
example during insertion into a jack
socket. Tensioning at the coil cord may
otherwise cause the test probe to snap
back resulting in possible injury.
•Measurement of insulation resistance and
equivalent leakage current (leakage current
alternative measuring methods)
Testing is conducted with up to 500 V.
Current limiting is utilized (I < 3.5 mA),
but if terminals L or N at the test socket
are touched, electrical shock may occur
which could result in consequential acci-
dents.
•Leakage current measurement during opera-
tion with line voltage: Please note that the
device under test is operated with line
voltage during measurement. Exposed
conductive parts may conduct danger-
ous touch voltage during testing. Do not
touch under any circumstances! (Mains
power is disconnected if leakage current
exceeds approx. 10 mA.)
The function test may only be per-
formed after the DUT has successfully
passed the safety test!
•Probe check for probe connector P1: Please
perform a probe check after each test.
If there is a fuse defect in test probe
P1 after having started the test, all
subsequent measurements which
are performed with this measure-
ment path might be erroneously
rated as good!
SECULIFE ST PRO Safety Precautions