
9309377
Setup
Calibration Factor
In this menu, you can recalibrate the meter for different
fluids.
1. If you have not already entered the dispensing
options menus, do steps 1 and 2 in Selecting
Dispensing Options on page 8.
2. Press the Auto/Reset button repeatedly until CAL
blinks and the current calibration factor is
displayed to indicate that you are in the Calibration
menu.
3. With CAL blinking, you can press the Auto/Reset
button and choose not to change the calibration
factor, which advances to the Auto Preset Amount
menu.
OR
You can recalibrate the meter as follows:
3a. If the unit of measurement is liters, dispense
exactly one liter of fluid into a calibrated
1-liter container. If the unit of measurement
is gallons, pints, or quarts, dispense exactly
one quart of fluid into a calibrated 1-quart
container. For proper calibration, you must
dispense the exact amount, according to the
markings on the container.
After dispensing, the displayed number is the
new calibration factor. The new calibration
factor must be within 20 percent of the
factory default value [in the range of 294 gal.
(310 liters) to 442 gal. (467 liters)], or it will
not be accepted.
The following table lists approximate
calibration factors for different fluids. Your
calibration number may vary slightly due to
temperature or flow rate.
Calibration Number
Fluid Quarts Liters
oil (10W–30) 368 389
gear lube 368 389
automatic
transmission
fluid
368 389
antifreeze 347 367
Calibration factors at 70_F (21_C) at 2.0 gpm (7.6 lpm).
3b. Press the Auto/Reset button to store the new
calibration factor and advance to the Auto
Preset Amount menu.
Auto Preset Amount
In this menu, you can specify an amount to be
displayed when you enter the Auto dispense mode.
Typically, you would enter the amount you most
frequently dispense.
1. If you have not already entered the dispensing
options menus, do steps 1 and 2 in Selecting
Dispensing Options on page 8.
2. Press the Auto/Reset button repeatedly until
AUTO blinks and the auto preset amount in
memory is displayed to indicate that you are in the
Auto Preset Amount menu. This is the amount
that is displayed when the Auto/Reset button is
pressed during normal operation.
3. With AUTO blinking and the stored auto preset
amount displayed, as shown below, do ONE of the
following two options:
DEnter a new auto preset amount by pressing
the 10 button to change the 10s digit,
the 1.0 button to change the 1s digit, and
the 0.1 button the change the first decimal
digit.
In the figure below, the arrows point from the
numbered buttons on the keypad to the
corresponding digits in the display.
NOTE: If you press and hold the 10, 1.0, or
0.1 buttons, the digits advance rapidly.
To store the displayed auto preset amount in
memory and advance to the Shut-Off Default
Amount menu, press the Auto/Reset button.
OR
DPress the Auto/Reset button.
The currently displayed auto preset amount is
retained, and the display advances to the
Shut-Off Default Amount menu.
! CAUTION
To prevent overfills and spills
Maximum Working Pressure 7 MPa (70 bar, 1000 PSI)
DDo not obstruct the trigger
DDo not leave unattended
DDo not operate below 32_F (0_C)
DRead instruction manual
06044A
AUTO
GAL