Instruction Manual
HI 93701
Free Chlorine ISM
www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
HI93701iswarrantedfortwoyears against defects in workman-
shipand materialswhenusedforits intendedpurposeandmaintained
accordingtoinstructions.
Thiswarrantyislimitedtorepairorreplacementfreeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required, contactthedealerfromwhomyou purchased the
instrument.Ifunderwarranty,reportthe model number, date of pur-
chase,serialnumberandthenatureof the failure. If the repair is not
coveredbythe warranty,youwillbenotified ofthechargesincurred.If
theinstrumentistobereturnedtoHannaInstruments,first obtain a
ReturnedGoodsAuthorizationNumberfrom the Customer Service de-
partmentand thensenditwithshipment costsprepaid.Whenshipping
anyinstrument, makesureitisproperly packagedforcompleteprotec-
tion.
Tovalidate yourwarranty,filloutandreturnthe enclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrights are reserved. Reproductionin whole or inpart is prohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket,Rhode Island, 02895 , USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
This instrument is in compliance with directives.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
The HI 93701 portable microprocessor meter measures the
free chlorine (Cl2) content in water and wastewater in the
0.00 to 2.50 mg/L (ppm) range.
Themeter usesan exclusivepositive-locking systemtoensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid or powder form and are supplied
in bottles or in packets. The amount of reagent is precisely
dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONS
Range 0.00 to 2.50 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading
TypicalEMC
Deviation ±0.01 mg/L
Light Source Light Emitting Diode @ 555 nm
Method Adaptation of the EPA recommended
DPDmethod330.5.The reactionbetween
freechlorineandtheDPDreagentcauses
a pink tint in the sample
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 80 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.)
REQUIRED REAGENTS
POWDER:
Code Description Quantity
HI 93701-0 DPD 1 packet
LIQUID:
Code Description Quantity
HI 93701A-F DPD1 Indicator 3 drops
HI 93701B-F DPD1Buffer 3 drops
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximum readableconcentration (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
Warranty
Preliminary Examination
General Description
Specifications Display Code Guide