
7
7
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
HI 92701 AND HI 92704HI 92701 AND HI 92704
HI 92701 AND HI 92704HI 92701 AND HI 92704
HI 92701 AND HI 92704
HI92701C HI92701F
HI92704C HI92704F
Range -200.0to 999.9°C -300.0to 999.9°F
1000to 1370°C 1000to2500°F
Resolution 0.1/1°C 0.2/1°F
Accuracy ±0.3% of reading ±0.3% of reading
(@25°C/77°F) or ±0.5°C or ±1°F
forone year, excluding probeerror
TypicalEMC Deviation ±5°C ±10°F
PowerSupply 4x1.5V AA size, 500 hours with 60' printing
interval. Socket for 12VDC adapter
AutoShut-off After 5 minutes of non-use
Channels 1for HI92701; 4for HI92704
Printer Low-power impact belt type, 14 characters per line
using 38 mm plain paper
PrintingIntervals 1, 2,5, 10,15,30, 60,120 and180minutes
Environment 0 to 50 °C (32 to 122°F) Max. 95% RH
Dimensions 220x 82x 66 mm(8.7 x3.2x 2.6")
Shipping Instrument: 500g (18 oz.)
Weight Kit: 1.4 kg (3.1 lb.)
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
HI 92801 AND HI 92804HI 92801 AND HI 92804
HI 92801 AND HI 92804HI 92801 AND HI 92804
HI 92801 AND HI 92804
HI92801C HI92801F
HI92804C HI92804F
Range -200.0to 999.9°C -300.0to 999.9°F
1000to 1370°C 1000to2500°F
Resolution 0.1/1°C 0.2/1°F
Accuracy ±0.3% of reading ±0.3% of reading
(@25°C/77°F) or ±0.5°C or ±1°F
forone year, excluding probeerror
TypicalEMC Deviation ±5°C ±10°F
PowerSupply 4x1.5V AA size, 500 hours with 60' printing
interval. Socket for 12VDC adapter
AutoShut-off After 5 minutes of non-use
Channels 1for HI92801; 4for HI92804
Printer Low-power impact belt type, 14 characters per line
using 38 mm plain paper
Printing/LoggingIntervals 1,2, 5,10, 15,30, 60, 120and 180minutes
Environment 0 to 50 °C (32 to 122°F) Max. 95% RH
Dimensions 220x 82x 66 mm(8.7 x3.2x 2.6")
Shipping Instrument: 500g (18 oz.)
Weight Kit: 1.4 kg (3.1 lb.)
Untitled-27
8
8
ON/OFF
TEMP
INITIAL PREPARATION
Each meter is supplied complete with batteries. Remove the back
cover, unwrap the batteries and install them while paying attention
to the polarity.
To prepare the instrument for use, connect a K-type thermocouple
temperature probe to the connector located on the top of the
instrument.
To switch the HI92701 and the HI92704 on,
press the ON/OFF key.
To switch the HI92801 and the HI92804 on,
press the TEMP key.
Choose the most appropriate temperature probe(s) for your applica-
tion (see page 30) and connect it (them) to the meter.
HI 92701 and HI 92801 have a single probe connection.
HI 92704 and HI 92804 have four probe connectors.
With the meter facing you, channel #1 is the first connector on the
top left hand side.
If the temperature probe is not connected to the instrument, the
meter will display and print "----" to alert the user.
Thisalso indicatesthe possibilityof abroken probecable. Thechannel
is displayed with "CH" and the number of channel (four channels
versionsonly).
Note: To ensure accurate readings it is recommended that the
temperature probes be connected to consecutive channels be-
ginning with channel #1. For example, if only one probe is to
be used, it must be connected to channel #1. If 2, 3 or 4
probes are to be used, they must be connected to channels 1,
2, 3, and 4 in ascending order.
CH
°C
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE