  9. Hioki 8715-01 User manual

Hioki 8715-01 User manual

POWER HiCORDER
Recorders
8715-01
Introducing easy-to-use power monitoring recorders!
See what you’ve been missing !
★Capture Spikes, Sags, and Surges
★Monitor and record power anomalies and fluctuations
easily, even unattended
★Analyze Leakage Currents as well as AC/DC Current up
to 2000 amps
★True-RMS with Trigger Function
★250µsec Transient Waveform Capture
Do you ever worry about power failure?
2
[ Setup ]
Measurement mode is set to Waveform, time axis to
2ms. Recording length is set to 20 division.
[ Channel setup ]
For current waveforms, after selecting a clamp-on
probe for the input type from Mode, select the range
of the clamp. Here, we select 9010 and 100A. Set
the range of the clamp-on probe in the same way.
[ Starting Measurement ]
Clamp the probe around the conductor to be
measured.
Press the START button.
The current waveform is
displayed on the screen.
To stop measuring, press
the STOP button.
What a Power Recorder Does
Simple Setup
There are many power-quality concerns in today's
PC-based businesses, and detecting power anomalies
is important for maintaining factory facilities and
equipment. The POWER HiCORDER 8715-01 is
monitoring recorders that can capture instantaneous
power drop-outs, brown-outs, spikes and sags, and
record voltage changes with simple operation, and even
monitor and record while totally unattended.
The compact B5 size and 1.6 kg (56.44 oz.) weight
provide excellent portability among worksites.
1. Simple setup. No worries about operability.
2. Accepts direct input of line voltages up to 400 V,
and up to 1000 V AC/ 2000 V DC with the use of the
DIFFERENTIAL PROBE 9322 accessory.
3. Voltage and current can be simultaneously measured on
four isolated channels.
4. Data immediately before an anomaly occurs can be
stored and recorded.
5. High quality data printouts on site.
6. High-speed A/D converter stores data in memory,
providing simple recording of events that cannot be
recorded by a pen recorder.
7. Two measurement functions
are available: Waveform
Measurement Mode (memory
mode), and RMS Trend
Measurement Mode (RMS
recorder & memory).
8. HIOKI clamp-on probes (voltage-
output type) can be directly
connected.
Example
To monitor the current waveform on a 220 V
power line.
Current Waveform Example
measured.
Press the START button.
The current waveform is
displayed on the screen.
To stop measuring, press
the STOP button.
Current Waveform Example
Record power anomalies accurately with simple operation !
- Aplications -
3
- Specifications -
Basic Specifications
Measurement Mode
Waveform Measurement mode (Power Waveform Observation),
RMS Trend Measurement mode (RMS Observation)
Input types and
number of channels
Input channels fixed, Analog 4 channels
Inter-channels and input-frame isolation
Maximum sampling rate
400 k sample/s (2.5 μs period)
Simultaneous sampling for 2 or 4 analog channels
Memory capacity
continuous 4-channels measurement (12-bit analog) × 64 kilo-
words/ ch
External memory
PC Card TYPE II slot × 1 : Flash ATA card (max.1 GB),
MS-DOS format. Memory contents : Settings data,
Measurement data (binary or text format), Screen data (BMP)
Backup function
Clock, waveform data and settings, Battery life approx. 5 years
(at 25°C/ 77°F).
External control
Terminal block : trigger input/output
Environment conditions
(no condensation)
Operation: +5°C/41°F to +40°C/104°F, 35% to 80% relative humidity.
Storage: -10°C/14°F to +50°C/122°F, 35% to 80% relative humidity.
Applicable standards
Safety: EN61010
EMC: EN61326, EN61000-3-2, EN61000-3-3
Power supplies
(1) AC Adapter model 9418-15 or 9418-10 (DC 12V ±10%)
(2) BATTERY PACK 9447 (AC adapter has priority when used in
combination with battery pack, fast recharge possible with AC adapter)
Power consumption
15 VA max.
Continuous operation time
Approx. 3 hours (with BATTERY PACK 9447, trigger standby at 23°C/ 73°F)
Charge time
Approx. 2 hours fast charge with power switch OFF (at 23°C/ 73°F)
Dimensions and
mass
approx. 280 mm (11.02 in) W × 170 mm (6.69 in) H × 52 mm
(2.05 in) D, 1.6kg (56.44 oz) (without batteries)
Supplied
accessories
Instruction manual ×1, Measurement guide ×1, Recording
paper ×1, Paper attachment ×2, Blank box ×1, Strap ×1, AC
ADAPTER 9418-15 ×1, CONNECTION CORD 9197 ×4,
Application disk ×1
Waveform Measurement (Power Waveform Observation) Mode
Time axis
200 and 400µs, 1, 2, 5, 10, and 20ms/division, time axis zoom
×2 to ×10; 3 settings, compression 1/2 to 1/50; 5 settings
Sampling period
1/80 of time axis ranges (minimum sampling period 2.5µs)
Recording length
20, 50, 100, 200, or 400 division (800 division at sequential save OFF)
Pre-trigger
Can record data from before the trigger point, 0/10/20/50, or 100%
of recording length
Other functions
Voltage axis normal (×1/2), magnified (×1), Left-right waveform
scrolling, Automatic, manual and partial (between A-B cursors) printing
Other Functions
General
Printing of settings including input range, trigger time, etc., cursor
measurement, start condition retention, auto setup, list & gauge
printing, power-save function, DMM function (voltage shown as
numerals on the display), auto saving, partial saving (between A-B cursors
or all data), sequential save.
DMM Function
Display update rate: 1 second
Display contents: RMS value (only measuring DC and 50/60Hz)
Display digit: 4 (last digit 0 to 4 is rounded zero, 5 to 9 is rounded five)
Voltage ranges: Auto-select only (five ranges from 10 mV to 100 V/ div.)
Accuracy: ±3 % rdg. ±5 dgt.
Input section
(accuracy at 23 ±5 ˚C/73 ±9 ˚F after 30 minutes warm-up time; accuracy guaranteed for 1 year)
Input
Isolated BNC terminal, Input item: Voltage, 9010, 9018, 9132,
3283, 3284, 3285, 9322 (selectable)
Measurement range
Voltage : 100, 200, 400 V line, or 100 V DC
9010 : 10, 20, 50, 100, 200, or 500 A
9018 : 10, 20, 50, 100, 200, or 500 A
9132 : 20, 50, 100, 200, 500, or 1000 A
3283 : 10 mA, 100 mA, 1, 10, or 200 A
3284 : 20 or 200 A
3285 : 200 or 2000 A
9322 : 400 or 600 V line
Max. sampling rate
400 kS/s (simultaneous sampling of all channels)
Accuracy, frequency
characteristics
±0.5 % f.s. (applied accuracy of clamp-on probe), DC to 50 kHz ±3 dB
Low-pass filter
OFF, 500 Hz, 5 kHz
Input resistance and
capacitance
1 MΩ, 7 pF approx. (at 100 kHz)
Input coupling
DC (f ixed)
Max. allowable input
450 V AC rms, DC CAT II (upper voltage which when applied to between
input pins does not damage them)
Max. rated voltage
to earth
450 V AC rms, DC (upper voltage which when applied to input channel
casing or between input channels does not damage them)
Recording and Display Section
Display
5.7-inch STN color LCD, with English/Japanese language selector
240 × 320 dots
Recording paper
112 mm (4.41 in) × 18 m (59.06 ft), thermal paper roll
Recording area
100 mm in full scale (10 divisions), 1 division = 10 mm (0.39 in)
Recording speed
Max. 1sec/division (with AC adapter use)
Trigger Function (only for anomalous waveform measurement and instantaneous power fluctuation recording)
Trigger source
Analog input CH1 - CH4 (8714 -01: CH1 - CH2), external (either ON or
OFF for each source), logical AND/OR of sources
Anomalous
waveform triggers
Level trigger: At preset voltage level, on either rising or falling edge
Window trigger: At entry or exit from preset upper and lower limits
Voltage drop trigger: Especially for 50/60 Hz commercial power,
when peak voltage falls below preset value
Period trigger: When period of rising or falling edge of measured
signal exceeds preset period
Waveform judgment trigger: Especially for 50/60Hz commercial
power, to monitor signals outside of judgment range in real time
(when time axis range 20ms/division is not applicable)
Power fluctuation
measurement triggers
(valid only when instantaneous
waveform measurement is
enabled)
Voltage drop trigger: Especially for 50/60Hz commercial power,
when peak voltage falls below preset value
RMS level trigger: At preset effective value, on either rising or
falling edge
RMS window trigger: At entry or exit from preset upper and lower
limit levels
Pre-Trigger
0, 10, 20, 50, 100% (for anomalous waveform measurement, instantaneous
power fluctuation recording)
Level setting resolution
0.25% f.s. (f.s.=10 divisions, 0.1% f.s. for use with waveform judgment trigger)
Trigger filter
Off/On (0.5 divisions fixed filter width)
RMS Trend Measurement (RMS Fluctuation Measurement) Mode
Measurement objective
Commercial mains power (50/60Hz)
Time axis
1/2/5 seconds/DIV (cannot realtime print); 10/30 seconds; 1/2/5/10/30
minutes; or 1hour/DIV, time axis compression 1/2 to 1/50; 5 settings
Sampling period
250 µs fixed (RMS value 800 data/second)
Recording length
Continuous measurement only (up to 200 division of data are stored internally)
RMS accuracy
±3 % f.s.
Waveform display
area
100 V Line: 75 to 125 V rms (standard)
200 V Line: 150 to 250 V rms (standard)
400 V Line: 275 to 525 V rms (standard)
Clamp: 0 A rms to f.s.value on clamp-on probe (r ms)
9322 Differential probe: 400 V Line; 275 to 525 V rms (standard)
9322 Differential probe: 600 V Line; 475 to 725 V rms (standard)
Other functions
Voltage axis normal (×1/2), magnified (×5), Left-right waveform scroll,
Automatic, manual and partial (between A-B cursors) printing, Hybrid
recording: prints logging data for each division together with waveform
RMS Trend Measurement (Instantaneous Waveform Measurement) Mode
Time axis
200 and 400 µs, 1, 2, 5, 10, and 20 ms/division, time axis zoom
×2 to ×10; 3 settings, compression 1/2 to 1/50; 5 settings
Sampling period
1/80 of time axis ranges (minimum sampling period 2.5µs)
Recording length
20, 50, 100, or 200 division (400 division at sequential save OFF)
Pre-trigger
Can record data from before the trigger point, 0/10/20/50, or 100%
of recording length
Other functions
Voltage axis normal (×1/2), magnified (×1), Left-right waveform scroll
Mass : approx. 1.6 kg (56.44 oz)
170 mm
(6.69 in)
280 mm
(11.02 in)
52 mm
(2.05 in)
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@hiokiusa.com
DISTRIBUTED BY
All information correct as of Jan. 5, 2009. All specifi cations are subject to change without notice. 8715-01E6-91B-00P Printed in Japan
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
1608-1610 Shanghai Times Square Office, 93 Huai Hai Zhong
Road
, Shanghai, P.R.China POSTCODE: 200021
TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360
E-mail: info-sh@hioki.com.cn
Beijing Office :
A-2602 Freetown, 58 Dong San Huan Nan Road
Beijing, P.R.China POSTCODE: 100022
TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090
E-mail: info-bj@hioki.com.cn
Guangzhou Office :
Room A-3206, Victory PlazaServices Center, No.103,
Tiyuxi Road, Guangzhou, P.R.China POSTCODE:510620
TEL +86-20-38392673/2676 FAX +86-20-38392679
E-mail: info-gz@hioki.com.cn
Up to 100/200 VAC measurement
(Input cable 9197 is supplied)
CONNECTION CORD 9790
(Thin Type) CAT II300 V, ultra-
flexible 2.8 mm (0.11 in) diameter test
lead cable, 1.5 m (4.92 ft) length
Note: Attachment clips sold separately.
ALLIGATOR CLIP 9790
-01
Red/black set attaches to the ends
of test leads (9790)
GRABBER CLIP 9790
-02
Red/black set attaches to the ends
of test leads (9790)
CONTACT PIN 9790-03
Red/black set attaches to the ends
of test leads (9790)
CONNECTION CORD 9197
For up to 200 VAC, 1 : 1
probe, 1.8 m (5.91 ft) length
CONNECTION CORD 9198
For up to 200 VAC, 1 : 1
probe, 1.7 m (5.58 ft) length
Tip Expanders
Attachment clips are sold separately
from CONNECTION CORD 9790.
Purchase the appropriate attachment
clips for your application separately
Tip Expanders
9790
-01
Tip Expanders
9790
-03
Tip Expanders
9790
-02
POWER HiCORDER 8715
-01 (4 ch)
AC ADAPTER 9418-15 ×1, CONNECTION CORD
9197 ×4, Recording paper ×1, Application disk, or other
accessories are included
Configuration of options Note: Product names appearing in this catalog are trademarks or registered trademarks of various companies.
Up to 200/400 VAC measurement
Input from 100 V power-concent
DIFFERENTIAL PROBE 9322
For up to 200 V, 400 VAC. Use with
the AC Adapter 9418-15, for power
supply. 1000 : 1 probe
CONCENT INPUT CORD 9448
Use for 100 V commercial line,
2m (6.56 ft) length
AC ADAPTER 9418-15
For powering Differential probe
9322, 100 to 240 V AC.
Current measurement *50/60 Hz line use
Current measurement *50/60 Hz line use
Power Supply (Sold separately)
Other options
Printer options
PC Software
CLAMP ON PROBE 9018
-50
Enables observation of AC current
waveforms. Input: 40 Hz to 3 kHz,
selectable 10 to 500 A ranges, Output:
0.2 V AC/ range
Carrying case
CARRYING CASE 9648
Hard case type, for storing options
CARRYING CASE 9391
Soft case type, for storing options
Holds more options than the 9648
hard case
BATTERY PACK 9447
7.2 V, 2400 mAh
CHARGE STAND 9643
Independent of main unit the 8715-01, use with
the AC ADAPTER 9418-15 to charge one Model
BATTERY PACK 9447.
AC ADAPTER 9418-15
Universal 100 to 240 V AC, 12 V
DC/ 2.5 A output
CONVERSION ADAPTER
9199
Banana-to-BNC, use to connect
to insulation-BNC terminal on
Input section
CLAMP ON PROBE 9132
-50
Enables observation of AC current
waveforms. Input: 40 Hz to 1 kHz,
selectable 20 to 1000 A ranges, Output:
0.2 V AC/ range
CLAMP ON AC/DC HiTESTER 3284
20 A or 200 A AC/DC, DC analog output,
waveform monitor output, 10 Hz to 2 kHz
response.
CLAMP ON AC/DC HiTESTER 3285
200 A or 2000 A AC/DC, DC analog output,
waveform monitor output, 10 Hz to 1 kHz
response.
CLAMP ON LEAK HiTESTER 3283
For leakage current measurement, 10 mA to
200 AAC ranges, DC analog output, waveform
monitor output, 40 Hz to 2 kHz response.
AC ADAPTER 9445
-02/-03
With the model 3283, 3284, 3285
continuous use.
For USA, Canada: 9445-02,
For EU: 9445-03.
CONVERSION ADAPTER 9199
Banana-to-BNC, use to connect to
insulation-BNC terminal on Input
section
OUTPUT CORD 9094
Required along with the 9199 adapter to
connect the model 3283, 3284, 3285 to
the 8715-01.
Input cords of included accessories
are the 9197 only. For current
measurement, please purchase
the optional clamp-on probe, or
clamp-on tester.
Supplied Accessories
Recommended
+
PAPER WINDER 220H
Paper width: 70 to 220 mm (2.76 to
8.66 in), using special-purpose AC
adapter
RECORDING PAPER
9234
112 mm (4.41 in) × 18 m
(59.06 ft), 10 rolls set
Not CE marked
Supplied Accessories
Removable storage (CF card)
Supplied with PC
Card adapter
PC CARD 256M 9727
(256 MB capacity)
PC CARD 512M 9728
(512 MB capacity)
PC CARD 1G 9729
(1 GB capacity)
PC Card Precaution
Use only PC Cards sold by HIOKI.
Compatibility and performance
are not guaranteed for PC cards
made by other manufacturers. You
may be unable to read from or save
data to such cards.
LINE SPLITTER CT101A
For 100 V/ 15 A, convenient for
measuring 100 V AC line current
with clamp-on probe
CONNECTION CORD 9217
Cord has insulated BNC connectors
at both ends, and connects to
insulated BNC connectors on input
terminal. 1.7 m (5.58 ft) length.
CONNECTION CORD 9165
Cord has metallic BNC connectors at
both ends, and connects to metallic BNC
connectors. 1.5 m (4.92 ft) length.
CONVERSION ADAPTER 9199
Banana-to-BNC, use to connect to
BNC terminal on Input Module
Not CE marked
Not CE marked
WAVE PROCESSOR 9335
Data conversion, print functions,
waveform display, compatible
with Windows 95/98/Me,
Windows NT 4.0/2000/XP, and
Windows Vista 32-bit type.
Recommended
+
Not CE marked
Combination example (1)
Main unit with memory 64 kW/ch
Recording Paper Battery pack Current probe Carrying case
4-channels, with current
measurement
8715-01 × 1 9234 × 1 (10 rolls) 9447 × 1 9018-50 × 4 9391 × 1

Other manuals for 8715-01

3

Other Hioki Measuring Instrument manuals

Hioki FT3432 User manual

Hioki

Hioki FT3432 User manual

Hioki PR8111 User manual

Hioki

Hioki PR8111 User manual

Hioki 8715-01 User manual

Hioki

Hioki 8715-01 User manual

Hioki PW3335 User manual

Hioki

Hioki PW3335 User manual

Hioki L2003 User manual

Hioki

Hioki L2003 User manual

Hioki 3280-10F User manual

Hioki

Hioki 3280-10F User manual

Hioki IM3523 User manual

Hioki

Hioki IM3523 User manual

Hioki 3157 User manual

Hioki

Hioki 3157 User manual

Hioki CM4371 User manual

Hioki

Hioki CM4371 User manual

Hioki PQ3100 User manual

Hioki

Hioki PQ3100 User manual

Hioki 3490 User manual

Hioki

Hioki 3490 User manual

Hioki PW6001 User manual

Hioki

Hioki PW6001 User manual

Hioki PW9100-03 User manual

Hioki

Hioki PW9100-03 User manual

Hioki LCR HiTESTER 3535 Instruction manual

Hioki

Hioki LCR HiTESTER 3535 Instruction manual

Hioki MR8875 User manual

Hioki

Hioki MR8875 User manual

Hioki PW3365-20 User manual

Hioki

Hioki PW3365-20 User manual

Hioki SM7420 User manual

Hioki

Hioki SM7420 User manual

Hioki 8715-01 User manual

Hioki

Hioki 8715-01 User manual

Hioki 8841 User manual

Hioki

Hioki 8841 User manual

Hioki 3532-50 User manual

Hioki

Hioki 3532-50 User manual

Hioki 9625 User manual

Hioki

Hioki 9625 User manual

Hioki CM3286-01 User manual

Hioki

Hioki CM3286-01 User manual

Hioki 9651 User manual

Hioki

Hioki 9651 User manual

Hioki 2300 Smart Site User manual

Hioki

Hioki 2300 Smart Site User manual

Popular Measuring Instrument manuals by other brands

Hanna Instruments Checker HI716 Instruction manua

Hanna Instruments

Hanna Instruments Checker HI716 Instruction manua

Bosch DLE 70 Professional Original instructions

Bosch

Bosch DLE 70 Professional Original instructions

ATI Technologies C12-17 Operation and maintenance manual

ATI Technologies

ATI Technologies C12-17 Operation and maintenance manual

Anyload 805BS Series manual

Anyload

Anyload 805BS Series manual

ABB AquaMaster IM9 installation manual

ABB

ABB AquaMaster IM9 installation manual

Toro Turf Guard Installation setup manual

Toro

Toro Turf Guard Installation setup manual

GDS GDS-68XP Operation and maintenance manual

GDS

GDS GDS-68XP Operation and maintenance manual

Sage RIO SIX instruction manual

Sage

Sage RIO SIX instruction manual

HT Italia PQA819 user manual

HT Italia

HT Italia PQA819 user manual

KMB SMV operating manual

KMB

KMB SMV operating manual

Stewart Warner B-755 installation instructions

Stewart Warner

Stewart Warner B-755 installation instructions

Shodex IEC SP-825 Operation manual

Shodex

Shodex IEC SP-825 Operation manual

Keysight B2900A Demo guide

Keysight

Keysight B2900A Demo guide

TESTO 475 instruction manual

TESTO

TESTO 475 instruction manual

GE N60 instruction manual

GE

GE N60 instruction manual

PASCO PS-3230 quick start guide

PASCO

PASCO PS-3230 quick start guide

UEi C161C instruction manual

UEi

UEi C161C instruction manual

Netech UPM 2000 user manual

Netech

Netech UPM 2000 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.