POWER HiCORDER
Recorders
8714-01,8715-01
Introducing easy-to-use power monitoring recorders!
See what you’ve been missing !
★Capture Spikes, Sags, and Surges
★Monitor and record power anomalies and fluctuations
easily, even unattended
★Analyze Leakage Currents as well as AC/DC Current up
to 2000 amps
★True-RMS with Trigger Function
★250µsec Transient Waveform Capture
Do you ever worry about power failure?
2
[ Setup ]
Measurement mode is set to Waveform, time axis to
2ms. Recording length is set to 20 division.
[ Channel setup ]
For current waveforms, after selecting a clamp-on
probe for the input type from Mode, select the range
of the clamp. Here, we select 9010 and 100A. Set
the range of the clamp-on probe in the same way.
[ Starting Measurement ]
Clamp the probe around the conductor to be
measured.
Press the START button.
The current waveform is
displayed on the screen.
To stop measuring, press
the STOP button.
What a Power Recorder Does
Simple Setup
There are many power-quality concerns in today's
PC-based businesses, and detecting power anomalies
is important for maintaining factory facilities and
equipment. The POWER HiCORDERs 8714-01 and
8715-01 are monitoring recorders that can capture
instantaneous power drop-outs, brown-outs, spikes and
sags, and record voltage changes with simple operation,
and even monitor and record while totally unattended.
The compact B5 size and 1.5kg (52.9oz.) weight provide
excellent portability among worksites. (8714-01)
1. Simple setup. No worries about operability.
2. Accepts direct input of line voltages up to 400 V, and
up to 1000 V AC/ 2000 V DC with the use of the
DIFFERENTIAL PROBE 9322 accessory.
3. Voltage and current can be simultaneously measured on
four isolated channels.
4. Data immediately before an anomaly occurs can be stored
and recorded.
5. High quality data printouts on site.
6. High-speed A/D converter stores data in memory,
providing simple recording of events that cannot be
recorded by a pen recorder.
7. Two measurement functions are
available: Waveform Measurement
Mode (memory mode), and RMS
Trend Measurement Mode (RMS
recorder & memory).
8. HIOKI clamp-on probes (voltage-
output type) can be directly
connected.
Example
To monitor the current waveform on a 220 V
power line.
Current Waveform Example
measured.
Press the START button.
The current waveform is
displayed on the screen.
To stop measuring, press
the STOP button.
Current Waveform Example
Record power anomalies accurately with simple operation !
- Aplications -
3
Two models: the two-channel (8714-01) and four-channel (8715-01)
- Specifications -
Basic Specifications
8714-01 8715-01
Measurement Mode
Waveform Measurement mode (Power Waveform Observation),
RMS Trend Measurement mode (RMS Observation)
Input types and
number of channels
Input channels fixed, Analog 2 channels
Inter-channels and input-frame isolation
Input channels fixed, Analog 4 channels
Inter-channels and input-frame isolation
Maximum sampling rate
400k sample/s (2.5μs period)
Simultaneous sampling for 2 or 4 analog channels
Memory capacity
continuous 2-channels measurement
(12-bit analog) × 64 kilo-words/ ch
continuous 4-channels measurement
(12-bit analog) × 64 kilo-words/ ch
External memory
PC Card TYPE II slot × 1 : Flash ATA card (max.1 GB), MS-DOS
format. Memory contents : Settings data, Measurement data
(binary or text format), Screen data (BMP)
Backup function
Clock, waveform data and settings, Battery life approx. 5 years (at
25°C/ 77°F).
External control
Terminal block : trigger input/output
Environment conditions
(no condensation)
Operation: +5°C/41°F to +40°C/104°F, 35% to 80% relative humidity.
Storage: -10°C/14°F to +50°C/122°F, 35% to 80% relative humidity.
Applicable standards
Safety: EN61010
EMC: EN61326, EN61000-3-2, EN61000-3-3
Power supplies
(1) AC Adapter model 9418-15 or 9418-10 (DC 12V ±10%)
(2) BATTERY PACK 9447 (AC adapter has priority when used in
combination with battery pack, fast recharge possible with AC adapter)
Power consumption
15VA max.
Continuous operation time
Approx. 3 hours (with BATTERY PACK 9447, trigger standby at 23°C/ 73°F)
Charge time
Approx. 2 hours fast charge with power switch OFF (at 23°C/ 73°F)
Dimensions and
mass
approx. 280 (11.02) W × 170 (6.69) H × 52 (2.05)
Dmm (inch), 1.5kg/ 52.91oz (without batteries)
approx. 280 (11.02) W × 170 (6.69) H × 52 (2.05)
Dmm (inch) , 1.6kg/ 56.44oz (without batteries)
Supplied
accessories
Instruction manual ×1, Measurement guide ×1, Recording paper ×
1, Paper attachment ×2, Blank box ×1, Strap ×1, AC ADAPTER
9418-15 ×1, CONNECTION CORD 9197 ×2 (8714-01), ×4 (8715-01),
Application disk ×1
Waveform Measurement (Power Waveform Observation) Mode
Time axis
200 and 400µs, 1, 2, 5, 10, and 20ms/division, time axis zoom
×2 to ×10; 3 settings, compression 1/2 to 1/50; 5 settings
Sampling period
1/80 of time axis ranges (minimum sampling period 2.5µs)
Recording length
20, 50, 100, 200, or 400 division (800 division at sequential save OFF)
Pre-trigger
Can record data from before the trigger point, 0/10/20/50, or 100%
of recording length
Other functions
Voltage axis normal (×1/2), magnified (×1), Left-right waveform scrolling,
Automatic, manual and partial (between A-B cursors) printing
Other Functions
General
Printing of settings including input range, trigger time, etc., cursor
measurement, start condition retention, auto setup, list & gauge
printing, power-save function, DMM function (voltage shown as numerals
on the display), auto saving, partial saving (between A-B cursors or all data),
sequential save.
DMM Function
Display update rate: 1 second
Display contents: RMS value (only measuring DC and 50/60Hz)
Display digit: 4 (last digit 0 to 4 is rounded zero, 5 to 9 is rounded five)
Voltage ranges: Auto-select only (five ranges from 10 mV to 100 V/ div.)
Accuracy: ±3% rdg. ±5dgt.
Input section
(accuracy at 23 ±5 ˚C/73 ±9 ˚F after 30 minutes warm-up time; accuracy guaranteed for 1 year)
Input
Isolated BNC terminal, Input item: Voltage, 9010, 9018, 9132,
3283, 3284, 3285, 9322 (selectable)
Measurement range
Voltage : 100, 200, 400V line, or 100V DC
9010 : 10, 20, 50, 100, 200, or 500A
9018 : 10, 20, 50, 100, 200, or 500A
9132 : 20, 50, 100, 200, 500, or 1000A
3283 : 10mA, 100mA, 1, 10, or 200A
3284 : 20 or 200A
3285 : 200 or 2000A
9322 : 400 or 600V line
Max. sampling rate
400kS/s (simultaneous sampling of all channels)
Accuracy, frequency
characteristics
±0.5% f.s. (applied accuracy of clamp-on probe), DC to 50kHz ±3dB
Low-pass filter
OFF, 500Hz, 5kHz
Input resistance and
capacitance
1MΩ, 7pF approx. (at 100kHz)
Input coupling
DC (fixed)
Max. allowable input
450V AC rms, DC CAT II (upper voltage which when applied to between
input pins does not damage them)
Max. rated voltage
to earth
450V AC rms, DC (upper voltage which when applied to input channel
casing or between input channels does not damage them)
Recording and Display Section
Display
5.7-inch STN color LCD, with English/Japanese language selector
240 × 320 dots
Recording paper
112mm (4.41in) × 18m (59.06ft), thermal paper roll
Recording area
100mm in full scale (10 divisions), 1 division = 10mm (0.39in)
Recording speed
Max. 1sec/division (with AC adapter use)
Trigger Function (only for anomalous waveform measurement and instantaneous power fluctuation recording)
Trigger source
Analog input CH1 - CH4 (8714-01: CH1 - CH2), external (either ON or
OFF for each source), logical AND/OR of sources
Anomalous
waveform triggers
Level trigger: At preset voltage level, on either rising or falling
edge
Window trigger: At entry or exit from preset upper and lower limits
Voltage drop trigger: Especially for 50/60Hz commercial power,
when peak voltage falls below preset value
Period trigger: When period of rising or falling edge of measured
signal exceeds preset period
Waveform judgment trigger: Especially for 50/60Hz commercial
power, to monitor signals outside of judgment range in real time
(when time axis range 20ms/division is not applicable)
Power fluctuation
measurement triggers
(valid only when instantaneous
waveform measurement is enabled)
Voltage drop trigger: Especially for 50/60Hz commercial power,
when peak voltage falls below preset value
RMS level trigger: At preset effective value, on either rising or
falling edge
RMS window trigger: At entry or exit from preset upper and lower
limit levels
Pre-Trigger
0, 10, 20, 50, 100% (for anomalous waveform measurement, instantaneous
power fluctuation recording)
Level setting resolution
0.25% f.s. (f.s.=10 divisions, 0.1% f.s. for use with waveform judgment trigger)
Trigger filter
Off/On (0.5 divisions fixed filter width)
RMS Trend Measurement (RMS Fluctuation Measurement) Mode
Measurement objective
Commercial mains power (50/60Hz)
Time axis
1/2/5 seconds/DIV (cannot realtime print); 10/30 seconds; 1/2/5/10/30
minutes; or 1hour/DIV, time axis compression 1/2 to 1/50; 5 settings
Sampling period
250µs fixed (RMS value 800 data/second)
Recording length
Continuous measurement only (up to 200 division of data are stored internally)
RMS accuracy
±3 % f.s.
Waveform display
area
100V Line: 75 to 125V rms (standard)
200V Line: 150 to 250V rms (standard)
400V Line: 275 to 525V rms (standard)
Clamp: 0A rms to f.s.value on clamp-on probe (rms)
9322 Differential probe: 400V Line; 275 to 525V rms (standard)
9322 Differential probe: 600V Line; 475 to 725V rms (standard)
Other functions
Voltage axis normal (×1/2), magnified (×5), Left-right waveform scroll,
Automatic, manual and partial (between A-B cursors) printing, Hybrid
recording: prints logging data for each division together with waveform
RMS Trend Measurement (Instantaneous Waveform Measurement) Mode
Time axis
200 and 400µs, 1, 2, 5, 10, and 20ms/division, time axis zoom
×2 to ×10; 3 settings, compression 1/2 to 1/50; 5 settings
Sampling period
1/80 of time axis ranges (minimum sampling period 2.5µs)
Recording length
20, 50, 100, or 200 division (400 division at sequential save OFF)
Pre-trigger
Can record data from before the trigger point, 0/10/20/50, or 100%
of recording length
Other functions
Voltage axis normal (×1/2), magnified (×1), Left-right waveform scroll
Mass (8714-01) : approx. 1.5kg/52.91oz
Mass (8715-01) : approx. 1.6kg/56.44oz
170 mm
(6.69 in)
280 mm
(11.02 in)
52 mm
(2.05 in)
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@hiokiusa.com
DISTRIBUTED BY
All information correct as of Dec. 21, 2007. All specications are subject to change without notice. 8714E4-7ZM Printed in Japan
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
1904 Shanghai Times Square Office, 93 Huai Hai Zhong Road
Shanghai, P.R.China POSTCODE: 200021
TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360
E-mail: info-sh@hioki.cn
Beijing Office :
A-2602 Freetown, 58 Dong San Huan Nan Road
Beijing, P.R.China POSTCODE: 100022
TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090
E-mail: info-bj@hioki.cn
Guangzhou Office :
Room 303, Profit Plaza, No.76, West Huangpu Road
Guangzhou, P.R.China POSTCODE: 510623
TEL +86-20-38392673/2676 FAX +86-20-38392679
E-mail: info-gz@hioki.cn
Input from power-concent
POWER HiCORDER 8714-01 (2ch model)
POWER HiCORDER 8715-01 (4ch model)
CARRYING CASE 9648
Hard case type, for storing
options
PAPER WINDER 220H
Paper width: 70 to 220 mm,
using special-purpose AC
adapter
RECORDING PAPER
9234
18 m/ 59.06 feet length,
10 rolls/ 1 set
CARRYING CASE 9391
Soft case type, for storing options
Holds more options than the 9648
hard case
Included accessories: AC ADAPTER 9418-15 ×1, Recording paper ×1, Blank box ×1, Strap ×1,
CONNECTION CORD 9197 (8714-01:×2, 8715-01:×4), Application disk ×1
CLAMP ON PROBE 9018-10
Input from 10 to 500A
40Hz to 3kHz for 0.2V AC output.
BNC terminal
CLAMP ON PROBE 9132-10
Input from 20 to 1000A
40Hz to 1kHz for 0.2V AC output.
BNC terminal
LINE SPLITTER CT101A
For 100V/ 15A,
convenient for measuring
100V AC line current with
clamp-on probe
Composition of options Note: Product names appearing herein are trademarks or registered trademarks of various companies.
CONNECTION CORD
9217
Insulation BNC-to-insulation
BNC, use to connect to
insulation-BNC terminal on
Input Module
CONNECTION CORD 9197
For up to 500V,
1.5 m length
CONCENT INPUT CORD
9448
Use for 100V commercial
line, 2m length
WAVE PROCESSOR 9335
Data conversion, print functions,
waveform display, compatible with
Windows 95/98/Me, Windows NT
4.0/2000/XP
PC CARD 128M 9726
(128MB capacity)
PC CARD 256M 9727
(256MB capacity)
PC CARD 512M 9728
(512MB capacity)
PC CARD 1G 9729
(1GB capacity)
Supplied with PC
Card adapter
CONVERSION ADAPTER
9199
Banana-to-BNC, use to
connect to insulation-BNC
terminal on Input section
CLAMP ON AC/DC HiTESTER
3284
20A or 200A AC/DC, with
analog output of 1V f.s. DC, and
waveform monitor output of 1V
f.s. AC at 10Hz to 2kHz, or DC.
Requires the AC ADAPTER
9445-02/-03
CLAMP ON LEAK HiTESTER
3283
For leakage current measurement,
includes 10mA to 200A ranges,
with analog output of 1V f.s. DC,
and waveform monitor output
of 1V f.s. AC at 40Hz to 2kHz.
Requires the AC ADAPTER
9445-02/-03
BATTERY PACK 9447
7.2V, 2400mAh
CHARGE STAND 9643
Independent of main unit the
8714-01/8715-01, use with the AC
ADAPTER 9418-15 to charge one
Model BATTERY PACK 9447.
AC ADAPTER 9418-15
Universal 100 to 240 V AC,
12 V DC/ 2.5 A output
AC ADAPTER 9445-02/-03
Use for the 3283, 3284,
3285, for USA/CSA
9445-02, for EU 9445-03
OUTPUT CORD 9094
Required along with the
9199 adapter to connect
the 3283/3284/3285 to the
8714-01 or 8715-01
CONVERSION ADAPTER
9199
Banana-to-BNC, use to
connect to insulation-BNC
terminal on Input section
+
CLAMP ON AC/DC HiTESTER
3285
200A or 2000A AC/DC, with
analog output of 1V f.s. DC, and
waveform monitor output of 1V
f.s. AC at 10Hz to 1kHz, or DC.
Requires the AC ADAPTER
9445-02/-03
AC ADAPTER 9418-15
Use for the 9322, 100 to 240
V AC
DIFFERENTIAL PROBE 9322
For inputs up to 2kV DC or
1kV AC, the 9322 requires
the AC ADAPTER 9418-15
1kV AC, 2kV DC input
Up to 450V AC input
Record Current Waveform Record Current Waveform
PC Card Precaution
Use only PC Cards sold
by HIOKI. Compatibility
and performance are not
guaranteed for PC cards made
by other manufacturers. You
may be unable to read from or
save data to such cards.
Not CE certified
Not CE certified
Not CE certified
Not CE certified
Input cords of included accessories are the 9197 only. For current
measurement, please purchase the optional clamp-on probe, or clamp-on
tester.
+
nCombination example: 2-channels, with current measurement
Main unit
Paper Battery pack Current up to 500A Carrying case
Model number ×
quantity
8714-01×1 9234×1 (10 rolls) 9447×1 9018-10×2 9391×1
nCombination example: 4-channels, with current measurement
Main unit
Paper Battery pack Current up to 500A Carrying case
Model number ×
quantity
8715-01×1 9234×1 (10 rolls) 9447×1 9018-10×4 9391×1

Other manuals for 8715-01

3

This manual suits for next models

1

Other Hioki Measuring Instrument manuals

Hioki DSM-8104 User manual

Hioki

Hioki DSM-8104 User manual

Hioki FT3424 User manual

Hioki

Hioki FT3424 User manual

Hioki IM3536 User manual

Hioki

Hioki IM3536 User manual

Hioki 8715-01 User manual

Hioki

Hioki 8715-01 User manual

Hioki 3662-20 User manual

Hioki

Hioki 3662-20 User manual

Hioki PW3390 Use and care manual

Hioki

Hioki PW3390 Use and care manual

Hioki PQ3198 User manual

Hioki

Hioki PQ3198 User manual

Hioki MR8875 User manual

Hioki

Hioki MR8875 User manual

Hioki DSM-8542 User manual

Hioki

Hioki DSM-8542 User manual

Hioki FT3424 User manual

Hioki

Hioki FT3424 User manual

Hioki 3280-10F User manual

Hioki

Hioki 3280-10F User manual

Hioki SM-8216 User manual

Hioki

Hioki SM-8216 User manual

Hioki 8847 User manual

Hioki

Hioki 8847 User manual

Hioki Memory HiCorder MR6000 User manual

Hioki

Hioki Memory HiCorder MR6000 User manual

Hioki 3283-20 User manual

Hioki

Hioki 3283-20 User manual

Hioki 9651 User manual

Hioki

Hioki 9651 User manual

Hioki PW3365-20 User manual

Hioki

Hioki PW3365-20 User manual

Hioki PW3360-21 User manual

Hioki

Hioki PW3360-21 User manual

Hioki 8715-01 User manual

Hioki

Hioki 8715-01 User manual

Hioki 3532-50 User manual

Hioki

Hioki 3532-50 User manual

Hioki 9695-02 User manual

Hioki

Hioki 9695-02 User manual

Hioki 9274 User manual

Hioki

Hioki 9274 User manual

Hioki 8835-01 User manual

Hioki

Hioki 8835-01 User manual

Hioki CM4373-50 User manual

Hioki

Hioki CM4373-50 User manual

Popular Measuring Instrument manuals by other brands

Klein Tools ET110 instruction manual

Klein Tools

Klein Tools ET110 instruction manual

Curtis 700 series manual

Curtis

Curtis 700 series manual

Keysight 16900 series installation guide

Keysight

Keysight 16900 series installation guide

PCB Piezotronics 356A12 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 356A12 Installation and operating manual

geo-FENNEL FL 240HV user manual

geo-FENNEL

geo-FENNEL FL 240HV user manual

Rainforest Automation EMU-2 quick start guide

Rainforest Automation

Rainforest Automation EMU-2 quick start guide

GANN Hydromette BL A plus operating instructions

GANN

GANN Hydromette BL A plus operating instructions

MICRO-EPSILON thicknessCONTROL MTS 8202.LLT operating instructions

MICRO-EPSILON

MICRO-EPSILON thicknessCONTROL MTS 8202.LLT operating instructions

Lutron Electronics YK-2001PHA Operation manual

Lutron Electronics

Lutron Electronics YK-2001PHA Operation manual

F&F OM-633 quick start guide

F&F

F&F OM-633 quick start guide

TIME MARK 26 quick start guide

TIME MARK

TIME MARK 26 quick start guide

FineTek EDX2 Operation manual

FineTek

FineTek EDX2 Operation manual

HP 3456A Operating and service manual

HP

HP 3456A Operating and service manual

SIGRIST TurBiScat PM 40 Brief instructions

SIGRIST

SIGRIST TurBiScat PM 40 Brief instructions

TOOLCRAFT LDM X100 operating instructions

TOOLCRAFT

TOOLCRAFT LDM X100 operating instructions

Endress+Hauser SpectraSensors OXY5500 Safety manual

Endress+Hauser

Endress+Hauser SpectraSensors OXY5500 Safety manual

Pfeiffer Vacuum GSD 350 operating instructions

Pfeiffer Vacuum

Pfeiffer Vacuum GSD 350 operating instructions

Ushio UIT-0 Series User's manual and troubleshooting guide

Ushio

Ushio UIT-0 Series User's manual and troubleshooting guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.