Hovertech AirWedge User manual

TABLE OF CONTENTS
Symbol Reference..................................................................................2
Intended Use and Precautions..........................................................2
Part Identication
– AirWedge® Adjustable Positioning Device...............................3
Part Identication – HT-Air® Air Supply .........................................3
HT-Air® Keypad Functions...................................................................3
Instructions for Use ...............................................................................4
Product Specications/Required Accessories.............................4
Cleaning & Preventive Maintenance..............................................5
Warranty Statement..............................................................................6
Returns and Repairs..............................................................................7
Adjustable Positioning Device
User Manual
Visit www.HoverMatt.com for other languages

User Manual
2 | HoverTech International AWManual, Rev. B
Intended Use and Precautions
INTENDED USE
The AirWedge® Adjustable Positioning Device is used to pneumatically
lift and position the head and chest of a patient who is lying in the
supine position without manual lifting. It may also be used to elevate
legs, feet, etc.
INDICATIONS FOR USE
Patients who are unable to position themselves for procedures or general
patient care
CONTRAINDICATIONS
Patients who are experiencing thoracic, cervical or lumbar fractures that are
deemed unstable
INTENDED CARE SETTINGS
Hospitals, long term or extended care facilities
PRECAUTIONS AIRWEDGE
• Ensure the AirWedge meets the patient’s needs before using.
• A clinical assessment should be carried out by a qualied medical
professional before positioning patients using the AirWedge.
• Make sure patient is centered on the AirWedge before inating.
• Caregiver(s) must ensure the patient is attended and stabilized while
on the AirWedge.
• Always use a minimum of two caregivers when operating the AirWedge.
• Product should only be used by trained personnel.
• Only use attachments and/or accessories that are authorized by
HoverTech International.
PRECAUTIONS
HOVERTECH INTERNATIONAL AIR SUPPLY
• Not for use in the presence of ammable anesthetics or in a hyperbaric
chamber or oxygen tent.
• Route the power cord in a manner to ensure freedom from hazard.
• Avoid blocking the air intakes of the Air Supply.
• Use this product only for its intended purpose as described in this
manual.
• CAUTION: Avoid electric shock. Do not open HoverTech International
Air Supply.
• WARNING: Reference product specic user manuals for operating
instructions.
Symbol Reference
DECLARATION OF CONFORMITY
TO MEDICAL DEVICE DIRECTIVE

AWManual, Rev. B www.HoverMatt.com | 3
User Manual
HT-Air® Keypad Functions
Transfer
Speed
High
Adjustable/
Pressure
Setting
Transfer
Speed
Low
Standby (Amber LED)
Power IndicationAdjustable Speed
Indication
(Green Flashing LED)
Green LEDs
The ADJUSTABLE keypad function has four dierent settings.
Each press of the button increases the air pressure and rate
of ination. The Green Flashing LED will indicate the ination
speed by the number of ashes (i.e. two ashes equals the
second ination speed).
STANDBY: Used to stop ination/air ow (Amber LED indicates
STANDBY mode).
Part Identication – AirWedge® Adjustable Positioning Device
Air Filter
Cover
Hose
Nozzle
Hose
Hose Release Button
Control Panel Label
Carrying Handle/
Attachment Hook
Power
Cord
Strap
Power Cord
(US/UK/Euro)
WARNING: The HT-Air is not compatible with DC power supplies.
The HT-Air is not for use with the HoverJack Battery Cart.
Part Identication – HT-Air® Air Supply
Rail Straps (4)
Ination/
Deation Valve
(2)
Head Chamber
Chest Chamber
Securing Flap

User Manual
4 | HoverTech International AWManual, Rev. B
Instructions for Use
1. Center the AirWedge on the surface where it is to be used and secure
with rail straps.
2. For head and chest elevation, center patient on the AirWedge, with
the patient’s head close to the top edge of the pillow. (The AirWedge
can also be used to elevate legs, feet, etc. For these and other uses,
place the AirWedge as needed and follow the ination procedure.)
3. To begin ination, hold the HT-Air air supply hose against the head
ination valve. Press the ADJUSTABLE button on the HT-Air keypad
to initiate air ow and select the appropriate speed. Partially inate
head pillow. Move hose to chest ination valve and inate to desired
height. Then, nish inating head pillow as needed.
4. The rmness can be reduced by pressing the diaphragm in each
ination valve to slowly deate the chamber.
5. When the procedure is complete, deate the AirWedge by pressing
the diaphragm in each ination valve until both chambers are fully
deated.
Product Specications/Required Accessories
AIRWEDGE ADJUSTABLE POSITIONING DEVICE
Material: Double-coated polyurethane nylon twill
Construction: RF Welded
Width: 22" (56 cm) (Deated)
Length: 30" (76 cm) (Deated)
Model #: HTPP-200
Weight Limit: 800 LBS/363 KG
REQUIRED ACCESSORY:
Model #: HTAIR1200 (North American Version) – 120V~, 60 Hz, 10A
Model #: HTAIR2300 (European Version) – 230V~, 50 Hz, 6A
Model #: HTAIR1000 (Japanese Version) – 100V~, 50/60 Hz, 12.5A
Model #: HTAIR2356 (Korean Version) – 230V~, 50/60 Hz, 6A
LATEX FREE

AWManual, Rev. B www.HoverMatt.com | 5
User Manual
Cleaning & Preventive Maintenance
In between patient uses, the AirWedge should be wiped down with a
cleaning solution used by your hospital for medical equipment disinfection.
A 10:1 bleach solution (10 parts water: one part bleach) or disinfectant
wipes can also be used. It is important to follow the cleaning solution
manufacturer’s instructions for use, including dwell time and saturation.
NOTE: Cleaning with bleach solution may discolor fabric.
Do not launder the AirWedge.
INFECTION CONTROL
Whatever the patient is lying on to keep the surface clean may be placed
on top of the AirWedge to help keep it clean. If desired, a disposable sheet
may be used to cover the AirWedge (available for separate purchase).
If the AirWedge is used on an isolation patient, the hospital should employ
the same protocols/procedures it utilizes for the bed mattress and/or for
linens in that patient room.
PREVENTIVE MAINTENANCE
Prior to use, a visual inspection should be performed on the AirWedge
to ensure that there is no visible damage that would render the
AirWedge unusable.
The AirWedge should be periodically inspected to ensure the following:
• All ination/deation valves are self-sealing with no evidence of leakage.
• All rails straps are attached to the AirWedge.
• All snaps are intact and operational.
• There are no punctures or tears in the AirWedge.
If any damage is found that would cause the AirWedge not to function as
intended, the AirWedge should be removed from use and returned to
HoverTech International for repair.
Product Specications/Required Accessories

User Manual
6 | HoverTech International AWManual, Rev. B
Warranty Statement
The AirWedge Adjustable Positioning Device is warranted to be free from
defects in materials and workmanship for (1) one year. Warranty begins on
date of in-service by a HoverTech International representative or shipment
date. In the unlikely event that a problem arises as a result of a defect in
materials or workmanship, we will promptly repair your item or replace it
if we feel that it cannot be repaired – at our expense and discretion using
current models or parts performing the equivalent function – upon receipt
of the original item to our repair department.
You must pre-notify HoverTech International by phone (800-471-2776).
Item is to be insured by you against loss during transportation and must
be shipped with transportation and/or broker charges prepaid. Should any
HoverTech International product be returned that is not covered under
warranty, there will be a minimum $100 service charge plus shipping costs.
This warranty is not an unconditional guarantee for the life of the product.
Our warranty does not cover product damage that may result from use
contrary to manufacturer’s instructions or specications, misuse, abuse,
tampering, or damage due to mishandling. Equipment that has been
neglected, improperly maintained, repaired or altered by someone other
than an authorized representative of manufacturer, or operated in anyway
contrary to the operating instructions, shall void this warranty.
This warranty does not cover normal“wear and tear.”Component parts,
particularly any optional equipment, valve caps, their attachments and cords,
will show wear with use over time and eventually may need to be refurbished
or replaced. This normal type of wear is not covered by our warranty, but we
will provide prompt, high quality repair service and parts at a nominal cost.
HoverTech International’s liability under this warranty and on any claim of
any kind for any loss or damage arising out of, connected with, or resulting
from the design, manufacture, sale, delivery, installation, repair or opera-
tion of its products, whether in contract or tort, including negligence, shall
not exceed the purchase price paid for the product and upon expiration of
the applicable warranty period, all such liability terminates. The remedies
which this warranty provides are exclusive and HoverTech International
shall not be liable for any incidental or consequential damages.
There are no warranties, expressed or implied, which extend beyond this
warranty statement. The provisions of these warranty clauses are in lieu of
all other warranties, expressed or implied, and of all other obligations or
liabilities on HoverTech International’s part and they neither assume nor
authorize any other person to assume for HoverTech International any oth-
er liability in connection with manufacturer sale or lease of said products.
HoverTech International makes no warranty of merchantability or tness
for a particular purpose. By accepting the goods, the buyer acknowledges
that buyer has determined the goods are suitable for the buyer’s purposes.
MANUFACTURER’S SPECIFICATIONS ARE SUBJECT TO CHANGE.

AWManual, Rev. B www.HoverMatt.com | 7
User Manual
Returns and Repairs
All products being returned to HoverTech International (HTI) must have
a Returned Goods Authorization (RGA) number issued by the company.
Please call (800) 471-2776 and ask for a member of the RGA Team who will
issue you an RGA number. Any product returned without an RGA number
will cause a delay in the repair time.
HTI’s warranty covers manufacturer’s defects in material and workmanship.
If a repair is not covered under warranty, a minimum repair fee of $100 per
item will be assessed, plus return shipping. A Purchase Order for the repair
charge should be provided by the facility at the time the RGA number is
issued, should the repair not be covered under warranty. Lead-time for
repairs is approximately 1-2 weeks, not including shipping time.
Returned products should be sent to:
HoverTech International
Attn: RGA # ___________
4482 Innovation Way
Allentown, PA 18109

User Manual
8 | HoverTech International AWManual, Rev. B
4482 Innovation Way
Allentown, PA 18109
800.471.2776
Fax 610.694.9601
www.HoverMatt.com
Table of contents
Other Hovertech Medical Equipment manuals

Hovertech
Hovertech Hoverjack HJBSC-200 User manual

Hovertech
Hovertech HOVERMATT T-Burg User manual

Hovertech
Hovertech HoverJack User manual

Hovertech
Hovertech HoverJack User manual

Hovertech
Hovertech HoverSling User manual

Hovertech
Hovertech HoverJack User manual

Hovertech
Hovertech HOVERJACK HJ3202T User manual

Hovertech
Hovertech HTAIR-2300 User manual

Hovertech
Hovertech hovermatt User manual

Hovertech
Hovertech Q2 Roller User manual