manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hygiena
  6. •
  7. Laboratory Equipment
  8. •
  9. Hygiena PCD4000 User manual

Hygiena PCD4000 User manual

INS0063 JUL 2020 Rev. J
Hygiena™Calibration Control Kit
Calibration verification of Hygiena™luminometers
Part No# PCD4000 (Kit Includes: 1 Positive Rod and 1 Negative Rod)
Description/ Intended Use:
Calibration control rods are used to verify calibration of Hygiena luminometers.
Positive Rod consists of C14 radioactive source that emits very low level of low
energy βradiation in a plastic scintillation matrix. Matrix is configured to give
constant light output at a high level for up to five years. Negative Rod is used
to check for possible background light getting into instrument or light detector
not calibrating correctly. Negative Rod produces low-light (0–4) RLU
background. It is recommended to verify calibration with Calibration Control Kit
(PCD4000) each week or every other week depending on use and Quality
Control Program requirements. Incorporating Calibration Control Kit into
overall Quality Control program will validate luminometer is within
specifications and operating correctly.
Directions:
Positive and Negative Rods are inserted and read in luminometer in same
manner as ATP test devices. Calibration rods do not need to be activated
like ATP test devices. Do not activate Calibration rods.
Using PCD4000 with EnSURE™Touch:
1. Turn on EnSURE Touch and open the pink calibration application and
touch the “Check Calibration Now” button
2. Select the Calibration Control Kit option
3. Insert the black Negative Calibration Rod into the chamber. Push rod in
gently so it reaches bottom of chamber, close the lid, and touch the “Run
Test” button
4. Record the result and press “Next” to continue. The results are also
stored on the instrument and can be synced to SureTrend Cloud.
5. Select the Calibration Control Kit option and Insert the red Positive
Control Rod into the chamber. Push rod in gently so it reaches bottom of
chamber, close the lid, and touch the “Run Test” button
6. Record the result and press “Next” to return to the calibration screen.
The results are also stored on the instrument and can be synced to
SureTrend Cloud.
7. Repeat steps 1-6 for each EnSURE Touch,
Using PCD4000 with EnSURE or SystemSURE Plus:
1. Open chamber and insert the red Positive Rod. Push rod in gently so it
reaches bottom of chamber.
2. Close lid and press “OK” to initiate measurement.
3. Record result. Repeat steps 1 and 2 two more times and record results.
Calculate average of three readings and enter in table provided below.
Record Positive Rod Reference No. Refer to Interpretation of Results.
4. Repeat steps 1 through 3 using the black Negative Rod.
5. Repeat steps 1 through 4 for each luminometer. Record readings in table
provided below. Additional tables are printed on back side of this sheet.
Instrument Serial No.:
Positive Rod Reference:
Test 1
Test 2
Test 3
Average
Negative Rod
Test 1
Test 2
Test 3
Average
Interpretation of Results:
Refer to Table 1 below for acceptable RLU ranges. Results should not vary by
more than ± 20% from Positive Rod Reference No. If results from Positive Rod
are outside RLU range or vary by > ±20% from Positive Rod Reference No.,
contact your Hygiena representative.
Table 1: In-Calibration RLU Ranges for Hygiena Luminometers
EnSURE
Touch
(RLU)
EnSURE
(RLU)
SystemSURE Plus
(RLU)
Negative
Rod
0-4
0-4
0-4
Positive Rod
130-230
80-160
40-80
Incorporating Calibration Control Kit into a Quality Control Program:
Calibration Control Kit is designed to be incorporated into a Quality Control
Program that monitors and tracks performance of Hygiena luminometers
and/or devices. EnSURE Touch tracks calibration checks in the Calibration
app and in SureTrend Cloud. For EnSURE and SystemSURE Plus, it is
recommended that a specific location be programmed in SureTrend for all
calibration checks to be performed and results to be stored. Results can be
viewed on the luminometer or in SureTrend.
Troubleshooting:
Several factors can cause RLU readings to be outside acceptable range.
Possible causes include:
•Dirty calibration rods. Clean outside of both rods with lint-free cloth that
has been dipped in isopropyl or ethyl alcohol. Air dry and repeat reading.
•Damaged calibration rods. Return damaged calibration rods to
Hygiena for replacement.
•Luminometer contamination. Results that have changed suddenly,
especially immediately after a number of sampling devices have been
run, can indicate contamination problems. Results from calibration rods
can increase or decrease as a result of foreign materials disturbing
optics chamber of luminometer. If contamination is suspected in
luminometer, remove read chamber and clean. On EnSURE Touch, you
can access the read chamber by unscrewing the cap at the bottom.
•Position of rod. It is important to make sure calibration rods are placed
in read chamber so rod touches bottom of chamber. If rod does not
reach bottom of chamber, rod will be too far from sensor, and will result
in inaccurate results (low measurements).
Storage & Shelf Life:
•Five (5) year shelf life. Refer to expiration date printed on label.
•Store at 15 –25 °C (59 –77 °F) in box rods were received in, or dark
container. Store away from light.
INS0063 JUL 2020 Rev. J
Hygiena™Calibration Control Kit
Calibration verification of Hygiena™luminometers
Part No# PCD4000 (Kit Includes: 1 Positive Rod and 1 Negative Rod)
Precautions:
•Limit exposure to light. Store calibration rods in dark or weak ambient
light. Kit box provided is recommended as appropriate dark storage. Do
not expose to direct sunlight or intense artificial light for too long before
use.
•Handle with care. Avoid crushing, cutting, or dropping calibration rods.
Any damage that changes shape or light transmission properties of
Positive Rod can affect assigned values. Do not use if scintillator tip of
device is cracked or damaged.
•Radioactive material. Not for human use. Introduction into foods,
beverages, cosmetics, drugs, or medicinal products, or into products
manufactured for commercial distribution is prohibited. Exempt quantities
should not be combined.
•For further safety instruction, refer to Calibration Control Kit Safety Data
Sheet (SDS).
Disposal:
•U.S. & Canada: Level of radioactive material in Positive Rod does not
require special hazardous waste treatment, per U.S. Federal regulations
governing use and transfer of exempted radioactive materials. Contents
of Positive Rod are exempt from NRC or Agreement State licensing
requirements. Positive Rod contains extremely small amounts of
radioactive Carbon-14, completely contained by outer housing of device.
There are no radioactive emissions on surface of device, and no special
precautions are needed in handling device to avoid exposure.
•To dispose of Positive Rod, remove or deface “Caution Radioactive
Materials” label on rod. Positive Rod may then be disposed in trash.
•For disposal requirements in other countries, check with your national or
local supplier.
NOTE: Make copies of this page as needed if more luminometers are in use.
Instrument Serial No.:
Instrument Serial No.:
Positive Rod Reference No.:
Positive Rod Reference No.:
Test 1
Test 2
Test 3
Average
Test 1
Test 2
Test 3
Average
Negative Rod
Negative Rod
Test 1
Test 2
Test 3
Average
Test 1
Test 2
Test 3
Average
Instrument Serial No.:
Instrument Serial No.:
Positive Rod Reference No.:
Positive Rod Reference No.:
Test 1
Test 2
Test 3
Average
Test 1
Test 2
Test 3
Average
Negative Rod
Negative Rod
Test 1
Test 2
Test 3
Average
Test 1
Test 2
Test 3
Average
Instrument Serial No.:
Instrument Serial No.:
Positive Rod Reference No.:
Positive Rod Reference No.:
Test 1
Test 2
Test 3
Average
Test 1
Test 2
Test 3
Average
Negative Rod
Negative Rod
Test 1
Test 2
Test 3
Average
Test 1
Test 2
Test 3
Average
INS0063 JUL 2020 Rev. J
Hygiena™Calibration Control Kit
Calibration verification of Hygiena™luminometers
Part No# PCD4000 (Kit Includes: 1 Positive Rod and 1 Negative Rod)
Hygiena Liability:
Hygiena will not be liable to user or others for any loss or damage whether
direct or indirect, incidental or consequential from use of this device. If this
product is proven to be defective, Hygiena’s sole obligation will be to
replace product or at its discretion, refund the purchase price. Promptly
notify Hygiena within 5 days of discovery of any suspected defect and
return product to Hygiena. Please contact Customer Service for a Returned
Goods authorization number.
Contact Information:
If more information is required, please visit us at www.hygiena.com or
contact us at:
Hygiena —Americas
Phone: 1.805.388.8007
E-mail: info@hygiena.com
Hygiena —International
Phone: +44 (0)1923 818821
E-mail: enquiries@hygiena.com

Other Hygiena Laboratory Equipment manuals

Hygiena Scigiene DB625 User manual

Hygiena

Hygiena Scigiene DB625 User manual

Hygiena SuperSnap SUS3000X User manual

Hygiena

Hygiena SuperSnap SUS3000X User manual

Hygiena EnSURE Touch User manual

Hygiena

Hygiena EnSURE Touch User manual

Hygiena BAX System Q7 Instruction Manual

Hygiena

Hygiena BAX System Q7 Instruction Manual

Hygiena Autosampler III User manual

Hygiena

Hygiena Autosampler III User manual

Hygiena BAX System X5 User manual

Hygiena

Hygiena BAX System X5 User manual

Hygiena INCUBATOR User manual

Hygiena

Hygiena INCUBATOR User manual

Hygiena RiboPrinter User manual

Hygiena

Hygiena RiboPrinter User manual

Hygiena BAX Q7 User manual

Hygiena

Hygiena BAX Q7 User manual

Hygiena Dualo 32 R2 Instruction sheet

Hygiena

Hygiena Dualo 32 R2 Instruction sheet

Hygiena Innovate Installation manual

Hygiena

Hygiena Innovate Installation manual

Hygiena DUP-1000 User manual

Hygiena

Hygiena DUP-1000 User manual

Hygiena foodproof KIT230127 User manual

Hygiena

Hygiena foodproof KIT230127 User manual

Hygiena Dualo 32 Instruction sheet

Hygiena

Hygiena Dualo 32 Instruction sheet

Hygiena UltraSnap US2020 User manual

Hygiena

Hygiena UltraSnap US2020 User manual

Popular Laboratory Equipment manuals by other brands

Hall Research Technologies 511-POH-30W user manual

Hall Research Technologies

Hall Research Technologies 511-POH-30W user manual

Heidolph Smart S-biosystems 801620 operating instructions

Heidolph

Heidolph Smart S-biosystems 801620 operating instructions

Beckman Coulter FP 1000 manual

Beckman Coulter

Beckman Coulter FP 1000 manual

BRIGHT M3500 MICROTOME Operating, Maintenance and Spare Parts Manual

BRIGHT

BRIGHT M3500 MICROTOME Operating, Maintenance and Spare Parts Manual

Live Photonic Sequencer LPS-200H user manual

Live Photonic

Live Photonic Sequencer LPS-200H user manual

Sheldon Shel-Lab SVAC1-2 Installation and operation manual

Sheldon

Sheldon Shel-Lab SVAC1-2 Installation and operation manual

Novasina LabMaster-aw operating instructions

Novasina

Novasina LabMaster-aw operating instructions

HARVEY Chemiclave EC5500 Service manual

HARVEY

HARVEY Chemiclave EC5500 Service manual

Optika Italy IM Series instruction manual

Optika Italy

Optika Italy IM Series instruction manual

WILKINS ZW209 Installation, Troubleshooting, Maintenance Instructions

WILKINS

WILKINS ZW209 Installation, Troubleshooting, Maintenance Instructions

Tecno-gaz EkstraPLUS PT002TTM Instructions for use

Tecno-gaz

Tecno-gaz EkstraPLUS PT002TTM Instructions for use

ibidi Heating System instructions

ibidi

ibidi Heating System instructions

TRENDnet TPE-113GI Quick installation guide

TRENDnet

TRENDnet TPE-113GI Quick installation guide

VPixx Technologies X-Rite i1Display Pro user manual

VPixx Technologies

VPixx Technologies X-Rite i1Display Pro user manual

Terragene Trazanto manual

Terragene

Terragene Trazanto manual

IKA RC 5 basic operating instructions

IKA

IKA RC 5 basic operating instructions

cytiva ReadyToProcess Mixer 20/50 operating instructions

cytiva

cytiva ReadyToProcess Mixer 20/50 operating instructions

Cleaver Scientific CSL-CELLAS instruction manual

Cleaver Scientific

Cleaver Scientific CSL-CELLAS instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.