IACER Srl 15 MNPG117-02
•to use the device in presence of signs of deterioration of the device itself,
cables and/or accessories: please contact the dealer or the manufacturer
following the instructions given in the paragraph Support. Control
carefully the integrity of the device before each use;
•to use the device close to flammable substances/gas/explosives, in
environments with high concentrations of oxygen, with aerosol-therapy
devices or in wet environments (use of the device is prohibited in
bathroom or shower areas or while showering/bathing);
•absolutely use a device that has been wet or has come into contact with
liquids before it has been checked by the manufacturer and / or service
center. Take care to prevent liquids from entering the ventilation slots;
•to use the device while driving or during the operation and control of
equipment/machinery;
•the use of the device in hyposensitive areas, on the carotid sinuses
(carotid), genitals, near the uterus and abdomen, in the area of the body
in which glands are present, on cancerous lesions. Do not direct the beam
towards or near the eyes;
•maintain the ultrasound head stationary on one point during the
therapy. We advise moving the head if the intensity is more than 0,5
W/cm²;
•to use pointed or sharp objects on the ultrasound head and the control
panel of the device.
Warning:
•before administering any treatment to a patient, you should become
acquainted with the operating procedures for each mode of treatment
available, as well as the indications, contraindications, warnings and
precautions. Consult other resources for additional information regarding
the application of Ultrasound;
•ultrasound should be routinely checked before each use to determine
that all controls function normally, especially that the intensity control
does properly adjust the intensity of the ultrasonic power output in stable
manner. Also, determine that the treatment time control does actually
terminate ultrasonic power output when the timer reaches zero;
•handle the handpiece with care to preserve its characteristics;
•the tendency to bleed is increased by the heat as more blood flows in the
area. Be careful when treating patients with bleeding disorders;
•avoid heating or overheating the capsule in cases of acute and subacute
arthritis.;
•pay attention to use connection cables with children/young people:
strangulation danger;