IKA C 200 User manual

Calorimeter System C 200
Operating instructions EN

3
Source language: German EN
Chapter Page
Declaration of conformity 5
1 Safety instructions 6
2 User information 8
2.1 Information regarding operating instructions 8
2.2 Warranty and liability 8
3 Transport, storage, place of installation 9
3.1 Conditions of transport and storage 9
3.2 Place of installation 9
3.3 Unpacking 9
3.4 Delivery scope 9
4 Installations and starting up 10
4.1 Calorimeter C 200 10
4.2 Installation 10
4.3 Switchin on the system 12
4.4 Display and operating elements 12
4.5 Configuring the system 13
4.6 System settings 14
4.7 Filling the calorimeter for the first time 15
4.8 Switchin off the system 16
4.9 Coding decomposition vessel 16
4.10 Oxygen station C 248 16
5 Calorimetric measurements 17
5.1 Determining the calorific value 17
5.2 Corrections 17
5.3 Information about the sample 18
5.4 Calibration 19
6 Preparing and performing measurements 19
6.1 Decomposition vessel C 5010 19
6.2 Preparing the decomposition vessel 20
6.3 Preparing the measurement 21
6.4 Performing the measurement 22
6.5 Cleaning the decomposition vessel 23
6.6 Errors in the measuring procedure 24
Contents

4 5
7 Service menue 26
7.1 Operation 26
7.2 Description of service menue options 26
8 Cleaning and maintenance 27
8.1 Inner vessel filter 27
8.2 Filler 27
8.3 Micro filter 27
8.4 Maintaining the water circulation 28
8.5 Decomposition vessels 28
8.6 Cleaning informations 28
9 IKA - Accessories and consumables 29
9.1 Accessories 29
9.2 Consumables 29
10 Technical data 30
Declaration of conformity
KONFORMITÄTSERKLÄRUNG DE
Wir erklären in alleiniger Verantwortung, dass dieses Produkt den Bestimmungen der Richtlinien 2014/35/EU,
2014/68/EU, 2014/30/EU und 2011/65/EU entspricht und mit den folgenden Normen und normative Dokumenten über-
einstimmt: EN 61010-1, EN 61010-2 und EN 61326-1.
Eine Kopie der vollständigen EU-Konformitätserklärung kann bei [email protected] angefordert werden.
DECLARATION OF CONFIRMITY EN
We declare under our sole responsibility that this product corresponds to the regulations 2014/35/EU, 2014/68/EU,
2014/30/EU and 2011/65/EU and conforms with the standards or standardized documents: EN 61010-1, EN 61010-2
and EN 61326-1.
DÉCLARATION DE CONFORMITÉ FR
Nous déclarons sous notre responsabilité que se prodiut est conforme aux réglementations 2014/35/UE, 2014/68/UE,
2014/30/UE et 2011/65/UE et en conformité avec les normes ou documents normalisés suivant: EN 61010-1,
EN 61010-2 et EN 61326-1.
Une copie de la déclaration de conformité UE complète peut être demandée en adressant un courriel à l’adresse [email protected].
DECLARACION DE CONFORMIDAD ES
Declaramos por nuestra responsabilidad propia que este producto corresponde a las directrices 2014/35/UE,
2014/68/UE, 2014/30/UE y 2011/65/UE y que cumple las normas o documentos normativos siguientes: EN 61010-1,
EN 61010-2 y EN 61326-1.
Si lo desea, puede solicitar una copia completa de la declaración de conformidad de la UE en la dirección de correo
electrónico [email protected].
DICHIARAZIONE DI CONFORMITÀ IT
Dichiariamo, assumendone la piena responsabilità, che il prodotto è conforme alle seguenti direttive 2014/35/UE,
2014/68/UE, 2014/30/UE e 2011/65/UE, in accordo ai seguenti regolamenti e documenti: EN 61010-1, EN 61010-2 e
EN 61326-1.
Una copia della dichiarazione di conformità UE completa può essere richiesta all’indirizzo [email protected].

6 7
Safety instructions
der such as splints, hay, straw etc. explode
when combusted! Always wet these materi-
als first! Highly flammable liquids with a low
vapour pressure (e.g. tetramethyl dihydrogen
disiloxane) must not directly touch the cotton
thread!
• Furthermore, toxic combustion residue in the
form of gases, ashes or condensation, for ex-
ample, is possible in the inner wall of the de-
composition vessel.
• When handling combustion samples, com-
bustion residue and auxiliary materials, please
observe the relevant safety regulations. The
following materials, for example, could pose a
risk:
– corrosive
– highly flammable
– explosive
– bacteriologically contaminated
– toxic
• Please observe the relevant regulations when
handling oxygen.
Warning: oxygen as a compressed gas is
oxidising; intensively aids combustion; can
react violently with flammable materials.
Do not use oil or grease!
• Caution - magnetism! Effects of the mag-
netic field have to be taken into account (e.g.
data storage media, cardiac pacemakers ...).
• When using stainless steel crucibles tho-
roughly check their condition after each expe-
riment. If the material gets thinner, the crucible
may catch fire and damage the decomposition
vessel. Crucibles must not be used for more
than 25 combustions for safety reasons.
• The decomposition vessel is manufactured
in accordance with the directive for pressure
equipment 97/23/EC.
This is indicated by the CE symbol with the ID
number of the notified body. The decompositi-
on vessel is a category III pressure device. The
decomposition vessel has undergone an EC
prototype test. The CE declaration of confor-
mity confirms that this decomposition vessel
corresponds to the pressure device described in
the EC prototype test certificate. The decompo-
sition vessel has undergone a pressure test
For your protection
• Read the operating instructions in full
before starting up and follow the safety
instructions.
• Keep the operating instructions in a place whe-
re they can be accessed by everyone.
• Ensure that only trained staff work with the ap-
pliance.
• Follow the safety instructions, guidelines,
occupational health and safety and acci-
dent prevention regulations.
• Wear your personal protective equipment
in accordance with the hazard category
of the medium to be processed.
• Calorimeter system C 200 may only be used to
determine the calorific value of solid and flu-
id materials. Only the appropriate original IKA
decomposition vessels may be used for this
purpose. For detailed information, please see
the operating instructions for the decompositi-
on vessels.
• The maximum extra energy added to the de-
composition vessel must not exceed 40,000 J
(select the test mass accordingly). The permit-
ted operating pressure of 230 bar must not be
exceeded. The maximum permitted operating
temperature must not exceed 50 °C.
•
Do not overfill the decomposition vessel with
sample. Only fill the decomposition vessel with
oxygen to a maximum pressure of 40 bar.
Check the set pressure on the pressure reducer.
Perform a leakage test before each combustion
process (follow operating instructions for de-
composition vessel!).
• Some materials tend to explode when combus-
ted (e.g. due to formation of peroxide), which
could cause the decomposition vessel to crack.
• Standard decomposition vessels may not
be used for testing explosive samples.
• If the burning behaviour of a material is unk-
nown, it must be tested before combustion in
the decomposition vessel (risk of explosion). If
you are burning unknown samples, leave the
room or keep your distance from the calorime-
ter.
• Benzoic acid may only be combusted in its
pressed form! Flammable dust and powder
must be first pressed. Oven-dry dust and pow
• For the purposes of these operating instruc-
tions a technical expert is someone
1.
who guarantees to conduct the tests properly
on the basis of his training, knowledge and ex-
perience gained through practical work,
2. who is sufficiently reliable,
3. who is not subject to any instructions in terms
of the test activity,
4. who has suitable test equipment if necessary,
5. who can provide appropriate proof of the re-
quirements listed in 1.
• National directives and laws must be observed
for operating pressure vessels!
•
Anyone operating a pressure vessel must keep it
in a proper condition, operate it properly, super-
vise it, carry out the necessary maintenance and
repair work immediately and implement the sa-
fety measures required in the circumstances.
• Pressure vessels must not be used if they have
defects which could pose a risk to staff or third
parties. The pressure equipment directive can
be obtained from Carl Heymanns or Beuth pu-
blishers.
•
The C 248 oxygen station must be set up at least
1.5 m far away from the calorimeter.
For protection of the equipment
• The voltage stated on the type plate must cor-
respond to the mains voltage.
• Removable parts must be refitted to the appli-
ance to prevent the infiltration of foreign ob-
jects, liquids etc.
• Protect the appliance and accessories from
bumps and impacts.
with test pressure of 330 bar and a leaka-
ge test with oxygen at 30 bar.
• Decomposition vessels are experiment autocla-
ves and must be tested by a technical expert
after each use.
Individual use is understood here to include a
series of experiments performed under roughly
the same conditions in terms of pressure and
temperature. Experiment autoclaves must be
operated in special chambers (C 2000, C 5000,
C 7000, C 200).
• The decomposition vessels must undergo re-
peated tests (internal tests and pressure tests)
performed by the technical expert. The fre-
quency of these tests is to be determined by
the operator on the basis of experience, type of
operation and the material used in the decom-
position vessel.
• The declaration of conformity becomes
invalid if mechanical modifications are
carried out to the experiment autoclaves
or if tightness can no longer be guaran-
teed as a result of major corrosion (e.g.
pitting by halogens).
• The threads on the body of the decomposition
vessel and the union nut are subject to conside-
rable stress in particular and must therefore be
checked regularly for wear.
• The condition and function of the seals must
be checked and ensured by way of a leakage
test (see operating instructions for decomposi-
tion vessel!).
• Only technical experts may perform pressure
tests and service work on the decomposition
vessel.
• We recommend that you send the decom-
position vessel to our factory for inspec-
tion, and if necessary, repair after 1000
tests or after one year or sooner depen-
ding on use.

8 9
User information
General hazard
This symbol indicates information which is essential for the safety of
your health. Failure to observe this information can cause damage to
health and injuries.
This symbol indicates information which is important for ensuring that
the appliance functions without any technical problems. Failure to
observe this information could damage the calorimeter system.
This symbol indicates information which is important for ensuring that calori-
metric measurements are performed eciently and for using the calorimeter
system. Failure to observe this information can result in inaccurate measu-
rements.
This symbol indicates references to the optimization of operatinal sequences
are characterized.
Attention - Risk of damage due to magnetism.
Numbers , , etc. indicate guidelines in the following sections which
must always be carried out in order.
TIP
Please read these operating instructions carefully.
IKA- WERKE only consider themselves to be
responsible for the safety, reliability and perfor-
mance of the appliance when
• the appliance has been operated in accordance
with the operating instructions,
• only persons authorised by the manufacturer
interfere with the appliance,
•
only original parts and original accessories are
used for repairs.
IKA- WERKE do not accept liability for damage
or costs resulting from accidents, misuse of the
appliance or unauthorised changes, repairs or
upgrades.
2.2 Warranty and liability
You have purchased an original IKA product,
manufactured to the latest, highest quality stan-
dards. According to IKA's conditions of warran-
ty, this product is guaranteed for 12 months.
To ensure the long-term precision and function
of the calorimeter system, we recommend that
you conclude a maintenance agreement (annu-
al maintenance) with IKA or an authorised IKA
workshop. If the first maintenance is carried out
within 12 months of purchase, the warranty will
be extended to 24 months.
In the case of a warranty claim, please contact
the responsible representative or the supplier.
You can also return the appliance directly to IKA-
WERKE.
Please enclose the sales invoice and the reason
for the claim and state the name of the contact
person at your company. Freight charges are to
be paid by the sender.
2.1 Information regarding operating instructions
Transport, storage, place of installation
• The (constant) room temperature should be
around 20 °C ... 25 °C
• The system must be installed on a level surface.
• An adequate power supply corresponding to
the nameplates on the system components.
• Oxygen supply (99.95 % pure oxygen, quality
3.5; pressure 30 bar) with pressure display and
shut-off device (C 29 reducing valve, acces-
sories)
1 x Power cable
1 x Operating instructions
1 x Water emptying hose (length: 1 m)
1 x Oxygen station C 248
1 x Measuring cup (2 l)
Please observe the respective country-specific re-
gulations for operating pressure equipment when
installing the appliance. A constant ambient tem-
perature is an important requirement for ensuring
the high measuring accuracy of the system. The
following conditions must therefore be fulfilled at
the place of installation:
• No direct solar radiation
• No draughts (e.g. beside windows, doors, air
conditioning)
• Sufficient distance to radiators and other heat
sources
• Please unpack the device carefully
• In the case of any damage a report must be
sent immediately (post, rail or forwarder)
1 x Basic device C 200
1 x Decomposition vessel C 5010
1 x Ignition adaptor
1 x Attachment set
1 x Table power supply
3.1 Conditions of transport and storage
3.2 Place of installation
3.3 Unpacking
3.4 Delivery scope
• Only the original packing may be used for
transportation.
• The appliance must be completely emptied be-
fore storing and transportation.
• The system must be protected against mecha-
nical impact, vibrations, dust deposits and cor-
rosive ambient air during transportation and
storage.
• It is also important to ensure that the relative
humidity does not exceed 80%.

10 11
If you use a thermostat to fill the tank, you can
also hang the drain hose in the thermostat. Check
that the thermostat is below the calorimeter. If
the thermostat is self-priming in the return line,
you can also install the thermostat beside the ca-
lorimeter.
Installation and starting up
4.1 Calorimeter C 200
4.2 Installation
Measuring cell cover
Tank fillers
Display
Keyboard
Fill level display
Connecting peripherals
When connecting peripherals make sure that
both they and the calorimeter are switched off at
ON/OFF switch .
All the connections for draining as well as the pe-
ripherals are on the back of the appliance.
Connecting the table power supply
Connect the calorimeter to the table power sup-
ply (4-pin plug ). Check that the voltage infor-
mation on the rating plate of the power supply
unit matches your mains supply. Connect the po-
wer line of the plug to the voltage source.
The unit is ready for service
when the mains plug has
been plugged in.
PC connector
Printer connector
ON/OFF switch
Table power supply connector
Table power supply
Connecting the water drain
Insert the emptying hose (included in package
contents) into plug-in coupling . Place this in
the mould so that it is sloping. This must always
be connected for operation.
If the inner vessel ever requires emergency drai-
nage, insert the emptying hose into plugin cou-
pling . To empty the tank container, insert the
emptying hose into plug-in coupling (see ser-
vice menu section 7.2).
Inner vessel
emergency drain
Tank drain
Inner vessel drain
TIP
Opening the calorimeter cover:
The calorimeter cover is opened by hand by lifting the cover by the gripping groove and raising it until
it automatically swivels to the right and locks in position.
Raise Swivel to right

12 13
Switch the calorimeter on at ON/OFF switch
(back of appliance). The appliance is now in
standby mode.
Press ON (F1) to work with the appliance. The
start screen will appear.
The operating console features the following ele-
ments:
Display elements during operation:
Status line: shows the current status of the
appliance.
Footer: shows what the function keys cur-
rently do.
Exception: a progress bar is displayed during
automatic measuring runs.
Reading value: shows the current tempera-
ture increase in minute intervals during the
measurement or the measurement result after
the measurement.
Current temperature value: shows the cur-
rent temperature in second intervals at the
sensor in the inner vessel.
Operating elements:
Function keys F1, F2, F3: what these keys
do depends on the operating state of the ap-
pliance. The footer of the display shows what
the function keys currently do.
Die Fußzeile des Displays zeigt die aktuelle Be-
legung der Funktionstasten an.
Numerical keypad: this is used to enter digits
and the decimal point into command lines.
Delete button: this key is used to delete the
last character entered.
Display elements in menu mode:
If you press the Menu (F3) key, a menu will ap-
pear in the display which allows you to enter set-
tings. There are six submenus:
GENERAL
CALIBRATION VALUES
UNIT OF MEASUREMENT
LANGUAGE
MEASURING PROCEDURE
SERVICE
Pressing the DOWN (F2) key selects the menu
line below. Double clicking (F2) allows you to
switch between UP and DOWN. The current di-
rection of movement is displayed in the footer.
BACK (F3) allows you to exit a menu without
confirming new settings.
Selection menu (e.g. combustible crucible)
4.3 Switching on the system
4.4 Display and operating elements
TIP
Menu header:
shows the name of the submenu and the cur-
rently selected menu line of the total number of
menu lines (for example, 2/5: you are in the se-
cond of five lines).
There are three different menu lines , and :
Submenu line:
“. . .“ this line indicates that a submenu is called
up with OK (F1). “. . . :“ indicates that it is a
selection menu.
If you press OK (F1) on a selection line in this
selection menu, this selection will be confirmed
and you will exit the menu. The current selection
is marked with ” = ” and is displayed after the ”:”
in the submenu line.
Command line:
this line allows you to directly enter numerical va-
lues using the keyboard. If you press OK (F1),
the menu settings will be saved and you will exit
the menu.
4.5 Configuring the system
In order to ensure that the appliance works
properly, you must set some parameters the
first time you use it.
Select language
MENU (F3)
UP/DOWN (F2) to "Language"
OK (F1)
UP/DOWN (F2) select desired language
(default: English)
OK (F1)
BACK (F3)
The set language will appear in the "Language"
submenu line.
Set date
MENU (F3)
UP/DOWN (F2) to "General"
OK (F1)
UP/DOWN (F2) to "Date"
Enter the date in the format dd.mm.yyyy
(e.g: 06.12.2005)
OK (F1)
BACK (F3)
Set time
MENU (F3)
UP/DOWN (F2) to "General"
OK (F1)
UP/DOWN (F2) to "Time"
Enter the time in the format hh:mm:ss
(e.g: 14:29:56)
OK (F1)
BACK (F3)
Reference
In order to calibrate the appliance you must state
the exact calorific value of the calibration subs-
tance used (usually benzoic acid).

14 15
MENU (F3)
UP/DOWN (F2) to "General"
OK (F1)
Enter the calorific value in the format xxxxx
(default: 26460)
OK (F1)
BACK (F3)
Calibration values
Once you have calibrated the appliance, you will
need to enter the calculated C-values (calibration
values) of all the decomposition vessels used.
MENU (F3)
UP/DOWN (F2) to "Calibration values"
OK (F1)
UP/DOWN (F2) select desired decompositi-
on vessel
Enter the C-value in the format xxxx
(default: 1)
OK (F1)
BACK (F3)
See the advice in section 5.4 Calibration.
4.6 System settings
In addition to the configuration described in sec-
tion 4.5, you can apply other settings which are
not necessary for the correct functioning of the
appliance or which are only required for special
applications.
Combustible crucible
If you are using a combustible crucible, you can
enter this here. The value for the external energy
QExternal1 will then be automatically reduced by
50 joules as no cotton thread is used. The com-
bustible crucible itself must be weighed and the
resulting calculated energy value manually ente-
red under QExternal2 (see section . 6.3, ) so
that it can be taken into account as external ener-
gy when calculating the calorific value.
MENU (F3)
UP/DOWN (F2) to "General"
OK (F1)
UP/DOWN (F2) to "Combustible crucible"
OK (F1)
Select the options "with" oder "without"
(default: without)
OK (F1)
BACK (F3)
Operator
You can assign each operator an ID number (0-9).
This number will appear on the results report.
MENU (F3)
UP/DOWN (F2) to "General"
OK (F1)
UP/DOWN (F2) to "Operator"
Enter ID code (default: 0)
OK (F1)
BACK (F3)
Unit of measure
You can set the unit of measure you want the
measuring results to be displayed in. The current-
ly selected unit of measure is marked with "=".
Time control
Calorimeter C 200 performs the measurement
automatically according to a set period of time
and calculates the provisional result. The measu-
ring time is set at 14 minutes.
4.7 Filling the calorimeter for the first time
Before using the calorimeter for the first time
you must fill the outer vessel with tap water.
Destilled and/or deactveted water may not be
used!
To do this, pour two litres of tap water into the
filler of tank (see section 4.1) using the measu-
ring cup provided.
The water must first be maintained at a constant
temperature. To ensure accurate results, the initi-
al temperature must not fluctuate too much.
Water temperature 18 °C - 25 °C with an ac-
curacy of ±1 °C during a measurement series.
You must now pump the water out of the tank
into the outer vessel:
MENU (F3)
1st FILL (F2)
Pumping procedure is started
Keep an eye on the water by pulling the filler of
tank (see section 4.1) out the top.
When the overflow causes the water to run back
in the side of the tank, the outer vessel is full and
you must switch off the pump again by pressing
STOP OK (F2) once more.
You should change the water if it has been sitting
for a long time, see section 8.3.
MENU (F3)
UP/DOWN (F2) to "Unit"
OK (F1)
UP/DOWN (F2) select the desired unit
(default: J/g)
OK (F1)
BACK (F3)
Measuring procedure
You can choose between four measuring proce-
dures:
MENU (F3)
UP/DOWN (F2) to "Measuring procedure"
OK (F1)
UP/DOWN (F2) select desired measuring
procedure
(default: Isoperibol)
OK (F1)
BACK (F3)
Isoperibol
Calorimeter C 200 automatically performs the
measurement according to the standard for iso-
peribol calorimeters and calculates the provisional
result for you. The measurement takes approx. 17
minutes with excellent result reproducibility.
Manual
This is ”student mode”. You work according to
the standard for isoperibol / isothermal calorime-
ters. You read the temperature values off yourself
every minute, calculate the temperature drift be-
fore ignition, ignite and end the experiment by
pressing F1, calculate the temperature drift after
the temperature increase and calculate the result
itself.
Dynamic
Calorimeter C 200 performs the measurement
automatically and calculates the provisional re-
sult. The measuring time is reduced to approx. 8
minutes thanks to a dynamic correction process.
It is still possible, however, to meet the accuracy
specifications of the international standards.
TIP

16 17
4.10 Oxygen station C 248
The C 200 does not have an integrated oxygen
filling point for the decomposition vessel. You
can use oxygen station C 248 for filling.
The C
248 oxygen station must be set up at least 1.5 m
far away from the calorimeter.
For information on
how to operate and connect the oxygen station,
please see the accompanying operating instruc-
tions.
Operation:
• Place decomposition vessel C 5010 on the
marked position.
• Move the lever down and centre the de-
composition vessel below the fill head .
• Lock the lever in place.
•
Fill the decomposition vessel for approx. 30
secs.
• Move the lever back to its original position.
4.8 Switching off the system
Standby mode
In order to switch off the calorimeter system, the
start screen must be displayed.
Press OFF (F1). The appliance will switch to
standby mode.
Switching off
Only switch the appliance off in standby mode.
To do this, press the ON/OFF switch (see sec-
tion 4.2) on the back of the appliance.
4.9 Coding decomposition vessels
You can use several decomposition vessels (max.
4) when working with calorimeter C 200. You
can code these to distinguish between them by
sticking the black coding strips into the special
slots on the decomposition vessels.
TIP
Slot for
coding strips
1 2
34
Decomposition vessel no.:
without coding
Lever
Fill head
In many cases, however, not just the combus-
tion products referred to in the standards are
produced. In such cases the fuel sample and the
combustion products must be analysed to pro-
vide data for a revised calculation. The standard
calorific value is then calculated from the measu-
red calorific value and the analysis data. The heat
value Hu is the same as the calorific value, minus
the condensation energy of the water contained
in the fuel and formed through combustion. The
heat value is the more important parameter from
a technical point of view because in all major,
technical applications only the heat value can be
evaluated in terms of energy.
The complete bases of calculation for the calori-
fic and heat value can be found in the relevant
standards (e.g.: DIN 51 900; ASTM D 240; ASTM
D 5865..). They are also contained in the CalWin
calorimeter software.
5.2 Corrections
Due to the nature of the system a combustion
test does not just produce the combustion heat
of the sample, but also heat from external energy
(QExternal).
This can fluctuate considerably in relation to the
heat quantity of the fuel sample.
The combustion heat of the cotton thread which
ignites the sample and the electric ignition energy
would distort the measurement. This influence is
taken into account in the calculation by way of a
correction value.
5.1 Determining the calorific value
Combustion is carried out in a calorimeter under
specific conditions. The decomposition vessel is
filled with a weighed fuel sample, the fuel sample
is ignited and the temperature increase in the ca-
lorimeter system measured. The specific calorific
value of the sample is calculated as follows:
Ho = (C * DT - QExt1 - QExt2) / m (1)
m
Weight of fuel sample
C
Heat capacity (C-value) of calorimeter system
DT
Calculated temperature increase of water in inner
vessel of measuring cell
QExt1
Correction value for the heat energy generated
by the cotton thread as ignition aid
QExt2
Correction value for the heat energy from other
burning aids
The decomposition vessel is filled with pure oxy-
gen (99.95 %) to optimise the combustion pro-
cess. The pressure of the oxygen atmosphere in
the decomposition vessel is max. 30 bar. Formula
(1) for the calorific value of a material requires
that combustion takes place under specifically
defined conditions. The relevant standards are
based on the following assumptions:
• The temperature of the fuel and its combustion
products is 25 °C.
• The water contained in the fuel before combus-
tion and the water formed whilst combusting
the hydrogenous compounds of the fuel is in
fluid form after combustion.
• The atmospheric nitrogen has not oxidised.
• The gaseous products after combustion consist
of oxygen, nitrogen, carbon dioxide and sulp-
hur dioxide.
• Solid materials may form (e.g. ashes).
Calorimetric measurements

18 19
Failure to observe these instructions could re-
sult in damage to the decomposition vessel.
Exploding decomposition vessels present a risk
of serious injury or death. When working with
unknown substances, select very small weighted
samples at the start in order to determine the na-
tural energy.
Normally solid materials in powder form can be
combusted directly. Materials which com-
bust quickly (e.g. benzoic acid) must not
be burnt loose. These materials tend to splash
and there is therefore no guarantee of complete
combustion.
Furthermore, this can damage the inner
wall of the decomposition vessel.
IKA pelleting press C 21 and IKA analytical mill A
11 basic (see Accessories) are available for sample
preparation.
Materials which are dicult to burn (materials
with a high mineral content, low caloric materi-
als) can often only be fully combusted using IKA
acetobutyrate capsules C 10, IKA combustion
bags C 12 or IKA combustible crucible C 14 (see
Accessories). It is also possible to use liquid bur-
ning aids such as paran oil.
Before filling the capsule or the combustion bag
with the substance to be determined, weigh
them to calculate the extra external energy ad-
ded by the burning aid from the weight and the
calorific value. This must be taken into account
in QExternal2. You should keep the amount of
burning aid used to a minimum.
Most fluid substances can weighed out directly
into the crucible. Highly volatile substances are
poured into combustion capsules (IKA gelatine
capsules C 9 or IKA acetobutyrate capsules C 10,
see Accessories) and combusted together with
the capsules.
The burning aids (e.g. cotton thread) must also
fully combust. If there is any unburnt residue, the
test must be repeated.
When working with unknown substances, select
very small weighted samples at the start in order
to determine the natural energy. If you are bur-
ning unknown samples, leave the room or keep
your distance from the calorimeter.
After combustion the water produced is collec-
ted and the decomposition vessel is thoroughly
rinsed with distilled water. The water used for
rinsing and the solution produced are combined
and examined for their acidity. If the sulphur con-
tent of the fuel and the nitric acid correction are
known, it is not necessary to analyse the water.
Note: in all the automatic calculations an extra
100 J have already been included for the electric
ignition energy. This value cannot be set.
Materials which are dicult to ignite or combust
are combusted together with a burning aid. The
burning aid is first weighed and then put into
the crucible with the sample. The additional heat
quantity can be determined from the weight of
the burning aid and its known specific calorific
value. You must correct the test result by this heat
quantity.
IKA- combustible crucible C 14 is a combustible
crucible, which can be used instead of a standard
crucible. The combustible crucible burns without
leaving any residue whatsoever. When using a
combustible crucible you do not need an extra
cotton thread. The crucible is placed directly on
the permanent ignition wire in the decompositi-
on vessel and ignited.
The cleanliness of the material used in the com-
bustible crucible prevents chemical contaminati-
on of the sample (no blank values).
Decomposition vessels, in which the combustible
crucible is used, must be fitted with an extra part
(support C 5010.4, see Accessories). The sample
is weighed out into the combustible crucible as
normal. In most cases no additional burning aid
is required because the combustible crucible itself
serves as a burning aid.
Virtually all of the materials to be studied contain
sulphur and nitrogen. Under the conditions in ca-
lorimetric measurements, sulphur and nitrogen
combust to SO2, SO3 and NOx. Together with
the water from combustion and moisture, sulp-
huric and nitric acid as well as heat of solution
are produced. In order to obtain the standard ca-
lorific value, the influence of the heat of solution
on the calorific value is corrected. The calculation
formulae depend on the standard used. These
are not taken into account in the calculation for
C 200. Use IKA's CalWin software for this.
5.3 Information about the sample
FIt is essential that the sample fully combusts
to ensure correct determination of the calorific
value. After each experiment check the crucible
and all the solid residue for signs of incomplete
combustion.
As a rule the weighted sample must be se-
lected in such a way that the temperature
increase during the measurement is below 4 K
and comes close to the temperature increase
of the calibration (max. extra energy: 40,000 J).
The system must be calibrated in every work
mode used.
If a calorimeter is operated with several de-
composition vessels, you will need to deter-
mine the heat capacity of the system for each
decomposition vessel.
Ensure that calibration is carried out under the
same conditions as the subsequent tests. If sub-
stances are used in the decomposition vessel in
combustion tests (e.g. distilled water or solu-
tions), you must use exactly the same amount of
this substance for calibration.
For more detailed information on calibration, ple-
ase see the relevant standards.
6.1 Decomposition vessel C 5010
If several decomposition vessels are used,
their individual parts must not be interchan-
ged (see Stamping individual parts).
To prolong the life of wearing parts (o-rings,
seals, etc.) we recommend that you always
work with a water trap.
5.4 Calibration
The calorimeter system must be calibrated before
accurate measurements are possible. This is done
by combusting tablets made of certified benzoic
acid (see Accessories) with a known calorific va-
lue. The heat quantity required to raise the tem-
perature of the calorimeter system by one Kelvin
is used to determine the heat capacity of the so-
called "C-value" of the system. For this calculati-
on the formula (1) (see section 5.1) is adapted:
C = (Ho * m + QExt1 + QExt2) / DT (2)
This value is used for determining the following
calorific values.
The heat capacity is determined by the measu-
ring cell and the decomposition vessel. It has a
significant influence on the calorific value to be
calculated and must be redetermined in particular
when using for the first time, after servicing and
when parts are replaced. A monthly control mea-
surement is recommended.
The term "measurements" below refers to both
the measurements to calibrate the calorimeter
system (calibration measurements) and the actual
measurements for determining the calorific value.
The difference lies in the calculation (cf. section
5, formulae (1) and (2)), whereas preparation and
performance are virtually identical.
Exact measurements are only possible when the
individual test steps are carried out carefully.
You must therefore follow the exact proce-
dure described in section 1 "For your safe-
ty” and in the following sections
Please also see section 5 "Calorimetric measure-
ments”.
Failure to observe these instructions could re-
sult in damage to the decomposition vessel.
Damaged decomposition vessels could crack!
Follow the operating instructions for the de-
composition vessel!
Preparing and performing measurements
Union nut
Oxygen valve
Cover
Crucible
Electric
ignition
contact
Ignition
wire
Crucible
holder

20 21
Fill the decomposition vessel using oxygen sta-
tion C 248 (see section 4.10).
Slide the ignition adaptor onto the decompo-
sition vessel.
For information on how to operate the decompo-
sition vessel, please see the operating instructions
provided.
6.2 Preparing the decomposition vessel
Prepare the decomposition vessel as follows:
Unscrew the union nut and remove the cover
using the handle.
Attach a cotton thread to the centre of the
ignition wire using a loop.
Weigh out the substance directly into the
crucible with an accuracy of 0.1 mg. Note the
weight or enter directly into the calorimeter. (See
section 6.3 "Preparing the measurement").
Insert the crucible into the crucible holder.
If necessary put some distilled water or a solution
into the decomposition vessel.
See section 5.3 "Information about the samp-
le" and section 1 "Safety precautions”.
Using tweezers, align the cotton thread so that
it hangs inside the crucible and is immersed in the
sample. This will ensure that the burning thread
ignites the sample during the ignition process.
Attention! Place the cover onto the lower
section and push down until it presses against the
stop piece in the lower section. Place the union
nut onto the lower section and tighten by hand.
Cover
Union nut
Handle
Cotton thread
Max. 1 mm
6.3 Preparing the measurement
The calorimeter is in ”waiting” mode.
Selecting MEASUREMENT (F2) will take you
to the "Prepare measurement" menu.
Enter the noted weighted sample in the
marked "Einwaage" (weighted sample) field with
an accuracy of 0.0001 g using the keyboard.
You can access all the other options by pressing
UP/DOWN (F2):
CALIBRATION
Enter "1" here to perform a calibration.
Check the other presettings:
DECOMPOSITION VESSEL
Enter the number of the decomposition vessel
used for this test. If you are only using one de-
composition vessel, you can leave the "1" in the
"decomposition vessel" field.
QEXTERNAL1
Correction value for the heat energy generated
by the cotton thread ignition aid. A default value
of 50 J is specified here. The default value for the
combustible crucible option is 0 J. If you use a
burning aid other than the C 710.4 IKA-cotton
thread, change this value.
Place the decomposition vessel into the inner
vessel of calorimeter C 200. The decomposition
vessel must be placed between the three locating
bolts.
Pour approx. 2 l tap water maintained at a con-
stant temperature into the tank using the measu-
ring cup. Keep an eye on the level indicator!
Attaching
the ignition
adaptor
Detaching
the ignition
adaptor
1.
2.
Locating bolt
Magnetic stirring bar

22 23
Close the cover by moving it to the left out
of the locking position until it slides down by its-
elf. The decomposition vessel comes into contact
with the igniters via the ignition adaptor.
The "Fill" message will appear.
The inner vessel will be filled with water (ap-
prox. 70 s). The measurement process will begin
as soon as it is full.
QEXTERNAL2
Correction value for the heat energy from other
burning aids e. g. combustible crucible). The de-
fault value is 0.
TESTNO
For each measurement the software automati-
cally assigns a number in the form of yymmddnn,
with jj representing the year, mm the month, dd
the day and nn a consecutive number. You can
also assign your own numbers to the measure-
ments.
Example: TestNo = 0509150 represents the first
measurement on 15 September 2005
Press OK (F1) to apply your entries.
6.4 Performing the measurement
Once you have done all the necessary prepa-
ratory work as described above, the message
”Storage filled?” (Tank filled?) will appear. Check
that the tank is filled with water maintained at
a constant temperature and confirm by pressing
CONTINUE (F1).
The message appears “Vessel safe locked?“
Make sure the fact that the the decomposition
vessel is duly locked and confirms you this with
OK (F1).
After in each case 1000 ignitions of an decompo-
sition vessel you are with the message “Vessel x :
1000 ignitions!” pointed out that this decompo-
sition vessel reached one point of maintenance
time and must be submitted of a security exami-
nation.
Confirming these references by you successively
OK(F1) and OK(F2) press.
This reference does not relieve you of the obliga-
tion to already constandly examine decompositi-
on vessel also before on wear and accomplish the
safety examination if neccessary.
The message appears “Close the cover“.
After the measurement open the cover to au-
tomatically empty the inner vessel. Remove the
decomposition vessel and the ignition adaptor. To
release tension in the decomposition vessel use
venting button under a fume hood or venting
station C 5030 available in our accessories range.
See also section 1 "Safety precautions".
Open the decomposition vessel and check the
crucible for signs of incomplete combustion. If
combustion is incomplete, discard the test result.
Repeat the test.
6.5 Cleaning the decomposition vessel
If you suspect that the combustion sample,
the combustion gases produced or the com-
bustion residue could be harmful to health,
wear personal protective equipment (e.g.
protective gloves, breathing masks) when
handling these materials. Harmful or polluting
combustion residue must be disposed of as
hazardous waste. Express reference is made
to the applicable regulations.
In order to obtain accurate measurements it is
essential that the decomposition vessel is clean
and dry. Impurities alter the heat capacity of the
decomposition vessel and thus cause inaccurate
measuring results. It is important to thoroughly
clean the inner walls of the vessel, the internal
fittings (brackets, electrodes etc.) and the com-
bustion crucible (inside and out!) after each com-
bustion test.
In most cases, you will only need to remove con-
densate from the inner walls of the vessel and the
internal fittings. It is sucient to thoroughly wipe
the parts with an absorbent, non-fibrous cloth. If
the decomposition vessel cannot be cleaned in
the above way (e.g. due to baking, pitting, cor-
rosion etc.), please contact the Technical Service.
a) The measurement process is fully automatic
for automatic measuring procedures (isoperibol,
dynamic and time-controlled, see section 4.6).
The result will appear once the measuring pro-
cess is complete.
b) With the manual measuring procedure the
user decides when ignition is to take place and
when the measurement is complete.
To start ignition press IGNITE (F1)
Press the same key END (F1) again to end the
measurement
In the case of manual ignition / completion, "Pre-
paring to ignite" or „Preparing to complete" will
appear in the status line. Ignition/completion is
only finished when this display disappears (max.
60 sec.).

24 25
Cause:
Sample not combusted.
(Cotton thread had no contact to sample)
User action:
Open cover and remove decomposition vessel. If
the cotton thread has not combusted, check the
ignition contacts and the ignition wire. Otherwi-
se, start a test and use ignition aid if necessary.
Cause:
Decomposition vessel may not have been filled
with oxygen.
User action:
Open the cover to cancel the test and perform a
new measurement.
Message:
A/D converter error
Cause:
An error has occurred in the temperature measu-
ring system.
User action:
If the measurement is still active, open the cover
to cancel the test. Then reinitialise the converter
(see section 7.2 ”ReInit”). If this does not work,
switch the appliance off and then back on again.
If this does not work either, please contact IKA-
Service.
The combustion residue in the crucible, e.g. soot
or ashes, should also be wiped away with an ab-
sorbent non-fibrous cloth.
6.6 Errors in the measuring procedure
Errors in the measuring procedure are shown in
the alarm line of the display and remain there un-
til they are appropriately acknowledged by the
user.
Message:
Cover open!
Cause:
The cover was opened during the measuring pro-
cess.
User action:
The inner vessel will be emptied automatically,
after which a new measurement can be started..
Note:
if the cover was opened within a minute of igni-
tion, automatic emptying will be delayed by ap-
prox. 2 mins for safety reasons.
Cause:
No ignition contact during the measurement.
User action:
Open the cover and check the ignition wire and
ignition contacts. If necessary, clean the ignition
contacts or change the ignition wire. Check that
the ignition adaptor is in the right position.
Message:
No temperature increase
Appears if the value specified for the temperature
increase is not reached within one minute of ig-
nition (0.05 K).
Cause:
Start temperature of inner vessel is not in range
22 ± 3 °C.
User action:
Open cover to cancel measurement, or press
CONTINUE (F1) to continue the measurement
anyway.
Note:
A measuring result obtained in this way does not
correspond to the standard conditions.
User action:
Check the water temperature for the tank. Check
the temperature shown. If it does not change
within 5 to 10 secs, reinitialise the A/D converter
(see section 7.2 ”ReInit”).
Message:
Drift unstable
Cause:
There is no magnetic rod in the inner vessel or
the magnetic rod is outside of the magnetic field.
User action:
Open the cover to cancel the test and check the
position of the magnetic rod. If necessary, insert
the rod or position properly (see section 6.2, ).
Cause:
The stirrer is not working.
User action:
Open the cover to cancel the test. Half fill the
inner vessel with water (via Service menu see sec-
tion 7.2) and then manually switch the stirring
motor on and off (also via Service menu). If the
stirring motor does not work, please contact the
IKA-Service.
Message:
Fill time exceeded
Cause:
No water in tank.
User action:
Check the fill level in the tank and top up with
water if necessary. Press CONTINUE (F1) to re-
peat the filling process and continue the measu-
rement.
Cause:
Filter in inner vessel is dirty.
User action:
Open the cover to cancel the test. Remove and
clean the inner vessel filter.
Cause:
Pump for filling inner vessel is not working.
User action:
Open the cover to cancel the test. Switch on
the pump in the service menu (see section 7.2
”Pump”) and check water jet in tank. If there is no
water flow, please contact IKA-Service.
Cause:
Fine filter clogged.
User action:
Clean the fine filter (see section 8.3 "Fine filter").
Message:
Start temperature outside

26 27
Please note that you must stop every started ac-
tion.
Exceptions:
The "Ignite” action ends automatically after a set
time period (approx. 2 seconds).
With actions "Reset” and "ReInit” you will au-
tomatically exit the menu after the actions have
been performed.
If you exit the service menu, all the started actions
will be stopped and the system will be restored to
its original state. This will allow you to continue
without any errors.
7.2 Description of service menu options
Ignition
This menu item allows you to check the ignition
function.
Requirement:
There must be a decomposition vessel without
sample, but with ignition thread in the inner ves-
sel and the cover must be closed.
Fill IV
This menu item allows you to fill the inner vessel
manually.
Requirement:
There must be enough water in the tank.
Empty IV
This menu item allows you to empty the inner
vessel.
Requirement: the emptying hose must be locked
in place in the plug-in coupling(section 4.2).
Stirrer
This menu item allows you to switch on the stir-
rer drive and check that the magnetic rod is also
rotating in the inner vessel.
Requirement:
There must be approx. 0.5 l water in the inner
vessel.
Meldung:
Füllzeit überschritten
7.1 Operation
This menu allows you to directly control and test
various actions and statuses of the calorimeter
without performing a measurement. There are
also menu items which can be used to start up
and shut down the appliance. The service menu
can only be run if the calorimeter displays the
start screen.
Perform the desired action using:
MENU (F3)
UP/DOWN (F2) to "Service"
OK (F1)
There are now eight actions to choose from
Select the desired action with UP/DOWN (F2
)
OK (F1) starts the action
OK (F1) stops the action again
Service menu Reset
This menu item allows you to restore the default
settings.
Reinit
This menu item allows you to reinitialise the
A / D converter.
There is also a filter in the filler. This prevents dirt
in the storage container. If the filter is dirty, remo-
ve the entire filler and clean the filter (see section
4.1, ).
8.3 Micro filter
In the back plate of the equipment is additionally
a micro filter. If the micro filter is dirty, the water
flow in the equipment is reduced. Thus the water
filling time of the inner vessel extends. Thereby
if the upper time limiting is exceeded by 120 se-
conds, the system indicates the disturbance and
breaks the measurement off.
Pump
Switch the pump on with this menu item. The
outer vessel will be filled and rinsed (see also sec-
tion 4.7).
Requirement:
Check that there is water in the tank.
Empty outer vessel
This menu item allows you to empty the outer
vessel.
Note:
Shut-down, completely empty appliance.
Requirement:
The emptying hose must be locked in place in
plug-in coupling (section 4.2).
In order to ensure trouble-free operation over a
long period of time, it is important to carry out
the following maintenance work to the calorime-
ter system.
Disconnect main plug prior
to cleaning!
8.1 Inner vessel filler
Remove the filter element from the inner vessel
and clean the inner vessel filter in clear water or
an ultrasonic bath if there is any dirt visible. Also
clean the inner vessel, by re-moving the positio-
ning mount for the decomposition vessel. Place
the filter back on the inner vessel connector after
cleaning.
Operation without the filter element causes dirt
in the valves and thus appliance failure.
8.2 Filler
Cleaning and maintenance

28 29
8.4 Maintaining the water circulation
In the case of discontinuous use with tap water
(single measurements with long gaps) a stabiliser
must be added to the water circulation to prevent
the formation of algae.
Add approx. 4 ml IKA-Aqua-Pro C 5003.1 to the
tank (see Accessories).
Switch the pump on via the service menu (sec-
tion “Pump“). Switch the pump off again after
30 - 60 s.
You can also add the 4 ml Aqua-Pro to the water
maintained at a constant temperature in the last
measurement (before long gaps in measuring).
If the appliance is not going to be in opera-
tion for a long period of time, it is advisable
to completely empty the calorimeter's water
circulation.
The water must be drained out before trans-
portation.
Empty the outer vessel via the service menu (sec-
tion 7.2 “Outer vessel“). The emptying hose must
be locked in place in the plug-in coupling (sec-
tion 4.2).
Empty the tank by locking the emptying hose in
place in the plug-in coupling (section 4.2). The
tank will empty automatically.
Press the locking knob on the plug-in coupling
to remove the hose.
8.5 Decomposition vessels
Please see the operating instructions for
C 5010 for information on decomposition
vessel maintenance!
8.6 Cleaning information
For cleaning disconnect the mains plug!
Use only cleaning agents which have been ap-
proved by IKA to clean IKA devices.
These are water (with tenside) and isopropanol.
• For cleaning the micro filter you must empty
the outer vessel (see section 7.2).
• Screw the cover of the micro filter down and
pull you out the filter element.
•
Clean the filter element with clear water.
• Insert the filter element after cleaning with the
opening inward again into the housing of the
micro filter and screw you on by hand the co-
ver.
• Fill the outer vessel again with water (see sec-
tion 4.7).
• Control the tightness after you the equipment
again filled (see section 6.6 “Fill time excee-
ded).
Spare parts order
When ordering spare parts, please give:
- device type.
- serial number, see type plate.
- position number and description of spare part,
see www.ika.com.
Repair
Please only send devices in for repair that
have been cleaned and are free of materials
which might present health hazards.
For repair, please request the “Safety Declara-
tion (Decontamination Certificate)” from IKA
or use the download printout of it from IKA web-
site at www.ika.com.
If your appliance requires repair, return it in its
original packaging. Storage packaging is not suf-
ficient when sending the device - also use appro-
priate transport packaging.
9.2 Consumables
C 710.4 Cotton thread, cut to length
(500 pieces)
C 5010.3 Ignition wire, replacement
(5 pieces)
C 5003.1 Aqua-Pro bath stabiliser
(30 ml)
C 4 Quartz dish
C 5 VA combustion crucible set
(25 pieces)
C 6 Quartz dish, large
C 710.2 VA combustion crucible set, large
(25 pieces)
C 9 Gelatine capsules
(100 pieces)
C 10 Acetobutyrate capsules
(100 pieces)
C 12 Combustion bag, 40 x 35 mm
(100 pieces)
C 12A Combustion bag, 70 x 40 mm
(100 pieces)
C 43 Benzoic acid
(NBS 39i, 30 g)
C 43A Benzoic acid
(100 g)
C 723 Benzoic acid blister pack
(50 pieces)
C 14 Combustible crucible
(100 pieces)
C 15 Paran strips
(600 pieces)
Comment:
- Do not place electrical appliances into the
cleaning agents for cleaning purposes.
- Stainless steel parts can be cleaned using stan-
dard stainless steel cleaning agents, but do not
use abrasives.
- We recommend that you wear protective gloves
for cleaning.
- The operator is responsible for ensuring approp-
riate decontamination in the event that dange-
rous material is spilt onto or into the appliance.
- Before using a cleaning or decontamination
method other than that recommended by the
manufacturer, check with the manufacturer that
the intended method will not destroy the appli-
ance.
9.1 Accessories
C 5010 Decomposition vessel,
standard
C 5010.4 Support for combustible
crucible
C 5010.5 Support for large crucible
C 5030 Venting station
C 5040 CalWin®, calorimeter software
C 248 Oxygen station
C 21 Pelleting press
C 29 Reducing valve
C 200.1 Measuring cup
A 11 basic Analytical mill
IKA - Accessories and consumables
TIP

30
Table power supply (external):
Rated voltage
Frequency
Input power max.
Hz
W
100 - 240 V AC
50 / 60
150
Calorimeter:
Rated voltage
Input power max. W
24 V DC 5A
150
Fuses (internal) 1x 2,5 AT
Perm. on-time Continuous operation
Protection class as per DIN EN 60 529 IP 20
Protection class III
Overvoltage category 2
Contamination level II
Perm. ambient temperature °C 20 ... 25 (constant)
Perm. ambient humidity %80
Operation at a terrestrial altitude 2000 m above sea level
Dimensions mm 400 x 400 x 400 (B x T x H)
Weight kg 21
Measurement ranges J40.000
Measuring mode / measurement times min approx. 17 /Isoperibol
approx. 8 /Dynamic
14 /Time control
approx. 17 /Manual
Interfaces 1 x parallel (Centronics)
1 x serial (RS 232)
Subject to technical changes!
Technical data

20000008742_EN_3548801c_C 200_032021
Technical specifications may be changed without prior notice.
IKAworldwide IKAworldwide /// #lookattheblue @IKAworldwide
Discover and order the fascinating products of IKA online:
www.ika.com
JAPAN
IKA Japan K.K.
Phone: +81 6 6730 6781
eMail: [email protected]
VIETNAM
IKA Vietnam Company Limited
Phone: +84 28 38202142
eMail: [email protected]
INDIA
IKA India Private Limited
Phone: +91 80 26253 900
eMail: [email protected]
UNITED KINGDOM
IKA England LTD.
Phone: +44 1865 986 162
eMail: [email protected]
MALAYSIA
IKA Works (Asia) Sdn Bhd
Phone: +60 3 6099-5666
eMail: [email protected]
CHINA
IKA Works Guangzhou
Phone: +86 20 8222 6771
eMail: [email protected]
POLAND
IKA Poland Sp. z o.o.
Phone: +48 22 201 99 79
eMail: [email protected]
USA
IKA Works, Inc.
Phone: +1 910 452-7059
eMail: [email protected]
KOREA
IKA Korea Ltd.
Phone: +82 2 2136 6800
eMail: [email protected]
BRAZIL
IKA Brasil
Phone: +55 19 3772 9600
eMail: [email protected]
IKA-Werke GmbH & Co. KG
Janke & Kunkel-Straße 10, 79219 Staufen, Germany
Phone: +49 7633 831-0, Fax: +49 7633 831-98
eMail: [email protected]
Other manuals for C 200
2
Table of contents
Other IKA Laboratory Equipment manuals

IKA
IKA ULTRA-TURRAZ UTL 1000/10 User manual

IKA
IKA HBR 4 control User manual

IKA
IKA RET control User manual

IKA
IKA HBR 4 control User manual

IKA
IKA RCT basic User manual

IKA
IKA RET control User manual

IKA
IKA HB digital User manual

IKA
IKA G-L User manual

IKA
IKA ULTRA -TURRAX Tube Drive User manual

IKA
IKA H 29 Manual
Popular Laboratory Equipment manuals by other brands

Efkon
Efkon EFAS user guide

ELITechGroup
ELITechGroup AEROSPRAY TB Applications manual

Applied Photonics
Applied Photonics LIBS-6 user manual

Mitutoyo
Mitutoyo MICROPAK-2LA user manual

Bayer HealthCare
Bayer HealthCare Clinitek Status Operator's manual

Neural Analytics
Neural Analytics Lucid In-Service Presentation