Jireh Microbe User manual

MICROBE
Manual Magnetic Scanner
BT0179 Rev 04.5

PAGE i of iv BT0179 Rev 04.5
SAFETY WARNINGS / PRECAUTIONS
DANGER! The MICROBE is designed for a specic use. Using the
MICROBE outside of its intended use could cause damage to the product.
Read and understand this manual before using.
KEEP THIS MANUAL – DO NOT LOSE
THIS MANUAL IS PART OF THE MICROBE AND MUST BE RETAINED FOR THE LIFE
OF THE PRODUCT. PASS ON TO SUBSEQUENT OWNERS.
Ensure any amendments are incorporated with this document.
WARNING! Can be harmful to pacemaker and ICD wearers. Stay
at least 25 cm (10 in) away.
WARNING! Do NOT operate scanner in an explosive environment.
Do NOT operate scanner in the presence of volatile substances.
The WEEE symbol indicates that the product must not be disposed of as
unsorted municipal waste, but should be collected separately.
(see - Frame bar on page 10)

PAGE ii of iv
Chapter1 Identification 1
1.1. Product brand 1
1.2. Manufacturer 1
Chapter2 Product Specifications 2
2.1. Intended Use 2
2.1.1. Operating limits 2
2.1.2. Operating environment 2
2.2. Dimensions and Weight 2
2.4. Environmental Sealing 3
2.5. Performance Specifications 3
Chapter3 Definitions 4
3.1. Definition of Symbols 4
Chapter4 System components 5
4.1. Component Identification 5
4.2. Tools 6
4.2.1. Included tools 6
4.2.2. Optional tools 6
4.3. Cart 7
4.3.1. Cart Body 7
4.3.2. Cart Handle 7
4.3.3. Brake Handle 7
4.3.4. Wheels 8
4.3.5. Umbilical Housing 8
4.3.6. Front Pivot 9
4.3.7. Rear Pivot 9
4.4. Pivot Buttons 10
4.5. Frame Bar 10
4.6. Slip Joint Probe Holder 11
4.6.1. Probe Holder Setup 11
4.6.2. Probe Holder Adjustment 13
4.6.3. Probe Holder Force Adjustment 13
4.6.4. Slip Joint Probe Holder Left/Right Conversion 15
4.7. Vertical Probe Holder 17
TABLE OF CONTENTS

PAGE iii of iv BT0179 Rev 04.5
4.7.1. Probe Holder Setup 17
4.7.2. Probe Holder Vertical Adjustment 19
4.7.3. Probe Holder Transverse Adjustment 20
4.7.4. Probe Holder Longitudinal Adjustment 21
4.7.5. Probe Holder Left/Right Conversion 22
4.8. Magnetic Wheel Kit 24
4.9. Pre-Amp Bracket 25
Chapter5 Preparation for Use 26
5.1. Configurations 26
5.1.1. One Probe Cantilever 26
5.1.2. Two Probe 26
5.1.3. Two Probe Cantilever 27
5.1.4. Four Probe 27
Chapter6 Operation 28
6.1. MICROBE Setup on a Scanning Surface 28
Chapter7 Maintenance 30
Chapter8 Troubleshooting 31
8.1. Technical Support 31
Chapter9 Spare Parts 32
9.1. MICROBE Cart 32
9.2. Kit Components 34
9.2.1. Encoder Connector Type 35
9.3. Accessories 35
9.3.1. Magnetic Wheel Kit 35
9.3.2. Pre-Amp Bracket 35
9.4. Probe Holders 36
9.4.1. Slip Joint Probe Holder Parts 36
9.4.2. Vertical Probe Holder Parts 37
9.5. Probe Holder Components 38
9.5.1. Arm Style 38
9.5.2. Yoke Style 38
9.5.3. Swing Arm Style 38
9.5.4. Pivot Button Style 38

PAGE iv of iv
9.6. Variable Components 39
9.6.1. Frame Bars 39
Chapter10 Disposal 40
Chapter11 Limited Warranty 41

PAGE 1 of 42 BT0179 Rev 04.5
Chapter 1
IDENTIFICATION
1.1. Product brand
This user manual describes the proper safety precautions, setup and use of the
MICROBE manual magnetic scanner.
1.2. Manufacturer
DISTRIBUTOR:
Jireh Industries Ltd.
53158 Range Road 224
Ardrossan, Alberta, Canada
T8E 2K4
Phone: 780.922.4534
jireh.com
MANUFACTURER:

PAGE 2 of 42
Chapter 2
PRODUCT SPECIFICATIONS
2.1. Intended Use
The MICROBE is a hand held scanner with built-in encoder and magnetic wheels. It
is designed to translate phased array and/or TOFD probes around ferrous piping and
vessels.
2.1.1. Operating limits
Minimum Maximum
Pipe/tube range, outer diameter
(2 probe)
3.8 cm (1.5 in) Flat
Pipe/tube range, outer diameter
(4 probe)
8.8 cm (3.5 in) Flat
Pipe/tube range, longitudinal 7.6 cm (3.0 in) Flat
Pipe/tube range, inner diameter 152.4 cm (60 in) Flat
2.1.2. Operating environment
The MICROBE is designed for use in industrial environments that are between
-20° C (-4° F) and 50° C (122° F).
2.2. Dimensions and Weight
A
B
C
Fig. 1 - Single wheel dimensions

PAGE 3 of 42 BT0179 Rev 04.5
A: (Fig. 1-A) 8 cm 3.15 in
B: (Fig. 1-B) 17.2 cm 6.8 in
C: (Fig. 1-C) 8.1 cm 3.2 in
D: (Fig. 2-D) 11 cm 4.3 in
Umbilical length (standard kit): 5 m (16.4 in)
Cart weight: 0.77 kg (1.7 lb)*
*Cart weight with umbilical housing but not including cabling.
2.4. Environmental Sealing
Dust-tight, water tight (not submersible).
2.5. Performance Specifications
X-Axis encoder resolution 19.1 counts/mm (485.1 counts/inch)
D
B
C
Fig. 2 - Double wheel dimensions

PAGE 4 of 42
Chapter 3
DEFINITIONS
3.1. Definition of Symbols
Instructions to ‘look here’ or to ‘see this part’
Denotes movement. Instructing user to carry out
action in a specified direction.
Indicates alignment axis, can also indicate
insertion or movement of parts.

PAGE 5 of 42 BT0179 Rev 04.5
Chapter 4
SYSTEM COMPONENTS
4.1. Component Identification
Fig. 3 - MICROBE Cart
BTA010
Fig. 4 - MICROBE/ROTIX Umbilical Housing
UMA012-
Fig. 5 - Frame Bar
BG0038-
Fig. 6 - Magnetic Wheel
BTS031
Fig. 7 - Slip Joint Probe Holder
PHA012-
Fig. 8 - Irrigation Kit
CMG007
Fig. 9 - Battery Powered Optical Guide
CXS080
Fig. 10 - 3/8 in Wrench
EA470
Fig. 11 - 3 mm Hex Wrench
EA414

PAGE 6 of 42
4.2. Tools
4.2.1. Included tools
The 3 mm hex driver (Fig. 14) is sucient for all typical operations and
adjustments of the MICROBE.
The 3/8 in wrench (Fig. 15) is used to remove and install buttons on the probe
holders.
4.2.2. Optional tools
Some specialized adjustments require tools that are not included with this kit.
Fig. 12 - Pre-Amp Bracket
CES029-
Fig. 13 - MICROBE Case
BTA002
Fig. 14 - 3 mm hex driver Fig. 15 - 3/8 in wrench
Fig. 16 - 1.5 mm hex
wrench
Fig. 17 - 2 mm
hex wrench
Fig. 18 - 2.5 mm
hex wrench
Fig. 19 - 3 mm
hex wrench

PAGE 7 of 42 BT0179 Rev 04.5
4.3. Cart
4.3.1. Cart Body
A housing for the main encoder and
provides a mounting base for probe holders
and the umbilical housing (Fig. 20).
4.3.2. Cart Handle
The handle is used to both manage cabling
and provide an ergonomic grip during use.
To remove the cart handle, simultaneously
depress the handle release catch and slide
forward (Fig. 21).
4.3.3. Brake Handle
Pivot the brake handle left or right to
engage or disengage the brake.
To remove the brake handle, pull straight
up on the central knob of the black lever.
(Fig. 22). To reinsert, align the spline on the
handle with the socket and press down until
it snaps into place.
NOTE: Pulling the brake handle sideways causes binding and makes the handle
dicult to remove.
Fig. 20 - Cart
1
2
Fig. 21 - Cart handle removal
Fig. 22 - Brake handle removal

PAGE 8 of 42
4.3.4. Wheels
The wheels of the cart are interchangeable.
To remove/install the cart wheels, insert the
provided 3 mm hex driver (Fig. 14) in the end of the
shaft opposite the
wheel you wish to
remove. Thread or
unthread the
desired wheel by
hand (Fig. 23). Additional magnetic wheels may be installed when added
magnetic attraction is necessary (see Magnetic Wheel Kit on page 24).
TIP: Be sure all wheels are tight as this can aect the brake and encoder
performance.
4.3.5. Umbilical Housing
The umbilical transmits encoder
signals to the user instrument.
To remove the umbilical, first remove
the cart handle (Fig. 21) and brake
handle (Fig. 22). Locate the black wing
knob at the bottom of the cart and
loosen one turn (Fig. 24).
With the knob loose, slide the umbilical
housing towards the rear of the cart
(Fig. 25).
TIP: The umbilical lock knob is only re-
quired to be loose, one turn should
be enough to remove the umbilical.
Fig. 23 - Umbilical housing removal
WARNING! MAGNETIC MATERIAL. The
magnetic wheels produce a magnetic eld which may
cause failure or permanent damage to items such as
watches, memory devices, CRT monitors, medical
devices or other electronics. People with pacemakers or
ICD’s must stay at least 25 cm (10 in) away.
Fig. 24 - Umbilical housing lock knob
Fig. 25 - Umbilical housing removal

PAGE 9 of 42 BT0179 Rev 04.5
TIP: If the umbilical does not slide freely from the body, you may need to push on the
wing knob to loosen the dovetail nut.
4.3.6. Front Pivot
The front pivot oers an attachment point for various frame bars.
To attach a frame bar, loosen both dovetail jaws enough to allow the frame bar
to be inserted and tighten the knobs (Fig. 26).
The angle of the bar can be adjusted by loosening the side mounted lever,
pivoting to the desired angle, and closing the lever again (Fig. 27).
4.3.7. Rear Pivot
The rear pivot is positioned at the rear of the cart body and provides an
attachment point for various frame bars. To mount a rear pivot follow these
steps:
1. Release the side mounted lever and rotate the pivot down 90° (Fig. 28).
2. Align the rear pivot with the pins at the rear of the cart (Fig. 29).
1
2
Fig. 26 - Frame bar attachment
1
2
Fig. 27 - Pivot
1
2
Fig. 28 - Rotate pivot 90° down Fig. 29 - Position on cart body

PAGE 10 of 42
3. Using the supplied 3 mm hex driver, screw the rear pivot to the cart (Fig. 30).
4. Rotate the pivot to the original position and lock the side mounted lever
(Fig. 31).
4.4. Pivot Buttons
Available in a variety of shapes
and sizes fitting various wedge
dimensions.
Use the supplied 3/8 wrench
(Fig. 15) to remove and install
pivot buttons in the desired hole
location (Fig. 32).
4.5. Frame Bar
Frame bars (Fig. 33) are used to mount
probe holders, probe positioning systems
and other accessories.
Frame bars are available in a variety of
lengths (see - Frame bar on page 10).
Fig. 30 - Screw to scanner body
2
1
Fig. 31 - Return pivot position
1
2
Fig. 32 - Pivot buttons
Fig. 33 - Frame bar

PAGE 11 of 42 BT0179 Rev 04.5
4.6. Slip Joint Probe Holder
AFrame Bar
BProbe Holder Adjustment Knob
CLatch
DSwing Arm Knob
EYo k e
FProbe Holder Arm Adjustment Knob
GProbe Holder Arm
HArm Clamp Screw
IPivot Buttons
4.6.1. Probe Holder Setup
To mount a UT wedge in the probe holder, follow these steps:
1. Rotate the probe holder adjustment knob and attach probe holder to a
frame bar (Fig. 35).
2. Use the probe holder adjustment knob to position the probe holder along
the frame bar (Fig. 36).
F
I
A
C
BD
E
G
H
Fig. 34 - Slip joint probe holder
Fig. 35 - Attach to frame bar Fig. 36 - Adjust on frame bar
Fig. 37 - Adjust swing arm Fig. 38 - Place pivot buttons

PAGE 12 of 42
3. Use swing arm knob to position the swing arm (Fig. 37).
TIP: The swing arm is typically used to adjust TOFD center to center distance
relative to the phased array probes on a four probe configuration (Fig. 33).
4. Using the supplied 3/8 in wrench (Fig. 15), place the pivot buttons as required
(Fig. 38).
TIP: If narrow scanning footprint is required, use pivot button holes closest to the
yoke. Wedge pivoting may be impeded when closer to the yoke.
5. Loosen the arm clamp screw and the probe holder arm adjustment knob
(Fig. 39), remove the outer probe holder arm from yoke.
6. Adjust inner probe holder arm as required to best centre the probe on the
yoke’s pivot axis (Fig. 39).
TIP: The probe holder yoke can accommodate many dierent probe and wedge
sizes of varying widths. It is best to centre the wedge with the yoke’s pivot
axis to reduce wedge tipping when scanning. Position the inner probe holder
arm accordingly with the centre of the yoke (Fig. 33).
7. Position the wedge on the inner
probe holder arm (Fig. 40).
8. Tighten the arm clamp screw
(Fig. 40).
9. Slide outer probe holder arm along
the yoke pinching the wedge in
place.
10. Tighten probe holder arm
adjustment knob (Fig. 41).
Fig. 39 - Adjust probe holder arms Fig. 40 - Place wedge
Fig. 41 - Pinch wedge with arm

PAGE 13 of 42 BT0179 Rev 04.5
4.6.2. Probe Holder Adjustment
To adjust the probe holder, follow these steps:
1. Ensure probe holder is in latched, upper position (Fig. 42). If the probe holder
is already latched, it will only move within the slip joint adjustment range and
have no spring tension.
2. Push the probe holder yoke down toward inspection surface until the
wedge is approximately 6 mm (¼ in) above the inspection surface (Fig. 43).
3. Lift probe slightly and press latch button (Fig. 44) to apply spring pressure to
the wedge.
4. Gently lower probe holder and wedge to the scanning surface (Fig. 45).
4.6.3. Probe Holder Force Adjustment
It is possible to adjust the tension of the probe holder spring.
NOTE: To perform this operation the 2 mm hex wrench (Fig. 17) and 3 mm hex
wrench (Fig. 19) is required.
Light 1 kg 2 lb
Medium 2 kg 4 lb
Heavy 3 kg 6 lb
When configured correctly, these
settings exert the indicated spring
force on the Probe.
Fig. 42 - Lift to latched position Fig. 43 - Lower to scanning surface
6
mm
approx.
Fig. 44 - Lift and press latch button Fig. 45 - Spring loaded scan position

PAGE 14 of 42
To adjust the probe holder’s force, follow these steps:
NOTE: Do not perform this operation on scanning surface.
1. Ensure the probe holder is in the upright latched position (Fig. 42).
2. Lift probe holder slightly and press the latch button (Fig. 46) to release the
probe holder the full 45° degrees.
3. Insert the short arm of a 3 mm hex wrench into the 3 mm slot (Fig. 47).
4. Place the 2 mm hex wrench into the force adjustment screw (Fig. 48).
5. Lightly press the long arm of the 3 mm hex wrench down. Using the 2 mm
hex wrench, loosen the force adjustment screw but do not remove it (Fig. 49).
6. Gently apply pressure on
the long leg of the 3 mm
hex wrench until the force
adjustment marker lines
up with the desired spring
tension. While keeping the
markers in line, tighten the
force adjustment screw
(Fig. 50).
Fig. 46 - Lift slightly and press Latch Fig. 47 - Unlatched position
Fig. 48 - Insert hex wrenches Fig. 49 - Press 3 mm hex wrench down
Heavy Medium
Light
Force Adj. Marker
Heavy Medium
Light
Force Adj. Marker
Fig. 50 - Choose desired tension

PAGE 15 of 42 BT0179 Rev 04.5
4.6.4. Slip Joint Probe Holder Left/Right Conversion
To reverse the probe holder, follow these steps:
1. Unscrew the yoke from the swing arm (Fig. 51).
2. Loosen the probe holder arm adjustment knob and arm clamp screw. Slide
the arms from the yoke (Fig. 52).
3. Flip the yoke 180° and reverse the probe holder arms (Fig. 53).
4. Place the pivot buttons on the inside of the probe holder arms (Fig. 54) using
a 3/8 in wrench (Fig. 15). Slide the arms onto the yoke and tighten the probe
holder arm adjustment knob and the arm clamp screw.
Fig. 51 - Unscrew yoke pivot screw Fig. 52 - Remove arms
Fig. 53 - Flip yoke and reverse arms Fig. 54 - Attach arms and move buttons
Other manuals for Microbe
1
Table of contents
Other Jireh Scanner manuals