Jorvet ELECTROSURGE J0470M User manual

J04
70M
ELECTR
OSURG
E
Radio-F
r
equency
Electr
osurger
y
Unit
Operation Instructions

ELECTROSURGE
J0470M
Hand Piece
Connection
(Black Socket)
Hand Piece
Indifferent Plate
Connection
(Red Socket)
Indifferent
Plate
Intensity Level
Waveform
Mode
Electrodes
Foot
Switch
Power
Active
Light

1
JORVET
ELECTR
OSURG
E
Radio F
r
equency
Electrosurgical
Unit
What Is Radio Frequency Electrosurgery?
Electrosurgery is a commonly used procedure that has been
used in
both human and veterinary surgery
for
over
20
years. Ultra-
high
frequency radio waves are transmitted through a fine wire electr
ode
to
a flat antenna
on
a ground plate placed under the animal.
The
high
frequency radio waves pass through tissue and make a
precise
surgical
incision just like a scalpel blade. The surgeon chooses
from
varying radio
waveforms that result
in
varying degrees: from
pure
cutting to
hemostasis. Radio frequency electrosurgery should
not be
confused
with electro
cauter
y
.
Electrocautery uses thermal energy
or
heat
that
simply burns the
tissue.
The advantages of electrosurgery are fine, precise incisions
with hemostasis of small “bleeders”, and less blood obstructing
the surgical field. This makes any procedure faster and easier.

2
What Do the Dial
Settings
Refer T
o?
The
two
dials
on
the front panel refer
to
Fig.
1
waveform mode
and
power
intensit
y
.
The waveform mode has different settings. This
dial
allows the selective
use
of
various waveforms that give varying degr
ees
of
cutting power and
hemostasis.
Cut
mode:
This selection produces a sine wave of continuous high frequency that
provides smooth cutting power. This creates the least
amount of
lateral
heat and tissue damage. This is the preferred mode
for thin
avian skin and
incisions where heavy bleeding is
not
a problem.
This
mode provides
for
the least amount
of hemostasis.
Coag 1
mode:
This selection produces a
periodic sine wave with brief rests
between cycles
. This waveform has less
cutting power, but coagulation
of
small capillary beds is
achieved.
Coag 2
mode:
This selection produces a partially-rectified s i n e wave. This provides the
most hemostasis
on
v
essels
up to 1/16” in
diameter
.
This mode can also
use the indirect
method of
coagulation. This method employs a fine
hemostat
to
grasp
the
vessel
cleanl
y
.
The active electrode can touch the
hemostat near
the
tip and coagulate that v
essel.

3
Getting
Star
t
ed
Remove the unit from the box and lay
it on
a flat
surface.
1)
Connect the female end of the power cord to the rear of the unit, and the
male end in a grounded electrical outlet.
2)
The male jack of the Black (or Green) surgery hand piece is inserted into
the
unit into
the black socket marked “ HANDPIECE”
on
the front
panel.
3) I
ndifferent Plate: This is the red cord. The
male
jack should
be
inserted
into the red female jack opening
marked “IND PLATE” on the front
panel.
4)
Connect the foot pedal in the rear ofthe unit
.
5) The Indifferent Plate should be placed under the animal close to
the procedure site. Conductive gel must be used to create uniform
skin contact and maximize surface contact with the plate. Failure
to use gel can result in plate burns to the animal in addition to
poor or no function.
6) Select an electrode and insert the straight shaft end into the top
opening of the handpiece. Be careful not to bend or damage the
electrode in any way. Do not attempt to unscrew the handpiece
as this will damage it.
7) Set the waveform dial to the desired mode.
8) Select the power intensity. 1 is lowest and 10 is highest. Always
start at the minimum level and gradually work up to a level where
cutting is smooth but no sparks are emitted.
9) The “ACTIVE” light should be lit when the footswitch is pressed. This
indicates that radio frequency waves are being sent.
10) The electrode should be gently held on the tissue. A smooth,
constant motion will produce less lateral heat and the cleanest
incision. The tissue should always be kept moist for better and
cleaner cutting. Cutting should be done without pressure, as this
can damage the tissue and tip.

4
Practice
It is highly recommended to
practice your technique before using the
unit
on
clinical cases. There are a number of
text
books
available
on
electrosurger
y
.
The wide range
of
applications
can be
found
by
reading
through
these
texts.
1)
Obtain
a fresh, lean cut
of
beef, such as round
or
sirloin
steak.
The meat should reach room temperature. Moisten the
meat
with a small amount
of saline.
2) Set up the unit and place the meat on top of the
indifferent plate.
3) Select an electrode and place it in the hand piece.
4)
Turn waveform mode
to
Coag
2.
5)
Set power level at #8
position.
6)
Press
on
the
foot pedal.
7) Do
several incisions and observe the results. This high setting
will
cause sparking and may char
tissue.
8)
Adjust power
to #1.
This
will
illustrate poor cutting
or
extreme
drag.
9)
Move the power setting
to
different levels and continue
to
practice.
10)
Try different waveform settings, but wait
10
seconds between
adjustments. The more
you
practice the easier the unit
will
perform
for you on
actual
patients.
Tip:
The cutting should
be
smooth with
no
sparking
or
r
esistance.

5
Cutting
The tissue should always
be
moist. Dry tissue can result
in surface
charring. Saline solutions are the best solution
to
use.
Cutting
involves
no
pressure. The hand should rest
on
a support
in
order
to
keep control
over the hand
pi
e
c
e
.
Remember to use a feather light
t
o
u
ch
.
Hemostasis
Small bleeders can easily
be
stopped with an electrosurgery
unit.
The vessel should have direct pressure first placed
on
a sponge
or
hemostat. The vessel should
be
clearly located and
not
co
v
er
ed
with
blood.
The waveform should
be in
Coag
1 or
Coag
2. Coag 2
is
preferred
for
larger bleeders. The
two
types
of
coagulation
electrodes
are the round ball electrodes and heavier needle.
An
indirect
method
uses a hemostat with a fine
tip.
The vessel is grasped
by
the hemostat.
The electrode is applied
to
the hemostat about
1”
from its
end.
Care and
Selection
of
Electr
odes
The electrodes can
be
autoclaved
or cold
sterilized. The
protective
rubber
on
the shaft should
be
intact.
If it
is cracked
or
worn a shock
or
burn may
be
felt
by
the
operator
.
Autoclaving
will
affect the
r
ubber
sleeve over
time.
NOTE:
Always clean
off
electrodes after use with a gentle
di
s
in
f
e
c
ta
n
t
.

6
Electrode Type
There are literally hundreds
of
different styles
of
electrodes available
from various manufacturers. The
J
or
V
et
Electrode is the
industry
standard
3/32”
shank. Other
manufactur
er
’
s
electrodes
will
work
on
the
J
or
V
et
unit.
GT1: Blade Electrode
GT2: Small Loop Electrode
GT3: Large Loop Electrode
GT4: Fine Wire Electrode
GT5: Heavy Wire Electrode
GT8: Small Ball (2.38mm) Electrode
GT9: Heavy Ball (5mm) Electrode
GT1 GT2 GT3 GT4 GT5 GT8 GT9

7
Electrode
St
yles
Ball end:
These are best
for
hemostasis
purposes.
Loop end:
These electrodes work nicely with skin tumors
or
gingival hyperplasia. The
loop
encircles and slices through tissue.
Circle and diamond
loop
configurations are
available.
Straight:
Thin wires work well
for
cutting. The heavier wires
work
well
for hemostasis.
Caution
1)
Radio frequency can interfere with heart pacemakers.
Consult
a
physician before exposing anyone wearing a pacemaker
to
radio
frequency
waves.
2)
Various
ECG
and monitoring equipment can
be
affected
by
radio
frequency waves. Place these items as far away from
the
Indifferent
Plate as
possible.
3)
When changing electrodes
do not
have the
foot pedal
depressed.
4)
Electrosurgery can give
off
sparks
and ignite flammable
liquids
or gases.
5)
Electrosurgery still requires anesthesia
or analgesia.
6)
The operator can inflict wounds, burns,
or
shocks
to
themselves
through inappropriate use. Just as misuse of a scalpel can cut
you, so can misuse of electrosurgery units.

8
.Trouble
Shooting
1)
Before starting, make sure all connections are correct.
R e m e m b e r : Hand piece into black socket; IND plate into red socket.
2) Too
much sparking: Adjust power
down by 2 positions.
3) Will notcut:
a)
Is the power light
on?
b)
Foot pedal
connected?
c)
ACTIVE light
on
when
foot
pedal is
depr
essed?
d)
Electrode properly
attached?
e)
Indifferent Plate attached and
properly placed?
Autoclaving Handpiece & Electrodes
The handpiece and electrodes should be thoroughly cleaned and
dried following a procedure. These can then be autoclaved in a
sterilization pouch at no more than 250° F for 10-15 minutes.

9
◆
SKU
J0470M
Specifications
◆110
volt (available
in 220v)
◆
Output power:
100 watts
◆
Output frequency:
2.0 mHz
◆Size:
Height:
3”
Width:
8 ½”
Depth: 9”
Weight:
7
½
lbs.
◆
Replacement Fuse:
12.5a 250v fuse.

10
____________________________________
____________________________________
____________________________________
____________________________________
____________________________________
____________________________________
____________________________________
____________________________________
____________________________________
____________________________________
____________________________________
____________________________________
Users Notes

11
LIMITED
WARRANTY
This Electrosurgy Unit is under warranty
for
a period
of 12-months
against defects
in
workmanship and materials.
The completed Warranty Card must be returned to
Jorgensen Laboratories, Inc. within 14 days of purchase.
Handpiece: 1 year
Indifference Plate: 90 days
Foot Switch: 90 days
Electrodes are
not
covered under
warrant
y
.
Pocket for Copy of Purchase Invoice
Table of contents
Other Jorvet Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Dräger
Dräger Atlan A300 Instructions for use

NeuroTek
NeuroTek Tac/AudioScan operating instructions

Boston Scientific
Boston Scientific Precision Spectra OMG Adapter manual

Bellman & Symfon
Bellman & Symfon BE9199 Medical device information

MyPurMist
MyPurMist mypurmist free quick start guide

Cincinnati Sub-Zero
Cincinnati Sub-Zero BLANKETROL III 233 Operation and technical manual