A/C icon
Battery Icon
Pump will not charge or hold a charge:
Check to ensure A/C Icon is illuminated- if not try another
cord.
If the AC indicator is illuminated, but not the battery
indicator, check the age of the battery. If available,
temporarily try a battery from another pump. If the pump
then functions, replace the battery (J1097d1).
If still not working call Technical Support .
Maintenance & Troubleshooting
AIR sensors
“AIR” Error - Clean air sensors
with cotton-tipped applicator
and alcohol. Try a new IV set. If
not resolved call Jorgensen
Technical Support.
Jorgensen Technical Support
800-525-5614