Jorvet J0563N Instruction Manual

JorVet Doppler Monitor Gen II
J0563N User Manual
Maintenance & Care
e JorVet Hand-Held Doppler can be operated in a temperature range of 50-113°F (10-
45°C) with humidity not exceeding 85%. Do not expose to extremely high or low tempera-
tures.
Special care is needed for the probe. Mechanical shock should be avoided and careful han-
dling is necessary as damage can occur if the probe hits a hard surface.
e probe can be cleaned with a so tissue or cotton ball moistened with water or dilute
alcohol. Do not scratch or use any abrasive cleaning materials. Do not use organic sol-
vents. Use only mild detergent or disinfecting products. Do not attempt to disconnect the
probe from the main unit by pulling on the cables, always use the metal connectors. If gel
is allowed to dry on the probe, soak the probe in mild detergent, water or dilute alcohol
to loosen the debris, do not scratch or use any abrasive cleaning materials on the probe to
remove the gel.
e main Doppler unit can be cleaned by wiping with a so cloth or paper towel lightly
moistened with a mild detergent, disinfectant or dilute alcohol. Take care not to get any
liquid in any of the connection points. Never submerge the Doppler unit in any type of
liquid.
Repairs - In the case of damage or any questions about the correct operations of the device,
please contact JorVet. ere are no user serviceable parts inside the unit. Do not open the
device; there are no elements inside designed to be operated by the user and doing so will
void the warranty.
• Doppler
• 8 MHz Transverse Linear Probe
• AC Adaptor
• Sphygmomanometer
• BP Cus (Set of 5)
• Headphones
• Ultrasound Gel
• Storage and Carrying Case
Package Contents
Warranty
is equipment is guaranteed for a period of one year aer the date of purchase when used
under normal conditions. Do not open the main Doppler unit; doing so will void the war-
ranty. In the event of any service needed during the warranty period, please contact JorVet.
If the warranty has expired, repair services are still oered for a fee; please contact JorVet
for details.
Jorgensen Laboratories, Inc.
1450 N. Van Buren Avenue
Loveland, CO 80538
(800) 525-5614
www.jorvet.com
[email protected]
Additonal Probes
Pencil Probe for exotics - J0563ND10
Esophageal probe - J0563ND11
Getting to Know Your Doppler
e pocket size Doppler veterinary blood ow detector is a simple device enabling a wide
range of easy and reliable diagnosis of blood ow in the vessels, auscultation on very small
animals, and blood pressure measurement.
Recharging Jack Probe Jack
Power Indicator
On / O Power
Membrane
Blood Pressure Measurement Instructions
1. Determine where you would like to obtain the blood pressure and gently position the
patient so you can easily access the artery. Common locations to obtain blood pressure
on pets is on the forelimb, using the ulnar artery above the metacarpal pad; the hind limb,
using either the saphenous artery above the metatarsal pad or the dorsal pedal artery on
the front of the metatarsals; or the coccygeal artery located on the underside of the tail.
2. Place appropriately sized cu above the vessel to be used. Once the cu is placed,
ensure that it is in good condition. Attach the sphygmomanometer and gently inate and
deate the cu to verify that it holds pressure.
3. Shave the hair over the vessel or wet the hair with alcohol. Ensuring the probe has good
contact with the skin will provide the best possible sound. Excess hair can block or mute
the signal from the probe, making it dicult to hear.
4. Apply a small amount of ultrasound gel to the Doppler probe. e gel conducts the
sound from the artery so the probe can receive it. Too much gel will make the probe
slippery and dicult to handle; too little gel will not conduct the sound properly. Place the
probe with gel on it over the vessel. Don’t press too hard as excessive pressure can occlude
the vessel. A swishing sound should be heard when the probe is placed properly. Move the
probe slightly over the vessel until you achieve the best sound.
5. Occlude the vessel by slowly inating the cu with the sphygmomanometer until you
can no longer hear pulses. Inate the cu a little bit more aer the pulse sounds are no
longer heard. Slowly deate the cu while watching the dial on the sphygmomanometer.
When the sound of the pulses returns, the number the needle on the sphygmomanometer’s
dial is pointing to is the systolic blood pressure.
6. As the user continues to deate the cu, a diastolic pressure can be obtained by noting
the position of the needle on the dial when a change in the Doppler’s tonality is heard.
Obtaining a diastolic pressure reading requires a lot of experience and good operating
conditions.
7. Ideally, the user will obtain 3-5 blood pressure readings to verify accuracy. Always
deate the sphygmomanometer completely and wait about 15-30 seconds in between read-
ings to allow for normal blood ow through the limb. Failure to do this can lead to falsely
high readings. Never leave the cu inated when not actively obtaining a blood pressure
reading!
Operating Instructions
During transport, the probe is disconnected from the main unit. e push-pull connectors
of the probe are connected on the side of the Doppler. e pair of connectors are symmet-
rical and, generally, can be connected in any order. e self-latching mechanism of the
connectors protects against accidental disconnect of the probe. To disconnect, always pull
by the metal part of the connector, never pull on the cord!
e JorVet BP Doppler is easy to use and was designed with simplicity in mind. It has
minimum controls – volume up/down and power on/o.
To turn on the device, press the ON button. To turn the device o, press the OFF button.
To increase the volume, press the UP arrow button. To decrease the volume, press the
DOWN arrow button.
e continuous green LED light indicates that the unit is powered on. A pulsating green
LED light indicates that the batteries should be recharged. When the unit hasn’t been
used for 4 minutes or the battery is low, the unit will power o automatically. To charge
the battery, plug the AC adaptor into the mini-jack on the Doppler unit and plug the AC
adaptor into an appropriate electrical outlet. e LED light should shine RED while the
battery charges. During charging, the red light becomes weaker and will disappear when
the battery is fully charged. A full battery charge time is 15 hours.
e mini-jack socket is made to accept the charging cable as well as the headphones. e
headphones allow the user to listen to the pulses privately. is is convenient if the Dop-
pler is being used in a noisy room or if the sounds from the Doppler frighten the patient.
Blood Pressure Measurement Instructions
Specications

Other Jorvet Medical Equipment manuals

Jorvet J0558X User manual

Jorvet

Jorvet J0558X User manual

Jorvet J1060Q User manual

Jorvet

Jorvet J1060Q User manual

Jorvet J-166M Manual

Jorvet

Jorvet J-166M Manual

Jorvet ELECTROSURGE J0470M User manual

Jorvet

Jorvet ELECTROSURGE J0470M User manual

Jorvet Vet Pro Pump Operating instructions

Jorvet

Jorvet Vet Pro Pump Operating instructions

Jorvet J0744emb User manual

Jorvet

Jorvet J0744emb User manual

Popular Medical Equipment manuals by other brands

Forge Medical VasoStat Instructions for use

Forge Medical

Forge Medical VasoStat Instructions for use

Drive Folding bath bench with back Assembly instructions

Drive

Drive Folding bath bench with back Assembly instructions

Nasco Healthcare C.H.A.R.L.I.E. LF01420U manual

Nasco Healthcare

Nasco Healthcare C.H.A.R.L.I.E. LF01420U manual

Animark PREGSCAN instruction manual

Animark

Animark PREGSCAN instruction manual

jablotron Nanny BM-02 manual

jablotron

jablotron Nanny BM-02 manual

ResMed Swift FX user guide

ResMed

ResMed Swift FX user guide

Electromed SmartVest instruction manual

Electromed

Electromed SmartVest instruction manual

RADIOMETER FLEX ABL90 Operator Training Guide

RADIOMETER

RADIOMETER FLEX ABL90 Operator Training Guide

Miele E 429/1 Fitting instructions

Miele

Miele E 429/1 Fitting instructions

Balanced Body Contrology Assembly instructions

Balanced Body

Balanced Body Contrology Assembly instructions

PamMobility PULSE 7M009001 NP Instructions for use and maintenance

PamMobility

PamMobility PULSE 7M009001 NP Instructions for use and maintenance

Atom Infa Warmer i Operation manual

Atom

Atom Infa Warmer i Operation manual

Atmos S 351 Service manual

Atmos

Atmos S 351 Service manual

Creative Medical POD-1a user manual

Creative Medical

Creative Medical POD-1a user manual

NIOX FeNO VERO Quick reference guide

NIOX

NIOX FeNO VERO Quick reference guide

ARJO HUNTLEIGH NIMBUS 4 Quick reference guide

ARJO HUNTLEIGH

ARJO HUNTLEIGH NIMBUS 4 Quick reference guide

Mastercare Master Turner Medium user guide

Mastercare

Mastercare Master Turner Medium user guide

Contec Medical Systems Co. CONTEC08E manual

Contec Medical Systems Co.

Contec Medical Systems Co. CONTEC08E manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.