KERN MLB50-3 User manual

MLB50-3-BA-defsi-0310
28
GB KERN MLB50-3
Version 1.0 05/2003
Operating Instructions
Moisture Analyzer
Table of contents:
1Technical data ________________________________________________________ 30
2Declaration of conformity _______________________________________________ 31
3Fundamental information (general) _______________________________________ 32
3.1 Intended use ___________________________________________________________ 32
3.2 Inappropriate use_______________________________________________________ 32
3.3 Protective measures _____________________________________________________ 32
3.4 Warning ______________________________________________________________ 32
3.5 Guarantee _____________________________________________________________ 33
3.6 Monitoring the test substances ____________________________________________ 33
4Fundamental safety information __________________________________________ 34
4.1 Observe the information in the operating instructions ________________________ 34
4.2 Staff training __________________________________________________________ 34
5Transport and storage __________________________________________________ 34
5.1 Acceptance check _______________________________________________________ 34
5.2 Packaging _____________________________________________________________ 34
6Unpacking, installation and commissioning_________________________________ 35
6.1 Place of installation, place of use __________________________________________ 35
6.1.1 Installation__________________________________________________________________ 35
6.1.2 List of items supplied _________________________________________________________ 35
6.2 Mains supply __________________________________________________________ 36
6.3 Connecting peripheral equipment _________________________________________ 36
6.4 Initial start-up _________________________________________________________ 36
6.5 Adjustment ____________________________________________________________ 37
6.6 Adjusting______________________________________________________________ 37

MLB50-3-BA-defsi-0310 29
7Moisture determination _________________________________________________ 38
7.1 Use ___________________________________________________________________ 38
7.2 Sample material ________________________________________________________ 38
7.3 Sample preparation _____________________________________________________ 38
7.4 Drying temperature _____________________________________________________ 39
7.5 Application table (recommendations) ______________________________________ 40
7.6 Temperature calibration / adjustment______________________________________ 41
8Setting the parameters __________________________________________________ 42
8.1 Temperature correction on “SubSt“ test object ______________________________ 43
8.1.1 Setting the correction factor ____________________________________________________ 44
8.2 Modes “modE“ _________________________________________________________ 45
8.3 Maximum drying time “IntEr“ ___________________________________________ 46
8.4 Drying temperature “tEmP“______________________________________________ 46
8.5 Data query time interval “Strob“ __________________________________________ 46
9Operation ____________________________________________________________ 46
9.1 Control elements _______________________________________________________ 46
9.1.1 Overview of the keyboard______________________________________________________ 46
9.1.2 Overview of the display screen __________________________________________________ 47
9.2 Operation _____________________________________________________________ 47
9.2.1 Weighing___________________________________________________________________ 47
9.2.2 Taring ( key) ___________________________________________________________ 48
9.3 Determination of humidity _______________________________________________ 49
9.3.1 Test accuracy________________________________________________________________ 50
10 RS232 serial interface ________________________________________________ 51
10.1 Print format ___________________________________________________________ 51
10.2 Changing languages_____________________________________________________ 53
11 Maintenance, upkeep, disposal _________________________________________ 53
11.1 Cleaning ______________________________________________________________ 53
11.2 Maintenance, upkeep____________________________________________________ 53
11.3 Disposal _______________________________________________________________ 53
12 Troubleshooting _____________________________________________________ 54

MLB50-3-BA-defsi-0310
30
1 Technical data
KERN MLB 50-3
Maximum load rating (max.) 50 g
Minimum for drying 0.02 g
Temperature range 50°C-160°C
Setting the temperature in 1°C stages
Pre-heating stage (boost) No
Read off (d) 0.001g /0.01%
Repeatability in weighing
(=standard divergence)
0.001 g
Repeatability during drying
(=standard divergence)
Initial weight 1g
Initial weight 10g
0.2%
0.02%
Please note: recommended minimum weight 2g
Adjusting weight not added 50g (F2)
Ambient conditions 15°C ... 35°C ambient temperature
max. 80% air humidity; not condensing
Shutdown criterion Automatic:
End of drying if the results do not alter during 3
strobe intervals (=heating intervals).
If a time limit has been set drying will shut down
after this period expires.
Adjustable strobe interval: 1 s – 59 s
Manual:
According to set time (10 min to 590 min [=9h 50
min]) (without strobe interval)
Display after drying % humidity = weight reduction
(=residual weight/start weight)*100%
% dry matter = residual weight
(=weight loss/start weight)*100%
% ATRO
(=weight loss/residual weight)*100%
Details are not possible in grammes during drying.
Including sampling trays 10
Protective hood no
Type of projector 2 halogen quartz projectors
200 watts each
Display LED display/ character height 15 mm
3 displays for: - drying status
(depending on configuration)
- temperature
- drying time

MLB50-3-BA-defsi-0310 31
2 Declaration of conformity
KERN & Sohn GmbH
Ziegelei 1
D – 72336 Balingen
Tel: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.com
Declaration of conformity
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE- Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con disitintivo CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms with the
following standards.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den
nachstehenden Normen übereinstimmt.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente
déclaration, est conforme aux normes citées ci-après.
Español Manifestamos en la presente que el producto al que se refiere esta declaración est´´a de
acuerdo con las normas siguientes
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è con-
forme alle norme di seguito citate.
Moisture Balance: KERN MLB 50-3
Mark applied EU Directive Standards
73/23EEC Low voltage
EN 60950
89/336EEC EMC EN 50081-1
EN 50082-1
Date: 01.05.2003 Signature:
Gottl. KERN & Sohn GmbH
Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149

MLB50-3-BA-defsi-0310
32
3 Fundamental information (general)
3.1 Intended use
The balance you have acquired serves to determine the weighing value of the mate-
rial to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the
material to be weighed is manually and carefully placed in the centre of the weighing
plate. The weighing value can be read off after a stable weighing value has been ob-
tained.
3.2 Inappropriate use
Do not use the balance for dynamic weighing. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“ in the balance. (Example: Slowly drain-
ing fluids from a container on the balance.)
Do not leave a permanent load on the weighing plate. This can damage the measur-
ing equipment.
Be sure to avoid impact shock and overloading the balance in excess of the pre-
scribed maximum load rating (max.), minus any possible tare weight that is already
present. This could cause damage to the balance.
Never operate the balance in hazardous locations. The series design is not explo-
sion-proof.
Structural alterations may not be made to the balance. This can lead to incorrect
weighing results, faults concerning safety regulations as well as to destruction of the
balance.
The balance may only be used in compliance with the described guidelines. Varying
areas of application/planned use must be approved by KERN in writing.
3.3 Protective measures
The moisture analyser of Safety Class 1 may only be connected through a carefully
installed socket with protective earth (PE). The protective effect must not be can-
celled out by the use of an extension lead without earth wire. For power supplies from
mains not provided with a protective earth, an equivalent protection meeting the valid
installation regulations must be provided by a qualified electrician.
3.4 Warning
Individual parts of the housing (e.g. ventilation grille ...) can become very hot during
operation. Avoid touching the apparatus except by the handles provided for this pur-
pose.
Sample materials which produce corrosive vapours (e.g. acids) can cause problems
of corrosion of components. The MLB50-3 should be used primarily for drying hy-
drous substances.
The MLB50-3 must not be used to analyse highly inflammable samples which pre-
sent an explosion hazard.

MLB50-3-BA-defsi-0310 33
3.5 Guarantee
The guarantee is not valid following
- non-observation of our guidelines in the operating instructions
- use outside the described applications
- alteration to or opening of the device
- mechanical damage and damage caused by media, liquids
- natural wear and tear
- inappropriate erection or electric installation
- overloading of the measuring equipment
3.6 Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight
must be checked at regular intervals within the scope of quality assurance. For this
purpose, the answerable user must define a suitable interval as well as the nature
and scope of this check. Information is available on KERN’s home page (www.kern-
sohn.com) with regard to the monitoring of balance test substances and the test
weights required for this. Test weights and balances can be adjusted quickly and at
a reasonable price in KERN’s accredited DKD calibration laboratory (return to na-
tional normal).

MLB50-3-BA-defsi-0310
34
4 Fundamental safety information
4.1 Observe the information in the operating instructions
Please read the operating instructions carefully before erecting and commissioning,
even if you already have experience with KERN balances.
4.2 Staff training
The device may only be operated and looked after by trained members of staff.
5 Transport and storage
5.1 Acceptance check
Please check the packaging immediately upon delivery and the device during un-
packing for any visible signs of external damage.
5.2 Packaging
Please retain all parts of the original packaging in case it should be necessary to re-
turn items at any time.
Only the original packaging should be used for return consignments.
Before despatch, disconnect all attached cables and loose/movable parts.
Apply any intended transport security devices. Secure all parts, e.g. glass windshield,
weighing plate, power unit etc., to prevent slipping and damage.

MLB50-3-BA-defsi-0310 35
6 Unpacking, installation and commissioning
6.1 Place of installation, place of use
The balance is constructed in such a way that reliable weighing results can be
achieved under normal application conditions.
By selecting the correct location for your balance, you will be able to work quickly and
precisely.
Therefore please observe the following at the place of installation:
- Place the balance on a firm, level surface;
- Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
- Protect the balance against direct draughts due to open windows and doors;
- Avoid jarring during weighing;
- Protect the balance against high humidity, vapours and dust;
- Do not expose the device to extreme dampness for longer periods of time.
Inadmissible bedewing (condensation of air moisture on the device) can occur
if a cold device is taken into a significantly warmer environment. In this case,
please acclimatise the device for approx. 2 hours at room temperature after it
has been disconnected from the mains.
- Avoid static charging of the material to be weighed, weighing container and
windshield.
Major display deviations (incorrect weighing results) are possible if electromagnetic
fields occur as well as due to static charging and instable power supply. It is then ne-
cessary to change the location.
6.1.1 Installation
Use the foot screws to level the balance until the air bubble in the bubble level is in
the prescribed circle.
6.1.2 List of items supplied
Standard accessories: Available optional accessories:
- Balance including humidity - Aluminium trays (item no.:MLB-A01)
identifier attachment - Glass fibre filter (item no.:RH-A02)
- Tray holder - Temperature calibration set
- 10 sampling trays - (item no.:MLB-A03)
- Mains cable

MLB50-3-BA-defsi-0310
36
6.2 Mains supply
A mains cable is used for the electric power supply.
Please check whether the balance voltage intake has been set correctly. The balance
may only be connected to the mains supply if the details on the balance (label) and
the customary on-site mains voltage are identical.
Important:
Does the designation (220 V 50 Hz) comply with the customary on-site mains voltage?
- Do not connect if there are differences in the mains voltage!
- The balance may be connected if in compliance.
6.3 Connecting peripheral equipment
The balance must be disconnected from the mains before connecting or disconnect-
ing additional equipment (printer, PC) to or from the data interface.
Only use KERN accessories and peripheral equipment with your balance. These
have been ideally coordinated to your balance.
6.4 Initial start-up
The accuracy of the balance depends on the local acceleration of the fall.
Please be sure to observe the information in the chapter on “adjusting”.
Use the main switch (above the power plug) to switch on the balance. ”tEmP “
will appear.
You have two choices:
1. Press the key, the balance will run through a ”tEst “ mode. “0.000 “ will appear
on the display screen after a few seconds. A short warm-up period of approx. 10
minutes is advisable for stabilisation purposes.
2. It is possible to start up a pre-heating process that pre-heats the weighing region
to a reasonable drying process temperature. This pre-heating process is only practi-
cal before initial drying if there are to be several consecutive drying procedures.
This procedure can be triggered whilst “tEmP “ is on the display screen by operating
the key. The balance switches automatically to weighing mode after a few
minutes.
Use the key to enter weighing mode.

MLB50-3-BA-defsi-0310 37
6.5 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated – in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location (
only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out during the initial start-up, after change in lo-
cation and variation of surrounding temperature. It is also recommendable to adjust
the balance periodically during weighing operation in order to obtain exact measured
values.
6.6 Adjusting
Using a precision weight, the accuracy of the balance can be checked at any time
and adjusted.
Method of approach when adjusting:
Ensure sound ambient conditions.
Use the main switch (above the power plug) to switch on the balance. ”tEmP “ will
appear on the display screen. Press the key, the balance will run through a ”tEst
“ mode. “0.000 “ will appear on the display screen after a few seconds. A short warm-
up period of approx. 10 minutes is advisable for stabilisation purposes.
Start up the adjusting process by operating the key. “------------- “ will appear on
the display screen for a few seconds.
The key must be pressed whilst “-------------“ is on the display screen.
Subsequently, “noCAL“ will appear on the display screen . “-LoAd-“ will appear on
the display screen after a few seconds, followed by the nominal value of the adjusting
weight.
The calibration process can be aborted during this period by pressing the key.
Now apply the adjusting weight and calibration will take place (please wait).
“ CAL “ will appear on the display screen, followed by ” rELoAd “ a few seconds
later.
Now remove the adjusting weight from the weighing plate.
The adjusting process is completed. The adjusting weight can be re-applied in order
to check the calibration. If the displayed value is not correct, adjusting should be re-
peated.

MLB50-3-BA-defsi-0310
38
7 Moisture determination
7.1 Use
The rapid determination of moisture content is of enormous significance wherever
moisture has to be added to or removed from a product during manufacture. For in-
numerable products, the moisture content is both a measure of quality and also an
important cost factor. When trading in industrial or agricultural products, as well as in
products in the chemical or foodstuffs field, firm limits for moisture content, defined by
sales agreements and standards, often apply.
7.2 Sample material
Samples with the following properties can usually be determined satisfactorily:
♦Granular or powdery, floatable solids.
♦thermally stable materials, which readily give off the moisture to be determined
without volatilising other constituents.
♦Liquids, which evaporate to dryness without forming a skin.
The determination can be difficult in samples which:
♦are viscous/sticky liquids.
♦tend to form a crust or skin readily on drying.
♦tend to decompose easily or liberate different constituents on heating.
7.3 Sample preparation
Special preparation of the samples is, in many cases, not required. Large solid bod-
ies, e.g. kernels and grains, should, however, be ground. Both the drying time and
the accuracy which can be achieved will be greatly influenced by the sample distribu-
tion. The result is that there are two contradictory requirements:
The lighter the load being weighed, the shorter the drying times which can be
achieved and the heavier the load, the more accurate will be the result.
Sample distribution:
Solid material
Distribute powdered and granular samples evenly over
the balance pan. Reduce the size of coarse samples
(grind, mortar)
Liquids
Apply tough and sticky products thinly. the use of circular
glass-fibre filters or quartz-sand is recommended.
Weight loss through spattering can be prevented by
covering the sample with a circular glass fibre filter. If
circular glass-fibre filters or quartz-sand are used the
weight must be subtract by tarring.

MLB50-3-BA-defsi-0310 39
7.4 Drying temperature
Consideration should be given to the following influencing factors when setting the
drying temperature:
Sample surface:
Contrary to powdery and granular samples, fluid and spreadable samples have a
smaller surface that tends to transfuse heat energy.
The use of a glass fibre filter improves heat insertion.
Colour of the sample:
Pale samples reflect more heat radiation than dark samples and therefore require a
higher drying temperature.
Availability of volatile matter:
The better and faster water or other volatile matter is available, the lower the setting
for the drying temperature. If the availability of water is poor (e.g. in plastics), the
water must be calcinated at a high temperature (the higher the temperature, the
higher the water vapour pressure).
Similar results can be achieved using other humidity identification methods (e.g.
cabinet dryer) by experimental optimisation of the setting parameters such as tem-
perature, heating grade and shut-down criteria.

MLB50-3-BA-defsi-0310
40
7.5 Application table (recommendations)
Preparation of the standard sample:
- Distribute the sample evenly on the aluminium tray.
- Reduce the sample to small pieces and distribute evenly on the aluminium
tray.
Preparation of special samples:
- A glass fibre filter can be used if the test material is sensitive or difficult to dis-
tribute (e.g. mercury).
- Apply the sample evenly to the glass fibre filter and cover with a second glass
fibre filter.
- The glass fibre filter can also be used for protection against material splashes
(every splash adulterates the final result).
MATERIAL Weight of
test item
(g)
Drying tem-
perature
(º C)
Date
query
interval
(s)
% humidity or
% solid body
Standard
divergence
Drying
time
(min)
Dry piece of
apple
5-8 100 10 76.5 0.1 10-15
Moist apple 5-8 100 10 7.5 5-10
Butter 2-5 138 15 16.3 0.1 4.5
Mustard 2-3 130 20 76.4 0.7 10
Ground cof-
fee
2-3 106 5 2.8 0.1 4
Cornflakes 2-4 120 15 9.7 5-7
Yogurt 2-3 110 15 86.5 4.5-6.5
Powdered
coffee cream
2-3 130 15 78.5 0.1 6-8
Cocoa pow-
der
2-3 106 20 0.1 0.1 2
Crisps 3-4 106 15 6.9 0.1 7.5
Margarine 3-4 138 20 16 0.1 10
Mayonnaise 1-2 138 20 56.5 0.4 10
Powdered
milk
2-4 90 15 5 0.2 6
Milk choco-
late
2-5 106 15 1.3 0.1 3.5
Red wine 3-5 100 15 97.4 0.1 15-20
Chocolate
powder
2-4 100 5 1.9 0.1 4
Sunflower oil 10-14 138 20 0.1 2
Instant soup
powder
2-3 80 15 3 0.2 4.5-7
Sugar 4-5 138 15 11.9 0.1 10
Milk 2-3 120 15 88 0.2 6-8
Flour 8-10 130 10 12.5 0.1 4-5
Cement 8-12 138 15 0.8 0.1 4-5
Paper 2-4 106 20 6.4 0.1 10
Sludge 11-12 130 15 80 90
Polyamide 2-5 138 20 2 0.2 75
Dust 5-10 104 10 7.3 0.3 8-15
Charcoal 8-10 120 10 3.8 0.1 8-10

MLB50-3-BA-defsi-0310 41
7.6 Temperature calibration / adjustment
The temperature can be checked and re-adjusted if necessary using the optional
temperature calibration set (MLB-A03).
Temperature calibration
- The balance is in normal weighing mode.
- Insert the temperature sensor through the balance cover opening and push in
close to the weighing plate (see pictures).
Guide bush for temperature sensor
-
-
- Press the key and then press the and keys simultaneously.
A “Code“ is now queried on the display screen.
- Use the arrow keys to enter the correct code (“002003“).
Numerical entry from 0-9
For changing the entry point on the display screen
- Press the key. “St_tE “ will appear on the display screen - the temperature
sensor setting programme.
- Press the key. The temperature inside the weighing region will now appear.
If this is not 20°C, please use the arrow keys to alter the temperature.
- Press the key. If a temperature of 20°C is entered, a number of approxi-
mately 4000 will appear on the display screen.
- The key is now used to start temperature calibration. The balance will now
heat up to 50% of its total power.

MLB50-3-BA-defsi-0310
42
- The balance should not be moved during this procedure in order to achieve a sta-
ble result. The balance will have achieved a stable temperature value after
approx. 15 minutes.
- The temperature in the weighing region should now have reached approx. 80°C.
- Press the key. “opEn“ will appear on the display screen.
- The cover can now be opened. + 2°C is now added on to the actual temperature,
e.g. if the thermometer is at 79°C, 79°C + 2°C = 81°C is entered.
2°C must be added to every temperature value.
- The thermometer can now be removed from the weighing region.
- The procedure is completed.
8 Setting the parameters
The following keys are used to set and select the parameters:
Function key – selects the parameters and adopts settings
Used to run through the various parameter settings, numerical entry
from 0-9
Used to change the entry point on the display screen
Parameter selection:
- The balance is in weighing mode. Press the key and ”SubSt“ will ap-
pear on the display screen.
(Details chapter: Temperature correction on ”SubSt “ test object)
- Press the key again and ”modE“ will appear on the display screen.
(Details chapter: Mode “modE “).
- Press the key again and ” intEr “ will appear on the display screen.
(Details chapter: Maximum drying time “IntEr “). Please note: This menu item
only appears in modes 4 - 7.
- Press the key again and ”tEmP“ will appear on the display screen.
(Details chapter: Drying temperature “ tEmP “)
- Press the key again and “Strob “ will appear on the display screen.
(Details chapter: Data query interval)

MLB50-3-BA-defsi-0310 43
8.1 Temperature correction on “SubSt“ test object
Certain materials give off more heat than is imported from outside.
Correction is necessary in this case in order to achieve an accurate measurement
result.
This setting allows the user to correct differences between the set humidity identifier
temperature and the actual temperature measured on the test item.
The balance is able to store 9 different correction factors.
These can be determined, set and selected as follows:
In order to set the appropriate correction factor it is necessary to carry out a drying
procedure and to measure the temperature in the interior of the test item.
Procedure:
The balance is in normal weighing mode. Apply a test quantity to the weighing plate.
Insert a thermo sensor into the test item.
The set drying parameters are:
- SubSt = 0
- modE = 7
- IntEr = 30:00 minutes or longer
- tEmP = a typical value for the sample
- Strob = 20 seconds
Press the key in order to display the parameters. Press the key again to
start the test. Calculation of the correction factor can be carried out if the temperature
in the test item proves to be stable after a certain period of time.
Formula: T (measured) – T (set)
Correction = ----------------------------------------------
T (set)
Example: The measured temperature amounts to 121°C, the temperature is set at
110°C, the correction is 0.10.
Typical materials with high temperature radiation:
Material Set temperature Measured
temperature
Correction factor
Flour 100°C 103.7°C 0.04
Coal dust 100°C 122°C 0.22
Damp tea leaves 100°C 120.5°C 0.20
Dry tea leaves 100°C 108.5°C 0.08
Cement 100°C 121°C 0.21

MLB50-3-BA-defsi-0310
44
8.1.1 Setting the correction factor
The balance is in weighing mode. Press the key and ” Subst “ will appear on
the display screen.
Use the key to select the correction factors (1-9).
Use the key to confirm selection. “ SubCoF “ will now appear on the display
screen and the temperature display flashes.
Now use the and keys to enter the determined correction factor. Use the
key to accept entry.
If the entered correction factor is > 0.99, “FALSE “ will appear on the display screen
and the entry will not be adopted. Repeat the entry using a new factor (<1) .
If you now activate the stored correction factor during testing, this will be allowed for.

MLB50-3-BA-defsi-0310 45
8.2 Modes “modE“
The following keys are used to set and select the parameters:
Function key – selects the parameters and adopts settings
Used to run through the various parameter settings, numerical entry
from 0-9
Used to change the entry point on the display screen
7 different modes are available.
Mode 1:
Percentage humidity (%)
Weight loss as against initial value
Drying is concluded if the weighing result of three consecutive heating intervals re-
mains unchanged.
Mode 2:
Percentage weight value (%) as against initial value
Drying is concluded if the weighing result of three consecutive heating intervals re-
mains unchanged.
Mode 3:
Percentage humidity as against solid body
Drying is concluded if the weighing result of three consecutive heating intervals re-
mains unchanged.
Mode 4:
Percentage humidity (%)
Weight loss as against initial value
Drying is concluded if the weighing result of three consecutive heating intervals re-
mains unchanged or if the time deadline has expired.
Mode 5:
Percentage weight value (%) as against initial value
Drying is concluded if the weighing result of three consecutive heating intervals re-
mains unchanged or if the time deadline has expired.
Mode 6:
Percentage humidity as against solid body
Drying is concluded if the weighing result of three consecutive heating intervals re-
mains unchanged or if the time deadline has expired.
Mode 7:
Percentage humidity (%)
Weight loss as against initial value
Drying is concluded if the time limit has expired.

MLB50-3-BA-defsi-0310
46
8.3 Maximum drying time “IntEr“
Drying modes 4 – 7 allow predefinition of the maximum drying period. This can be
predefined between 10 minutes and 9 hours 50 minutes.
8.4 Drying temperature “tEmP“
The drying temperature can be selected between 50°C and 160°C.
8.5 Data query time interval “Strob“
The “Strob“ interval time is the time between two points in time of the query result
data transfer via the RS-232 interface.
When using modes 1 – 6 the test is concluded if 3 query results are the same. This is
the case if no further moisture is able to escape from the sample.
Note:
The following keys are used to set and select the parameters (chapter 8.3 – 8.5):
Function key – selects the parameters and adopts settings
Used to run through the various parameter settings, numerical entry
from 0-9
Used to change the entry point on the display screen
9 Operation
9.1 Control elements
9.1.1 Overview of the keyboard
Function key – selects the parameters and adopts settings
Used to run through the various parameter settings, numerical entry
from 0-9
Used to change the entry point on the display screen
Transmits the test or weighing results to a printer via the RS-232 inter-
face.
Starts the drying procedure.
Sets the display to zero in weighing mode. The drying process can be
aborted by pressing the stop key.

MLB50-3-BA-defsi-0310 47
9.1.2 Overview of the display screen
9.2 Operation
9.2.1 Weighing
Use the main switch (above the power plug) to switch on the balance. A short warm-
up period of approx. 10 minutes is advisable for stabilisation purposes.
Press the key and wait for the “0“ indication. Now it is ready for use.
Important: If the display does not show “0“ press the key.
Only now (!) place object on the weighing pan. Make sure that the weighing object
does not stripe or touch the housing or base. Now the weight will be indicated.
If the object should be heavier than the weighing range allowance, “FULL“ (over-
load) will appear on the display.
Indicator for:
-Set time
-Test time
Indicator for:
-Temperature selection
-Process temperature
-Entr
y
of correction factor
Indicator for :
-Drying mode selection
(“modE“)
-Correction factor selection
(“subCoF“)
Main display panel:
- Function display
- Error messages
- Play back result
- Adjusting functions etc.
Table of contents
Other KERN Measuring Instrument manuals

KERN
KERN ORA 10 BA User manual

KERN
KERN KFE-TM User manual

KERN
KERN RPOMAT MSB 80 User manual

KERN
KERN ORD 45BM, ORD 92BM, ORD 85BM, ORD 92HM, ORD 1SM, ORD 3SM, ORD 2WM, ORD 1PM, ORD 2UM, ORD 5UM, ORD 6 US, ORD... User manual

KERN
KERN ORA 1WA/WB User manual

KERN
KERN MSF 200N User manual

KERN
KERN ORA 62 BA User manual

KERN
KERN TMBA-A01-A User manual

KERN
KERN MLB-A03 User manual

KERN
KERN MLS 50-3...N series User manual