Leader LV 5800A User manual

LV 5800A / LV 5800
MULTI MONITOR
INSTRUCTION MANUAL
(ETHERNET)

TABLE OF CONTENTS
1. INTRODUCTION.......................................................................................................... 1
1.1 About Network .......................................................................................................................... 1
1.2 Trademark Acknowledgments................................................................................................... 1
2. ETHERNET.................................................................................................................. 2
2.1 Remote Control Using TELNET................................................................................................ 2
2.2 FTP File Transfer ...................................................................................................................... 3
2.3 Command Entry Procedure ...................................................................................................... 4
2.4 TELNET Commands................................................................................................................. 5
2.4.1 LV 5800(A) Commands (MULTI MONITOR)...................................................................... 5
2.4.2 LV 58SER01A Commands (SDI INPUT) ........................................................................... 9
2.4.3 LV 58SER02 Commands (EYE PATTERN unit) .............................................................. 25
2.4.4 LV 58SER03 Commands (COMPOSITE VIDEO INPUT) ................................................ 30
2.4.5 LV 58SER04 Commands (MPEG DECODER) ................................................................ 35
2.4.6 LV 58SER40A Commands (DIGITAL AUDIO).................................................................. 43
2.5 FTP Commands...................................................................................................................... 52
3. SNMP......................................................................................................................... 53
3.1 Description.............................................................................................................................. 53
3.2 SNMP Version Supported....................................................................................................... 53
3.3 Setup ...................................................................................................................................... 53
3.3.1 Configuring the SNMP Manager...................................................................................... 53
3.3.2 Setting the LV 5800(A) .................................................................................................... 54
3.4 MIB ......................................................................................................................................... 55
3.4.1 Standard MIB .................................................................................................................. 55
3.4.2 Enterprise MIB................................................................................................................. 60
3.5 Enterprise Trap ....................................................................................................................... 88
3.5.1 Configuring the SNMP Manager...................................................................................... 88
3.5.2 Specific Trap.................................................................................................................... 89
3.5.3 Variable Binding List ........................................................................................................ 90
4. FIRMWARE REVISION HISTORY............................................................................. 92

1. INTRODUCTION
1
1. INTRODUCTION
This manual explains how to control an LV 5800(A) remotely over an Ethernet. For other
explanations and notes, see the main LV 5800(A) instruction manual.
1.1 About Network
The remote control capability operation via the Ethernet is confirmed only under the local
network environments; it is not guaranteed under other environments.
1.2 Trademark Acknowledgments
Windows is a registered trademark of Microsoft Corporation in the United States and/or other
countries.
Dolby is a trademark of Dolby Laboratories.

2. ETHERNET
2
2. ETHERNET
You can set nearly all settings that can be specified using the front panel keys by transferring the
command to LV 5800(A) using the PC that is connected via Ethernet network.
In addition, the various data that is detected by LV 5800(A) can be transferred to the PC that is
connected via Ethernet network.
See the main LV 5800(A) instruction manual for Ethernet settings (*1) to the main frame.
*1 For IP Address settings to LV 5800(A), press the SYS key to display the system menu. Then, press F•2
(PLATFORM SETUP) to display the tab menu to set up IP Address.
2.1 Remote Control Using TELNET
To control the LV 5800(A) remotely using a controller such as a PC connected to the network,
TELNET (*2) is used.
For a description on how to start (*3) TELNET, see the instruction manual that came with your
PC. The login name and password cannot be changed on the LV 5800(A).
When the TELNET connection is established, “login:” appears on the screen. Enter “LV5800”.
LV5800 login: LV5800
Next “Password:” appears. Enter “LV5800“.
Actually, letter of password is not displayed.
LV5800 login: LV5800
Password: LV5800
Appears「LV5800>」.
LV5800 login: LV5800
Password: LV5800
LV5800>
*2 Flow control for TELNET remote control
If you are remotely controlling the LV 5800(A) via TELNET, enable the TELNET flow control. If your
TELNET client does not support flow control, the LV 5800(A) may not operate correctly if commands are
sent at high speeds.
If you cannot use flow control, (1) allow approximately 1-second interval between each command or (2)
send the REMOTE:REPLY command described in section 2.4.1, “LV 5800(A) Commands (MULTI
MONITOR)” to enable return values and perform software handshaking.
*3 To start TELNET on Windows 7
On the task bar, click Start, and then click Run.
Type “TELNET” followed by a space and then the LV 5800(A) IP address. Click OK.
(To use TELNET, open Control Panel, click Turn Windows features on or off under Program and Features,
and select the Telnet Client check box.)

2. ETHERNET
3
2.2 FTP File Transfer
To transfer files from LV 5800(A) to PC connected to the network, FTP is used. For a
description on how to start (*1) FTP, see the instruction manual that came with your PC.
The user name and password cannot be changed on the LV 5800(A).
When the FTP starts, “User” appears on the screen. Enter “LV5800”.
Connected to xxx.xxx.xxx.xxx
220 FTP Server ready
User(xxx.xxx.xxx.xxx:(none)): LV5800
Next “Password:” appears. Enter “LV5800“.
Actually, letter of password is not displayed.
Connected to xxx.xxx.xxx.xxx
220 FTP Server ready
User(xxx.xxx.xxx.xxx:(none)): LV5800
331 Password required
Password: LV5800
Appears「ftp>」.
Connected to xxx.xxx.xxx.xxx
220 FTP Server ready
User(xxx.xxx.xxx.xxx:(none)): LV5800
331 Password required
Password: LV5800
230 Logged in
ftp>
*1 To start FTP on Windows 7
On the task bar, click Start, and then click Run.
Type “FTP” followed by a space and then the IP address. Click OK.

2. ETHERNET
4
2.3 Command Entry Procedure
The system of control commands used on the Ethernet network follows the menu structure.
Reading this manual and the instruction manuals of each Unit of the LV 5800(A) will facilitate
the process.
●Entry Method of the Command
The entry method of the command is shown below. The control commands are entered
using the following syntax: the command followed by a space followed by commands.
The parameter can be used either of: none, 1, or 2 pieces.
[Command] + [Space] + [Parameter1] + [Space] + [Parameter2]
* You can use uppercase or lowercase characters for commands.
* You must specify a unit number for commands that start with [SYS:UNIT:]. Enter a unit number (1 to 6)
in [Parameter 1]. To check which unit is installed to a unit number (n), use the [SYS:UNITn_INFO]
command.
* Measurement display commands that start with [WFM], [VECT], [PIC], [AUDIO], [STATUS], and [EYE]
apply only to the display specified by the [DISPLAY] command. If you enter a command with a mode
different from the current display mode, the command is discarded.
●Control Command Entry Example
The command entry example is shown below.
・When there is no parameter.
STATUS:RESET [Enter]
(The error of the status screen is reset.)
・When there is a 1 parameter.
WFM:GAIN:MAG X5 [Enter]
(Sets the gain of the video signal waveform display to 5 times.)
・When there are 2 parameters.
SYS:UNIT:LINKFORMAT 1 SINGLE [Enter]
(Sets the link format of SDI Input installed in unit 1 to single link.)

2. ETHERNET
5
2.4 TELNET Commands
2.4.1 LV 5800(A) Commands (MULTI MONITOR)
●Front Panel LV 5800(A) (MULTI MONITOR)
Command Parameter1 Description
DISPLAY 1 Selects first area
2 Selects second area
3 Selects third area
4 Selects fourth area
? Queries the selected area
MULTI ON Selects multi screen display
OFF Selects 1 screen display
? Queries the multi screen display
INPUT:UNIT UNIT1 Selects unit 1
UNIT2 Selects unit 2
UNIT3 Selects unit 3
UNIT4 Selects unit 4
? Queries the selected unit
INPUT:CH A Selects channel A
B Selects channel B
? Queries the input channel
MODE WFM Selects the video signal waveform
VECT Selects vector waveform
PIC Selects picture display
AUDIO Selects audio display
STATUS Selects status display
EYE Selects eye pattern display
? Queries the measurement mode
EXT INT Selects internal sync signal
EXT Selects external sync signal
? Queries the sync signal
RCLL 1 to 60 Recall presets memory
MAKE CAPTURE Captures the displayed screen to the internal memory of the
LV 5800(A)

2. ETHERNET
6
●PLATFORM SETUP (GENERAL SETUP) LV 5800(A) (MULTI MONITOR)
Command Parameter1 Description
SYS:PLATFORM:SOURCE_MODE SINGLE Sets the single input mode.
MULTI Sets the multi input mode. (Doesn’t interlock with the
line selection.)
MULTI_LINE Sets the multi input mode. (Interlocks with the line
selection.)
? Queries the input mode.
SYS:PLATFORM:DISPLAY:
MULTI_MODE
2MULTI Sets the 2 screen display when the multi screen is
selected
4MULTI Sets the 4 screen display when the multi screen is
selected
? Queries the multi screen display
SYS:PLATFORM:WINDOW_MARKER OFF Sets the selected window to a colorlessness frame.
BLUE Sets the selected window to a blue frame.
WHITE Sets the selected window to a white frame.
? Queries the frame color of selected window.
SYS:PLATFORM:CAPTURE_MODE SCREEN Sets the capture mode to screen capture.
VIDEO_FRAME
Sets the capture mode to frame capture.
? Queries the capture mode.
SYS:PLATFORM:DISPLAY:INFO:FORMAT ON The format is displayed.
OFF The format is not displayed.
? Queries the presence of format display setting.
SYS:PLATFORM:DISPLAY:INFO:DATE OFF The date is not displayed.
YMD
Displays the date in order of the year/ month / day.
MDY Displays the date in order of the month/ day/ year.
DMY
Displays the date in order of the day/ month / year.
? Queries the presence of date display setting and display
type.
SYS:PLATFORM:DISPLAY:INFO:TIME ON The time is displayed.
OFF Time is not displayed.
? Queries the presence of time display setting.
SYS:PLATFORM:DISPLAY:INFO:COLOR ON The color system is displayed.
OFF The color system is not displayed.
? Queries the presence of color system display
SYS:PLATFORM:DISPLAY:INFO:INPUT ON Number of input unit and the input channel are
displayed.
OFF Number of input unit and the input channel are not
displayed.
?
Queries the presence of setting for displaying both the
number of input unit and the input channel.
SYS:PLATFORM:DISPLAY:BACKLIGHT HIGH Brightens the backlight.
LOW Dims the backlight.
? Queries the brightness of the backlight.
SYS:PLATFORM:DISPLAY:AUTO_OFF OFF The backlight is always turned on.
5MIN The backlight is shut off 5 minutes after pressing the last
key.
30MIN The backlight is shut off 30 minutes after pressing the
last key.
60MIN The backlight is shut off 60 minutes after pressing the
last key.
? Queries the presence of auto shutoff setting of the
backlight and the time of shutoff of the backlight.

2. ETHERNET
7
●PLATFORM SETUP (REMOTE) LV 5800(A) (MULTI MONITOR)
Command Parameter1 Description
SYS:PLATFORM:REMOTE_MODE BIT Recalls the presets from No.1 to 8 by bit.
BINARY Recalls the presets from No.1 to 60 by binary code.
? Queries the remote mode.
SYS:PLATFORM:ALARM:POLARITY POSITIVE High is output at the time of error detection
NEGATIVE Low is output at the time of error detection
?
Queries the output polarity at the time of error detection.
SYS:PLATFORM:ALARM1:UNIT UNIT1 Outputs the error of unit No.1 to alarm 1.
UNIT2 Outputs the error of unit No.2 to alarm 1.
UNIT3 Outputs the error of unit No.3 to alarm 1.
UNIT4 Outputs the error of unit No.4 to alarm 1.
ALL Outputs the error of all units to alarm 1.
? Outputs the error of unit from No.1 to 4 to alarm 1.
SYS:PLATFORM:ALARM2:UNIT UNIT1 Outputs the error of unit No.1 to alarm 2.
UNIT2 Outputs the error of unit No.2 to alarm 2.
UNIT3 Outputs the error of unit No.3 to alarm 2.
UNIT4 Outputs the error of unit No.4 to alarm 2.
ALL Outputs the error of unit from No.1 to 4 to alarm 2.
? Queries the unit that outputs the error to alarm 2.
SYS:PLATFORM:ALARM3:UNIT UNIT1 Outputs the error of unit No.1 to alarm 3.
UNIT2 Outputs the error of unit No.2 to alarm 3.
UNIT3 Outputs the error of unit No.3 to alarm 3.
UNIT4 Outputs the error of unit No.4 to alarm 3.
ALL Outputs the error of unit from No.1 to 4 to alarm 3.
? Queries the unit that outputs the error to alarm 3.
SYS:PLATFORM:ALARM4:UNIT UNIT1 Outputs the error of unit No.1 to alarm 4.
UNIT2 Outputs the error of unit No.2 to alarm 4.
UNIT3 Outputs the error of unit No.3 to alarm 4.
UNIT4 Outputs the error of unit No.4 to alarm 4.
ALL Outputs the error of unit from No.1 to 4 to alarm 4.
? Queries the unit that outputs the error to alarm 4.
SYS:PLATFORM:ALARM1:CH A Outputs the error of A channel to alarm 1.
B Outputs the error of B channel to alarm 1.
A/B Outputs the error of A/B channels to alarm 1.
? Queries the channel that outputs the error to alarm 1.
SYS:PLATFORM:ALARM2:CH A Outputs the error of A channel to alarm 2.
B Outputs the error of B channel to alarm 2.
A/B Outputs the error of A/B channels to alarm 2.
? Queries the channel that outputs the error to alarm 2.
SYS:PLATFORM:ALARM3:CH A Outputs the error of A channel to alarm 3.
B Outputs the error of B channel to alarm 3.
A/B Outputs the error of A/B channels to alarm 3.
? Queries the channel that outputs the error to alarm 3.
SYS:PLATFORM:ALARM4:CH A Outputs the error of A channel to alarm 4.
B Outputs the error of B channel to alarm 4.
A/B Outputs the error of A/B channels to alarm 4.
? Queries the channel that outputs the error to alarm 4.
SYS:PLATFORM:ERROR_BEEP ENABLE Sounds the Beep Alarm when error is detected.
DISABLE Does not sound the Beep Alarm when error is detected.
? Queries whether to sound Beep Alarm when error is
detected

2. ETHERNET
8
●DATE&TIME LV 5800(A) (MULTI MONITOR)
Command Parameter1 Description
SYS:DATE YYYY/MM/DD,hh:mm:ss Sets up the date and time.
? Queries the date and time
●SYSTEM INFORMATION LV 5800(A) (MULTI MONITOR)
Command Parameter1 Description
SYS:UNIT1_INFO
-
Queries the unit 1 type
SYS:UNIT2_INFO - Queries the unit 2 type
SYS:UNIT3_INFO
-
Queries the unit 3 type
SYS:UNIT4_INFO - Queries the unit 4 type
SYS:UNIT5_INFO
-
Queries the unit 5 type
SYS:UNIT6_INFO - Queries the unit 6 type
●LCD OFF LV 5800(A) (MULTI MONITOR)
Command Parameter1 Description
SYS:LCD_OFF - Shuts off the LED.
●INITIALIZE LV 5800(A) (MULTI MONITOR)
Command Parameter1 Description
SYS:INIT
-
Initializes the LV 5800(A).
●Miscellaneous LV 5800(A) (MULTI MONITOR)
Command Parameter1 Description
REMOTE:HELP - Displays the list of the commands.
REMOTE:REPLY (*1) ON Replies for the command.
OFF Does not reply for the command.(Default setting)
*1 Return value when REMOTE:REPLY is ON
If the LV 5800(A) is in the factory default condition (REMOTE:REPLY = OFF), the LV 5800(A) does not return
a value even when a command has an error during TELNET remote control. It only returns a value for queries
using a “?” parameter. This is done to maintain compatibility with the LV 5700.
By setting the REMOTE:REPLY command parameter to ON, the LV 5800(A) will return a value as follows:
OK: If the command is processed properly.
ERR1: If the number of parameters is incorrect.
ERR2: If a parameter value is out of range.
ERR3: If the function is not enabled in the current condition.
If the command itself is wrong, the LV 5800(A) will return the following character string.
-rbash:the incorrect command:command not found

2. ETHERNET
9
2.4.2 LV 58SER01A Commands (SDI INPUT)
●UNIT SETUP (GENERAL SETUP) LV 58SER01A (SDI INPUT)
Command Parameter1 Parameter2 Description
SYS:UNIT:LINK_FORMAT 1,2,3,4 (*1) SINGLE Sets the link format to single link.
DUAL Sets the link format to dual link.
2K Sets the link format to 2048 × 1080 (2K)
of dual link.
? Queries the link format
SYS:UNIT:MANUAL_SELECT 1,2,3,4 AUTO Sets the format automatically at the time
of single link.
MANUAL Sets the format manually at the time of
single link.
? Queries the format at the time of single
link.
SYS:UNIT:MANUAL_FORMAT 1,2,3,4 1080I/60 Sets the format to 1080i/60 at the time
of single link.
1080I/59.94 Sets the format to 1080i/59.94 at the
time of single link.
1080I/50 Sets the format to 1080i/50 at the time
of single link.
1080SF/30 Sets the format to 1080PsF/30 at the
time of single link.
1080SF/29.97 Sets the format to 1080PsF/29.97 at the
time of single link.
1080SF/25 Sets the format to 1080PsF/25 at the
time of single link.
1080SF/24 Sets the format to 1080PsF/24 at the
time of single link.
1080SF/23.98 Sets the format to 1080PsF/23.98 at the
time of single link.
1080P/30 Sets the format to 1080p/30 at the time
of single link.
1080P/29.97 Sets the format to 1080p/29.97 at the
time of single link.
1080P/25 Sets the format to 1080p/25 at the time
of single link.
1080P/24 Sets the format to 1080PsF/24 at the
time of single link.
1080P/23.98 Sets the format to 1080PsF/23.98 at the
time of single link.
720P/60 Sets the format to 720p/60 at the time of
single link.
720P/59.94 Sets the format to 720p/59.94 at the
time of single link.
720P/50 Sets the format to 720p/50 at the time of
single link.
720P/30 Sets the format to 720p/30 at the time of
single link.
720P/29.97 Sets the format to 720p/29.97 at the
time of single link.
720P/25 Sets the format to 720p/25 at the time of
single link.
*1 Enters the unit number in which LV 58SER01A is being installed.

2. ETHERNET
10
Command Parameter1 Parameter2 Description
SYS:UNIT:MANUAL_FORMAT 1,2,3,4 720P/24 Sets the format to 720p/24 at the time of
single link.
720P/23.98 Sets the format to 720p/23.98 at the
time of single link.
525I/59.94 Sets the format to 525i/59.94 at the time
of single link.
625I/50 Sets the format to 625i/50 at the time of
single link.
SYS:UNIT:I_PSF 1,2,3,4 INTERLACE Sets the format display type to the
interlace at the time of single link.
SEG.FRAM Sets the format display type to the
segmented frame at the time of single
link.
? Queries the format display type at the
time of single link.
SYS:UNIT:DUAL:SYSTEM 1,2,3,4 GBR_444 Sets the color system to GBR (4:4:4) at
the time of dual link.
YCBCR_422 Sets the color system to YCbCr (4:2:2)
at the time of dual link.
? Queries the color system at the time of
dual link.
SYS:UNIT:DUAL:DEPTH 1,2,3,4 10B Sets the number of quantization bits per
pixel to 10-bit at the time of dual link.
12B Sets the number of quantization bits per
pixel to 12-bit at the time of dual link.
? Queries the number of quantization bits
per pixel at the time of dual link.
SYS:UNIT:DUAL:SCAN 1,2,3,4 1080I Sets the scan mode to the interlace at
the time of dual link.
1080PSF Sets the scan mode to the
segment-frame at the time of dual link.
1080P Sets the scan mode to the progressive
at the time of dual link.
? Queries the scan mode at the time of
dual link.
SYS:UNIT:INFO:TIME_CODE 1,2,3,4 REAL Use the built-in time function of main
frame for the time display.
LTC Use the LTC (time code) for the time
display.
VITC Use the VITC (time code) for the time
display.
D-VITC Use the D-VITC (time code) for the time
display.
? Queries the time display mode being
displayed.
SYS:UNIT:SELECT_OUTPUT 1,2,3,4 A OUTPUT SDI A/B becomes the reclock
output of A channel during single link.
A/B OUTPUT SDI A/B becomes the reclock
output of active channel during single
link.
? Queries the setting of the OUTPUT SDI
A/B during single link.

2. ETHERNET
11
●UNIT SETUP (ERROR SETUP1) LV 58SER01A (SDI INPUT)
Command Parameter1 Parameter2 Description
SYS:UNIT:ERROR:TRS 1,2,3,4 ON Detects the TRS errors.
OFF Does not detect the TRS errors.
? Queries the presence of TRS error
detection.
SYS:UNIT:ERROR:HD_LINE 1,2,3,4 ON Detects the line number errors.
OFF Does not detect the line number errors.
? Queries the presence of line number
error detection.
SYS:UNIT:ERROR:HD_CRC 1,2,3,4 ON Detects the CRC errors.
OFF Does not detect the CRC errors.
? Queries the presence of CRC error
detection.
SYS:UNIT:ERROR:SD_EDH 1,2,3,4 ON Detects the EDH errors.
OFF Does not detect the EDH errors.
? Queries the existence of EDH error
detection function.
SYS:UNIT:ERROR:ILLEGAL_CODE 1,2,3,4 ON Detects the illegal code errors.
OFF Does not detect the illegal code errors.
? Queries the presence of illegal code
error detection.
SYS:UNIT:ERROR:CABLE 1,2,3,4 ON Detects the cable errors.
OFF Does not detect the cable errors.
? Queries the presence of cable error
detection.
SYS:UNIT:ERROR:HD_CABLE 1,2,3,4 LS_5CFB Sets the cable type to FS_5CFB for
HD-SDI input.
1694A Sets the cable type to 1694A for
HD-SDI input.
L_7CHD Sets the cable type to L-7CHD for
HD-SDI input.
? Queries the cable type for HD-SDI
input.
SYS:UNIT:ERROR:HD_LENGTH 1,2,3,4 5 to 200 Sets the cable length (m) considered as
an error for the HD-SDI input.
? Queries the cable length considered as
an error for HD-SDI input.
SYS:UNIT:ERROR:HD_WARN 1,2,3,4 5 to 200 Sets the cable length (m) considered as
a warning for the HD-SDI input.
? Queries the cable length considered as
a warning for HD-SDI input.
SYS:UNIT:ERROR:SD_CABLE 1,2,3,4 L_5C2V Sets the cable type to L_5C2V for
SD-SDI input.
8281? Sets the cable type to 8281 for SD-SDI
input.
1505A Sets the cable type to 1505A for SD-SDI
input.
? Queries the cable type for SD-SDI input.
SYS:UNIT:ERROR:SD_LENGTH 1,2,3,4 50 to 300 Sets the cable length (m) considered as
an error for the SD-SDI input.
? Queries the cable length considered as
an error for SD-SDI input.
SYS:UNIT:ERROR:SD_WARN 1,2,3,4 50 to 300 Sets the cable length (m) considered as
a warning for the SD-SDI input.
? Queries the cable length considered as
a warning for SD-SDI input.

2. ETHERNET
12
●UNIT SETUP (ERROR SETUP2) LV 58SER01A (SDI INPUT)
Command Parameter1 Parameter2 Description
SYS:UNIT:ERROR:ANC:PARITY 1,2,3,4 ON Detects the parity errors in the ancillary
data.
OFF Does not detect the parity errors in the
ancillary data.
? Queries the presence of parity error
detection in the ancillary data.
SYS:UNIT:ERROR:ANC:CHECKSUM 1,2,3,4 ON Detects the checksum errors in the
ancillary data.
OFF Does not detect the checksum errors in
the ancillary data.
? Queries the presence of checksum error
detection in the ancillary data.
SYS:UNIT:ERROR:AUDIO:BCH 1,2,3,4 ON Detects the BCH errors in the embedded
audio.
OFF Does not detect the BCH errors in the
embedded audio.
? Queries the presence of BCH error
detection in the embedded audio.
SYS:UNIT:ERROR:AUDIO:DBN 1,2,3,4 ON Detects the DBN errors in the embedded
audio.
OFF Does not detect the DBN errors in the
embedded audio.
? Queries the presence of DBN error
detection in the embedded audio.
SYS:UNIT:ERROR:AUDIO:PARITY 1,2,3,4 ON Detects the parity errors in the embedded
audio.
OFF Does not detect the parity errors in the
embedded audio.
? Queries the presence of parity error
detection in the embedded audio.
SYS:UNIT:ERROR:AUDIO:INHIBIT 1,2,3,4 ON Detects the inhibit errors in the embedded
audio.
OFF Does not detect the inhibit errors in the
embedded audio.
? Queries the presence of inhibit error
detection in the embedded audio.

2. ETHERNET
13
●UNIT SETUP (ERROR SETUP3) LV 58SER01A (SDI INPUT)
Command Parameter1 Parameter2 Description
SYS:UNIT:ERROR:GAMUT:LPF 1,2,3,4
HD1M_SD1M Sets the LPF for gamut error detection to 1
MHz.
HD2.8M_SD1M
Sets the LPF for gamut error detection to
2.8 MHz as for HD-SDI and to 1 MHz as for
SD-SDI.
OFF Turns off the LPF for gamut error detection.
? Queries the LPF setting for gamut error
detection.
SYS:UNIT:ERROR:GAMUT 1,2,3,4
ON Detects the gamut errors.
OFF Does not detect the gamut errors.
? Queries the presence of gamut error
detection.
SYS:UNIT:ERROR:GAMUT:UPPER 1,2,3,4
90.8 to 109.4 Sets the upper limit (%) of gamut error
? Queries the upper limit of the gamut error.
SYS:UNIT:ERROR:GAMUT:LOWER 1,2,3,4
-7.2 to 6.1 Sets the lower limit (%) of gamut error
? Queries the lower limit of the gamut error.
SYS:UNIT:ERROR:GAMUT:AREA 1,2,3,4
0.1 to 5.0 Sets the area ratio (%) for gamut error
detection.
? Queries the area ratio for gamut error
detection.
SYS:UNIT:ERROR:GAMUT:DURATION 1,2,3,4
1 to 60 Sets the duration (frames) of detection for
gamut error.
? Queries the duration of detection for gamut
error.
SYS:UNIT:ERROR:C_GAMUT 1,2,3,4
ON Detects the composite gamut errors.
OFF Does not detect the composite gamut
errors.
? Queries the presence of composite gamut
error detection.
SYS:UNIT:ERROR:C_GAMUT:SETUP 1,2,3,4
0P Sets the setup to exclude when converting
the composite.
7.5P Sets the setup to 7.5 % when converting
the composite.
? Queries the setup value when converting
the composite.
SYS:UNIT:ERROR:C_GAMUT:UPPER 1,2,3,4
90.0 to 135.0 Sets the upper limit (%) of composite
gamut error
? Queries the upper limit of the composite
gamut error.
SYS:UNIT:ERROR:C_GAMUT:LOWER 1,2,3,4
-40.0 to 20.0 Sets the lower limit (%) of composite
gamut error
? Queries the lower limit of the composite
gamut error.
SYS:UNIT:ERROR:C_GAMUT:AREA 1,2,3,4
0.1 to 5.0 Sets the detection area (%) for composite
gamut error.
? Queries the detection area (%) for
composite gamut error.
SYS:UNIT:ERROR:C_GAMUT:DURATION 1,2,3,4
1 to 60 Sets the duration of detection (frames) for
gamut error.
? Queries the duration of detection for gamut
error.

2. ETHERNET
14
●UNIT SETUP (ERROR SETUP4) LV 58SER01A (SDI INPUT)
Command Parameter1 Parameter2 Description
SYS:UNIT:ERROR:FREEZE 1,2,3,4
ON Detects the freeze errors.
OFF Does not detect the freeze errors.
? Queries the presence of freeze
error detection.
SYS:UNIT:ERROR:FREEZE:UPPER 1,2,3,4
0 to 100 Sets the detection area (upper) for
freeze error. (%)
? Queries the detection area (upper)
for freeze error.
SYS:UNIT:ERROR:FREEZE:LOWER 1,2,3,4
0 to 100 Sets the detection area (lower) for
freeze error. (%)
? Queries the detection area (lower)
for freeze error.
SYS:UNIT:ERROR:FREEZE:LEFT 1,2,3,4
0 to 100 Sets the detection area (left side)
for freeze error. (%)
? Queries the detection area (left
side) for freeze error.
SYS:UNIT:ERROR:FREEZE:RIGHT 1,2,3,4
0 to 100 Sets the detection area (right side)
for freeze error. (%)
? Queries the detection area (right
side) for freeze error. (%)
SYS:UNIT:ERROR:FREEZE:DURATION 1,2,3,4
2 to 300 Sets the duration of detection
(frames) for freeze error.
? Queries the duration of detection
for freeze error.
SYS:UNIT:ERROR:BLACK 1,2,3,4
ON Detects the black errors.
OFF Does not detect the black errors.
? Queries the presence of black error
detection.
SYS:UNIT:ERROR:BLACK:LEVEL 1,2,3,4
0 to 100 Sets the detection level (%) for
black error.
? Queries the detection level for
black error.
SYS:UNIT:ERROR:BLACK:AREA 1,2,3,4
1 to 100 Sets the detection area (%) for
black error.
? Queries the detection area for black
error.
SYS:UNIT:ERROR:BLACK:DURATION 1,2,3,4
1 to 300 Sets the duration of detection
(frames) for black error.
? Queries the duration of detection
for black error.

2. ETHERNET
15
●UNIT SETUP (ERROR SETUP5) LV 58SER01A (SDI INPUT)
Command Parameter1 Parameter2 Description
SYS:UNIT:ERROR:LEVEL 1,2,3,4
ON Detects the level errors.
OFF Does not detect the level errors.
? Queries the presence of level error
detection.
SYS:UNIT:ERROR:LEVEL:RUMA:UPPER 1,2,3,4
-51 to 766 Sets the upper limit (mV) of luminance
level.
? Queries the upper limit of luminance
level.
SYS:UNIT:ERROR:LEVEL:RUMA:LOWER 1,2,3,4
-51 to 766 Sets the lower limit (mV) of luminance
level.
? Queries the lower limit of luminance
level.
SYS:UNIT:ERROR:LEVEL:CHROMA:UPPER 1,2,3,4
-400 to 399 Sets the upper limit (mV) of chroma
difference level.
? Queries the upper limit of chroma
difference level.
SYS:UNIT:ERROR:LEVEL:CHROMA:LOWER 1,2,3,4
-400 to 399 Sets the lower limit (mV) of chroma
difference level.
? Queries the lower limit of chroma
difference level.
●WFM LV 58SER01A (SDI INPUT)
Command
Parameter 1
Description
WFM - Displays the video signal waveform.
WFM:CH1
ON
Displays CH1.
OFF Does not display CH1
?
Queries the presence of CH1 display.
WFM:CH2 ON Displays CH2.
OFF
Does not display CH2
? Queries the presence of CH2 display.
WFM:CH3
ON
Displays CH3.
OFF Does not display CH3
?
Queries the presence of CH3 display.
WFM:OVLAY ON Sets overlay display.
OFF
Sets parade display.
? Queries the display mode.

2. ETHERNET
16
●WFM - INTEN/SCALE LV 58SER01A (SDI INPUT)
Command Parameter1 Description
WFM:INTEN:WFM -128 to 127 Adjusts the intensity of the video waveforms.
? Queries the intensity of the video waveforms.
WFM:COLOR WHITE Sets the waveform display color to white.
GREEN Sets the waveform display color to green.
MULTI Sets the waveform display color as follows: Y: Yellow, Cb:
Cyan, Cr: Magenta, G: Green, B: Blue, and R: Red.
? Queries the waveform display color.
WFM:INTEN:SCALE -8 to 7 Adjusts the intensity of the scale.
? Queries the intensity of the scale.
WFM:SCALE:UNIT HDV_SDP Sets the scale unit to V for HD-SDI and % for SD-SDI.
HDV_SDV Sets the scale unit to V.
HDP_SDP Sets the scale unit to %.
? Queries the scale unit.
WFM:SCALE:COLOR75P ON Displays the 75% scale.
OFF Does not display the 75% scale.
? Queries the presence of the 75% scale display.
WFM:SCALE:COLOR WHITE Sets the scale color to white.
YELLOW Sets the scale color to yellow.
CYA N Sets the scale color to cyan.
GREEN Sets the scale color to green.
MAGENTA Sets the scale color to magenta.
RED Sets the scale color to red.
BLUE Sets the scale color to blue.
? Queries the scale color.
●WFM - GAIN/SWEEP LV 58SER01A (SDI INPUT)
Command
Parameter1
Description
WFM:GAIN:VAR CAL Sets the gain mode of the waveform to a constant (x1).
VAR Sets the gain mode of the waveform to variable (×0.200 to
×2.000).
? Queries the gain mode setting.
WFM:GAIN:VAL
0.200 to 2.000
Sets the variable gain value of the waveform.
? Queries the variable gain value of the waveform.
WFM:GAIN:MAG
X1
Sets the magnification of the waveform to x1.
X5 Sets the magnification of the waveform to x5.
?
Queries the magnification of the waveform.
WFM:SWEEP:SWEEP H Sets the sweep mode to the line display.
V
Sets the sweep mode to the field or frame display.
? Queries the sweep mode setting.
WFM:SWEEP:H_SWEEP
1H
Sets 1 line display.
2H Sets 2 line display.
?
Queries the line display setting.
WFM:SWEEP:V_SWEEP 1V Sets 1 frame display if the input signal is progressive.
Sets 1 field display if the input signal is interlace or
segmented frame.
2V
Sets 1 frame display.
? Queries the field or frame display setting

2. ETHERNET
17
Command Parameter1 Description
WFM:SWEEP:H_MAG X1 In case of line display, set the sweep magnification so that
the waveform fits on the screen.
X10 In case of line display, set the sweep magnification to x10
with respect to x1.
X20 In case of line display, set the sweep magnification to x20
with respect to x1.
ACTIVE In case of line display, display to magnify the waveform in the
active period.
BLANK In case of line display, display to magnify the waveform in the
H blanking period.
? Queries the sweep magnification mode in case of line
display.
WFM:SWEEP:V_MAG X1 In case of field (frame) display, set the sweep magnification
so that the waveform fits on the screen.
X20 In case of field (frame) display, set the sweep magnification
to x20 with respect to X1.
X40 In case of field (frame) display, set the sweep magnification
to x40 with respect to X1.
? Queries the sweep magnification mode in case of field
(frame) display.
WFM:SWEEP:FIELD
FIELD1
Displays the waveform of field 1.
FIELD2 Displays the waveform of field 2.
?
Queries the selected field.
●WFM - LINE SELECT LV 58SER01A (SDI INPUT)
Command Parameter1 Description
WFM:LINE_SELECT ON Enables the line select function.
OFF Disables the line select function.
? Queries the line select function.
WFM:LINE_FIELD FIELD1 Sets the selection range of the line select to field 1.
FIELD2 Sets the selection range of the line select to field 2.
FRAME Sets the selection range of the line select to frame.
? Queries the selection range of the line select.
WFM:LINE_NUMBER Example)
1 to 1125
(Depending on
the format)
Sets the line of the line select.
? Queries the selected line of the line select.
●WFM - ARRANGE LV 58SER01A (SDI INPUT)
Command
Parameter1
Description
WFM:FILTER:NORMAL FLAT Sets the filter to Flat for the component display.
LOWPASS
Sets the filter to LPF for the component display.
? Queries the filter setting for the component display.
WFM:FILTER:COMPOSITE
FLAT
Sets the filter to Flat when displaying the pseudo-composite.
FLAT_LUM Sets the filter to Flat and to parade display of the luminance
signal when displaying the pseudo-composite.
LUM_CHROMA Sets the filter to luminance signal and to parade display of
the color signal when displaying the pseudo-composite.
? Queries the filter setting when displaying the
pseudo-composite.

2. ETHERNET
18
Command Parameter1 Description
WFM:BLANKING:NORMAL
REMOVE
Displays the active period for the component display.
H_VIEW Displays the active period and the horizontal blanking period
for the component display.
V_VIEW Displays the active period and the vertical blanking period for
the component display.
ALL_VIEW Displays the entire input signal for the component display.
? Queries the blanking period display for the component
display.
WFM:BLANKING:COMPOSITE REMOVE Displays the active period when displaying the
pseudo-composite.
V_VIEW Displays the active period and the vertical blanking period
when displaying the pseudo-composite.
? Queries the blanking period display when displaying the
pseudo-composite.
WFM:PERSISTENCE
ON
Sets the persistence display to enabe.
OFF Sets the persistence display to disable.
INFINIT
Overlays the waveform.
? Queries the persistence display setting.
WFM:PERSIST_CLEAR
-
Clears the overlaid waveforms.
WFM:SPECIAL_FORM NORMAL Sets the SPECIAL FORM display to disable.
TIMING
Sets the SPECIAL FORM display to TIMING.
4Y_PARADE Sets the SPECIAL FORM display to 4Y PARADE.
?
Queries the SPECIAL FORM display setting.
●WFM - COLOR SYSTEM LV 58SER01A (SDI INPUT)
Command Parameter1 Description
WFM:MATRIX YCBCR Sets the color matrix to YCbCr display.
GBR Sets the color matrix to GBR display.
RGB Sets the color matrix to RGB display.
COMPOSITE Sets the color matrix to pseudo-composite display.
? Queries the color matrix setting.
WFM:MATRIX:YGBR ON Displays the luminance signal when displaying RGB.
OFF Does not display the luminance signal when displaying RGB.
? Queries the presence of luminance signal when displaying
RGB.
WFM:MATRIX:YRGB ON Displays the luminance signal when displaying RGB.
OFF Does not display the luminance signal when displaying RGB.
? Queries the presence of luminance signal when displaying
RGB.
WFM:MATRIX:COMPOSITE:FORMAT
AUTO Selects NTSC or PAL automatically when displaying the
pseudo-composite.
NTSC Displays NTSC when displaying the pseudo-composite.
PAL Displays PAL when displaying the pseudo-composite.
? Queries the display format of the pseudo-composite display.
WFM:MATRIX:SETUP 0P Sets the setup to exclude when displaying the
pseudo-composite.
7.5P Sets the setup to 7.5% when displaying the
pseudo-composite.
? Queries the setup value of the pseudo-composite display.
This manual suits for next models
1
Table of contents
Other Leader Monitor manuals

Leader
Leader LV5307 User manual

Leader
Leader LV 5490 User manual

Leader
Leader LV 5800 User manual

Leader
Leader LV 58SER21 User manual

Leader
Leader LV 5380 User manual

Leader
Leader LV5600 Series Installation guide

Leader
Leader LBO-51MA User manual

Leader
Leader LV5600 Series User manual

Leader
Leader LV 5380 OP72 User manual

Leader
Leader LV 5490 User manual